Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2010 May 10.
Published in final edited form as: Int J Lab Hematol. 2008 Apr 14;31(5):496–504. doi: 10.1111/j.1751-553X.2008.01066.x

The effect of anticoagulant, storage temperature and dilution on cord blood hematology parameters over time

K J FREISE *, R L SCHMIDT , E L GINGERICH , P VENG-PEDERSEN *, J A WIDNESS
PMCID: PMC2866065  NIHMSID: NIHMS199748  PMID: 18422712

SUMMARY

The objective of the study was to determine whether selected hematologic parameters measured on umbilical cord blood samples using an automated hematology analyzer (Sysmex XE-2100) were affected by (i) anticoagulant (the specimens were collected in EDTA vs. sodium heparin), (ii) temperature (the specimens were maintained at 4° C vs. room temperature for up to 72 h) and (iii) 1 : 5 dilution vs. undiluted using the manufacturer's diluting solution. Use of heparin, instead of EDTA, had little effect on the hematologic results (n = 8) except for lower platelet and progenitor cell counts. Results were remarkably stable for 72 h at either room temperature or 4° C except for modest red blood cell swelling at 24 h. Specimens of blood diluted at 1 : 5 had an immediate small, but significant change on white cell count (+13.3%), reticulocyte count (−11.2%) and reticulocyte hemoglobin content (−19.6%). Diluted samples did not change further over 4 h at room temperature. With a 1 : 5 dilution, analysis of 40 μl of cord blood stored for 3 days at room temperature may provide useful hematologic information with little phlebotomy loss.

Keywords: Neonate, reticulocytes, red blood cell parameters, automated hematology analyzer, heparin, EDTA

INTRODUCTION

Phlebotomy blood loss for clinical laboratory testing in the neonatal intensive care unit (NICU) is a major contributor to anemia and the resultant red blood cell (RBC) transfusions administered to critically ill, very low birth weight (VLBW) infants (Lin et al., 2000; Madan et al., 2005; Widness et al., 2005). Indeed, reduction in the volume of blood drawn for laboratory testing has been shown to reduce the number of red blood cell (RBC) transfusions (Madan et al., 2005; Widness et al., 2005). Therefore, the development of laboratory methodologies that require smaller blood volumes for clinical and research testing is an important objective for improving neonatal care by preventing anemia and thereby avoiding the risks and cost associated with RBC transfusion.

Hematology analyzers utilizing flow cytometry are routinely used to provide rapid, reliable and accurate determination of hematology parameters in the clinical setting. Generally, blood samples are collected in tubes containing the recommended anticoagulant, ethylenediaminetetraacetic acid (EDTA), and sent to the laboratory for immediate analysis. In the setting of the neonatal intensive care unit (NICU), 250–500 μl samples of whole blood are typically required for hematologic testing. In the NICU setting, blood collected for clinical laboratory testing (i) is often drawn in heparin or anticoagulants other than EDTA and (ii) is often drawn in volumes that are in excess of that needed. If properly saved and labeled, the limited volumes of leftover anticoagulated whole blood ordered for clinically indicated purposes could be retrieved, diluted and analyzed to yield valuable clinical and/or research data without the need for drawing additional blood. Available samples are commonly collected in sodium heparin containing tubes for analysis of blood gases, electrolytes and other blood chemistries.

Previous studies in adults have demonstrated the stability of many hematologic parameters over time (Paxton & Bendele, 1993; Peng et al., 2001; deBaca et al., 2006; Hedberg & Lehto, 2008) and have documented which parameters are affected when sodium heparin is used in place of EDTA (Paxton & Bendele, 1993; Peng et al., 2001). None of the studies have investigated these effects on fetal or neonatal blood, nor have they investigated the effect of sample dilution on hematologic parameters. Furthermore, these effects have not been evaluated for the most recent hematologic variables analyzed with the newer, highly sophisticated hematology analyzers. Examples of these newer measurements include the hematopoietic progenitor cell (HPC) count, nucleated red blood cell (NRBC) count, immature reticulocyte fraction (IRF) and reticulocyte hemoglobin content (Ret He).

Therefore, using fetal umbilical cord blood as a surrogate for neonatal blood, the objective of the current study was to determine the effect of EDTA vs. sodium heparin on selected hematologic parameters kept at different storage temperatures for up to 72 h. We also sought to examine the effect of a 1 : 5 dilution of cord blood on hematologic parameters using the manufacturer's cellular diluting solution while maintaining samples at room temperature for 4 h. The findings of these studies have important clinical and research implications for critically ill, VLBW infants relative to reducing the volume of blood required for laboratory testing and the resultant number of RBC transfusions.

MATERIALS AND METHODS

Eight study subjects born after uncomplicated pregnancy, labor and delivery were enrolled. This included six term and two preterm (<37 weeks postmenstrual age) infants. All procedures were approved by the University of Iowa Human Subject Internal Review Board Committee and parental written consent was obtained for all study subjects.

Immediately after delivery of the placenta, fetal cord blood samples were collected in EDTA and sodium heparin containing Vacutainer® tubes (BD Vacutainer® sodium heparin, 143 USP Units, Reference #367874, and K2 EDTA, 10.8 mg, Reference #367899, Franklin Lakes, NJ, USA). Only tubes that filled completely were included. Blood samples from individual collection tubes were split into twelve 0.5 ml aliquots and stored at room temperature (~20° C) or 4° C (six aliquots per storage temperature). Selected hematologic parameters were determined serially on individual aliquots over 72 h, the period of time over which stored, retrieved blood remaining from clinical analyses might commonly be available. Separate aliquots from each anticoagulant tube at the two temperatures were measured in manual mode at 0, 4, 8, 24, 48 and 72 h using the Sysmex XE-2100 automatic hematology analyzer (Sysmex Corporation, Kobe, Japan). The analyzer was calibrated and quality controlled according to the manufacturer's recommendations. Immediately after collection, aliquots from both the EDTA and heparin samples were diluted 1 : 5 with the cellular suspension medium CELLPACK (Sysmex Corporation). The diluted EDTA and heparin blood samples were aliquoted in five 0.2 ml volumes and stored at room temperature for up to 4 h, the estimated period of time before diluted blood samples might commonly be analyzed. Aliquots of the 1 : 5 diluted blood samples were analyzed in capillary mode after 0, 0.5, 1, 2 and 4 h using the same hematology analyzer.

The effects of anticoagulant and dilution at time 0 were analyzed for the hematologic parameters using a linear mixed effects model, with subject as a random effect and anticoagulant, dilution and their interaction as fixed effects. The effects of anticoagulant and storage temperature over time for undiluted cord blood hematologic results were analyzed using a linear mixed effects repeated measures model, with subject as a random effect and anticoagulant, temperature, time and their interactions as fixed effects. The effect of anticoagulant over time on diluted cord blood was analyzed similarly using subject as a random effect, and using anticoagulant, time and their interactions as fixed effects.

The correlation between repeated measurements on the same subject for both diluted and undiluted blood was modeled as a Toeplitz (diagonal constant) variance–covariance matrix structure, based on the Akaike's Information Criterion (Akaike, 1974). The null hypothesis that |μ1 − μ0| ≤5%μ0 was tested against the alternative hypothesis that |μ1 − μ0| > 5%μ0 using Student's t-test, where μ1 is the mean value of the treatment combination of interest and μ0 is the mean baseline value of the control treatment. A difference of larger than 5% from μ0 was selected for the alternative hypothesis as smaller differences are not generally of clinical and/or biological relevance. For the effect of anticoagulant and dilution at time 0 analysis, the baseline control treatment was the mean value of the EDTA/undiluted/time 0 treatment for both the 4° C and room temperature samples. Similarly, for the effect of anticoagulant and storage temperature over time for undiluted blood analysis and for the effect of anticoagulant over time on diluted blood analysis, the baseline control treatments were the ‘EDTA/4° C/undiluted/time 0 treatment’ and ‘EDTA/diluted/time 0 treatment’, respectively. Statistically significant differences were determined at α = 0.05 probability of type I error rate. Dunnett's post hoc adjustment was performed for multiple comparisons. All statistical analyses were conducted using sas® for Windows (Version 9.1.3, Service Pack 4; SAS® Institute Inc., Cary, NC, USA).

RESULTS

Collection of cord blood with sodium heparin as the anticoagulant did not result in any statistically significant effects on any of hematologic parameters that were greater than 5% of the EDTA values of undiluted blood samples measured immediately after collection except for platelets, immature platelet fraction (IPF) and HPC (Figure 1). For undiluted samples, the platelet count was lower and the IPF was higher when heparin was used as the anticoagulant rather than EDTA. Similarly, dilution of fresh cord blood with CELLPACK did not affect any of the hematologic parameters more than 5% of the EDTA–undiluted blood parameters values except for reticulocyte count and Ret He (Figure 1). With dilution, reticulocyte count dropped from 214 ± 10.8 (mean ± standard error of the mean) to 190 ± 8.88×103 cells/μl (−11.2%) and Ret He dropped from 32.6 ± 0.415 to 26.3 ± 0.510 (−19.6%), irrespective of the anticoagulant. Additionally, there was a statistically significant effect of dilution greater than 5% of the control treatment on white blood cell (WBC) counts when a statistical test was conducted across both diluted EDTA and heparin treatments simultaneously. However, the individual EDTA/diluted and heparin/diluted WBC count effects were not statistically greater than 5% of the control treatment. Unlike reticulocytes and Ret He, however, dilution with CELLPACK increased the WBC count across anticoagulants by 13.3%.

Figure 1.

Figure 1

Mean hematologic values immediately after collection normalized by the mean EDTA/undiluted/time 0 h sample value. Horizontal line indicates the normalized value of 1.0 and bars represent the normalized standard error (n = 8). Hematologic parameters missing for the diluted samples are not reported by the Sysmex XE-2100 automatic hematology analyzer when operated in capillary (i.e. diluted) mode.

In general, the undiluted cord blood hematologic parameters changed little over the 72 h postcollection period irrespective of the storage temperature or the anticoagulant (Table 1). Exceptions to this were the WBC count, immature granulocyte (IG) count, HPC count, IPF, mean corpuscular volume (MCV) (Figure 2c), platelets (Figure 2d) and Ret He. Other than platelets, IG count, IPF and MCV, none of these changes were statistically greater than 5% of the EDTA/4° C/undiluted/time 0 h value until after 24 h after blood collection. While the 5% difference limit seems large relative to quality control variances for hematologic parameters, in the NICU even greater variances are observed for capillary blood sampling relative to venous or arterial sampling (Linderkamp et al., 1977; Rivera, 1982). For all hematologic parameters evaluated other than HPC count, whole blood collected into EDTA tubes and stored at 4° C did not change more than 5% of the time 0 h value through 72 h of storage. When blood was collected into heparin containing tubes, storage at room temperature was superior to storage at 4° C after 48 to 72 h for WBC count, IG count, IPF and Ret He. Conversely, for MCV storage at 4° C was superior to storage at room temperature over time irrespective of anticoagulant. The RBC and reticulocyte count did not change statistically over time, irrespective of the anticoagulant or storage temperature (Figure 2a,b, respectively). In contrast, MCV slowly increased over time with either EDTA or heparin as anticoagulants when stored at room temperature.

Table 1.

Mean undiluted cord blood hematologic parameter results by anticoagulant, storage temperature and duration of storage (n = 8)

Time (h)
Variable Anticoagulant Temperature 0 4 8 24 48 72
WBC (/μl) EDTA 4° C 9418 9464 9604 9894 9943 9725
RT 9619 9533 9480 9544 9751 10084
Heparin 4° C 9110 9219 8904 8645 7991 7511*
RT 9003 9036 9166 9131 8970 9238
IG (/μl) EDTA 4° C 178 188 203 175 156 135
RT 205 198 180 179 188 129
Heparin 4° C 179 213 296 370* 416* 431*
RT 183 168 164 164 176 199
HPC (/μl) EDTA 4° C 9.14 15.1 12.6 14.0 16.0 20.3*
RT 5.43 10.7 10.0 11.7 12.0 12.0
Heparin 4° C 2.14 2.57 1.00 2.00 4.29 4.14
RT 2.43 1.00 0.71 0.43 0.43 0.43
Platelets (103/μl) EDTA 4° C 271 244 243 240 250 255
RT 275 282 282 279 278 277
Heparin 4° C 198* 49.8* 60.9* 53.4* 72.0* 85.9*
RT 191* 171* 180* 161* 204* 199*
MPV (fl) EDTA 4° C 10.2 10.4 10.5 10.4 10.7 10.6
RT 10.2 10.5 10.4 10.6 10.8 10.9
Heparin 4° C 10.0 10.4 10.4 10.5 10.8 11.2
RT 9.7 10.3 10.3 10.4 9.9 10.1
IPF (%) EDTA 4° C 3.00 3.74 3.65 4.19 4.06 4.30
RT 3.20 3.26 3.74 3.74 4.45 4.99
Heparin 4° C 4.63 16.1* 13.8* 14.8* 16.1* 18.0*
RT 4.51 7.75 7.56 8.39 5.60 6.64
Hgb (g/dl) EDTA 4° C 15.9 15.9 15.8 16.0 15.9 15.9
RT 15.8 15.8 15.7 16.0 15.9 16.0
Heparin 4° C 15.4 15.4 15.3 15.5 15.6 15.6
RT 15.4 15.4 15.4 15.4 15.5 15.0
RBC (106/μl) EDTA 4° C 4.32 4.32 4.32 4.34 4.35 4.33
RT 4.31 4.27 4.28 4.28 4.29 4.28
Heparin 4° C 4.19 4.21 4.20 4.24 4.26 4.26
RT 4.23 4.22 4.19 4.23 4.21 4.20
MCV (fl) EDTA 4° C 108 108 108 108 109 110
RT 108 110 111 116* 121* 123*
Heparin 4° C 106 106 106 107 108 108
RT 106 107 108 114 119* 121*
RDW (%) EDTA 4° C 16.8 16.7 16.6 16.6 16.4 16.3
RT 16.8 16.8 16.9 17.3 17.4 17.4
Heparin 4° C 16.6 16.6 16.5 16.3 16.2 16.1
RT 16.6 16.6 16.6 17.1 17.1 17.0
MCH (pg) EDTA 4° C 36.9 36.7 36.7 36.7 36.6 36.7
RT 36.7 36.9 36.7 37.3 37.1 37.3
Heparin 4° C 36.8 36.6 36.4 36.7 36.5 36.6
RT 36.4 36.6 36.6 36.5 36.9 35.6
Ret. He (pg) EDTA 4° C 32.3 31.6 31.6 32.6 34.4 35.1
RT 32.4 33.1 33.2 33.7 34.3 34.7
Heparin 4° C 32.8 32.5 32.6 33.7 35.3* 35.7*
RT 33.1 33.7 33.9 34.4 34.4 34.1
Ret. (103/μl) EDTA 4° C 216 214 214 217 217 210
RT 215 215 217 219 220 226
Heparin 4° C 213 215 216 223 233 237
RT 212 214 212 217 216 241
IRF (%) EDTA 4° C 37.9 37.9 37.9 38.3 39.5 39.0
RT 40.1 39.7 41.4 39.0 40.1 39.6
Heparin 4° C 37.2 37.4 38.4 38.1 36.9 36.6
RT 38.0 37.3 38.6 38.0 36.6 34.5
NRBC (/μl) EDTA 4° C 960 956 995 1011 1078 1048
RT 973 953 998 901 895 879
Heparin 4° C 868 926 961 943 901 904
RT 878 886 855 856 864 851

RT, room temperature.

*

Absolute value of the difference from the mean EDTA/4°C/undiluted/time 0 h treatment significantly greater than 5% (P < 0.05).

Figure 2.

Figure 2

Mean RBC (a), reticulocytes (b), MCV (c) and platelets (d) vs. time for EDTA − 4° C (Inline graphic), EDTA – room temperature (Inline graphic), heparin − 4° C (Inline graphic) and heparin – room temperature (Inline graphic) – undiluted treatments. Bars represent the standard error (n = 8).

Following 1 : 5 dilution in CELLPACK, there were no statistically significant changes in the hematologic parameters studied which were greater than 5% of mean EDTA/diluted/time 0 h value over the 4 h study period when stored at room temperature (Table 2). As noted above, the only statistically significant effect was that of heparin as the anticoagulant on platelets.

Table 2.

Mean diluted cord blood hematologic parameter results by anticoagulant and duration of storage (n = 8)

Storage period (h)
Variable Anticoagulant 0 0.5 1 2 4
WBC (/μl) EDTA 10846 10896 10908 10898 11020
Heparin 10216 10196 10335 10310 10334
Platelets (103/μl) EDTA 304 306 306 300 311
Heparin 229* 222* 226* 228* 229*
IPF (%) EDTA 4.48 5.04 4.46 4.39 4.73
Heparin 5.40 5.09 4.74 5.05 4.95
Hgb (g/dl) EDTA 15.8 15.8 15.8 15.7 15.8
Heparin 15.6 15.5 15.4 15.5 15.5
RBC (106/μl) EDTA 4.33 4.35 4.35 4.32 4.34
Heparin 4.25 4.24 4.21 4.27 4.24
MCV (fl) EDTA 110 110 110 110 110
Heparin 109 109 108 108 109
MCH (pg) EDTA 36.6 36.4 36.3 36.4 36.3
Heparin 36.7 36.6 36.6 36.4 36.6
Ret He (pg) EDTA 26.1 26.5 26.6 27.7 28.2
Heparin 26.2 26.5 26.9 27.3 28.1
Reticulocytes (103/μl) EDTA 188 196 196 191 201
Heparin 190 188 192 199 196
IFR (%) EDTA 40.8 44.1 41.1 42.3 40.4
Heparin 41.8 42.4 42.6 40.8 44.0
*

Absolute value of the difference from the mean EDTA/diluted/time 0 h treatment significantly greater than 5% (P < 0.05).

DISCUSSION

In this study, we demonstrate that the Sysmex XE-2100 hematology analyzer is capable of reducing physician-ordered laboratory blood loss by nearly an order of magnitude (from 250–500 μl commonly requested in by hospital laboratories to 40 μl while providing hematologic results for the same expansive group of analytes. In using umbilical cord blood at delivery as a surrogate for neonatal blood, we also demonstrated that with few exceptions that anticoagulated whole blood results are unaffected by (i) storage for up to 72 h at 4° C, (ii) storage for up to 24 h room temperature, after which cell swelling occurs as indicated by greater MCV and (iii) the use heparin over EDTA, the recommended anticoagulant for hematologic parameters, except for lower platelet and HPC counts and increases in IPF with heparin. The findings of this study have potentially important clinical and research implications for reducing iatrogenic phlebotomy loss in the critical care setting without sacrificing either the accuracy or the scope of laboratory test results.

As technology advances, the capability of chemistry and hematology analyzers used in clinical settings continues to expand while the sample volumes required by these instruments continue to decrease. In the intensive care setting, an important consequence of these technologic advances is less iatrogenic blood loss. As a result, patients are less prone to develop anemia and require fewer RBC transfusions. Despite these technologic advances, even greater progress is needed, particularly in the NICU setting where iatrogenic blood loss and resultant anemia among VLBW infants remains as the primary reason for RBC transfusions (Lin et al., 2000; Madan et al., 2005; Widness et al., 2005).

For measurement of all hematologic parameters, EDTA is the recommended anticoagulant (Haematology, 1993) Nonetheless, we found that analysis of cord blood immediately after collection using heparin as the anticoagulant had no effect on the primary hematologic parameters other than lower platelet counts (Figure 1). Similar findings have been reported for hematologic parameters in adults (Reinhart et al., 1990; Kawamoto et al., 2000). The effect of heparin on lowering progenitor cell counts and increasing IPF has not been previously reported in adults perhaps because both are relatively recent reported additions of some hematologic analyzers.

While dilution of fresh cord blood with CELLPACK initially affected only reticulocyte count, Ret He and WBC count (Figure 1), hematology parameters did not vary further over 4 h at room temperature following dilution (Table 2). This indicates that anticoagulated whole blood samples do not need to be analyzed immediately following dilution to obtain accurate results. Despite the effects of dilution on these hematology parameters, the ability to analyze hematologic parameters with as little as 40 μl of whole blood when using a 1 : 5 dilution and analysis in capillary mode on the Sysmex XE-2100 instrument offers the advantages of markedly decreasing phlebotomy loss and of allowing capillary blood draws as a reliable alternative to venous sampling (2005). If similar laboratory findings are demonstrated in adults, these features may also be of importance for the elderly in nursing homes and for critically ill, VLBW premature infants, who by virtue of their extremely small size and need for frequent laboratory monitoring, quickly become clinically anemic to the extent that they require frequent RBC transfusions in the early weeks of life (Lin et al., 2000; Madan et al., 2005; Widness et al., 2005). Recent studies by our group in VLBW infants have demonstrated that reducing the volume of blood drawn for laboratory testing may reduce the number of RBC transfusions (Madan et al., 2005; Widness et al., 2005).

In the NICU setting, blood collected for clinical laboratory testing is (i) often drawn in heparin or anticoagulants other than EDTA and (ii) often drawn in volumes in excess of that needed. Hence, the reduced blood volume needed when using dilutions and capillary mode allows for substantially reduced blood volumes to be collected and combined with the stability of the hematologic values over several days (Tables 1 and 2) permits ‘leftover’ anticoagulated blood from other routine clinical tests to be retrieved, diluted and analyzed for clinical and/or research purposes. As in the critical care setting, a majority of samples are collected in heparin, this provides a rich source of data without incurring additional blood loss.

Though most hematologic parameters were stable over 72 h, the MCV increase indicated cell swelling over time when undiluted blood was stored at room temperature; however, it did not change when stored at 4° C (Figure 2c). The temperature-dependent swelling in RBC for both EDTA and heparin is consistent with previous reports in humans (Brittin et al., 1969; Paxton & Bendele, 1993; deBaca et al., 2006; Hedberg & Lehto, 2008) and dogs at room temperature (Medaille, Briend-Marchal & Braun, 2006). Both the undiluted RBC and reticulocyte counts did not substantially change over 72 h regardless of anticoagulant and storage temperature (Table 1 and Figure 2a,b, respectively). The stability of reticulocytes is consistent with other reports using Sysmex instrumentation demonstrating that reticulocytes drawn in EDTA from adults are stable for 48 h when stored at either room temperature or 4° C (Peng et al., 2001; deBaca et al., 2006). Other reports have demonstrated reticulocyte stability between 1 and 3 days when stored at 4° C or 24° C (Perry et al., 1996). Of note, in our study, key clinical hematologic parameters including hemoglobin concentration, mean corpuscular hemoglobin (MCH), RBC distribution width (RDW), mean platelet volume (MPV), IRF and NRBC were also not affected by anticoagulant, storage temperature or storage time. Our results of RDW, MPV, IRF and NRBC are in variance from those of deBaca et al. (2006), who have reported significant changes/variability in these parameters upon storage of EDTA-anticoagulated blood at room temperature for over 24 h. The observed variances may, however, be attributed to the differences in sample size (8 vs. 40) and specimen type (normal vs. normal and abnormal) utilized in the two studies. Recently, Hedberg & Lehto (2008) reported changes in RDW at 48 and 72 h of EDTA samples stored at room temperature (but not in IRF).

Not surprisingly, some of the newer hematologic parameters that have inherently low cell numbers were more variable than the more ‘classical’ parameters. In particular, the IG, HPC and NRBC count were substantially more variable (Figure 1). Accordingly, detecting changes in the mean value using heparin over time required larger percent changes (Tables 1 and 2). Given the high variability of these parameters, caution is required when interpreting change over time, particularly when heparin is used as the anticoagulant.

In conclusion, we have demonstrated that with its capability of utilizing 40 μl anticoagulated whole blood samples diluted 1 : 5 with CELLPACK, the Sysmex XE-2100 hematology analyzer is capable of reducing iatrogenic laboratory blood loss by nearly an order of magnitude (from 250–500 μl to 40 μl). The results provided include nearly the same expansive panel hematologic determinations as determined on undiluted whole blood samples. Using umbilical cord blood at delivery as a surrogate for neonatal blood, we also demonstrated that the use of sodium heparin instead of EDTA as anticoagulant has little effect on the measured hematologic parameters of cord blood other than lowering platelet and HPC counts and increasing IPF. In general, the primary hematologic parameters were stable over 72 h at both room temperature and 4° C storage temperatures, except for swelling of RBCs after 24 h of room temperature storage. However, dilution of cord blood 1 : 5 in CELL-PACK had a significant initial effect on WBC count, reticulocyte count and reticulocyte hemoglobin amount values. Of note, after dilution with CELL-PACK, no further changes in any measured hematology values were observed through 4 h of storage at room temperature. The stability of the hematologic parameters over time and the ability to reduce the blood volume required for hematologic testing has important clinical and research implications for improving patient care by reducing anemia and the need for and the risks associated with RBC transfusions. This is particularly true of VLBW premature infants for whom iatrogenic blood loss remains as the primary cause of symptomatic anemia.

ACKNOWLEDGEMENTS

This work is supported by the United States Public Health Service National Institute of Health Grants P01 HL46925 and Grant M01-RR-59 from the National Center for Research Resources, General Clinical Research Center Program. The Sysmex XE-2100 automatic hematology analyzer used was provided on-loan from Sysmex Corporation, Kobe, Japan.

Abbreviations

EDTA

ethylenediaminetetraacetic acid

Hgb

hemoglobin

HPC

hematopoietic progenitor cells

IG

immature granulocytes

IPF

immature platelet fraction

IRF

immature reticulocyte fraction

MCH

mean corpuscular hemoglobin

MCV

mean corpuscular volume

MPV

mean platelet volume

NICU

neonatal intensive care unit

NRBC

nucleated red blood cell

RBC

red blood cell

RDW

red blood cell distribution width

Ret He

reticulocyte mean corpuscular hemoglobin

VLBW

very low birth weight

WBC

white blood cell

REFERENCES

  1. Akaike H. Automatic control: a new look at the statistical model identification. IEEE Transactions. 1974;19:716–723. [Google Scholar]
  2. deBaca ME, Gulati G, Kocher W, Schwarting R. Effects of storage of blood at room temperature on hemato-logic parameters measured on Sysmex XE-2100. Laboratory Medicine. 2006;37:28–36. [Google Scholar]
  3. Brittin GM, Brecher G, Johnson CA, Elashoff RM. Stability of blood in commonly used anticoagulants. Use of refrigerated blood for quality control of the Coulter Counter Model S. American Journal of Clinical Pathology. 1969;52:690–694. doi: 10.1093/ajcp/52.6.690. [DOI] [PubMed] [Google Scholar]
  4. Haematology I.C.f.S.i. Recommendations of the International Council for Standardization in Haematology for Ethylenediaminetetraacetic Acid Anticoagulation of blood for blood cell counting and sizing. International Council for Standardization in Haematology: Expert Panel on Cytometry. American Journal of Clinical Pathology. 1993;100:371–372. doi: 10.1093/ajcp/100.4.371. [DOI] [PubMed] [Google Scholar]
  5. Hedberg P, Lehto T. Aging stability of complete blood count and white blood cell differential parameters analyzed by Abbott CELL-DYN Sapphire hematology analyzer. International Journal of Laboratory Hematology. 2008 doi: 10.1111/j.1751-553X.2007.01009.x. (in press). doi: 10.1111/j.1751-553X.2007. 01009.x. [DOI] [PubMed] [Google Scholar]
  6. Kawamoto T, Hiino M, Takubo T, Tatsumi N. Low-molecular-weight heparin as a multipurpose anticoagulant for laboratory testing. Osaka City Medical Journal. 2000;46:37–53. [PubMed] [Google Scholar]
  7. Lin JC, Strauss RG, Kulhavy JC, Johnson KJ, Zimmerman MB, Cress GA, Connolly NW, Widness JA. Phlebotomy overdraw in the neonatal intensive care nursery. Pediatrics. 2000;106:E19. doi: 10.1542/peds.106.2.e19. [DOI] [PubMed] [Google Scholar]
  8. Linderkamp O, Versmold HT, Strohhacker I, Messow-Zahn K, Riegel KP, Betke K. Capillary-venous hematocrit differences in newborn infants. I. Relationship to blood volume, peripheral blood flow, and acid-base parameters. European Journal of Pediatrics. 1977;127:9–14. doi: 10.1007/BF00465560. [DOI] [PubMed] [Google Scholar]
  9. Madan A, Kumar R, Adams MM, Benitz WE, Geaghan SM, Widness JA. Reduction in red blood cell transfusions using a bedside analyzer in extremely low birth weight infants. Journal of Perinatology. 2005;25:21–25. doi: 10.1038/sj.jp.7211201. [DOI] [PubMed] [Google Scholar]
  10. Medaille C, Briend-Marchal A, Braun JP. Stability of selected hematology variables in canine blood kept at room temperature in EDTA for 24 and 48 hours. Veterinary Clinical Pathology. 2006;35:18–23. doi: 10.1111/j.1939-165x.2006.tb00083.x. [DOI] [PubMed] [Google Scholar]
  11. Paxton H, Bendele T. Effect of time, temperature, and anticoagulant on flow cytometry and hematological values. Annals of the New York Academy of Sciences. 1993;677:440–443. doi: 10.1111/j.1749-6632.1993.tb38811.x. [DOI] [PubMed] [Google Scholar]
  12. Peng L, Yang H, Jiang H, Su J, Peng Z. Automated reticulocyte counting using the Sysmex RAM-1. Clinical and Laboratory Haematology. 2001;23:97–102. doi: 10.1046/j.1365-2257.2001.00365.x. [DOI] [PubMed] [Google Scholar]
  13. Perry ES, Moore RH, Berger TA, Billups LC, Maybee DA, Salata KF, Lippert LE. In vitro and in vivo persistence of reticulocytes from donor red cells. Transfusion. 1996;36:318–321. doi: 10.1046/j.1537-2995.1996.36496226144.x. [DOI] [PubMed] [Google Scholar]
  14. Reinhart WH, Haeberli A, Stark J, Straub PW. Influence of blood withdrawal and anticoagulant on clotting activity, hematologic data, and certain rheologic measurements. Journal of Laboratory and Clinical Medicine. 1990;115:98–103. [PubMed] [Google Scholar]
  15. Rivera LM. Postnatal persistence of capillary-venous differences in hematocrit and hemoglobin values in low-birth-weight and term infants. Pediatrics. 1982;70:956–957. [PubMed] [Google Scholar]
  16. Sysmex XE-2100 . Operator's Manual Automated Hematology Analyzer XE-2100 Main Unit. Sysmex Corporation; Kobe, Japan: 2005. [Google Scholar]
  17. Widness JA, Madan A, Grindeanu LA, Zimmerman MB, Wong DK, Stevenson DK. Reduction in red blood cell transfusions among preterm infants: results of a randomized trial with an in-line blood gas and chemistry monitor. Pediatrics. 2005;115:1299–1306. doi: 10.1542/peds.2004-1680. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES