Abstract
Following our established theoretical model to deal with the second-harmonic generation (SHG) excited by a linearly polarized focused beam in type I collagen, in this paper, we further quantitatively characterize the differences between SHG emissions in type I collagen excited by collimated and focused beams. The effects of the linear polarization angle (α) and the fibril polarity characterized by the hyperpolarizability ratio ρ on SHG emission has been compared under collimated and focused beam excitation, respectively. In particular, SHG emission components along the i axis
(i = x,y,z), the induced SHG emission deviation angle γij, and the detected SHG signals (I2ω,ij) in the ij plane by rotating the applied polarizer angle φij have been investigated (i = x, x, y; j = y, z, z). Results show that under our simulation model, SHG emission in the xy plane, such as I2ω,x ,I2ω,y ,γxy and I2ω,xy varying as polarization angle (α) under collimated and focused light, presents no significant difference. The reverse of the fibril polarity has induced great impact on I2ω,x ,γxy and I2ω,xy in both collimated and focused light. I2ω,x and γxy show similarity, but I2ω,xy at α = 30° demonstrates a slight difference in focused light to that in collimated light. Under focused light, the reverse of fibril polarity causes obvious changes of the collected SHG intensity I2ω,xz and I2ω,yz at a special polarization angle α = 60° and γxz, γyz along α.
Keywords: Collimated beam, Focused beam, Polarization (α), SHG intensity
Introduction
Second-harmonic generation (SHG) is a nonlinear optical effect first recognized by Franken et al. in crystals in 1961, the year shortly after the demonstration of the laser [1]. In this process, two near-infrared incident photons are scattered by a material with noncentrosymmetric structural features into one emerging visible photon, which is at exactly half the excitation wavelength (twice the energy). This well-known nonlinear optical effect was then identified in biological systems; fibrillous collagen was identified about 10 years later in 1971 [2].
In such a nonlinear optical process, as it occurs by scattering rather than absorption and re-emission as in two-photon fluorescence processes, there is no energy loss during excitation to emission; SHG thus has unique advantages such as no photobleaching and no photoxicity caused in the specimen. SHG is the intrinsic signal induced by the interaction of photons on the optical properties of specimen itself, no additional staining on fluorochrome is required for its enhancement. Meanwhile, in light of the coherence preservation of the excitation light, SHG usually carries the directionality information that is associated with the characteristics of the specimen [3, 4].
SHG signals therefore have been widely exploited for imaging in biological tissues, especially in fibrillous collagen type I, one of the strongest SHG producers of biological specimens. SHG was first noticed by Fine and Hansen in fibrillous collagen type I early in 1971 based on the use of linearly polarized collimated light [2]. In 1978, Gannaway and Sheppard successfully demonstrated an SHG phenomenon through a microscope [5]. Since then, more researchers have been interested in taking advantage of the microscopic SHG signals to achieve three-dimensional high-resolution optical imaging in biological specimens. A great number of experiments of microscopic SHG imaging through fibrillous collagen type I have been conducted [6, 7]. At the same time, a series of theories of dealing with SHG based on linearly polarized focused light have been developed [8–10]. Yet, in a theoretical model to deal with polarization effects of focused light on SHG emission in fibrillous collagen type I, the type of biological specimen with nonplanar (but linear) homogenous distribution of scatterers (dipoles) has not been well established. On the contrary, such effects studies in collimated light have been well explored [11, 12]. Accordingly, how focused light influences SHG in fibrillous collagen type I has not been comprehensively compared to that of collimated light before. Therefore, based on our newly established theoretical model to deal with the SHG excited by a linearly polarized focused beam in type I collagen [13], in this paper, we further characterize the differences between SHG emissions in type I collagen excited by collimated and focused light. In particular, the effects of the linear polarization angle (α) and the fibril polarity characterized by the parameter of hyperpolarizability ratio (ρ) on SHG emission have been compared under collimated and focused beam excitation, respectively.
Theory of SHG in type I collagen under collimated and focused beam
Based on the structural features of type I collagen, it is assumed to be composed of highly organized fibrils that have a cylindrical rod-like shape. In this paper, the collagen fibrils are hypothetically of zero thickness and are composed of many infinitesimally small subunits (dipoles), each possessing C6 symmetry. A schematic coordinate system to describe SHG emission from such fibrils is shown in Fig. 1. The subunits (dipoles) align in the x direction and extend along the x direction. The x axis is thus also the polar symmetrical axis of the dipoles that conforms to cylindrical symmetry. The incident excitation field
propagates toward the z axis with a linearly polarized direction in the xy plane and a polarization angle α from the x axis. We suppose that the effective volume density NV of dipoles and the effective excited volume V under collimated and focused beams are the same, which means that the total number of dipoles N = NVV that contribute to the generation of second-harmonic light is identical for both sets of conditions.
Fig. 1.
Schematic diagram to characterize SHG emission from type I collagen fibrils under linearly polarized beam of collimated light and focused light, respectively. Collagen fibrils composed of dipoles are assumed to be aligned along the x direction. The light is linearly polarized in the xy plane
with angle α from the x direction and propagates in the z direction. Collimated light trace is represented by a dashed line, and the solid line represents a focused light trace
Electric dipole moment of the single dipole induced by fundamental electrical field
The overall electric dipole moment (or polarization) induced by a fundamental electric field is [14]
![]() |
1 |
Here,
is the permanent dipole moment and the frequency-dependent parameters α, β, and γ are the linear polarizability and first and second hyperpolarizability, respectively. The contribution to SHG is related to the third term by a single dipole (scatterer) in the fibrils, which can be described as [15]:
![]() |
2 |
βijk represents the first hyperpolarizabilities. The subscripts ijk refer to three axes in an orthogonal coordinate system; ε is the polarization direction. Because of Kleinman and cylindrical symmetry, only two terms in β related to μ remain:
![]() |
3 |
![]() |
4 |
where Eω,x and Eω,y are the fundamental field strengths polarized in the x and y directions, respectively.
SHG emission electric field induced by effective total dipole moment
Dipole moment thus induces the second-harmonic electrical field, the configuration of
is as follows [16]
![]() |
5 |
We define
, where ε0 is the free-space permittivity, c is the speed of light, ω is the frequency of the fundamental beam. The direction of the vector
denotes the SHG emission direction. ψ represents the angle between the
axis and the emission direction
of SHG. Sin ψ thus illustrates the projection relationship between the direction of the excited SHG dipole moment
and the emitted electric field
of SHG, and it is defined by the solid angle in Cartesian coordinates:
![]() |
6 |
where θ is the angle between the direction of SHG propagation and the direction of propagation of the incident beam (z axis), and ϕ is the angle between the emission plane (which is defined as the plane constituted by the direction of incident beam propagation and the direction of SHG propagation, as shown in Fig. 1) and the xz plane.
SHG electric field strength from collagen fibrils when illuminated by a collimated laser beam
When collimated light is applied, emitted SHG propagates in the direction of incident beam with no deviation, thus ψ = π/2. As a result, for the whole quantity N of excited scatterers (dipoles), the induced SHG electrical field is
![]() |
7 |
Correspondingly, according to Eqs. 3 and 4, the electric field of SHG along the x- and y-axes can be respectively written as
![]() |
8 |
![]() |
9 |
SHG electric field strength from collagen fibrils when illuminated by a focused laser beam
The single electric dipole moment induced by the fundamental electrical field has the same expressions as those demonstrated in Eqs. 3 and 4; however, under the focused laser beam case, the driving field
has a more complicated form, which can be well approximated by a three-dimensional (3D) Gaussian profile in the focus area as follows:
![]() |
10 |
is the electric field strength at the central point of the focused beam. kω = 2πnω/λω is the wave vector in a specimen with refractive index nω. ξ is the wave vector reduction factor due to focusing that accounts for a reduction in axial momentum by conversion to lateral momentum within the focus. wxy and wz are the focal ellipse in the lateral and axial directions, respectively. According to Eq. 5, the distribution of the SHG electric field induced by whole excited dipoles N at an observation point (r, θ, φ) is
![]() |
11 |
where
![]() |
12 |
N = VNV , and
is the active SHG volume [17].
![]() |
13 |
![]() |
14 |
![]() |
15 |
The amplitude and direction of SHG intensity
The intensity of SHG produced has the following relation with the electric field:
![]() |
16 |
SHG intensity illuminated by a collimated beam
Based on Eqs. 8 and 9, the x and y components of the SHG intensity under collimated beam excitation can be easily obtained as
![]() |
17 |
![]() |
18 |
The total SHG intensity thus has the following relationship with the linear polarization angle α under the collimated beam:
![]() |
19 |
SHG intensity illuminated by a focused beam
Due to the distribution of the SHG electrical field in three dimensions in the emission space, the total SHG intensity under the focused beam is the integration in solid angle as
![]() |
20 |
where
![]() |
21 |
Accordingly, three components of SHG intensity based on the coordinates x-, y-, and z-axes have been obtained (as for collimated light, we represent the excited SHG by the x and y components; for focused light, we present it in our previous paper [13] by the components perpendicular (s) and parallel (p) to the emission plane). In order to compare the SHG by focused light to that from collimated light, we project the components in the p and s directions to the x, y, and z directions. This detailed derivation process is revealed in the Appendix.
![]() |
22 |
![]() |
23 |
![]() |
24 |
Polarization direction of SHG emission
In this paper, the polarization direction of the emitted SHG is to be expressed through three deviation angles γij relative to the i axis in the Cartesian coordinate system, where when i is taken to be x, x, and y, j is correspondingly taken to be y, z, and z, respectively. In the ij plane, the maximum SHG intensity appears at the deviation angle γij, given by
![]() |
25 |
If a polarizer is applied that is parallel in the ij plane and has an angle ϕij with the i axis based on Malus’s law, the collected SHG emission signals as the angle of polarizer ϕij will be
![]() |
26 |
Comparison of SHG emission excited by collimated and focused beam
Influence of the linear polarization angle α
In this section, first, we discuss SHG emission intensity components along the x-, y-, and z axes as the linear polarization angle α excited by the collimated and focused beams.
To do this, we introduce a parameter ρ as the ratio of hyperpolarizability as
, and at the same time, we suppose ρ = 2.6 and βxyy = 1. Meanwhile,
in collimated light and
in focused light are defined for simplicity and comparison.
Figure 2a demonstrates the SHG intensity
as the polarization angle α excited by the collimated beam when ρ = 2.6, based on Eqs. 17, 18, and 19. We notice that I2ω,x has two symmetrical maximum values at α = 0° and 180° as well as two symmetrical minimum values at α = 90° and 270°, while I2ω,y shows four maximum values at
and four minimum values at
. I2ω has two maximum and two minimum values at the same excitation angles α as those of I2ω,x . It is clear that I2ω,x has a more dominant contribution than that of I2ω,y to the total SHG intensityI2ω.
Fig. 2.
Effects of the polarization angle α on the SHG emission intensity along the x direction
, the y direction
, and the total component
under collimated light when ρ = 2.6 (a). Effects of polarization angle α on the SHG emission intensity along the x direction
, y direction
, z direction
, and the total component
under focused light when ρ = 2.6 (b)
In the case of an excitation beam that is focused, SHG emission is three-dimensional, in which case there is an additional component along the z direction except components confined within two-dimensional xy plane as those in collimated case. The cause is due to the appearance of the additional angle modulation term A(θ,ϕ) as demonstrated in Eq. 14, which results in an off z axis propagation direction of SHG emission. Based on Eqs. 22, 23, and 24, SHG intensities I2ω,x , I2ω,y , I2ω,z and I2ω are shown as a function of α (Fig. 2b) after excitation by a focused incident beam. We notice that I2ω,x and I2ω,y keep the similar shapes as those in collimated beam under rotation of the polarization angle α. However, the maximum values appearing at α = 0° and 180° as well as the minimum values appearing at α = 90° and 270° of I2ω,x no longer coincide with those of total SHG intensity I2ω observed in the collimated excitation beam, since I2ω,z now contributes to the total SHG intensity I2ω . I2ω,z varies less steeply along α and its strength is much smaller than I2ω,x .
Figure 3a shows the influence of the fundamental polarization angle α (0°~360°) on the SHG polarization deviation angle γxy in a collimated beam. It is noticed that γxy changes periodically as a function of α. γxy has its maximum deviation degree of 21° at α = 60°, 120°, 240°, and 300° and the minimum deviation degree γxy = 0° at α = 0°, 90°, 180°, and 270°. Figure 3b shows the influence of α on γij in the case of focused beam. The influence of α on γxy is similar to that under the collimated beam case. In the xz plane, the polarization angle α has an even and negligible effect on the SHG deviation angle γxz. In the yz plane, four maximum angles γyz (around 86°) appear when polarization angle α is 0°, 90°, 180°, and 270° and when the polarization angle α is 60°, 120°, 240°, and 300°. γyz presents four minimum angles (around 15°). We notice that α has different impact range on γyz, γxy, and γxz, and γyz is the most significantly affected; its change range is about 70°, while that of γxy is about 21° and that of γxz is almost zero. In addition, we also notice that α has a nonsymmetrical effect on both γyz and γxy from x and y axes. When α is off from the x axis, the deviation angle γyz and γxy decreases or increases slowly to the peak value, while when α is off from the y axis, the deviation angle γyz and γxy decreases or increases dramatically to the peak value. In other words, the turning points of α locate far away from the x axis and closer to the y axis rather than in the middle of the x and y axes.
Fig. 3.
Effects of excitation polarization angle α (0°~360°) on the SHG polarization deviation angle γxy in a collimated beam when ρ = 2.6 (a). Effects of excitation polarization angle α (0°~360°) on SHG polarization deviation angle γxy, γxz, and γyz in a focused beam when ρ = 2.6 (b)
Finally, in practice, SHG emission signals are usually collected by applying a polarizer in front of the detector. To understand the influence of the polarizer direction ϕij on the collected SHG signals for reference to practical applications, the collected SHG signals varying as the polarizer direction ϕij rotating from 0° to 360° under different fundamental polarization angles α = 0°, 30°, 60°, and 90° have been investigated.
Figure 4a demonstrates the SHG intensity
as a function of the rotation of the angle ϕxy of the polarizer under α = 0°, 30°, 60°, and 90° in collimated beam. We notice that when the fundamental polarization α is along the x axis (α = 0°) or the y axis (α = 90°), SHG emission has a similar mode where the maximum SHG intensity appears at ϕxy = 0° and 180° and no SHG emission collected at ϕxy = 90° and 270°. However, the strength of collected SHG signals at α = 90° is much weaker than that when α = 0°, which indicates that SHG has the most efficiency when the illumination electric-field polarization is orientated parallel to the principal axis of the fibrils (α = 0°). As the illumination beam polarizes at α = 30° and 60°, the collected SHG signals present a slight peak intensity deviation of 9° and 21°, respectively, from the fibrils axis.
Fig. 4.
a SHG intensity I2ω as a function of the rotation of the angle ϕxy of the polarizer under α = 0°, 30°, 60°, and 90° in collimated beam when ρ = 2.6. b SHG intensity I2ω,xy as a function of the rotation of the angle ϕxy under α = 0°, 30°, 60°, and 90° in focused beam when ρ = 2.6. c SHG intensity I2ω,xz as a function of the rotation of the angle ϕxz under α = 0°, 30°, 60°, and 90° in focused beam when ρ = 2.6. d SHG intensity I2ω,yz as a function of the rotation of the angle ϕyz under α = 0°, 30°, 60°, and 90° in focused beam when ρ = 2.6
Figure 4b–d demonstrates the SHG intensity I2ω,xy , I2ω,xz , and I2ω,yz as the rotation of the angle ϕxy, ϕxz, and ϕyz of the polarizer under α = 0°, 30°, 60°, and 90°, respectively, in focused beam. In the xy plane (Fig. 4b), collected SHG emission mode as the rotation of angle ϕxy is similar to that under collimated beam, but the derivation angle of the peak intensity has a slight change to be 8° under focused beam rather than 9° in collimated beam at α = 30°. Figure 4c shows the collected SHG emission I2ω,xz as the rotation of the polarizer ϕxz in the xz plane. It demonstrates that the SHG peak intensity that could be collected has an almost constant deviation at any polarization angle α, which is 4° at α = 0°, 90°, and 5° at α = 30° and 60° from the fibrils axis (x axis) towards the z axis. The SHG emission I2ω,yz in the yz plane is demonstrated in Fig. 4d, where we notice that the SHG peak intensity appears at the direction almost along the z axis (ϕyz = 86°) when the fundamental polarization α is along the x axis (α = 0°) or y axis (α = 90°). The SHG peak intensity deviates from the y axis at 30° and 14° at polarization α = 30° and α = 60° respectively.
Influence of the fibril polarity characterized by hyperpolarizability ratio ρ
β is the hyperpolarizability of the collagen fibrils which is related to the electronic transition in the material; thus, the ratio
is a reflection of the biological structure and chemical features of the collagen. It has been proved that ρ increases with the increasing of age and
has been predicted to vary between −3 to 3 in collagen [18]. The negative value of ρ suggests that the polarity of the fibrils reverses from the excited electrical field. In this section, to understand the influence of the fibril polarity on SHG emission under collimated and focused beam, the negative value of ρ = −2.6 is applied for the exploration of SHG emission.
Figure 5a shows the SHG intensity
as a function of the polarization angle α excited by collimated beam when ρ = −2.6. In this situation, I2ω,x is the component that has been affected more obviously compared to that of ρ = 2.6. The shape of I2ω,x as α changes becomes more complicated; two additional peak values of I2ω,x appear at, α = 90° and 270°; there are no at α = 0° and 180°. Accordingly, the total SHG intensity presents two peaks of the same size of I2ω,x at α = 90°and 270° instead of two minimum values when ρ = 2.6. Also, in the focused beam case when ρ = −2.6 as shown in Fig. 5b, I2ω,x is heavily affected. The intensity distribution at different polarization angles α becomes different from that at ρ = 2.6, where two peak intensities suddenly appear at α = 90° and 270°. Nevertheless, the fibril polarity has a similar influence on I2ω,x under collimated and focused beam conditions.
Fig. 5.
a Effects of fibril polarity (ρ = −2.6) on SHG emission intensity along the x direction
, y direction
, and the total component
under collimated light at excitation polarization angle α = 0°, 30°, 60°, and 90°. b Effects of fibril polarity on SHG emission intensity along the x direction
, y direction
, z direction
, and the total component
under focused light at excitation polarization angle α = 0°, 30°, 60°, and 90°
Figure 6a shows the SHG polarization deviation angle γij (in the xy plane) induced by excited polarization angle α (0°~360°) when ρ = −2.6 in the collimated beam. We notice that in the collimated case, the reverse of fibril polarity to the excited electrical field has little impact on the overall shape of the SHG polarization deviation angle γxy along all the excited polarization angle α; however, the maximum deviation angle γxy reaches to 90° instead of 21°. In the focused beam case, as shown in Fig. 6b, also, the fibril polarity has an impact on the SHG maximum deviation angle γxy, which reaches to 90°. Fibril polarity has no apparent impact on γyz except to cause a decrease in the minimum value from 14° to 4°, while it does have significant impact on the SHG polarization deviation angle γxz, where γxz has a similar pattern as γxy, a fully different shape compared to the flat curve when ρ = 2.6.
Fig. 6.
Effects of fibril polarity (ρ = −2.6) on the SHG polarization deviation angle γxy in collimated beam as excitation polarization angle α varies from 0°~360° (a). Effects of fibril polarity (ρ = −2.6) on SHG polarization deviation angle γxy, γxz, and γyz in focused light as excitation polarization angle α varies from 0°~360° (b)
Figure 7a demonstrates the influence of fibril polarity (ρ = −2.6) on collected SHG intensity distribution I2ω,xy by rotating of the angle ϕxy of the polarizer under α = 0°, 30°, 60°, 90° in collimated beam. When α = 0° and 90°, the fibril polarity has no visible impact on I2ω,xy . ϕxy changes from 9° to 15° at α = 30° and from 21° to 90° at α = 60° from ρ = 2.6 to ρ = − 2.6. The collected SHG intensity distribution
as rotation of the polarizer angle ϕij in the corresponding xy, xz, and yz planes in the case of focused beam have been demonstrated in Fig. 7b–d, respectively. As shown in Fig. 7b, we notice that the fibril polarity in the focused beam has a similar impact on I2ω,xy as that in the collimated beam when α = 0° and 90° as well as α = 60°. However, the angle of polarizer ϕxy corresponding to the collected peak SHG emission shifts from 8° to 13° at α = 30° in focused light rather than 9° to 15° in collimated light. Comparing Fig. 7c to Fig. 7c, we see that when α = 60°, I2ω,xz is the one that has been influenced greatly by the fibril polarity, the corresponding ϕxz for the collected peak SHG emission shifts from 5° to 85°. By the comparison of Fig. 7d to Fig. 4d, we realize that fibril polarity has no influence on I2ω,yz at polarization angle α = 0° and a slight influence at α = 90° that causes a shift of ϕyz = 86° to ϕyz = 85°. In addition, the reverse of the fibril polarity causes the shift of ϕyz from 28° and 14° to 30° and 4° at α = 30° and 60°, respectively.
Fig. 7.
a SHG intensity I2ω as the rotation of the angle ϕxy of polarizer under α = 0°, 30°, 60°, and 90° in collimated beam when ρ = −2.6. b SHG intensity I2ω,xy as the rotation of the angle ϕxy under α = 0°, 30°, 60°, and 90° in focused beam when ρ = −2.6. c SHG intensity I2ω,xz as the rotation of the angle ϕxz under α = 0°, 30°, 60°, and 90° in focused beam when ρ = −2.6. d SHG intensity I2ω,yz as the rotation of the angle ϕyz under α = 0°, 30°, 60°, and 90° in focused beam when ρ = −2.6
Discussion and conclusions
Base on our established theory of SHG emission in collagen type I as the variation of the excitation polarization angle α under focused light, in this paper, we further characterize the differences between SHG emissions in type I collagen influenced by excitation polarization angle α and fibril polarity excited by collimated and focused beams, respectively.
The comparison of the emission SHG excited by collimated beams and focused beams demonstrates the following results. When the illumination is collimated, SHG propagates along the direction of the incident beam; thus, the SHG electric field is confined to be polarized in the xy plane, and two components of SHG emission along the x and y directions are included. However, when the illumination is focused, SHG propagates along a defined solid angle, which can be split into components along three axes; an extra component along the z axis exists in the focused beam.
On SHG emission components along the x and y directions, focusing induces a change of the relationship of I2ω,x with the total SHG emission I2ω , where the maximum SHG emission at α = 0° and 180° as well as the minimum SHG emission at α = 90° and 270° of I2ω,x are no longer coincident with I2ω . Obviously, I2ω,z has an additional contribution to I2ω . Here, we should mention that Tiaho et al. [19] have their experimental conclusion that SHG has the maximum magnitude when the polarization angle of incident light is 40°, while our simulation results show that SHG has its maximum magnitude when the polarization angle is 0°. The apparent discrepancy is due to the difference of the prerequisite conditions. In the paper by Tiaho et al., they introduce a concept of the effective orientation angle θe of the harmonophores, which means that the angle was formed by the scatterers (dipoles) with the fibril axes. It is around 50–60°. In our theoretical simulation study, as we assume that the scatterers (dipoles) are distributed uniformly along the fibrils axes, based on this concept, the effective orientation angle is 0°. Because the initial polarization direction of the dipoles is different, the maximum SHG emission direction is different.
On the xy plane, the deviation angles γxy of the SHG emission along excitation polarization angle α has a similar pattern under collimated and focused beams where the maximum deviation angle γxy = 21° appears when α locates off the x axis 60° and the minimum deviation angle γxy = 0° appears when α locates along the x and y axes. The deviation angles γxz and γyz in the focused beam show different affected patterns by excitation polarization angle α. α has a non-symmetrical affect on γyz from the x and y axes. When α is off from the x axis, γyz decreases slowly to the peak value, while when α is off from the y axis, γyz decreases dramatically to the peak value. On the other hand, α has almost uniform influence on γyz.
The collected SHG emission signals vary as the variation angle ϕ of the polarizer as well as the polarization angle α. In xy plane, the variation pattern under all demonstrated cases of polarization angle α shows similarity under collimated and focused beams, which shows that the collected SHG emission signals have peaks locating at two symmetrical angles ϕxy. However, there is a slight difference. The angle ϕxy correspondingly shifts from 0–9–21–0° in collimated light, yet 0–8–21–0° in focused light at different polarization angles α of 0–30–60–90°. In the xz and yz planes, the collected SHG emission signals show different patterns of the shift angles ϕxz and ϕyz in focused beam.
The fibril polarity has caused significant influence on I2ω,x in both collimated and focused light cases. Two additional peak values of I2ω,x appear at α = 90° and 270° with no peaks appearing at α = 0° and 180°. No visible effect is found on I2ω,y and I2ω,z in collimated light and focused light. Also, fibril polarity has induced the maximum value of γxy to increase from 21° to 90° in both cases. Moreover, in the focused light case, fibril polarity causes dramatic changes of γxz and affects the minimum values of γyz, decreasing them from 14° to 4°.
Generally, the effect of fibril polarity on I2ω,ij is different at different polarization angle α in both collimated and focused light case. The reverse of fibril polarity has slight different influence on I2ω,xy at α = 30° comparing that in focused light to that in collimated light, where the peak intensity of I2ω,xy shifts from 8° to 13° in focused light and 9° to 15° in collimated light. However, at α = 60°, it has a significant influence on I2ω,xy in both collimated light and focused light, which induces the peak intensity of I2ω,xy shifting from 21° to 90°. Influence that is more obvious occurs on I2ω,xz and I2ω,yz at α = 60° in focused light.
Acknowledgement
The authors gratefully thank the National Natural Science Foundation of China (Grant Nos. 30470495 and 30940020) for their support.
Appendix
We apply a matrix to represent SHG components along parallel (p) and perpendicular (s) directions of emission plane to the x, y, and z direction:
![]() |
27 |
![]() |
28 |
![]() |
29 |
![]() |
30 |
while
![]() |
31 |
Therefore,
![]() |
32 |
![]() |
33 |
![]() |
34 |
![]() |
35 |
![]() |
36 |
Finally,
![]() |
37 |
![]() |
38 |
![]() |
39 |
As a result, we can derive emission SHG intensity in the x, y, and z direction components
![]() |
40 |
![]() |
41 |
![]() |
42 |
Contributor Information
Ying Chang, Email: yingchang_1984@yahoo.cn.
Xiaoyuan Deng, Email: xiaoyuandeng@gmail.com.
References
- 1.Franken PA, Hill AE, Peters W.Generation of optical harmonics Phys. Rev. Lett. 19617118. 10.1103/PhysRevLett.7.1181961PhRvL...7..118F [DOI] [Google Scholar]
- 2.Fine S, Hansen WP.Optical second harmonic generation in biological tissues Appl. Opt. 1971102350. 10.1364/AO.10.0023501971ApOpt..10.2350F [DOI] [PubMed] [Google Scholar]
- 3.Zipfel WR, Williams RM, Webb WW. Nonlinear magic: multiphoton microscopy in the biosciences. Nat. Biotechnol. 2003;21(11):1368–1376. doi: 10.1038/nbt899. [DOI] [PubMed] [Google Scholar]
- 4.Helmchen F, Denk W. Deep tissue two-photon microscopy. Nat. Methods. 2005;2(12):932–940. doi: 10.1038/nmeth818. [DOI] [PubMed] [Google Scholar]
- 5.Gannaway JN, Sheppard CJR. Second-harmonic imaging in the scanning optical microscope. Opt. Quantum. Electron. 1978;10:435–439. doi: 10.1007/BF00620308. [DOI] [Google Scholar]
- 6.Erikson A, Ortegren J, Hompland T, Davies CDL, Lindgren M. Quantification of the second-order nonlinear susceptibility of collagen I using a laser scanning microscope. J. Biomed. Opt. 2007;12(4):44002. doi: 10.1117/1.2772311. [DOI] [PubMed] [Google Scholar]
- 7.Odin C, Guilbert T, Alkilani A, Boryskina PO, Fleury V, Grand LY.Collagen and myosin characterization by orientation field second harmonic microscopy Opt. Express 20081616151–16157. 10.1364/OE.16.0161512008OExpr..1616151O [DOI] [PubMed] [Google Scholar]
- 8.Williams RM, Zipfel WR, Webb WW. Interpreting second harmonic generation images of collagen I fibrils. Biophys. J. 2005;80(2):1377–1386. doi: 10.1529/biophysj.104.047308. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Stoller P, Reiser KM, Celliers PM, Rubenchik AM. Polarzation-modulated second harmonic generation in collagen. Biophys. J. 2002;82:3330–3342. doi: 10.1016/S0006-3495(02)75673-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Yew EYS, Sheppard CJR.Second harmonic generation polarization microscopy with tightly focused linearly and radially polarized beams Opt. Commun. 2007275453–457. 10.1016/j.optcom.2007.03.0652007OptCo.275..453Y [DOI] [Google Scholar]
- 11.Roth S, Freund I.Second harmonic generation in collagen J. Chem. Phys. 197970041637–1643. 10.1063/1.4376771979JChPh..70.1637R [DOI] [Google Scholar]
- 12.Freund I, Deutsch M, Sprecher A.Connective tissue polarity. Optical second-harmonic microscopy, crossed-beam summation, and small-angle scattering in rat-tail tendon Biophys. J. 198650693–712. 10.1016/S0006-3495(86)83510-X1986BpJ....50..693F [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Chang Y, Chen CS, Chen JX, Jin Y, Deng XY. Theoretical simulation study of linearly polarized light on microscopic second harmonic generation (SHG) in collagen type I. J. Biomater. Appl. 2009;14(4):44016. doi: 10.1117/1.3174427. [DOI] [PubMed] [Google Scholar]
- 14.Willetts A, Rice JE, Burland DM.Problems in the comparison of theoretical and experimental hyperpolarizabilities J. Chem. Phys. 19929797590–7599. 10.1063/1.4634791992JChPh..97.7590W [DOI] [Google Scholar]
- 15.Moreaux L, Sandre O, Mertz J.Membrane imaging by second-harmonic generation microscopy J. Opt. Soc. Am. B 2000171685–1694. 10.1364/JOSAB.17.0016852000OSAJB..17.1685M [DOI] [Google Scholar]
- 16.Hecht E. Optics. Beijing: Higher Education Press; 2005. p. 46. [Google Scholar]
- 17.Mertz J, Moreaux L.Second-harmonic generation by focused excitation of inhomogeneously distributed scatterers Opt. Commun. 2001196325–330. 10.1016/S0030-4018(01)01403-12001OptCo.196..325M [DOI] [Google Scholar]
- 18.Odin C, Grand YL, Renault A, Gailhouste L, Baffet G.Orientation fields of nonlinear biological fibrils by second harmonic generation microscopy J. Microsc. 200822932–38. 10.1111/j.1365-2818.2007.01868.x2416922 [DOI] [PubMed] [Google Scholar]
- 19.Tiaho F, Recher G, Rouede D.Estimation of helical angles of myosin and collagen by second harmonic generation imaging microscopy Opt. Express 2007151912286–12295. 10.1364/OE.15.0122862007OExpr..1512286T [DOI] [PubMed] [Google Scholar]

















































