Multibeam field emission x-ray system with half-scan reconstruction algorithm

Yang Lu 1,a), Hengyong Yu 2, Guohua Cao 3, Jun Zhao 4, Ge Wang 5, Otto Zhou 6
PMCID: PMC2909302  PMID: 20831085

Abstract

Purpose: In this article, the authors propose a multibeam field emission x-ray (MBFEX) system along with a half-scan fan-beam reconstruction algorithm.

Methods: The proposed system consists of a linear CNT-based MBFEX source array, a single large area detector that is divided into same number of segments as the number of x-ray beams, a multihole collimator that aligns each beam with a corresponding detector segment, and a sample rotation stage. The collimator is placed between the source and the object to restrict the x-ray radiations through the target object only. In this design, all the x-ray beams are activated simultaneously to provide multiple projection views of the object. The detector is virtually segmented and synchronized with the x-ray exposure and the physiological signals when gating is involved. The transmitted x-ray intensity from each beam is collected by the corresponding segment on the detector. After each exposure, the object is rotated by a step angle until sufficient data set is collected. The half-scan reconstruction formula for MBFEX system is derived from the conventional filtered backprojection algorithm. To demonstrate the advantages of the system and method in reducing motion artifacts, the authors performed simulations with both standard and dynamic Shepp–Logan phantoms.

Results: The numerical results indicate that the proposed multibeam system and the associated half-scan algorithm can effectively reduce the scanning time and improve the image quality for a time-varying object.

Conclusions: The MBFEX technique offers an opportunity for the innovation of multisource imaging system.

Keywords: multibeam field emission x-ray system (MBFEX), half-scan algorithm, image reconstruction

INTRODUCTION

Microcomputed tomography (micro-CT) has recently emerged as a powerful noninvasive imaging tool for biomedical imaging of small animals, which has increasingly played a critical role in phenotyping, drug discovery, and in providing a basic understanding of mechanisms of disease. However, motion-induced artifacts blur the micro-CT image and result in deteriorated spatial resolution than the nominal values. The physiological motions of a mouse are about ten times faster than humans. To minimize the motion blurs, at least ten times faster temporal resolution than that of humans is necessary. This poses a particular challenge for micro-CT system design because it requires not only short exposure (hence high temporal resolution) and high fluency rate, but also small focal spot size to achieve high system resolution without compromising the imaging quality.1

To address the abovementioned conflicting requirements for both high spatial and temporal resolution, Dr. Otto Zhou’s group at the University of North Carolina at Chapel Hill recently developed a dynamic micro-CT scanner based on a carbon nanotube (CNT) field emission microfocus x-ray source that can potentially deliver spatial resolution better than 100 μm and 50 ms temporal resolution.2 Compared to the conventional thermionic microfocus x-ray sources, a primary advantage of the field emission x-ray source is its electronic programmability and the easiness for gating, even for fast nonperiodic physiological signals, including respiratory and cardiac motions. Very recently, the performance of this scanner has been improved to 80 μm spatial resolution and 15 ms temporal resolution.3 The key idea is that with programmability of the field emission x-ray source, the motion-induced artifacts can be minimized by prospective gating even for free-breathing mice, i.e., the image acquisition is synchronized with the animal’s physiological motions. With prospective gating, however, the micro-CT scanning time becomes much longer. To acquire a single 3D data set, prospective and respiratory gated micro-CT normally takes about 10 min. In a recent study with cine cardiac micro-CT, in which as many as ten data sets were acquired of the same mouse corresponding to ten phases in a cardiac cycle, the total scanning time was as long as a few hours.4 With such long scanning time, the imaged animal is under more stress and becomes more difficult to handle, which in turn worsens the reproducibility of gating, and therefore compromises the image quality.

The long scanning time with prospectively gated micro-CT can be greatly reduced using the multisource configuration. In history, the dynamic spatial reconstructor5 used 14 x-ray tubes mounted on a gantry around a patient. Although high temporal resolution was demonstrated for cardiac imaging, the system is practically undesirable in terms of cost, size, and maintenance. Nowadays, most of current micro-CT systems use only a single source-detector pair and one projection is acquired within one physiological cycle. Recent technological development has led to some dual source-detector micro-CT systems (SkyScan, Model 1178, Kartuizersweg 3B, 2550 Kontich, Belgium and VAMP GmbH, TomoScope, Henkestrabe 91, 91052 Erlangen, Germany) that can reduce the scanning time by a factor of 2 because two projections can be taken simultaneously. In principle, if N projection images can be taken simultaneously, the scanning time can be reduced by a factor of N. However, we have not seen any micro-CT system with more than two conventional thermionic x-ray sources. This is partially due to the high cost associated with multiple (≥3) conventional thermionic x-ray sources and limited space for integration of many source-detector pairs. Interestingly, research in the past few years indicates that the CNT x-ray technology allows effective control of x-ray beams both temporally and spatially.6 The feasibility of utilizing the CNT-based multibeam field emission x-ray (MBFEX) technology for stationary or multibeam tomography imaging systems has been demonstrated for digital breast tomosynthesis,7 real-time tomosynthesis guidance for radiation therapy,8 and novel computed tomography geometry.9 Recently, a proof-of-concept multibeam micro-CT scanner operating in the sequential imaging mode has also been demonstrated.10

Here we propose to develop a high-speed micro-CT system using the CNT-based field emission x-ray source array. The key benefit of the proposed system is to simultaneously collect multiple projections, thus proportionally shortening the imaging time for prospective gated small animal imaging. In this paper, we derive a half-scan fan-beam reconstruction formula for this new imaging geometry and evaluate the new formula with numerical simulations.

This paper is organized as follows. In Sec. 2, we will discuss the configuration of the MBFEX micro-CT system, including characteristics of the MBFEX system, and the detector segmentation for different x-ray beams. In Sec. 3, we will derive the formulas for projection rebinning. In Sec. 4, we will perform three numerical experiments to showcase the advantages of the proposed half-scan algorithm. In Sec. 5, we will discuss the related issues and conclude the paper.

SYSTEM DESIGN

System development

Ideally, it is best to develop a stationary micro-CT scanner using hundreds of distributed x-ray beams and multiple detectors.11 However, development of such a stationary scanner is limited by the practical constraints with available x-ray source and detector technology, the lack of suitable image reconstruction algorithm, and the potential issues with system geometry and scattering. Therefore, we propose to develop a micro-CT system that has a few distributed x-ray beams and requires only limited steps of rotation. As illustrated in Fig. 1, the proposed system consists of a linear CNT-based MBFEX source array, a single large area detector that is divided into the same number of segments as the number of x-ray beams, a multihole collimator that aligns each beam with a corresponding detector segment, and a sample rotation stage. The collimator is placed between the source and the object to restrict the x-ray radiations through the target object only. In this design, all the x-ray beams are activated simultaneously to provide multiple projection views of the object. The detector is virtually segmented and synchronized with the x-ray exposure and the physiological signals when gating is involved. The transmitted x-ray intensity from each beam is collected by the corresponding segment on the detector. After each exposure, the object is rotated by a step angle until sufficient data set is collected.

Figure 1.

Figure 1

The schematic of the MBFEX system.

Detector segmentation

Unlike single source geometry, in this MBFEX system, multiple x-ray pixels share the same detector. Hence, the detector area is virtually split into several subregions corresponding to different x-ray pixels, and each of them does not overlap with others. Furthermore, to maximize the usage of the detector, the gaps between different subregions should be as narrow as possible. The optimal solution is illustrated as in Fig. 2. Given the detector length Ld, the distance Ls between x-ray pixels, and the radius r of ROI, the optimal solution can be achieved through changing the source-detector distance D and the source-object distance R0. A quick calculation shows

R0=DLs+rD2+(Ls+12Ld)2Ls+12Ld. (2.1)

If the number of the x-ray pixels is larger than 3, i.e., 5 or more, Lsi and Ld(i+1), i=2,3…, can be calculated iteratively as (see Appendix A)

Lsi=B1+B124A1C12A1,Ld(i+1)=B2+B224A2C22A2, (2.2)

where A1=(R0D)2r2, B1=2Ldi(R02R0Dr2), C1=(R02r2)L2diD2r2, A2=R02r2, B2=2Lsi(R02R0Dr2), and C2=[(R0D)2r2]L2siD2r2.

Figure 2.

Figure 2

The optimal design of MBFEX system with three x-ray pixels.

RECONSTRUCTION ALGORITHM

Data transform from real detector to virtual detector

In the MBFEX system, the detector is not perpendicular to the central beams of the x-ray pixels S±1. Therefore, the projections collected on the detector are not equal-spatial or equal-angular. For computation accuracy, the projections should be transformed from the real detectors to the virtual detectors, which are perpendicular to the center beams of the x-ray pixels. Let the object center be the coordinate origin, Si=(Ri cos λi,Ri sin λi) be each source position, Pr be any point on the real detector with coordinate ur, Pv be any point on the virtual detector with coordinate uv, Dod be the distance between the object center and the detector, D be the distance between the source S0 and the detector, and Ri,i∊{0,±1} be the source-to-virtual detector distances (see Fig. 3).

Figure 3.

Figure 3

Illustration of the data transform from the real detector to the virtual detector. (a) and (b) are for pixels S1 (S−1) and S0, respectively.

In Cartesian coordinates, Pr=(xr,yr) and Pv=(xv,yv) can be expressed as

{xr=uryr=Dod}and{xv=uv sinλiyv=uv cosλi}. (3.1)

By the aforementioned construction we have

(PrSi)=k(PvSi), (3.2)

where k is a real constant.

Substituting Eq. 3.1 into Eq. 3.2, it yields

{ur+Ri cosλi=k(uv sinλi+Ri cosλi)Dod+Ri sinλi=k(uv cosλi+Ri sinλi)}, (3.3)

which implies

ur=uv sinλi+Ri cosλiuv cosλi+Ri sinλi(Ri sinλi+Dod)Ri cosλi. (3.4)

Any beam irradiated from S0 is equivalent to a beam started from the source S0, which has a larger scanning radius R1 [see Fig. 3b]. Let α and β be the angle between the beam and the central beam for S0 and S0, respectively. By construction,

α=tan1urD,β=tan1uvR1. (3.5)

By the law of sines, we get

sin(πα)R1=sinβR0

and

β=sin1R0 sinαR1. (3.6)

Substituting Eq. 3.6 into Eq. 3.5, we get the relationship between ur and uv. That is,

ur=D tan(sin1R1 sin(tan1uvR1)R0),γ=βα. (3.7)

where γ determines the position of S0.

Reconstruction formula

The reconstruction formula for MBFEX system is derived from the conventional filtered backprojection algorithm.12 It can be stated as follows:

f(x,y)=12λsλeR2(Rs)2dλω(λ,u)g(λ,u)h(RtRsu)RR2+u2du, (3.8)

where λs and λe are the starting and ending points of the backrprojection segment, h is the convolution kernel, g(λ,u) is the projection specified by the angle λ of the scanning trajectory and the coordinate u on the detector, R is the scanning radius, and s and t are the rotated coordinate system defined as

[st]=[cosλsinλsinλcosλ][xy]. (3.9)

The weight function ω is from Parker’s half-scan weighting scheme.13 Necessary changes have been made for the equal-spatial detector.

ω(λ,u)={sin2(π4λδtan1(uR)),0λ2δ2tan1(uR)1,2δ2tan1(uR)λπ2tan1(uR)sin2(π4π+2δλδ+tan1(uR)),π2tan1(uR)λπ+2δ}, (3.10)

where λes+π+2δ and δ=tan−1(|umax|∕R) is half the full fan angle of the central-scanning plane.

The images can be exactly reconstructed as long as the projections cover the angular range Δ=π+2 sin−1(rR1). Two cases are possible (see Fig. 4). For fixed r and R1, if R012(R12rR1), Case A occurs. That is, the x-ray pixel S1 moves from s1 to s1 across s−1 when the pixel S−1 moves from s−1 to s1. Clearly, the trajectories of S1 and S−1 cover the range Δ. Thus we could reconstruct the object without using the projections related to source S0. If R0<12(R12rR1), Case B occurs. That is, the pixel S1 moves from s1 to s1 when the source S−1 moves from s−1 to s1, and arc s1s1 are not covered by the trajectories of S1 and S−1. Notice that the pixel S0 is also moving from s0 to s0 during the scanning. From the discussion in Sec. 3A, the projections from pixel S0 with radius R0 can be transformed to those of a virtual source with radius R1, and the trajectory of the virtual x-ray pixel covers the arc s1s1. Hence, the missing projections can be compensated by the transformed data over the arc s1s1.

Figure 4.

Figure 4

Two possible cases when using half-scan algorithm. (a) Case A: The trajectories of S1 and S−1 cover the minimum angle range; (b) Case B: The trajectories of S0, S1, and S−1 cover the minimum angle range.

NUMERICAL EXPERIMENTS

To demonstrate the advantages of the proposed system and reconstruction scheme, we designed three numerical experiments with the well-known Shepp–Logan phantom. In the simulation, the MBFEX source had three x-ray pixels. The length of the detector was 300 mm for Case A and 550 mm for Case B. Each detector had 800 pixels per row. Given the object size r=35 mm, the source to detector distance D, the x-ray pixel to x-ray pixel distance Ls, and the source to object distance R0 can be calculated by Eq. 2.1. Then, we determined that if Case A (Case B) will occur by the condition discussed in Sec. 3B. The parameters used in the simulations were listed in Table 1. Actually, we should avoid the occurrence of Case B in practice because it may lead to a very large size of detector or the MBFEX source will be too close to the object. For all the simulations, we assumed that the x-ray pixels were identical (focal spot sizes, current, and kVp) to each other except for the locations.

Table 1.

Parameters for numerical experiments.

  Case A Case B
D 800 mm 450 mm
R0 600 mm 350 mm
Ls 292.5 mm 568.5 mm
Ld 300 mm 550 mm
r 35 mm 35 mm
Detector elements per row 800 800
Projections per round 800 800

In the first experiment, we used conventional half-scan algorithm to reconstruct the image, i.e., the data transform step discussed in Sec. 3A is omitted. A reconstruction with our algorithm was also performed to verify the correctness of the proposed method. From Fig. 7, we observed that the reconstructed image was severely distorted, which indicated that x-ray source to detector element variations had great impact on the reconstruction. Fortunately, this issue can be readily fixed by employing a data rebinning process, as discussed in Sec. 3A. After introducing the virtual detectors, different x-ray pixels can be seen as one x-ray source with different starting positions, so the reconstruction step can be treated as what we do for single source case.

Figure 7.

Figure 7

Reconstructed images of the stationary Shepp–Logan phantom and their profiles. The projections were collected in the same MBFEX system for Case A, but the reconstruction algorithm for (a) was the conventional half-scan algorithm and for (c) was the proposed algorithm. (b) and (d) are profiles along y=0 and the dotted and solid lines are for the phantom and the reconstructed image, respectively. The display window is [0.99, 1.04].

In the second experiment, noise was considered when generating the projections. According to the Lambert–Beer law, the number of transmitted photons obeys the Poisson distribution and it is mainly determined by the number of photons emitted from the x-ray source to each detector cell. Setting the photon number N at different dose levels and following the steps described in Ref. 14, we obtained the projections associated with the Poisson noise, then reconstructed the image with our proposed algorithm. There was no observable difference in the simulation results between the MBFEX system and single source system at the same dose level (see Fig. 8).

Figure 8.

Figure 8

Images reconstructed for a stationary Shepp–Logan phantom with different noise levels. Photon numbers N for left, middle, and right columns are 1.0×108, 1.0×107, and 1.0×106, respectively. The first row was reconstructed by the proposed algorithm for Case A, while the second row by the conventional half-scan algorithm for single source case. The display window is [0.99, 1.04].

The last experiment was to demonstrate that the MBFEX system is an effective way to reduce the scanning time, thus improving the image quality. We simulated a modified dynamic Shepp–Logan phantom, of which an eclipse centered at (7.7 mm, 0) rotated clockwise at 40° per round.

The algorithm was implemented as follows:

  • (1)

    Projection generation. Assuming that the phantom was centered at the origin, the MBFEX source and the detector were rotated simultaneously and continuously. The exposure time for each x-ray pixel is short enough to be ignored (compared to the detector readout time). 800 views were uniformly distributed over 2π for a complete cycle scanning. At each view, projections (or the line integral of the phantom density) were calculated by analytic method for every detector pixel (see the first row of Fig. 5). In Fig. 5, the horizontal axis and the vertical axis in each image are for different views and detector bins, respectively.

  • (2)

    For pixels S1 and S−1, use Eq. 3.4 to transform the projections from the real detector to the virtual detector. For source S0, use Eq. 3.7 to do the transformation. See the second row of Fig. 5.

  • (3)

    Compare the value of 12(R12rR1) and R0, then determine the required projections of each source; assume the starting angle of S1 is 0. In Case A, the required data are [0,2 cos−1(R0R1)] for S1 and [2 cos−1(R0R1),Δ] for S−1. In Case B, the required data are [0,Δ−2 cos−1(R0R1)], [2 cos−1(R0R1),Δ], and [Δ−2 cos−1(R0R1),2 cos−1(R0R1)] for pixels S1, S−1, and S0, respectively.

  • (4)

    Combine the separated projections into one complete data set and apply the half-scan weighting function 3.10 (see Fig. 6). The projections are combined in sequence, i.e., first S1, then S0 and S−1. The edges are visible if the projections are combined without rebinning.

  • (5)

    Use Eq. 3.8 to reconstruct the object.

Figure 5.

Figure 5

Projections on the real detectors and virtual detectors for Case B. From left to right, the columns are projections distributed on [0,0.385π], [0.325π,0.710π] ([0.385π,0.649π] after rebinning), and [0.649π,1.034π]. They correspond to x-ray pixels S1, S0, and S−1, respectively. The first row is for the projections collected on the real detectors, while the second row is for the rebinned projections on the virtual detectors. After rebinning in each image, the horizontal axis is for the views ranging from 0 to 2π and the vertical axis is for the detector bins.

Figure 6.

Figure 6

Combined projections for half-scan reconstruction. Only 414 views were used in the reconstruction (153 views of S1, 155 views of S−1, and 106 views of S0). Left: Combined projections without rebinning; right: Combined projections after rebinning.

Simulation results are shown in Figs. 789. For comparison, in Fig. 9, images reconstructed by the conventional full-scan, half-scan algorithms from single source system are also displayed. All the images were obtained without any motion correction process. Clearly, although the motion artifacts were still visible, they were significantly reduced with the optimized MBFEX system configuration.

Figure 9.

Figure 9

Images reconstructed by (a) conventional full-scan algorithm; (b) conventional half-scan algorithm; (c) proposed half-scan algorithm with MBFEX system for Case A; and (d) proposed half-scan algorithms with MBFEX system for Case B. The display window is [0.99, 1.04].

DISCUSSION AND CONCLUSION

Compared to the single source geometry, the reduction of scanning time in the MBFEX system is not only determined by the algorithm we use but also affected by the ratio of Ls and R0 (or the angle φ formed by pixels S−1, S1 and the center of the object). Let ρ be the scanning time reduction; then for the full-scan algorithm ρf=φ∕(2π) and for half-scan algorithm ρh=φ∕(π+2 sin−1(rR1)). Clearly, the larger φ is, the less scanning time will be, and the proposed MBFEX system associated with the half-scan algorithm can effectively reduce the scanning time.

Although the proposed algorithm has greatly reduced the redundant data, the data redundancy is still inevitable. For example, in Fig. 4a, the data corresponding to the angular range s1s1 is obtained by both pixels S−1 and S1, but only one of them is required for the algorithm. Recent research shows the utilization of the redundant data may help eliminating the artifacts and improving the noise uniformity,15, 16 thus yielding higher image quality. In the MBFEX system, a simple and effective way to fully utilize the redundant data is to give an equal weight to the overlapping projections, i.e., 1∕2 to the data collected over the range s1s1 in Fig. 4a. At the current stage, we are more interested in developing fan-beam algorithms for MBFEX system. However, the proposed algorithm could be readily extended to the 3D case in the well-known Feldkamp framework. For readers’ interest, the cone-beam algorithm for MBFEX system is stated in Appendix B.

As indicated in Fig. 1, the MBFEX source array in our designs is linear in geometry. Understandably, a MBFEX source array in arc geometry or multisegment arrangement can provide better system geometry, which can result into more uniform geometry across different beams and potentially packing more beams with the same detector size. The reason that a linear MBFEX source instead of a curved source is used is the fabrication cost. The benefits from a curved MBFEX source are much more outweighed by the engineering costs. Alternatively, since MBFEX technology can tightly pack multiple x-ray pixels into a single linear x-ray tube, one might wonder why not simply use one such linear MBFEX tube tightly packed with dozens of x-ray pixels and turn on the desired x-ray beams for different designs. Our preliminary studies show the smallest pitch that two microfocused x-ray pixels can be packed together is ≥10 mm. With this constraint in mind, together with the limit on detector size, a linear MBFEX source array with dozens of x-ray pixels is not a good choice because it does not give fine enough freedom in adjusting the locations of x-ray pixels required in the design indicated in Fig. 1.

The MBFEX technique offers an opportunity for the innovation of multisource imaging system. The benefit of MBFEX system, such as its improved image quality for dynamic object, has been discussed in this paper. Other merits or shortcomings, i.e., its performances with compressive-sensing-based interior tomography, noise, and the dose∕scattering reduction related issues will be studied in the future work.

ACKNOWLEDGMENTS

This work was supported in part by the National Science Foundation of China (Grant Nos. 30570511 and 30770589), the National High Technology Research and Development Program of China (Grant No. 2007AA02Z452), the National Basic Research Program of China (Grant No. 2010CB834300), NSF∕MRI Program (Grant No. CMMI-0923297), NSF∕NIBIB (Grant Nos. EB002667 and EB011785), and NIH–NCI (Grant No. U54CA119343).

APPENDIX A: DERIVATION OF EQUATION 2.2

First, let us consider the case when there are only three x-ray pixels. Denote the origin as P0=(0,0), the center of the detector as Pd0=(0,R0D), and Pd2=(−Ld2,R0D), we have the equation of the line Ps1Pd2

y=DLs1+Ld2(x+Ld2)+R0D (A1)

and the distance r between the origin and line Ps1Pd2 is

r=DLd2Ls1+Ld2+R0D(DLs1+Ld2)2+1. (A2)

Equation A2 yields

R0=DLs+rD2+(Ls+12Ld)2Ls+12Ld, (A3)

where the relationships Ld2=12Ld and Ls1=Ls have been used.

Second, let us consider the case when more sources are involved. Let Psi=(Lsi,R0) be the source position, Pdi=(−Ldi,R0D), Pd(i+1)=(−Ld(i+1),R0D), and i=1,2… be the edge of the detector coverage of source Si (see Fig. 10). Hence, we have the line PsiPdi

y=DLsi+Ldi(x+Ldi)+R0D (A4)

and the line PsiPd(i+1)

y=DLsi+Ld(i+1)(x+Ld(i+1))+R0D. (A5)

Since Ldi is known and by the fact that the distance between P0 and line PsiPd(i+1) is r, we can first calculate Lsi by

Lsi=B1+B124A1C12A1, (A6)

where A1=(R0D)2r2, B1=2Ldi(R02R0Dr2), and C1=(R02r2)L2diD2r2.

Figure 10.

Figure 10

The optimal design of the MBFEX system with N (N≥3) x-ray pixels.

Then Ld(i+1) can be computed by

Ld(i+1)=B2+B224A2C22A2, (A7)

where A2=R02r2, B2=2Lsi(R02R0Dr2), and C2=[(R0D)2r2]L2siD2r2.

APPENDIX B: CONE-BEAM ALGORITHM FOR MBFEX SYSTEM

Following the idea we derived from the fan-beam algorithm, before reconstruction, the projection data should be transformed from the real detector to a virtual detector. In 3D case, each projection collected on a two-dimensional detector array is denoted as (u,v). We have derived the data transform formulae in u-axes, so the remaining work is to locate the projection in v-axes.

In Fig. 3a, the distance between S1 and Pr can be computed as

S1Pr=(R1 cosλ1+ur)2+(R1 sinλ1+Dod)2 (B1)

and the distance between S1 and Pv is

S1Pv=R12+uv2. (B2)

Since

vvvr=S1PvS1Pr, (B3)

substituting Eqs. B1, B2 into Eq. B3 gives

vr=vv(R1 cosλ1+ur)2+(R1 sinλ1+Dod)2R12+uv2. (B4)

The reconstruction formula for 3D case is derived from the general Feldkamp algorithm by Wang et al.17 It can be described as follows:

f(x,y)=12λsλeR2(Rs)2dλω(λ,u)g(λ,u,v)h(RtRsu)RR2+u2+v2du (B5)

and the weight function are changed accordingly

ω(λ,u)={sin2(π4λδtan1(uR)),0λ2δ2tan1(uR)1,2δ2tan1(uR)λπ2tan1(uR)sin2(π4π+2δλδ+tan1(uR)),π2tan1(uR)λπ+2δ}, (B6)

where λ=λ1+v2R2, R=R2+v2, δ=tan−1(|umax|∕R), λes+π+2δ, and δ=arctan(rR) is half the full fan angle of the central-scanning plane.

References

  1. Badea C., Hedlund L., and Johnson G., “Micro-CT with respiratory and cardiac gating,” Med. Phys. 31, 3324–3329 (2004). 10.1118/1.1812604 [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Cao G., Lee Y. Z., Peng R., Liu Z., Rajaram R., Calderon-Colon X., An L., Wang P., Phan T., Sultana S., Lalush D. S., Lu J. P., and Zhou O., “A dynamic micro-CT scanner based on a carbon nanotube field emission x-ray source,” Phys. Med. Biol. 54, 2323–2340 (2009). 10.1088/0031-9155/54/8/005 [DOI] [PubMed] [Google Scholar]
  3. Cao G., Calderon-Colon X., Wang P., Burk L., Lee Y., Rajaram R., Sultana S., Lalush D., Lu J., and Zhou O., “A dynamic micro-CT scanner with a stationary mouse bed using a compact carbon nanotube field emission x-ray tube,” Proc. SPIE 7258, 72585Q–1 (2009). 10.1117/12.813817 [DOI] [PubMed] [Google Scholar]
  4. Badea C. T., Fubara B., Hedlund L. W., and Johnson G. A., “4-D micro-CT of the mouse heart,” Mol. Imaging 4, 110–116 (2005). [DOI] [PubMed] [Google Scholar]
  5. Robb R. A., Hoffman E. A., Sinak L. J., Harris L. D., and Ritman E. L., “High-speed three-dimensional x-ray computed tomography: The dynamic spatial reconstructor,” Proc. IEEE 71, 308–319 (1983). 10.1109/PROC.1983.12589 [DOI] [Google Scholar]
  6. Zhang J., Yang G., Cheng Y., Gao B., Qiu Q., Lee Y., Lu J., and Zhou O., “Stationary scanning x-ray source based on carbon nanotube field emitters,” Appl. Phys. Lett. 86, 184104 (2005). 10.1063/1.1923750 [DOI] [Google Scholar]
  7. Qian X., Rajaram R., Calderon-Colon X., Yang G., Phan T., Lalush D., Lu J., and Zhou O., “Design and characterization of a spatially distributed multibeam field emission x-ray source for stationary digital breast tomosynthesis,” Med. Phys. 36, 4389–4399 (2009). 10.1118/1.3213520 [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Maltz J., Sprenger F., Fuerst J., Paidi A., Fadler F., and Bani-Hashemi A., “Fixed gantry tomosynthesis system for radiation therapy image guidance based on a multiple source x-ray tube with carbon nanotube cathodes,” Med. Phys. 36, 1624–1636 (2009). 10.1118/1.3110067 [DOI] [PubMed] [Google Scholar]
  9. Zhang T., Schulze D., Xu X., and Kim J., “Tetrahedron beam computed tomography (TBCT): A new design of volumetric CT system,” Phys. Med. Biol. 54, 3365–3378 (2009). 10.1088/0031-9155/54/11/006 [DOI] [PubMed] [Google Scholar]
  10. Peng R., Zhang J., Calderon-Colon X., Wang S., Sultana S., Chang S., Lu J., and Zhou O., “Stationary micro-CT scanner using a distributed multi-beam field emission x-ray source: A feasibility study,” Proc. SPIE 7258, 725847 (2009). 10.1117/12.811467 [DOI] [Google Scholar]
  11. Quan E. and Lalush D., “A faster OSC algorithm for iterative reconstruction in a rotation-free micro-CT system,” Phys. Med. Biol. 54, 1061–1072 (2009). 10.1088/0031-9155/54/4/016 [DOI] [PubMed] [Google Scholar]
  12. Kak A. and Slaney M., Principles of Computerized Tomographic Imaging (IEEE Press, New York, 1999). [Google Scholar]
  13. Parker D., “Optimal short scan convolution reconstruction for fan beam CT,” Med. Phys. 9, 254–257 (1982). 10.1118/1.595078 [DOI] [PubMed] [Google Scholar]
  14. Yu H., Ye Y., Zhao S., and Wang G., “Local ROI reconstruction via generalized FBP and BPF algorithms along more flexible curves,” Int. J. Biomed. Imaging 2006, 1–7 (2006). 10.1155/IJBI/2006/14989 [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Mori S., Endo M., Obata T., Kishimoto R., Kato H., Kandatsu S., Tsujii H., and Tanada S., “Noise properties for three weighted Feldkamp algorithms using a 256-detecotor row CT-scanner: Case study for hepatic volumetric cine imaging,” Eur. J. Radiol. 59, 289–294 (2006). 10.1016/j.ejrad.2006.02.007 [DOI] [PubMed] [Google Scholar]
  16. Heuscher D., Brown K., and Noo F., “Redundant data and exact helical cone-beam reconstruction,” Phys. Med. Biol. 49, 2219–2238 (2004). 10.1088/0031-9155/49/11/008 [DOI] [PubMed] [Google Scholar]
  17. Wang G., Lin T., Cheng P., and Shinozaki D., “A general cone-beam reconstruction algorithm,” IEEE Trans. Med. Imaging 12, 486–496 (1993). 10.1109/42.241876 [DOI] [PubMed] [Google Scholar]

Articles from Medical Physics are provided here courtesy of American Association of Physicists in Medicine