DISC VARIANCE OF THE HARKEN MITRAL PROSTHESIS: THE CONTRIBUTION OF ASSOCIATED AORTIC REGURGITATION

Michael D Yarnoz 1, David Hueter 1, John R McCormick 1, Harrison Black 1, Robert L Berger 1
PMCID: PMC287682  PMID: 15216088

Abstract

Four cases of severe mitral regurgitation due to disc variance of the Harken disc prosthesis in the mitral position are described.

The valve occluder actually escaped into the left atrium in two patients, and neither survived despite emergency valve replacement. In the other two, disc malfunction was identified by flouroscopy, the prosthesis was replaced, and both patients survived. All four patients had associated aortic regurgitation, which most likely contributed to erosion of the disc edges. It is suggested that patients with the Harken disc prosthesis undergo periodic evaluation to detect abnormal disc motion.

Full text

PDF
383

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Beall A. C., Jr, Morris G. C., Jr, Noon G. P., Guinn G. A., Reul G. J., Jr, Lefrak E. A., Greenberg S. D. An improved mitral valve prosthesis. Ann Thorac Surg. 1973 Jan;15(1):25–34. doi: 10.1016/s0003-4975(10)64930-2. [DOI] [PubMed] [Google Scholar]
  2. Berger K., Sauvage L. R., Wood S. J. Long-range alterations in prosthetic valves. Cardiovasc Clin. 1973;5(1):185–205. [PubMed] [Google Scholar]
  3. Berger R. L., Saini V. K. A continuous suture technique for mitral valve replacement. Ann Thorac Surg. 1973 Nov;16(5):505–508. doi: 10.1016/s0003-4975(10)65025-4. [DOI] [PubMed] [Google Scholar]
  4. Berroya R. B., Escano F. B., Jr Mitral disc-valve variance. Thorax. 1972 Jan;27(1):87–89. doi: 10.1136/thx.27.1.87. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Björk V. O., Bök K., Holmgren A. Significance of position and opening angle of the Björk-Shiley tilting disc valve in mitral surgery. Scand J Thorac Cardiovasc Surg. 1973;7(3):187–201. doi: 10.3109/14017437309135562. [DOI] [PubMed] [Google Scholar]
  6. Braunwald N. S., Detmer D. E. A critical analysis of the status of prosthetic valves and homografts. Prog Cardiovasc Dis. 1968 Sep;11(2):113–132. doi: 10.1016/0033-0620(68)90022-4. [DOI] [PubMed] [Google Scholar]
  7. Carey J. S., Hughes R. K. A complication of mitral-valve replacement with caged-lens prosthesis. Ann Thorac Surg. 1968 Jul;6(1):77–81. doi: 10.1016/s0003-4975(10)65993-0. [DOI] [PubMed] [Google Scholar]
  8. Harken D. E., Matloff J. M., Zuckerman W., Chaux A. A new mitral valve. J Thorac Cardiovasc Surg. 1968 Mar;55(3):369–382. [PubMed] [Google Scholar]
  9. Oparah S. S., Keefe J. F., Ryan T. J., Berger R. L. Mitral valve replacement with a turtleneck-disc prosthesis. J Thorac Cardiovasc Surg. 1975 Apr;69(4):568–574. [PubMed] [Google Scholar]
  10. Ridolfi R. L., Hutchins G. M. Detection of ball variance in prosthetic heart valves by liver biopsy. Johns Hopkins Med J. 1974 Mar;134(3):131–140. [PubMed] [Google Scholar]
  11. Roberts W. C., Bulkley B. H., Morrow A. G. Pathologic anatomy of cardiac valve replacement: a study of 224 necropsy patients. Prog Cardiovasc Dis. 1973 May-Jun;15(6):539–587. doi: 10.1016/s0033-0620(73)80024-6. [DOI] [PubMed] [Google Scholar]
  12. Sharma S. D., Easley R. M., Jr, Zaroff L. I., Goldstein S. Mitral disc variance (Harken prosthesis). Am Heart J. 1973 Nov;86(5):681–686. doi: 10.1016/0002-8703(73)90347-5. [DOI] [PubMed] [Google Scholar]
  13. Shaw T. R., Gunstensen J., Turner R. W. Sudden mechanical malfunction of Hammersmith mitral valve prostheses due to wear of polypropylene. J Thorac Cardiovasc Surg. 1974 Apr;67(4):579–583. [PubMed] [Google Scholar]
  14. Stinson E. B., Griepp R. B., Shumway N. E. Clinical experience with a porcine aortic valve xenograft for mitral valve replacement. Ann Thorac Surg. 1974 Oct;18(4):391–401. doi: 10.1016/s0003-4975(10)64376-7. [DOI] [PubMed] [Google Scholar]
  15. Vasko J. S., Leighton R. F. Acute massive mitral regurgitation resulting from disc-valve malfunction. Ann Thorac Surg. 1968 Dec;6(6):564–569. doi: 10.1016/s0003-4975(10)66124-3. [DOI] [PubMed] [Google Scholar]

Articles from Cardiovascular Diseases are provided here courtesy of Texas Heart Institute