Skip to main content
American Journal of Physiology - Heart and Circulatory Physiology logoLink to American Journal of Physiology - Heart and Circulatory Physiology
. 2010 Apr 30;299(1):H62–H69. doi: 10.1152/ajpheart.00922.2009

Dynamic control of maximal ventricular elastance via the baroreflex and force-frequency relation in awake dogs before and after pacing-induced heart failure

Xiaoxiao Chen 1, Javier A Sala-Mercado 2, Robert L Hammond 2,3, Masashi Ichinose 2, Soroor Soltani 1, Ramakrishna Mukkamala 1, Donal S O'Leary 2,
PMCID: PMC2904132  PMID: 20435845

Abstract

We investigated to what extent maximal ventricular elastance (Emax) is dynamically controlled by the arterial baroreflex and force-frequency relation in conscious dogs and to what extent these mechanisms are attenuated after the induction of heart failure (HF). We mathematically analyzed spontaneous beat-to-beat hemodynamic variability. First, we estimated Emax for each beat during a baseline period using the ventricular unstressed volume determined with the traditional multiple beat method during vena cava occlusion. We then jointly identified the transfer functions (system gain value and time delay per frequency) relating beat-to-beat fluctuations in arterial blood pressure (ABP) to Emax (ABP→Emax) and beat-to-beat fluctuations in heart rate (HR) to Emax (HR→Emax) to characterize the dynamic properties of the arterial baroreflex and force-frequency relation, respectively. During the control condition, the ABP→Emax transfer function revealed that ABP perturbations caused opposite direction Emax changes with a gain value of −0.023 ± 0.012 ml−1, whereas the HR→Emax transfer function indicated that HR alterations caused same direction Emax changes with a gain value of 0.013 ± 0.005 mmHg·ml−1·(beats/min)−1. Both transfer functions behaved as low-pass filters. However, the ABP→Emax transfer function was more sluggish than the HR→Emax transfer function with overall time constants (indicator of full system response time to a sudden input change) of 11.2 ± 2.8 and 1.7 ± 0.5 s (P < 0.05), respectively. During the HF condition, the ABP→Emax and HR→Emax transfer functions were markedly depressed with gain values reduced to −0.0002 ± 0.007 ml−1 and −0.001 ± 0.004 mmHg·ml−1·(beats/min)−1 (P < 0.1). Emax is rapidly and significantly controlled at rest, but this modulation is virtually abolished in HF.

Keywords: autonomic nervous system, beat-to-beat variability, system identification, Treppe effect, ventricular contractility


the control of ventricular contractility importantly contributes to the modulation of cardiac output and therefore cardiovascular homeostasis. Specific control mechanisms involved include the arterial baroreflex and force-frequency relation (i.e., the Treppe or Bowditch effect). Both of these mechanisms may occur simultaneously. For example, in response to a fall in arterial blood pressure (ABP), there will be a baroreflex-mediated sympathoexcitation, which could increase the inotropic state. The concurrent tachycardia could also increase the inotropic state independently via the force-frequency relation. Previous studies have described these mechanisms through various indexes of ventricular contractility including Frank-Starling curves (9, 17, 44), cardiac sympathetic nerve discharge (10), maximal temporal derivative of ventricular pressure (8, 17, 26, 51), and maximal ventricular elastance (Emax) (2, 3, 12, 18, 23, 24, 27, 48). However, Emax is generally recognized as the most specific index available.

Nearly all investigations of Emax control have focused on its steady-state performance (e.g., gain values). To our knowledge, only two studies have examined its dynamic behavior (e.g., time constants and delays). Sunagawa and colleagues (18) characterized the dynamic properties of the baroreflex control of Emax by identifying the transfer function relating randomly perturbed carotid sinus pressure to estimated beat-to-beat fluctuations in Emax of anesthetized, vagotomized dogs. Their identified transfer function exhibited the expected negative feedback dynamics with an overall time constant indicative of sympathetic nervous mediation and could be parameterized as a second-order delay system. This same group then characterized the dynamic properties of Emax control via the efferent baroreflex limb, while also obtaining information about the force-frequency relation, by identifying the transfer functions relating the randomized left and right stellate ganglion stimulations to the estimated beat-to-beat fluctuations in Emax of isolated canine hearts with and without atrial pacing (27). Their identified transfer functions could also be represented with second-order delay systems and suggested that Emax is controlled by the left sympathetic nerve through direct inotropic action and by the right sympathetic nerve via both direct inotropic action and indirect chronotropic effects. While these two studies provide unique and valuable insights, they were performed in either isolated hearts or anesthetized animals with acute surgical trauma, which may markedly affect the dynamic control of Emax. The open-loop identification approach may also alter the compensatory mechanisms.

Hallmark features of heart failure (HF) are low basal Emax and depressed ability to control Emax. Furthermore, reflexes such as the arterial baroreflex are depressed in HF (2, 6, 33). Olivier and Stephenson (33) attributed the majority of the baroreflex impairment in the maintenance of ABP to the loss of the cardiac output component of the reflex. Other reflexes that depend on substantial cardiac responses, such as the muscle metaboreflex during exercise, are also markedly affected in HF because of the loss of the cardiac output component (42). Attenuation or loss of the ability to rapidly change Emax could contribute importantly to this impaired ability to raise cardiac output. To our knowledge, no study has assessed the effects of HF on the dynamic control of Emax.

In this study, we aimed 1) to separately characterize the dynamic control of Emax via the arterial baroreflex and force-frequency relation during normal closed-loop operation in conscious, chronically instrumented animals and 2) to determine in what way the two dynamic Emax control mechanisms are altered by HF. To accomplish these aims, we mathematically analyzed naturally occurring, beat-to-beat hemodynamic variability from fully conscious dogs before and after pacing-induced HF. A preliminary version of this study has been reported in abbreviated form (7).

MATERIALS AND METHODS

We collected continuous hemodynamic data from dogs standing quietly before and after rapid chronic pacing-induced HF. We then analyzed the spontaneous hemodynamic variability as follows. First, rather than using single beat methods (45, 49) to estimate the Emax fluctuations on a beat-to-beat basis as previously done, we employed a simple and potentially more reliable method in which the ventricular unstressed volume (V0) was assumed to be relatively constant and determined with the traditional multiple beat method during vena cava occlusion. We then jointly identified the transfer function relating beat-to-beat fluctuations in ABP to Emax (ABP→Emax) to characterize the dynamic properties of the arterial baroreflex and the transfer function relating beat-to-beat fluctuations in heart rate (HR) to Emax (HR→Emax) to characterize the dynamic properties of the force-frequency relation (see, e.g., Refs. 16, 20, 35, for background on transfer functions.) Finally, we statistically compared the results before and after HF.

Hemodynamic Data

The collection of the hemodynamic data for analysis in this study is described in detail elsewhere (42). Below, we briefly describe the instrumentation employed and the relevant aspects of the executed protocol. All procedures were approved by the Wayne State University Animal Investigation Committee and conformed to the National Institutes of Health guidelines.

We studied five healthy adult mongrel dogs (∼20–25 kg). We installed chronic instrumentation in each dog through two recovery surgeries. The instrumentation included a fully implanted, high-fidelity pressure transducer (Data Sciences International) for left ventricular (LV) pressure (LVP) measurement; two pairs of sonomicrometry crystals (Sonometrics) for LV short-axis and long-axis dimensions (DSA and DLA) measurement and subsequent computation of LV volume (LVV) via the modified ellipsoid formula (π/6) × DSA × DSA × DLA; a fluid-filled catheter that was connected to a standard extracorporeal pressure transducer (Abbott) for ABP measurement; an ultrasonic flow probe (Transonic Systems) for aortic flow rate measurement; two hydraulic vascular occluders (In Vivo Metrics) for superior and inferior vena cava occlusion; and three stainless steel sutures (Ethicon) that served as electrodes for ventricular pacing. (For studies unrelated to the present investigation, we also instrumented each dog during the second surgery for measurement of hindlimb blood flow via a retroperitoneal approach.) After allowing each dog at least 2 wk to recover from the thoracotomy and at least 1 wk to recover from the follow-up surgery, we continuously recorded the hemodynamic data at a sampling frequency of ∼300 Hz during a baseline period and multiple transient vena cava occlusions while the dog was standing quietly. Thereafter, we connected the ventricular pacing leads to a pacemaker set at a rate of 240 beats/min for ∼30 days to induce a moderate level of HF. We then discontinued the ventricular pacing and likewise recorded the hemodynamic data. We did not record data within the HF induction period.

Data Analysis

Determination of spontaneous beat-to-beat variability.

We estimated Emax on a beat-to-beat basis from the hemodynamic data during the baseline periods according to the method illustrated in Fig. 1. First, we employed the traditional multiple beat method for determining Emax, that is, linear regression on the end-systolic LVP-LVV points during transient vena cava occlusion (41). The slope and x-intercept of the resulting line respectively represent the average Emax and the LV V0. Assuming a constant V0, we then computed the LV elastance (LVE) waveform from the LVP and LVV waveforms during the baseline periods by dividing the former waveform by the difference between the latter waveform and V0 (averaged over the respective multiple transient vena cava occlusions to reduce noise). Finally, we established beat-to-beat Emax by identifying the maximum of the LVE waveform over each beat.

Fig. 1.

Fig. 1.

Method for estimating spontaneous beat-to-beat maximal ventricular elastance (Emax) variability from left ventricular pressure (LVP) and volume (LVV) measurements during transient vena cava occlusion (left) and a baseline period (right). V0 is the ventricular unstressed volume and is assumed to be constant (i.e., not modulated by control mechanisms), and LVE is the time-varying left ventricular elastance. t, Time.

We calculated the corresponding HR and ABP on a beat-to-beat basis by detecting each beat from the aortic flow rate waveform and averaging the ABP waveform over each beat during the baseline periods. We used HR rather than the standard cardiac cycle interval for the analysis, because the force-frequency relation has customarily been reported in terms of the former variable (12, 24). However, the use of the latter variable would not have materially altered the results that follow. We then converted the spontaneous Emax, ABP, and HR beat sequences to time series by first forming stepwise continuous time signals in which the value for each beat is held for the duration of the beat and then sampling these signals to 1 Hz using an anti-aliasing filter whose impulse response is a unit-area rectangular pulse of 2-s width.

Identification of transfer functions.

From these three time series, we identified the ABP→Emax and HR→Emax transfer functions according to the block diagram shown in Fig. 2. The perturbing noise source NEmax in the block diagram is also estimated and represents the residual variability in Emax due to, for example, an Emax estimation error as well as other control mechanisms such as the cardiopulmonary baroreflex. We mathematically represented the block diagram with the following dual-input, autoregressive exogenous input model:

Emax(t)=i=1paiEmax(ti)+i=qrbiABP(ti)+i=0mciHR(ti)+WEmax(t).

Here, t indicates discrete time; the three sets of unknown parameters (ai, bi, and ci) define the ABP→Emax and HR→Emax transfer functions; the unmeasured residual error WEmax together with the set of parameters (ai) specify NEmax; and the unknown model order terms, p, q, r, and m, limit the number of parameters to be estimated (22). For a fixed model order, we estimated the parameters in closed form from 3–6-min stationary intervals of zero-mean fluctuations in the ABP, HR, and Emax time series by linear least squares minimization of WEmax (22). We established the model order by setting p and r to, respectively, 2 and q (i.e., second-order delay representation of the arterial Emax baroreflex) and determining q and m from the time series (rather than any previous data) via a minimization of the standard minimum description length criterion (22). See discussion for the enumeration and justification of the major assumptions underlying the model.

Fig. 2.

Fig. 2.

Block diagram for identifying the transfer functions relating fluctuations in arterial blood pressure (ABP) to Emax (ABP→Emax) and fluctuations in heart rate (HR) to Emax (HR→Emax) from spontaneous beat-to-beat ABP, HR, and Emax variability. NEmax represents the residual variability in Emax not accounted for by the ABP and HR fluctuations and is also estimated.

Statistical comparisons.

We computed the mean and power spectra [via standard autoregressive modeling (22)] of the ABP, HR, and Emax time series. Our calculation of the HR spectra essentially corresponds to the efficient algorithm outlined in Ref. 4. We then parameterized the computed power spectra in terms of low-frequency (0.04–0.15 Hz), high-frequency (0.15–0.50 Hz), and total powers and the identified transfer functions in terms of gain value, time delay, and overall time constant [determined with a robust rectangular-based method (21)]. Finally, we employed paired t-tests to compare the scalar parameters mainly before and after HF but also within an experimental condition.

RESULTS

Table 1 shows the group average mean of the ABP, HR, and Emax time series as well as the group average V0 before and after pacing induced HF. Figure 3 illustrates the group average power spectra of these time series, and Table 2 provides their low-frequency, high-frequency, and total powers. Mean ABP and mean Emax decreased, whereas mean HR and V0 increased, from the control condition to the HF condition. (Mean Emax, as determined with the traditional multiple beat method during vena cava occlusion, revealed a similar result.) The Emax spectra appeared more narrowband than their ABP and HR counterparts. All of the power spectra looked depressed following the induction of HF. The corresponding low-frequency, high-frequency, and total powers all decreased, but only the low-frequency power of the ABP spectrum reached the P = 0.1 level of significance. The HR spectrum in particular also appeared shifted toward lower frequencies after HF induction, but this shift was likewise statistically insignificant.

Table 1.

Group average mean of the hemodynamic variables before and after HF

Hemodynamic Variable Mean Control HF
ABP, mmHg 111.4 ± 2.3 86.6 ± 1.5
HR, beats/min 109.4 ± 5.9 128.3 ± 3.9*
Emax, mmHg/ml 5.5 ± 0.1 3.1 ± 0.1*
V0, ml 10.5 ± 4.7 23.3 ± 10.4*

Values are means ± SE. HF, heart failure; ABP, arterial blood pressure; HR, heart rate; Emax, maximal ventricular elastance; V0, ventricular unstressed volume.

*

P < 0.05;

P < 0.005.

Fig. 3.

Fig. 3.

Group average power spectra of the spontaneous beat-to-beat ABP, HR, and Emax fluctuations (means ± SE) from 5 conscious dogs before and after pacing induced heart failure (HF). bpm, Beats/min.

Table 2.

Group average powers of the beat-to-beat fluctuations in the hemodynamic variables before and after HF

Hemodynamic Variable Power Control HF
ABP, mmHg2
    Low frequency 7.51 ± 1.11 2.86 ± 0.46*
    High frequency 4.35 ± 2.63 1.16 ± 0.48
    Total 22.59 ± 4.58 10.33 ± 2.31
HR, (beats/min)2
    Low frequency 25.56 ± 3.48 20.96 ± 8.49
    High frequency 64.93 ± 33.08 20.58 ± 11.68
    Total 116.03 ± 36.80 65.62 ± 21.05
Emax, mmHg2/ml2
    Low frequency 0.0080 ± 0.0025 0.0026 ± 0.0004
    High frequency 0.0041 ± 0.0020 0.0024 ± 0.0008
    Total 0.0290 ± 0.0093 0.0115 ± 0.0037

Values are means ± SE. Low frequency is 0.04–0.15 Hz, and high frequency is 0.15–0.50 Hz.

*

P < 0.05.

Figure 4 illustrates the group average ABP→Emax and HR→Emax transfer functions in terms of the magnitude and phase responses as well as the intuitive step responses before and after HF. The computation of the magnitude and phase responses at frequencies outside the range of the 3–6-min time series was feasible because of the use of a parametric model. Table 3 shows the gain values and the overall time constants of these transfer functions. The transfer functions during the control condition are precisely defined as follows:

HABPEmax(ω)=0.0013e2jω11.3858ejω+0.4585e2jω
HHREmax(ω)=0.0038+0.0113ejω0.0055e2jω0.0016e3jω+0.0004e4jω+0.0004e5jω11.3858ejω+0.4585e2jω

where H indicates transfer function and ω indicates normalized frequency.

Fig. 4.

Fig. 4.

Group average ABP→Emax and HR→Emax transfer functions in terms of magnitude and phase responses as well as step responses (means ± SE) from the 5 conscious dogs before and after HF.

Table 3.

Group average gain values and overall time constants of the transfer functions relating beat-to-beat fluctuations in ABP to Emax (ABP→Emax) and beat-to-beat fluctuations in HR to Emax (HR→Emax) before and after HF

Transfer Function Parameters Control HF
GABPEmax, ml−1 −0.023 ± 0.012 −0.0002 ± 0.007
GHREmax, mmHg·ml−1·(beats/min)−1 0.013 ± 0.005 −0.0010 ± 0.004
τABPEmax, s 11.2 ± 2.8*
τHREmax, s 1.7 ± 0.5*

Values are means ± SE. G, gain value.

*

P < 0.05, two values marked with asterisk are significantly different from each other.

P < 0.10.

Overall time constant (τ) was not well defined because of the small transfer function magnitude (see Fig. 4).

During the control condition, the ABP→Emax and HR→Emax transfer functions exhibited negative and positive gain values, respectively. Thus Emax would decrease (increase) in the steady state in response to an increase (decrease) in mean ABP but in the absence of any change to mean HR as a result of the arterial baroreflex, whereas Emax would increase (decrease) in the steady state in response to an increase (decrease) in mean HR without any change to mean ABP because of the force-frequency relation. The two transfer functions were similar in that they both behaved as low-pass filters. Thus Emax would reach steady state without oscillating in response to a change in mean HR or mean ABP. However, the ABP→Emax transfer function was more sluggish. This transfer function showed a notable time delay and an overall time constant that was over five times as large as that of the HR→Emax transfer function. Thus Emax would reach steady state faster in response to a change in mean HR than a perturbation to mean ABP.

During the HF condition, the ABP→Emax and HR→Emax transfer functions were markedly blunted, except for the higher-frequency regime of the latter transfer function. The gain values of the two transfer functions were, in particular, reduced to negligible values. The time constants and delays were not well defined because of the small transfer function magnitudes. Thus Emax would not change much transiently and especially in the steady state in response to a variation in either mean ABP or mean HR.

DISCUSSION

This study is the first to examine the dynamic control of Emax in conscious subjects. Furthermore, this study is the first to determine the effects of HF on the dynamic control of Emax in isolated hearts, anesthetized preparations, or conscious subjects. Through mathematical analysis of spontaneous beat-to-beat hemodynamic variability, we were able to reveal quantitatively the strength and temporal (or frequency) characteristics of the arterial baroreflex and force-frequency relation in modulating ventricular contractility during normal closed-loop operation in healthy and HF conditions. In particular, we found that both of these mechanisms significantly and independently controlled Emax with response times on the order of seconds during healthy conditions. Furthermore, the strength of both the arterial baroreflex and force-frequency relation was severely reduced in HF, not only in the steady state but also in the transient phase. This reduced ability to elicit beat-to-beat changes in Emax likely contributes importantly to the loss of the cardiac output component in the reflex cardiovascular control mechanisms seen in HF (33, 42).

Dynamic Emax Control

Control condition.

The ABP→Emax transfer function showed negative feedback dynamics during the control condition (see Fig. 4), which is consistent with the arterial baroreflex mechanism. More specifically, the transfer function behaved as a low-pass filter (see Fig. 4) with a time delay and overall time constant of 2.6 ± 0.5 and 11.2 ± 2.8 s, respectively (see Table 3). These dynamics are indicative of the well-known sympathetic nervous mediation of the arterial Emax baroreflex (5). Sunagawa and colleagues (18) reported that the transfer function relating randomly perturbed carotid sinus pressure to beat-to-beat Emax fluctuations in anesthetized, vagotomized dogs similarly exhibited low-pass frequency characteristics with a time delay and overall time constant of 2.3 and 11 s, respectively. However, the gain value of −0.085 ml−1 that they found is nearly four times as large as the −0.023 ± 0.012 ml−1 ABP→Emax gain value that we obtained (see Table 3). This difference is in accord with a previous study showing that the steady-state carotid sinus baroreflex control of the normalized maximal temporal derivative of ventricular pressure was smaller in conscious than in anesthetized dogs (51).

The HR→Emax transfer function indicated that HR changes directionally cause the same Emax changes during the control condition (see Fig. 4), which is consistent with the force-frequency relation. In particular, when compared with the ABP→Emax transfer function, this transfer function likewise acted as a low-pass filter (see Fig. 4) but with a minimal time delay and a significantly smaller overall time constant of 1.7 ± 0.5 s (see Table 3). These dynamics are coherent with the belief that the force-frequency relation is mediated by fast intracellular calcium handling mechanisms (though the more sluggish sympathetic nervous system does appear to set its operating point and thus its gain value) (34, 38). The HR→Emax gain value of 0.013 ± 0.005 mmHg·ml−1·(beats/min)−1 (see Table 3) is about three times smaller than the 0.03 mmHg·ml−1·(beats/min)−1 gain value previously obtained from isolated canine hearts at the same mean HR (24). Deviations in these values are likely due at least in part to the different experimental conditions.

These results indicate that the combined effect of the arterial baroreflex and force-frequency relation can significantly increase ventricular contractility and thereby importantly aid in pressure homeostasis. In response to pressure challenges, changes in autonomic activity elicit both changes in ventricular performance as well as changes in HR. For example, if mean ABP fell by 10 mmHg, our results predict that the arterial baroreflex would increase Emax by 0.23 ml/mmHg. Based on recent studies from our laboratory on dynamic baroreflex control of HR using the same animal model, HR would be expected to increase by nearly 60 beats/min in response to this pressure insult (43). Our results here predict that this tachycardia would, in turn, cause the force-frequency relation to increase Emax by nearly 0.8 ml/mmHg. Together, the direct baroreflex-mediated increase in ventricular contractility coupled with that elicited indirectly by the force-frequency relation would increase Emax by nearly 20%. Using data from previous studies on Emax control (18, 24) would overestimate this change. Since the time constants governing the baroreflex control of HR and Emax are similar (5), the time for Emax to reach steady state in response to the fall in ABP would be within 1 min. Finally, the standard deviation of Emax variability because of ABP fluctuations was about the same as that due to HR fluctuations during healthy conditions (result not shown). Thus ABP and HR fluctuations contribute roughly the same to the genesis of normal Emax variability.

HF condition.

The ABP→Emax transfer function including its gain value was blunted essentially to zero following the induction of HF (see Fig. 4 and Table 3). This alteration is likely due to the diminished total baroreflex arch functioning in HF (14, 31, 52), as well as to depressed sympathetic nervous responsiveness during the pathophysiological condition (i.e., sympathetic nervous activity is always overexcited and therefore does not respond to oscillations in ABP and its other inputs). Thus not only are baroreflex-mediated changes in HR smaller in HF, but the abolition of any change in ventricular contractility will further act to attenuate the strength of the arterial baroreflex in the control of ABP as baroreflex-mediated changes in cardiac output will be depressed (15).

The HR→Emax transfer function and especially its gain value were depressed following the induction of HF (see Fig. 4 and Table 3). Several previous studies have shown that the force-frequency relation was reduced or even inverted in the steady state in isolated myocardium from failing human hearts (11, 29); conscious, but autonomically blocked, dogs following pacing induced HF (2); and HF patients (11, 13). The mechanisms involved appear to be abnormal intracellular calcium handling (37). Another potential mechanism for the diminished strength of the force-frequency relation in our study may be the significant increase in mean HR (i.e., shift in operating point; see Table 3) as previously shown in isolated canine hearts (24). The abolishment of the ABP→Emax transfer function and the marked attenuation of the HR→Emax transfer function coupled with lower baroreflex control of HR likely contribute to the markedly attenuated control of cardiac output in HF (15, 33).

Beat-to-Beat Hemodynamic Variability

The power spectra of the beat-to-beat ABP, HR, and Emax fluctuations during the control condition showed a high-frequency peak around 0.25 Hz but did not reveal a low-frequency peak near 0.1 Hz (see Fig. 3), which has previously been reported in conscious dogs (40). These power spectra all showed a tendency to decrease during HF (see Fig. 3 and Table 2). A number of previous studies have shown that HR variability is reduced in HF including after rapid chronic pacing in conscious dogs (36) due to aberrant autonomic nervous system functioning, and at least one study has demonstrated that ABP variability is smaller in HF patients than control subjects (39). We are not aware of any previous study addressing Emax variability during HF. The lack of statistical significance of the reductions in spectral powers in our study may be due to the smaller sample size as well as an error in the beat-to-beat Emax estimates. Importantly, however, the identified transfer functions coupling the beat-to-beat fluctuations were able to show greater levels of statistical significance, since a major portion of the beat-to-beat Emax estimation error was likely uncorrelated with the independently measured ABP and HR fluctuations and therefore ascribed to the perturbing noise term NEmax rather than the transfer functions (see Fig. 2). Indeed, it is for this reason that we have focused the study on the transfer functions rather than the beat-to-beat variability.

Data Analysis: Assumptions, Limitations, and Strengths

We employed Emax as the quantitative index of ventricular contractility, thereby assuming that the ventricular end-systolic pressure-volume relationship was linear. Although nonlinearity has been observed over wide pressure and volume ranges (32, 50), the relationship indeed appears to be linear over more physiological pressure and volume ranges in conscious dogs (46, 47) as well as humans in health (25) and with severe HF (1).

We used V0 determined with the traditional multiple beat method during vena cava occlusion to estimate the requisite beat-to-beat Emax during the corresponding baseline period (see Fig. 1). We therefore assumed that V0 was constant within an experimental condition of a dog (i.e., not modulated by control mechanisms). This assumption originates from the classic studies of Sagawa and colleagues demonstrating that inotropic agents selectively alter Emax (e.g., Ref. 41) as well as subsequent work showing that average Emax, but not V0, is controlled over a wide HR range via the force-frequency relation (24). Although single-beat methods for estimating Emax on a beat-to-beat basis do not involve such a constant V0 assumption, they are based on other, seemingly more stringent, assumptions such as that LVE function, normalized in amplitude and time over each beat, is invariant (45) or that peak isovolumic pressure at end-diastolic volume can be extrapolated from the isovolumic phase LVP of ejecting beats (49). These methods have been verified during large perturbations to Emax but not smaller, spontaneous Emax changes. We did attempt to implement several variants of the single-beat method based on an invariant normalized LVE function to estimate beat-to-beat Emax. However, the resulting Emax fluctuations were fast and large, even extending to nonphysiological, negative values. For example, the arterial baroreflex gain value determined from these fluctuations was often positive.

We characterized dynamic Emax control via the arterial baroreflex and the force-frequency relation through the ABP→Emax and HR→Emax transfer functions, respectively. Thus we assumed that these control mechanisms were linear and time invariant. Sunagawa and colleagues (18, 27) showed that the coherence of the transfer functions relating randomly perturbed carotid sinus pressure and randomized left and right stellate ganglion stimulations to estimated beat-to-beat fluctuations in Emax during stable conditions was between 0.5 and 0.8 over the lower-frequency regime. Since the spontaneous hemodynamic fluctuations analyzed herein were smaller and likewise obtained at rest, our linearity and time invariance assumption appears to be quite tenable. However, the trade-off is that the transfer functions are only valid over the limited range of the naturally occurring hemodynamic variability.

Finally, we jointly identified the ABP→Emax and HR→Emax transfer functions from spontaneous beat-to-beat ABP, HR, and Emax fluctuations using a new dual-input autoregressive exogenous input model (see initial equation and Fig. 2). We therefore made the following additional assumptions. First, the transfer functions may be succinctly represented with pole-zero structures. Such structures have proven successful in representing various physiological control systems in the past (e.g., Refs. 28, 30). Second, the residual error term WEmax in the model is a white noise process that is uncorrelated with the HR fluctuations and a similar process inducing the ABP fluctuations, whereas the model order term q is nonnegative to enforce causality in the closed-loop path between the ABP and Emax fluctuations. This assumption is needed for a unique and consistent estimation of the model parameters (ai, bi, and ci) (53). Although a lack of correlation may not strictly hold, the baroreflex does have a time delay (5) and is therefore causal. Third, the model order terms p and r are set to 2 and q, respectively. This assumption arises from the aforesaid nonparametric modeling results of Sunagawa and colleagues (18, 27) obtained from anesthetized, vagotomized dogs and isolated canine hearts. (Note that, like the model parameters, the model order terms q and m were estimated from the measured fluctuations.)

With the above assumptions, we were able obtain results uniformly consistent with known qualitative physiology. A novelty of these results is in quantitatively revealing how Emax is dynamically controlled during relevant physiological conditions. Furthermore, the HR→Emax transfer function may not have been reported before even during nonphysiological conditions.

GRANTS

This work was supported by the National Heart, Lung, and Blood Institute Grants HL-55473, HL-080568, and HL-095819 and the National Science Foundation Grant CAREER 0643477.

DISCLOSURES

No conflicts of interest, financial or otherwise, are declared by the author(s).

REFERENCES

  • 1.Aroney CN, Herrmann HC, Semigran MJ, Dec GW, Boucher CA, Fifer MA. Linearity of the left ventricular end-systolic pressure-volume relation in patients with severe heart failure. J Am Coll Cardiol 14: 127–134, 1989 [DOI] [PubMed] [Google Scholar]
  • 2.Asanoi H, Ishizaka S, Joho S, Kameyama T, Inoue H, Sasayama S. Altered inotropic and lusitropic responses to heart rate in conscious dogs with tachycardia-induced heart failure. J Am Coll Cardiol 27: 728–735, 1996 [DOI] [PubMed] [Google Scholar]
  • 3.Asanoi H, Ishizaka S, Kameyama T, Sasayama S. Neural modulation of ventriculoarterial coupling in conscious dogs. Am J Physiol Heart Circ Physiol 266: H741–H748, 1994 [DOI] [PubMed] [Google Scholar]
  • 4.Berger RD, Akselrod S, Gordon D, Cohen RJ. An efficient algorithm for spectral analysis of heart rate variability. IEEE Trans Biomed Eng 33: 900–904, 1986 [DOI] [PubMed] [Google Scholar]
  • 5.Berger RD, Saul JP, Cohen RJ. Transfer function analysis of autonomic regulation. I. Canine atrial rate response. Am J Physiol Heart Circ Physiol 256: H142–H152, 1989 [DOI] [PubMed] [Google Scholar]
  • 6.Bhargava V, Shabetai R, Mathiasen R, Dalton N, Hunter JJ, Ross J., Jr Loss of adrenergic control of the force-frequency relation in heart failure secondary to idiopathic or ischemic cardiomyopathy. Am J Cardiol 81: 1130–1137, 1998 [DOI] [PubMed] [Google Scholar]
  • 7.Chen X, Mukkamala R, Sala-Mercado JA, Hammond RL, Ichinose M, Soltani S, O'Leary DS. Dynamic control of maximal ventricular elastance in conscious dogs before and after pacing-induced heart failure. Conf Proc IEEE Eng Med Biol Soc 2009: 5328–5331, 2009 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.De Pauw M, Vilaine JP, Heyndrickx GR. Role of force-frequency relation during AV-block, sinus node block and beta-adrenoceptor block in conscious animals. Basic Res Cardiol 99: 360–371, 2004 [DOI] [PubMed] [Google Scholar]
  • 9.Downing SE, Gardner TH. Reflex regulation of ventricular work performance. Yale J Biol Med 39: 73–89, 1966 [PMC free article] [PubMed] [Google Scholar]
  • 10.Downing SE, Sonnenblick EH. Effects of continuous administration of angiotensin II on ventricular performance. J Appl Physiol 18: 585–592, 1963 [DOI] [PubMed] [Google Scholar]
  • 11.Feldman MD, Gwathmey JK, Phillips P, Schoen F, Morgan JP. Reversal of the force-frequency relationship in working myocardium from patients with end-stage heart-failure. J Appl Cardiol 3: 273–283, 1988 [DOI] [PubMed] [Google Scholar]
  • 12.Freeman GL, Little WC, O'Rourke RA. Influence of heart rate on left ventricular performance in conscious dogs. Circ Res 61: 455–464, 1987 [DOI] [PubMed] [Google Scholar]
  • 13.Hasenfuss G, Holubarsch C, Hermann HP, Astheimer K, Pieske B, Just H. Influence of the force-frequency relationship on haemodynamics and left ventricular function in patients with non-failing hearts and in patients with dilated cardiomyopathy. Eur Heart J 15: 164–170, 1994 [DOI] [PubMed] [Google Scholar]
  • 14.Hirsch AT, Dzau VJ, Creager MA. Baroreceptor function in congestive heart failure: effect on neurohumoral activation and regional vascular resistance. Circulation 75: IV36–IV48, 1987 [PubMed] [Google Scholar]
  • 15.Ichinose M, Sala-Mercado JA, O'Leary DS, Hammond RL, Coutsos M, Ichinose T, Pallante M, Iellamo F. Spontaneous baroreflex control of cardiac output during dynamic exercise, muscle metaboreflex activation, and heart failure. Am J Physiol Heart Circ Physiol 294: H1310–H1316, 2008 [DOI] [PubMed] [Google Scholar]
  • 16.Karu ZZ. Signals and Systems Made Ridiculously Simple Huntsville, AL: ZiZi, 1995 [Google Scholar]
  • 17.Kass DA. Force-frequency relation in patients with left ventricular hypertrophy and failure. Basic Res Cardiol 93, Suppl 1: 108–116, 1998 [DOI] [PubMed] [Google Scholar]
  • 18.Kubota T, Alexander J, Jr, Itaya R, Todaka K, Sugimachi M, Sunagawa K, Nose Y, Takeshita A. Dynamic effects of carotid sinus baroreflex on ventriculoarterial coupling studied in anesthetized dogs. Circ Res 70: 1044–1053, 1992 [DOI] [PubMed] [Google Scholar]
  • 19.Kudo T, Mikuniya A, Suto N, Okubo T, Yamamoto T, Okumura K. Cardiac sympathetic stimulation increases cardiac contractility but decreases contractile efficiency in canine hearts in vivo. Jpn Circ J 62: 925–932, 1998 [DOI] [PubMed] [Google Scholar]
  • 20.Lathi BP. Linear Systems and Signals New York, NY: Oxford University Press, 2004 [Google Scholar]
  • 21.Lathi BP. Signal Processing and Linear Systems Carmichael, CA: Berkeley Cambridge, 1998 [Google Scholar]
  • 22.Ljung L. System Identification: Theory for the User Englewood Cliffs: Prentice Hall, 1987 [Google Scholar]
  • 23.Matsuura W, Sugimachi M, Kawada T, Sato T, Shishido T, Miyano H, Nakahara T, YI , Alexander J, Jr, Sunagawa K. Vagal stimulation decreases left ventricular contractility mainly through negative chronotropic effect. Am J Physiol Heart Circ Physiol 273: H534–H539, 1997 [DOI] [PubMed] [Google Scholar]
  • 24.Maughan WL, Sunagawa K, Burkhoff D, Graves WL, Jr, Hunter WC, Sagawa K. Effect of heart rate on the canine end-systolic pressure-volume relationship. Circulation 72, 654–659, 1985 [DOI] [PubMed] [Google Scholar]
  • 25.Mehmel HC, Stockins B, Ruffmann K, von Olshausen K, Schuler G, Kubler W. The linearity of the end-systolic pressure-volume relationship in man and its sensitivity for assessment of left ventricular function. Circulation 63: 1216–1222, 1981 [DOI] [PubMed] [Google Scholar]
  • 26.Miura T, Miyazaki S, Guth BD, Kambayashi M, Ross J., Jr Influence of the force-frequency relation on left ventricular function during exercise in conscious dogs. Circulation 86: 563–571, 1992 [DOI] [PubMed] [Google Scholar]
  • 27.Miyano H, Nakayama Y, Shishido T, Inagaki M, Kawada T, Sato T, Miyashita H, Sugimachi M, Alexander J, Jr, Sunagawa K. Dynamic sympathetic regulation of left ventricular contractility studied in the isolated canine heart. Am J Physiol Heart Circ Physiol 275: H400–H408, 1998 [DOI] [PubMed] [Google Scholar]
  • 28.Mukkamala R, Mathias JM, Mullen TJ, Cohen RJ, Freeman R. System identification of closed-loop cardiovascular control mechanisms: diabetic autonomic neuropathy. Am J Physiol Regul Integr Comp Physiol 276: R905–R912, 1999 [DOI] [PubMed] [Google Scholar]
  • 29.Mulieri LA, Hasenfuss G, Leavitt B, Allen PD, Alpert NR. Altered myocardial force-frequency relation in human heart failure. Circulation 85: 1743–1750, 1992 [DOI] [PubMed] [Google Scholar]
  • 30.Mullen TJ, Appel ML, Mukkamala R, Mathias JM, Cohen RJ. System identification of closed-loop cardiovascular control: effects of posture and autonomic blockade. Am J Physiol Heart Circ Physiol 272: H448–H461, 1997 [DOI] [PubMed] [Google Scholar]
  • 31.Niebauer MJ, Holmberg MJ, Zucker IH. Aortic baroreceptor characteristics in dogs with chronic high output failure. Basic Res Cardiol 81: 111–122, 1986 [DOI] [PubMed] [Google Scholar]
  • 32.Noda T, Cheng CP, De Tombe PP, Little WC. Curvilinearity of LV end-systolic pressure-volume and dP/dtmax-end-diastolic volume relations. Am J Physiol Heart Circ Physiol 265: H910–H917, 1993 [DOI] [PubMed] [Google Scholar]
  • 33.Olivier NB, Stephenson RB. Characterization of baroreflex impairment in conscious dogs with pacing-induced heart failure. Am J Physiol Regul Integr Comp Physiol 265: R1132–R1140, 1993 [DOI] [PubMed] [Google Scholar]
  • 34.Opie LH. Normal and abnormal cardiac function. In: Heart Disease: A Textbook of Cardiovascular Medicine (6th ed.), edited by Braunwald E, Zipes DP, Libby P. Philadelphia, PA: W. B. Saunders, 2001, p. 443 [Google Scholar]
  • 35.Oppenheim AV, Willsky AS, Hamid S. Signals and Systems Upper Saddle River, NJ: Prentice Hall, 1996 [Google Scholar]
  • 36.Piccirillo G, Ogawa M, Song J, Chong VJ, Joung B, Han S, Magri D, Chen LS, Lin SF, Chen PS. Power spectral analysis of heart rate variability and autonomic nervous system activity measured directly in healthy dogs and dogs with tachycardia-induced heart failure. Heart Rhythm 6: 546–552, 2009 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Pieske B, Kretschmann B, Meyer M, Holubarsch C, Weirich J, Posival H, Minami K, Just H, Hasenfuss G. Alterations in intracellular calcium handling associated with the inverse force-frequency relation in human dilated cardiomyopathy. Circulation 92: 1169–1178, 1995 [DOI] [PubMed] [Google Scholar]
  • 38.Piot C, Lemaire S, Albat B, Sequin J, Nargeot J, Richard S. High frequency-induced upregulation of human cardiac calcium currents. Circulation 93: 120–128, 1996 [DOI] [PubMed] [Google Scholar]
  • 39.Radaelli A, Perlangeli S, Cerutti MC, Mircoli L, Mori I, Boselli L, Bonaita M, Terzoli L, Candotti G, Signorini G, Ferrari AU. Altered blood pressure variability in patients with congestive heart failure. J Hypertens 17: 1905–1910, 1999 [DOI] [PubMed] [Google Scholar]
  • 40.Rimoldi O, Furlan R, Pagani MR, Piazza S, Guazzi M, Pagani M, Malliani A. Analysis of neural mechanisms accompanying different intensities of dynamic exercise. Chest 101: 226S–230S, 1992 [DOI] [PubMed] [Google Scholar]
  • 41.Sagawa K, Suga H, Shoukas AA, Bakalar KM. End-systolic pressure/volume ratio: a new index of ventricular contractility. Am J Cardiol 40: 748–753, 1977 [DOI] [PubMed] [Google Scholar]
  • 42.Sala-Mercado JA, Hammond RL, Kim JK, McDonald PJ, Stephenson LW, O'Leary DS. Heart failure attenuates muscle metaboreflex control of ventricular contractility during dynamic exercise. Am J Physiol Heart Circ Physiol 292: H2159–H2166, 2007 [DOI] [PubMed] [Google Scholar]
  • 43.Sala-Mercado JA, Ichinose M, Hammond RL, Coutsos M, Ichinose T, Pallante M, Iellamo F, O'Leary DS. Spontaneous baroreflex control of heart rate versus cardiac output: altered coupling in heart failure. Am J Physiol Heart Circ Physiol 294: H1304–H1309, 2008 [DOI] [PubMed] [Google Scholar]
  • 44.Sarnoff SJ, Gilmore JP, Brockman SK, Mitchell JH, Linden RJ. Regulation of ventricular contraction by the carotid sinus. Its effect on atrial and ventricular dynamics. Circ Res 8: 1123–1136, 1960 [DOI] [PubMed] [Google Scholar]
  • 45.Senzaki H, Chen CH, Kass DA. Single-beat estimation of end-systolic pressure-volume relation in humans. A new method with the potential for noninvasive application. Circulation 94: 2497–2506, 1996 [DOI] [PubMed] [Google Scholar]
  • 46.Sodums MT, Badke FR, Starling MR, Little WC, O'Rourke RA. Evaluation of left ventricular contractile performance utilizing end-systolic pressure-volume relationships in conscious dogs. Circ Res 54: 731–739, 1984 [DOI] [PubMed] [Google Scholar]
  • 47.Spratt JA, Tyson GS, Glower DD, Davis JW, Muhlbaier LH, Olsen CO, Rankin JS. The end-systolic pressure-volume relationship in conscious dogs. Circulation 75: 1295–1309, 1987 [DOI] [PubMed] [Google Scholar]
  • 48.Suga H, Sagawa K, Kostiuk DP. Controls of ventricular contractility assessed by pressure-volume ratio, Emax. Circ Res 10: 582–592, 1976 [DOI] [PubMed] [Google Scholar]
  • 49.Sunagawa K, Yamada A, Senda Y, Kikuchi Y, Nakamura M, Shibahara T, Nose Y. Estimation of the hydromotive source pressure from ejecting beats of the left ventricle. IEEE Trans Biomed Eng 27: 299–305, 1980 [DOI] [PubMed] [Google Scholar]
  • 50.van der Velde ET, Burkhoff D, Steendijk P, Karsdon J, Sagawa K, Baan J. Nonlinearity and load sensitivity of end-systolic pressure-volume relation of canine left ventricle in vivo. Circulation 83: 315–327, 1991 [DOI] [PubMed] [Google Scholar]
  • 51.Vatner SF, Higgins CB, Franklin D, Braunwald E. Extent of carotid sinus regulation of the myocardial contractile state in conscious dogs. J Clin Invest 51: 995–1008, 1972 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Wang W, Chen JS, Zucker IH. Carotid sinus baroreceptor sensitivity in experimental heart failure. Circulation 81: 1959–1966, 1990 [DOI] [PubMed] [Google Scholar]
  • 53.Wellstead PE, Edmunds JM. Least-squares identification of closed-loop systems. Int J Control 21: 689–699, 1975 [Google Scholar]

Articles from American Journal of Physiology - Heart and Circulatory Physiology are provided here courtesy of American Physiological Society

RESOURCES