Abstract
We obtain quantitative phase reconstructions from differential phase contrast images obtained with a scanning transmission x-ray microscope and 2.5 keV x rays. The theoretical basis of the technique is presented along with measurements and their interpretation.
X-ray imaging techniques have long capitalized on absorption contrast, governed by the imaginary component β of the refractive index. More recently phase contrast techniques have exploited the higher contrast offered by the real component δ of the refractive index at x-ray energies [1-13]. Full-field methods have considerable success in quantitative phase reconstruction [6-12]. However, these methods often cannot reconstruct specimens with significant absorption [6,12], usually greater than about 10%. Interferometric methods [4] may further be limited by phase-wrapping effects to determine the phase only modulo 2π radians. Both the capabilities and the measurement geometry determine the application of these techniques; for example, quantitative full-field methods have been employed for tomography [10-13].
In this Letter we describe a quantitative technique for phase imaging with a scanning transmission x-ray microscope (STXM). Compatibility with the STXM geometry will enable the technique to be combined with fluorescence microscopy to determine elemental concentrations from a single x-ray measurement. We advance the imaging theory for this system, and use this to determine the conditions for differential phase contrast imaging. The technique is robust in the presence of absorption, intensity fluctuations, and noise. The effects of a novel differential absorption contrast (DAC) term are described.
The transmitted intensity has been used in electron [14], x-ray [15-19], and optical [20] scanning transmission microscopes to obtain differential phase contrast (DPC) images. The electron technique was recently adapted for x-rays and extended to provide quantitative information [5]. These analyses use a weak-specimen approximation to invert the contrast transfer function, and require the phase shift δkt and absorption βkt to be less than 0.1, which places severe restrictions on their application. The technique outlined in this Letter allows quantitative reconstruction of specimens with arbitrary total phase shift and without phase-wrapping effects.
In a STXM one usually employs a single element detector to obtain a transmission image of a specimen. Figure 1 shows a typical optical arrangement used for a STXM equipped instead with an annular quadrant detector (AQD). Other investigations have used CCDs [21], quadrant [14] and three-segment [22] designs, and dedicated configurations optimized for combined differential interference contrast and DPC imaging [5,19]. Our current detector is optimized for DPC imaging [23].
The amplitude in the focal plane (with coordinate ) is related to the amplitude at the zone plate [24]:
(1) |
where f is the focal length of the lens and k the wave number. We assume coherent and uniform illumination of the zone plate. The effect of incomplete coherence is a broadening of the focus with a commensurate loss of resolution. The effect on the recovered phase of the slightly nonuniform illumination typical of STXM is negligible.
We approximate the amplitude at the detector plane located a distance z downstream of the focal plane using Fraunhofer propagation, which is justified as z is typically several focal lengths [24]. Accordingly, the intensity in the detector plane with the specimen absent is
(2) |
We introduce a specimen into the focal plane by multiplying the focal-plane amplitude by the specimen function where t is the specimen thickness and is the complex refractive index decrement. Expanding the phase and absorption terms in a Taylor series about the point illuminated by the beam (without loss of generality chosen to be 0) gives
(3) |
Ignoring second-order terms for the present, the intensity in the detector plane is given by
(4) |
The constant absorption term preceding the integral describes the absorption contrast used in most STXM measurements. The gradient terms within the integral are responsible for differential phase contrast (DPC) and differential absorption contrast (DAC), respectively. The DAC contribution is negligible when is large and the probe is small. In particular, the DAC term is negligible when it does not vary appreciably over the probe dimensions. As 95% of the intensity falls within the first 4 maxima of the focal spot [25], this condition requires
(5) |
where B4 ≈ 13.324 is the fourth zero of the Bessel function and δRN = 50 is the finest zone width of the zone plate. Presuming constant β and examining the limit in terms of a thickness variation, we require the thickness gradient ∇(t)0 ≪ 144 for carbon at 2.5 keV [26]. When DAC can be neglected, the intensity in the detector plane becomes
(6) |
We have performed a change of variable in the detector plane given by , which follows from the Fourier shift theorem, and identified the resulting function of as the no-specimen intensity determined in Eq. (2). The change of variable describes a shift of the intensity in the detector plane due to the specimen phase gradient. The angular deflection is , in agreement with the predictions of a simple refractive treatment [27]. It is interesting to note that deflection angles are typically of order 1 μrad, the objective lens N. A. 5 mrad, and the specimen wedge angle 1 rad.
The second-order terms in the Taylor series expansion [Eq. (3)] are even, and so their Fourier transforms are even. These terms redistribute the amplitude symmetrically about the shifted center, with negligible effects on the center-of-mass of the intensity distribution. While third and higher odd orders can shift the center-of-mass of the intensity distribution, their effect is negligible due to the use of a focused probe, which restricts the contribution of these terms to the small values of . Our wave-propagation simulations show that the interaction of the specimen with the beam shifts the intensity in the detector plane as expected but also gives rise to intensity fringes. These fringes result from the higher order terms in Eq. (6), and, as discussed, do not affect the location of the center-of-mass of the intensity.
We use the quadrant detector to quantify the deflection of the intensity distribution, and define the signal with horizontal and vertical components Sx = (IR − IL)/ITOT and Sy = (IR − IL)/ITOT respectively, where IT, IB, IR, and IL refer to the counts recorded by the top, bottom, right, and left halves, and ITOT is the sum of all counts. Normalizing the signal to the total intensity accounts for the average specimen absorption term in Eq. (6) and removes the effect of source intensity variations. Equation (6) describes the intensity in the detector plane as a shifted image of the objective pupil function magnified by the factor . In the x-ray regime the shift is small and the intensity reaching the top half of the detector is
(7) |
where Δy is the vertical component of the beam deflection and RZP and RCS are the radii of the zone plate and the central stop, respectively. Similar expressions can be derived for IB, IR, and IL. The DPC signal is given by
(8) |
We have treated the Sx values to correct for detector misalignment and beam drift by requiring that each row of the image sums to zero, which is valid for an isolated specimen. A similar normalization was applied to Sy. Other treatments exist in cases where the specimen is not isolated. Figure 2 shows Sx for a measurement of 5-μm-diameter polystyrene spheres [28], taken at beam line 2-ID-B at the Advanced Photon Source. The x-ray probe was formed using a 160-μm-diameter zone plate, a 40-μm-diameter central stop, and a 30-μm-diameter order-sorting aperture; the x-ray photon energy was 2.5 keV. The radius of the focal spot is approximately 60 nm [29] and the depth of focus approximately ±10 μm [30], and so the 5 μm spheres are entirely within the focal plane. The specimen was prepared by placing a small drop of a 1:1 ethanol:water suspension of the spheres onto an electron microscopy grid. Most of the excess solution was wicked away with an absorbent cloth, leaving the spheres behind.
The reconstruction of gradient maps is a general physical problem. The Hartmann sensor has been used in observational astronomy and adaptive optics; reconstructions have used physical constraints to optimize orthonormal basis sets with least-squares fitting routines [31]. Various matrix approaches have been used [32,33], but these are computationally intensive. We use instead a Fourier integration technique [34,35]. The Fourier derivative theorem relates the two dimensional integral to the directional derivatives by
(9) |
where denotes Fourier transformation, and u and v are reciprocal-space coordinates of the forward Fourier transform. The indeterminate zero-frequency term of Eq. (9) diverges due to the zero value of the denominator. In practice we prevent numerical instability by setting the zero-frequency element of the forward Fourier transform to zero, and the corresponding element of u + iv to a nonzero value. We normalize the integrated phase so the average of the perimetric values is zero.
The integrated phase is given by the real part of Eq. (9). Nonzero elements in the imaginary part of Eq. (9) result from small “contradictions” in the derivatives, leaking power from the real to the imaginary component. As such, the imaginary component reflects errors in the reconstruction, and provides feedback for the accuracy of the integral. Here the imaginary part of Eq. (9) is less than about 10% of the real part, and the contradictions occur mostly at the perimeter of the spheres and at the ethanol:water meniscus, where DAC and higher order effects may be discernible. However, the influence of these few values on the integrated phase is mitigated by the use of a small step size.
Figure 3 presents the reconstructed thickness for the cluster of spheres. We interpreted the reconstructed phase as a thickness using the Henke tabulation [36] to determine δ = 3.81 × 10−5, using C8H8 for the molecular formula and a density of 1.05 g/cm3 for polystyrene. The uncertainty associated with these assumptions is below about 10%. The data of Fig. 4 show the thickness profile along a line passing through one of the spheres. Also shown on this plot is the thickness profile that one would expect for a 5-μm sphere added to a small sinusoidal background. The small deviations between the measured and calculated data (indicated) are due to the residual solution. The phase excursion of a single sphere is a little above 2.5 rad.
The two-dimensional integration is overconstrained, and determines a solution consistent with both directional derivatives. As a result there is a nonlocal relationship between the determined phase and the measured data, with the effect that the procedure is robust to noise. The detector signal-to-noise is of order 4000 [23], which corresponds here to a phase gradient of about 1.5 × 10−5 rad/nm or a carbon thickness gradient of about 1 part in 60. Because of the noise insensitivity of the integration we expect that this is a conservative estimate of the measurement sensitivity.
We have measured an object with dimensions and composition similar to a typical biological specimen. Assuming an average composition for, e.g., a cellular matrix will allow direct determination of cellular volumes and therefore also of trace elemental concentrations when used in conjunction with scanning fluorescence x-ray microscopy.
Acknowledgments
M. d. J. thanks B. F. Smith and T. S. Munson for discussions of numerical integration routines, C. Rau for help with the STXM, B. Tieman for computing assistance, and J. Arko for mechanical design and fabrication. Use of the Advanced Photon Source was supported by the U.S. Department of Energy, Basic Energy Sciences, Office of Energy Research, under Contract No. DE-AC02-06CH11357.
Footnotes
PACS numbers: 42.30.Rx, 07.79.–v, 07.85.Tt
References
- [1].Snigirev A, et al. Rev. Sci. Instrum. 1995;66:5486. [Google Scholar]
- [2].Wilkins SW, et al. Nature (London) 1996;384:335. [Google Scholar]
- [3].Cloetens P, et al. J. Phys. D. 1996;29:133. [Google Scholar]
- [4].Bonse U, Hart M. Appl. Phys. Lett. 1965;6:155. [Google Scholar]
- [5].Hornberger B, Feser M, Jacobsen C. Ultramicroscopy. 2007;107:644. doi: 10.1016/j.ultramic.2006.12.006. [DOI] [PubMed] [Google Scholar]
- [6].Nugent KA, Gureyev TE, Cookson DF, Paganin D, Barnea Z. Phys. Rev. Lett. 1996;77:2961. doi: 10.1103/PhysRevLett.77.2961. [DOI] [PubMed] [Google Scholar]
- [7].Paganin D, Nugent KA. Phys. Rev. Lett. 1998;80:2586. [Google Scholar]
- [8].Allman BE, et al. J. Opt. Soc. Am. A. 2000;17:1732. doi: 10.1364/josaa.17.001732. [DOI] [PubMed] [Google Scholar]
- [9].Pfeiffer F, Weitkamp T, Bunk O, David C. Nature Phys. 2006;2:258. [Google Scholar]
- [10].Barty A, Nugent KA, Roberts A, Paganin D. Opt. Commun. 2000;175:329. [Google Scholar]
- [11].McMahon PJ, et al. Opt. Commun. 2003;217:53. [Google Scholar]
- [12].Cloetens P, et al. Appl. Phys. Lett. 1999;75:2912. [Google Scholar]
- [13].Pfeiffer F, Kottler C, Bunk O, David C. Phys. Rev. Lett. 2007;98:108105. doi: 10.1103/PhysRevLett.98.108105. [DOI] [PubMed] [Google Scholar]
- [14].Dekkers NH, de Lang H. Optik (Jena) 1974;41:452. [Google Scholar]
- [15].Kaulich B, et al. Opt. Express. 2002;10:1111. doi: 10.1364/oe.10.001111. [DOI] [PubMed] [Google Scholar]
- [16].Thieme J, Schmahl G, Rudolph D, Umbach E, editors. X-ray Microscopy and Spectromicroscopy. Springer; Berlin: 1998. [Google Scholar]
- [17].Kagoshima Y, et al. Jpn. J. Appl. Phys. 2004;43:L1449. [Google Scholar]
- [18].Gianoncelli A, et al. Appl. Phys. Lett. 2006;89:251 117. [Google Scholar]
- [19].Feser M, et al. Nucl. Instrum. Methods Phys. Res., Sect. A. 2006;565:841. doi: 10.1016/j.nima.2006.05.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Amos WB, Reichelt S, Cattermole DM, Laufer J. J. Microsc. 2003;210:166. doi: 10.1046/j.1365-2818.2003.01189.x. [DOI] [PubMed] [Google Scholar]
- [21].Morrison G, Eaton WJ, Barrett R, Charalambous P. J. Phys. IV (France) 2003;104:547. [Google Scholar]
- [22].McCallum BC, Landauer MN, Rodenburg JM. Optik (Jena) 1995;101:53. [Google Scholar]
- [23].Hornberger B, et al. J. Synchrotron Radiat. to be published. [Google Scholar]
- [24].Goodman JW. Introduction to Fourier Optics. McGrawHill; New York: 1968. [Google Scholar]
- [25].Born M, Wolf E. Principles of Optics. Cambridge University Press; Cambridge, England: 2002. [Google Scholar]
- [26].Evaluating the same expression for gold at 2.5 keV and 10 keV requires ∇(t)0 ≪ 1 and ∇(t)0 ≪ 20, respectively, and so DAC contributions can be significant.
- [27].de Jonge MD, et al. Nucl. Instrum. Methods Phys. Res., Sect. A. 2007;582:218. doi: 10.1016/j.nima.2007.08.111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [28]. Fluka Micro particles based on polystyrene no. 79633, http://www.sigmaaldrich.com/
- [29].Quiney HM, et al. Nature Phys. 2006;2:101. [Google Scholar]
- [30].Attwood D. Soft X-Rays and Extreme Ultraviolet Radiation. Cambridge University Press; Cambridge, England: 1999. [Google Scholar]
- [31].Rimmer MP. Appl. Opt. 1974;13:623. doi: 10.1364/AO.13.000623. [DOI] [PubMed] [Google Scholar]
- [32].Herrmann J. J. Opt. Soc. Am. 1980;70:28. [Google Scholar]
- [33].Southwell WH. J. Opt. Soc. Am. 1980;70:998. [Google Scholar]
- [34].Arnison MR, et al. J. Microsc. 2004;214:7. doi: 10.1111/j.0022-2720.2004.01293.x. [DOI] [PubMed] [Google Scholar]
- [35].Kottler C, David C, Pfeiffer F, Bunk O. Opt. Express. 2007;15:1175. doi: 10.1364/oe.15.001175. [DOI] [PubMed] [Google Scholar]
- [36].Henke BL, Gullikson EM, Davis JC. At. Data Nucl. Data Tables. 1993;54:181. [Google Scholar]