Abstract
Unlike other CARS-based spectroscopy techniques, nonlinear interferometric vibrational spectroscopy (NIVS) is linear in analyte concentration and has a Raman lineshape free of non-resonant background distortions. We use spontaneous Raman scattering as a high accuracy benchmark for NIVS. As a challenging comparison, we examine spectra in the CH stretching region of 6 lipid samples. Singular value decomposition and reference to an independent chemical assay are used to directly compare NIVS and spontaneous Raman scattering. We demonstrate that NIVS can determine the relative degree of unsaturation in six different lipid samples as accurately as spontaneous Raman spectroscopy, but 200 times faster. A skin tissue sample is mapped out to demonstrate quantitative lipid-protein differentiation with spatial resolution.
Spontaneous Raman spectroscopy is a convenient method for measuring vibrational spectra with well-defined line shapes and a linear dependence on sample concentration. Visible/NIR excitation allows penentration through many common solvents to access analytes. Thus Raman spectroscopy is a useful quantitative analytical tool. The drawback of spontaneous Raman spectroscopy is the low signal level and long averaging time required to obtain satisfactory signals.
Nonlinear Raman spectroscopies such as Coherent Anti-Stokes Raman Scattering (CARS)1 offer much greater sensitivity by substituting a stimulated process for a spontaneous process. A sequence of pump-Stokes-pump pulses produces an anti-Stokes output pulse whenever a vibrational frequency Ω is in resonance (figure 1 inset). Classical CARS used picosecond narrow bandwidth lasers for both the pump and Stokes pulses,1 producing good spectral resolution but requiring spectral scanning to cover the spectrum. In 2002, Chen and coworkers,2 Xie and coworkers3, and Muller and coworkers4 reported multiplex CARS (mCARS), which uses a broadband Stokes pulse so CARS spectra can be dispersed onto a detector array, parallelizing data collection. In 2004, Saykally and coworkers5 replaced the narrowband pump pulse by a chirped broadband pulse, maintaining high resolution while simplifying pulse generation to a single fs laser source.
The major drawback of these CARS techniques is that they measure only the output intensity |ECARS|2~| χ(3)|2. The consequences are two-fold: the signal depends quadratically on the amount of sample; and the line shape is distorted by non-resonant background in a difficult-to-control manner.6 As a result, CARS intensity-based techniques have been very useful for imaging general molecular content (e.g. lipid content in tissues, membranes or medical samples),7–10 but not for quantitative analysis (e.g. capable of distinguishing small differences between different lipids).
The non-resonant background and linearity problems were solved in 2004 by Boppart and coworkers with the introduction of nonlinear interferometric vibrational spectroscopy (NIVS), and its related imaging technique, nonlinear interferometric vibrational imaging (NIVI).11 Like chirped CARS, NIVS uses a fs Stokes pulse to provide broadband spectral coverage and pulse chirping to derive all pulses from a single fs source while maintaining high resolution. The key to NIVS is the addition of a reference pulse (also obtained from the same fs source), which interacts with the CARS pulse in an interferometer to yield a signal linearly proportional to χ(3).11 NIVS measures both the real and imaginary components of the suscpetibility independently, instead of just measuring its overall magnitude-squared. The NIVS signal is thus linear in concentration and has the simple Raman lineshape suitable for quantitative analysis of mixtures or subtle spectral differences between compounds.
Here we directly compare NIVS with spontaneoous Raman spectroscopy as the ‘gold standard’ by obtaining spectra of six different oil samples with varying degrees of unsaturation. Quantitative decomposition of the NIVS and Raman spectra by singular value decomposition, and calibration with unsaturation data obtained independently by chemical titration of the oil samples, show that NIVS has the same ability as spontaneous Raman spectroscopy to distinguish subtle vibrational spectral features and determine relative degree of unsaturation. NIVS achieves the required signal to noise in less than 100 ms, which is over 200 times faster than Raman spectroscopy on a state-of-the-art Raman microscope. NIVS could thus be suitable for high-throughput applications where rapid and accurate spectral measurements and a quantitative determination of sample content are required. We map protein and lipid content in a skin sample to demonstrate a spectroscopic imaging application.
Experimental and computational section
Oil samples and chemical assay
Six commercially available vegetable oils (coconut, olive, safflower, peanut, corn, soybean) with varying proportions of saturated and unsaturated fatty acids were studied (table 1). The degree of unsaturation was measured by an iodine addition assay to provide data independent of the Raman and NIVS spectroscopies. For each assay, Voil = 0.250 ml of oil was dissolved in 20.0 ml of chloroform. A 25.0 ml volume of Hanus solution (Iodine bromide in glacial acetic acid, Sigma-Aldrich) was then added and kept in the dark for an hour. IBr is consumed by the C=C double bonds of the unsaturated fatty acids in this reaction. Then, 20.0 ml of 10 % KI was added to convert excess IBr into I2, which was titrated against 0.1 M sodium thiosulfate. A 1 ml of 0.1 M sodium thiosulfate is equivalent to 0.05 mmoles of I2. By subtracting the titrant volume VT from the volume Vblank of a blank chloroform sample, the number of moles of IBr consumed by the C=C bonds in the oil is calculated. The number of C=C bonds per ml of the oil is obtained using
Table 1.
Veg. Oil | NC=C/ml I2 assay ×1.44·1021 | Saturated | Oleic 1(C=C) | Linoleic 2(C=C) | α-Linoleic 3(C=C) | NC=C/FA Calc. | NC=C/FA I2 assay |
---|---|---|---|---|---|---|---|
Coconut | 0.18 | 91 | 6 | 2 | - | 0.10 | 0.11 |
Olive | 1.00 | 16 | 71 | 10 | 1 | 0.91 | 0.76 |
Safflower | 1.04 | 7 | 78 | 15 | - | 1.08 | 0.80 |
Peanut | 1.23 | 13 | 48 | 32 | - | 1.12 | 0.95 |
Corn | 1.89 | 13 | 28 | 58 | 1 | 1.47 | 1.43 |
Soybean | 1.92 | 15 | 24 | 54 | 7 | 1.53 | 1.45 |
(1) |
where NA is the Avagadro’s number. We also used the measured density and the average molecular weight of each oil (based on standard fatty acid compositions)12 to obtain the average number of C=C bonds per fatty acid in the oil.
Raman spectroscopy
The Raman spectra of the oils were acquired on a commercial confocal Raman microscope (Senterra, Bruker) using 532 nm excitation at 10 mW and an exposure time of 20 seconds per complete spectrum. Sample films were sandwiched between a slide and coverslip separated by a 125 μm thick spacer. Raman signal was collected in a back scattering geometry using an objective with a focal volume of ~2 μm in the lateral and ~6 μm in the axial directions. Spectral processing involves the correction for detector spectral sensitivity, removal of cosmic ray spikes, and baseline subtraction by polynomial fitting.
Nonlinear interferometric vibrational spectroscopy (NIVS)
The detailed theory of CARS and NIVS is discussed in the literature.13–16 For analytical purposes, NIVS differs from CARS in two key aspects: it produces an output signal that is linearly proportional to the amount of sample, and interferometric phase reconstruction leads to reproducible lineshapes identical to spontaneous Raman lineshapes. We discuss these aspects and our instrumentation in more detail here.
For a Raman pump field envelope EP(ω) and Stokes field envelope ES(ω) (Fig. 1 inset), the generated CARS output field is given by11, 14
(2) |
χ(3)(Ω) is the third order susceptibility of the substance studied at vibrational frequency Ω. The term in parentheses is the vibrational coherence created in the analyte by the pump and Stokes fields. Good spectral coverage requires a broadband Stokes field, and good spectral resolution usually entails the use of a narrowband pump field. If the ultrashort Stokes field is approximated by a delta function that gates the pump field at t = 0 when ω′ = ω0, the vibrational coherence simplifies to
(3) |
If in addition the temporal envelope EP (t) of the pump field in eq. (3) is constant while the vibrational coherence decays (i.e. ), eq. (2) reduces to
(4) |
The two conditions required for eq. (4) to hold can be satisfied as discussed in the instrumentation subsection.
Experiments usually detect the CARS intensity |ECARS (ω)|2, but this produces a distorted lineshape from off-resonant contributions to the signal, as well as a quadratic concentration dependence. A signal with a Raman lineshape (imaginary χ(3)) and a linear concentration dependence can be obtained if both the CARS amplitude and phase are extracted from the CARS output field by Fourier transform spectral interferometry,17 resulting in Nonlinear Interferometric Vibrational Spectrocopy (NIVS).
NIVS works by reconstructing the full complex CARS field from a real-valued spectral interferogram. To create the interferogram, the CARS output field ECARS (ω) is mixed with a transform limited reference pulse Eref in a Mach-Zehnder interferometer (Fig. 1). The resulting spectral interferogram I(ω) (figure 2A) contains DC components from the CARS and reference fields, and the phase-sensitive cross term that will yield the NIVS signal:
(5) |
τ is the time delay between the CARS and reference pulses, essential for the reconstruction of the complex CARS field. The |Eref |2 term can be rejected by background subtraction of the known reference pulse. τ permits the rejection of the |ECARS|2 term by temporally separating the DC and phase sensitive components as follows. Letting , we can express the inverse Fourier transform of eq. (5) as (figure 2B)
(6) |
The first term is a slowly varying DC component symmetric about t = 0 while the last two terms are time reversed from each other (eq. (5) is real). Because of causality, there is no signal prior to the Stokes pulse initiating vibrational coherence. After shifting the time origin to be the center of the Stokes pulse (figure 2C), one can impose χ(t)=0 for t < −T, where T is greater than the Stokes pulse width. If the delay τ of the Mach-Zehnder interferometer is chosen larger than T, the two time-reversed terms in eq. (6) do not overlap with each other or with the DC component (figure 2B). The function χ(t−τ) thus obtained by shifting and truncation is Fourier-transformed to recover χ(ω) ∝ χ(3)(Ω)*ES (ω0 − Ω)Eref (ω)*. Both χ(ω) and χ(3)(Ω) are linear in analyte concentration, but χ(ω) is weighted by the known field envelopes ES and Eref.
In the absence of electronic resonances at ω, the non-resonant CARS background is real and can be rejected by discarding the real part of χ or χ(3). The NIV spectrum is given by the imaginary part, and matches the Raman spectrum, except for Raman depolarization effects.
Instrumentation
The use of narrowband pump and broadband Stokes pulses is typically required to attain good spectral resolution and broad spectral range simultaneously. However, NIVS achieves this by deriving the pump and Stokes pulses from the same ultrafast broadband source (figure 1), but chirping the pump pulse so ω′ = ω0 + bt in eq. (2). The spectral components of EP(ω′) are spread out in time, and time-gating by the ultrashort Stokes pulse at t = 0 ensures that only EP with ω′ = ω0 contributes to the the generation of vibrational coherence, thus retaining the narrowband spectral resolution.11, 15 The chirp of the pump pulse causes a chirp of the CARS output field in eq (4), which must be computationally removed (step C to D in figure 2). The chirped pump field is described mathematically as
(7) |
in the time domain.18 The constant a determines the Gaussian pulse envelope and b determines the chirp rate. Multiplying eq. (4) by the conjugate phase factor e+ibt2 removes the chirp in the CARS field and restores one-to-one correspondence between the anti-Stokes and vibrational frequencies. As dicussed above eq. (4), a must be small enough so |EP(t)| remains nearly constant while the vibrational coherence decays.
Figure 1 shows a schematic of our NIVS experimental setup, which is based on a 250 kHz regeneratively amplified oscillator. A modelocked Ti:Sapphire laser (MIRA, 82 MHz) seeds the regenerative amplifier (Coherent, RegA-9000, 250 kHz) which produces microjoule pulses. 10% of this output was used as the Raman pump, and linearly chirped to ~ 6 ps by passing it through an 85 cm block of BK7 glass. The remaining 90% of the output was used to pump an optical parametric amplifier (Coherent, OPA-9450), which produced the Stokes (idler) and reference (signal) pulses.
The pump and the Stokes fields were focused onto the sample by a microscopic objective filled to NA 0.3 (numerical aperture). A slide and coverslip with a 125 μm spacer identical to those used in the Raman experiment were used for direct comparison. The focal volume was ~2 μm in lateral and ~9 μm in axial directions. The nonlinear interaction of these fields with the sample generates the higher frequency CARS signal. The forward directed CARS signal was collected by an objective and was spectrally filtered from the pump, Stokes, and sample fluorescence, which all lie at longer wavelengths. The spectrally filtered CARS beam was then made collinear with the reference beam at a 50:50 cube beamsplitter. The CARS and the reference beams were spatially filtered at a pinhole to imporve beam quality for robust interference. The mixed beams were then dispersed by a grating (1200 groves/mm) to yield the spectral interferogram, which was recorded by a line camera (DALSA P2, 2048 pixels).
The colors of the pump and Stokes pulses were tuned to target the C-H vibrational spectral range of 2800 – 3100 cm−1. About 10 mW of pump and 1 mW of Stokes power were used and the NIV spectra were acquired at a line rate of 1 kHz. The acquisition time for the NIV spectra reported here was 100 ms to achieve the same signal to noise ratio as the Raman spectrum (500:1 for the peak at 2857 cm−1 relative to the rms fluctuations of the baseline). We used 0.3 NA (numerical aperture) objectives for beam focusing and signal collection throughout. The pump, Stokes, and the reference fields were parallel-polarized.
Calibration and processing
The CARS spectrum from a reference sample (acetone, or silicone as in figure 2) was monitored on a spectrometer (Ocean Optics QE65000), inline with the experimental setup, to calibrate the angular dispersion of wavelength on the line camera. The reference power spectrum was subtracted from the calibrated spectral interferogram to discard the bulk of the DC component. The wavelengths were then converted to Raman shifts by referencing to an approximate instantaneous pump frequency ω0 (figure 2A).
The interferogram was then inverse Fourier transformed to obtain the time domain polarization (figure 2B). The residual DC and the negative time components were zeroed and the time origin was shifted to the peak of the CARS signal. The polarization then needed to be multiplied by the conjugate phase ibt2 to correct for the pump-induced chirp. The chirp rate b was obtained by minimizing the width of the vibrational resonance of acetone at 2925 cm−1. This procedure was iterated to optimize the values of ω0 and b in such a way that the peak was observed at 2925 cm−1 with minimum width. The position and width of the 2907 cm−1 resonance of silicone can also be used for the same purpose.
The chirp-corrected polarization (figure 2C) was then Fourier transformed to yield χ(ω) (figure 2D). The analysis presented in Results does not require field envelope normalization, which introduces additional noise. Rather than dividing χ(ω) by ES and Eref to compare with the Raman lineshape, we multiplied the Raman lineshape by ES and Eref. Of course the NIVS signal-to-noise ratio decreases at the edges of the field envelopes, where less laser power was available to excite the sample.
The NIV spectrum must be corrected for the dispersion in the Stokes beam due to the transmissive optics in the beam path. This dispersion essentially leads to a linear scaling of the vibrational spectrum due to the interaction of the chirped pump frequencies (within the narrow bandwidth around ω0) with the chirped Stokes frequencies (within the broad bandwidth). We used the silicone vibrational peaks at 2907 and 2965 cm−1 to correct for this scaling.
Results
As an assay independent of Raman and NIVS, table 1 lists the number of C=C bonds per ml (NC=C/ml) in each oil obtained by iodine titration. Both NIVS and Raman data were calibrated against these absolute unsaturation values as detailed below. Table 1 also shows the average number of C=C bonds per fatty acid (NC=C/FA) obtained by combining the iodine assay with a measurement of the oil densities and average molecular masses based on chain length distribution.12 The measured NC=C/FA agrees with typical literature values based on fatty acid composition within the range of sample-to-sample variations expected of such oils.
Our main goal is to assess whether NIVS can determine the relative degree of unsaturation of different oils with the same relative accuracy as Raman spectroscopy. Using the setup shown in figure 1, we obtained NIV spectra of six thin film samples of cooking oils (table 1) in the challenging C-H stretching spectral range (2730–3000 cm−1), where methyl, methylene and =CH- stretching bands overlap. The polarized CARS output field interfered with a reference field in a Mach-Zehnder interferometer, allowing reconstruction of the NIVS signal. NIVS has the same χ(3) lineshape as Raman spectroscopy and is linear in the analyte volume. For comparison, Raman spectra were acquired over the same spectral range on a Raman microscope. Quantitative results can be obtained by Raman spectroscopy in this spectral range, but not by chirped CARS, multiplex CARS, or similar techniques that detect only the anti-Stokes intensity, and thus suffer from non-resonant line shape distortion and a nonlinear dependence of the signal on the sample concentration.
Figure 3 compares the isotropic Raman and NIV spectra. For direct comparison with the NIVS signal, the Raman spectra were multiplied by the Stokes and reference field envelopes. All spectra were normalized to the same peak height at 2856 cm−1. The C-H stretching spectral features of typical unsaturated fatty acids19 (listed in Table 2) in cooking oils are readily observed in the NIV spectra. For comparable signal to noise ratios (500:1) the NIV spectra were acquired 200 times faster than the Raman spectra: 100 ms vs. 20 s.
Table 2.
Methyl oleate | Methyl linoleate | Methyl α-linoletae | Vibrational assignments |
---|---|---|---|
3003 | 3012 | 3015 | =C-H stretch |
2960 | 2951 | 2951 | C-H stretch (-CH3 asym) |
2929 | 2924 | 2935 | C-H stretch (-CH3 sym) |
2897 | 2903 | 2911 | C-H stretch (-CH3 sym) |
2876 | 2877 | 2880 | C-H stretch (-CH2 asym) |
2856 | 2855 | 2855 | C-H stretch (-CH2 sym) |
2734 | 2732 | 2729 |
The relative intensity variation of the two main peaks in the isotropic Raman and polarized NIV spectra is due to the inherent polarization sensitivity of NIVS. The linearly polarized reference field acts analogous to an analyzer in polarization-CARS experiments.3, 20 Our reference field was polarized parallel to the pump and Stokes polarization, corresponding to an analyzer set for the polarized signal. Since the symmetric CH2 streching mode signal (2856 cm−1, depolarization ~1/5) is better aligned with the reference than the asymmetric C-H2 mode signal (2885 cm−1, depolarization ~3/4),3 the peak at 2856 cm−1 appears relatively more intense than in the isotropic Raman spectra.
The intensity of the peak at 3010 cm−1 relative to the main saturated CH stretching peaks at 2856 and 2885 cm−1 represents the degree of unsaturation (number of C=C double bonds) in the cooking oils (see table 2). The Raman signal intensity of different oils in this region can be explained quantitatively by their relative degree of unsaturation. We want to assess whether NIVS can determine the degree of unsaturation as accurately as Raman spectroscopy.
To verify the quantifiable information content of the NIV spectra we applied singular value decomposition (SVD)21 to the six NIV spectra and separately to the six Raman spectra. SVD is a linear decomposition technique and can be applied because both signals are linear in analyte volume. The purpose of SVD is to construct a set of orthonormal basis spectra Bi (Ω) such that each observed spectrum S(Ω) can be reconstructed exactly as
(8) |
with a different coefficient vector C for each spectrum. Not all of the basis functions are correlated with the degree of unsaturation, and we can find an optimal parameter
(9) |
to compare the predictions made by NIVS and Raman spectroscopy for the degree of relative unsaturation. Larger values of ai weight the basis function Bi more.
Only two basis functions (B2 and B4) made a significant contribution to the unsaturated peak at 3010 cm−1 (Figure 4A). Figure 4B shows the coefficients C2 and C4 of the different oils obtained from NIVS. A linear combination U = a2C2 + a 4C4 (dotted line in figure 4B) that best represents the degree of unsaturation from table 1 (NC=C/ml) was then obtained by least squares fitting of the coefficients a2 and a4. Since the NIV spectra provide only the relative degree of unsaturation, we scaled U to match the absolute NC=C/ml obtained for olive oil from the iodine assay (table 1).
The Raman spectra were subjected to the same analysis. Figure 5 shows a plot of both the NIVS (circles) and Raman (triangles) unsaturation NC=C/ml against that obtained from the iodine assay. Both spectroscopies produce excellent linear fits with slopes that are virtually identical. The linearity of the fits confirms the linear quantifiability of NIV spectra, which is as good as that of Raman spectra within the uncertainities of the iodine assay. The NIVS measurement is over 200 times faster than the spontaneous Raman measurement.
To demonstrate lipid/protein differentiation by NIVS in an image (NIVI), we scanned a 50 μm thick section of pig tissue at the dermis/subcutaneous fat layer boundary. Figure 6 shows the total resonant Raman signal in a 500×500 μm sample (< 5 μm resolution), as well as NIVS spectral images color-coded by lipid content and protein content (as a fraction of maximum content observed) based on characteristic Raman peaks at 2855 and 2935 cm−1.
Discussion
Many CARS-related techniques have been developed to improve the spectral fidelity and imaging capability of the technique. For example, frequency modulation CARS, heterodyne CARS, and interferometric CARS, use ps pulses good for imaging, but without the spectral range needed for chemometric analysis.6, 16 To obtain broadband coverage, Xie and coworkers combined mCARS with polarization detection to suppress the nonresonant background better. This approach definitely improves the resonant vibrational contrast and is excellent for imaging of general lipid content.3 However, a residual background remains in the mCARS spectra when compared to the spontaneous Raman spectra22 and the signal is not linear in population. In contrast, NIVS achieves complete background rejection, verified by comparison to Raman spectra, as well as quantifiable linearity (Fig. 5). The use of polarization selection to reject background also applies to NIVS, as the nonresonant background has a depolarization of 1/3 and polarization selection is inherent in the anti-Stokes-reference beam interaction with parallel polarization. Further optimization of the reference polarization could be useful for detecting weak transitions with various polarization ratios.
Computational tools have been developed to extract the Raman lineshape from CARS intensities, such as the maximum entropy method.23 These tools can greatly improve ordinary CARS lineshapes, but still have to make assumptions, such as the imaginary part of χ(3) being much smaller than the off-resonant background, yet far removed from resonance. It is currently not clear how such procedures affect spectra when signal-to-noise ratios in excess of 100 are desired. Similar computational ‘clean-up’ may turn out to be useful to improve the lineshape further even on NIVS data.
Of the nonlinear Raman spectroscopy tools developed in the last decade, only femtosecond stimulated Raman scattering (FSRS)24 can be compared with NIVS as an analytical tool. In FSRS, a white-light continuum provides broad spectral coverage, and a picosecond narrowband pump pulse achieves a reported spectral resolution of 10 cm−1, comparable to the resolution reported here. The signal is linear in sample concentration. The subtraction required to produce Raman lineshapes from FSRS can create small negative artifacts, but the lineshapes are far superior to intensity-detected CARS lineshapes. A quantitative comparison of signal-to-noise ratios (S/N) is not possible, but for neat cyclohexane (similar density of CH2 groups and Raman cross sections compared to neat oil) and a pathlength of 1 cm, S/N ~300:1 were achieved in 100 ms (100 pulses).24 This can be compared with a S/N of 500:1 achieved by NIVS with a 0.0125 cm pathlength in 100 ms. Thus NIVS appears to perform better in that regard. The authors of ref. 24 note that saturation of the S/N in FSRS should not occur if difference spectra are measured instead of the absolute spectra recorded there, and of course their technique has the advantage of very high time resolution when lower S/N is sufficient.
NIVS-based imaging (NIVI) could be a useful tool for analytical imaging applications. For example, Krafft et al.25 discuss the need for combining Raman spectroscopy spectral quality and CARS microscopy imaging for various bioanalytical and biomedical applications. They had to implement Raman spectroscopy and CARS microscopy separately. NIVS based imaging combines the best of these two techniques, allowing fast imaging with spectral pixels that are linear in analyte and have reproducible lineshapes. For lipid studies specifically, such quantification would be useful.7, 8 Instead of just identifying lipids in tissues, spectral signatures of unsaturation, such as from omega-3 fatty acids, could be studied. There has been some interest in the effects of omega-3 fatty acids on cancer cell lipid metabolism.26
Along the lines of the skin tissue scan shown in Figure 6, we recently developed chemometric imaging useful for the pathology of cancer tissue using intensity analysis of NIV imaging instead of the SVD analysis outlined here.27 The underlying idea is that carcinomas have a very different ratio of lipid to protein content, which can be readily quantified if accurate lineshapes can be measured. Adaptation of SVD-decomposed NIV imaging to cancer diagnosis could allow small (sub-0.1 mm) cancerous incursions to be identified in thin sections (<0.1 mm) of healthy tissue, judging from the S/N and high linearity of the signal obtained in our oil study presented here.
NIVS could be further improved by increasing its bandwidth, or by allowing the interferometry to be recorded in a single shot to increase time resolution. A number of schemes have been developed in these directions.28–30
Acknowledgments
This work was supported by a grant from the National Institutes of Health, National Cancer Institute, R21/R33 CA115536 (S. B. and M. G.). Z. J. was supported by the Beckman Fellows Program. M. G. and P. C. were additionally supported by the James R. Eiszner Chair.
References
- 1.Shen YR. Principles of Nonlinear Optics. Wiley Interscience; New York: 1984. [Google Scholar]
- 2.Chen PC, Joyner CC, Patrick ST, Benton KF. Anal Chem. 2002;74:1618–1623. doi: 10.1021/ac011129+. [DOI] [PubMed] [Google Scholar]
- 3.Cheng JX, Volkmer A, Book LD, Xie XS. J Phys Chem B. 2002;106:8493–8498. [Google Scholar]
- 4.Muller M, Schins JM. J Phys Chem B. 2002;106:3715–3723. [Google Scholar]
- 5.Knutsen KP, Johnson JC, Miller AE, Petersen PB, Saykally RJ. Chem Phys Lett. 2004;387:436–441. [Google Scholar]
- 6.Evans CL, Xie XS. Ann Rev Anal Chem. 2008;1:883–909. doi: 10.1146/annurev.anchem.1.031207.112754. [DOI] [PubMed] [Google Scholar]
- 7.Wang HW, Fu Y, Huff TB, Le TT, Wang HF, Cheng JX. Vibr Spectrosc. 2009;50:160–167. doi: 10.1016/j.vibspec.2008.11.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Le TT, Huff TB, Cheng JX. BMC Cancer. 2009:9. doi: 10.1186/1471-2407-9-42. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Rinia HA, Bonn M, Muller M, Vartiainen EM. ChemPhysChem. 2007;8:279–287. doi: 10.1002/cphc.200600481. [DOI] [PubMed] [Google Scholar]
- 10.Windbergs M, Jurna M, Offerhaus HL, Herek JL, Kleinebudde P, Strachan CJ. Anal Chem. 2009;81:2085–2091. doi: 10.1021/ac8020856. [DOI] [PubMed] [Google Scholar]
- 11.Jones GW, Marks DL, Vinegoni C, Boppart SA. Opt Lett. 2006;31:1543–1545. doi: 10.1364/ol.31.001543. [DOI] [PubMed] [Google Scholar]
- 12.Hilditch TP. Brit J Nutr. 1949;3:347–354. [Google Scholar]
- 13.Oron D, Dudovich N, Silberberg Y. Phys Rev Lett. 2002:89. doi: 10.1103/PhysRevLett.89.273001. [DOI] [PubMed] [Google Scholar]
- 14.Marks DL, Boppart SA. Phys Rev Lett. 2004:92. doi: 10.1103/PhysRevLett.92.123905. [DOI] [PubMed] [Google Scholar]
- 15.Knutsen KP, Messer BM, Onorato RM, Saykally RJ. J Phys Chem B. 2006;110:5854–5864. doi: 10.1021/jp052416a. [DOI] [PubMed] [Google Scholar]
- 16.Cheng JX, Xie XS. J Phys Chem B. 2004;108:827–840. [Google Scholar]
- 17.Lepetit L, Cheriaux G, Joffre M. Journal of the Optical Society of America B-Optical Physics. 1995;12:2467–2474. [Google Scholar]
- 18.Siegman AE. Lasers. University Science Books; New York: 1986. [Google Scholar]
- 19.Frank CJ, Redd DCB, Gansler TS, McCreery RL. Anal Chem. 1994;66:319–326. doi: 10.1021/ac00075a002. [DOI] [PubMed] [Google Scholar]
- 20.Cheng JX, Book LD, Xie XS. Opt Lett. 2001;26:1341–1343. doi: 10.1364/ol.26.001341. [DOI] [PubMed] [Google Scholar]
- 21.Press WH, Teukolsky SA, Vetterling WT, Flannery BP. Numerical Recipes: The Art of Scientific Computing. 3. Cambridge University Press; New York: 2007. [Google Scholar]
- 22.Volkmer A. J Physics D. 2005;38:R59–R81. [Google Scholar]
- 23.Vartiainen EM, Rinia HA, Muller M, Bonn M. Opt Express. 2006;14:3622–3630. doi: 10.1364/oe.14.003622. [DOI] [PubMed] [Google Scholar]
- 24.McCamant DW, Kukura P, Yoon S, Mathies RA. Rev Sci Instr. 2004;75:4971–4980. doi: 10.1063/1.1807566. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Krafft C, Dietzek B, Popp J. Analyst. 2009;134:1046–1057. doi: 10.1039/b822354h. [DOI] [PubMed] [Google Scholar]
- 26.Freudiger CW, Min W, Saar BG, Lu S, Holtom GR, He CW, Tsai JC, Kang JX, Xie XS. Science. 2008;322:1857–1861. doi: 10.1126/science.1165758. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Benalcazar WA, Chowdary PD, Jiang Z, Marks DL, Chaney EJ, Gruebele M, Boppart SA. IEEE J Sel Top Quant Elec. 2010 doi: 10.1109/JSTQE.2009.2035537. [DOI] [Google Scholar]
- 28.Lee YJ, Cicerone MT. Opt Express. 2009;17:123–135. doi: 10.1364/oe.17.000123. [DOI] [PubMed] [Google Scholar]
- 29.Lim SH, Caster AG, Leone SR. Opt Lett. 2007;32:1332–1334. doi: 10.1364/ol.32.001332. [DOI] [PubMed] [Google Scholar]
- 30.Kee TW, Zhao HX, Cicerone MT. Opt Express. 2006;14:3631–3640. doi: 10.1364/oe.14.003631. [DOI] [PMC free article] [PubMed] [Google Scholar]