Table 7.

Studies of single-channel nocturnal oximetry performed at home

Author, Year, Country Number of subjects Name of Device Reference Diagnostic Criteria Oximeter scoring criteria Sn (%) Sp (%) LR+ LR−
Williams et al,5 1991, USA 40 referred to sleep clinic Biox 3700 IV*, Nelllcor N100 AI > 10 ODI 4% 58 92 10.6 0.16
Oximetry plus clinical score 85 92 7.2 0.46

Series et al,6 1993, Canada 240 sleep clinic patients with possible OSA Biox IVa AHI ≥ 10 Visual scoring 98 48 1.9 0.04

Gyulay et al,7 1993, Australia 98 with possible OSA Biox 3700§ AHI ≥ 15 Clinical assessment 79 50 1.6 0.42
ODI 4% 40 98 20.0 0.61
ODI 3% 51 90 5.1 0.54
ODI 2% 65 74 2.5 0.47
CT < 90 93 51 1.9 0.14
Visual inspection 72 88 6.0 0.32

Ryan et al,8 1995, UK 69 with possible OSA Minolta Pulsox-7** AHI ≥ 15 ODI 4% > 15 32 100 0.68

Olson et al,9 1999, Australia 793 with possible OSA Biox 4700†† AHI ≥ 15 Delta Index (< 0.6) 88 40 1.5 0.29
CT 90 ≥ 1% 75 46 1.4 0.54
AHI ≥ 30 Delta Index (< 0.6) 93 34 1.4 0.21
CT 90 ≥ 1% 84 44 1.5 0.36

Wiltshire et al,10 2001, England 100 with possible OSA Biox 3740‡‡ ODI ≥ 10 ODI ≥ 10 41 100 0.59
ODI ≥ 15 ODI ≥ 15 35 100 0.65

AHI, Apnea hypopnea Index; AI, Apnea Index; ODI 4%, 3%, 2%, oxygen desaturation index per hour to 4%, 3%, 2%; CT < 90, cumulative time below 90% saturation; Delta Index, the average of the absolute differences of oxygen saturation between successive intervals; Sn, sensitivity; Sp, specificity; LR+, positive likelihood ratio; LR−, negative likelihood ratio;

*

Ohmeda, Boulder, CO, USA;

Nellcor Inc, Hayward, CA, USA;

Ohmeda, Boulder, CO, USA;

§

Ohmeda, Boulder, CO, USA;

**

Minolta, Diessenhofen, Switzerland;

‡‡

Ohmeda, Boulder, CO, USA;

‡‡

Ohmeda, Boulder, CO, USA