Abstract
Purpose: The authors describe a new technique to determine the system presampled modulation transfer function (MTF) in digital radiography using only the detector noise response.
Methods: A cascaded-linear systems analysis was used to develop an exact relationship between the two-dimensional noise power spectrum (NPS) and the presampled MTF for a generalized detector system. This relationship was then utilized to determine the two-dimensional presampled MTF. For simplicity, aliasing of the correlated noise component of the NPS was assumed to be negligible. Accuracy of this method was investigated using simulated images from a simple detector model in which the “true” MTF was known exactly. Measurements were also performed on three detector technologies (an x-ray image intensifier, an indirect flat panel detector, and a solid state x-ray image intensifier), and the results were compared using the standard edge-response method. Flat-field and edge images were acquired and analyzed according to guidelines set forth by the International Electrotechnical Commission, using the RQA 5 spectrum.
Results: The presampled MTF determined using the noise-response method for the simulated detector system was in close agreement with the true MTF with an averaged percent difference of 0.3% and a maximum difference of 1.1% observed at the Nyquist frequency (fN). The edge-response method of the simulated detector system also showed very good agreement at lower spatial frequencies (less than 0.5 fN) with an averaged percent difference of 1.6% but showed significant discrepancies at higher spatial frequencies (greater than 0.5 fN) with an averaged percent difference of 17%. Discrepancies were in part a result of noise in the edge image and phasing errors. For all three detector systems, the MTFs obtained using the two methods were found to be in good agreement at spatial frequencies less than 0.5 fN with an averaged percent difference of 3.4%. Above 0.5 fN, differences increased to an average of 20%. Deviations of the experimental results largely followed the trend seen in the simulation results, suggesting that differences between the two methods could be explained as resulting from the inherent inaccuracies of the edge-response measurement technique used in this study. Aliasing of the correlated noise component was shown to have a minimal effect on the measured MTF for the three detectors studied. Systems with significant aliasing of the correlated noise component (e.g., a-Se based detectors) would likely require a more sophisticated fitting scheme to provide accurate results.
Conclusions: Results indicate that the noise-response method, a simple technique, can be used to accurately measure the MTF of digital x-ray detectors, while alleviating the problems and inaccuracies associated with use of precision test objects, such as a slit or an edge.
Keywords: modulation transfer function (MTF), resolution, digital radiography, linear systems analysis, noise power spectrum (NPS), noise, edge, image quality
INTRODUCTION
The modulation transfer function (MTF) is the most widely accepted measure of the spatial resolution response of medical x-ray imaging detectors and has proved to be a valuable tool for determining and comparing detector performance.1, 2, 3, 4 Although the methods used to measure the MTF have changed slightly over the past several decades,5, 6, 7, 8, 9, 10, 11, 12, 13, 14 the underlying premise has remained a constant: A precisely machined object or test device is imaged and the subsequent response of the detector is used to obtain the MTF using several steps, which generally include determination of the line spread function (LSF), taking the Fourier transform of the LSF, and use of various noise reduction and fitting techniques.6, 11, 12, 15, 16, 17, 18, 19, 20
The slit11 and edge12 response methods are the two most commonly used and accepted techniques for measurement of the MTF, with the edge-response method being preferred (with adoption by the IEC) (Ref. 21) for various reasons, including simpler construction and less sensitivity to misalignment.12 However, it is well documented that the edge-response method is also vulnerable to a host of potential problems that could influence the measurement and result in inaccuracies. Inaccuracies may result from errors in the calculated edge angle,12, 22, 23 noise,12, 23, 24 influences of scattered radiation,24, 25 use of finite-element differentiation,10, 22 profile misregistration and phasing errors,22, 23, 26 truncation of the LSF tails and incorrect normalization,27 and windowing and processing.12, 20, 22 Difficulties also arise when comparing measurements with published results of other investigators, as slight differences in object, acquisition, and∕or processing techniques could easily alter the results.23, 24, 27, 28, 29, 30, 31, 32 Significant variations have even been shown to occur when different individuals measured the MTF of the same data set.32 The aim of this work is to develop a new method for measurement of the MTF of digital radiographic systems that is less susceptible to such influences, by using the intrinsic noise response of the detector.
Cascaded linear systems methods33 have been used to accurately predict the signal and noise transfer properties of a wide variety of digital x-ray imaging technologies including x-ray image intensifiers,34 direct and indirect flat panel detectors (FPD),35, 36 and CCD∕EMCCD-based detectors.37, 38, 39, 40 Careful inspection of the resultant noise response of such systems from a theoretical framework, as described by the two-dimensional noise power spectrum (NPS), indicates that the resolution response (i.e., the MTF) is inherently included in such information.41, 42 In this work, we present an exact relationship using a generalized cascaded linear systems analysis. This relationship was then utilized to measure the presampled detector MTF from the noise response alone, without use of a test object.
The accuracy of the noise-response method was investigated using image simulations of a modeled detector system, in which the “true” MTF was known exactly. Measurements were also performed on several different clinical imaging technologies, including an x-ray image intensifier (XII) and a FPD, as well as a custom-built solid-state x-ray image intensifier (SSXII),40, 43, 44, 45 which is an electron-multiplying CCD (EMCCD)-based detector with very high-resolution capabilities (greater than 10 lp∕mm). Results from the noise-response method were compared with those obtained using the standard edge-response method.
Potential benefits of this method, including elimination of the requirement for a test object (which requires careful construction and precise alignment) and the ability to acquire the MTF in all directions simultaneously, will be discussed. Effects of potential noise sources that have not been considered in the theoretical development of the noise-response method and aliasing will also be discussed.
METHODS
Theoretical development of the noise-response method
Cascaded linear systems theory models the overall response of linear and shift-invariant detector systems as a contributing sum of individual stages, each of which consists of a single interaction process.33 Progression of quanta through the imaging chain is generally serial in nature, in which the output of one stage is the subsequent input of the next. Parallel processes occur when x-ray converting phosphors are implemented, resulting from the emission and reabsorption of secondary x rays.38, 46 Parallel processes also occur in photoconductive (direct) x-ray detectors.47 Cascaded linear systems theory has been an extremely useful tool for characterizing and optimizing detector performance and has been used to accurately predict the signal and noise transfer properties of a wide variety of digital x-ray imaging technologies.33, 34, 35, 36, 37, 38, 39, 40 Excellent agreement has been observed between theoretical calculations and measured results, even with the required estimation of many system performance parameters.
Parallel cascade theory was used to generalize the response from digital x-ray detectors, which were considered to consist of four fundamental stages: X-ray absorption, x-ray-to-light conversion, light-to-electron conversion, and digitization. Each stage consists of several interaction substages, where the quanta undergo one of three processes:48 Amplification, blurring, or addition of noise. A brief summary of the parallel cascade model analysis is presented below. Additional details can be seen in a number of works.33, 34, 35, 36, 37, 38, 39, 40
Amplification processes may involve binary selection or conversion of quanta from one form to another. The signal at the output of stage i of an amplification process is described by
(1a) |
where gi is the average gain of stage i and Φi−1(u,v) is the signal in the spatial frequency coordinates (u,v) at stagei−1 (i.e., at the input of stage i). The NPS at stage i is given by48
(1b) |
where σgi2 is the gain variance of stage i, Φi−1 is the average signal at stage i−1, and NPSi−1(u,v) is the NPS at stagei−1 in the spatial frequency coordinates (u,v). It follows that consecutive binomial processes can be combined into a single binomial process with a probability equal to that of the product of the probabilities of each individual stage.37 Blurring processes may be either stochastic (caused by the spatial spreading or scattering of stochastic quanta) or deterministic (e.g., pixelization resulting from digital sampling). The signal at the output of stage i of a blurring process is given by
(2a) |
where Ti is the MTF at stage i. The NPS after a stochastic blurring stage can be written as48
(2b) |
and for a deterministic blurring stage
(2c) |
An additive noise stage does not change the signal as the images are assumed to have been offset corrected. The NPS after an additive noise stage is given by
(3) |
Using Eqs. 1a, 1b, 2a, 2b, 2c, 3 at each appropriate substage of the imaging chain enables calculation of the frequency-dependent signal and noise at the detector output.
The NPS for a generic digital x-ray detector (the stages of which are consistent with a wide range of detector technologies) can then be written at each stage as described below.
Stage 1. X-ray absorption
The first stage in any x-ray detector is the absorption of incident x rays, which consists of a single binomial-gain substage, in which the x rays either are or are not absorbed. The signal and NPS at the output of stage 1 are
(4a) |
and
(4b) |
where g1 is the quantum detection efficiency of the absorber, and Φ0 is the incident x-ray fluence.
Stage 2. X-ray-to-light conversion
For indirect detectors, there is an additional stage in which the x rays are intermediately converted to light via a phosphorescent material. This process involves amplification from the conversion of x-ray photons to light photons and stochastic blurring processes from the light spreading in the phosphor. Parallel events may occur (depending on the x-ray energy and phosphor) as a result of the production of secondary x rays.46, 47 The signal and NPS at the output of stage 2 can be written as38
(5a) |
(5b) |
where T2(u,v) is the MTF associated with the stochastic spreading of light in the phosphor, W is the average work energy required to liberate one image forming quanta, E is the incident x-ray photon energy, Ek is the energy of the characteristic x rays produced in the phosphor, Pk is the electron-shell participation fraction, ωk is the electron-shell fluorescent yield, fk is the absorption probability from the electron-shell transition, and Tk is the MTF associated with the secondary x-ray reabsorption at a remote location. For clarity, the energy response was averaged both for the incident polychromatic x-ray spectra and for the atomic response of the x-ray converting phosphor. However, this is generally not required, and in practice energy differences are considered.
Using the following substitutions:
(6) |
(7) |
(8) |
where gPE is the mean gain of the photoelectric process, TPE(u,v) is the photoelectric MTF, and AS(u,v) is the frequency-dependent Swank noise, Eqs. 5a, 5b can be simplified to
(9a) |
(9b) |
Eqs. 6, 7, 8 are consistent with those described by Hajdok et al.49
Stage 3. Light-to-electron conversion
Prior to digitization, light photons (or x rays in the case of direct detectors) are converted into electrons. Light photons undergo a binomial selection process, in which they either are or are not converted into an electron. This conversion may also include stochastic blurring processes, for instance, light spreading in fiber optic components. The signal and NPS at the output of stage 3 are described by
(10a) |
(10b) |
where g3 is the light-to-electron conversion factor (with consecutive binomial processes combined into one binomial process) and T3(u,v) is the MTF associated with stochastic spreading of light in the conversion process.
Stage 4. Digitization
The final stage for any digital x-ray detector is digitization. Digitization occurs with the conversion of electrons to digital numbers (DNs), an amplification process. The signal is accumulated in pixel elements, resulting in deterministic blurring. Noise extending beyond the Nyquist frequency is aliased to lower frequencies. Digitization also involves the addition of electronic noise resulting from the digital readout process. The signal and NPS at the output of stage 4 are50
(11a) |
(11b) |
where38
(11c) |
is the presampled NPS, Δx and Δy are the pixel width in the horizontal and vertical directions, n and m are the number of replicates, g4 is the electron-to-digital-number conversion factor which is assumed to have no variation,51T4(u,v) is the aperture MTF associated with integration of the signal over a finite region (e.g., the sinc function of the detector element size), and NPSAdd(u,v) is the additive electronic noise.
The system MTF (TSys) and the effective system gain in units of DN per absorbed x-ray photon can be written as
(12) |
and
(13) |
Combining Eqs. 11a, 11b, 11c, 12, 13 we can then simplify to the following expressions:
(14a) |
(14b) |
where
(15a) |
and
(15b) |
are the correlated and uncorrelated noise components, respectively, and Φ4 is the average digital detector value for an input of Φ0 quanta∕mm2.
Other investigators have shown that sum of uncorrelated noise and its alias produces a white spectrum in the frequency domain.52, 53 Hence, the digital output NPS can be written as52
(16) |
where FP is the pixel fill factor. If the aliasing of the correlated noise is negligible, or, similarly, if
(17) |
which is a reasonable assumption for indirect detectors,54, 55 Eq. 16 simplifies to53
(18) |
Potential errors introduced by making this assumption will be discussed further in the results section.
Equation 18 is the output digital NPS, the form of which is generic for a wide range of detector technologies. The NPS consists of additive contributions from primary quantum noise, Poisson excess noise, secondary quantum noise, and additive electronic noise.53 In general, a structure noise term which scales proportionally to Φ42 should also be included, as a result of variations in detector sensitivity. However, all images used in this analysis were corrected for gain variations using an average of 30 flat-field images at each exposure level, resulting in a negligible structure noise component. Hence, this term was omitted. The slope of a linear fit of the NPS plotted versus mean signal (Φ4) then goes as
(19) |
and can be used to separate the additive instrumentation noise from the quantum noise sources, which scale proportionally with the detector entrance exposure. Equation 19 can then be used to determine the system MTF TSys, as described below in Sec. 2B.
Experimental procedure
The procedure used for determining the presampling MTF from the detector noise response is described below.
Step 1: Preparation of the imaging setup
A standardized x-ray spectrum56 was used to facilitate comparisons as the response of digital x-ray detectors has been shown to have a significant dependence on the energy of the incident x-ray beam.57 The desired spectra were achieved using a specified thickness of added aluminum filtration (Alloy 1100) (Ref. 58) and by adjusting the kVp of the generator to achieve a specific half value layer. A standard measurement geometry was also used, as defined by the IEC (Ref. 21) to minimize the effects of scatter, with as large a source-to-image distance as the imaging system would allow (greater than 100 cm) and with the added filtration placed as close as possible to the source. To simplify the measurement, the x-ray scatter reduction grid was removed (when appropriate). Clinical digital x-ray detectors typically implement highly nonlinear image processing in an attempt to improve perceived image quality. This image processing could potentially affect the measurement and was disabled to avoid any complications.
Step 2: Measurement of the NPS
NPS measurements were also done as prescribed by the IEC.21 Flat-field images were acquired using the standard spectrum and geometry with no test object in the beam. Responses of the detectors used in this study were linear relative to the number of input quanta, so the implementation of a conversion function (for linearization) was unnecessary. A central region of interest (ROI) was selected in the image for analysis and broken into 25 half-overlapping 256×256 pixel regions. To improve statistics of the measurement, 30 flat field images (providing more than 17×106 independent image pixels) were used for analysis. The NPS was then calculated using59
(20) |
where M is the number of ROIs analyzed, I(xi,yj) is the signal at pixel location (xi,yj), is the average signal at pixel location (xi,yj), and un and vk are the spatial frequency values in the horizontal and vertical directions, respectively, which are sampled at an interval of fN∕128.
Step 3: Separation of additive instrumentation noise
Additive instrumentation noise is generally “white” (i.e., a constant as a function of spatial frequency) and does not contain information of the detector MTF. The additive instrumentation noise is constant as a function of exposure, whereas the quantum noise scales proportionally with the detector entrance exposure. A linear regression fitted to a plot of NPS versus signal was used to separate the two noise components [Eq. 18] with the slope representing the quantum noise per unit signal. To provide a good estimate of regression coefficients, the NPS was measured across a range of exposures, which encompassed a majority of the dynamic range of the detector.
Step 4: Determination of the MTF
The system MTF [TSys(u,v)] was then determined from the quantum noise component of the NPS, using Eq. 19. A functional form of TSys(u,v), based on underlying physical principles, was used in the fit of the slope or quantum noise per unit signal as a function of spatial frequency. Fitting functions used for this analysis were based on a Gaussian mixture model60 and the complementary error function (ERFC) (Ref. 61) and are shown below, in one dimension for simplicity
(21) |
and
(22) |
where As(f) is the calculated frequency-dependent Swank factor, hi corresponds to the ith fitting parameter, and N is the number of Gaussian forms in the Gaussian mixture model. Derivation and calculation of AS(u,v) has been described by Hadjok et al.,49, 62, 63 demonstrating good agreement with Monte Carlo and measured results. A total of four different fitting functions [Eq. 21 with N=1, 2, and 3 and Eq. 22] were used and the one providing the best fit was used for subsequent analysis. Different fitting functions were used to ensure very high precision in the fitting technique. Frequencies were sampled at an interval spacing of fN∕128. The MTF was determined directly from the fitting function
() |
but could also be determined indirectly using the slope data and fit parameters h1 and h5
(24) |
which is more robust in allowing the data to deviate somewhat from the fitted result. If results provided by Eqs. 23, 24 were to show a significant discrepancy, it would be indicative that a more appropriate functional form should be sought. The fit parameters h1 and h5 correspond to and ΔxΔyg4∕FP, respectively, and may be used to determine the system gain and the electron-to-digital-number conversion factor (g4) if the pixel size is known. More advanced methods could also be used to extract TSys, including quadratic optimization. However, for this analysis the use of a fitting technique proved sufficient.
Image simulations
To evaluate the accuracy of the noise-response method, images were simulated using MATLAB (version 7.0.1, MathWorks, Natick, MA) for a simplified detector model in which the MTF was known exactly. The detector model was based on a high-resolution, high-sensitivity EMCCD-based SSXII and is described further below.
Thirty flat-field images were simulated at six different exposure levels [17, 27, 43, 53, 67, and 83 μR (4.4, 7.0, 11, 14, 17, and 21 nC∕kg)] according to the following prescription. First, a number of incident x-ray photons, based on Poisson statistics, were randomly generated per pixel (32×32 μm2) for a 1000×1000 image matrix. The number of x-ray photons absorbed in the phosphor was then determined using a binomial selection process with a success rate of 0.77. A conversion gain of 500 (assumed to be Poisson distributed for simplicity) was used to determine the subsequent number of light photons generated per absorbed x-ray photon. The blur associated with the conversion process was taken to be a single Gaussian with a standard deviation of 28 μm (full width half maximum of 66 μm). Light photons were then converted to electrons using binomial selection with a success rate of 0.5. Electrons were in turn “digitized” using an electron-to-digital-number conversion factor of 0.8 e−∕DN and electronic noise was added (Gaussian with zero-mean) with a standard deviation of 10 DN. The true MTF of this simulated detector system is given by
(25) |
where σ is the standard deviation of the Gaussian blurring function.
Edge images were also simulated for the detector model described above. Images of an edge test object were generated in a manner similar to that described by Carton et al.24 The edge was centered in the image, oriented at an angle of 2° relative to the pixel rows, and assumed to absorb 95% of the incident x rays. An average of ten frames was used to reduce the noise content in the images.
Experimental measurements
For further validation of the method, a comparison was made between the MTFs determined for digital x-ray detectors by the noise-response method and an established measurement technique. Images were acquired for three different detectors, covering the gamut of what we believe to be the past, present, and future of medical x-ray imaging technologies: An XII, an indirect FPD, and a SSXII. The XII was a model CAS-8000 V (Toshiba Medical Systems Corp., Tustin, CA) with an estimated CsI:Na phosphor thickness of 500 μm. The XII was operated in 4.5 in. magnification mode with an effective pixel size of 119 μm. A Varian Paxscan 2020+FPD (Palo Alto, CA) was used with a CsI:Tl phosphor thickness of 600 μm and a pixel pitch of 194 μm. The SSXII40, 43, 44, 45 was a prototype developed by our laboratory for high-resolution, high-sensitivity imaging, and had a 375 μm thick CsI:Tl phosphor and an effective pixel size of 32 μm. Each imaging system was prepared as described in Sec. 2B. The RQA5 spectrum was used for all measurements by adding 21 mm of aluminum (alloy 1100) to a 76 kVp x-ray beam, providing an approximate half value layer of 7.1 mm of Al.
For the noise-response analysis, 30 flat-field images were acquired at six different mA s values spanning a majority of the 12 bit dynamic range of the detectors. Images were also acquired of an opaque edge, slightly angulated at 1°–3° relative to both the horizontal and vertical directions of the pixel matrix. All images were gain and offset corrected. MTF measurements for both the noise-response and edge-response methods were done on three separate occasions to gauge the uncertainty of the measurements.
RESULTS AND DISCUSSION
Image simulations
The NPS was determined for the simulated flat-field images in a central 768×768 region, providing 25 overlapping ROIs per image or 750 overlapping ROIs for the entire 30 image sequence. The combined horizontal and vertical response was taken by averaging along the zero frequency axes and the three adjacent rows∕columns. This averaged NPS is shown in Fig. 1 for the six different incident x-ray exposures used.
Figure 1.
Horizontally and vertically averaged NPS determined for the simulated detector system for six different incident x-ray exposures.
The NPS was plotted as a function of signal at each spatial frequency (with a sampling interval of fN∕128 or 0.12 cycles∕mm), and the data were fitted with a linear regression to separate the quantum noise and the additive electronic noise, as shown in Fig. 2 for four representative frequencies. In total, 128 frequencies were fitted (from 0 to 15.6 cycles∕mm). Overall, the fit was very good with an average r2 value of 0.996. The 95% confidence intervals of the regression coefficients, taken as a percentage of the regression coefficients, were also averaged for all spatial frequencies, and the average was 5.6%. This indicates that the use of 30 flat-field images acquired at six different exposures should provide sufficiently high precision for the subsequent analysis.
Figure 2.
The NPS of the simulated flat-field images as a function of digital signal for four representative spatial frequencies (0.0, 4.0, 8.0, and 12.0 cycles∕mm) selected for illustration. The slopes of the linear regressions (the quantum NPS per unit signal) were used to determine the MTF and are given along with their 95% confidence interval.
In this manner, the quantum noise per unit signal was obtained as a function of spatial frequency. Four functions were then fit to the quantum noise distribution, as described by Eq. 21 (with N=1, 2, and 3) and Eq. 22. Figure 3 shows the quantum noise per signal level with the best fit which was obtained using a single Gaussian (N=1) functional form for TSys. Good agreement was observed at all spatial frequencies and the fit was shown to largely overlay the simulation data. Averaged percent differences were determined to be 1.9%, 5.2%, and 4.3% for spatial frequencies ranging from 0 to 5, 5 to 10, and 10 to 15 cycles∕mm, respectively. The system MTF obtained from this fit is shown on a semilogarithmic plot in Fig. 4. Also shown in this plot is the true MTF, which was known exactly [Eq. 25]. The two curves are in very good agreement with an averaged percent deviation of 0.3% (mean difference of 0.001) and a maximum of 1.1% at the Nyquist frequency.
Figure 3.
Slope of the linear regression of the NPS versus signal (the quantum NPS per unit signal) plotted as a function of spatial frequency on a semilogarithmic scale for the simulated detector system. Largely overlaying the slope data is the resulting fit with the functional form described by Eq. 21.
Figure 4.
The presampled MTF for the simulated detector system determined using the noise-response and edge-response methods, shown on a semilogarithmic plot. The true MTF of the detector was known exactly and is also shown in the plot.
For comparison, the MTF determined using the simulated edge images is also shown in Fig. 4. The edge-response MTF was determined using standard methods12 with no image processing [i.e., no smoothing of the edge spread function (ESF) and subsequent LSF and no use of a windowing function] as such processing would inherently affect the results. Good agreement was observed up to 12.5 cycles∕mm with an averaged percent deviation of 2.7% (mean difference of 0.016). The ROI selected around the edge was 3.2 cm wide (i.e., the entire FOV), which resulted in an unavoidable truncation of the LSF tails, which is an inherent problem of the edge-response method, especially for small FOV detectors. However, because no low-frequency effects were modeled, this truncation did not significantly affect the accuracy of the edge-response method. The noise-response method does not rely on the intermediate determination of the LSF and therefore is not susceptible to the use of a finite ROI size. Above 12.5 cycles∕mm, the edge-response MTF was shown to diverge from the true MTF with an averaged percent deviation of 35% and a maximum of over 100%. This divergence was found to result from both noise in the ESF and the LSF and from phase errors23 (which result from the resampling of the ESF at regular intervals, which is a requirement of the fast Fourier transform)20 and is indicative of the inherent error in the edge-response MTF procedure used for this analysis.22
Experimental measurements
A similar analysis was also done for the three actual detector systems as that described in Secs. 2B, 3A. Zero-frequency NPS values were omitted from the fitting procedure due to the potential for a sharp increase at very low spatial frequencies as a result of artifacts.64 The combined horizontal and vertical response was taken by averaging along the zero frequency axes and the three adjacent rows∕columns. For each detector, the quantum NPS per unit signal was fitted with Eqs. 21, 22. The fitting function that provided the best fit was used for further analysis. Careful attention was paid to ensure that a good fit was obtained at all spatial frequency values (which were sampled at an interval of fN∕128) in order to avoid potential systematic errors resulting from poor fitting techniques. Table 1 summarizes the goodness of fits for each of the three detectors. The 95% confidence intervals of the regression coefficients, taken as a percentage of the regression coefficients, were averaged for all spatial frequencies, and the average was less than 4% for all three detectors. The “best fit” fitting function matched the data within 1% across all spatial frequencies for each of the three detectors studied, which was a slight improvement over assuming the same functional form [e.g., using only Eq. 22 as the fitting function] for each detector. As a result, errors due to the uncertainties in the linear regressions and fitting technique were negligible. The phosphor thickness was varied by ±100 μm from the manufacturer specified value in order to assess potential errors resulting from the frequency-dependent Swank factor estimate, which is dependent on the phosphor thickness. Results suggest that the corresponding error has less than a 3% effect on the measured MTF at the Nyquist frequency (even less at lower frequencies), which is in agreement with the assertion that the Swank factor is largely insensitive to converter thickness for quantum efficiency values greater than 0.5.62 Others have shown that different methodologies for measuring the NPS (differing in the number of ROIs, ROI sizes, detrending techniques, and methods used for extraction of 1D data from the 2D NPS) provide consistent results with less than a 2% variability,65 suggesting that the slightly different techniques used for measuring the NPS would not significantly affect the results.
Table 1.
Summary of the goodness of fit of the parameters of the noise-response method for each of the three detector systems. The average 95% confidence intervals for the regression coefficients, obtained by fitting the NPS versus signal to a linear curve at each spatial frequency, are given in the second column. Four different fitting functions were then used to fit the resulting slope data as a function of spatial frequency and the function giving the best fit that was used to provide the presampled MTF is indicated in the third column. The average percent deviation between the slope data and the best fit function is given in the last four columns for four frequency intervals.
Detector | 95% confidence intervals of regression coefficients (percent of coefficient) (%) | Best fit function | Fitting function agreement with slope data (Averaged % deviation) | |||
---|---|---|---|---|---|---|
0–0.25 fN | 0.25–0.5 fN | 0.5–0.75 fN | 0.75–1.0 fN | |||
XII | 3.1 | Gaussian | −0.59 | 0.75 | −0.95 | 0.18 |
FPD | 2.0 | Gaussian | 0.72 | 0.55 | 0.21 | 0.04 |
SSXII | 3.6 | ERFC | −0.08 | −0.2 | 0.01 | 0.39 |
Figures 567 show the measured presampled MTF using the noise-response and edge-response methods for the XII, FPD, and SSXII, respectively, plotted up to the Nyquist frequency, which was different for the three detectors. Error bars are representative of the standard deviation of the three measurements. Uncertainty in the edge-response method was significantly larger than that of the noise-response method (the error bars of which were generally less than the thickness of the curve), perhaps in part due to slight variations in the orientation of the edge test object and differences in the response of the procedure itself (as mentioned in Sec. 1). The two methods were shown to largely agree within experimental uncertainty for all three detectors; however, systematic differences were observed.
Figure 5.
The XII presampled MTF measured using the noise and edge-response methods plotted up to the Nyquist frequency of 4.2 cycles∕mm.
Figure 6.
The FPD presampled MTF measured using the noise and edge-response methods plotted up to the Nyquist frequency of 2.5 cycles∕mm.
Figure 7.
The SSXII presampled MTF measured using the noise and edge-response methods plotted up to the Nyquist frequency of 15.6 cycles∕mm.
Figure 8 shows the percent difference between the measured MTF from the edge-response method and the noise-response method for the simulated detector system and the XII, FPD, and SSXII. Also shown is the percent difference between the noise-response MTF of the simulated detector compared with the true MTF. To facilitate comparisons between detectors, the spatial frequencies were normalized to the appropriate fN. Close agreement with the true MTF was observed when using the noise-response method, with a maximum percent difference of 1.1%, whereas the edge-response MTF was shown to deviate by more than 50%. Measurements on the three detectors show very good agreement between the two methods up to one half of the Nyquist frequency (fN), with an averaged percent difference of 3.4%. Above fN∕2, the edge-response method increasingly diverges from the noise-response method, with an average percent difference of 20%. Deviations of the experimental results largely follow the trend of the simulation results, suggesting that the discrepancies between the two methods are due to the inaccuracies of the edge-response procedure implemented for this study. Difficulty arises when attempting to correct for this error as it is inherently dependent on the particular images used in the analysis. Simulation images, in which the true MTF is known, provide the only scenario where the error in the MTF measurement can be accurately assessed. However, simulation images are significantly simpler than actual images which include scatter and variations in object uniformity in addition to many other physical effects (as described in Sec. 1). Also, the edge-response method for the SSXII was shown to exhibit oscillatory behavior at higher spatial frequencies. Others have shown that when the presampled MTF is negligible for frequencies above 2fN, phasing errors in the MTF may become sinusoidal in nature,23 which could account for these oscillations. Oscillatory behavior also arises from truncation of the LSF tails15 which results from use of a finite ROI.
Figure 8.
Percent difference of the MTF determined using the noise-response method and the edge-response method for the simulated detector system, relative to the true MTF. Also shown is the percent difference of the MTF measured using the edge-response method relative to the noise-response method for the XII, FPD and SSXII.
It should be noted that any aliasing not being properly taken into account (as described in Sec. 2A) or Lubberts effect42 would tend to disproportionally increase the NPS at higher spatial frequencies, resulting in a corresponding inflation in the measured MTF at higher spatial frequencies when using the noise-response method, as demonstrated by Eq. 16. However, for each of the three detector systems studied in this analysis, the edge-response MTF was always larger than the noise-response MTF indicating that these effects were likely minimal. To estimate the error resulting from the assumption that the correlated noise alias was negligible, a fit including these effects was performed, using the following equation [which is a combination of Eqs. 16, 21]:
(26) |
where ±100 frequency replicates were considered. The MTF was then obtained as otherwise described in Sec. 2B. The calculated error in the MTF [taken to be the percent difference between the MTF determined using Eqs. 21, 22, 26] was found to be 3.5% maximally at the Nyquist frequency for the FPD, and less at lower spatial frequencies and for the other two detectors. Improved precision could result from including the effects of aliased correlated noise. Further investigation is required to determine whether or not the noise-response method can provide an accurate MTF for heavily aliased systems, such as direct detectors.
Accurate MTF measurements were obtained using the noise-response method on three very different detector technologies, using the same generalized cascaded-systems model, supporting the notion that the generalized form of the NPS, as given in Eq. 14b, accurately describes a wide range of detectors. Cascaded-systems models by others, for a wide variety of detector technologies (including direct detectors), also indicate general agreement with Eq. 14b. Therefore, verification of Eq. 14b appears to be unnecessary for well-behaved systems. In each instance, the NPS contains a correlated and uncorrelated component which scales proportionally with exposure, with the correlated component being passed by the system MTF. Appropriate isolation of this component enables recovery of the system MTF from the NPS. Admittedly, it is difficult to foresee all physical processes and to include each in the model. Other potential effects that have not been explicitly included, such as lag or detector multiplicative noise, would scale the NPS equally at all spatial frequencies,66 and therefore would not affect the measured MTF, as they would get absorbed in the fitting coefficients h1 and h5 in Eqs. 21, 22. Afterglow should have the same frequency components as the light (or electrons) emitted “immediately” after x-ray absorption and therefore would also not affect the measurement. Only noise sources that were to vary with spatial frequency would affect the measurement, and it is difficult to envision any such processes.
The two-dimensional NPS provides the noise response in all directions. As such, the MTF measured using the noise-response method can be determined not only in the horizontal and vertical directions (an average of which was used in this analysis for comparison with the edge-response method), but at any arbitrary angle relative to the pixel matrix. This is contrary to standard methods, which are inherently one dimensional and rely on slight angulations relative to the pixel rows or columns to provide a “finely sampled” response, hence limiting them to the orthogonal or near-orthogonal directions.11, 21, 23 Accurate characterization of the two-dimensional MTF could prove useful for improved system evaluation and observer performance models.14
CONCLUSIONS
A new method for determination of the presampled MTF, the “noise-response method,” has been described and evaluated. The accuracy of this method was demonstrated using both simulated and experimental data sets. For the simulated image set which used a simple detector model for which the true MTF was known exactly, excellent agreement was obtained with the MTF determined using the noise-response method, with a maximum deviation of 1.1%. Comparison measurements were also made on this simulated data set with the established edge-response method, and these showed deviations greater than 35% from the true MTF. Experimental measurements on a range of detector technologies (including an XII, FPD, and SSXII) demonstrated agreement between the noise-response and edge-response methods within experimental uncertainty, with discrepancies likely resulting from errors inherent in the edge-response MTF procedure.
Compared to current measurement methods, the noise-response method simplifies the MTF determination by eliminating the need for manufacture and alignment of precisely machined test objects, thereby eliminating inaccuracies that result from the use of such objects and subsequent analysis of the resulting images. Further, the two-dimensional MTF is readily obtained with the noise-response method, whereas traditional edge and slit methods are inherently one-dimensional.
ACKNOWLEDGMENTS
This work was supported in part by the National Institutes of Health under Grant Nos. RO1-EB002873 and RO1-EB008425 and an equipment grant from Toshiba Medical Systems Corp.
References
- Rossmann K., “Point spread-function, line spread-function, and modulation transfer function, tools for the study of imaging systems,” Radiology 93, 257–272 (1969). [DOI] [PubMed] [Google Scholar]
- Cunningham I. A., “Applied linear-systems theory,” in Handbook of Medical Imaging: Physics and Psychophysics, edited by Beutel J., Kundel H. L., and Metter R. L. V. (SPIE, Bellingham, 2000), Vol. 1, pp. 82–156. [Google Scholar]
- Samei E. and Flynn M. J., “An experimental comparison of detector performance for computer radiography systems,” Med. Phys. 29, 447–459 (2002). 10.1118/1.1449873 [DOI] [PubMed] [Google Scholar]
- Lazzari B., Belli G., Gori C., and Turco M. R. D., “Physical characteristics of five clinical systems for digital mammography,” Med. Phys. 34, 2730–2743 (2007). 10.1118/1.2742498 [DOI] [PubMed] [Google Scholar]
- Coltman J. W., “The specification of imaging properties by response to a sine wave input,” J. Opt. Soc. Am. 44, 468–469 (1954). 10.1364/JOSA.44.000468 [DOI] [Google Scholar]
- Judy P. F., “The line spread function and modulation transfer function of a computer tomographic scanner,” Med. Phys. 3, 233–236 (1976). 10.1118/1.594283 [DOI] [PubMed] [Google Scholar]
- Schneiders N. J. and Bushong S. C., “Single-step calculation of the MTF from the ERF,” Med. Phys. 5, 31–33 (1978). 10.1118/1.594401 [DOI] [PubMed] [Google Scholar]
- Sones R. A. and Barnes G. T., “A method to measure the MTF of digital x-ray systems,” Med. Phys. 11, 166–171 (1984). 10.1118/1.595493 [DOI] [PubMed] [Google Scholar]
- Droege R. T. and Rzeszotarski M. S., “An MTF method immune to aliasing,” Med. Phys. 12, 721–725 (1985). 10.1118/1.595654 [DOI] [PubMed] [Google Scholar]
- Cunningham I. A. and Fenster A., “A method for modulation transfer function determination from edge profiles with correction for finite-element differentiation,” Med. Phys. 14, 533–537 (1987). 10.1118/1.596064 [DOI] [PubMed] [Google Scholar]
- Fujita H., Tsai D. Y., Itoh T., Doi K., Morishita J., Ueda K., and Ohtsuka A., “A simple method for determining the modulation transfer function in digital radiography,” IEEE Trans. Med. Imaging 11, 34–39 (1992). 10.1109/42.126908 [DOI] [PubMed] [Google Scholar]
- Samei E., Flynn M. J., and Reimann D. A., “A method for measuring the presampled MTF of digital radiographic systems using an edge test device,” Med. Phys. 25, 102–113 (1998). 10.1118/1.598165 [DOI] [PubMed] [Google Scholar]
- Bradford C. D., Peppler W. W., and Waidelich J. M., “Use of a slit camera for MTF measurements,” Med. Phys. 26, 2286–2294 (1999). 10.1118/1.598742 [DOI] [PubMed] [Google Scholar]
- Fetterly K. A., Hangiandreou N. J., Schueler B. A., and Ritenour E. R., “Measurement of the presampled two-dimensional modulation transfer function of digital imaging systems,” Med. Phys. 29, 913–921 (2002). 10.1118/1.1472498 [DOI] [PubMed] [Google Scholar]
- Doi K., Strubler K., and Rossmann K., “Truncation errors in calculating the MTF of radiographic screen-film systems from the line spread function,” Phys. Med. Biol. 17, 241–250 (1972). 10.1088/0031-9155/17/2/010 [DOI] [PubMed] [Google Scholar]
- Yin F. F., Giger M. L., and Doi K., “Measurement of the presampling modulation transfer function of film digitizers using a curve fitting technique,” Med. Phys. 17, 962–966 (1990). 10.1118/1.596463 [DOI] [PubMed] [Google Scholar]
- Reichenbach S. E., Park S. K., and Narayanswamy R., “Characterizing digital image acquisition devices,” Opt. Eng. 30, 170–177 (1991). 10.1117/12.55783 [DOI] [Google Scholar]
- Boone J. M. and Seibert J. A., “An analytical edge spread function model for computer fitting and subsequent calculation of the LSF and MTF,” Med. Phys. 21, 1541–1545 (1994). 10.1118/1.597264 [DOI] [PubMed] [Google Scholar]
- Stenman A. and Gustafsson F., “Adaptive smoothing methods for frequency-function estimation,” Automatica 37, 675–685 (2001). [Google Scholar]
- Maidment A. D. and Albert M., “Conditioning data for calculation of the modulation transfer function,” Med. Phys. 30, 248–253 (2003). 10.1118/1.1534111 [DOI] [PubMed] [Google Scholar]
- International Electrotechnical Commission, “Medical electrical equipment: Characteristics of digital x-ray imaging devices–Part 1: Determination of the detective quantum efficiency,” IEC Report No. 62220-1 (International Electrotechnical Commission, Geneva, Switzerland, 2003).
- Greer P. B. and van Doorn T., “Evaluation of an algorithm for the assessment of the MTF using an edge method,” Med. Phys. 27, 2048–2059 (2000). 10.1118/1.1288682 [DOI] [PubMed] [Google Scholar]
- Buhr E., Gunther-Kohfahl S., and Neitzel U., “Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image,” Med. Phys. 30, 2323–2331 (2003). 10.1118/1.1598673 [DOI] [PubMed] [Google Scholar]
- Carton A. K., Vandenbroucke D., Struye L., Maidment A. D., Kao Y. H., Albert M., Bosmans H., and Marchal G., “Validation of MTF measurement for digital mammography quality control,” Med. Phys. 32, 1684–1695 (2005). 10.1118/1.1921667 [DOI] [PubMed] [Google Scholar]
- Neitzel U., Buhr E., Hilgers G., and Granfors P. R., “Determination of the modulation transfer function using the edge method: Influence of scattered radiation,” Med. Phys. 31, 3485–3491 (2004). 10.1118/1.1813872 [DOI] [PubMed] [Google Scholar]
- Steckner M. C., Drost D. J., and Prato F. S., “A cosine modulation artifact in modulation transfer function computations caused by the misregistration of line spread profiles,” Med. Phys. 20, 469–473 (1993). 10.1118/1.597040 [DOI] [PubMed] [Google Scholar]
- Friedman S. N. and Cunningham I. A., “Normalization of the modulation transfer function: The open-field approach,” Med. Phys. 35, 4443–4449 (2008). 10.1118/1.2977536 [DOI] [PubMed] [Google Scholar]
- Cunningham I. A. and Reid B. K., “Signal and noise in modulation transfer function determinations using the slit, wire, and edge techniques,” Med. Phys. 19, 1037–1044 (1992). 10.1118/1.596821 [DOI] [PubMed] [Google Scholar]
- Morishita J., Doi K., Bollen R., Bunch P. C., Hoeschen D., Sirand-rey G., and Sukenobu Y., “Comparison of two methods for accurate measurement of modulation transfer functions of screen-film systems,” Med. Phys. 22, 193–200 (1995). 10.1118/1.597456 [DOI] [PubMed] [Google Scholar]
- Rogge F., Bosmans H., and Marchal G., “Practical MTF calculation in digital mammography: A multicenter study,” Proc. SPIE 5368, 761–769 (2004). 10.1117/12.535823 [DOI] [Google Scholar]
- Samei E., Ranger N. T., J. T.DobbinsIII, and Chen Y., “Intercomparison of methods for image quality characterization. I. Modulation transfer function,” Med. Phys. 33, 1454–1465 (2006). 10.1118/1.2188816 [DOI] [PubMed] [Google Scholar]
- Neitzel U., Gunther-Kohfahl S., Borasi G., and Samei E., “Determination of the detective quantum efficiency of a digital x-ray detector: Comparison of three evaluations using a common image data set,” Med. Phys. 31, 2205–2211 (2004). 10.1118/1.1766421 [DOI] [PubMed] [Google Scholar]
- Cunningham I. A., Westmore M. S., and Fenster A., “A spatial-frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systems,” Med. Phys. 21, 417–427 (1994). 10.1118/1.597401 [DOI] [PubMed] [Google Scholar]
- Rowlands J. A. and Taylor K. W., “Absorption and noise in cesium iodide x-ray image intensifiers,” Med. Phys. 10, 786–795 (1983). 10.1118/1.595357 [DOI] [PubMed] [Google Scholar]
- Zhao W., Ji W. G., Debrie A., and Rowlands J. A., “Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: Characterization of a small area prototype detector,” Med. Phys. 30, 254–263 (2003). 10.1118/1.1538233 [DOI] [PubMed] [Google Scholar]
- Siewerdsen J. H., Antonuk L. E., el-Mohri Y., Yorkston J., Huang W., Boudry J. M., and Cunningham I. A., “Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology,” Med. Phys. 24, 71–89 (1997). 10.1118/1.597919 [DOI] [PubMed] [Google Scholar]
- Williams M. B., Simoni P. U., Smilowitz L., Stanton M., Phillips W., and Stewart A., “Analysis of the detective quantum efficiency of a developmental detector for digital mammography,” Med. Phys. 26, 2273–2285 (1999). 10.1118/1.598741 [DOI] [PubMed] [Google Scholar]
- Ganguly A., Rudin S., Bednarek D. R., and Hoffmann K. R., “Micro-angiography for neuro-vascular imaging. II. Cascade model analysis,” Med. Phys. 30, 3029–3039 (2003). 10.1118/1.1617550 [DOI] [PubMed] [Google Scholar]
- Vedantham S., Karellas A., and Suryanarayanan S., “Solid-state fluoroscopic imager for high-resolution angiography: Parallel-cascaded linear systems analysis,” Med. Phys. 31, 1258–1268 (2004). 10.1118/1.1689014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kuhls-Gilcrist A., Yadava G., Patel V., Jain A., Bednarek D. R., and Rudin S., “The solid state x-ray image intensifier (SSXII): An EMCCD-based x-ray detector,” Proc. SPIE 6913, 69130K-1–69130K-10 (2008). 10.1117/12.772724 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rossmann K., “Modulation transfer function of radiographic systems using fluorescent screens,” J. Opt. Soc. Am. 52, 774–777 (1962). 10.1364/JOSA.52.000774 [DOI] [Google Scholar]
- Lubberts G., “Random noise produced by x-ray fluorescent screens,” J. Opt. Soc. Am. 58, 1475–1482 (1968). 10.1364/JOSA.58.001475 [DOI] [Google Scholar]
- Rudin S., Kuhls A. T., Yadava G. K., Josan G. C., Wu Y., Chityala R. N., Rangwala H. S., Ionita C. N., Hoffmann K. R., and Bednarek D. R., “New light-amplifier-based detector designs for high spatial resolution and high sensitivity CBCT mammography,” Proc. SPIE 6142, 614263 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kuhls-Gilcrist A., Yadava G., Patel V., Bednarek D. R., and Rudin S., “Progress in electron-multiplying CCD (EMCCD) based, high-resolution, high-sensitivity x-ray detector for fluoroscopy and radiography,” Proc. SPIE 6510, 651047 (2007). 10.1117/12.709079 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kuhls-Gilcrist A., Bednarek D. R., and Rudin S., “Component analysis of a new solid state x-ray image intensifier (SSXII) using photon transfer and instrumentation noise equivalent exposure (INEE) measurements,” Proc. SPIE 7258, 725842 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yao J. and Cunningham I. A., “Parallel cascades: New ways to describe noise transfer in medical imaging systems,” Med. Phys. 28, 2020–2038 (2001). 10.1118/1.1405842 [DOI] [PubMed] [Google Scholar]
- Zhao W., Ji W. G., and Rowlands J. A., “Effects of characteristic x rays on the noise power spectra and detective quantum efficiency of photoconductive x-ray detectors,” Med. Phys. 28, 2039–2049 (2001). 10.1118/1.1405845 [DOI] [PubMed] [Google Scholar]
- Rabbani M., Shaw R., and Metter R. V., “Detective quantum efficiency of imaging systems with amplifying and scattering mechanisms,” J. Opt. Soc. Am. 4, 895–901 (1987). 10.1364/JOSAA.4.000895 [DOI] [PubMed] [Google Scholar]
- Hajdok G., Yao J., Battista J. J., and Cunningham I. A., “Signal and noise transfer properties of photoelectric interactions in diagnostic x-ray imaging detectors,” Med. Phys. 33, 3601–3620 (2006). 10.1118/1.2336507 [DOI] [PubMed] [Google Scholar]
- Giger M. L., Doi K., and Metz C. E., “Investigation of basic imaging properties in digital radiography. 2. Noise Wiener spectrum,” Med. Phys. 11, 797–805 (1984). 10.1118/1.595583 [DOI] [PubMed] [Google Scholar]
- Janesick J. R., Scientific Charge-Coupled Devices (SPIE, Bellingham, 2001). 10.1117/3.374903 [DOI] [Google Scholar]
- Zhao W. and Rowlands J. A., “Digital radiology using active matrix readout of amorphous selenium: Theoretical analysis of detective quantum efficiency,” Med. Phys. 24, 1819–1833 (1997). 10.1118/1.598097 [DOI] [PubMed] [Google Scholar]
- Mackenzie A. and Honey I. D., “Characterization of noise sources for two generations of computed radiography systems using powder and crystalline photostimulable phosphors,” Med. Phys. 34, 3345–3357 (2007). 10.1118/1.2750973 [DOI] [PubMed] [Google Scholar]
- Zhao W., Ristic G., and Rowlands J. A., “X-ray imaging performance of structured cesium iodide scintillators,” Med. Phys. 31, 2594–2605 (2004). 10.1118/1.1782676 [DOI] [PubMed] [Google Scholar]
- Siewerdsen J. H., Antonuk L. E., el-Mohri Y., Yorkston J., Huang W., and Cunningham I. A., “Signal, noise power spectrum, and detective quantum efficiency of indirect-detection flat-panel imagers for diagnostic radiology,” Med. Phys. 25, 614–628 (1998). 10.1118/1.598243 [DOI] [PubMed] [Google Scholar]
- International Electrotechnical Commission, “Medical diagnostic x-ray equipment: Radiation conditions for use in the determination of characteristics,” IEC Report No. 61267 (International Electrotechnical Commission, Geneva, Switzerland, 1997).
- Fetterly K. A. and Hangiandreou N. J., “Effects of x-ray spectra on the DQE of a computed radiography system,” Med. Phys. 28, 241–249 (2001). 10.1118/1.1339883 [DOI] [PubMed] [Google Scholar]
- Ranger N. T., Samei E., J. T.DobbinsIII, and Ravin C. E., “Measurement of the detective quantum efficiency in digital detectors consistent with the IEC 62220-1 standard: Practical considerations regarding the choice of filter material,” Med. Phys. 32, 2305–2311 (2005). 10.1118/1.1929187 [DOI] [PubMed] [Google Scholar]
- Dobbins J. T., “Image quality metrics for digital systems,” in Handbook of Medical Imaging: Physics and Psychophysics, edited by Beutel J., Kundel H. L., and Metter R. L. V. (SPIE, Bellingham, 2000), Vol. 1, pp. 161–222. [Google Scholar]
- Kyprianou I. S., Badano A., Gallas B. D., and Myers K. J., “Singular value description of a digital radiographic detector: Theory and measurements,” Med. Phys. 35, 4744–4756 (2008). 10.1118/1.2975222 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Burgess A. E., “An empirical equation for screen MTFs,” Med. Phys. 5, 199–204 (1978). 10.1118/1.594473 [DOI] [PubMed] [Google Scholar]
- Hajdok G., Battista J. J., and Cunningham I. A., “Fundamental x-ray interaction limits in diagnostic imaging detectors: Spatial resolution,” Med. Phys. 35, 3180–3193 (2008). 10.1118/1.2924219 [DOI] [PubMed] [Google Scholar]
- Hajdok G., Battista J. J., and Cunningham I. A., “Fundamental x-ray interaction limits in diagnostic imaging detectors: Frequency-dependent Swank noise,” Med. Phys. 35, 3194–3204 (2008). 10.1118/1.2936412 [DOI] [PubMed] [Google Scholar]
- Williams M. B., Mangiafico P. A., and Simoni P. U., “Noise power spectra of images from digital mammography detectors,” Med. Phys. 26, 1279–1293 (1999). 10.1118/1.598623 [DOI] [PubMed] [Google Scholar]
- J. T.DobbinsIII, Samei E., Ranger N. T., and Chen Y., “Intercomparison of methods for image quality characterization. II. Noise power spectrum,” Med. Phys. 33, 1466–1475 (2006). 10.1118/1.2188819 [DOI] [PubMed] [Google Scholar]
- Friedman S. N. and Cunningham I. A., “A moving slanted-edge method to measure the temporal modulation transfer function of fluoroscopic systems,” Med. Phys. 35, 2473–2484 (2008). 10.1118/1.2919724 [DOI] [PubMed] [Google Scholar]