Abstract
Background
Colonoscopy (CSPY) for colorectal cancer screening has several limitations. Colon Capsule Endoscopy (PillCam Colon, CCE) was compared to CSPY under routine screening conditions.
Methods
We performed a prospective, single-center pilot study at a University Hospital. Data were obtained from November 2007 until May 2008. Patients underwent CCE on Day 1 and CSPY on Day 2. Outcomes were evaluated regarding sensitivity and specificity of polyp detection rate, with a significance level set at >5 mm.
Results
59 individuals were included in this study, the results were evaluable in 56 patients (males 34, females 22; median age 59). CCE was complete in 36 subjects. Polyp detection rate for significant polyps was 11% on CSPY and 27% on CCE.
6/56 (11%) patients had polyps on CSPY not detected on CCE (miss rate).
Overall sensitivity was 79% (95% confidence interval [CI], 61 to 90), specificity was 54% (95% CI, 35 to 70), positive predictive value (PPV) was 63% and negative predictive value (NPV) was 71%. Adjusted to significance of findings, sensitivity was 50% (95% CI, 19 to 81), specificity was 76% (95% CI, 63 to 86), PPV was 20% and NPV was 93%.
Conclusion
In comparison to the gold standard, the sensitivity of CCE for detection of relevant polyps is low, however, the high NPV supports its role as a possible screening tool.
Trial Registration
Background
The incidence of colorectal cancer [(CRC), standardized to age and world population] is 20-45/100.000 for men und 15-30/100.000 for women[1]. It is increasing with age[2] and the cumulative lifetime risk both for men and for women arises to 6%[3]. Regardless of much improved diagnostics, 50% of patients with CRC die from it. The majority (90%) of CRC develop from benign adenomatous polyps. Current evidence points to the important role of screening colonoscopy with subsequent polypectomy (CSPY), reducing the risk of developing CRC by 76-90%[4-6]. Also, early recognization and removal of carcinomatous lesions is crucial, as 5-year survival rates for DUKES stage A (UICC 0 to 1) are above 90%, whereas overall 5-year-survival rate is 62%[7].
Screening guidelines recommending CSPY every 10 years for asymptomatic patients above 50 years with negative family history have been established in several European countries (eg. France, Italy, Germany) as well as in the United States in order to detect polyps and prevent progression into cancer[4,8,9]. Nevertheless, the average miss rates vary from 13% to 27%[10,11] (adenoma ≤1 cm), and around 5% for carcinoma[10]. Moreover, CSPY is an investigator-dependent and risk-bearing (risk of perforation 0.18%, with polypectomy 0.32%)[12] procedure, which requires adequate bowel preparation in order to be cost-effective[13]. Patient acceptance remains the main limiting factor to wide distribution of screening. Correct compliance for screening was observed in 37.1% of subjects and 62.9% for a single CSPY ever in life (retrospective follow-up data in the US (either on yearly FOBT, 5-yearly sigmoidoscopy or 10-yearly CSPY)[14].
For the reason of limited resources, in Switzerland, to date, CRC screening is not regularly promoted by the health-care system[15,16].
Taking all of the above mentioned into account, Colon capsule endoscopy (CCE; PillCam Colon®, Given Imaging Ltd., Yoqneam, Israel) might be a novel method for the screening of large populations.
As a further development to the well established PillCamSB®, two wide-angle cameras provide an extended view of lesions.
Two previous feasibility studies have shown its possible value for tertiary care centers in a study setting[17,18] as well as practicability in a private practice[19], however, a recent multicenter study evaluating both detection of polyps as well as cancer detected a limited sensitivity[20]. We report the findings of a single center study comparing the performance of CCE with CSPY for the detection of colorectal polyps and cancer.
Methods
Study design
This was a prospective, single center pilot analysis comparing the efficacy of CCE against routine screening colonoscopy at a tertiary care center (University Hospital of Basel, Switzerland). Patient enrolment was from November 5, 2007, to May 7, 2008. All patients provided written informed consent. The study was approved by the Ethics Committee and registered at http://www.clinicaltrials.gov (NCT00991003). The study was partly funded by the Nycomed Fund of the University Hospital Basel, Switzerland and by Given Imaging Ltd., Yoqneam, Israel.The authors designed the study, gathered and analyzed the data; the sponsors did not participate in design or conduct of the study nor did they review or approve the data.
Patients and Data Collection
Every patient who presented at or was referred to our center was considered. Men and women above the age of 50 years without symptoms (indication for screening) or with lower gastrointestinal signs and symptoms and individuals younger than 50 years with positive family history for CRC, minimum 18 years were included in this study. Exclusion criteria were CRC in the patient's history, cardiac pacemaker, contraindications for sodium phosphate solution (Colophos®) and risk factors for capsule retention including surgical intestinal anastomosis, Crohn's Disease, diverticulitis and radiologically suspected bowel obstruction.
General patient characteristics were assembled, including demographics, family history and recent surgery, as well as bowel habits. Patient acceptance was assessed with a 5-item questionnaire upon completion of the second examination (Table S1, additional file 1).
Adverse events were recorded on days 1 and 2 of the study. Additionally, some technical data such as completeness of the examinations (inspection of the whole large intestine from ileocoecal valve to anus), colon transit time, location of the capsule at the time of the first Colophos® booster dose and capsule excretion were recorded.
Study Definitions and Outcomes
The aim was to evaluate this novel method (CCE) for performance as a screening tool compared to CSPY in asymptomatic patients under routine screening conditions. The proclaimed benefit would be an increase in acceptance of screening for CRC and an augmented detection rate of adenoma and/or carcinoma. Significance was defined as polyp size >5 mm, with the hypothesis that detection rate on CCE corresponds with results of CSPY. The primary endpoint was the number of cancerous lesions and polyps detected on CCE compared to CSPY. Secondary endpoints were completeness of the exam, patient acceptance and adherence to preparation regimen. Subanalyses regarding effect of bowel preparation on polyp detection on CCE and accuracy of detection with respect to histopathology were performed.
Interventions
Patients underwent CCE on day 1 and CSPY on day 2. The examinations were carried out by different physicians, with blinding of results until both examinations had been completed and until interobserver evaluation was finished. CSPY was performed by one of eight different gastroenterologists and intubation of the terminal ileum was not required. CCE was read by two of two gastroenterologists at our department. Segmental unblinding was not feasible as the study was performed during the routine setting at a University Hospital.
The PillCam® Colon Capsule is 11 mm × 31 mm in size (Figure 1). It is equipped with two wide-angle (156°) cameras acquiring pictures from both ends of the capsule at a rate of 4 frames per second (2 pictures per second and camera). Activation of the device is automatical on package removal. A sleeping mode of 1 hour 45 minutes is entered upon 10 minutes after ingestion. Detailed procedure has been described elsewhere. Data analysis was performed using RAPID software.
The capsule video was read in three steps: identification of anatomical sites was done during "Quick View Mode" at a rate of 20 frames per second. In a second and third step the pictures were read at approximately 8 frames per second as described before. CCE was read only once in order to simulate a routine clinical setting. Detected lesions were reviewed by an additional physician who assessed their presence, size and localisation, with consecutive interobserver agreement.
Standard CSPY with or without polypectomy was performed under Propofol sedation using existing hospital protocols for preparation, and procedural and post-procedural care[21].
Polyps were estimated by size (greater than 5 mm, ≤5 mm) and location (right or left hemicolon). Polyps detected on CSPY were removed and histologically examined. Other pathologies such as diverticulae, angioectasia and hemorrhoids and their location were also noted but not considered as relevant.
Colon Preparation, Level of Cleansing and Propulsion of the Capsule
For colon cleansing we applied our department's standard preparation procedure for CSPY including low-fibre diet and Macrogolum (PEG, Cololyt®; Spirig Pharma, Egerkingen, Switzerland), and added an oral motility agent (Motilium®; Janssen-Cilag AG, Baar, Switzerland), Phospho Soda-boosters (Colophos®; Spirig Pharma, Egerkingen, Switzerland) and a suppository (Bisacodylum 10 mg, Prontolax®; Streuli Pharma, Uznach, Switzerland), Table 1.
Table 1.
Time (hours) | Action |
Day -3 | |
all day | Low-fibre diet |
Day -2 | |
all day | liquid diet |
Day -1 | |
all day | clear liquids only |
18:00-19:30 | 2 Liters of Cololyt® (1 cup every 10-15 minutes) |
Exam day | |
7:00-8:30 | 2 Liters of Cololyt® (1 cup every 10-15 minutes) |
11:00 | 20 mg Domperidone (Motilium®) with a cup of water |
11:15 | Ingestion of PillCam® Colon with a cup of water |
until 13:00 | no food/liquid ingestion |
13:00 | Real time view assessment if capsule left stomach, then Booster dose of 45 ml Colophos® + 1 liter of water |
17:00 | Real time view assessment if capsule left stomach, then Booster dose of 30 ml Colophos® + 1 liter of water *) |
19:30 | 10 mg Bisacodylum rectal suppository *) |
22:00 | End of examination |
Day +1 | Conventional colonoscopy |
*) This action was taken only if PillCam® Colon had not been excreted by this time. |
Quality of colon preparation was assesed using a 3 point scale in both CCE and CSPY, Table 2.
Table 2.
Grade | Description |
---|---|
1 | clean, small amount of feces/dark fluid, good assessability |
2 | moderate preparation due to remaining feces/dark fluid, limited assessability but clean enough to preclude significant lesions |
3 | poor preparation, enough feces/fluid to strongly limit assessability |
Statistical analysis
CCE was compared to CSPY which was considered to be the gold standard. As this is a pilot study, calculation regarding statistical power did not apply. Sensitivity, specificity, positive predictive value and negative predictive value were calculated per patient, for the included number of patients (n = 56). Percentage values were rounded to the nearest full number. Examinations were excluded from statistical analysis if the capsule had not reached the colon during the recording time. The equivalence between the results of CCE and CSPY was calculated using Wilcoxon signed-rank test and χ2-test. A p-value < 0.05 was considered statistically significant. Baseline characteristics, cause of referral, colonoscopic and capsule-endoscopic findings, technical data and results of the patient questionnaire were documented on a datasheet using Excel (Microsoft, Redmont, Washinton, USA). Statistical analyses were performed using SPSS software package (SPSS Inc., Chicago, Illinois, USA).
Results
Patients' characteristics
A total of 59 Patients were enrolled in this study. Three patients had to be excluded from the data analysis because the capsule did not reach the colon during examination time (1 remained in stomach, 2 in small bowel).
The exams of 56 patients (34 male, 22 female, mean age 60 years, median 59 years, range 38-84 years, 86% 50 years or older) were analyzed, indications are shown in table 3.
Table 3.
Screening | 23 (41%) |
Family history | 4 (7%) |
Follow-up after polypectomy | 3 (5%) |
Abdominal pain | 8 (14%) |
Bleeding frank | 8 (14%) |
Anemia | 2 (4%) |
Change in bowel habits | 6 (11%) |
Diarrhea | 1 (2%) |
Weight loss | 1 (2%) |
Completion of examinations and technical problems
The capsule was excreted within 10 hours after ingestion and within battery duration in 36 patients (64%). In 2 (4%) cases wake-up was too late (past the ileocoecal valve) and in 2 (4%) transmission interference occurred from capsule to data recorder, probably due to external interference (multiple nominations). The mean colon recording time in the 36 complete examinations was 3 hours 9 minutes (median 2 hours 48 minutes). Mean and median transmission times in all 56 subjects were 4 hours 35 minutes and median 4 hours 52 minutes, respectively.
CSPY was complete in all 56 patients, with a rate of 68% (n = 38) for intubation of the terminal ileum.
Colon Preparation
Level of cleansing on CCE was good in 15 (27%), moderate in 30 (54%) and poor in 11 (20%) cases. On CSPY, 7 (13%) subjects were well prepared, 38 (68%) moderately and 11 (20%) poorly. 34 (61%) of the patients were reported to have the same cleansing level in both colonoscopy and CCE, 22 (39%) had different cleansing levels in both examinations.
Polyp detection
Polyp detection rate (per-patient) was 50% (n = 28) for CSPY and 62% (n = 35) for CCE. Significant size polyps were diagnosed in 6 patients (11%) on CSPY, resp. 15 patients (27%) on CCE. 11% (n = 6) patients had polyps of any size on CSPY that were not detected on CCE (miss rate) (Wilcoxon signed-rank test p > 0.05: not significant).
13/56 (23%) patients had findings of any size on CCE that were not verified on CSPY, 2 (4%) were of significant size. Patients did not undergo repeat colonoscopy.
For polyps of any size, CCE showed a sensitivity of 79% (95% CI, 61 to 90), specificity 54% (95% CI, 35 to 70), PPV of 63% and NPV of 71% for polyps of any size. For overall polyp size, detection of polyps on CCE and on CSPY was independent with statistical significance (p = 0.013) on Pearson's χ2-Test, indicating differences in the detection rate for polyps on both examinations (nominal data). For relevant polyps (>5 mm) there was a correspondence in the detection rates of both methods (p > 0.05). The sensitivity was 50% (95% CI, 19 to 81), the specificity was 76% (95% CI, 63 to 86), the PPV was 20% and the NPV was 93%, Table 4.
Table 4.
For polyps of any size | For polyps >5 mm | |
---|---|---|
Sensitivity % (95% CI) | 79 (61 - 90) | 50 (19 - 81) |
Specificity % (95% CI) | 54 (35 - 70) | 76 (36 - 86) |
Positive predictive value % | 63 | 20 |
Negative predictive value % | 71 Χ2: p = 0.013 |
93 Χ2: p > 0.05 |
Presentation of different size-polyps on CCE/CSPY is shown in Figure 2.
Effect of cleansing level on polyp detection
To assess whether the cleansing level determined accuracy of polyp detection, we subanalyzed the 15 (27%) patients classified as good cleanliness level and the 11 (20%) with a pooply prepared bowel. Sensitivity did not differ significantly in respect to cleansing, although PPV and NPV were both 100% in good bowel preparation.
Analysis according to histopathology
Subanalysis with respect to histopathology revealed excellent sensitivity for CCE for detection of significant size tubular and tubulo-villous adenoma.
Per-patient-prevalence of adenoma was 27% (n = 15) on CSPY. Numbers are too small to calculate for sensitivity and specificity, but all (n = 3) of the detected tubulo-villous adenoma were detected by CCE. Tubular adenoma were detected in 18% (a total of 10 patients on CSPY, all size), one the two detected tubular adenoma of significant size on CSPY was classified as ≤5 mm on CCE. One of the two detected serrated adenoma was not seen on CCE.No high-grade dysplasia or cancerous lesion was found on either type of examination. Overall prevalence of hyperplastic polyps was 23% (n = 13). A detailed analysis regarding number, size, location and histology of polyps is shown in Table S6 (additional file 2).
Adverse events
One patient had an allergic skin reaction to the adhesive tape of the electrodes during CCE. One patient presented with abdominal pain after polypectomy (during CSPY). A colonic perforation was ruled out.
Patients' acceptance
Patients' acceptance was assessed from the given questionnaires. Out of a total of 56 questionnaires, 53 (95%) were returned. No patient reported troubles swallowing the capsule. 16 patients (30%) felt restricted during daily-life activities by carrying the electrodes and the recorder. 21 (40%) patients preferred CCE to CSPY, while 20 (38%) preferred CSPY and 12 (23%) had no preference.
Also, 50 (94%) patients would recommend CCE. 44 (83%) subjects would prefer to undergo a capsule colonoscopy again in ten years' time for screening purposes compared to 8 patients (15%) who declined and 1 patient (2%) who was indecisive.
Discussion
Overall, this single center pilot trial comparing CCE to CSPY for screening of colorectal cancer showed feasible results in imaging for colonic polyps.
Detection rate of polyps of any size on CSPY was 50%, with an adenoma detection rate of 27% (any size).Those are comparable to a large prospective, community-based study[22].
For the primary endpoint, CCE had a low sensitivity (50%). The used size graduation for significance (as defined as >5 mm in our study and >6 mm/>3 polyps of any size in the previous two pilot studies), is debatable: these definitions are adaptations from studies on virtual colonoscopy (CT and MR colonography) where the significance level was set at >10 mm, showing a sensitivity > 75% and specificity > 90% for both examinations[23-25]. To date, there is evidence that virtual colonoscopy is not an accurate alternative for diagnostically relevant small polyps[26].
Additional polyp detection (of significant size) was 4%. We partly ascribe this to altered size estimation and multiple detection of polyps on CCE[27], and to the generally low detection rate of significant size polyps on colonoscopies in our study (assumption from experience).
However, although numbers are small, CCE showed excellent detection of significant size adenoma (tubuluar and tubulo-villous).
From the 41/56 patients that were diagnosed positive (polyps of any size) on colonoscopy, 6 patients had a negative CCE. Reasons for this miss rate have been discussed above, but clearly, CCE needs improvement to the gold standard.
Several findings on CCE (23%, all polyp size) were not confirmed on CSPY. This could also be a false negative rate for CSPY. As CSPY was referred to as the gold standard, we did not include a second look endoscopy to assess this finding, nor did we analyze underlying patterns why polyps were missed on CSPY.
The negative predictive value of 93% for significant polyps fulfills demands for a screening exam[28], although the meaning is altered by a PPV of 20 to 63%.
A recently published study[20] showed comparable results for sensitivity and negative predictive value, although patients with known history of both polyps and colon carcinoma were included.
Second, the preparation regimen suggested by Given Imaging has much been improved, though it still remains more intense than bowel preparation for colonoscopy.
We found a lower excretion rate for CCE (64%, n = 36) than the two previous pilot studies. This might be caused by an additional fasting time of almost 4 hours between ingestion of the second 2 liters of Macrogolum and initiation of CCE, as motility studies have shown an enhanced colonic propulsion of the capsule through PEG. Also, the Israelian study[18] was able to use Tegaserod (Zelmac®, Novartis, Basel, Switzerland) which in Switzerland was discontinued in October 2001.
In our study, we found a good level of cleansing only in 37% (n = 15) patients on CCE. As we regularly use a three-point-scale for graduation of cleansing level this was adapted to CCE for better utility.
We also had a low percentage of good cleansing grades for CSPY (13%, n = 7). Our results for poor bowel preparation are consistent with other reports on colon cleansing, where the mean percentage of poorly cleaned bowels varies between 20-25% (20%, n = 11, in our study)[30,31], but in marked contrast to previous pilot studies that showed an excellent/good cleansing level of 88%[17] resp. 84.4%[18].
Grades for cleansing on CSPY are not described in the two mentioned pilot studies. We do explain our findings with the altered assessment scale and, as mentioned above, at least for CCE, with the break in between ingestion of the second 2 liters of PEG and ingestion of the capsule.
Sensitivity for CCE was not improved by good levels-of-cleansing[17,18,20], however, this might be due to a low number of patients in this subanalysis.
Bowel preparation is also crucial for an efficient colonoscopy. Inadequate preparation correlates with higher polyp miss rates and higher costs due to lower completion rates with need for repeated colonoscopy[25,11].The ideal bowel preparation for CCE is yet to be defined, and timing seems to be important. As for now, if preparation for capsule is more complex than for CSPY, with the procedure itself bearing a risk of volume overload especially in predisposed patients, how large is the benefit of CCE?
Third, 8-10 hours of validated capsule battery duration were not sufficient for all patients, especially those with transit disorders. In a recent motility study for patients with severe intestinal motor disorders, using PillCam SB[32], it was shown, that occurrence of less contractile activity, more static periods and also more content retention was statistically significant. Although these findings apply to small bowel observations, it seems probable to translate affected intestinal motility to the colon and rectum, therefore emphasizing the need for a prolonged battery time.
Fourth, CCE has a good acceptance: 94% of patients (n = 50) would recommend CCE as a screening method to somebody who has so far rejected undergoing screening by CSPY, and 83% (n = 44) would adhere to the suggested screening regimen suggesting a 10-year repeat CCE. Interestingly, on postinterventional assessment, 20 patients (38%) preferred CSPY to CCE which is contradictory to preinterventional data[33-35] and might be due to propofol usage for sedation[36].
Conclusion
In summary, PillCam Colon provides a screening solution which is minimally invasive, safe and does not require sedation. It is very well accepted by patients and recommended to people who have so far denied CRC screening programs. It is an easy to perform examination with an excellent negative predictive value for application in screening purposes under routine conditions. However, diagnostic accuracy for relevant size polyps (i.e. sensitivity) is low. In order to have an assigned part in CRC screening, capsule endoscopy needs to overcome its main limitations, i.e. adaption in preparation and optimizing of technical aspects.
Competing interests
The authors declare that they have no competing interests.
Authors' contributions
JBP designed the study and its concept, obtained funding, acquired, analyzed and interpreted the data, performed the statistical analysis and drafted the manuscript. SP acquired, analyzed and interpreted the data and performed the statistical analysis. SP participated in analysis and interpretation of data and administratively and technically supported the study. CB designed the study and its concept, administratively, technically and materially supported the study, supervised it and critically revised the manuscript for important intellectual content. LD administratively, technically and materially supported the study, supervised it and critically revised the manuscript for important intellectual content. All authors read and approved the final manuscript.
Pre-publication history
The pre-publication history for this paper can be accessed here:
Supplementary Material
Contributor Information
Julia B Pilz, Email: julia.pilz@ksli.ch.
Susanne Portmann, Email: susanne.portmann@stud.unibas.ch.
Shajan Peter, Email: shajan@uab.edu.
Christoph Beglinger, Email: beglinger@tmr.ch.
Lukas Degen, Email: ldegen@uhbs.ch.
Acknowledgements
The authors thank Prof. Dr. Jürgen Drewe, University Hospital Basel, for review of statistical analysis.
References
- Parkin DM, Whelan SL, Ferlay J, Raymand L, Young J, (eds) Cancer incidence in five continents. Lyon: VII. IARC Scientific Publications; 1997. International Agency for Research on Cancer. WHO and International Association of cancer registries. [Google Scholar]
- Krebsmortalität in der Schweiz. (Last accessed 10/15/2009. http://asrt.ch/asrt/newstat/in4ch8605.pdf
- Hawk ET, Levin B. Colorectal cancer prevention. J Clin Oncol. 2005;23(2):378–91. doi: 10.1200/JCO.2005.08.097. [DOI] [PubMed] [Google Scholar]
- Winawer SJ, Fletcher RH, Miller L, Godlee F, Stolar MH, Mulrow CD, Woolf SH, Glick SN, Ganiats TG, Bond JH, Rosen L, Zapka JG, Olsen SJ, Giardello FM, Sisk JE, Van Antwerp R, Brown-Davis C, Mrciniak DA, Mayer RJ. Colorectal cancer screening - clinical guidelines and rationale. Gastroenterology. 1997;112:594–642. doi: 10.1053/gast.1997.v112.agast970594. [DOI] [PubMed] [Google Scholar]
- Winawer SJ, Zauber AG, O'Brien MJ, Ho MT, Gottlieb L, Sternberg SS, Waye JD, Bond JH, Schapiro M, Panish JF, Ackroyd F, shike M, Kurtz RC, Hornsby-Lewis L, Gerdes H, Stewart ET. Prevention of colorectal cancer by colonoscopic polypectomy. N Engl J Med. 1993;329(27):1977–81. doi: 10.1056/NEJM199312303292701. [DOI] [PubMed] [Google Scholar]
- Thiis-Evensen E, Hoff GS, Sauar J, Langmark F, Majak BM, Vatn MH. Population-based surveillance by colonoscopy: effect on the incidence of colorectal cancer. Telemark Polyp Study I. Scand J Gastroenterol. 1999;34(4):414–20. doi: 10.1080/003655299750026443. [DOI] [PubMed] [Google Scholar]
- American Cancer Society available at. http://www.cancer.org
- Levin B, Lieberman DA, McFarland B, Andrews KS, Brooks D, Bond J, Dash C, Giardello FM, Glick S, Johnson D, Johnson CD, Levin TR, Pickhardt PJ, Rex DK, Smith RA, Thornson A, Winawer SJ. Screening and surveillance for the early detection of colorectal cancer and adenomatous polpys, 2008: A joint guideline from the American Cancer Society, the US Multi-Society Task Force on Colorectal Cancer, and the American College of Radiology. Gastroenterol. 2008;134(5):1570–95. doi: 10.1053/j.gastro.2008.02.002. [DOI] [PubMed] [Google Scholar]
- Rex DK, Johnson DA, Anderson JC, Schoenfeld PS, Burke CA, Inadomi JM. American College of Gastroenterology. American College of Gastroenterology guidelines for colorectal cancer screening 2009 (corrected) Am J Gastroenterol. 2009;104(3):739–50. doi: 10.1038/ajg.2009.104. [DOI] [PubMed] [Google Scholar]
- Rex DK, Cutler CS, Lemmel GT, Rhamani EY, Clark DW, Helper DJ, Lehman GA, Mark DG. Colonic miss rates of adenomas determined by back-to-back colonoscopies. Gastroenterology. 1997;112(1):24–28. doi: 10.1016/S0016-5085(97)70214-2. [DOI] [PubMed] [Google Scholar]
- Bressler B, Paszat LF, Vinden C, Li C, He J, Rabeneck L. Colonoscopic miss rates for right-sided colon cancer: a population-based analysis. Gastroenterology. 2004;127(2):452–6. doi: 10.1053/j.gastro.2004.05.032. [DOI] [PubMed] [Google Scholar]
- Frühmorgen P. Aktuelle Gastroenterologische Diagnostik. Berlin: Springer; 1985. Rekto- und Koloskopie. In: Blum AL, Siewert JR, Ottenjann R, Lehr L (eds) pp. 266–275. [Google Scholar]
- Rex DK, Imperiale TF, Latinovich DR, Bratcher LL. Impact of bowel preparation on efficiency and cost of colonoscopy. Am J Gastroenterol. 2002;97(7):1696–700. doi: 10.1111/j.1572-0241.2002.05827.x. [DOI] [PubMed] [Google Scholar]
- Peterson NB, Murff HJ, Ness RM, Dittus RS. Colorectal cancer screening among men and women in the United States. J Womens Health (Larchmt) 2007;16(1):57–65. doi: 10.1089/jwh.2006.0131. [DOI] [PubMed] [Google Scholar]
- Schoepfer A, Marbet UA. Colonoscopic findings fo symptomatic patients aged 50 to 80 years suggest that work-up of tumour suspicious symptoms hardly reduces cancer-induced mortality. Swiss Med Wkly. 2005;135(45-46):679–83. doi: 10.4414/smw.2005.11033. [DOI] [PubMed] [Google Scholar]
- Oncosuisse. http://www.oncosuisse.ch
- Schoofs N, Devière J, Van Gossum. PillCam colon capsule endoscopy compared with colonoscopy for colorectal tumor diagnosis: a prospective pilot study. Endoscopy. 2006;38(10):971–7. doi: 10.1055/s-2006-944835. [DOI] [PubMed] [Google Scholar]
- Eliakim R, Fireman Z, Gralnek IM, Yassin K, Waterman M, Kopelman Y, Lachter J, Koslowsky B, Adler SN. Evaluation of the PillCam Colon capsule in the detection of colonic pathology: results of the first multicenter, prospective, comparative study. Endoscopy. 2006;38(10):963–70. doi: 10.1055/s-2006-944832. [DOI] [PubMed] [Google Scholar]
- Sieg A, Friedrich K, Sieg U. Is PillCam COLON Capsule Endoscopy ready for colorectal cancer screening? A prospective feasibility study in a Community gastroenterology practice. Am J Gastroenterol. 2009;104:848–854. doi: 10.1038/ajg.2008.163. [DOI] [PubMed] [Google Scholar]
- Van Gossum A, Munoz-Navas M, Fernandez-Urien I, Carretero C, Gay G, Delvaux M, Lapalus MG, Ponchon T, Neuhaus H, Philipper M, Costamagna G, Riccioni ME, Spada C, Petruzziello L, Fraser C, Postgate A, Fitzpatrick A, Hagenmuller F, Keuchel M, Schoofs N, Devière J. Capsule endoscopy versus colonoscopy for the detection of polyps and cancer. N Engl J Med. 2009;361(3):264–70. doi: 10.1056/NEJMoa0806347. [DOI] [PubMed] [Google Scholar]
- Heuss LT, Schnieper P, Drewe J, Pflimlin E, Beglinger C. Risk stratification and safe administration of propofol by registered nurses supervised by the gastroenterologist: a prospective observational study of more than 2000 cases. Gastrointest Endosc. 2003;57(6):664–71. doi: 10.1067/mge.2003.191. [DOI] [PubMed] [Google Scholar]
- Barclay RL, Vicari JJ, Doughty AS, Johanson JF, Greenlaw. Colonic withdrawal time and adenoma detection durng screening colonoscopy. N Engl J Med. 2006;355(24):2533–41. doi: 10.1056/NEJMoa055498. [DOI] [PubMed] [Google Scholar]
- Hara AK, Johnson CD, Reed JE, Ahlquist DA, Nelson H, MacCarty RL, Harmsen WS, Ilstrup DM. Detection of colorectal polyps with CT colography: initial assessment of sensitivity and specificity. Radiology. 1997;205:59–65. doi: 10.1148/radiology.205.1.9314963. [DOI] [PubMed] [Google Scholar]
- Schoenenberger AW, Bauerfeind P, Krestin GP, Debatin JF. Virtual colonoscopy with magnetic resonance imaging: in vitro evaluation of a new concept. Gastroenterology. 1997;112:1863–70. doi: 10.1053/gast.1997.v112.pm9178678. [DOI] [PubMed] [Google Scholar]
- Royster AP, Fenlon HM, Clarke PD, Nunes DP, Ferrucci JT. CT colonoscopy of colorectal neoplasms: Two-dimensional and three-dimensional virtual-reality techniques with colonoscopic correlation. Am J Roentgen. 1997;169:1237–42. doi: 10.2214/ajr.169.5.9353434. [DOI] [PubMed] [Google Scholar]
- Rosman AS, Korsten MA. Meta-analysis comparing CT-colonography, air-contrast barium enema, and colonoscopy. Am J med. 2007;120(3):203–210. doi: 10.1016/j.amjmed.2006.05.061. e4. [DOI] [PubMed] [Google Scholar]
- Brinch K, Larsson HB, Madsen JL. A deconvolution technique for processing small intestine transit data. Eur J Nucl Med. 1999;26(3):272–6. doi: 10.1007/s002590050388. [DOI] [PubMed] [Google Scholar]
- Rösch T, Eickhoff A, Fritscher-Ravens A, Eliakim R, Arber N. The new scopes - broadening the colonoscopy marketplace. Digestion. 2007;76(1):42–50. doi: 10.1159/000108393. [DOI] [PubMed] [Google Scholar]
- Fireman Z, Kopelman Y, Fish L, Sternberg A, Scapa E, Mahaina E. Effect of oral purgatives on gastric and small bowel transit time in capsule endoscopy. Isr Med Assoc J. 2004;6(9):521–3. [PubMed] [Google Scholar]
- Froehlich F, Wietlisbach V, Gonvers JJ, Burnand B, Vader JP. Impact of colonic cleansing on quality and diagnostic yield of colonoscopy: the European Panel of Appropriateness of Gastrointestinal Endoscopy European Multicenter Study. Gastrointest Endosc. 2005;61(3):378–84. doi: 10.1016/S0016-5107(04)02776-2. [DOI] [PubMed] [Google Scholar]
- Ness RM, Manam R, Hoen H, Chalasani N. Predictors of inadequate bowel preparation for colonoscopy. Am J Gastroenterol. 2001;96(6):1797–802. doi: 10.1111/j.1572-0241.2001.03874.x. [DOI] [PubMed] [Google Scholar]
- Malagelada C, De Iorio F, Azpiroz F, Accarino A, Segui S, Radeva P, Malagelada JR. New insight into intestinal motor function via noninvasive endoluminal image analysis. Gastroenterology. 2008;135(4):1155–62. doi: 10.1053/j.gastro.2008.06.084. [DOI] [PubMed] [Google Scholar]
- Wee CC, McCarthy EP, Phillips RS. Factors associated with colon cancer screening: the role of patient factors and physician counseling. Prev Med. 2005;41(1):23–9. doi: 10.1016/j.ypmed.2004.11.004. [DOI] [PubMed] [Google Scholar]
- Meissner HI, Breen N, Klabunde CN, Vernon SW. Patterns of colorectal cancer screening uptake among men and women in the United States. Cancer Epidemiol Biomarkers Prev. 2006;15(2):389–394. doi: 10.1158/1055-9965.EPI-05-0678. [DOI] [PubMed] [Google Scholar]
- Tangka FK, Molinari NA, Chattopadhyay SK, Seeff LC. Market for colorectal cancer screening by endoscopy in the United States. Am J Prev Med. 2005;29(1):54–60. doi: 10.1016/j.amepre.2005.03.006. [DOI] [PubMed] [Google Scholar]
- Lazzaroni M, Bianchi Porro G. Preparation, premedication and surveillance. Endoscopy. 2005;37(2):101–9. doi: 10.1055/s-2004-826149. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.