Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2011 Jul 1.
Published in final edited form as: Radiat Res. 2010 Jul;174(1):91–100. doi: 10.1667/RR1933.1

Computational Electromagnetic Analysis in a Human Head Model with EEG Electrodes and Leads Exposed to RF-Field Sources at 915 MHz and 1748 MHz

Leonardo M Angelone a,b,1, Giorgi Bit-Babik c, Chung-Kwang Chou c
PMCID: PMC2917271  NIHMSID: NIHMS220260  PMID: 20681803

Abstract

An electromagnetic analysis of a human head with EEG electrodes and leads exposed to RF-field sources was performed by means of Finite-Difference Time-Domain simulations on a 1-mm3 MRI-based human head model. RF-field source models included a half-wave dipole, a patch antenna, and a realistic CAD-based mobile phone at 915 MHz and 1748 MHz. EEG electrodes/leads models included two configurations of EEG leads, both a standard 10–20 montage with 19 electrodes and a 32-electrode cap, and metallic and high resistive leads. Whole-head and peak 10-g average SAR showed less than 20% changes with and without leads. Peak 1-g and 10-g average SARs were below the ICNIRP and IEEE guideline limits. Conversely, a comprehensive volumetric assessment of changes in the RF field with and without metallic EEG leads showed an increase of two orders of magnitude in single-voxel power absorption in the epidermis and a 40-fold increase in the brain during exposure to the 915 MHz mobile phone. Results varied with the geometry and conductivity of EEG electrodes/leads. This enhancement confirms the validity of the question whether any observed effects in studies involving EEG recordings during RF-field exposure are directly related to the RF fields generated by the source or indirectly to the RF-field-induced currents due to the presence of conductive EEG leads.

INTRODUCTION

Electroencephalography (EEG) in combination with exposure to mobile phone or equivalent sources has been used to model neurological effects of mobile phone exposure (1). However, there are some questions regarding the effect of EEG conductive leads on the RF-field exposure. When conductive EEG leads are exposed to an RF field (e.g., RF coil in MRI, mobile phones), induced currents may be generated along the leads; such currents, when conducted into the head through the interface between EEG electrode and epidermis, may be associated with changes in the RF power distribution in the anatomical structures underneath the electrode, including epidermis, fat, skull and brain (24). For relatively high-power RF-field sources, such as MRI RF coils (i.e., power of the order of tens of watts), changes in the RF-field power inside the head may raise thermally related safety issues (59), whereas for lower-power RF-field sources, such as mobile phones (i.e., power of the order of hundreds of mW), the induced currents from the EEG leads into the head may affect dosimetry (10) and the proper setup for the EEG measurements (1).

Numerical studies based on the Finite-Difference Time-Domain (FDTD) method (11) have been widely used to study RF-field dosimetry (1216) in terms of the specific absorption rate (SAR) (17). The FDTD method has been shown to provide results with a 20% accuracy compared to experimental setups (8), and it has been selected as one of the primary methods for SAR evaluation by the IEEE International Committee on Electromagnetic Safety (18).

SAR distribution in the human head is strongly affected by the inhomogeneous dielectric properties of the human head tissues (15, 16, 19, 20); the use of numerical head models with increasing spatial resolution and anatomical detail (10, 15, 21, 22) allowed improved precision in estimation of local SAR with respect to thin anatomical structures of the human head, including those close to the EEG electrodes such as epidermis, subcutaneous tissue and bone marrow. Simulation and experimental studies performed with medical implants or EEG leads exposed to an RF field (10, 21, 23, 24) showed local SAR peaks near the metallic implants or EEG electrodes, with values depending on the lead orientation (10), shielding effect of EEG leads (21, 25), radiofrequency (21, 25), number of EEG electrodes/leads (10, 21), and resistivity of EEG leads (2, 3, 17, 26). In the present study, we performed a systematic numerical analysis of the effects of these variables on the electromagnetic field and SAR in a 1-mm3 resolution human head model with EEG electrodes/leads and exposed to RF fields from a number of sources, including a mobile phone.

METHODS

Numerical Head Model

Anatomical MRI data were acquired on a healthy 37-year-old right-handed adult male volunteer. Informed consent was obtained in accordance with Massachusetts General Hospital policies. Data were acquired with a quadrature birdcage transmit/receive head coil on a 1.5 T scanner (General Electric, Milwaukee, WI). Data were collected with a T1-weighted 3D-SPGR sequence (TR/TE = 24/8 ms) with 124 slices, 1.3 mm thick (matrix size 256 × 192, FOV 256 mm). The volume data were resampled to isotropic voxels with dimensions of 1 × 1 × 1 mm3. Segmentation was then applied on this dataset volume. Twenty-six non-brain anatomical structures were segmented by means of morphometric analysis.2 Three brain structures (cerebrospinal fluid, gray and white matter) were also segmented with an automatic segmentation algorithm (27, 28) and coregistered with the non-brain structures. The final head model (Fig. 1) consisted of a total of 29 anatomical structures (Table 1). Each anatomical structure was assigned biophysical properties (Table 1) at 915 and 1748 MHz, frequencies within the commonly used transmit bands for mobile phones (19, 2931). The electrical parameters were considered: (a) linear with electric field, (b) isotropic, (c) dispersive and (d) heterogeneous in space (32).

FIG. 1.

FIG. 1

Three-dimensional view of MRI-based human head model wearing EEG electrodes/leads used for the study. The models of the three RF-field sources used (dipole, patch antenna and mobile phone) are shown. Thirty-two EEG electrodes with two realistic configurations of EEG leads (Set 1 and Set 2) were evaluated.

TABLE 1.

Anatomical Structures of MRI-Based Head Model and Electrical Properties at 915 MHz and 1748 MHz

Anatomical structural entities Density (kg/m3) 915 MHzd
1748 MHzd
σ (S/m) εr σ (S/m) εr
Brain
 Gray matter 1030a 0.94 52.72 1.36 50.19
 White matter 1030a 0.59 38.88 0.89 37.09
 Cerebrospinal fluid 1010a 2.41 68.63 2.89 67.28
Non-brain
 Adipose 920a 0.11 11.33 0.18 11.04
 Air (respiration/digestion/sinus) 1.3b 0.00 1.00 0.00 1.00
 Aqueous humor 1010a,c 1.64 68.90 2.00 68.60
 Bone (facial) 1850b 0.14 12.45 0.27 11.81
 Connective tissue 1100a 0.78 42.65 1.25 40.34
 Cornea 1076c 1.39 55.23 1.83 52.87
 CSF subarachnoid 1010a 2.41 68.63 2.89 67.28
 Bone marrow 1080b 0.34 20.79 0.57 19.41
 Dura 1030a 0.96 44.42 1.29 42.97
 Ear/pinna 1100a 0.78 42.65 1.25 40.34
 Epidermis 1100a 0.86 41.40 1.16 38.96
 Inner table (skull) 1850b 0.14 12.45 0.27 11.81
 Lens 1100c 0.79 46.57 1.12 45.41
 Muscle 1040d 0.97 55.96 1.36 54.51
 Nasal structures 1100a 0.78 42.65 1.25 40.34
 Nerve 1040a 0.57 32.53 0.83 30.93
 Orbital fat 920a 0.11 11.33 0.18 11.04
 Outer table (skull) 1850b 0.14 12.45 0.27 11.81
 Retina/choroid/sclera 1170a 1.17 55.27 1.57 53.65
 Soft tissue 1100a 0.78 42.65 1.25 40.34
 Spinal cord 1040a 0.57 32.53 0.83 30.93
 Subcutaneous tissue 1100a 0.78 42.65 1.25 40.34
 Subcutaneous fat 920a 0.11 11.33 0.18 11.04
 Teeth 1850b 0.14 12.45 0.27 11.81
 Tongue 1040d 0.97 55.96 1.36 54.51
 Vitreous humor 1010a,c 1.64 68.90 2.00 68.60
a

Ref. (30).

b

Ref. (6).

c

Ref. (31).

d

Ref. (19).

Numerical Model of RF-Field Sources

Three RF-field sources were modeled: a half-wave dipole, a patch antenna and a mobile phone.

Half-wave dipole

A dipole was modeled with a perfectly electric conductive (PEC) 164-mm-long wire, placed 15 mm between the feedpoint and head surface (Fig. 1). The dipole was center-fed, and simulations were performed at 915 MHz.

Patch antenna

A patch antenna was modeled as in ref. (10) using a rectangular block (192 × 136 mm2, 10 mm thick) 61 mm from the head. The patch was composed of a dielectric material (σ = 0.0001 S/m, εr = 1.5), placed between a ground plate (192 × 136 mm2) of perfectly conductive material on one side and a metallic resonant plate (140 × 90 mm2) on the other side, facing the head (Fig. 1). The two copper plates were connected with a thin metallic wire, where a 915 MHz sinusoidal source was placed.

Mobile phone

A CAD model of a commercially available mobile phone was used. The distance between head and feedpoint of the phone was 15 mm, as for the dipole. The minimum distance between the phone and the head was 3 mm. Simulations with the phone were performed at 915 MHz and 1748 MHz.

Numerical Model of EEG Electrodes and Leads

EEG electrodes

Thirty-two EEG electrodes were designed with Circuit Maker (Altium Inc., San Diego, CA) on a 2D mask and imported into Matlab (Mathworks, Natick, MA) (26). The mask was coregistered with the axial slice of the largest diameter on the head model. EEG electrodes present in the mask were projected and placed on top of the surface of the head model (“epidermis” of the model). EEG electrodes were modeled as discs (radius = 7 mm, height = 3 mm). The model was then imported into the XFDTD software (Remcom Co., State College, PA) to complete the model by including the EEG leads. Two different types of EEG caps were modeled (Fig. 1): Set 1 and Set 2.

Set 1

EEG leads modeled as PEC wires were connected separately to each electrode and were oriented around the head (Set 1 in Fig. 1) (10, 21, 25). Leads were modeled using three to nine segments, depending on the position with respect to the head, with distance from the head between 1 mm and 11 mm. Two different subsets were simulated to model two typical configurations of connections between EEG electrodes and amplifier through conductive leads: 19 leads, following the standard 10–20 montage (33), and an extended 32-lead configuration (3537). For this set, the case of leads with increased resistivity (ρ = 0.01 Ωm) was also evaluated for comparison (26, 35).

Set 2

Thirty-two leads connecting to each electrode and following the head curvature, never touching the head surface, were bundled at the Cz position (33, 35, 37), exiting vertically from the head for 53 mm (Set 2 in Fig. 1).

FDTD Simulations

FDTD simulations were performed using commercially available software (XFDTD, Remcom Inc., State College, PA). The models were imported into the XFDTD software, and a uniform padding of 40 free-space cells with seven layers of perfectly matching (PML) boundary was used to model open space beyond the simulation domain. A harmonic source (1 V peak-to-peak, 50 Ω feedpoint resistance) was used to model the RF-field excitation in all cases. Calculations were run until steady-state conditions were reached with −40 dB of convergence criterium as evaluated with the XFDTD software, corresponding to approximately 12 cycles of sinusoidal signal. Simulations were performed on a Dual Core 1.83 GHz computer with 4 GB of RAM. Electric-field amplitude, induced currents, 1-g average (SAR1g), 10-g average (SAR10g), and whole-head average SAR (SARw) were computed with the XFDTD software (38). The results were normalized to 0.25 W of net input power at 915 MHz and 0.125 W at 1748 MHz.

The SARs calculated with EEG electrodes/leads were compared voxel by voxel (i.e., 1-mm3 spatial resolution) to the corresponding values calculated without EEG electrodes as follows (26):

Pi=SARiLEADSARiNOLEAD=PoweriLEADPoweriNOLEAD (1)

where SARiLEAD and SARiNOLEAD are the SARs in the ith voxel with or without EEG electrodes/leads, respectively. This ratio is equivalent to the ratio of the RF-field power absorbed in the ith voxel with or without EEG electrodes/leads, respectively.

RESULTS

The RF field induced visible currents along the EEG leads for both sets with all RF-field sources modeled (Fig. 2) with related SAR increases underneath the EEG electrodes (Fig. 3).

FIG. 2.

FIG. 2

Sagittal view of induced currents for the control case (no leads) and the two sets of EEG electrodes/leads. The sources were a half-wave dipole antenna, a patch antenna and a mobile phone at 915 MHz. Values are normalized for 0.25 W of available input power. The induced currents along the conductive leads are visible. Color scale from 0 dB to −70 dB.

FIG. 3.

FIG. 3

Coronal view of SAR for the control case (no leads) and the two sets of EEG electrodes/leads. The sources were a half-wave dipole antenna, a patch antenna and a mobile phone at 915 MHz. All sources were on the left of the head (i.e., right in the image as for radiological convention). Values normalized for 0.25 W of available input power. Arrows indicate local increases in SAR underneath the EEG electrodes.

Volumetric Analysis of Changes in SAR

Table 2 reports the volumetric analysis as described in Eq. (1) for different anatomical structures present underneath the EEG electrodes, namely epidermis, subcutaneous fat, skull (i.e., outer and inner table of skull), bone marrow and brain (i.e., gray and white matter). Figure 4 shows cross sectional views of the Pi patterns for Set 1 and Set 2. The largest increase of 1279-fold was in the epidermis with Set 1 exposed to the dipole. Local changes in the brain were observed with Set 1 with the dipole [i.e., max (Pi) = 46, avg (Pi) = 1.2 and 5% of brain volume with Pi > 2] and patch antenna [i.e., max (Pi) = 21, avg (Pi) = 1 and 4% of brain volume with Pi > 2]. For the mobile phone at 915 MHz (Table 2 and Fig. 4), there was a peak increase of 859-fold in the epidermis and 48-fold in the brain with Set 1 and 14% of brain volume wih Pi > 2. Smaller changes were observed with the mobile phone at 1748 MHz, with a 52-fold maximum increase in the epidermis, 14-fold in the brain, and 4% of brain volume with Pi > 2. The enhancements for Set 2 were all smaller than those for Set 1. For example, for exposure to the mobile phone at 915 MHz, there was a peak increase of 460-fold in the epidermis but smaller changes in the internal structures, with a 6.6-fold peak increase and 0.8% Pi > 2 in the brain structures.

TABLE 2.

Peak and Mean Changes in RF-Field Power with and without Leads

Dipole 915 MHz Pi
Epidermis, dermis Subcutaneous fat Skull Bone marrow Brain
Set 1 max (Pi) 1279.0 80.8 66.2 18.8 45.9
mean (Pi) 0.78 1.05 1.04 0.89 1.16
Volume (1 mm3) 16990 25348 23546 6988 70193
Pi > 2
Volume with Pi > 2 percentage of total volume 4.4% 6.6% 7.3% 5.1% 5.1%
Set 2 max (Pi) 366.0 75.0 54.9 14.8 9.1
mean (Pi) 0.81 1.05 1.07 1.01 1.03
Volume (1 mm3) 7163 8764 12718 3960 18485
Pi > 2
Volume with Pi > 2 percentage of total volume 1.9% 2.3% 3.9% 2.9% 1.3%

Patch 915 MHz Pi
Epidermis, dermis Subcutaneous fat Skull Bone marrow Brain

Set 1 max (Pi) 270.0 163.2 8.5 10.0 21.1
mean (Pi) 0.69 0.99 0.92 0.86 1.00
Volume (1 mm3) 9276 17612 13472 3744 56905
Pi > 2
Volume with Pi > 2 percentage of total volume 2.4% 4.6% 4.2% 2.7% 4.1%
Set 2 max (Pi) 430.7 83.6 58.0 16.6 23.5
mean (Pi) 0.77 1.03 1.05 1.02 0.98
Volume (1 mm3) 5650 9799 11075 3849 10061
Pi > 2
Volume with Pi > 2 percentage of total volume 1.5% 2.6% 3.4% 2.8% 0.7%

Mobile phone 915 MHz Pi
Epidermis, dermis Subcutaneous fat Skull Bone marrow Brain

Set 1 max (Pi) 859.1 454.0 31.0 21.9 47.8
mean (Pi) 1.04 1.34 1.33 1.27 1.36
Volume (1mm3) 25922 41755 43417 19549 195016
Pi > 2
Volume with Pi > 2 percentage of total volume 6.7% 10.9% 13.4% 14.3% 14.2%
Set 2 max (Pi) 459.5 83.9 34.5 8.2 6.6
mean (Pi) 0.87 1.07 1.04 1.03 1.01
Volume (1 mm3) 7983 8698 8691 3675 11603
Pi > 2
Volume with Pi > 2 percentage of total volume 2.1% 2.3% 2.7% 2.7% 0.8%

Mobile Phone 1748 MHz Pi
Epidermis, dermis Subcutaneous fat Skull Bone marrow Brain

Set 1 max (Pi) 52.2 15.9 10.2 5.3 14.1
mean (Pi) 0.89 1.3 1.10 1.13 1.08
Volume (1 mm3) 26826 40065 17791 8315 56447
Pi > 2
Volume with Pi > 2 percentage of total volume 7.0% 10.5% 5.5% 6.1% 4.1%
Set 2 max (Pi) 31.0 8.9 9.0 2.2 7.2
mean (Pi) 0.74 1.0 1.00 0.97 0.98
Volume (1mm3) 2276 2614 3331 91 5505
Pi > 2
Volume with Pi > 2 percentage of total volume 0.6% 0.7% 1.0% 0.1% 0.4%

Notes. Values for Set 1 and Set 2 for anatomical structures underneath EEG electrodes are shown. The volume with Pi higher than 2 (i.e. over 100% increase in RF power compared to no leads) is provided as the absolute value as well as a percentage of the overall volume of the anatomical structure.

FIG. 4.

FIG. 4

Changes in RF-field power with and without leads (Pi) for 915 MHz mobile phone exposure with the two sets of EEG electrodes/leads modeled. Coronal (from the back to the front of the head), axial (from the top of the head to the neck), and sagittal (from right to the left of the head) are shown. The phone was placed on the left of the head. Significant increases in Pi were observed for Set 1 near the EEG leads and in all the anatomical structures underneath the EEG electrode near the source (coronal and axial view) and on the back of the head near the leads (sagittal view). Changes for Set 2 were localized mainly underneath the electrode in Cz position, where the leads were bundled.

Antenna Performance and Peak SAR Evaluation

For both head models with EEG leads and with all RF-field sources at 915 MHz there were no significant changes in antenna performance (i.e., efficiency, impedance at source) and whole-head average SAR (Table 3).

TABLE 3.

Dipole Patch and Mobile Phone at 915 MHz and 1748 MHz for Head without Leads and with Two Sets of 32 EEF Electrodes/Leads Modeled in this Study

Dipole - 915 MHz

Head no electrodes Set 1 32 electrodes/leads Set 2 32 electrodes/leads
Net input power (W) 0.25 0.25 0.25
Power dissipated (W) 0.18 0.18 0.18
Radiation efficiency 28.8% 26.2% 29.2%
Impedance (Z) 55.65 + j42.62 54.83 + j46.34 55.07 + j42.73
Peak 1 g average SAR (W/kg) 1.72 1.94 1.73
Peak 10 g average SAR (W/kg) 1.20 1.37 1.22
Whole-head SAR (W/kg) 0.035 0.036 0.034

Patch - 915 MHz

Head no electrodes Set 1 32 electrodes/leads Set 2 32 electrodes/leads

Net input power (W) 0.25 0.25 0.25
Power dissipated (W) 0.13 0.14 0.13
Radiation efficiency 49.2% 45.4% 49.5%
Impedance (Z) 2.68 + j30.89 2.12 + j30.58 2.63 + j30.98
Peak 1 g average SAR (W/kg) 0.54 1.29 0.54
Peak 10 g average SAR (W/kg) 0.31 0.37 0.31
Whole-head SAR (W/kg) 0.022 0.023 0.022

Mobile phone - 915 MHz

Head no electrodes Set 1 32 electrodes/leads Set 2 32 electrodes/leads

Net input power (W) 0.25 0.25 0.25
Power dissipated (W) 0.19 0.21 0.19
Radiation efficiency 21.9% 17.1% 22.1%
Impedance (Z) 64.08 + j124.33 68 + j129.35 64.11 + j124.12
Peak 1 g average SAR (W/kg) 0.82 0.86 0.82
Peak 10 g average SAR (W/kg) 0.44 0.54 0.44
Whole-head SAR (W/kg) 0.015 0.016 0.015

Mobile phone - 1748 MHz

Head no electrodes Set 1 32 electrodes/leads Set 2 32 electrodes/leads

Net input power (W) 0.13 0.13 0.13
Power dissipated (W) 0.11 0.10 0.10
Radiation efficiency 23.130% 18.400% 19.4%
Impedance (Z) 16.8 + j121 15.82 + j120 16.03 + j120.7
Peak 1 g average SAR (W/kg) 0.19 0.21 0.20
Peak 10 g average SAR (W/kg) 0.10 0.11 0.10
Whole-head SAR (W/kg) 0.002 0.002 0.002

Dipole

There was a 13% increase in peak 1-g average SAR and a 14% increase in peak 10-g average SAR with electrode Set 1 compared to the case of no electrodes/leads. No changes (<1%) in peak 1-g average and 10-g average were observed with Set 2 (Table 3). The peak SAR was on the epidermis under the EEG lead near the dipole (see Fig. 3) for Set 1, whereas it was in the epidermis at the Cz position, where the leads were bundled, for Set 2.

Patch

There was a 2.4× increase in peak 1-g average and 20% in peak 10-g average SAR with Set 1 compared to the case of no electrodes/leads. No changes (<1%) in peak 1-g average and 10-g average were observed with Set 2 (Table 3).

Mobile Phone

There was a 5% increase in the peak 1-g average and 20% in the peak 10-g average SAR with Set 1 compared to the case with no electrodes/leads. No changes were observed with Set 2 (Table 3). Similar to the dipole and patch antenna cases, the peak was in the epidermis under the EEG electrode closest to the source for Set 1, whereas it was at the Cz position for Set 2 (see Fig. 3). Overall, the effects of EEG leads were reduced at 1748 MHz (Table 3) compared to 915 MHz. There was a 10% increase in peak 1-g average and 10-g average SAR with Set 1 and 4% with Set 2 compared to the case of no electrodes/leads. The peak SARs were similar for 19 and 32 electrodes/leads (Table 4). As for the 32 electrodes, the peaks SARs for 19 electrodes were under the EEG electrode/leads closest to the source. Table 5 shows changes within 10% for peak 1-g average and 10-g average SAR with high resistive leads compared to the no-lead case. The effect of increased resistivity was more evident for Set 1, with a peak 10-g average SAR of 0.49 W/kg compared to 0.54 W/kg with conductive leads (Table 3) and 0.44 W/kg without leads.

TABLE 4.

Variable Number of Electrodes

Mobile phone - 915 MHz

Head no electrodes Set 1 19 electrodes/leads Set 2 32 electrodes/leads
Net input power (W) 0.25 0.25 0.25
Power dissipated (W) 0.19 0.21 0.21
Radiation efficiency 21.9% 17.2% 17.1%
Impedance (Z) 64.08 + j124.33 67.84 + j129.18 68 + j129.35
Peak 1 g average SAR (W/kg) 0.82 0.86 0.86
Peak 10 g average SAR (W/kg) 0.44 0.53 0.54
Whole-head SAR (W/kg) 0.015 0.016 0.016

Notes. Head model with 19 and 32 EEG electrodes leads, with leads bundled as in Set 1. RF-field source was mobile phone at 915 MHZ. SARs are normalized to net input power equal to 0.25 W.

TABLE 5.

Variable Resistivity

Mobile phone - 915 MHz
Head no electrodes Set 1 32 electrodes/leads ρ = 0.01 Ωm Set 2 32 electrodes/leads ρ = 0.01 Ωm
Net input power (W) 0.25 0.25 0.25
Power dissipated (W) 0.19 0.20 0.20
Radiation efficiency 21.9% 21.1% 21.7%
Impedance (Z) 64.08 + j124.33 201 + j0.954 62.24 + j125.27
Peak 1 g average SAR (W/kg) 0.82 0.83 0.79
Peak 10 g average SAR (W/kg) 0.44 0.49 0.44
Whole-head SAR (W/kg) 0.015 0.016 0.015

DISCUSSION

When conductive leads are exposed to a radiofrequency field, induced currents are generated along the leads (2). The effects of such induced currents can be seen in two perspectives: (a) potential safety issues related to local increase of electric-field deposition at the interface between the EEG electrode and the human head, historically measured in terms of specific absorption rate (17) and (b) effects on electromagnetic-field distribution related to the EM scattering of the leads; i.e., the EEG leads cannot be considered proper “RF-transparent” measurement probes. In this perspective, changes in the EMF due to the EEG leads can be considered as “noise” with respect to the measurements of EEG for RF-field exposure of a specific antenna. Schmid et al. (10) presented a new exposure system for studies requiring EEG recording on human subjects exposed to GSM900-and UMTS-equivalent sources. Although the study showed minimal interference when EEG leads were placed orthogonally to the source, the study raised potential issues for different lead orientations or numbers of EEG electrodes/leads. Since the amount of current induced depends on the frequency and the specific geometry of the leads with respect to the source, estimation of such currents can be cumbersome for complex geometries, such as conductive leads placed on the head of human subjects during electroencephalography recordings. In such cases, an anatomically precise computational model can be useful to evaluate the interactions between the leads and the RF-field source. Several MRI-based human head models are now available (15). The model used in this study is characterized by a 1-mm3 isotropic spatial resolution and 29 anatomical structures, which allowed the modeling with geometrical accuracy the thin structures of interests for the problem of EEG leads, such as epidermis, where EEG electrodes are connected to the head, and bone marrow, where potential local increases in SAR are possible (21).

Previous studies used peak 1-g average and 10-g average SAR to evaluate changes in SAR with EEG leads exposed an RF field generated by mobile phones (10, 25). The regulatory limit for localized exposure to an RF field generated by mobile phones set by the Federal Communications Commission (FCC) is 1.6 W/kg for peak 1-g average SAR (39, 40). The limit set by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) (41) and the American National Standard Institute (ANSI) and IEEE (42) is 2 W/kg for peak 10-g average SAR. It has been shown that the temperature rise in tissue correlates well with an average tissue mass of 10 g (43, 44). This study showed that the peak 1-g average and peak 10-g average SARs with mobile phones were below the corresponding exposure limits. Thus, given the relatively low input power of mobile phones as opposed, for example, to the RF coil used in MRI (21), the conductive EEG leads are not expected to generate excessive temperature rises during exposure to the RF field of mobile phones.

The whole-head average SAR and 10-g average SAR did not vary significantly (i.e., less than 10%) in the different cases considered. The peak 1-g average SAR varied less than 10% with dipole and mobile phone, while a twofold change compared to the case of no leads was observed with Set 1 and the patch antenna. The results of the study confirmed the peculiar local behavior of RF power in the presence of conductive leads (45), with changes occurring in volumes of a few mm3 rather than over the whole head. While relatively coarse parameters such as whole-head average SAR, 10-g average SAR or even 1-g average SAR did not show significant changes in the RF-field power with and without leads, a more comprehensive volumetric assessment of changes in RF power over the entire head showed significant local increases of RF-field energy absorption in anatomical structures underneath the EEG electrodes.

The resolution of the numerical model used in this study is one of the highest among the anthropomorphic numerical models for RF-field dosimetry (15); this allowed a 1-mm3-resolution volumetric assessment of RF-field power. Numerical errors related to staircasing of small structures may still be present (8, 46), and a proper validation of these results with geometrically matched experimental measurements may be needed. The SARs found in this study were in line with previous published work. The simulations with the dipole and mobile phone were in line with canonical results.3 The peak 10-g average SAR with the patch antenna and the head model without leads was 1.2 W/kg (1 W of net input power), similar to the value of 1.02 W/kg simulated and 0.92 W/kg measured with an antenna of same dimensions and at the same distance from the head reported in ref. (10). Simulations with a mobile phone at 915 MHz and a head model without leads resulted in a peak 10-g average of 1.77 W/kg (1 W of net input power) compared to 3.4 W/kg in ref. (15). Since the study in ref. (15) was performed with the SAM phantom, for a more direct comparison, simulations with an electrically homogeneous model (σ = 0.97 S/m, εr = 55.6, ρ = 1040 kg m−3) (19, 29) based on the same head model used in this study were performed with the mobile phone at 915 MHz. The peak 10-g average SAR for this homogeneous model was 3.18 W/kg (1 W of net input power), within 10% of the value in ref. (15).

Our simulations showed that the presence of conductive EEG leads combined with specific configurations of an RF-field source created a reduction of the RF field (i.e., “shielding effect”) in anatomical structures near the EEG electrodes, similar to that reported in refs. (21) and (25). In particular, the value of Pi was less than one in the epidermis or bone marrow with the dipole at 915 MHz. The shielding effect was greater with the patch antenna. The results of this study also confirm that the leads with higher resistivity (ρ = 0.01 Ωm), used in previous studies to minimize interactions with other RF-field sources (35), helped reduce the currents induced in the leads and the related local SAR on the epidermis and surrounding structures.

CONCLUSIONS

A 1-mm3 anatomically precise human head model with two realistic sets of EEG electrodes and leads was used to evaluate potential changes in SAR during exposure to several RF-field sources, including a mobile phone. Significant changes induced by the conductive EEG leads at the electrode sites with respect to the control case of a human head without EEG leads. Observed SAR peaks were below exposure limits, suggesting that no potential thermal issues may be present when using EEG leads with a mobile phone or equivalent source. However, local enhancement of the electromagnetic field and SAR over those without leads was present over the entire head, including gray and white matter structures. Changes in the SAR were different for two realistic configurations of EEG caps and were present for both the standard 10–20 montage and the extended 32-EEG electrodes. High-resistive leads reduced the currents induced in the leads and local SAR in the head.

The results of this study indicate that the effect of RF-field exposure on EEG is not a simple issue to resolve. Since EEG recording measures very small electrical variations in the brain, it is important to realize that an induced current is carried by the conductive leads into the head. Whether any observed effect is directly related to the RF fields or is caused indirectly by the current induced by the RF field due to the presence of conductive EEG leads must be determined to understand the biological effects of RF-field exposure.

Acknowledgments

This study was supported by the Mobile Manufacturer Forum, the NCRR (grant P41RR14075), the MIND Institute, and the Athinoula A. Martinos Center for Biomedical Imaging. Giorgi Bit-Babik and C-K. Chou are employees of Motorola whose research focuses on the safe use of electromagnetic energy. They contributed their time on this project with the support of the company. The authors would like to thank Bruce Rosen, Giorgio Bonmassar, Christos Vasios, Nikos Makris, David Kennedy, Jonathan Kaiser, George Papadimitriou at the A. Martinos Center for Biomedical Imaging and Mark Cronin-Golomb and Sergio Fantini at Tufts University for the help and the useful insights during the completion of the study. The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services.

Footnotes

2

L. M. Angelone, RF Dosimetry At Ultra High-Field MRI. Ph.D. Dissertation, Biomedical Engineering Department, Tufts University, Medford, MA, 2008.

3

M. Siegbahn, G. Bit-Babik, J. Keshvari, A. Christ, B. Derat, V. Monebhurrun, C. Penney and T. Wittig, An international inter-laboratory comparison of mobile phone SAR calculation with CAD-based models. In Abstracts of the Bioelectromagnetics Society Annual Meeting, p. 22, Davos, Switzerland, 2009.

References

  • 1.Valentini E, Curcio G, Moroni F, Ferrara M, De Gennaro L, Bertini M. Neurophysiological effects of mobile phone electromagnetic fields on humans: a comprehensive review. Bioelectromagnetics. 2007;28:415–432. doi: 10.1002/bem.20323. [DOI] [PubMed] [Google Scholar]
  • 2.Guy AW. Biophysics – Energy absorption and distribution. AGARD Lect Ser. 1974;78:1–14. [Google Scholar]
  • 3.Chou CK, Guy AW. Carbon-loaded Teflon electrodes for chronic EEG recordings in microwave research. J Microwave Power. 1979;14:399–404. doi: 10.1080/16070658.1979.11689175. [DOI] [PubMed] [Google Scholar]
  • 4.Chou CK. RF heating of metallic implants during MRI. In: Ng K-H, Montgomery ND, Tan L-K, editors. International EMF Conference 2007: Electromagnetic Fields, Bioeffects Research, Medical Applications, and Standards Harmonization. University of Malaysia Press; Kuala Lumpur: 2008. pp. 30–37. [Google Scholar]
  • 5.Ibrahim TS, Kangarlu A, Chakeress DW. Design and performance issues of RF coils utilized in ultra high field MRI: experimental and numerical evaluations. IEEE Trans Biomed Eng. 2005;52:1278–1284. doi: 10.1109/TBME.2005.847564. [DOI] [PubMed] [Google Scholar]
  • 6.Collins CM, Liu W, Wang J, Gruetter R, Vaughan JT, Ugurbil K, Smith MB. Temperature and SAR calculations for a human head within volume and surface coils at 64 and 300 MHz. J Magn Reson Imaging. 2004;19:650–656. doi: 10.1002/jmri.20041. [DOI] [PubMed] [Google Scholar]
  • 7.Finelli DA, Rezai AR, Ruggieri PM, Tkach JK, Nyenhuis JA, Hrdlicka G, Sharan A, Gonzalez-Martinez J, Stypulkowski PH, Shellock FG. MR imaging-related heating of deep brain stimulation electrodes: in vitro study. Am J Neuroradiol. 2002;23:1795–1802. [PMC free article] [PubMed] [Google Scholar]
  • 8.Gajsek P, Walters TJ, Hurt WD, Ziriax JM, Nelson DA, Mason PA. Empirical validation of SAR values predicted by FDTD modeling. Bioelectromagnetics. 2002;23:37–48. doi: 10.1002/bem.96. [DOI] [PubMed] [Google Scholar]
  • 9.Shellock FG. Reference Manual for Magnetic Resonance Safety, Implants and Devices. Biomedical Research Publishing Group; Los Angeles: 2008. [Google Scholar]
  • 10.Schmid G, Cecil S, Goger C, Trimmel M, Kuster N, Molla-Djafari H. New head exposure system for use in human provocation studies with EEG recording during GSM900- and UMTS-like exposure. Bioelectromagnetics. 2007;28:636–647. doi: 10.1002/bem.20340. [DOI] [PubMed] [Google Scholar]
  • 11.Taflove A, Hagness SC. Computational Electrodynamics: The Finite-Difference Time-Domain Method. Artech House; Boston: 2000. [Google Scholar]
  • 12.Collins CM, Li S, Smith MB. SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil. Magn Reson Med. 1998;40:847–856. doi: 10.1002/mrm.1910400610. [DOI] [PubMed] [Google Scholar]
  • 13.Ibrahim TS, Mitchell C, Abraham R, Schmalbrock P. In-depth study of the electromagnetics of ultrahigh-field MRI. NMR Biomed. 2007;20:58–68. doi: 10.1002/nbm.1094. [DOI] [PubMed] [Google Scholar]
  • 14.Bit-Babik G, Guy AW, Chou CK, Faraone A, Kanda M, Gessner A, Wang J, Fujiwara O. Simulation of exposure and SAR estimation for adult and child heads exposed to radiofrequency energy from portable communication devices. Radiat Res. 2003;163:580–590. doi: 10.1667/rr3353. [DOI] [PubMed] [Google Scholar]
  • 15.Kainz W, Christ A, Kellom T, Seidman S, Nikoloski N, Beard B, Kuster N. Dosimetric comparison of the specific anthropomorphic mannequin (SAM) to 14 anatomical head models using a novel definition for the mobile phone positioning. Phys Med Biol. 2005;50:3423–3445. doi: 10.1088/0031-9155/50/14/016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Gandhi OP, Chen XB. Specific absorption rates and induced current densities for an anatomy-based model of the human for exposure to time-varying magnetic fields of MRI. Magn Reson Med. 1999;41:816–823. doi: 10.1002/(sici)1522-2594(199904)41:4<816::aid-mrm22>3.0.co;2-5. [DOI] [PubMed] [Google Scholar]
  • 17.NCRP. Report No. 67. National Council on Radiation Protection and Measurements; Bethesda, MD: 1981. Radiofrequency Electromagnetic Fields: Properties, Quantities and Units, Biophysical Interaction, and Measurement. [Google Scholar]
  • 18.Standard P1528. 1-201x, IEEE, Draft, SCC-34, SC-2, WG-2 Computational Dosimetry. IEEE; New York: 2009. Draft Recommended Practice for Determining the Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Body from Wireless Communications Devices, 30 MHz–6 GHz: General Requirements for using the Finite Difference Time Domain (FDTD) Method for SAR Calculations. [Google Scholar]
  • 19.Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues: I. Literature survey. Phys Med Biol. 1996;41:2231–2249. doi: 10.1088/0031-9155/41/11/001. [DOI] [PubMed] [Google Scholar]
  • 20.IEEE. Clause 5: Phantom models Standard. Vol. 1528. IEEE; New York: 2003. Recommended Practice for Determining the Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement Techniques. [Google Scholar]
  • 21.Angelone L, Potthast A, Segonne F, Iwaki S, Belliveau J, Bonmassar G. Metallic electrodes and leads in simultaneous EEG-MRI: Specific absorption rate (SAR) simulation studies. Bioelectromagnetics. 2004;25:285–295. doi: 10.1002/bem.10198. [DOI] [PubMed] [Google Scholar]
  • 22.Makris N, Angelone L, Tulloch S, Sorg S, Kaiser J, Kennedy D, Bonmassar G. MRI-based anatomical model of the human head for specific absorption rate mapping. Med Biol Eng Comput. 2008;46:1239–1251. doi: 10.1007/s11517-008-0414-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Ho HS. Safety of metallic implants in magnetic resonance imaging. J Magn Reson Imaging. 2001;14:472–477. doi: 10.1002/jmri.1209. [DOI] [PubMed] [Google Scholar]
  • 24.Chou CK, McDougall JA, Chan KW. RF heating of implanted spinal fusion stimulator during magnetic resonance imaging. IEEE Trans Biomed Eng. 1997;44:367–73. doi: 10.1109/10.568912. [DOI] [PubMed] [Google Scholar]
  • 25.Hamblin DL, Anderson V, McIntosh RL, McKenzie RJ, Wood AW, Iskra S, Croft RJ. EEG electrode caps can reduce SAR induced in the head by GSM900 mobile phones. IEEE Trans Biomed Eng. 2007;54:914–920. doi: 10.1109/TBME.2007.893486. [DOI] [PubMed] [Google Scholar]
  • 26.Angelone LM, Vasios CE, Wiggins G, Purdon PL, Bonmassar G. On the effect of resistive EEG electrodes and leads during 7 T MRI: simulation and temperature measurement studies. Magn Reson Imaging. 2006;24:801–812. doi: 10.1016/j.mri.2006.01.006. [DOI] [PubMed] [Google Scholar]
  • 27.Dale AM, Fischl B, Sereno MI. Cortical surface-based analysis. I. Segmentation and surface reconstruction. Neuroimage. 1999;9:179–194. doi: 10.1006/nimg.1998.0395. [DOI] [PubMed] [Google Scholar]
  • 28.Segonne F, Dale AM, Busa E, Glessner M, Salat D, Hahn HK, Fischl B. A hybrid approach to the skull stripping problem in MRI. Neuroimage. 2004;22:1060–1075. doi: 10.1016/j.neuroimage.2004.03.032. [DOI] [PubMed] [Google Scholar]
  • 29.Federal Communications Commission (FCC) Body Tissue Dielectric Parameters Tool. [Available online at http://www.fcc.gov/fcc-bin/dielec.sh]
  • 30.Li QX, Gandhi OP. Thermal implications of the new relaxed IEEE RF safety standard for head exposures to cellular telephones at 835 and 1900 MHz. IEEE Trans Microwave Theory Tech. 2006;54:3146–3154. [Google Scholar]
  • 31.DeMarco SC, Lazzi G, Liu W, Weiland JD, Humayun MS. Computed SAR and thermal elevation in a 0.25-mm 2-D model of the human eye and head in response to an implanted retinal stimulator – part I: models and methods. IEEE Trans Antennas Propagation. 2003;51:2274–2285. [Google Scholar]
  • 32.Vorst AV, Rosen A, Kotsuka Y. RF/Microwave Interaction with Biological Tissues. Wiley; New York: 2006. [Google Scholar]
  • 33.Klem GH, Luders HO, Jasper HH, Elger C. The ten-twenty electrode system of the international federation of clinical neurophysiology. Electroencephalogr Clin Neurophysiol. 1999;(Suppl 52):3–6. [PubMed] [Google Scholar]
  • 34.Mullinger K, Debener S, Coxon R, Bowtell R. Effects of simultaneous EEG recording on MRI data quality at 1.5, 3 and 7 Tesla. Int J Psychophysiol. 2008;67:178–188. doi: 10.1016/j.ijpsycho.2007.06.008. [DOI] [PubMed] [Google Scholar]
  • 35.Vasios CE, Angelone LM, Purdon PL, Ahveninen J, Belliveau JW, Bonmassar G. EEG/(f)MRI measurements at 7 Tesla using a new EEG cap (“InkCap”) Neuroimage. 2006;33:1082–1092. doi: 10.1016/j.neuroimage.2006.07.038. [DOI] [PubMed] [Google Scholar]
  • 36.Lemieux L, Allen PJ, Franconi F, Symms MR, Fish DR. Recording of EEG during fMRI experiments: patient safety. Magn Reson Med. 1997;38:943–952. doi: 10.1002/mrm.1910380614. [DOI] [PubMed] [Google Scholar]
  • 37.Regan D. Human Brain Electrophysiology: Evoked Potentials and Evoked Magnetic Fields in Science and Medicine. Elsevier; New York: 1989. [Google Scholar]
  • 38.Caputa K, Okoniewski M, Stuchly MA. An algorithm for computations of the power deposition in human tissue. IEEE Antennas Propagation Mag. 1999;41:102–107. [Google Scholar]
  • 39.Federal Communications Commission (FCC) Office Engineering Technologies. OET Bull. Supplement C. Vol. 65. Washington, DC: 2001. Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency Electromagnetic Fields. [Google Scholar]
  • 40.American National Standard Institute (ANSI) American National Standard Safety Levels with respect to Human Exposure to Radio Frequency Electromagnetic Fields, 300 kHz to 100 GHz. Standard C95.1. 1982 [Google Scholar]
  • 41.International Commission on Non-Ionizing Radiation Protection (ICNIRP) Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz) Health Phys. 1998;74:494–522. [PubMed] [Google Scholar]
  • 42.American National Standard Institute (ANSI)/IEEE. 3 kHz to 300 GHz. Standard C95.1. IEEE; New York: 2006. Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields. [Google Scholar]
  • 43.Hirata A, Shirai K, Fujiwara O. On averaging mass of SAR correlating with temperature evaluation due to a dipole antenna. Prog Electromagn Res. 2008;84:221–237. [Google Scholar]
  • 44.McIntosh RI, Anderson V, McKenzie RJ. The use of temperature as a metric for the assessment of RF safety. Radiat Prot Australasia. 2008;25:9–21. [Google Scholar]
  • 45.Bassen H, Kainz W, Mendoza G, Kellom T. MRI-induced heating of selected thin wire metallic implants –laboratory and computational studies – findings and new questions raised. Minim Invasive Ther Allied Technol. 2006;15:76–84. doi: 10.1080/13645700600640931. [DOI] [PubMed] [Google Scholar]
  • 46.Railton CJ, Schneider JB. An analytical and numerical analysis of several locally conformal FDTD schemes. IEEE Trans Microwave Theory Tech. 1999;47:56–66. [Google Scholar]

RESOURCES