Trimming of a Broken Migrated Biliary Metal Stent with the Nd:YAG Laser

I Zuber-Jerger 1,*, F Kullmann 1
PMCID: PMC2895170  PMID: 20651959

Abstract

Biliary metal stents are a permanent solution for bile duct stenosis. Complications can arise when the stent migrates, breaks or is overgrown by tumour. The following case demonstrates how a Nd:YAG laser can be used to solve these problems. A 93-year-old man presented with jaundice and fever. Two years earlier a 6-cm metal stent had been implanted into a postinflammatory stenosis of the common bile duct after recurrent cholangitis and repetitive plastic stenting. Duodenoscopy showed that the stent was broken. It had migrated about 3 cm into the duodenum, leading to kinking of the stent and breaking of the wires. The stent was also occluded. It was necessary to purge the common bile duct and to introduce a second stent. However, the only way to reach the papilla was through the broken wires. Placing a second stent this way was impossible. Thus we trimmed the stent with a Nd:YAG laser. The piece that had migrated into the duodenum was retrieved. Now the papilla could be reached. The rest of the metal stent was purged with NaCl 0.9%. A second metal stent was placed. Since an Nd:YAG laser is part of the equipment of most endoscopy units, it can be used to trim a broken or migrated biliary metal stent.

Key Words: Broken migrated biliary metal stent, Nd:YAG laser, Trimming

Background

Stenosis of the common bile duct [1,2, 3], the pancreatic duct [4] or the oesophagus [5, 6] can be resolved by the implantation of a metal stent. Gastrointestinal metal stents have proven to be a long-lasting solution in situations where surgery is impossible or denied. Complications arising from metal stents are migration, breaking and closure by bolus or tumour ingrowth. The most common complication is tumour ingrowth, which arises in nearly every second patient with oesophagus carcinoma and oesophageal metal stent [7]. This complication can sometimes be avoided by the implantation of a covered metal stent [8], and if occurring it can usually be solved by tumour debridement with laser and/or placement of a second stent [7, 8]. In the biliary tree the problem of tumour ingrowth into a metal stent is usually solved by the placement of a second covered stent.

Breaking and migration of metal stents in the oesophagus and common bile duct can lead to serious clinical problems, such as as complete closure of the bile duct leading to cholestasis, or of the duodenum or oesophagus leading to either ileus or complete inability of food intake. Gastrointestinal haemorrhage or ulceration or perforation due to duodenal erosion by the stent can arise [9,10,11]. The problem of migration and breaking of a stent can sometimes be solved by trimming the stent with laser and taking it out [12, 13]. We present the case of a patient with a broken migrated biliary metal stent leading to cholestasis and incomplete closure of the duodenum that could be solved by trimming the stent with an Nd:YAG laser.

Case Presentation

A 93-year-old man presented to the emergency room with fever and jaundice. Two years earlier an uncovered Zilver metal stent (length 6 cm, diameter 10 mm) had been implanted into a postinflammatory stenosis of the common bile duct because of recurrent cholangitis. Plastic stents had been occluded every second week and had thus not been able to avoid cholangitis.

At presentation the patient complained of chills and abdominal pain. Clinical examination revealed jaundice and painful palpation of the epigastrium. Fever rose up to 39°C. Laboratory tests showed elevated bilirubin (2.6 mg/dl) and gamma-GT (1,340 U/l), neutrophilia (82.5%) and an elevated CRP (66.48 mg/dl). Duodenoscopy revealed that the stent was totally occluded with detritus and had migrated into the duodenum about 3 cm. This migration had led to a kinking of the stent and the wires of the stent were broken at the curve (fig. 1). Thus, the broken biliary metal stent occluded the common bile duct completely and the duodenum partly. The papilla was inaccessible due to the stent. Ulceration of the duodenal wall with gastrointestinal bleeding or perforation had to be expected. Retrieving the stent seemed unavoidable, but we feared complications such as bleeding due to mucosa maceration and perhaps even perforation of the bile duct when forcefully retrieving a metal stent that was securely impacted in the bile duct after two years in place.

Fig. 1.

Fig. 1

Two years after implantation the metal stent is completely occluded with detritus and has migrated 3 cm into the duodenum. The migration has led to a kinking of the stent, resulting in breaking of the wires (arrow).

We thus tried to trim the stent at the papilla with an Nd:YAG laser with 308 J and 50 W. We managed to cut the wires exactly at the papilla. The piece of stent cut off by the laser rested in the duodenum after relaxation of the small intestine with butylscopolamine (fig. 2) and was taken out with a snare. The common bile duct was purged through the stent with NaCl 0.9% and a second metal stent (Ultraflex uncovered, length 8 cm, diameter 10 mm) was placed (fig. 3). Injection of contrast dye through a standard catheter showed that the biliary tree was free of detritus and stones and that the stream of bile was adequate. The patient was discharged symptom-free. No biliary complications have arisen in the last two years.

Fig. 2.

Fig. 2

The part of the metal stent that was cut off rests in the duodenum.

Fig. 3.

Fig. 3

Endoscopic view on the papilla with the correctly placed second metal stent.

Conclusion

A Nd:YAG laser is part of the equipment of most endoscopy units. It is used predominantly for malignant oesophageal stenosis [7, 8]. It can also be used to trim broken or migrated metal stents [12, 13], avoiding clinical complications such as bleeding and perforation and allowing, like in our case, further stenting of the common bile duct to resolve cholestasis.

References

  • 1.Srikureya W, Chang KJ. Endoscopic palliation of pancreatic adenocarcinoma. Curr Opin Gastroenterol. 2005;21:601–605. doi: 10.1097/01.mog.0000172750.05280.02. [DOI] [PubMed] [Google Scholar]
  • 2.Brugge WR. Endoscopic techniques to diagnose and manage biliary tumours. J Clin Oncol. 2005;23:4561–4565. doi: 10.1200/JCO.2005.19.729. [DOI] [PubMed] [Google Scholar]
  • 3.Siriwardana HP, Siriwardana AK. Systematic appraisal of the role of metallic endobiliary stents in the treatment of benign bile duct structure. Ann Surg. 2005;242:10–19. doi: 10.1097/01.sla.0000167761.70021.4d. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Eisendrath P, Deviere J. Expandable metal stents for benign pancreatic duct obstruction. Am J Gastroenterol. 1999;9:547–554. [PubMed] [Google Scholar]
  • 5.Kozarek RA. Endoscopic palliation of esophageal malignancy. Endoscopy. 2005;35:S9–S13. doi: 10.1055/s-2003-41538. [DOI] [PubMed] [Google Scholar]
  • 6.Von Rahden BH, Stein HJ. Staging and treatment of advanced esophageal cancer. Curr Opin Gastroenterol. 2005;21:472–477. [PubMed] [Google Scholar]
  • 7.Sargeant IR, Loizou LA, Tulloch M, Thorpe S, Brown SG. Recanalization of tube overgrowth: a useful new indication for laser in palliation of malignant dysphagia. Gastrointest Endosc. 1992;38:165–169. doi: 10.1016/s0016-5107(92)70383-6. [DOI] [PubMed] [Google Scholar]
  • 8.McManus K, Khan I, McGuigan J. Self-expanding oesophageal stents: strategies for re-intervention. Endoscopy. 2001;33:601–604. doi: 10.1055/s-2001-15312. [DOI] [PubMed] [Google Scholar]
  • 9.Isayama H, Komatsu Y, Tsujino T, Sasahira N, Hirano K, Toda N, Nakai Y, Yamamoto N, Tada M, Yoshida H, Shiratori Y, Kawabe T, Omata M. A prospective randomized study of ‘covered’ versus ‘uncovered’ diamond stents for the management of distal malignant biliary obstruction. Gut. 2004;53:729–734. doi: 10.1136/gut.2003.018945. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Marano BJ, Bonanno CA. Metallic biliary endoprosthesis causing duodenal perforation and acute upper gastrointestinal bleeding. Gastrointest Endosc. 1994;40:257–258. doi: 10.1016/s0016-5107(94)70188-1. [DOI] [PubMed] [Google Scholar]
  • 11.Ee H, Laurence BH. Haemorrhage due to erosion of a metal biliary stent through the duodenal wall. Endoscopy. 1992;24:431–432. doi: 10.1055/s-2007-1010514. [DOI] [PubMed] [Google Scholar]
  • 12.Shaver CP, Brady P, Pinkas H. You don't have to retrieve to relieve: How and when to trim a self-expanding metal biliary stent. Endoscopy. 2004;36:833. doi: 10.1055/s-2004-825836. [DOI] [PubMed] [Google Scholar]
  • 13.Vanbierbliet G, Piche T, Caroli-Bose FX, Dumas R, Peten EP, Huet PM, Tran A, Demarquay JF. Endoscopic argon plasma trimming of biliary and gastrointestinal metallic stents. Endoscopy. 2005;37:434–438. doi: 10.1055/s-2005-860989. [DOI] [PubMed] [Google Scholar]

Articles from Case Reports in Gastroenterology are provided here courtesy of Karger Publishers

RESOURCES