Reliability of patients measuring blood pressure at home: prospective observational study

Alain Nordmann 1, Beate Frach 1, Tobias Walker 1, Benedict Martina 1, Edouard Battegay 1
PMCID: PMC28268  PMID: 10541509

Measuring blood pressure at home is recommended to distinguish sustained hypertension from white coat hypertension and to monitor treatment.1,2 Blood pressure measured at home is a better predictor of 24 hour ambulatory blood pressure monitoring, target organ damage, and cardiovascular mortality than that measured by a doctor.35 We aimed to assess the accuracy of patients measuring blood pressure at home, to identify predictors of poor accuracy, and to determine if poor accuracy impairs assessment of hypertension.

Subjects, methods, and results

We asked 54 consecutive patients with hypertension or suspected hypertension (aged 30-83 years), who had been referred for 24 hour ambulatory blood pressure monitoring, to measure their blood pressure at home twice daily between 0600 and 1000 for 30 days, and to record the time and readings. All agreed to participate. A nurse instructed each patient in the use of fully automated oscillometric blood pressure monitors with integrated memory devices (Omron-IC, Advance AG, Switzerland). To comply with the hospital's ethics committee, we informed patients (only cursorily) about the device's memory capacity. They were unaware that we would compare the device's stored measurements with their reported measurements. We classified measurements as correct if the timing of measurements (within 15 minutes) and single self reported values were identical to those recorded by the device, or if the average of multiple measurements differed by ≤3 mm Hg for systolic or diastolic pressure.

Overall, 2915 (89.9%) of 3240 requested measurements were performed of which 2121 (72.8%) were reported correctly. Thirty four patients (63%, 95% confidence interval 49% to 76%) reported 80%-100% of measurements correctly, 20 (37%) reported <80% correctly, and 12 (22%) reported ≤50% correctly. The table lists the types of mistakes. Differences of >5 mm Hg between means of reported versus recorded systolic (range 5-28 mm Hg) or diastolic (5-11 mm Hg) home blood pressure values occurred in only eight patients (15%, 7% to 27%), and there was no preference for inadequately reporting lower or higher values.

Using multiple regression analysis we identified low educational level as the only independent predictor of poor reporting accuracy (<80% of measurements correct, P=0.004). Patients who had had 8 or less school years had a relative risk of 3.39 (χ2 test, 1.54 to 7.46) for reporting ≤80% of measurements correctly compared with patients of a higher educational level.

Comment

Our study suggests that reporting accuracy of blood pressure measurements taken at home is acceptable in most patients, but that some patients of low educational level may have poor reporting accuracy, which may affect the assessment of blood pressure.

There are some potential limitations with generalising our study. Although the number of participants was small, we suggest that most patients accurately measure blood pressure at home for two reasons. Firstly, we found only eight patients (15%, 7% to 27%) with poor reporting accuracy and, secondly, we rarely identified differences of >5 mm Hg in mean systolic or diastolic blood pressure between reported and recorded measurements. Participants were informed about the capability of the device to record measurements and this may have influenced reporting accuracy. Participants were not aware of our aim to compare their measurements with those recorded by the device, therefore a substantially biased reporting accuracy seems unlikely.

In summary, it is possible but unlikely that poor reporting accuracy of blood pressure measurements taken at home affects the assessment of hypertension. In less educated patients ambulatory blood pressure monitoring may be preferable to self reported measurements.

Table.

Types of mistakes in incorrectly reported measurements of blood pressure (n=794) taken at home (27% of all performed measurements)

Mistake No (%) of mistakes No of patients (n=54)
Timing 270 (34) 27
Averaging 231 (29) 25
Single measurements: 209 (26) 23
 False high systolic values 50 (6) 14
 False high diastolic values 51 (6) 18
 False low systolic values 48 (6) 13
 False low diastolic values 60 (8) 16
Invented measurements 57 (7) 9
Measurements not reported 13 (2) 7
Miscellaneous 14 (2) 7

Acknowledgments

We thank Mrs Verena Brenneisen for technical assistance and Advance AG, Switzerland for supplying the Omron-IC monitors.

Footnotes

Funding: EB was supported by a Swiss Clinicians Opting for Research grant (32-31948.91) from the Swiss National Science Foundation.

Competing interests: None declared.

References

  • 1.The Sixth Report of the Joint National Committee on prevention, detection, evaluation, and treatment of high blood pressure. Arch Intern Med. 1997;157:2413–2446. doi: 10.1001/archinte.157.21.2413. [DOI] [PubMed] [Google Scholar]
  • 2.American College of Physicians. Automated ambulatory blood pressure and self-measured blood pressure monitoring devices. Their role in the diagnosis and management of hypertension. Ann Intern Med. 1993;118:889–892. doi: 10.7326/0003-4819-118-11-199306010-00010. [DOI] [PubMed] [Google Scholar]
  • 3.Campbell NRC, Bass M, Chockalingam A, Lebel M, Milkovich L.for the Canadian Coalition for High Blood Pressure Prevention and Control. Self measurement of blood pressure: benefits, risks and interpretation of readings Can J Cardiol 1995suppl H18–22. [PubMed] [Google Scholar]
  • 4.Julius S, Meija A, Jones K, Krause L, Schork N, van de Ven C, et al. “White coat” versus “sustained” borderline hypertension in Tecumseh, Michigan. Hypertension. 1990;16:617–623. doi: 10.1161/01.hyp.16.6.617. [DOI] [PubMed] [Google Scholar]
  • 5.Okubo T, Imai Y, Tsuji I, Nagai K, Kato J, Kikuchi N, et al. Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population-based observation in Ohasama, Japan. J Hypertens. 1998;16:971–975. doi: 10.1097/00004872-199816070-00010. [DOI] [PubMed] [Google Scholar]

Articles from BMJ : British Medical Journal are provided here courtesy of BMJ Publishing Group

RESOURCES