Abstract
An abdominal left ventricular assist device (ALVAD) is undergoing controlled clinical trials in our institution. The ALVAD is pneumatically-actuated, synchronously or asynchronously with an external console and is interposed between the apex of the left ventricle and the infrarenal abdominal aorta. It is an order of magnitude more effective than conventional intraaortic balloon pumping. Thus far, we have implanted this pump in 21 patients (15 males and six females). The average age has been 50. The duration of cardiopulmonary bypass with intensive pharmacologic support and IABP until ALVAD implantation has been nearly 4 hours. The plasma hemoglobins prior to ALVAD implantation have averaged 168 mg%. The platelet counts at implantation have averaged 68,000 mm3. The average duration of ALVAD support has been in excess of one day and the longest trial extended for one week. We have been able to remove the pump after ventricular recovery in two instances and effected cardiac allografting in one instance of ALVAD dependency. We have found that (1) the profoundly depressed left (and right) ventricles can recover if totally supported with this device; (2) the device can function in the presence of ventricular fibrillation and/or standstill; (3) the device can effectively replace both left and right ventricular function in the presence of normal pulmonary vascular resistance and microcirculatory hemodynamics; and (4) in the presence of impending multiple organ failure, procrastination in use is to be avoided.
Full text
PDF













Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Brewer M. A., Igo S. R., Holub D. A., Hibbs C. W., Fuqua J. M., Edmonds C. H., Trono R., Norman J. C. Algorithm for evaluation of postcardiotomy or infarction low output syndromes during mechanical circulatory support (IABP/ALVAD). Trans Am Soc Artif Intern Organs. 1978;24:288–297. [PubMed] [Google Scholar]
- Naifeh J. G., Thompson P. A., Johnson M. D., Hibbs C. W., Fuqua J. M., Norman J. C. Performance of an abdominal left ventricular assist device during induced tachycardias and dysrhythmias. J Thorac Cardiovasc Surg. 1976 Aug;72(2):175–181. [PubMed] [Google Scholar]
- Norman J. C., Dacso C. C., Reul G. J., Massin E. K., Klima T., Kahan B. D., Keiser J. T., Cooley D. A. Partial artificial heart (ALVAD) use with subsequent cardiac and renal allografting in a patient with stone heart syndrome. Artif Organs. 1978 Nov;2(4):413–420. doi: 10.1111/j.1525-1594.1978.tb01632.x. [DOI] [PubMed] [Google Scholar]
- Norman J. C., Fuqua J. M., Hibbs C. W., Edmonds C. H., Igo S. R., Cooley D. A. An intracorporeal (abdominal) left ventricular assist device. Initial clinical trials. Arch Surg. 1977 Dec;112(12):1442–1451. doi: 10.1001/archsurg.1977.01370120032003. [DOI] [PubMed] [Google Scholar]
- Norman J. C. Intracorporeal partial artificial hearts: initial clinical trials. Heart Lung. 1978 Sep-Oct;7(5):788–803. [PubMed] [Google Scholar]
- Norman J. C. Partial artificial hearts: mechanical cloning of the ventricle. Artif Organs. 1978 Aug;2(3):235–243. doi: 10.1111/j.1525-1594.1978.tb03459.x. [DOI] [PubMed] [Google Scholar]
- Norman John C., Bernhard William F. CRITERIA, PROTOCOLS AND REPORTING FORMS FOR INITIAL LEFT VENTRICULAR ASSIST DEVICE CLINICAL TRIALS. Cardiovasc Dis. 1975;2(4):438–445. [PMC free article] [PubMed] [Google Scholar]
- Robinson W. J., Benedict B. S., Daly D. T., Hughes D. A., Migliore J. J., Hibbs C. W., Igo S. R., Norman J. C. An abdominal left ventricular assist device: preclinical studies. Ann Thorac Surg. 1975 May;19(5):540–551. doi: 10.1016/s0003-4975(10)64430-x. [DOI] [PubMed] [Google Scholar]
- Turina M., Boslo R., Senning A. Clinical application of paracorporeal uni- and biventricular artificial heart. Trans Am Soc Artif Intern Organs. 1978;24:625–631. [PubMed] [Google Scholar]