Skip to main content
Journal of Korean Neurosurgical Society logoLink to Journal of Korean Neurosurgical Society
. 2010 Mar 31;47(3):221–223. doi: 10.3340/jkns.2010.47.3.221

Rod Migration into the Posterior Fossa after Harms Operation : Case Report and Review of Literatures

Hyoung-Joon Chun 1, Koang Hum Bak 1,✉, Tae Hoon Kang 1, Hyeong-Joong Yi 1
PMCID: PMC2851083  PMID: 20379477

Abstract

C1 lateral mass and C2 pedicle (C1LM-C2P) fixation is a relatively new technique for atlantoaxial stabilization. Complications from C1LM-C2P fixation have been rarely reported. The authors report unilateral rod migration into the posterior fossa as a rare complication after this posterior C1-C2 stabilization technique. A 23-year-old man suffered severe head trauma and cervical spine injury after vehicle accident. He was unconscious for 2 months and regained consciousness. He underwent C1LM-C2P fixation for stabilization of type II odontoid process fracture described by Harms. The patient recovered without a major complication. Twenty months after operation, brain computed tomogram performed at psychology department for disability evaluation showed rod migration into the right cerebellar hemisphere. The patient had mild occipital headache and dizziness only regarding the misplaced rod. He refused further operation for rod removal. To our knowledge, this complication is the first report regarding rod migration after Harms method. We should be kept in mind the possibility of rod migration, and C1LM-C2P fixation should be performed with meticulous technique and long-term follow-up.

Keywords: Atlantoaxial fixation, Harms technique, Migration, Odontoid process fracture, Rod

INTRODUCTION

Several methods have been introduced for surgical correction of atlantoaxial instability (AAI) irrespective of causes14). Among them, techniques using screw system are regarded to have higher fusion rate than posterior wiring or clamp system14,17). Harms and Melcher9) technique of screw fixation with C1 lateral mass and C2 pedicle (C1LM-C2P) has been popularly used due to lower risk of vertebral artery (VA) injury since 2001. However, complications of this method have been less reported than those of transarticular screw fixation (TAF) because of short application period except common complications such as infection, malposition, nonunion, and vascular injury16). We present a rare complication of rod migration into the cerebellum that was found 20 months after Harms operation in a patient with atlantoaxial instability due to type II odontoid process fracture with pertinent literature reviews.

CASE REPORT

A 23-year-old man was brought in to an emergency room for stuporous mentality following severe vehicle collision. Admission head computed tomogram (CT) showed traumatic subarachnoid hemorrhage and intraventricular hemorrhage, and cervical X-ray and CT scan revealed type II odontoid process fracture. (Fig. 1) The patient was admitted in intensive care unit (ICU) and he recovered consciousness 2 months later. C1LM-C2P (Vertex, Medtronic Sofamor Danek, TN, USA) fixation combined with sublaminar wiring was performed for odontoid process fracture (Fig. 2).

Fig. 1.

Fig. 1

Preoperative lateral cervical X-ray (A) and computed tomography (B) show type II odontoid process fracture and resultant atlantoaxial instability.

Fig. 2.

Fig. 2

According to Harms technique, C1 lateral mass and C2 pedicle screw fixation was performed with sublaminar wiring using cadaveric bone. Immediate postoperative X-rays revealing good reduction between C1 and C2.

After atlantoaxial stabilization, his neck pain improved and he was discharged 3 weeks later. Follow-up cervical radiographics (Fig. 3) showed well-aligned instruments and solid bony fusion at 6 months postoperatively. After discharge, he was routinely followed up at the psychology department for impaired intelligence. On 20 months after operation, repeated brain CT scan (Fig. 4A) performed for disability evaluation showed rod migrated into the posterior fossa. Subsequent cervical X-ray (Fig. 4B, C) revealed migration of the rod into the cerebellum. The patient had mild occipital headache and dizziness only. We recommended surgical removal of rod device removal, but he refused operation.

Fig. 3.

Fig. 3

Follow-up radiographics on 6 months postoperatively shows good atlantoaxial stabilization.

Fig. 4.

Fig. 4

On 20 months after surgical procedure, the patient complained of headache and dizziness. Head computed tomography scan (A) and cervical X-rays (B and C) show migrated rod into the cerebellum.

DISCUSSION

The first description of surgical treatment for AAI appeared in the literature using heavy silk thread in 191015). Other posterior wiring techniques and clamp system were followed14). Grob and Magerl7) presented TAF technique and demonstrated higher fusion rate by more rigid fixation than posterior wiring methods. In a recent decade, Harms technique using cervical polyaxial screw rod system has been widely adopted2,9,11).

Harms method, C1LM-C2P fixation, has several advantages over TAF technique. First, individual placement of polyaxial screw in C1 and C2 allows direct manipulation of C1 and C2, enabling the following reduction and fixation9). Second, C1LM-C2P fixation has biomechanically superior or at least same stability when compared with TAF on the all dimensions of neck motion3,12,13). Third, the risk of VA injury is lower than TAF2,9,14). Additionally, Harms technique can be sufficiently applied when removal of posterior element of C1 or C2 is required for surgical decompression2).

Regardless of specific methods and instruments, a variety of complications have been reported following cervical spine surgery. Amongst, the migration of screw into the gastrointestinal tract4,5,19) or oral cavity extrusion6) after anterior approach with screw and plate system were rarely reported. Abumi et al.1) demonstrated the instrumentation-related complications in posterior cervical pedicle screw fixation including VA injury, radiculopathy caused by screw malposition, loosening of screw, and infection. A quite rare complication, rod migration to the cerebellum on 4 years later, was anecdotally reported in C4-C5 fusion surgery using Harrington instrumentation20).

However, complication with regard to C1LM-C2P fixation has seldom been reported so far. Because of the relatively scanty data and lack of long-term follow-up period, the exact incidence of complications has not been presented. Complications that related to the anatomical proximity were reported, such as hypoglossal nerve palsy10) and occipital neuralgia8,18). But problem associated with hardware failure, such as shown in the present case, has not been demonstrated yet. Because rod-screw system is familiar to spine surgeon, surgeons do not typically design consecutive long-term surveillance if patients do not complain the specific symptom.

In the current case, we have no culprit that the migration of the rod was attributed to any operative technical errors. However, in authors' opinion, there may be two possible hypotheses of rod migration. First, the cervical polyaxial screw rod systems on both sides were not tight enough to hold a rod against neck motion in this patient. The loose rod on one side may have propelled in cephalad direction, and then eroded the occipital bone, and finally located into the posterior fossa. The other rod seemed not tightly fixed either, but still locked to screw head. Second, the cause of rod migration may be due to the fusion failure, such as nonunion or resorption of graft materials, using cadaveric bone. During 20 months, the rod loosening was gradually developed from continuous mechanical load, under nonunion state.

The patient did not have corresponding neurologic complaints because he suffered severe brain trauma and residual sequelae. If the rod migrated medially, the result could have been fatal.

CONCLUSION

The clinicians should be alert that failure of assembled instrumentation failure can occur without obvious causes. In addition, regardless of neurologic symptoms of patient long-term follow-up should be performed until at least 3 years by annual basis following identification of bony fusion.

References

  • 1.Abumi K, Shono Y, Ito M, Taneichi H, Kotani Y, Kaneda K. Complications of pedicle screw fixation in reconstructive surgery of the cervical spine. Spine (Phila Pa 1976) 2000;25:962–969. doi: 10.1097/00007632-200004150-00011. [DOI] [PubMed] [Google Scholar]
  • 2.Aryan HE, Newman CB, Nottmeier EW, Acosta FL, Jr, Wang VY, Ames CP. Stabilization of the atlantoaxial complex via C-1 lateral mass and C-2 pedicle screw fixation in a multicenter clinical experience in 102 patients : modification of the Harms and Goel technique. J Neurosurg Spine. 2008;8:222–229. doi: 10.3171/SPI/2008/8/3/222. [DOI] [PubMed] [Google Scholar]
  • 3.Claybrooks R, Kayanja M, Milks R, Benzel E. Atlantoaxial fusion : a biomechanical analysis of two C1-C2 fusion techniques. Spine J. 2007;7:682–688. doi: 10.1016/j.spinee.2006.08.010. [DOI] [PubMed] [Google Scholar]
  • 4.Fountas KN, Kapsalaki EZ, Machinis T, Robinson JS. Extrusion of a screw into the gastrointestinal tract after anterior cervical spine plating. J Spinal Disord Tech. 2006;19:199–203. doi: 10.1097/01.bsd.0000164164.11277.49. [DOI] [PubMed] [Google Scholar]
  • 5.Gazzeri R, Tamorri M, Faiola A, Gazzeri G. Delayed migration of a screw into the gastrointestinal tract after anterior cervical spine plating. Spine (Phila Pa 1976) 2008;33:268–271. doi: 10.1097/BRS.0b013e31816b8831. [DOI] [PubMed] [Google Scholar]
  • 6.Geyer TE, Foy MA. Oral extrusion of a screw after anterior cervical spine plating. Spine (Phila Pa 1976) 2001;26:1814–1816. doi: 10.1097/00007632-200108150-00019. [DOI] [PubMed] [Google Scholar]
  • 7.Grob D, Magerl F. [Surgical stabilization of C1 and C2 fractures.] Orthopade. 1987;16:46–54. [PubMed] [Google Scholar]
  • 8.Gunnarsson T, Massicotte EM, Govender PV, Raja Rampersaud Y, Fehlings MG. The use of C1 lateral mass screws in complex cervical spine surgery : indications, techniques, and outcome in a prospective consecutive series of 25 cases. J Spinal Disord Tech. 2007;20:308–316. doi: 10.1097/01.bsd.0000211291.21766.4d. [DOI] [PubMed] [Google Scholar]
  • 9.Harms J, Melcher RP. Posterior C1-C2 fusion with polyaxial screw and rod fixation. Spine (Phila Pa 1976) 2001;26:2467–2471. doi: 10.1097/00007632-200111150-00014. [DOI] [PubMed] [Google Scholar]
  • 10.Hong JT, Lee SW, Son BC, Sung JH, Kim IS, Park CK. Hypoglossal nerve palsy after posterior screw placement on the C-1 lateral mass. Case report. J Neurosurg Spine. 2006;5:83–85. doi: 10.3171/spi.2006.5.1.83. [DOI] [PubMed] [Google Scholar]
  • 11.Kim YS, Lee JK, Kim JH, Kim SH. Post-traumatic atlantoaxial rotatory dislocation in an adult treated by open reduction and C1-C2 transpedicular screw fixation. J Korean Neurosurg Soc. 2007;41:248–251. [Google Scholar]
  • 12.Kuroki H, Rengachary SS, Goel VK, Holekamp SA, Pitkänen V, Ebraheim NA. Biomechanical comparison of two stabilization techniques of the atlantoaxial joints: transarticular screw fixation versus screw and rod fixation. Neurosurgery. 2005;56:151–159. doi: 10.1227/01.neu.0000144838.01478.35. discussion 151-159. [DOI] [PubMed] [Google Scholar]
  • 13.Melcher RP, Puttlitz CM, Kleinstueck FS, Lotz JC, Harms J, Bradford DS. Biomechanical testing of posterior atalntoaxial fixation techniques. Spine (Phila Pa 1976) 2002;27:2435–2440. doi: 10.1097/00007632-200211150-00004. [DOI] [PubMed] [Google Scholar]
  • 14.Menendez JA, Wright NM. Techniques of posterior C1-C2 stabilization. Neurosurgery. 2007;60:S103–S111. doi: 10.1227/01.NEU.0000249220.50085.E4. [DOI] [PubMed] [Google Scholar]
  • 15.Mixter SJ, Osgood RB. IV. Traumatic Lesions of the Atlas and Axis. Ann Surg. 1910;51:193–207. doi: 10.1097/00000658-191002000-00004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ondra SL, Marzouk S, Ganju A, Morrison T, Koski T. Safety and efficacy of C2 pedicle screws placed with anatomic and lateral C-arm guidance. Spine (Phila Pa 1976) 2006;31:E263–E267. doi: 10.1097/01.brs.0000214882.34674.be. [DOI] [PubMed] [Google Scholar]
  • 17.Papagelopoulos PJ, Currier BL, Hokari Y, Neale PG, Zhao C, Berglund LJ, et al. Biomechanical comparison of C1-C2 posterior arthrodesis techniques. Spine (Phila Pa 1976) 2007;32:E363–E370. doi: 10.1097/BRS.0b013e318060cc65. [DOI] [PubMed] [Google Scholar]
  • 18.Rhee WT, You SH, Kim SK, Lee SY. Troblesome occipital neuralgia developed by C1-C2 Harms construct. J Korean Neurosurg Soc. 2008;43:111–113. doi: 10.3340/jkns.2008.43.2.111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Sahjpaul RL. Esophageal perforation from anterior cervical screw migration. Surg Neurol. 2007;68:205–209. doi: 10.1016/j.surneu.2006.09.032. discussion 209-210. [DOI] [PubMed] [Google Scholar]
  • 20.Yablon IG, Cowan S, Mortara R. The migration of a Harrington rod after cervical fusion. Spine (Phila Pa 1976) 1993;18:356–358. doi: 10.1097/00007632-199303000-00009. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Korean Neurosurgical Society are provided here courtesy of The Korean Neurosurgical Society