Skip to main content
Journal of Diabetes Science and Technology logoLink to Journal of Diabetes Science and Technology
. 2010 May 1;4(3):505–513. doi: 10.1177/193229681000400302

Evolution of Diabetes Insulin Delivery Devices

Jean-Louis Selam 1
PMCID: PMC2901025  PMID: 20513314

Abstract

The first manufactured insulin pump was introduced in the 1970s and the first insulin pens in 1985; since then, many improvements have been made to both devices. The advantages of pens over syringes have been confirmed in numerous studies and include greater accuracy, ease of use, patient satisfaction, quality of life, and adherence. United States claims database analyses indicate that the improved adherence made possible by use of an insulin pen has the potential to reduce diabetes care costs when compared with using a vial and syringe. Features of certain advanced pump models include the ability to connect wirelessly to a blood glucose meter or to a subcutaneous interstitial glucose sensor for semicontinuous glucose-driven insulin rate adjustment. A new trend in the design of insulin pumps is the tubing-free patch pump that adheres directly to the skin. The low rate of insulin pen usage in the United States compared with European countries and the fact that many patients report that they are not offered the option of an insulin pen by their physician suggest that there is a need to increase patient and provider awareness of the currently available devices for insulin administration.

Keywords: continuous subcutaneous insulin infusion, diabetes, insulin delivery, insulin pen, insulin pump

Introduction

The publication of the results of the landmark Diabetes Control and Complications Trial (DCCT) in 1993 clearly demonstrated the need for intensified methods of blood glucose (BG) control in type 1 diabetes to prevent complications such as retinopathy, nephropathy, and neuropathy.1 Five years later, the importance of intensive glycemic control to prevent microvascular complications in type 2 diabetes was shown by the United Kingdom Prospective Diabetes Study (UKPDS).2 However, the need for more convenient, safer, and more effective methods of insulin administration had been apparent long before the DCCT and UKPDS results were published.3 When insulin was first discovered in the early 1920s, the method of delivery used large glass syringes and reusable needles, both of which needed sterilization by boiling after each use. Needles were sharpened with a pumice stone so they could be reused. For over 50 years, vial and syringe remained the only delivery option available for routine clinical use. The first manufactured insulin pump was introduced in the 1970s, while the first manufactured insulin pen, the NovoPen® (Novo Nordisk), was introduced in 1985.4 Since then, many improvements and innovations have been made to both insulin pumps and pen devices. Furthermore, insulin analogs have become available that enable both continuous subcutaneous insulin infusion (CSII) using an insulin pump and insulin therapy using multiple daily injections (MDI) to more closely match physiologic insulin patterns.58

For various reasons that are unrelated to the scientific evidence base, the rate of adoption of insulin pens and pumps has differed greatly between the United States and Europe. Insulin pumps are more widely used on the American side of the Atlantic than on the European side, whereas insulin pens are used as an alternative to syringes by the majority of diabetes patients in Europe but by only approximately 15% of diabetes patients in the United States.9,10 The faster development of insulin pumps in the United States may be due to the fact that the United States was the country where the first manufactured insulin pump was invented.10 Additionally, the publication of the DCCT results greatly contributed to the rapid growth of CSII use in the United States, because almost half of the DCCT patients in the intensive treatment arm had been treated with CSII. As initial instruction for use of CSII takes considerably longer than that for an insulin pen, the use of CSII in the United States may also be facilitated by the availability of certified diabetes educators, who have the time and expertise to educate patients in the correct use of this technology.

Both insulin pens and insulin pumps can offer benefits to patients, including the potential for improved clinical outcomes. However, in a survey of 600 patients using insulin for the treatment of type 2 diabetes in the United States, many patients reported that they had not been offered the option of an insulin pen by their physician.11 Together with the low rate of insulin pen use in the United States compared with European countries, this suggests that there is a need to increase provider awareness of the benefits and limitations of the currently available devices for insulin administration in type 2 diabetes so that patients are informed of the range of options available and are thus able to choose the device that best suits their individual circumstances. Therefore, this article reviews the benefits and limitations of insulin pens and pumps in the treatment of diabetes.

Methods

This review is based on a literature search of the PubMed database using the following search strategy: “(diabetes or insulin or insulins) and (pen or pens or pump or pumps or CSII or continuous subcutaneous insulin infusion)”. Health economic papers were identified by adding the search term “cost or economic.” Searches were limited to articles published in English between January 1, 1985, and September 29, 2009. Priority was given to meta-analyses, systematic reviews, practice guidelines, and controlled clinical trials. Additional articles were identified from the reference lists of review articles.

Benefits and Limitations of Insulin Pens Versus Vial and Syringe

Insulin injection using vial and syringe delivery has the potential for several problems, including the inconvenience of carrying several materials and preparing the syringe, the adverse psychological and social impact of using a syringe (because syringes are associated with sickness and drug abuse), use of the incorrect insulin product, and failure to administer accurate doses. The development of insulin pens has therefore focused on ways to counter such problems. Several disposable and reusable pen devices have been developed that provide options for delivering rapid- and long-acting insulins and insulin premixes. Table 1 lists the pen devices that are currently available in the United States. The advantages of insulin pens over syringes have been confirmed in numerous studies.1222 These advantages, which include greater accuracy, convenience, patient preference, and adherence, are discussed here.

Table 1.

Insulin Pen Delivery Devices Available in the United Statesa

Refillable pens (manufacturer) Prefilled disposable pens (manufacturer)
Autopen® 24 (Owen Mumford) FlexPen (Novo Nordisk)
Autopen Classic AN3800 (Owen Mumford) Humalog® KwikPen™ (Eli Lilly)
Autopen Classic AN3810 (Owen Mumford) Humalog Pen (Eli Lilly)
HumaPen LUXURA HD (Eli Lilly) SoloSTAR® (sanofi-aventis)
HumaPen MEMOIR (Eli Lilly)
NovoPen 3 (Novo Nordisk)
NovoPen 4 (Novo Nordisk)
NovoPen Junior (Novo Nordisk)
OptiClik® (sanofi-aventis)
a

Compatible insulin analogs for pen devices: Novo Nordisk = insulin detemir, insulin aspart, and biphasic insulin aspart 70/30; Eli Lilly = insulin lispro, insulin lispro mix 75/25, and insulin lispro mix 50/50; and sanofi-aventis = insulin glargine, insulin glulisine. The Autopen Classic takes Eli Lilly insulin cartridges, and the Autopen 24 takes sanofi-aventis insulin cartridges. The NovoPen models use 3 ml PenFill® cartridges.

Accuracy, Ease of Use, and Patient Preference

In a study of syringes and pens used by children with type 1 diabetes, pens were more accurate than syringes in measuring out insulin at low insulin doses (<5 U).20 At doses above 5 U, pens and syringes had similar accuracies. In another study, pens were found to be more accurate than syringes at doses of 1 and 2 U.16 In a survey of 507 insulin users, 89% of 479 respondents (not all patients answered all survey questions) considered an insulin pen to be more socially acceptable than a vial and syringe; 86% of 475 respondents indicated that a pen was easier to use; and 86% of 488 respondents said that it took less time to prepare and administer injections with a pen.14 Similar responses were found in a survey of nurses in a community hospital after implementation of insulin pen devices.22 The majority of nurses stated that insulin pens were more convenient than vials/syringes. In addition, implementation of insulin pen devices did not increase the nurses' time spent to teach patients to self-inject insulin and did not increase insulin-related needle stick injuries.

Korytkowski and colleagues17 assessed patient preference for an insulin pen versus vial and syringe in a randomized, open-label, crossover study in 121 adults with type 1 or type 2 diabetes. Patients were randomized to use either a prefilled pen or vial/syringe to administer an insulin analog premix regimen for four weeks, followed by four weeks' use of the other injection device. In summary, 74% of patients indicated a preference for the pen over the vial/syringe (compared with 20% who preferred the vial/syringe), 85% considered the pen more discreet for use in public (compared with 9% for the vial/syringe), 74% considered it easier to use overall (compared with 21% for the vial/syringe), and 85% found the insulin dose scale on the pen easier to read (compared with 10% for the vial/syringe). The quality-of-life benefits of insulin pens compared with syringes have also been confirmed in other studies using generic quality-of-life scales.18,21

Adherence

Adherence to the appropriate insulin therapy is a major element of good glycemic control, and there is evidence that insulin pens can improve patient adherence compared with vial and syringe delivery.12,19 Lee and associates19 analyzed U.S. managed care claims data for 1156 subjects with type 2 diabetes. This study found that medication adherence (measured by the medication possession ratio [MPR]) significantly improved from 62% to 69% (p < .01) after conversion from regular human or analog insulin injection using a vial and syringe to a prefilled insulin analog pen (containing either insulin aspart or biphasic insulin aspart 70/30). In a similar study by Cobden and coworkers12 of 486 subjects who switched from vial and syringe to an insulin pen prefilled with biphasic insulin aspart 70/30, the MPR increased from 59% to 68% (p < .01). However, it should be noted that, although the MPR is a well-established measure of adherence, it is not possible to confirm with claims data that patients are correctly or accurately administering their drugs, and it is also not possible to include factors such as drug sharing or wastage.

Health Economics of Insulin Pens

Insulin analogs supplied in cartridges or prefilled pens have a higher per unit of insulin cost than do insulin analogs supplied in vials. For example, one vial (1000 U) of insulin glulisine costs $105.95, which equates to a cost of 10.6 cents per unit of insulin. Five prefilled insulin pens containing insulin glulisine (total of 1500 U) have a total cost of $201.01, equating to a cost of 13.4 cents per unit of insulin (26% more than the cost of insulin glulisine supplied in a vial). (Prices are the retail prices available to consumers at www.drugstore.com as of March 18, 2010. These prices are without health insurance coverage. Co-pays for pens and vials are similar for most health insurance plans.) However, most pen devices now have good formulary coverage, so cost should not be a limitation to the patient or physician. Data from the two studies that analyzed U.S. managed care claims data indicate that the improved adherence made possible by use of an insulin pen has the potential to reduce diabetes care costs (not including the cost of insulin) when compared with vial/syringe delivery, despite higher prescription costs for pen delivery.12,19

In the study by Lee and colleagues,19 in addition to improved medication adherence in patients who converted from vial/syringe therapy to a prefilled insulin analog pen, the likelihood of experiencing a hypoglycemic event significantly decreased after conversion (odds ratio [OR] = 0.50; 95% confidence interval [CI], 0.37–0.68; p < .05). There were also significant decreases in hypoglycemia-attributable emergency department visits (OR = 0.44; 95% CI, 0.21–0.92; p < .05) and physician visits (OR = 0.39; 95% CI, 0.24–0.64; p < .05). Total mean all-cause annual treatment costs were reduced by $1590 per patient (from $16,359 to $14,769; p < .01). Annual hypoglycemia-attributable costs were reduced by $788 per patient (from $1415 to $627; p < .01), predominantly as a result of decreased hospitalization costs (from $857 to $288; p < .01). Annual diabetes-attributable costs were reduced by $600 per patient (from $8827 to $8227; p < .01). There were similar findings in the study by Cobden and associates,12 with significant decreases observed in the likelihood of hypoglycemic events and in treatment costs after conversion to a prefilled pen containing an insulin analog premix.

Another study assessed patients with type 2 diabetes enrolled in the North Carolina Medicaid program and found that initiating insulin therapy with an insulin pen was associated with significant reductions in health care resource utilization and associated costs compared with starting insulin therapy using a vial and syringe.23 In this study, diabetes-related medication adherence was comparable with the two delivery methods, with an adherence rate of 53% for patients initiating insulin with a pen compared with a rate of 50% in patients using a syringe. However, total annualized health care costs were significantly lower for patients using an insulin pen than for those using a syringe ($14,857.42 versus $31,764.78; p < .05). Cost reductions with pen therapy compared with vial/syringe use were seen in hospital costs ($1195.93 versus $4965.31; p < .05), diabetes-related costs ($7324.37 versus $13,762.21; p < .05), and outpatient costs ($7795.98 versus $13,103.51; p < .05).

Glycated Hemoglobin

Although two studies have reported that switching from vial/syringes to prefilled insulin analog pens improved adherence as measured by the MPR, no rigorous, controlled studies to date have shown that insulin pen use is associated with greater reductions in glycated hemoglobin (A1C) as compared with vial and syringe use. One small study in 23 homeless patients found that switching from vial and syringe to a reusable insulin pen improved glycemic control at 3 and 6 months.24 In a study in 72 patients with type 1 diabetes who switched from vial and syringe injections to four or five injections per day with an insulin pen, glycemic control improved at follow-up (9–13 months after the switch) only in those patients who has previously been receiving one or two injections per day.25 When these patients were followed up for a further five years, metabolic control was found to deteriorate over time.26 However, the lack of a control group in these studies means that the effects of the natural history of the disease and of regression to the mean cannot be excluded.

As mentioned earlier (in the Accuracy, Ease of Use, and Patient Preference section), in the randomized, open-label, crossover study conducted by Korytkowski et al.,17 patients with type 1 or type 2 diabetes were randomized to use either a prefilled pen or vial/syringe to administer biphasic insulin aspart 70/30 for four weeks, followed by four weeks' use of the other injection device. No statistically significant differences were found between the two devices in mean fasting plasma glucose, serum fructosamine, or four-point glucose profile.

Other Refinements

Over the past 20 years, insulin pens have been constantly refined, with certain newer models offering advantages over older ones. For example, the latest improved FlexPen® (Novo Nordisk) requires a lower injection force while maintaining dose accuracy when compared with the older, original FlexPen.27 Another example is the inclusion of a memory function in the HumaPen® MEMOIR™ device (Eli Lilly), which records the date, time, and amount of the previous 16 doses (including priming doses), so that patients and healthcare providers can see exactly how much insulin the patient last took and when. Finer needles and safety needles that are associated with reduced pain perception have also been developed for use with insulin pens.28,29 Disposable prefilled pens (which many patients find more convenient than the reusable cartridge-type pens) are now available for all insulin analogs. Many current insulin pen models also allow backward dialing to correct misdialed doses without wasting insulin. Two models allow the dose to be adjusted in half-unit increments (HumaPen LUXURA™ HD [Eli Lilly] and NovoPen Junior [Novo Nordisk]).

Limitations Versus Vial and Syringe

Apart from their higher prescription cost, the main limitation of pens compared with syringes is the inability for patients to mix their own insulin formulations (i.e., neutral protamine Hagedorn insulin mixed with regular insulin). However, three different premixed biphasic insulin analogs are available for use in prefilled and reusable pens. Furthermore, the mixing of insulin preparations is known to be highly inaccurate when performed by elderly patients.30

Benefits and Limitations of Insulin Pumps

Improvements in insulin pump technology are also having an impact in providing an alternative option for insulin delivery in patients failing to achieve glycemic control using a MDI regimen and in other selected patients. The brick-sized devices of decades past have been replaced by small pumps no bigger than a pager. Modern external insulin pumps weigh less than 4 oz and consist of an insulin reservoir, a small battery-operated pump, and a computerized control mechanism. Pumps deliver a continuous infusion of insulin (usually a rapid-acting insulin analog) via a cannula that is placed subcutaneously. Pumps are programmed to deliver both basal and bolus doses. Premeal or snack bolus doses can be selected to cover the user's estimated carbohydrate intake at mealtime and to correct for out-of-range BG readings. All pumps have occlusion and near-empty alarms. Pumps are also supplied with multiple basal delivery profiles that allow the patient to select different basal infusion rates based on differences in daily or weekly schedules. For example, a patient might require a different basal pattern on weekdays compared with weekends, or a schoolchild might need to adjust if the school day involves sporting activities.

Table 2 provides an overview of the features of currently available insulin pumps. Many advanced models (e.g., OneTouch® Ping™, OmniPod®, and MiniMed Paradigm®) can connect wirelessly to BG meters. The MiniMed Paradigm can also connect wirelessly to a disposable sub-cutaneous interstitial-glucose sensor for semicontinuous glucose-driven insulin rate adjustment; this system is currently the only integrated pump and continuous glucose monitoring system available. Currently available continuous interstitial glucose monitoring systems are not as accurate as current home glucose meters but are useful for providing patients with the ability to monitor changes in glucose levels between finger stick readings.

Table 2.

Key Features of Insulin Pumps Available in the United Statesa

Insulin pump model Accu-Chek® Spirit OneTouch Ping MiniMed Paradigm 522/722 DANA Diabecare IIS OmniPod Insulin Management System Nipro Amigo®
Manufacturer Disetronic Medical Systems AG Animas Corporation Medtronics, Inc. Sooil Development Insulet Nipro Diabetes Systems
Basal programs Five profiles with 24-hourly basal rates each; temporary basal rate in 10% increments from 0% to 200%, and 15 min increments from 15 min to 24 h 12 basal rates in four personalized programs; temporary rate from 30 min to 24 h in 30 min intervals or 10% increments Three profiles with up to 48 rates each Four profiles with 24 rates per profile; temporary basal rate in 25% increments ±100% Seven profiles with 24 rates each Four profiles with 48 rates available per profile; temporary basal rate in 10% increments from 10% to 200% or 15 min increments from 15 min to 24 h
Basal range 0.1–25 U/h in 0.1 U increments 0.025–25 U/h in 0.025 U increments 0.05–35 U/h 0.00–16 U/h 0.05–30 U/h in 0.05 U increments 0–30 U/h in 0.05 U increments
Smallest bolus 0.1 U 0.05 U 0.05 U 0.1 U 0.05 U 0.05 U
Overdelivery alarm No Yes Yes, self-tests and safeguards help prevent overdelivery Yes, internal cross checks No, safety systems monitor delivery and perform safety checks on pod and PDM Yes, internal processors continually monitor pump function to prevent overinfusion and underinfusion
Reservoir size 315 U 200 U 176 or 300 U 300 U 200 U 300 U
Display features Reversible display; backlit display Color screen Backlight Backlight; energy-saving sleep mode Color screen on PDM controller Backlight
Connection Standard luer-lock Standard luer-lock Proprietary Proprietary Integrated infusion set with no tubing required Standard luer-lock
Waterproof IPX8 (60 min at 2.5 m) Waterproof (up to 12 ft for 24 h) Splash resistant IPX8 IPX8 (30 min at 8 ft) IPX8 (35 min at 1 m)
Additional features Standard, advanced, or custom selectable user menus; side-mounted tactile buttons; audible or vibrating bolus confirmation and alerts;supports infrared wireless data transfer Includes a meter/remote that works wirelessly with the pump; audible or vibrating pump alerts; integrated food database Interacts wirelessly with continuous glucose monitor as part of the MiniMed Paradigm REAL-Time System; optional remote control at additional cost; audible or vibrating alerts Icon-based interface Tube-free, disposable system device applied directly to body with adhesive; uses wireless PDM for managing insulin delivery;integrated food database; built-in BG meter in PDM Pump casing is shatter resistant; audible or vibrating alerts and button feedback
a

The Solo MicroPump (Medingo, Ltd.) has received FDA approval but, as of the time of this writing, is not yet available for sale. PDM, personal diabetes manager

A new trend in the design of insulin pumps is the tubing-free “patch” pump. The only currently available patch pump is the OmniPod. The OmniPod pump/reservoir unit adheres directly to the skin and contains an integrated infusion set and automated inserter. The pump/reservoir unit communicates wirelessly with a separate controller that includes an integrated BG meter. Benefits of this patch pump design that have been reported by patients include the ability to wear the pump in the shower and the greater convenience of a tubing-free system. In one small study, 90% of patients (18 of 20) preferred using the OmniPod's automated cannula insertion system versus inserting with their current infusion sets.31 Use of a patch pump may be particularly beneficial in adolescents, as 52% of 48 adolescents in one study reported that they disconnected their (conventional design) pump for exercise.32

Like the OmniPod, the OneTouch Ping also comes with a separate wireless controller that includes an integrated BG meter and integrated food database for bolus calculations. The OneTouch Ping, which uses a conventional (i.e., nonpatch) pump design, can be controlled from both the pump itself as well as from the wireless controller.

It is expected that more patch pumps will come onto the market in the future. The Solo™ MicroPump (Medingo, Ltd.) is a patch pump that has already received U.S. Food and Drug Administration (FDA) approval but, as of the time of this writing, is not yet available for sale. Another likely future advance in the development of CSII technology is the development of more accurate continuous glucose monitoring systems for use in combination with insulin pumps.

Continuous Subcutaneous Insulin Infusion in Type 1 Diabetes

Among patients with type 1 diabetes, the principal indications for CSII include patients who are unable to achieve acceptable glycemic control using MDI, patients with histories of frequent or severe hypoglycemia, and patients who need more intensive management because of microvascular complications.33,34 Since the introduction of long-acting insulin analogs, the “dawn phenomenon” has become a less frequent indication for CSII.34 However, pump therapy is not only costly, but requires a high level of motivation and commitment to diabetes self-management, with frequent checks of BG levels throughout the day, a responsibility that not all patients with diabetes are willing or able to undertake. In addition, some patients, particularly adolescents, may be self-conscious about being attached to a foreign object.35

When used in CSII, rapid-acting insulin analogs have been shown to produce a modest but significantly greater reduction in A1C compared with regular human insulin and are preferred by patients.5 Table 3 provides an overview of the insulins approved for pump therapy, including the maximum time allowed in the insulin reservoir.

Table 3.

Types of Insulin Used in Pump Therapy

Insulin U.S. brand name (manufacturer) Approved age groups in the United States Maximum time allowed in reservoir
Rapid-acting insulin analogs
Insulin aspart NovoLog® (Novo Nordisk) Children and adults Six days
Insulin glulisine Apidra® (sanofi-aventis) Children and adults 48 h
Insulin lispro Humalog (Eli Lilly) Children and adults 48 h
Regular human insulin Humulin® R (Eli Lilly) Children and adults 48 h
Novolin® R (Novo Nordisk) Children and adults 48 h

The benefits of providing continuous delivery of a rapid-acting insulin analog may be substantial for selected patients.36 Compared with MDI, the potential advantages of insulin pump therapy in type 1 diabetes include a lower A1C, a reduced total daily insulin dose, a reduced risk of hypoglycemia, lower BG variability, elimination of the need for daily injections, and increased flexibility in meal timing and size.8,3739 A meta-analysis of 11 randomized controlled trials (RCTs) comparing CSII (using rapid-acting insulin analogs) with MDI in type 1 diabetes found that CSII was associated with a significantly lower A1C compared with MDI (standardized difference in mean: 0.3 percentage points in favor of CSII; 95% CI, 0.1–0.4; p < .001).40 No significant difference was observed in the rate of severe hypoglycemia. All 11 RCTs included in this meta-analysis enrolled patients failing on MDI who were randomized to continue with the same MDI regimen or switch to CSII. The results of this meta-analysis therefore support the principal indication of CSII as being patients unable to achieve acceptable glycemic control using MDI.

Another meta-analysis included three RCTs that compared CSII and optimized MDI therapy using rapid-acting analogs in adults with type 1 diabetes.41 The pooled estimated A1C reduction with CSII compared with MDI was 0.35 percentage points in favor of CSII (95% CI, −0.10 to 0.80; p = .08). There was no significant difference between CSII and MDI in the rate of hypoglycemic events. Importantly, a greater relative benefit of CSII was observed in patients with higher baseline A1C, suggesting that CSII may be particularly beneficial in patients with the poorest initial glycemic control.42

Continuous Subcutaneous Insulin Infusion in Type 2 Diabetes

Another meta-analysis assessed CSII versus MDI in type 2 diabetes. This meta-analysis included four RCTs that were of at least 12 weeks' duration and found that CSII did not produce any significant improvement of A1C compared with MDI (standardized difference in mean: 0.09 percentage points; 95% CI, −0.08 to 0.26; p = .31).43 Current evidence thus shows no clear benefits of CSII over MDI in the general type 2 diabetes population. Further research is required to investigate whether CSII may be useful in specific groups of type 2 diabetes patients, such as patients with marked insulin resistance; after failure of other intensified insulin regimens; during preconception, pregnancy, and lactation; following transplantation; and in cases of insulin allergy.44

Complications of Insulin Pump Therapy

Insulin pumps may undermedicate or overmedicate if they malfunction or are used improperly. Device problems that have been reported to the FDA include alarm problems, loosening and/or occlusion of the catheters, bent cannula, and screen display problems.45 Potential complications of CSII therapy therefore include diabetic ketoacidosis (DKA) and hypoglycemia.46 However, more reliable pumps and improved patient education have greatly reduced these risks. As with MDI therapy, DKA should be preventable through the use of published DKA prevention guidelines that recommend frequent monitoring of urine or serum ketones and BG, with appropriate intervention when ill.47

While infusion-site infections are uncommon, irritation or inflammation at the infusion site are common complications of using an insulin pump,48 though their incidence has been reduced by the introduction of more modern infusion sets (for example, sets that use a Teflon cannula) and by better patient education. Adherence with the advised infusion site preparation and cannula insertion techniques, and with the recommended site duration and site rotation schedule, may minimize dermatologic complications.49

Health Economics of Continuous Subcutaneous Insulin Infusion

To date, only one cost-effectiveness analysis comparing CSII with MDI in patients with diabetes in the United States has been published.50 A previously validated health economic model (the CORE Diabetes Model) was used to determine the incremental cost-effectiveness ratio of CSII compared with MDI using published clinical and cost data. The primary input variable was change in A1C and was assumed to be an improvement of −0.9% in children/young adults and −1.2% in adults for CSII compared with MDI. A series of Markov constructs simulated the progression of diabetes-related complications. The time horizon for the simulation was set to 60 years to capture the remainder of a type 1 diabetes patient's lifetime. Continuous subcutaneous insulin infusion was associated with an improvement in quality-adjusted life years (QALYs) gained of 1.061 versus MDI for adults and 0.799 versus MDI for children/young adults. The incremental cost-effectiveness ratio for CSII versus MDI in adults was $16,992 per QALY gained and in children/young adults was $27,195 per QALY gained. Assuming a cost-effectiveness threshold of $50,000 per QALY gained, CSII is thus estimated to be a cost-effective option for U.S. patients with type 1 diabetes.

Conclusions

Important advances have been made both in the ability to mimic physiologic insulin secretion using insulin analogs tailored to this purpose and in devices to administer these insulins. The advantages of pens over syringes include greater accuracy and ease of use and improved patient satisfaction, quality of life, and adherence. United States claims database analyses indicate that the improved adherence made possible by use of an insulin pen has the potential to reduce diabetes care costs when compared with using a vial and syringe. A cost-effectiveness analysis using a Markov model has estimated that CSII is a cost-effective option for U.S. patients with type 1 diabetes, with an incremental cost-effectiveness ratio for CSII versus MDI of $16,992 per QALY gained in adults and $27,195 per QALY gained in children/young adults. However, even with the most sophisticated insulin delivery devices, BG control still falls short of normality for most patients.51 The development of more physiologic routes of insulin administration, use of an artificial pancreas, or pancreas/beta-cell transplantation are therefore important ongoing areas of research. Moreover, the low rate of insulin pen usage in the United States compared with European countries and the fact that many patients report that they are not offered the option of an insulin pen by their physician suggest that there is a need to increase patient and provider awareness of the latest devices for insulin administration to enable the maximum number of patients to benefit from the important developments of the past three decades.

Acknowledgment

Writing and editorial support was provided by Janet Stephenson and Reza Sayeed of Bioscript Stirling Ltd.

Abbreviations

A1C

glycated hemoglobin

BG

blood glucose

CI

confidence interval

CSII

continuous subcutaneous insulin infusion

DCCT

Diabetes Control and Complications Trial

DKA

diabetic ketoacidosis

FDA

Food and Drug Administration

MDI

multiple daily injections

MPR

medication possession ratio

OR

odds ratio

QALY

quality-adjusted life year

RCT

randomized controlled trial

UKPDS

United Kingdom Prospective Diabetes Study

References

  • 1.The Diabetes Control Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993;329(14):977–986. doi: 10.1056/NEJM199309303291401. [DOI] [PubMed] [Google Scholar]
  • 2.UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33) Lancet. 1998;352(9131):837–853. [PubMed] [Google Scholar]
  • 3.Selam JL, Charles MA. Devices for insulin administration. Diabetes Care. 1990;13(9):955–979. doi: 10.2337/diacare.13.9.955. [DOI] [PubMed] [Google Scholar]
  • 4.Rex J, Jensen KH, Lawton SA. A review of 20 years' experience with the NovoPen family of insulin injection devices. Clin Drug Investig. 2006;26(7):367–401. doi: 10.2165/00044011-200626070-00001. [DOI] [PubMed] [Google Scholar]
  • 5.Colquitt J, Royle P, Waugh N. Are analogue insulins better than soluble in continuous subcutaneous insulin infusion? Results of a meta-analysis. Diabet Med. 2003;20(10):863–866. doi: 10.1046/j.1464-5491.2003.01018.x. [DOI] [PubMed] [Google Scholar]
  • 6.Radermecker RP, Scheen AJ. Continuous subcutaneous insulin infusion with short-acting insulin analogues or human regular insulefficacy, safety, quality of life, and cost-effectiveness. Diabetes Metab Res Rev. 2004;20(3):178–188. doi: 10.1002/dmrr.447. [DOI] [PubMed] [Google Scholar]
  • 7.Sheldon B, Russell-Jones D, Wright J. Insulin analogues: an example of applied medical science. Diabetes Obes Metab. 2009;11(1):5–19. doi: 10.1111/j.1463-1326.2008.01015.x. [DOI] [PubMed] [Google Scholar]
  • 8.Bruttomesso D, Filippi A, Costa S. Is there still a place for CSII in the treatment of type 1 diabetes? Comparing CSII with MDI after the arrival of insulin analogues. Infusystems Int. 2008;7(4):25–28. [Google Scholar]
  • 9.Marcus A. Diabetes care – insulin delivery in a changing world. Medscape J Med. 2008;10(5):120. [PMC free article] [PubMed] [Google Scholar]
  • 10.Selam JL. CSII in Europe: an analysis of articles published in Infusystems International. Infusystems Int. 2007;6(2):14–16. [Google Scholar]
  • 11.Rubin RR, Peyrot M. Factors affecting use of insulin pens by patients with type 2 diabetes. Diabetes Care. 2008;31(3):430–432. doi: 10.2337/dc07-1899. [DOI] [PubMed] [Google Scholar]
  • 12.Cobden D, Lee WC, Balu S, Joshi AV, Pashos CL. Health outcomes and economic impact of therapy conversion to a biphasic insulin analog pen among privately insured patients with type 2 diabetes mellitus. Pharmacotherapy. 2007;27(7):948–962. doi: 10.1592/phco.27.7.948. [DOI] [PubMed] [Google Scholar]
  • 13.Davis EM, Christensen CM, Nystrom KK, Foral PA, Destache C. Patient satisfaction and costs associated with insulin administered by pen device or syringe during hospitalization. Am J Health Syst Pharm. 2008;65(14):1347–1357. doi: 10.2146/ajhp070636. [DOI] [PubMed] [Google Scholar]
  • 14.Graff MR, McClanahan MA. Assessment by patients with diabetes mellitus of two insulin pen delivery systems versus a vial and syringe. Clin Ther. 1998;20(3):486–496. doi: 10.1016/s0149-2918(98)80058-1. [DOI] [PubMed] [Google Scholar]
  • 15.Kadiri A, Chraibi A, Marouan F, Ababou MR, el Guermai N, Wadjinny A, Kerfati A, Douiri M, Bensouda JD, Belkhadir J, Arvanitis Y. Comparison of NovoPen 3 and syringes/vials in the acceptance of insulin therapy in NIDDM patients with secondary failure to oral hypoglycaemic agents. Diabetes Res Clin Pract. 1998;41(1):15–23. doi: 10.1016/s0168-8227(98)00055-2. [DOI] [PubMed] [Google Scholar]
  • 16.Keith K, Nicholson D, Rogers D. Accuracy and precision of low-dose insulin administration using syringes, pen injectors, and a pump. Clin Pediatr (Phila) 2004;43(1):69–74. doi: 10.1177/000992280404300109. [DOI] [PubMed] [Google Scholar]
  • 17.Korytkowski M, Bell D, Jacobsen C, Suwannasari R FlexPen Study Team. A multicenter, randomized, open-label, comparative, two-period crossover trial of preference, efficacy, and safety profiles of a prefilled, disposable pen and conventional vial/syringe for insulin injection in patients with type 1 or 2 diabetes mellitus. Clin Ther. 2003;25(11):2836–2848. doi: 10.1016/s0149-2918(03)80337-5. [DOI] [PubMed] [Google Scholar]
  • 18.Lee IT, Liu HC, Liau YJ, Lee WJ, Huang CN, Sheu WH. Improvement in health-related quality of life, independent of fasting glucose concentration, via insulin pen device in diabetic patients. J Eval Clin Pract. 2009;15(4):699–703. doi: 10.1111/j.1365-2753.2008.01086.x. [DOI] [PubMed] [Google Scholar]
  • 19.Lee WC, Balu S, Cobden D, Joshi AV, Pashos CL. Medication adherence and the associated health-economic impact among patients with type 2 diabetes mellitus converting to insulin pen therapy: an analysis of third-party managed care claims data. Clin Ther. 2006;28(10):1712–1725. doi: 10.1016/j.clinthera.2006.10.004. discussion 1710–11. [DOI] [PubMed] [Google Scholar]
  • 20.Lteif AN, Schwenk WF. Accuracy of pen injectors versus insulin syringes in children with type 1 diabetes. Diabetes Care. 1999;22(1):137–140. doi: 10.2337/diacare.22.1.137. [DOI] [PubMed] [Google Scholar]
  • 21.Rubin RR, Peyrot M. Quality of life, treatment satisfaction, and treatment preference associated with use of a pen device delivering a premixed 70/30 insulin aspart suspension (aspart protamine suspension/soluble aspart) versus alternative treatment strategies. Diabetes Care. 2004;27(10):2495–2497. doi: 10.2337/diacare.27.10.2495. [DOI] [PubMed] [Google Scholar]
  • 22.Davis EM, Bebee A, Crawford L, Destache C. Nurse satisfaction using insulin pens in hospitalized patients. Diabetes Educ. 2009;35(5):799–809. doi: 10.1177/0145721709340057. [DOI] [PubMed] [Google Scholar]
  • 23.Pawaskar MD, Camacho FT, Anderson RT, Cobden D, Joshi AV, Balkrishnan R. Health care costs and medication adherence associated with initiation of insulin pen therapy in medicaid-enrolled patients with type 2 diabetes: a retrospective database analysis. Clin Ther. 2007;29:1294–1305. [PubMed] [Google Scholar]
  • 24.Wilk T, Mora PF, Chaney S, Shaw K. Use of an insulin pen by homeless patients with diabetes mellitus. J Am Acad Nurse Pract. 2002;14(8):372–379. doi: 10.1111/j.1745-7599.2002.tb00138.x. [DOI] [PubMed] [Google Scholar]
  • 25.Hörnquist JO, Wikby A, Andersson PO, Dufva AM. Insulin-pen treatment, quality of life and metabolic control: retrospective intra-group evaluations. Diabetes Res Clin Pract. 1990;10(3):221–230. doi: 10.1016/0168-8227(90)90065-2. [DOI] [PubMed] [Google Scholar]
  • 26.Wikby A, Stenström U, Andersson PO, Hörnquist J. Metabolic control, quality of life, and negative life events: a longitudinal study of well-controlled and poorly regulated patients with type 1 diabetes after changeover to insulin pen treatment. Diabetes Educ. 1998;24(1):61–66. doi: 10.1177/014572179802400108. [DOI] [PubMed] [Google Scholar]
  • 27.Pfützner A, Reimer T, Hohberg C, Frøkjaer LP, Jørgensen C. Prefilled insulin device with reduced injection force: patient perception and accuracy. Curr Med Res Opin. 2008;24(9):2545–2549. doi: 10.1185/03007990802329264. [DOI] [PubMed] [Google Scholar]
  • 28.McKay M, Compion G, Lytzen L. A comparison of insulin injection needles on patients' perceptions of pain, handling, and acceptability: a randomized, open-label, crossover study in subjects with diabetes. Diabetes Technol Ther. 2009;11(3):195–201. doi: 10.1089/dia.2008.0054. [DOI] [PubMed] [Google Scholar]
  • 29.Diglas J, Feinböck C, Irsigler K, Winkler F, Egger T, Weitgasser R, Pieber T, Lytzen L. Reduced pain perception with Pen Mate™, an automatic needle insertion device for use with an insulin pen. Pract Diabetes Int. 2005;16(2):39–41. [Google Scholar]
  • 30.Coscelli C, Calabrese G, Fedele D, Pisu E, Calderini C, Bistoni S, Lapolla A, Mauri MG, Rossi A, Zappella A. Use of premixed insulin among the elderly. Reduction of errors in patient preparation of mixtures. Diabetes Care. 1992;15(11):1628–1630. doi: 10.2337/diacare.15.11.1628. [DOI] [PubMed] [Google Scholar]
  • 31.Zisser H, Jovanovic L. OmniPod Insulin Management System: patient perceptions, preference, and glycemic control. Diabetes Care. 2006;29(9):2175. doi: 10.2337/dc06-0986. [DOI] [PubMed] [Google Scholar]
  • 32.Burdick J, Chase HP, Slover RH, Knievel K, Scrimgeour L, Maniatis AK, Klingensmith GJ. Missed insulin meal boluses and elevated hemoglobin A1c levels in children receiving insulin pump therapy. Pediatrics. 2004;113(3 Pt 1):e221–e224. doi: 10.1542/peds.113.3.e221. [DOI] [PubMed] [Google Scholar]
  • 33.Rodbard HW, Blonde L, Braithwaite SS, Brett EM, Cobin RH, Handelsman Y, Hellman R, Jellinger PS, Jovanovic LG, Levy P, Mechanick JI, Zangeneh F AACE Diabetes Mellitus Clinical Practice Guidelines Task Force. American Association of Clinical Endocrinologists medical guidelines for clinical practice for the management of diabetes mellitus. Endocr Pract. 2007;13(Suppl 1):1–68. doi: 10.4158/EP.13.S1.1. [DOI] [PubMed] [Google Scholar]
  • 34.DeVries JH, Dayan CM. Indications and current criteria for insulin pump therapy in Europe. Infusystems Int. 2006;5(4):30–32. [Google Scholar]
  • 35.Potti LG, Haines ST. Continuous subcutaneous insulin infusion therapy: a primer on insulin pumps. J Am Pharm Assoc (2003) 2009;49(1):e1–e13. doi: 10.1331/JAphA.2009.08122. [DOI] [PubMed] [Google Scholar]
  • 36.Bode BW. Use of rapid-acting insulin analogues in the treatment of patients with type 1 and type 2 diabetes mellitus: insulin pump therapy versus multiple daily injections. Clin Ther. 2007;29(Suppl D):S135–S144. doi: 10.1016/j.clinthera.2007.12.013. [DOI] [PubMed] [Google Scholar]
  • 37.Plotnick LP, Clark LM, Brancati FL, Erlinger T. Safety and effectiveness of insulin pump therapy in children and adolescents with type 1 diabetes. Diabetes Care. 2003;26(4):1142–1146. doi: 10.2337/diacare.26.4.1142. [DOI] [PubMed] [Google Scholar]
  • 38.Bode BW, Sabbah HT, Gross TM, Fredrickson LP, Davidson PC. Diabetes management in the new millennium using insulin pump therapy. Diabetes Metab Res Rev. 2002;18(Suppl 1):S14–S20. doi: 10.1002/dmrr.205. [DOI] [PubMed] [Google Scholar]
  • 39.Retnakaran R, Zinman B. Continuous subcutaneous insulin infusion versus multiple daily injections: insights from clinical trials. Infusystems USA. 2006;3(2):9–11. [Google Scholar]
  • 40.Monami M, Lamanna C, Marchionni N, Mannucci E. Acta Diabetol. 2009. Continuous subcutaneous insulin infusion versus multiple daily insulin injections in type 1 diabetes: a meta-analysis. [Epub ahead of print.] [DOI] [PubMed] [Google Scholar]
  • 41.Retnakaran R, Hochman J, DeVries JH, Hanaire-Broutin H, Heine RJ, Melki V, Zinman B. Continuous subcutaneous insulin infusion versus multiple daily injections: the impact of baseline A1c. Diabetes Care. 2004;27(11):2590–2596. doi: 10.2337/diacare.27.11.2590. [DOI] [PubMed] [Google Scholar]
  • 42.Retnakaran R, DeVries JH, Hanaire-Broutin H, Heine RJ, Melki V, Zinman B. Continuous subcutaneous insulin infusion versus multiple daily injections: modeling predicted benefits in relationship to baseline A1c. Diabetes Care. 2005;28(7):1835–1836. doi: 10.2337/diacare.28.7.1835. [DOI] [PubMed] [Google Scholar]
  • 43.Monami M, Lamanna C, Marchionni N, Mannucci E. Continuous subcutaneous insulin infusion versus multiple daily insulin injections in type 2 diabetes: a meta-analysis. Exp Clin Endocrinol Diabetes. 2009;117(5):220–222. doi: 10.1055/s-0028-1119405. [DOI] [PubMed] [Google Scholar]
  • 44.Jankovec Z, Rusavy Z. Insulin pump treatment in type 2 diabetes. Infusystems USA. 2008;5(4):29–32. [Google Scholar]
  • 45.Cope JU, Morrison AE, Samuels-Reid J. Adolescent use of insulin and patient-controlled analgesia pump technology: a 10-year Food and Drug Administration retrospective study of adverse events. Pediatrics. 2008;121(5):e1133–e1138. doi: 10.1542/peds.2007-1707. [DOI] [PubMed] [Google Scholar]
  • 46.Hanas R, Ludvigsson J. Hypoglycemia and ketoacidosis with insulin pump therapy in children and adolescents. Pediatr Diabetes. 2006;7(Suppl 4):32–38. doi: 10.1111/j.1399-543X.2006.00169.x. [DOI] [PubMed] [Google Scholar]
  • 47.Wolfsdorf J, Glaser N, Sperling MA. American Diabetes Association. Diabetic ketoacidosis in infants, children, and adolescents: a consensus statement from the American Diabetes Association. Diabetes Care. 2006;29(5):1150–1159. doi: 10.2337/diacare.2951150. [DOI] [PubMed] [Google Scholar]
  • 48.Conwell LS, Pope E, Artiles AM, Mohanta A, Daneman A, Daneman D. Dermatological complications of continuous subcutaneous insulin infusion in children and adolescents. J Pediatr. 2008;152(5):622–628. doi: 10.1016/j.jpeds.2007.10.006. [DOI] [PubMed] [Google Scholar]
  • 49.Conwell LS, Daneman D. Complications of insulin pump therapy in children and adolescents. Infusystems USA. 2008;5(4):25–28. [Google Scholar]
  • 50.St Charles M, Lynch P, Graham C, Minshall ME. Value Health. 2008. A cost-effectiveness analysis of continuous subcutaneous insulin injection versus multiple daily injections in type 1 diabetes patients: a third-party US payer perspective. [Epub ahead of print.] [DOI] [PubMed] [Google Scholar]
  • 51.Pickup J. How good is glycaemic control during CSII? Do we still need an artificial pancreas? Infusystems Int. 2006;5(4):25–29. [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES