NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2010 Feb 22.
Published in final edited form as: J Neurosci Methods. 2008 Jul 1;174(1):103–105. doi: 10.1016/j.jneumeth.2008.06.020

A method for achieving an order-of-magnitude increase in the temporal resolution of a standard computer monitor

Matthew P Fiesta 1, David M Eagleman 2,3
PMCID: PMC2826426  NIHMSID: NIHMS69225  PMID: 18640156

Abstract

As the frequency of a flickering light is increased, the perception of flicker is replaced by the perception of steady light at what is known as the critical flicker fusion threshold (CFFT). This threshold provides a useful measure of the brain’s information processing speed, and has been used in medicine for over a century both for diagnostic and drug efficacy studies. However, the hardware for presenting the stimulus has not advanced to take advantage of computers, largely because the refresh rates of typical monitors are too slow to provide fine-grained changes in the alternation rate of a visual stimulus. For example, a cathode ray tube (CRT) computer monitor running at 100 Hz will render a new frame every 10 msec, thus restricting the period of a flickering stimulus to multiples of 20 msec. These multiples provide a temporal resolution far too low to make precise threshold measurements, since typical CFFT values are in the neighborhood of 35 msec. We describe here a simple and novel technique to enable alternating images at several closely-spaced periods on a standard monitor. The key to our technique is to programatically control the video card to dynamically reset the refresh rate of the monitor. Different refresh rates allow slightly different frame durations; this can be leveraged to vastly increase the resolution of stimulus presentation times. This simple technique opens new inroads for experiments on computers that require more finely-spaced temporal resolution than a monitor at a single, fixed refresh rate can allow.

Keywords: Temporal resolution, Computer monitor, Frame rate, Flicker fusion

1. Introduction

Although medicine utilizes many measures of spatial vision (for example, eye charts, perimetry, etc), there are fewer ways to measure temporal aspects of vision (Eagleman, 2001, 2008; Eagleman et al., 2004; Eagleman et al., 2005) – that is, the brain’s speed of information processing. One of the most useful and popular is the measure of the flicker fusion threshold: the frequency at which a flickering light is first perceived to be steady (Seitz et al., 2004; Simonson and Brozek, 1952; Tyler, 1975, 1985, 1989, 1991; Tyler and Hamer, 1990, 1993). Flicker fusion measurements in medicine have been used for over a century, both in early diagnosis of diseases such as schizophrenia (Black et al., 1975; Saucer and Sweetbaum, 1958), Alzheimer’s (Curran et al., 2004) and glaucoma (Curran and Wattis, 2000; Tyler, 1981), as well as in the measurement of drug efficacy in the pharmaceutical industry (Curran and Wattis, 2000; Danjou et al., 1992). However, even while knowledge about the factors influencing flicker fusion has increased, the basic method for measuring CFFT still requires a custom-built hardware device – such as a light (e.g. LED, lightbulb) attached to an alternating power source. Rapid visual measures have not taken advantage of the advancement of desktop computers, primarily because commercially available monitors are rarely capable of refresh rates faster than 100 Hz. The resulting problem is that one is restricted to a large step size when changing the period of an alternating visual stimulus: at 100 Hz, a new frame is rendered every 10 msec, thus restricting the period of a simulated flickering stimulus (e.g. alternating between displayed and not displayed) to multiples of 20 msec. For example, one can present a flickering stimulus at 50 Hz (1 frame on, 1 frame off), at 25 Hz (2 frames on, 2 frames off), or at 16.7 Hz (3 frames on, 3 frames off). Since typical CFFT values range in the neighborhood of 35 – 50 Hz (Herrick, 1974; Simonson and Brozek, 1952; Tyler, 1989, 1991), the large step size in temporal resolution makes a detailed CFFT measurement impracticable using a computer and standard monitor.

2. A novel method for achieving high resolution alternation rates on a computer display

We describe here a simple and novel technique to enable alternating images with many possible periods on a standard CRT monitor – e.g. 20, 21.1, 22.2, 23.5, 25, 26.7, 27.8, 28.6, 30.8, 33.3, 35.7, 36.4, 40, 42.1 msec. The key element to our technique is to control the video card from within a stimulus presentation program to dynamically reset the refresh rate of the monitor. For example, a refresh rate of 100 Hz allows periods of 20 ms (50 Hz). By dynamically resetting this refresh rate from 100 Hz to 95 Hz, we achieve periods of 21 ms (47.5 Hz). Switching the monitor refresh rate to 90 Hz allows a period of 22.2 ms (45 Hz), and so on. Figure 1 shows the proliferation of alternation periods available on a standard monitor by employing this technique.

Figure 1.

Figure 1

Achieving an 8-fold increase in possible alternation frequencies available on a standard computer monitor. (A) Using a standard 100 Hz refresh rate, only a handful of alternation rates are available. (B) By dynamically switching between refresh rates available with standard video cards, many intermediate alternation rates of a visual stimulus can be achieved. The rates displayed to the right of each bar (and coded by the different colors) represent the spectrum of refresh rates that are commonly available. When utilizing this technique, it is important to ensure the monitor is capable of displaying at the refresh rates generated by the video card. By using the most modern video cards and lowering the screen resolution, one can achieve even finer temporal resolution than indicated in panel B (see text).

Depending on the demands of the experiment, the screen video resolution (width χ height) can be leveraged to achieve even higher refresh rates than the ones used in Fig 1 – for example, one can often include refresh rates of 140, 160, and/or 200 Hz by lowering the screen resolution to 800×600 pixels. This technique works with all video cards: even while different video card brands sometimes use different rates, all use the same standard protocol for communicating between motherboards and monitors.

3. An example application

We wanted to construct a computerized test to determine a subject’s baseline CFFT with high resolution, based on a portable physical device we had previously built using a computer chip and LED arrays (Stetson et al., 2007). Specifically, we had found that when a character and its negative are presented at a slow rate of alternation (as in Figure 2A), subjects will have no trouble identifying the character. But as the rate of alternation increases, a threshold is reached at which the information is presented too rapidly, and the letter can no longer be discriminated from a uniform display (Figure 2B)(Stetson et al., 2007). In this latter case, the character and its negative will perceptually overlap as though they were presented simultaneously. The sum of a character and its negative is a blank field, indistinguishable from any other character and its negative. In one version of our experiment, we presented 4-digit random numbers at different alternation rates on different trials, and subjects reported the numbers they were able to detect. In this way, we could quickly extract a measure of CFFT that was robust against malingering (i.e., if the alternation rate is above a subject’s perceptual fusion threshold, the subject had only a 1 in 10,000 chance of guessing all four numbers correctly).

Figure 2.

Figure 2

Using alternating images to measure the rate at which observers can report numbers and how that compares over different brain regions. (A) When a character is alternated slowly with its negative image, it is easy to identify. (B) As the rate of alternation speeds up, the patterns fuse into a uniform field, indistinguishable from any other character and its negative.

To translate this measurement technique from physical LED arrays onto the computer monitor, we faced a challenge. As described above, if a monitor is refreshed at 100 Hz, the minimum duration for displaying an image is 10 ms, translating to a minimum alternation period (i.e., the image and then its negative, Figure 2) of 20 ms. The next closest alternation rate could be achieved by extending the display duration to 20 ms, yielding a period of 40 ms. Since a threshold period is typically ~35 ms, the temporal resolution is too poor to be useful.

To test subjects at alternation periods that were not multiples of 20 ms, we needed a way to dynamically change the monitor refresh rate while the program was running. Using Visual C++, we programmed an executable called refresh which communicates with the video card to change the monitor refresh rate on the fly (code available here: www.eaglemanlab.net/flickercode; note that after downloading, the location of the executable must be placed in the Matlab path prior to use).

For illustration, we now sketch an example using PsychToolbox for Matlab (Brainard, 1997). If the experiment being performed required a period of 22.2 msec (obtained by setting a 90 Hz refresh rate and presenting each frame twice), the refresh call would be used as follows:

  • eval(sprintf(‘!refresh 90’))

    while(condition)

    • Screen(mainWindow, ‘WaitBlanking’)

      [stimulus code here…]

    end %while loop

In the first line, the eval Matlab command runs the executable refresh with the argument 90 (for 90 Hz). This sets the monitor to the desired refresh rate. One then ensures that the drawing of the graphics is synchronized with the scanning of the phosphor gun by calling the command Screen(mainWindow, ‘WaitBlanking’) (see Brainard, 1997 and psychtoolbox.org). One can then define the stimulus generation code that will presented at the desired rate.

Note that some tasks call for dynamically choosing the refresh rate within the course of the program. In this case, one can write the string to be run by the eval command in the following manner:

  • evalstr = sprintf('!refresh %d', refreshRateVariable);

    eval(evalstr);

This allows the variable refreshRateVariable (which passes the desired rate as an argument to refresh) to be dynamically specified elsewhere in the code.

Discussion

The past few decades have witnessed remarkable progress in new hardware techniques for accurately measuring flicker fusion, and a host of new discoveries (Tyler, 1985, 1989, 1991; Tyler and Hamer, 1990, 1993). However, as a result of the relatively slow refresh rates of computer monitors (as compared to stand-alone, custom-built electronic hardware devices) the ability to easily measure flicker fusion with a standard computer monitor has been elusive. The technique described here exploits the different refresh rates available on monitors to surmount this limitation and widen the possibilities for temporal visual processing experiments on computers.

We have demonstrated a new method for obtaining a dramatic increase in temporal resolution using a standard monitor. We hope this technique will be useful for a wide range of applications, both within and beyond psychophysics. As one example, this technique can be used to pilot hardware devices (e.g. ones that use LED or LCD displays), whereas in the past investigators have had no choice but to build hardware devices with very rapid base clock speeds. Our method enables investigators to pilot hypotheses in software prior to the possibly unnecessary development of a hardware prototype. It is hoped that this simple software method will be of benefit in speeding research in temporal visual processing using standard computer equipment.

Acknowledgments

We thank Chess Stetson for help on this project throughout, and acknowledge funding from RO1 NS053960 (DME) and contract W911NF-07-2-0023 from the Army Research Laboratory awarded to The Center for Strategic and Innovative Technology of The University of Texas at Austin (DME).

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  1. Black S, Franklin LM, de Silva FP, Wijewickrama HS. The flicker-fusion threshold in schizophrenia and depression. The New Zealand Medical Journal. 1975;81:244–246. [PubMed] [Google Scholar]
  2. Brainard DH. The Psychophysics Toolbox. Spatial vision. 1997;10:433–436. [PubMed] [Google Scholar]
  3. Curran S, Wattis J. Critical flicker fusion threshold: a potentially useful measure for the early detection of Alzheimer's disease. Hum Psychopharmacol. 2000;15:103–112. doi: 10.1002/(SICI)1099-1077(200003)15:2<103::AID-HUP149>3.0.CO;2-7. [DOI] [PubMed] [Google Scholar]
  4. Curran S, Wilson S, Musa S, Wattis J. Critical Flicker Fusion Threshold in patients with Alzheimer's disease and vascular dementia. International journal of geriatric psychiatry. 2004;19:575–581. doi: 10.1002/gps.1134. [DOI] [PubMed] [Google Scholar]
  5. Danjou P, Warot D, Hergueta T, Lacomblez L, Bouhours P, Puech AJ. Comparative study of the psychomotor and antistress effects of ritanserin, alprazolam and diazepam in healthy subjects: some trait anxiety-independent responses. International clinical psychopharmacology. 1992;7:73–79. [PubMed] [Google Scholar]
  6. Eagleman DM. Visual illusions and neurobiology. Nature Reviews Neurosci. 2001;2:920–926. doi: 10.1038/35104092. [DOI] [PubMed] [Google Scholar]
  7. Eagleman DM. In: Temporality, empirical approaches. In Oxford Companion to Consciousness. Wilken P, editor. Oxford: Oxford University Press; 2008. [Google Scholar]
  8. Eagleman DM, Jacobson JE, Sejnowski TJ. Perceived luminance depends on temporal context. Nature. 2004;428:854–856. doi: 10.1038/nature02467. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Eagleman DM, Tse PU, Buonomano D, Janssen P, Nobre AC, Holcombe AO. Time and the brain: how subjective time relates to neural time. J Neurosci. 2005;25:10369–10371. doi: 10.1523/JNEUROSCI.3487-05.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Herrick RM. Frequency thresholds for two-flash flicker and critical flicker: why they differ. Perception and Psychophysics. 1974 [Google Scholar]
  11. Saucer RT, Sweetbaum H. Perception of the shortest noticeable dark time by schizophrenics. Science. 1958;127:698–699. doi: 10.1126/science.127.3300.698. [DOI] [PubMed] [Google Scholar]
  12. Seitz AR, Nanez JE, Holloway SR, Watanabe T. Visual experience can substantially alter critical flicker fusion thresholds. Human Psychopharmacology Clinical and Experimental. 2004;20:55–60. doi: 10.1002/hup.661. [DOI] [PubMed] [Google Scholar]
  13. Simonson E, Brozek J. Flicker fusion frequency; background and applications. Physiological reviews. 1952;32:349–378. doi: 10.1152/physrev.1952.32.3.349. [DOI] [PubMed] [Google Scholar]
  14. Stetson C, Fiesta MP, Eagleman DM. Does time really slow down during a frightening event? PLoS One. 2007;2:1295. doi: 10.1371/journal.pone.0001295. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Tyler CW. Analysis of visual modulation sensitivity: two components in flicker perception. Vision Res. 1975;15:843–848. doi: 10.1016/0042-6989(75)90264-3. [DOI] [PubMed] [Google Scholar]
  16. Tyler CW. Specific deficits of flicker sensitivity in glaucoma and ocular hypertension. Investigative ophthalmology & visual science. 1981;20:204–212. [PubMed] [Google Scholar]
  17. Tyler CW. Analysis of visual modulation sensitivity. II. Peripheral retina and the role of photoreceptor dimensions. Journal of the Optical Society of America. 1985;2:393–398. doi: 10.1364/josaa.2.000393. [DOI] [PubMed] [Google Scholar]
  18. Tyler CW. Two processes control variations in flicker sensitivity over the life span. Journal of the Optical Society of America. 1989;6:481–490. doi: 10.1364/josaa.6.000481. [DOI] [PubMed] [Google Scholar]
  19. Tyler CW. Analysis of normal flicker sensitivity and its variability in the visuogram test. Investigative ophthalmology & visual science. 1991;32:2552–2560. [PubMed] [Google Scholar]
  20. Tyler CW, Hamer RD. Analysis of visual modulation sensitivity. IV. Validity of the Ferry-Porter law. Journal of the Optical Society of America. 1990;7:743–758. doi: 10.1364/josaa.7.000743. [DOI] [PubMed] [Google Scholar]
  21. Tyler CW, Hamer RD. Eccentricity and the Ferry-Porter law. J Opt Soc Am A Opt Image Sci Vis. 1993;10:2084–2087. doi: 10.1364/josaa.10.002084. [DOI] [PubMed] [Google Scholar]