Abstract
We introduce and demonstrate a method of measuring small force gradients acting on a harmonic oscillator in which the force-gradient signal of interest is used to parametrically up-convert a forced oscillation below resonance into an amplitude signal at the oscillator’s resonance frequency. The approach, which we demonstrate in a mechanically detected electron spin resonance experiment, allows the force-gradient signal to evade detector frequency noise by converting a slowly modulated frequency signal into an amplitude signal.
Many precision measurements rely on registering a signal of interest as a change in the amplitude or frequency of an oscillator. In theory, the ultimate precision of such measurements is limited by quantum-mechanical measurement noise; in practice, the precision achievable in an oscillator measurement is often limited by thermomechanical position fluctuations,1, 2 detector noise,1 or environmental fluctuations.3, 4, 5 If one is using the oscillator to detect a time-varying force, then the requirements for achieving thermally limited or quantum-limited sensitivity can be relaxed by using parametric amplification6, 7, 8 to raise both the displacement signal and the oscillator’s position-fluctuation noise above the detector’s noise floor. Here we propose and demonstrate using parametric amplification to evade both detector and surface frequency noise when using a cantilever to detect a force-gradient signal.
Forces acting on microcantilevers are routinely measured at the thermomechanical limit, where the smallest detectable force is set by the force fluctuations, PF=4kbT Γ, giving rise to the friction Γ experienced by the oscillator. Given the finite sensitivity of displacement sensors, however, achieving thermally limited sensitivity in a force measurement usually requires modulating the signal force at or near the cantilever frequency, fc, where the resulting displacement is amplified by the mechanical quality factor of the cantilever. In many cases, such modulation is inconvenient or impossible for fc≳1 kHz. In magnetic resonance force microscopy (MRFM), for example, force modulation at fc is often impossible because of the sample’s unfavorable spin relaxation times.9, 10 In electric force microscopy, force detection is inconvenient because of the undesirably long natural response time of the cantilever near resonance1, 11 and the finite charging time of the sample. These shortcomings are obviated by detecting the signal as a (slowly modulated) force gradient,1, 9, 10, 11 δk. In a force-gradient experiment, the cantilever is driven into self-oscillation via positive feedback1 and the (modulated) force-gradient signal shifts the instantaneous frequency of the cantilever, δfc≈fcδk∕2kc, where kc is the cantilever spring constant.
Achieving thermally limited sensitivity in a force-gradient experiment remains challenging, however. This is illustrated in Fig. 1, in which we plot the power spectrum of cantilever frequency fluctuations seen in three representative magnetic resonance force microscope experiments (detailed below). In the first experiment (solid black line), the cantilever was brought to a height h=30 nm above a gold-coated surface, driven to a root-mean-square amplitude ofxrms=73 nm, and the tip-sample potential adjusted to minimize tip charge and therefore frequency noise. The observed frequency noise (solid black line) is a sum of following three contributions: (1) thermomechanical position fluctuations (dotted line) with power spectrum , (2) detector noise (dashed line) having a power spectrum , where is the detector noise written as an equivalent cantilever displacement fluctuation, and (3) surface noise (dotted-dashed line) arising from uncompensated tip charge coupling to fluctuating electric field gradients produced by the sample.5
Figure 1.
Power spectrum of cantilever frequency fluctuations with (peaks S1 and S2) and without (solid baseline) a modulated force-gradient signal present. Dashed-dotted black line: surface noise. Dotted black line: thermal noise . Dashed line: detector noise . Noise at frequencies away from the signal peak in the blue and green traces has been removed for clarity.
In the second and third experiments, the magnetization of unpaired electron spins in the sample was modulated to create the force-gradient signal at 6.28 Hz (blue line) and 100 Hz (green line), respectively, in Fig. 1. We can see in the figure that even though the modulation frequency was chosen optimally, the noise in the 6.28 Hz experiment was nevertheless dominated by surface noise; consequently, the observed signal-to-noise ratio S1∕NS is smaller than the thermally limited signal-to-noise ratio S1∕NT by a factor of 14. Modulating at 100 Hz does avoid surface noise but the noise is dominated instead by detector noise and the observed signal-to-noise ratio is even worse: S2∕ND is smaller than S2∕NT by a factor of nearly 103.
To achieve a thermally limited signal-to-noise ratio in this representative force-gradient experiment would require modulating at a frequency fmod⪢80 Hz and operating with a detector having a position sensitivity of , 26 times smaller than we currently achieve using optical fiber interferometry. Here we introduce a parametric amplification scheme that (1) is compatible with a force-gradient measurement; (2) can be used with a modulated signal, allowing the signal to evade the effect of surface frequency noise; and (3) converts a frequency signal to an amplitude signal at fc, evading detector frequency noise. In contrast with other applications of parametric amplification in which an externally supplied force gradient amplifies a small force signal, here the signal of interest acts as the amplifier.
As in previous work,10 a force-gradient signal was generated by interacting unpaired electron spins in a gold-coated film of TEMPAMINE with a high-compliance magnetic tipped cantilever. Experiments were carried out in vacuum (P=10−6 mbar) and at cryogenic temperatures (T=8 K). The cantilever had fc=4829 Hz, kc=7.8×10−4 N∕m, a mechanical quality factor Q=3.8×104, and a nickel tip of radius r=2 μm. A (swept) magnetic field of B0=0.50 T to 0.85 T was applied to polarize the sample spins and a pulsed 17.28 GHz transverse magnetic field from a half-wave microstripline resonator was applied to saturate sample spins. Spin magnetization μz in the sample interacted with the magnetic field from the magnetic tip to create a force gradient, Δkspin, which shifted the resonance frequency of the cantilever;9, 10 where the sum is over all spins in resonance at the given applied field. A small potential, Vdc≈1 V, was applied between the cantilever and the gold sample coating to control the charge on the cantilever tip. The cantilever was driven below resonance at a frequency fd=48fc∕49=4730 Hz by applying an oscillating voltage Vdc=33 Vrms from a waveform generator (Agilent 33250A) to a nearby wire. The force applied to the cantilever from the drive wire is
| (1) |
where C′=∂C∕∂x is the derivative of the wire-cantilever capacitance with respect to the direction of cantilever motion. The resulting amplitude of motion at frequency fd, xd=χ(fd)Fd∕kc≈C′VdcVacχ(fd)∕kc, with the susceptibility and Fd the component of Fwire oscillating at fd; xd was 99 nm for experiments here. During the experiment, the effective Q of the cantilever was reduced to Qeff=3×103, using negative feedback applied by a piezo at the cantilever base. The sample spin magnetization, μz, was modulated at a frequency fp=fc∕49=fd∕48=98.55 Hz by pulsing the microwave field in synchrony with the cantilever motion.10
The equation of motion governing the cantilever displacement x is
| (2) |
where we have expressed frequencies in angular units, ωp=ωc−ωd and includes only the first Fourier component of the pulse modulation. We look for a solution of the form and are particularly interested in two coefficients: and , the amplitude of the spin-induced oscillation at the cantilever’s resonance frequency. We find, for Qeff⪢1,
| (3) |
The central prediction of Eqs. 2, 3 is that the spin-induced spring constant modulation δkspin at frequency fp acts to up-convert some of the oscillation at frequency fd to an oscillation at frequency fc. The data in Fig. 2 verify this prediction. For reference, Fig. 2a shows the power spectrum of thermomechanical motion of the nascent cantilever. In Fig. 2b we see the thermomechanical motion peak near 4829 Hz broadened by negative feedback and, in addition, a large peak near 4730 Hz due to the applied drive. In Fig. 2c the microwaves have been turned on and an additional narrow peak can be seen, near 4829 Hz, on top of the damped thermomechanical motion. This narrow peak demonstrates the up-conversion of a force-gradient frequency signal at fp to an amplitude signal at fc.
Figure 2.
Power spectral density of cantilever motion: (a) thermomechanical displacement fluctuations; (b) with negative feedback active while the cantilever was driven at a frequency fd=fc−fp=4730 Hz; and (c) identical condition to (b) but with a microwave field pulsed at fp=98.55 Hz. Conditions: B0=0.5925 T and h=200 nm.
We next demonstrated that our parametric up-conversion technique was effective at evading detector noise, allowing us to modulate fast enough to also avoid surface noise. Figure 3a is a plot, versus external magnetic field, of the Fourier component at fp=6.28 Hz of the spin-induced cantilever frequency shift. This modulation rate was chosen to minimize the frequency noise due to the surface and the detector (Fig. 1). The shape of the resulting magnetic resonance signal arises from considering that spins are in resonance with the sum of the tip field and the external field and weighting each spin in resonance by the second derivative of the tip field at the location rj of each spin.10 Increasing the modulation rate to fp=100.57 Hz, so that surface frequency noise will no longer affect the measurement, does not improve the signal-to-noise [Fig. 3b]. This is as expected given the large detector noise apparent at f=100 Hz in Fig. 1.
Figure 3.
Cantilever magnetic resonance recorded via modulated force-gradient detection and parametric up-conversion amplitude detection. (a) Optimal frequency-shift measurement (fp=6.28 Hz, RMS amplitude 99 nm and a background of −14.8 mHz subtracted). (b) Surface-noise evading frequency shift measurement [fp=100.57 Hz but otherwise identical to (a)]. (c) Detection of magnetic resonance via parametric up-conversion of a frequency-shift signal (open circles, fp=98.55 Hz and a 0.53 nm background subtracted.) The solid line is the signal predicted from Eq. 3 and the frequency-shift signal in (a), scaled to account for the difference in modulation frequencies (Ref. 12) (scale factor=0.70). Conditions: 2.5 mT∕pt field step and detection bandwidth b=1 Hz.
Using a spin modulation rate of fp=98.55 Hz and the parametric up-conversion scheme described above, in contrast, succeeds in evading surface noise. The up-converted amplitude experiment, Fig. 3c, has a signal-to-noise ratio equivalent to that seen in the optimal frequency-shift experiment, Fig. 3a. The observed signal (circles) agrees quantitatively with the signal predicted (line) using Eq. 3, calculated using the observed frequency-shift signal in Fig. 3a, xd, and Qeff as inputs.12
Given the power spectrum of frequency fluctuations observed in Fig. 1, we initially expected the up-conversion experiment to have a signal-to-noise closer to the thermomechanical limit (S2∕NT in Fig. 1). We found, however, that the signal-to-noise in the up-conversion experiment depended critically on the quality of the sine wave used to drive the cantilever off-resonance at fd. To explain this finding, we calculated the power spectrum Pδxn(ωc) of cantilever amplitude noise arising from voltage fluctuations in the drive source by adding a voltage-noise term, δVn(t), to Eq. 1. Voltage fluctuations will be a negligible source of noise when , the power spectral density of cantilever thermomechanical fluctuations. Meeting this condition requires a source with noise voltage power spectrum that satisfies
| (4) |
We conclude that the drive oscillator will contribute negligibly to cantilever position fluctuations at resonance if its voltage amplitude noise is much smaller than −97 dBc∕Hz at fc=4829 Hz and much smaller than −125 dBc∕Hz at 9559 and 98.55 Hz. Although the amplitude noise for our untuned audiofrequency drive oscillator is unspecified, we note that this level of amplitude noise is challenging to achieve even with a high-Q tuned radiofrequency oscillator.
In conclusion, we have introduced an approach for detecting minute force gradients acting on an harmonic oscillator and have demonstrated that the approach enables the measurement to evade surface and detector frequency noise. We have thus achieved a similar result to Budakian et al.,13 but without the need to modulate spin magnetization at fc and in an experiment which is conceptually simpler, easier to implement, and applicable to other forms of scanned probe microscopy beyond MRFM. Calculations indicate that further improvements in the electrical noise in the drive oscillator should enable the technique to better approach thermally limited sensitivity. Although we have chosen to demonstrate this technique using MRFM, we believe the approach is generally applicable to any oscillator force-gradient measurement.
Acknowledgments
We acknowledge National Institutes of Health Grant No. 5R01GM-070012 and Army Research Office MultiUniversity Research Initiative Grant No. W911NF-05-1-0403. This work was performed in part at the Cornell NanoScale Science and Technology Facility, a member of the National Nanotechnology Infrastructure Network, which is supported by National Science Foundation Grant No. ECS-0335765. E.W.M. and S.L. contributed equally to this work.
References
- Albrecht T. R., Grütter P., Horne D., and Rugar D., J. Appl. Phys. 69, 668 (1991). 10.1063/1.347347 [DOI] [Google Scholar]
- Hutter J. and Bechhoefer J., Rev. Sci. Instrum. 64, 1868 (1993). 10.1063/1.1143970 [DOI] [Google Scholar]
- Stipe B. C., Mamin H. J., Stowe T. D., Kenny T. W., and Rugar D., Phys. Rev. Lett. 87, 096801 (2001). 10.1103/PhysRevLett.87.096801 [DOI] [PubMed] [Google Scholar]
- Kuehn S., Loring R. F., and Marohn J. A., Phys. Rev. Lett. 96, 156103 (2006). 10.1103/PhysRevLett.96.156103 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yazdanian S. M., Hoepker N., Kuehn S., Loring R. F., and Marohn J. A., Nano Lett. 9, 2273 (2009). 10.1021/nl9004332 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rugar D. and Grütter P., Phys. Rev. Lett. 67, 699 (1991). 10.1103/PhysRevLett.67.699 [DOI] [PubMed] [Google Scholar]
- Carr D. W., Evoy S., Sekaric L., Craighead H. G., and Parpia J. M., Appl. Phys. Lett. 77, 1545 (2000). 10.1063/1.1308270 [DOI] [Google Scholar]
- Zalalutdinov M., Olkhovets A., Zehnder A., Ilic B., Czaplewski D., Craighead H. G., and Parpia J. M., Appl. Phys. Lett. 78, 3142 (2001). 10.1063/1.1371248 [DOI] [Google Scholar]
- Garner S. R., Kuehn S., Dawlaty J. M., Jenkins N. E., and Marohn J. A., Appl. Phys. Lett. 84, 5091 (2004). 10.1063/1.1762700 [DOI] [Google Scholar]
- Moore E. W., Lee S. -G., Hickman S. A., Wright S. J., Harrell L. E., Borbat P. P., Freed J. H., and Marohn J. A., Proc. Natl. Acad. Sci. U.S.A. 106, 22251 (2009). 10.1073/pnas.0908120106 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kikukawa A., Hosaka S., and Imura R., Appl. Phys. Lett. 66, 3510 (1995). 10.1063/1.113780 [DOI] [Google Scholar]
- See supplementary material at http://dx.doi.org/10.1063/1.3465906 for measurement of the scale factor.
- Budakian R., Mamin H. J., and Rugar D., Appl. Phys. Lett. 89, 113113 (2006). 10.1063/1.2349311 [DOI] [Google Scholar]



