Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2010 Jan 2.
Published in final edited form as: Curr Opin Obstet Gynecol. 2006 Feb;18(1):64–70. doi: 10.1097/01.gco.0000192965.29449.da

Advances in CAD for diagnosis of breast cancer

Lubomir Hadjiiski 1, Berkman Sahiner 1, Heang-Ping Chan 1
PMCID: PMC2800983  NIHMSID: NIHMS128169  PMID: 16493263

Abstract

Purpose of Review

Computer-Aided Diagnosis (CAD) is a technology used for detection and characterization of cancer. Although CAD is not limited to a single type of cancer, a large number of CAD systems to date have been designed and used for breast cancer. The aim of this review is to discuss the current state of the CAD systems for breast cancer diagnosis, their application as a second reader in clinical practice, and the studies that evaluated the effect of CAD on radiologists’ performance.

Recent Findings

A large number of CAD applications are being developed for different imaging modalities. The main clinical use of CAD to date is for screen-film mammography due to the commercially available FDA approved systems. Many studies showed that CAD improves radiologists’ performance. A large number of academic institutions have devoted substantial research effort to develop CAD methods.

Summary

CAD systems will play an increasingly important role in the clinic as a second reader. Clinical trials have shown that CAD can improve the accuracy of breast cancer detection. Preclinical studies have demonstrated the potential of CAD to improve classification of malignant and benign lesions. An increased number of CAD systems are being developed for different breast imaging modalities.

Keywords: Computer aided diagnosis, breast cancer, mammography, detection, characterization

Introduction

Mammography is currently the only recommended imaging method for breast cancer screening [1, 2]. Mammography is especially valuable as an early detection tool because it can identify breast cancer at an early stage, before physical symptoms develop [3]. Early detection via mammographic screening and physical examination has demonstrated improved survival from breast cancer in randomized controlled trials [1, 2]. However, the high sensitivity of mammography is accomplished at a cost of low specificity. In order to reduce false-negative diagnosis, lesions with greater than 2% chance of being malignant will be recommended for biopsy [4]. As a result, only 15–30% of patients referred to biopsy are found to have malignancy [46]. Unnecessary biopsies not only cause patient anxiety and morbidity but also increase health care costs. It is therefore important to improve the accuracy of interpreting mammographic lesions, thereby increasing the positive predictive value of mammography.

Computer-Aided Diagnosis (CAD) is becoming an increasingly important tool to assist radiologists in the mammographic interpretation process. Current CAD systems in clinical use serve as a second reader for breast cancer detection. CAD systems for classification of malignant and benign lesions are under development by a number of research groups.

For characterization of breast lesions detected either clinically or at screening, ultrasound and MR exams are frequently recommended. Encouraged by the promising results of CAD for mammographic interpretation, many research groups have started to investigate the use of CAD for the interpretation of ultrasound and MR breast images.

This literature review is mainly focused on publications published since 2004 that include CAD observer studies and CAD methods.

CAD for screen film mammography

The largest number of applications of CAD currently is for the lesion detection on screen film mammography.

Helvie et al. [7**] conducted a pilot prospective clinical trial of a noncommercial CAD system developed at the University of Michigan for screening mammography. A total of 2389 screening cases were read by 13 MQSA–qualified radiologists in two academic institutions. Of the 11 cancers detected at screening in this patient cohort, the CAD system detected 10, and the radiologists also detected 10. One of the cancers detected by the CAD system was not seen by the radiologist initially (10% increase in the detection rate). There was 10% increase in the call-back rate (from 14.4% to 15.8%) for the study cases with CAD. In one of the institutions where the call-back rate for non-study cases was available, this rate increased from 10.3% for non-study cases to 13.5% for study cases before the CAD marks were displayed.

In a medical center setting Birdwell et al. [8**] evaluated 8682 cases without and with CAD. They performed 165 interventions and found 29 cancers. Twenty one cancers were detected both by the CAD and the radiologists, six were detected only by radiologists and 2 were detected only by CAD. Radiologist reading with CAD resulted in 7.4% increase in the cancers detected. The increase of the recall rate because of CAD was 7.6%. A modest increase in the recall rate was observed for the study cases without CAD compared with that of the previous similar time period when radiologists interpreted the screening mammograms without having the CAD system present.

One interesting observation in the above two studies was that the radiologists’ call-back rate for the study cases increased even before the CAD marks were displayed, indicating that they might become more vigilant when they were aware that their reading would be compared with a second reading.

Gur et al. [9**] reported the results of a prospective study of CAD in their academic setting. Their study design compared the changes in breast cancer detection rate and recall rates between two periods of time, before and after the CAD system was introduced into their screening mammography practice. They found no statistically significant changes in the average recall and breast cancer detection rates between the 56432 cases read by 24 radiologists before the introduction of the CAD system and the 59139 cases read after its introduction. These findings therefore appeared to differ from those of the previous studies. The outcome of a study for evaluating the effects of CAD is expected to depend on a number of factors, such as the expertise of the radiologists using the system, the radiologists’ vigilance in interpreting the CAD masks, and the study design. The study design used by Gur et al. [9**] was different from the sequential reading design of the above studies in that the statistics without and with CAD were obtained from different periods of time. The change, if any, in a radiologist’s decision caused by CAD was not recorded for individual cases or individual readers. This design thus yields a global average of the effects from many factors that may influence the cancer detection rate. For example, Gur et al. [9**] reported that the proportion of patients who were screened for the first time decreased from about 40% to 30% during the entire study period. Potential differences in the demographics of the two patient cohorts recruited in the two periods might also contribute variances to the breast cancer detection and recall rates. It is not known whether any differences between the two patient cohorts might have masked the potential change in the breast cancer detection rate due to CAD.

Hadjiiski et al. [10**] evaluated the effects of CAD on radiologists’ characterization of masses on serial mammograms using 253 temporal image pairs (138 malignant and 115 benign) obtained from 96 patients who had masses on serial mammograms. Eight radiologists and 2 breast imaging fellows assessed the temporal pairs with and without computer aid. Receiver operating characteristic (ROC) methodology was used for analysis of classification performance. The average area under the ROC curve, Az for radiologists’ estimates of the likelihood of malignancy was 0.79 without CAD and improved to 0.84 with CAD. The improvement was statistically significant (P =0 .005). On the basis of BI-RADS assessments, it was estimated that with CAD, each radiologist reduced 0.7% of unnecessary biopsies on average and correctly recommended 5.7% additional biopsies.

Destounis et al. [11**] retrospectively evaluated the role of CAD in reducing the rate of false negative findings on screening mammograms considered normal at initial double reading. The CAD correctly marked 71% of the 52 actionable findings read as negative in previous screening year. This shows the potential of CAD to reduce FN rate at double reading.

The effect of CAD on the detection of clinically unsuspected breast cancers was studied by Butler et al. [12**]. Breast cancers occurred at location other than the site of the presenting clinical finding in 15 % (30/197) of the patients in whom cancer was detected. CAD identified 87% of these incidentally detected cancers and may be useful as an aid to the radiologists.

In an observer experiment with 5 radiologists and 185 patients, Marx et al. [13**] recorded an increase in short term call back rates ranging between 5% and 7% when the cases were evaluated with CAD. With CAD they also observed a reduced number of recommended unnecessary biopsies (range from 12% to 34%).

Baker et al. [14*] studied the consistency of a commercial CAD system by scanning 10 times the mammograms and evaluating the prompts from the CAD system. He concluded that there was an inconsistency in the CAD analysis for the breast cancers detected at screening; however, the CAD system was reasonably consistent in the overall number of cancers identified at different runs. Greater variability was observed for the false-positive marks compared to true-positive marks.

Two commercial and a in-house CAD schemes were compared by Gur et al. [15*] on 219 patients. The mass detection rate for the systems varied between 67% and 72%, and the differences were not statistically significant. The differences in the FP rate were statistically significant, however, ranging from 1.08 to 1.68 per four view examinations.

CAD for full field digital mammography

Full field digital mammography (FFDM) is becoming important imaging modality. Wei et al. [16**] proposed a multistage CAD for detection of masses on FFDM. They reported a case-based sensitivity of 70%, 80%, and 90% at 0.85, 1.31, and 2.14 FP marks/image. McLoughlin et al. [17**] developed a method for noise equalization in FFDM images and showed that the proposed square root noise model improved the performance of their CAD algorithm for detection of microcalcification clusters on FFDM images.

CAD for breast Ultrasound

A number of CAD schemes were proposed for characterization of malignant and benign breast lesions on ultrasound scans.

Sahiner et al. [18**] designed a CAD system for characterization of breast masses as malignant and benign on three-dimensional (3-D) ultrasound scans using 102 biopsied masses. Four radiologists evaluated the 3-D ultrasound volumes in an observer study. The CAD system based on 3-D segmentation method achieved an area under ROC curve, Az, of 0.92 for testing. The Az values of the four radiologists ranged from 0.84 to 0.92. The accuracy of the classifier designed for differentiating malignant and benign breast masses on 3-D US volumes in this study was similar to that of experienced breast radiologists.

A two stage CAD system which first detected the lesions on US images and then classified the lesion by distinguishing the cancer from all other lesions was developed by Drukker et al. [19**]. Four hundred cases were used for training and 485 cases were used for testing. The Az for the task of distinguishing the true lesions and false-positives was 0.94 and 0.91 for the training and testing data sets, respectively. For the task of classifying cancer from all other detections, the Az was 0.87 and 0.81 for the training and testing data sets, respectively.

Sahiner et al. [20**] designed a combined 3-D US and mammographic CAD system using 67 patients and performed an observer study with 5 radiologists. The Az of the combined CAD system was 0.91. The radiologists improved significantly (p=0.03) their average performance to an Az of 0.95 when using combined CAD.

Drukker et al. [21**] also designed a CAD system combining an ultrasound and mammographic information. The system was evaluated on a data set of 100 lesions from 97 patients having available mammograms and US images. The Az for the combined system was 0.92, which represented a statistically significant improvement compared to the mammography system alone (Az=0.77) and the ultrasound system alone (Az=0.88).

Additional CAD systems for classification of breast masses as malignant or benign on US images were reported by Joo et al. [22*], Chang et al. [23], Moon et al. [24*], and Song et al. [25]. The CAD systems were based on artificial intelligence techniques (support vector machines (SVM) and artificial neural networks (ANN)) and developed with data sets of different sizes (from 54 cases to 584 cases). The classification accuracy Az obtained with these CAD systems ranged from 0.86 to 0.95.

CAD for breast MRI

Recently MRI is used more frequently for diagnosis of breast cancer. A number of studies have been reported which explored the feasibility of CAD use for characterization of breast lesions as malignant and benign using contrast enhanced MRI.

Deurloo et al. [26**] designed a computer system to estimate the likelihood of malignancy of lesions on contrast enhanced MR images. They used 100 lesions (in 78 patients) to train the computer system. In addition they had the descriptive rating (benign, probably benign, indeterminate, suspicious, or highly suggestive of malignancy) of the lesions on MR images by an experienced radiologists. The scores from the computer system and the radiologist were merged by using the logistic regression analysis in a combined model. The model was tested on an independent set of 72 clinically and mammographically occult lesions in 60 patients. The Az values for the radiologist’s readings, the computer system and the combined model were 0.86, 0.85 and 0.91, respectively. The study demonstrated that the complementary information from a radiologist and a computer system has the potential to increase the overall performance for clinically and mammographically occult lesions.

Malignant and benign classification on T1-weighted 3D spoiled gradient echo sequence without and with intravenously injected contrast was performed by Chen et al. [27**]. A data set of 121 cases (77 malignant and 44 benign) was used. Features related to morphology, enhancement kinetics, and time course of enhancement-variation over the lesion were extracted from both the hand outlined and automatic computer segmented lesion contours. The classification accuracy in terms of Az using a linear classifier was 0.80 and 0.86 for hand and computer segmented contours, respectively.

Nattkemper et al. [28*] developed a classifier to distinguish between malignant and benign breast masses using MRI data from 74 cases (49 malignant and 25 benign). They extracted contour and wash-out type features determined by radiologists from dynamic contrast enhanced magnetic resonance imaging time-course data. Using Support Vector Machines (SVM) classifier they obtained an Az of 0.88.

DeMarini et al. [29*] evaluated 15 patients with 16 newly diagnosed locally advanced breast cancers using MRI before and after neoadjuvant chemotherapy. CAD assessments, including presence or absence of significant enhancement, enhancement profiles, and maximum sizes, were recorded. Prior to chemotherapy, all tumors demonstrated CAD-assessed significant enhancement. Following chemotherapy, the results showed a decrease in washout enhancement in patients with minimal residual malignancy at pathology. However, 7 of 16 tumors showed no residual significant enhancement after chemotherapy, but all had residual disease at pathology. In this patient population CAD for breast MRI may complement but should not replace the careful assessment of tumors by the radiologist.

CAD for new modalities - breast tomosynthesis

Digital breast tomosynthesis (DBT) is a promising new modality allowing 3D imaging of the breast.

Chan et al. [30*, 31**] have developed a CAD system for breast masses on DBT mammograms. Twenty-six cases obtained with a prototype DBT system were used. The CAD system first screened the 3D volume for mass candidates by gradient field analysis. Each candidate was segmented from the structured background. Morphological and texture features were extracted. A feature classifier was designed to differentiate true masses from normal tissues. The CAD system was trained and tested using a leave-one-case-out method. The classifier obtained an area under the test ROC curve of 0.91±0.03. The CAD system achieved a sensitivity of 85% at 2.2 false positives/case. The results demonstrate the feasibility of this approach to the development of a CAD system for DBT mammograms.

Reiser et al. [32**] presented initial results for a computerized mass lesion detection scheme for DBT images. The algorithm used a radial gradient index feature for the initial lesion detection and for segmentation of lesion candidates. A set of features was extracted for each segmented partition. Performance of two- and three dimensional features was compared. For gradient features, the additional dimension provided no improvement in classification performance. For shape features, classification using 3D features was improved compared to the 2D equivalent features. The system was designed using a leave-one-case-out method on 21 DBT cases. The preliminary overall performance was 76% sensitivity at 11 false positives per case.

CAD development – new methods and approaches for breast CAD

Automated interval change analysis of breast lesions is an important direction of CAD development. Filev et al. [33*], Marias et al. [34*] and Timp et al. [35*] proposed automatic methods for registration of corresponding masses on temporal pairs of mammograms that will assist a CAD system in interval change analysis.

Breast density is a risk factor for developing breast cancer. Wei et al. [36**] used an automated CAD system for estimation of the breast density on mammograms. The system estimated breast density using histogram analysis [37]. They evaluated the correlation between the percent mammographic dense area and the percent glandular tissue volume as estimated from MR images. The correlation was high, which indicated that changes in mammographic density may be a useful indicator of changes in fibroglandular tissue volume in the breast.

Li et al. [38**] developed computerized texture analysis of mammograms that provided radiographic descriptors of mammographic parenchymal patterns, which may be useful for identifying women at high risk for breast cancer and for monitoring the treatment of breast cancer patients.

Brem et al. [39**] studied the effect of breast density on CAD performance. They found no statistically significant difference in breast cancer detection in dense and nondense breasts. However the false-positive rate was lower in nondense breasts than in dense breasts for mass detection. The authors suggested that CAD may be particularly advantageous in patients with dense breasts, in which mammography is most challenging.

Brem et al. [40*, 41*] also studied the effect of cancer size and tumor histopathology on CAD detection performance. They found that CAD detection performance does not strongly depend on cancer size and tumor histopathology, and that the CAD system correctly marked a large majority of the biopsy-proven breast cancers.

An area of active CAD research is the improvement of breast mass detection algorithms. Catarious et al. [42*] and Timp et al. [43*] developed methods to improve the segmentation of the breast masses using iterative linear segmentation and segmentation based on dynamic programming, respectively. Wei et al. proposed a dual CAD system approach to detection of subtle masses [44*]. An improvement in detection sensitivity by 2 % to 10 % without an increase in FP marks was achieved.

Soo et al. [45*] used a commercial system to detect amorphous calcifications, which usually are difficult to detect for the radiologists. The case sensitivity of CAD was found to be relatively low at 51% with 2 FP per case.

Breast lesion characterization is also an important topic of CAD research. Bilska-Wolak et al. [46*] designed a likelihood ratio classifier for classification of masses as malignant and benign and tested its performance on an independent set. A test Az of 0.90 was obtained. Lim et al. [47] developed a classifier using generalized dynamic fuzzy neural networks and texture features. They obtained an Az of 0.87 for classification of malignant and benign masses.

Paquerault et al. [48*], Kallergi et al. [49*], Leichter et al. [50*], Wei et al. [51*], Papadopoulos et al. [52] presented methods for classification of microcalcifications on malignant and benign. The classification methods used included LDA, ANN, SVM, and Bayesian ANN. The classification accuracy ranged from an Az of 0.80 to 0.98. Nakayama et al. [53*] performed computer classification of microcalcification clusters on five different histological types. The classification accuracy ranged from 75% to 100%.

To facilitate automated analysis of mammograms for CAD system, a number of breast segmentation methods have been developed. Zhou et al. [54*] reported a new automated method for breast nipple detection. Ferrari et al. [55*] and Kwok et al. [56*] reported computerized methods for pectoral muscle segmentation.

Conclusions

CAD is an active area of research and development in medical imaging and diagnostic radiology. Recently, there are an increased number of CAD applications. To date the largest number of CAD systems have been developed for screening mammography. However, an increasing number of applications for US and MRI breast imaging are being evaluated. Although most clinical applications are devoted to breast cancer detection at present, it can be expected that CAD characterization will be an important component of the next-generation CAD systems.

References and recommended reading

  • 1.Feig SA, D'Orsi CJ, Hendrick RE, et al. American College of Radiology guidelines for breast cancer screening. AJR Am J Roentgenol. 1998;171:29–33. doi: 10.2214/ajr.171.1.9648758. [DOI] [PubMed] [Google Scholar]
  • 2.Cady B, Michaelson JS. The life-sparing potential of mammographic screening. Cancer. 2001;91:1699–1703. doi: 10.1002/1097-0142(20010501)91:9<1699::aid-cncr1186>3.0.co;2-w. [DOI] [PubMed] [Google Scholar]
  • 3.American Cancer Society. Cancer Facts & Figures 2005. 2005 www.cancer.org.
  • 4.Sickles EA. Periodic mammographic follow-up of probably benign lesions: results in 3184 consecutive cases. Radiology. 1991;179:463–468. doi: 10.1148/radiology.179.2.2014293. [DOI] [PubMed] [Google Scholar]
  • 5.Kopans DB. The positive predictive value of mammography. AJR Am J Roentgenol. 1992;158:521–526. doi: 10.2214/ajr.158.3.1310825. [DOI] [PubMed] [Google Scholar]
  • 6.Adler DD, Helvie MA. Mammographic biopsy recommendations. Curr Op Radiol. 1992;4:123–129. [PubMed] [Google Scholar]
  • 7. Helvie MA, Hadjiiski LM, Makariou E, et al. Sensitivity of noncommercial computer-aided detection system for mammographic breast cancer detection - A pilot clinical trial. Radiology. 2004;231:208–214. doi: 10.1148/radiol.2311030429. A pilot prospective clinical trial of a CAD system with 2389 screening cases and 13 radiologists showed 10% increase in detection rate and 10% increase in the call-back rate.
  • 8. Birdwell RL, Bandodkar P, Ikeda DM. Computer-aided Detection with Screening Mammography in a University Hospital Setting. Radiology. 2005;236:451–457. doi: 10.1148/radiol.2362040864. A clinical study based on 8682 cases showed 7.4% increase in cancer detection rate and 7.6% increase the recall rate when radiologists were using CAD.
  • 9. Gur D, Sumkin JH, Rockette HE, et al. Changes in breast cancer detection and mammography recall rates after the introduction of a computer-aided detection system. J National Cancer Institute. 2004;96:185–190. doi: 10.1093/jnci/djh067. No statistically significant changes in the recall and breast cancer detection rates between the 56432 cases read by 24 radiologists before the introduction of the CAD system and the 59139 cases read after its introduction were reported in this study.
  • 10. Hadjiiski LM, Chan HP, Sahiner B, et al. Improvement of Radiologists' Characterization of Malignant and Benign Breast Masses in Serial Mammograms by Computer-Aided Diagnosis: An ROC Study. Radiology. 2004;233:255–265. doi: 10.1148/radiol.2331030432. The effects of CAD on radiologists’ characterization of masses on serial mammograms using 253 temporal image pairs were evaluated. With CAD the radiologists improved both their estimates of likelihood of malignancy and biopsy recommendations.
  • 11. Destounis SV, DiNitto P, Logan-Young W, et al. Can Computer-aided Detection with Double Reading of Screening Mammograms Help Decrease the False-Negative Rate? Initial Experience. Radiology. 2004;232:578–584. doi: 10.1148/radiol.2322030034. This study showed the potential of CAD to reduce FN rate of false negative findings on screening mammograms considered normal at initial double reading.
  • 12. Butler SA, Gabbay RJ, Kass DA, et al. Computer-aided detection in diagnostic mammography: Detection of clinically unsuspected cancers. AJR Am J Roentgenol. 2004;183:1511–1515. doi: 10.2214/ajr.183.5.1831511. The effect of CAD on the detection of clinically unsuspected breast cancers was studied. CAD identified 87% of the incidentally detected cancers, other than the site of the presenting clinical finding, and may be useful as an aid to the radiologists.
  • 13. Marx C, Malich A, Facius M, et al. Are unnecessary follow-up procedures induced by computer-aided diagnosis (CAD) in mammography? Comparison of mammographic diagnosis with and without use of CAD. European Journal of Radiology. 2004;51:66–72. doi: 10.1016/S0720-048X(03)00144-X. In an observer experiment with 5 radiologists and 185 patients, it was recorded an increase in short term call back rates and also a reduced number of recommended unnecessary biopsies when the cases were evaluated with CAD.
  • 14. Baker JA, Lo JY, Delong DM, et al. Computer-aided detection in screening mammogrpahy: Variability in cues. Radiology. 2004;233:411–417. doi: 10.1148/radiol.2332031200. The consistency of a commercial CAD system was studied by scanning 10 times the mammograms.
  • 15. Gur D, Stalder JS, Hardesty LA, et al. Computer-aided Detection Performance in Mammographic Examination of Masses: Assessment. Radiology. 2004;233:418–423. doi: 10.1148/radiol.2332040277. Two commercial and a in-house CAD schemes were compared on 219 patients.
  • 16. Wei J, Sahiner B, Hadjiiski LM, et al. Computer aided detection of breast masses on full field digital mammograms. Med Phys. 2005;32:2827–2838. doi: 10.1118/1.1997327. A multistage CAD system for the detection of masses on full field digital mammography was proposed in this study.
  • 17. McLoughlin KJ, Bones PJ, Karssemeijer N. Noise Equalization for Detection of Microcalcification Clusters in Direct Digital Mammogram Images. IEEE Trans Medical Imaging. 2004;23:313–320. doi: 10.1109/TMI.2004.824240. A method for noise equalization in FFDM images was developed and evaluated with CAD.
  • 18. Sahiner B, Chan HP, Roubidoux MA, et al. Computerized characterization of breast masses on 3-D ultrasound volumes. Med Phys. 2004;31:744–754. doi: 10.1118/1.1649531. A CAD system for characterization of breast masses as malignant and benign on 3D ultrasound scans was designed and evaluated using 102 cases. The accuracy of the CAD system was similar to that of 4 experienced breast radiologists.
  • 19. Drukker K, Giger ML, Vyborny CJ, et al. Computerized Detection and Classification of Cancer on Breast Ultrasound. Acad Radiol. 2004;11:526–535. doi: 10.1016/S1076-6332(03)00723-2. A two stage CAD system which first detect the lesions on US images and then classified the lesion by distinguishing the cancer from all other lesions was developed using 400 cases for training and 485 cases for testing. The classification Az was 0.87 and 0.81 for the training and testing data sets, respectively.
  • 20. Sahiner B, Chan HP, Hadjiiski LM, et al. Radiological Society of North America. RSNA Program Book 2004; 2004. The effect of a multi-modality computer classifier on radiologists' accuracy in characterizing breast masses using mammograms and volumetric ultrasound images: An ROC study; p. 447. 447-447. A combined 3-D US and mammographic CAD system was designed by using 67 patients and performed an observer study with 5 radiologists. The radiologists improved significantly their average performance when using combined CAD.
  • 21. Drukker K, Horsch K, Giger ML. Multimodality Computerized Diagnosis of Breast Lesions Using Mammography and Sonography. Acad Radiol. 2005;12:970–979. doi: 10.1016/j.acra.2005.04.014. A CAD system combining an ultrasound and mammographic information was designed and evaluated on 100 lesions. The accuracy of the combined system was statistically significant improved compared to the mammography and US systems alone.
  • 22. Joo S, Yang YS, Moon WK, et al. Computer-Aided Diagnosis of Solid Breast Nodules: Use of an Artificial Neural Network Based on Multiple Sonographic Features. IEEE Trans Medical Imaging. 2004;23:1292–1300. doi: 10.1109/TMI.2004.834617. Design of a CAD system for classification of breast masses as malignant or benign on US images was reported.
  • 23.Chang RF, Wu WJ, Moon WK, et al. Automatic ultrasound segmentation and morphology based diagnosis of solid breast tumors. Breast Cancer Research and Treatment. 2005;89:179–185. doi: 10.1007/s10549-004-2043-z. [DOI] [PubMed] [Google Scholar]
  • 24. Moon WK, Chang RF, Chen CJ, et al. Solid Breast Masses: Classification with Computer-aided Analysis of Continuous US Images Obtained with Probe Compression. Radiology. 2005;236:458–464. doi: 10.1148/radiol.2362041095. Evaluation of the accuracy of continuous US images obtained during probe compression and CAD for classification of biopsy-proved benign and malignant breast tumors were introduced.
  • 25.Song JH, Venkatesh SS, Conant EA, et al. Comparative Analysis of Logistic Regression and Artificial Neural Network for Computer-Aided Diagnosis of Breast Masses. Acad Radiol. 2005;12:487–495. doi: 10.1016/j.acra.2004.12.016. [DOI] [PubMed] [Google Scholar]
  • 26. Deurloo EE, Muller SH, Peterse JL, et al. Clinically and Mammographically Occult Breast Lesions on MR Images: Potential Effect of Computerized Assessment on Clinical Reading. Radiology. 2005;234:693–701. doi: 10.1148/radiol.2343031580. The study demonstrated that the complementary information from a radiologist and a computer system on contrast enhanced MR images has the potential to increase the overall performance for clinically and mammographically occult lesions.
  • 27. Chen W, Giger ML, Lan L, et al. Computerized interpretation of breast MRI: Investigation of enhancement-variance dynamics. Medical Physics. 2004;31:1076–1082. doi: 10.1118/1.1695652. Malignant and benign classification on T1-weighted 3D spoiled gradient echo sequence without and with intravenously injected contrast was performed. The classification accuracy in terms of Az using a linear classifier was 0.80 and 0.86 for hand and computer segmented contours, respectively.
  • 28. Nattkemper TW, Arnrich B, Lichte O, et al. Evaluation of radiological features for breast tumour classification in clinical screening with machine learning methods. Art Intell Med. 2005;34:129–139. doi: 10.1016/j.artmed.2004.09.001. SVM based malignant and benign classification of breast masses using MRI data was reported.
  • 29. DeMartini WB, Lehman CD, Peacock S, et al. Computer-Aided Detection Applied to Breast MRI: Assessment of CAD-Generated Enhancement and Tumor Sizes in Breast Cancers Before and After Neoadjuvant Chemotherapy. Acad Radiol. 2005;12:806–814. doi: 10.1016/j.acra.2005.03.055. CAD for newly diagnosed locally advanced breast cancers with MRI before and after neoadjuvant chemotherapy.
  • 30. Chan HP, Wei J, Sahiner B, et al. 7th International Workshop on Digital Mammography. Durham, NC: 2004. Computerized detection of masses on digital tomosynthesis mammograms - A preliminary study. (in press) A study proposing a CAD system for breast masses on DBT mammograms.
  • 31. Chan HP, Wei J, Sahiner B, et al. Computer-aided detection system for breast masses on digital tomosynthesis mammograms - Preliminary experience. Radiology. 2005 doi: 10.1148/radiol.2373041657. (in press) Development of a CAD system for breast masses detection on DBT mammograms using 26 cases and 3D features. The CAD system achieved a sensitivity of 85% at 2.2 false positives/case. The results demonstrate the feasibility of this approach to the development of a CAD system for DBT mammograms.
  • 32. Reiser I, Nishikawa RM, Giger ML, et al. Computerized detection of mass lesions in digital breast tomosynthesis images using two- and three dimensional radial gradient index segmentation. Technology in Cancer Research and Treatment. 2004;3:437–441. doi: 10.1177/153303460400300504. Presentation of an initial results for a computerized mass lesion detection scheme for DBT images. The algorithm uses a radial gradient index feature for the initial lesion detection and for segmentation of lesion candidates. For shape features, classification using 3D features was improved compared to the 2D equivalent features.
  • 33. Filev P, Hadjiiski LM, Sahiner B, et al. Comparison of similarity measures for the task of template matching of masses on serial mammograms. Med Phys. 2005;32:515–529. doi: 10.1118/1.1851892. Similarity measures applied to automatic registration of corresponding masses on temporal pairs of mammograms were introduced.
  • 34. Marias K, Behrenbruch C, Parbhoo S, et al. A Registration Framework for the Comparison of Mammogram Sequences. IEEE Trans Medical Imaging. 2005;24:782–790. doi: 10.1109/TMI.2005.848374. An automatic methods for registration of corresponding masses on temporal pairs of mammograms.
  • 35. Timp S, Van Engeland S, Karssemeijer N. A regional registration method to find corresponding mass lesions in temporal mammogram pairs. Medical Physics. 2005;32:2629–2638. doi: 10.1118/1.1984323. CAD for automatic registration of corresponding masses on temporal pairs of mammograms based on the combined method.
  • 36. Wei J, Chan HP, Helvie MA, et al. Correlation between Mammographic Density and Volumetric Fibroglandular Tissue Estimated on Breast MR Images. Med Phys. 2004;31:933–942. doi: 10.1118/1.1668512. Evaluation of the correlation between the percent mammographic dense area and the percent glandular tissue volume as estimated from MR images. High correlation was obtained.
  • 37.Zhou C, Chan HP, Petrick N, et al. Computerized image analysis: Estimation of breast density on mammograms. Medical Physics. 2001;28:1056–1069. doi: 10.1118/1.1376640. [DOI] [PubMed] [Google Scholar]
  • 38. Li H, Giger ML, Olopade OI, et al. Computerized Texture Analysis of Mammographic Parenchymal Patterns of Digitized Mammograms. Acad Radiol. 2005;12:863–873. doi: 10.1016/j.acra.2005.03.069. Computerized texture analysis of mammograms that provided radiographic descriptors of mammographic parenchymal patterns, which may be useful for identifying women at high risk for breast cancer and for monitoring the treatment of breast cancer patients.
  • 39. Brem RE, Hoffmeister JW, Rapelyea JA, et al. Impact of breast density on computer-aided detection for breast cancer. AJR Am J Roentgenol. 2005;184:439–444. doi: 10.2214/ajr.184.2.01840439. The article investigates the effect of breast density on CAD performance. The authors suggested that CAD may be particularly advantageous in patients with dense breasts, in which mammography is most challenging.
  • 40. Brem RF, Hoffmeister JW, Zisman G, et al. A computer-aided detection system for the evaluation of breast cancer by mammographic appearance and lesion size. AJR Am J Roentgenol. 2005;184:893–896. doi: 10.2214/ajr.184.3.01840893. The article investigates the effect of lesion appearance and cancer size on CAD detection performance.
  • 41. Brem RF, Rapelyea JA, Zisman G, et al. Evaluation of Breast Cancer with a Computer-Aided Detection System by Mammographic Appearance and Histopathology. Cancer. 2005;104:931–935. doi: 10.1002/cncr.21255. The effect of lesion appearance and tumor histopathology on CAD detection performance was studied in this article.
  • 42. Catarious DM, Baydush AH, Floyd CE. Incorporation of an iterative, linear segmentation routine into a mammographic mass CAD system. Medical Physics. 2004;31:1512–1520. doi: 10.1118/1.1738960. Development of method that improves the segmentation of the breast masses using iterative linear segmentation.
  • 43. Timp S, Karssemeijer N. A new 2D segmentation method based on dynamic programming applied to computer aided detection in mammography. Medical Physics. 2004;31:958–971. doi: 10.1118/1.1688039. Introduction of method that improves and segmentation based on dynamic programming.
  • 44. Wei J, Sahiner B, Hadjiiski LM, et al. A dual computer-aided detection (CAD) system for improvement of mass detection on mammograms. RSNA Program Book. 2004;2004:491. In this study a dual CAD system approach to detection of subtle masses was proposed.
  • 45. Soo MS, Rosen EL, Xia JQ, et al. Computer-Aided Detection of Amorphous Calcifications. AJR Am J Roentgenol. 2005;184:887–892. doi: 10.2214/ajr.184.3.01840887. A commercial system was used to detect amorphous calcifications, which usually are difficult to be detected by the radiologists.
  • 46. Bilska-Wolak AO, Floyd CE, Lo JY, et al. Computer Aid for Decision to Biopsy Breast Masses on Mammography: Validation on New Cases. Acad Radiol. 2005;12:671–680. doi: 10.1016/j.acra.2005.02.011. In this study a likelihood ratio classifier was designed for classification of masses as malignant and benign and tested its performance on an independent set.
  • 47.Lim WK, Er MJ. Classification of mammographic masses using generalized dynamic fuzzy neural networks. Medical Physics. 2004;31:1288–1295. doi: 10.1118/1.1708643. [DOI] [PubMed] [Google Scholar]
  • 48. Paquerault S, Yarusso LM, Papaioannou J, et al. Radial gradient-based segmentation of mammographic microcalcifications: Observer evaluation and effect on CAD performance. Medical Physics. 2004;31:2648–2657. doi: 10.1118/1.1767692. In this study a classification of microcalcifications on malignant and benign based on radial gradient-based segmentation was performed.
  • 49. Kallergi M. Computer-aided diagnosis of mammographic microcalcification clusters. Med Phys. 2004;31:314–326. doi: 10.1118/1.1637972. Classification of microcalcifications on malignant and benign based on artificial neural network was included in this article.
  • 50. Leichter I, Lederman R, Buchbinder SS, et al. Computerized Evaluation of Mammographic Lesions: What Diagnostic Role Does the Shape of the Individual Microcalcifications Play Compared with the Geometry of the Cluster? AJR Am J Roentgenol. 2004;2004:705–712. doi: 10.2214/ajr.182.3.1820705. The shape of individual microcalcification and the geometry of clusters of microcalcifications were used for classification on malignant and benign.
  • 51. Wei L, Yang Y, Nishikawa RM, et al. A Study on Several Machine-Learning Methods for Classification of Malignant and Benign Clustered Microcalcifications. IEEE Trans Medical Imaging. 2005;24:371–380. doi: 10.1109/tmi.2004.842457. Classification of microcalcifications on malignant and benign using support vector machines based methods were covered in this article.
  • 52.Papadopoulos A, Fotiadis DI, Likas A. Characterization of clustered microcalcifications in digitized mammograms using neural networks and support vector machines. Art Intell Med. 2005;34:141–150. doi: 10.1016/j.artmed.2004.10.001. [DOI] [PubMed] [Google Scholar]
  • 53. Nakayama R, Uchiyama Y, Watanabe R, et al. Computer-aided diagnosis scheme for histological classification of clustered microcalcifications on magnification mammograms. Medical Physics. 2004;31:789–799. doi: 10.1118/1.1655711. A computer classification of microcalcification clusters on five different histological types was performed.
  • 54. Zhou C, Chan HP, Paramagul C, et al. Computerized nipple identification for multiple image analysis in computer-aided diagnosis. Medical Physics. 2004;31:2871–2882. doi: 10.1118/1.1800713.A new automated method for breast nipple detection was reported in this study.
  • 55. Ferrari RJ, Rangayyan RM, Desautels JEL, et al. Automatic Identification of the Pectoral Muscle in Mammograms. IEEE Trans Medical Imaging. 2004;23:232–245. doi: 10.1109/tmi.2003.823062. A computerized methods for pectoral muscle segmentation was reported.
  • 56. Kwok SM, Chandrasekhar R, Attikiouzel Y, et al. Automatic Pectoral Muscle Segmentation on Mediolateral Oblique View Mammograms. IEEE Trans Medical Imaging. 2004;23:1129–1140. doi: 10.1109/TMI.2004.830529. An additional study reporting automatic methods for pectoral muscle segmentation.

RESOURCES