Abstract
Purpose: Small animal micro-CT imaging is being used increasingly in preclinical biomedical research to provide phenotypic descriptions of genomic models. Most of this imaging is coincident with animal death and is used to show the extent of disease as an end point. Longitudinal imaging overcomes the limitation of single time-point imaging because it enables tracking of the natural history of disease and provides qualitative and, where possible, quantitative assessments of the effects of an intervention. The pulmonary system is affected by many disease conditions, such as lung cancer, chronic obstructive pulmonary disease, asthma, and granulomatous disorders. Noninvasive imaging can accurately assess the lung phenotype within the living animal, evaluating not only global lung measures, but also regional pathology. However, imaging the lung in the living animal is complicated by rapid respiratory motion, which leads to image based artifacts. Furthermore, no standard mouse lung imaging protocols exist for longitudinal assessment, with each group needing to develop their own systematic approach.
Methods: In this article, the authors present an outline for performing longitudinal breath-hold gated micro-CT imaging for the assessment of lung nodules in a mouse model of lung cancer. The authors describe modifications to the previously published intermittent isopressure breath-hold technique including a new animal preparation and anesthesia protocol, implementation of a ring artifact reduction, variable scanner geometry, and polynomial beam hardening correction. In addition, the authors describe a multitime-point data set registration and tumor labeling and tracking strategy.
Results:In vivo micro-CT data sets were acquired at months 2, 3, and 4 posturethane administration in cancer mice (n=5) and simultaneously in control mice (n=3). 137 unique lung nodules were identified from the cancer mice while no nodules were detected in the control mice. A total of 411 nodules were segmented and labeled over the three time-points. Lung nodule metrics including RECIST, Ortho, WHO, and 3D volume were determined and extracted. A tumor incidence rate of 30.44±1.93 SEM for n=5 was found with identification of nodules as small as 0.11 mm (RECIST) and as large as 1.66 mm (RECIST). In addition, the tumor growth and doubling rate between months 2–3 and 3–4 were calculated. Here, the growth rate was slightly higher in the second period based on the 3D volume data (0.12±0.13 to 0.13±0.17 μl) but significantly less based on the linear diameter metrics [RECIST (0.33±0.19 to 0.17±0.18 mm); Ortho (0.24±0.15 to 0.16±0.15 mm)], indicating the need to understand how each metric is obtained and how to correctly interpret change in tumor size.
Conclusions: In conclusion, micro-CT imaging provides a unique platform for in vivo longitudinal assessment of pulmonary lung cancer progression and potentially tracking of therapies at very high resolutions. The ability to evaluate the same subject over time provides for a sensitive assay that can be carried out on a smaller sample size. When integrated with image processing and analysisroutines as detailed in this study, the data acquired from micro-CT imaging can now provide a very powerful assessment of pulmonary disease outcomes.
Keywords: micro-CT, mouse lung imaging, lung cancer, small animal imaging, breath-hold imaging
INTRODUCTION
There are now many examples utilizing “micro” imaging modalities for the assessment of normal and diseased pulmonary phenotypes including micro-MRI for emphysema,1, 2 asthma,3 and lung cancer;4 micro-CT for emphysema,5, 6, 7, 8 pulmonary fibrosis,9 and lung cancer;10, 11, 12, 13, 14, 15 micro-PET for lung cancer;16, 17 and recently, fluorescence molecular tomography for visualization of inflammation.18
Longitudinal pulmonary imaging in small animals, especially the mouse, is useful for tracking disease progression and the response of the disease process to potential therapies. However, very few longitudinal pulmonary studies have been presented to date using micromodalities. Froese et al.6 presented serial micro-CT scans from Smad3 knockout mice and showed the superior utility of three-dimensional density measurements as opposed to two-dimensional and histologic assessment as a surrogate for development of airspace enlargement. Hori et al.13 described the use of micro-CT for serial assessment of lung lesions in a mouse model of lung cancer and presented quantitative analysis of intrasubject tumor growth heterogeneity over a period of 20 weeks. Fushiki et al. presented the utility of micro-CT quantitative imaging for monitoring of lung cancer progression and suppression with confirmation using bioluminescence and histology.12, 19
Performing longitudinal mouse lung imaging to provide valuable quantitative data in the microscale range (10–100 μm pixels) consistently can be difficult, as there are no standard mouse lung imaging protocols, with each group needing to develop and refine their own systematic approach. Micro-CT imaging is a useful modality for assessment of pulmonary structure in small animals, especially the mouse. Since micro-CT is an x-ray based imaging modality, if calibrated correctly the image content can be directly related to the density of the object under assessment. This fundamental property has been exploited for many years in clinical imaging, where it serves to qualitatively show, for instance, that muscle is different to bone or lung. Quantitative micro-CT scanning is now being employed in small animal micro-CT imaging for assessment of bone density, lung density, and cardiac output.20
Depending on the application and the technical requirements such as resolution, sensitivity, and acquisition speed, various forms of respiratory gating have been implemented. These include synchronous gating of an anesthetized free-breathing mouse21 and prospective gating of intubated and ventilated mice at either end inspiratory, end expiratory,10, 11, 22, 23, 24 or during controlled breath-holds.25 Further, retrospective gating has been implemented during the reconstruction process26, 27, 28, 29 and recently, a retrospective phase-correlated approach where subject-specific optimization led to greater reduction of motion blurring.30
Recently, we introduced an intermittent isopressure breath-hold (IIBH) gating technique for in vivo imaging of the lung using a commercial micro-CT system.31 In brief, the mouse is intubated, paralyzed, and ventilated using a computer controlled ventilator and x-ray projections are acquired during periodic breath-holds at user defined airway pressures. This approach provided high-resolution micro-CT imaging of an almost motion-free lung in vivo and has now been used by other groups with encouraging results.25, 32
In this study, we present in vivo micro-CT imaging for longitudinal assessment of a lung cancer model and provide details for animal preparation, anesthesia, image acquisition, processing, and analysis. Investigation of a urethane induced mouse lung cancer model shows the unique ability of micro-CT imaging to identify and track individual lung nodule growth. As compared to previous studies describing the assessment of lung cancer, examples shown in this study have been evaluated at a very early stage due to the resolution advantage—Within 8 weeks of urethane administration for the lung cancer model. The implementation of the IIBH technique in conjunction with appropriate calibration and processing routines provided for a highly reproducible assay, enabling semiautomated segmentation and tracking of individual lung nodules over time. In addition, nodules could be accurately identified and related to their unique anatomical location such as their confining lobe, a feature that requires motion-free high-resolution imaging for segmentation of lobar fissure.
MATERIALS AND METHODS
Micro-CT imaging
A modified version of the IIBH micro-CT imaging technique, as described by Namati et al.,31 was used for the longitudinal micro-CT scans in this study. In brief, the micro-CT x-ray source was set to 60 kVp and 500 μA tube current. The CCD detector was set to 500 ms exposure, binning of 2, and the system was configured to acquire 720 projections over 200° (half-scan) with 100 bright and dark calibration images. Each projection produced a 1536×1024 pixel image with a pixel spacing dimension of 28×28 μm2. This protocol provides a MTF resolution at 10% of 55 μm. The MTF was calculated as previously described31 using a 10 μm tungsten wire phantom (QRM, Nuremburg, Germany). Respiratory gating was performed using a combination of a computer controlled Flexivent (Scireq, Montreal, Quebec, Canada) ventilator, LABVIEW program, and electronic solenoid triggering system. X-ray projections are essentially acquired at periodic breath-holds at a constant user defined airway pressure. Total scan time for each data set was 45 min.
Animal preparation
Experimental procedures carried out in this study had been approved by the University of Iowa Animal Care Committee.
Cancer mice
Normal A∕J mice (n=8) aged between 6 and 8 weeks and weighing 20–22 g were used in this study. Each mouse was initially sedated using 3%–5% isofluorane, the mice were then weighed and n=5 were injected I.P. with 1 mg∕g urethane (Sigma-Aldrich, St Louis, MO) dissolved in 0.9% NaCl saline. Mice were tagged with ear tags (Harvard Apparatus, Massachusetts, USA) and placed back into their respective cages to recover. As urethane is a known anesthetic agent, mice administered with urethane required approximately 30 min longer to regain consciousness.
Ethyl carbamate, the formal chemical name for urethane, was a common anesthetic agent used on animals. Once discovered to be the cause of lung adenomas in mice and rats, its use as an anesthetic agent diminished.33 It was later proposed and demonstrated that vinyl carbamate epoxide, an oxidized off product of ethyle carbamate through two catalysations by the enzyme cytochrome P450 2E1, was the mutagenic and carcinogenic agent that would then react with DNA. The products of this reaction have miscoding properties and are considered the predominant source of the DNA lesions formed from ethyl carbamate.34, 35, 36, 37
Prior to imaging, mice were initially sedated using 3%–5% isofluorane, weighed, and injected intramuscular (IM) in the right hind leg with 1.0 mg∕kg medetomidine hydrochloride and 50.0 mg∕kg ketamine. This anesthetic regiment was determined empirically in order to reduce total anesthesia while maintaining sufficient sedation for restraining mice during the intubation phase. Mice were intubated with a 22-gauge 1 in. Teflon catheter through the visual aid of a custom miniature fiber optic bronchoscope. Unlike previously described intubation techniques using a single optical fiber to illuminate the vocal chords,38, 39 here we constructed a flexible fiber optic bronchoscope using 10 000 optical fibers enabling clear visualization of the vocal chords through an external viewfinder. The fiber optic bronchoscope was constructed using a 0.45 mm 10 000 fiber image guide (Fujikura, Atlanta, GA) and a 0.35 mm∕2 mm focal length GRIN lens (NSG, Somerset, NJ). The bronchoscope was fed through the 22-gauge catheter providing a unique ability to identify the trachea and feed the intubation tube over the bronchoscope. An image of the vocal chords of a mouse using the microbronchoscope is shown in Fig. 1. Clear identification of the vocal chords and trachea provides for swift placement of the intubation tube. A snake light was also placed under the chest resulting in light travel through the trachea to enable fast identification of the vocal chords.
Figure 1.
(a) Example image using the microbronchoscope revealing the vocal chords of a mouse during intubation. (b) Dotted outline of vocal chords; the central glow is the diffusion of light through the chest cavity from an external snake light.
Triple antibiotic ointment was placed over the eyes to maintain moisture during the imaging phase, as the anesthesia inhibits spontaneous blinking. Mice were placed supine onto a polystyrene bed and mounted onto the micro-CT carbon fiber animal stage. The limbs were secured using tape and ECG and temperature sensors were attached for physiologic monitoring using a BioVet C1 data acquisition system (Supertron Technologies, Newark, NJ). Respiratory paralysis was achieved by administration of 0.5 mg∕kg of rocuronium IM in the left hind leg. The tracheal catheter was then connected to a computer controlled Flexivent ventilator (Scireq, Montreal, Quebec,Canada) via 6 ft of 1∕8 in. ID tubing. The mice were ventilated at 120 br∕min and a tidal volume of 15 ml∕kg. Isofluorane was set to 1%–2% for maintenance of sedation throughout the imaging protocol.
A heat lamp and cylindrical chamber was used to maintain body temperature between 35° and 37° centigrade during scanning to reduce hypothermia, which is a major contributor to mortality during anesthesia. Finally, 0.5 ml of PBS was injected subcutaneously above the abdomen in order to prevent dehydration during the imaging period.
We have now examined over 50 mice using the aforementioned protocol and found a decrease in mortality from approximately 30% to <10% after implementing the new anesthesia, temperature maintenance, and intubation approach. In addition, our recovery period decreased from 2 to 3 h postimaging when using ketamine and xylazine 87.5 mg∕kg∕12.5 mg∕kg in combination with pancuronium 0.1 mg∕g to 15–30 min when using ketamine and medetomidine 50.0 mg∕kg∕1 mg∕kg with 0.5 mg∕kg rocuronium as detailed above.
Image processing
Micro-CT images underwent several processing steps prior to reconstruction.
Ring artifact reduction
Ring artifacts are commonly observed in reconstructed micro-CT images due to the inconsistent signals between detector elements. The conventional approach for minimizing ring artifacts is a flat-field calibration step. Two sets of images are acquired to perform the flat-field calibration. A dark field is acquired with no x rays and represents the detector dark current and digitization noise while a bright field is acquired with the x rays on and provides an unattenuated projection. In this study, each projection x ray is log calibrated against a dark field and a bright field image acquired at the beginning of the scan. However, as these are normally acquired at the beginning of the scan, slight changes to the detector sensitivity over the duration of the mouse scan can result in ring artifacts in the reconstructed images. This effect is clearly evident if the system has not been allowed to reach a steady state temperature.
Several prereconstruction40, 41 and postreconstruction42, 43, 44 techniques have been previously described. Here, we perform a prereconstruction low pass filtering step similar to Refs. 40, 41 on the raw sinograms prior to reconstruction.
For each sinogram we calculate the average intensity profile across each row using
| (1) |
where RSx is the average row sinogram value at column x. We then perform a low pass filter on this profile using a mean filter of length 9 and subtract this filtered version with the original averaged profile RSx. The final profile which represents the difference of the average row sinogram profile and the smoothed sinogram profile is subtracted from every row in the sinogram. The filter width is related to the smoothing factor and preservation of object features. A value of 9 was found experimentally to result in a qualitative reduction of spikes representing line artifacts while preserving changes in object attenuation. Rivers described a similar technique40 using a low pass filter in the Fourier domain and Boin et al.41 using a moving average filter in the real space.
A reconstructed example of using only the flat-field calibration versus the addition of this filtering step is shown in Figs. 2a, 2b, respectively. In addition, Fig. 2c represents the difference image between Figs. 2a, 2b.
Figure 2.
(a) Reconstructed micro-CT image with no ring artifact reduction. (b) Sinogram low pass filtering ring artifact correction. (c) Difference image between (a) and (b).
Variable geometry correction
The Siemens MicroCAT II incorporates a stationary specimen stage and a rotating source and detector couple. Imperfections in the gantry in addition to gravitational forces result in source-to-object-to-detector distance variations leading to suboptimal reconstructed images. Here we present a technique for determining this variation in terms of a dynamic center offset, source-to-object, and source-to-detector distance look up table that can be incorporated into the Feldkamp reconstruction algorithm (Exxim CC, Pleasanton, CA). The lookup table was created using a custom 12.7 mm brass ball bearing phantom scanned using the in vivo imaging protocol outline above. The projection images were processed and the area of the ball was converted into a translational shift versus projection angle. The area-to-translational shift was calculated using a prior calibration step that included acquiring images of the ball bearing phantom at two known relative distances and applying the concept of similar triangles as illustrative in Fig. 3.
Figure 3.
[(a) and (c)] Schematic geometric illustration for calculating the source-to-object and source-to-detector distance using the ball bearing phantom x-ray projections at two known distances. Ball phantom x-ray projection, (b) position 1: Closer to detector; (d) position 2: Closer to source.
Here, a is the known ball phantom diameter, d is the known translation of the phantom toward the detector, and finally the known detector pixel dimension is used to calculate b1 and b2. Using similar triangles, x and y are calculated using the following equations:
| (2) |
| (3) |
The value for the center offset, source-to-object, and source-to-detector distance was calculated for each angle and averaged over ten complete scans. A standard deviation of 0.473 mm was measured for the translation shift (mm) from the piercing point over 360°. The mean of the standard deviation across ten scans for the translation shift at each corresponding angle was measured to be 0.0141 mm, indicating the reproducibility of the shift between scans. An example of the result using the dynamic center offset, source-to-object, and source-to-detector distance reconstruction for both an ex vivo fixed and in vivo breath-hold lung data set are illustrated in Figs. 4a, 4b and Figs. 4c, 4d, respectively. Here the fixed sample was scanned with a binning of 1 resulting in a reconstructed pixel dimension of 14×14 μm2 and illustrates the improvement while eliminating motion related effects. The difference between the in vivo images is subtle, but still evident as indicated by the arrows to the vessel.
Figure 4.
(a) Standard reconstruction and (b) reconstruction using a dynamic center offset, source-to-object, and source-to-detector distance look up table. The arrow indicates a region that is significantly improved using the dynamic lookup tables. Scale bar=2 mm.
Beam hardening correction
Beam hardening is a common problem associated with the use of a polychromatic x-ray source. When performing quantitative analysis of density and, in particular, regional differences in density within and across subjects, reducing the effects of beam hardening are important. In addition to a hardware filter (e.g., 0.5 mm aluminum plate), there are also several computational techniques that have been previously introduced to reduce beam hardening. The simplest is a widely implemented a priori, nonlinear polynomial correction.45 This approach was implemented here using a custom 10, 20, and 30 mm water equivalent cylindrical phantom. The phantom was scanned using the in vivo imaging protocol outline above and a third order polynomial was fit to the x-ray attenuation versus thickness. These coefficients were then used in combination with the EXXIM Cobra (Exxim CC, Pleasanton, CA) reconstruction algorithm. Figure 5 represents the density line profile across the center of the 10, 20, and 30 mm phantom in the axial plane (a) before and (b) after correction. A second order polynomial regression for each reveals a clear improvement in the corrected versus noncorrected profiles. Here the R2 value was 0.1213, 0.1890, and 0.2231 versus 0.0445, 0.0002, and 0.0239 for the corrected. Subjectively, we can see the density profiles for the different size cylinders have been flattened and the difference in the average density between the profiles has been reduced.
Figure 5.
Density line profile across the center of a 10, 20, and 30 mm beam hardening phantom in the axial plane (a) before and (b) after beam hardening correction.
Hounsfield unit calibration
The attenuation values from reconstructed micro-CT images were calibrated and scaled into Hounsfield units (HU) by ensuring that water is 0 HU and air is −1000 HU. Attenuation values representing each pixel were initially normalized and truncated into integer values by the following expression:
| (4) |
where μ(x,y) represents the high precision floating point attenuation number for each pixel (x,y) before conversion, CT(x,y) is the CT number in Hounsfield units, and μwater is the attenuation coefficient for water.
HU calibration and scaling was performed using a water equivalent plastic phantom prior to each imaging session. Calibration into HU provides a convenient scale for segmentation and assessment of tissue with unique density values. In addition, HU calibrations allows for quantitative comparison between data sets and time-points.
Longitudinal lung registration
Registration between multiple time-point in vivo micro-CT data sets was performed using the AMIRA VISUALIZATION software platform. Data sets were registered through an iteratively optimized rigid transformation process based on mutual information. Registration was performed on the entire data set in three dimensions; however, only voxels bound between −1024 and 500 HU were sampled for calculating the mutual information. This reduced the bias due to high contrast structures other than the lung, such as the ribs and spine (1000HU+). The registration process was performed iteratively by varying the resampling factor of the reference data set from 16× to 1×. After each iteration, the previously calculated coordinate transform (three translation and three rotation parameters) of the model data set was used as the starting point for the next optimization. The final registered data sets was then resampled into the reference lattice using a Lanczos interpolation algorithm and finally exported as RAW 16 bit images.46 A flow diagram of this process is shown in Fig. 6. In all cases, the second time-point data set was used as the reference for registration.
Figure 6.
Rigid registration flow diagram for registering multitime-point micro-CT data sets to one another.
Tumor segmentation
In vivo micro-CT image data sets were imported into a prototype oncology program (OncoCare, Siemens, NJ, USA). Tumors were located manually and traced using a semiautomated segmentation routine.47, 48 First a straight line or stroke across the cross-section of the tumor was drawn, keeping each end of the line close to the boundary of the tumor. The program then automatically performs three main tasks. First, the algorithm learns the rough gray level distribution of the lesion from the stroke using a multiregion fuzzy connectedness algorithm. Second, the lesion is segmented on the current 2D plane and on 2D planes orthogonal to it using a shortest path algorithm. At the end of each segmentation, the gray level distribution for the lesion and the background is calculated and fed back into the algorithm. Finally, a random walker algorithm utilizing the seeds generated from the 2D segmentations are used to compute the 3D segmentation. Here, the algorithm determines the probability that a random walker initiating its walk from each voxel first arrives at a lesion seed before arriving at a background seed. The automated tracings were then manually inspected and borders were altered, as needed, using a series of manual tools.
Each tumor was given a unique label as shown in Table 1.
Table 1.
Nodule labeling nomenclature.
| Confining lobe | Anatomical position | Special attributes |
|---|---|---|
| RL1=right cranial lobe | S=solitary | A=nodule surrounded by atelectasis |
| RL2=right accessory lobe | H=heart | D=nodule has disappeared |
| RL3=right caudal lobe | P=pleura | |
| RL4=right middle lobe | L=lobe fissure | |
| LL1=left lung |
As an example, RL3_4_P would represent the fourth nodule located in the right caudal lobe adjacent to the pleura. Identification of anatomical structures such as lobar fissures provides a powerful means for compartmentalizing results for regional assessment of tumor incidence and growth rate.
Multiple size metrics for each segmented tumor were exported for further analysis, including tumor diameter based on the Response Evaluation Criteria in Solid Tumors (RECIST) (14–16), orthogonal (Ortho) tumor diameter, the World Health Organization (WHO) response criteria (14–16), 3D tumor volume, mean tumor HU density of voxels within the volume, and standard deviation of HU density within each tumor volume. This data was exported and formatted for further statistical analysis using SPSS (SPSS Inc., Chicago, IL).
The WHO criterion provides a bidimensional metric based on the product of the maximum diameter in an axial image by the largest perpendicular diameter in the same image, the RECIST criteria provides a unidimensional metric using the largest diameter in a single plane, while the Ortho metric provides the maximum orthogonal diameter. In the two later cases measurements are a diameter value and commonly limited to the transverse plane.
Intraobserver and interobserver study
To assess the reproducibility and variation in measuring tumor metrics using the aforementioned segmentation tools, intraobserver and interobserver studies were carried out. Each observer (n=3) segmented tumors (n=32) from a single mouse at the final time-point using the semiautomated algorithm. The original expert observer also retraced the tumors from the same data set two additional times in two different sittings separated by a period of three weeks.
Tumor labels across the observers and sittings were consolidated and statistical analysis was undertaken to determine the reliability of repeated measures. The intraclass correlation coefficient (ICC) and coefficient of variation (CV) were determined for both the intraobserver and interobserver results and shown in Tables 4, 5, respectively.
Table 4.
Intraobserver intraclass correlation coefficient and coefficient of variation results for repeated (n=3) segmentation and measurement of tumors (n=32) by the same observer.
| Variable | Mean | ICC | CV (%) | ||
|---|---|---|---|---|---|
| Estimate | 95% CI | Estimate | 95% CI | ||
| Volume (μl) | 0.271 | 0.97 | (0.94, 0.98) | 23.1 | (19.7, 27.9) |
| RECIST (mm) | 0.826 | 0.97 | (0.95, 0.99) | 6.8 | (5.8, 8.2) |
| Ortho (mm) | 0.621 | 0.96 | (0.93, 0.98) | 9.1 | (7.8, 11.0) |
| WHO (mm2) | 0.602 | 0.97 | (0.95, 0.99) | 14.6 | (12.4, 17.6) |
Table 5.
Interobserver intraclass correlation coefficient and coefficient of variation results for segmentation and measurement of tumors (n=32) by three observers.
| Variable | Mean | ICC | CV (%) | ||
|---|---|---|---|---|---|
| Estimate | 95% CI | Estimate | 95% CI | ||
| Volume (μl) | 0.260 | 0.95 | (0.90, 0.97) | 30.5 | (26.0, 36.9) |
| RECIST (mm) | 0.813 | 0.95 | (0.91, 0.97) | 9.0 | (7.7, 10.9) |
| Ortho (mm) | 0.606 | 0.95 | (0.91, 0.97) | 10.6 | (9.0, 12.8) |
| WHO (mm2) | 0.580 | 0.96 | (0.92, 0.98) | 18.6 | (15.5, 21.9) |
RESULTS
Lung cancer four-month three-time-point study
Micro-CT data sets from each urethane induced A∕J mouse scanned at 2, 3, and 4 months posturethane administration were registered with one another as outlined in the methods. Axial images revealing the progressive growth of a pulmonary nodule for one of the urethane mice are shown in Fig. 7.
Figure 7.
Axial images from the same urethane induced A∕J mouse at 2, 3, and 4 months posturethane administration, acquired using the in vivo micro-CT imaging technique. A growing pulmonary nodule located in the right azygous lobe can be identified at each time-point as indicated with the arrow (window=3000, level=250).
Using the registered in vivo micro-CT data sets of the same urethane induced A∕J mouse at 2, 3, and 4 months, individual tumors were segmented using a semiautomated approach and visualized as shown in Fig. 8.
Figure 8.
[(a), (b), (d), and (f)] Axial and coronal three-dimensional reconstructions of a mouse lung tumor (RL2_2_P) as identified at 2, 3, and 4 months posturethane administration, respectively.
Quantification of segmented tumors as shown in Fig. 8 was undertaken and common tumor metrics including the RECIST, Ortho, WHO, and 3D volumes were extracted. The RECIST, Ortho, WHO, 3D diameter, and 3D volume measurements for three out of a total of 31 nodules for one urethane induced A∕J mouse have been plotted in Figs. 9b, 9c, 9d, 9e, 9f, respectively. A combined graph including the RECIST, Ortho, WHO, and 3D diameter has also been generated in Fig. 9a. The 3D diameter was calculated based on the measured 3D volume and provides a “linear” representation for comparative purposes. In addition, the standard deviation from the intraobserver study for each metric are shown as error bars for the RECIST, Ortho, WHO, and 3D volume graphs.
Figure 9.
(b) RECIST (mm), (c) Ortho (mm), (d) WHO (mm2), (e) 3D diameter, and (f) 3D volume (μl) growth rate for the tumor shown in Fig. 8 plus two out of the 31 total tumors from the same mouse. (a) Combined RECIST, Ortho, WHO, and 3D diameter growth rates. Error bars represent the ± standard deviation from the intraobserver study for each metric, respectively.
Table 2 represents an overview of the statistics for tumors identified and tracked over months 2, 3, and 4 posturethane administration in lung cancer mice (n=5). 137 unique lung nodules were identified and a total of 411 nodules were segmented and labeled over the three-time-point study. Nodules as small as 0.11 mm and as large as 1.66 mm were identified and segmented. A mean RECIST (mm), Ortho (mm), WHO (mm2), and 3D Volume (μl) of 0.65±0.01, 0.48±0.01, 0.36±0.02, and 0.13±0.01 (SEM) were measured, respectively, from all tumors across all mice over the three time-points.
Table 2.
Overview of lung nodule metrics for identified and tracked nodules over months 2, 3, and 4 posturethane administration for n=5 mice.
| Scan month | Volume (μl) | RECIST (mm) | Ortho (mm) | WHO (mm2) | HU (mean) | HU (std) | |
|---|---|---|---|---|---|---|---|
| Month 2, N=137 | Mean | 0.025 | 0.384 | 0.278 | 0.121 | −299 | 126 |
| Std. error of mean | 0.004 | 0.019 | 0.016 | 0.014 | 23 | 4 | |
| Minimum | 0.003 | 0.112 | 0.101 | 0.020 | −653 | 56 | |
| Maximum | 0.212 | 0.860 | 0.711 | 0.611 | 36 | 216 | |
| Month 3, N=137 | Mean | 0.110 | 0.640 | 0.461 | 0.329 | −242 | 158 |
| Std. error of mean | 0.012 | 0.022 | 0.018 | 0.024 | 16 | 4 | |
| Minimum | 0.003 | 0.250 | 0.141 | 0.035 | −560 | 71 | |
| Maximum | 0.571 | 1.204 | 0.922 | 1.031 | 116 | 250 | |
| Month 4, N=137 | Mean | 0.206 | 0.773 | 0.584 | 0.502 | −206 | 176 |
| Std. error of mean | 0.022 | 0.023 | 0.020 | 0.034 | 15 | 5 | |
| Minimum | 0.003 | 0.180 | 0.158 | 0.033 | −690 | 78 | |
| Maximum | 1.515 | 1.664 | 1.373 | 2.285 | 176 | 411 | |
Table 3 represents an overview of the tumor growth and doubling rates from all nodules tracked over months 2, 3, and 4 posturethane administration in lung cancer mice (n=5). The mean and standard deviations for growth and doubling rates have been calculated for each segment, i.e., months 2–3 and 3–4 for each metric.
Table 3.
Tumor growth and doubling rate from nodules over months 2, 3, and 4 posturethane administration for n=5 mice.
| Volume (μl) | RECIST (mm) | Ortho (mm) | WHO (mm2) | ||
|---|---|---|---|---|---|
| Growth rate, months 2–3 | Mean (per month) | 0.124 | 0.336 | 0.249 | 0.292 |
| Std deviation | 0.138 | 0.198 | 0.153 | 0.234 | |
| Growth rate, months 3–4 | Mean (per month) | 0.131 | 0.178 | 0.169 | 0.236 |
| Std deviation | 0.174 | 0.182 | 0.152 | 0.268 | |
| Doubling rate, months 2–3 | Mean (months) | 1.3 | 2.3 | 2.2 | 1.3 |
| Std deviation | 1.6 | 2.2 | 1.0 | 1.2 | |
| Doubling rate, months 3–4 | Mean (months) | 1.2 | 13.9 | 6.5 | 5.0 |
| Std deviation | 4.9 | 85.3 | 28.6 | 18.6 |
Tables 4, 5 represent the results for the intraobserver and interobserver study, respectively. Here, the intraclass correlation coefficient and the coefficient of variation have been calculated and presented for the segmentation and measurement of 32 tumors. Table 4 was compiled from the repeated measures (n=3) from an expert observer while Table 5 was compiled from the single execution by three different observers.
DISCUSSION
Many disease conditions generate negative implications on normal lung structure and function, yet comprehensive longitudinal assessment of this organ is difficult because of its heterogeneity and significant motion in vivo. In this paper, we have presented a new methodology for longitudinal micro-CT lung imaging and associated quantitative descriptions and have applied these successfully to tracking the growth of lung nodules in a mouse model of lung cancer.
In our previous paper describing the IIBH technique, the reproducibility of measuring lung volume and air fraction using lung density was assessed in the same mice through acquisition and analysis of multiple micro-CT scans for four different gating techniques. Based on the two standard deviation Bland–Altman analysis, the IIBH technique was found to be more reproducible for both lung volume (38.13 μl versus 11.9 μl) and air content (13% versus 3.1%) metrics for nongated versus IIBH gating, respectively. In addition, the short-term reproducibility of both the image acquisition using the IIBH technique and image analysis based on calibrated HU density values were presented.31
This previous study provided a good foundation for the longitudinal assessment of tumor size using the semiautomated segmentation approach carried out here. Both the lung volume and density are important features that affect both the real tumor size and the tumor segmentation algorithm described in Sec. 2. In particular, when a user selects a region of interest by drawing a line across the tumor, the algorithm first uses the different density information to determine the correct seed locations for the lesion versus normal tissue. Based on this approach, it is critical that lung density values are reproducible across time-points. In addition, differences in lung volume can clearly affect the real tumor size due to distending forces from the surrounding parenchyma at different volumes or inflation pressures. Therefore, maintenance and reproducibility of lung volume and density are of primary concern for semiautomated tumor size measurements across time-points.
In this study, significant development was undertaken to the anesthesia and animal preparation protocols, where both the mortality and efficient recovery of the mouse postimaging were critically important for longitudinal imaging. A review of established anesthesia procedures49, 50, 51 as well as empirical testing was undertaken. In particular, the IIBH technique required paralysis of the mouse for stable breath-hold periods. Here rocuronium was utilized as opposed to pancuronium as the half life was much shorter (∼60 vs ∼200 min). This allowed for faster recovery (15–30 min as opposed to 2–3 h) and spontaneous breathing postmicro-CT scanning. In addition, mice were intubated and therefore required a short acting general anesthetic and analgesic that would cover both the intubation phase and the animal preparation period. Here we used a combination of medetomodine hydrochloride and ketamine. Medetomidine was utilized as opposed to xylazine, due to its higher affinity for alpha-2 receptors and the availability of a reversal agent (atipamezole) for fast postrecovery. A custom built microbronchoscope was also constructed in order to aid in fast and minimal risk intubations. Finally, maintaining an optimal body temperature during imaging increased the metabolism of the anesthetic and paralytic agents aiding in fast postrecovery. This combination of animal preparation and anesthesia resulted in a lower mortality (<10% as opposed to ∼30%), enabling high-resolution breath-hold imaging for longitudinal studies.
Although reconstructed images are, by their nature, artificial, practical steps can be taken to reduce inconsistencies and artifacts associated with micro-CT images within and across specimens. Such steps are paramount if quantitative analysis is to be made with a degree of certainty. Here we presented additional processing steps including ring artifact reduction, variable scanner geometry correction, and polynomial beam hardening correction to increase the resolution and consistency of the reconstructed images. With the ubiquitous use of micro-CT systems for research, such steps are now either available or will become standard features in commercial systems.
The prereconstruction ring artifact correction approach was chosen as opposed to postreconstruction techniques through qualitative assessment of the final reconstructed image and quantitative comparison of processing time. It was found that using a mean filter on the average row sinogram profile provided both a fast and effective approach as opposed to a Fourier domain low pass filtering step as previously described by Rivers,40 which also induced additional artifacts due to the Gibbs phenomenon. A similar result was also described by Boin et al.41 who found that filtering in the real space provided for qualitatively better images.
Cone beam computed tomography (CBCT) has become a well established technique for portable tomographic imaging in the clinical setting. Previous studies have described geometric calibration techniques for C-arm systems to improve reconstructed images.52, 53 The EXXIM reconstruction platform used in the Siemens micro-CAT II system was originally designed for C-arm CBCT reconstruction and has several features that are not utilized in the Siemens micro-CT user interface. Here we describe a simple technique to create the dynamic source-to-object, source-to–detector, and center offset lookup tables for integration with the EXXIM reconstruction platform. Although the Siemens gantry proved to be quite stable with less than 1 mm of shift at the piercing point, an improvement in images acquired from a fixed sample can be appreciated as illustrated in Figs. 4a, 4b. A fixed sample was used here to eliminate the effects of motion providing a better comparison for illustrative purposes. The benefits of using this geometric correction on the in vivo data sets were subtle but still evident as shown in Figs. 4c, 4d. Once the lookup table has been calculated, there is limited overhead in performing the reconstruction with the geometric correction.
Investigation of lung cancer at an early stage is of great importance, as our understanding of the early progression of this disease is limited. Evaluation of mouse models of lung cancer have provided further understanding of the genetic processes involved that lead to known phenotypes.54, 55 The ability to assess the same lung tumor from a very early stage in vivo using micro-CT imaging will enable a new approach to experimental studies. Here we have shown the utility of micro-CT imaging to identify and track individual tumor growth rates with high spatial resolution. Through implementation of a 3D rigid registration process, nodules could be confidently identified across time-points. Using a prototype oncology package, tumors were successfully segmented and important metrics were extracted, including the 3D volume for each nodule.
From the RECIST and Ortho growth plots in Figs. 9b, 9c, we can see that although the nodule shown in Fig. 7 (RL2_2_L) does not appear to have grown between months 3 and 4, the multidimensional WHO data in Fig. 9d, and more prominently the volumetric measurement and plot in Fig. 9f, reveal a clear increase in size over this period. Furthermore, the three-dimensional reconstruction (Fig. 8) of the tumor clearly presents the increasing size when visualized as a volume.
The 3D diameter Fig. 9e was calculated based on the 3D volume to provide a linear measure for comparison to the RECIST and Ortho metrics as shown in Fig. 9a. Since we expect the tumors in this model at the sizes reported here to be spherical, it is not surprising that the 3D diameter is comparable to the RECIST and Ortho metrics. However, in the case of nonspherical tumors, we would expect that the 3D volume and 3D diameter value to be significantly different to the RECIST and Ortho metrics, with the volume measures providing greater sensitivity.
A mean overview of the tumors identified and tracked in this study is presented in Table 2. Here we can identify the smallest and largest nodule measured to be 0.11 and 1.66 mm, respectively, using the RECIST criteria. It can also be seen that the mean Hounsfield value is increasing over months 2, 3, and 4 indicating the tumors are becoming more solid. In addition, the standard deviation is increasing over this period, revealing a greater intratumor and∕or intertumor tissue heterogeneity. Tumor growth and doubling rates were also calculated for each tumor across all mice with the average and standard deviation reported in Table 3. The tumor growth rate based on the volume measures shows a slight increase between months 2–3 (0.12±0.13 μl) and 3–4 (0.13±0.17 μl), while the RECIST (0.33±0.19 to 0.17±0.18 mm), Ortho (0.24±0.15 to 0.16±0.15 mm), and WHO (0.29±0.23 to 0.23±0.26 mm2) measures all show a decrease in growth rate between the same periods. The doubling rate reveals a similar trend, where the rate is decreasing for the volume measure and increasing for the RECIST, Ortho, and WHO measures. In addition, the doubling rate, which is a normalized value of time but not size, is quite different across the metrics particularly for the second period. This is most evident when comparing the volume versus the RECIST and Ortho measures. The underestimation of the growth rate using the unidimensional metrics is quite evident from these results and reveals the greater sensitivity of volume measurements for this task. However, it should be noted, as reported and discussed below, that volume measures are also more sensitive to observer and∕or segmentation variability.
The reliability of the semiautomated segmentation approach for measuring lung nodules was found to be very high based on the intraclass correlation coefficient for both the interobserver (>0.96) and intraobserver (>0.95) studies as shown in Tables 4, 5. This was not surprising as the algorithm was quite robust and the two additional observers were trained by the expert observer prior to the study to identify and segment tumors. The coefficient of variation reveals the increased sensitivity of both the volume (23.1% and 30.5%) and WHO (14.6% and 18.6%) metrics versus the linear RECIST (6.8% and 9.0%) and Ortho (9.1% and 10.6%) for the intraobserver and interobserver studies, respectively. This increased sensitivity using such two-dimensional and three-dimensional metrics further emphasizes the importance of understanding how each metric is obtained and the increasing sensitivity of the measures due to segmentation∕observer variability as the dimensionality increases. Overall, these results still indicate the strong reproducibility of using the image analysis technique described in this study to measure lung nodules.
In a previous paper describing longitudinal assessment of lung nodules in mice using micro-CT imaging,56 the smallest identifiable nodule documented was approximately 0.2 mm without the use of respiratory gating. In a more recent study by Hori et al.,13 the smallest nodule identified was 0.5 mm using synchronized respiratory gating and cast doubt over the minimum size detectable as documented in the previous study with no gating. Here, using the breath-hold gating technique, we were able to identify nodules down to 0.11 mm (RECIST) in diameter although the majority of initial nodules were detected at >0.20 mm, clearly providing a major advantage for assessment of lung nodules at a very early stage.
A fundamental advantage of using the IIBH technique for high-resolution lung imaging is the ability to identify structures of interest at an earlier stage. Here we have performed assessment of the urethane induced lung cancer model at significantly earlier stages to previous studies. Garbow et al.4 began assessment of Urethane induced tumor mice at 8 months posturethane administration, Hori et al.13 began at 10 weeks posturethane administration, while here we present identification and tracking of nodules at 8 weeks posturethane administration. Also, both prior studies identified less than three times the average number of lesions as found in previous histologic examinations of the same model.57 While here in this study, we found similar tumor incidence rates of 20–40 per mouse (30.44±1.93 SEM, n=5).
Radiation dose is an important factor when planning and executing micro-CT imaging and more so when undertaking longitudinal assessment of disease progression. Each scan performed in the study presented here administered approximately 0.85 Gy of radiation to the mouse as calculated using the Siemens dosage calculator. Mole had previously reported that mice can neutralize daily whole-body doses between 0.25 and 0.5 Gy.31, 58 However, those experiments were performed on CBA mice while A∕J mice were used in this study. In addition, the criterion for neutralization was that of survival rather than implying complete restoration to the status quo ante.59 A study by Safwat et al.60 measured the radiation-induced lung damage threshold for mice between 12 and 15.8 Gy with a greater threshold for lower rates of dosage administration (0.71 versus 0.08 Gy∕min). Sequential radiation therapy has also been reported in the order of five 10 Gy daily doses.61 Carlson et al. recently presented a study on radiation dose due to micro-CT imaging on subcutaneous and simulated internal tumors. They concluded that the dosage levels were likely below the threshold to affect tumor growth using their ultrafast protocol; however, further investigation into other imaging protocols such as those presented here need to be undertaken. Based on these findings along with our low dosage rate of 0.018 Gy∕min, we do not expect adverse or therapeutic effects on the mice or lung tumors in this longitudinal study; however, considering the differences in sensitivity between mice strains and lung cancer models, further studies should be undertaken in order to characterize potential dosage effects.
In conclusion, micro-CT imaging provides a unique platform for in vivo longitudinal assessment of lung cancer progression at very high resolutions. In addition, the ability to evaluate the same subject over time provides for a sensitive assay that can be carried out on a smaller sample size. When integrated with image processing and analysis routines as detailed here, the data acquired from micro-CT imaging can now provide a very powerful assessment of pulmonary disease progression.
ACKNOWLEDGMENTS
The authors would like to thank Osama Saba of Siemens Medical for providing the prototype oncology application, OncoCare. In addition, the authors would like to thank Jered Sieren for his CT support and Bridget Zimmerman for her expertise in the statistical analysis. This project was supported in part via NIH Grant No. HL-ROl-HL080285.
References
- Olsson L. E. et al. , “Measurement of MR signal and T2∗ in lung to characterize a tight skin mouse model of emphysema using single-point imaging,” J. Magn. Reson Imaging 25(3), 488–494 (2007). 10.1002/jmri.20840 [DOI] [PubMed] [Google Scholar]
- Driehuys B. and Hedlund L. W., “Imaging techniques for small animal models of pulmonary disease: MR microscopy,” Toxicol. Pathol. 35(1), 49–58 (2007). 10.1080/01926230601132048 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Driehuys B. et al. , “3He MRI in mouse models of asthma,” Magn. Reson. Med. 58(5), 893–900 (2007). 10.1002/mrm.21306 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Garbow J. R. et al. , “Quantitative monitoring of adenocarcinoma development in rodents by magnetic resonance imaging,” Clin. Cancer Res. 14(5), 1363–1367 (2008). 10.1158/1078-0432.CCR-07-1757 [DOI] [PubMed] [Google Scholar]
- Ford N. L. et al. , “Quantifying lung morphology with respiratory-gated micro-CT in a murine model of emphysema,” Phys. Med. Biol. 54(7), 2121–2130 (2009). 10.1088/0031-9155/54/7/018 [DOI] [PubMed] [Google Scholar]
- Froese A. R. et al. , “Three-dimensional computed tomography imaging in an animal model of emphysema,” Eur. Respir. J. 30(6), 1082–1089 (2007). 10.1183/09031936.00000507 [DOI] [PubMed] [Google Scholar]
- Kawakami M. et al. , “Sequential and quantitative analysis of a murine model of elastase-induced emphysema,” Biol. Pharm. Bull. 31(7), 1434–1438 (2008). 10.1248/bpb.31.1434 [DOI] [PubMed] [Google Scholar]
- Postnov A. A. et al. , “In vivo assessment of emphysema in mice by high resolution x-ray microtomography,” J. Microsc. 220, 70–75 (2005). 10.1111/j.1365-2818.2005.01510.x [DOI] [PubMed] [Google Scholar]
- Shofer S. et al. , “A micro-computed tomography-based method for the measurement of pulmonary compliance in healthy and bleomycin-exposedmice,” Exp. Lung Res. 33(3–4), 169–183 (2007). 10.1080/01902140701364458 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cavanaugh D. et al. , “In vivo respiratory-gated micro-CT imaging in small-animal oncology models,” Mol. Imaging 3(1), 55–62 (2004). 10.1162/153535004773861723 [DOI] [PubMed] [Google Scholar]
- Cody D. D. et al. , “Murine lung tumor measurement using respiratory-gated micro-computed tomography,” Invest. Radiol. 40(5), 263–269 (2005). 10.1097/01.rli.0000160070.67270.05 [DOI] [PubMed] [Google Scholar]
- Fushiki H. et al. , “Quantification of mouse pulmonary cancer models by microcomputed tomography imaging,” Cancer Sci. 100(8), 1544–1549 (2009). 10.1111/j.1349-7006.2009.01199.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hori Y. et al. , “Periodic analysis of urethane-induced pulmonary tumors in living A/J mice by respiration-gated x-ray microcomputed tomography,” Cancer Sci. 99(9), 1774–1777 (2008). 10.1111/j.1349-7006.2008.00834.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- Paulus M. J. et al. , “High resolution x-ray computed tomography: An emerging tool for small animal cancer research,” Neoplasia 2(1-2), 62–70 (2000). 10.1038/sj.neo.7900069 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Savai R. et al. , “Evaluation of angiogenesis using micro-computed tomography in a xenograft mouse model of lung cancer,” Neoplasia 11(1), 48–56 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ambrosini V. et al. , “Assessment of a chemically induced model of lung squamous cell carcinoma in mice by 18F-FDG small-animal PET,” Nucl. Med. Commun. 28(8), 647–652 (2007). 10.1097/MNM.0b013e32823f9ffa [DOI] [PubMed] [Google Scholar]
- Memon A. A. et al. , “Positron emission tomography (PET) imaging with [llC]-labeled erlotinib: A micro-PET study on mice with lung tumor xenografts,” Cancer Res. 69(3), 873–878 (2009). 10.1158/0008-5472.CAN-08-3118 [DOI] [PubMed] [Google Scholar]
- Haller J. et al. , “Visualization of pulmonary inflammation using noninvasive fluorescence molecular imaging,” J. Appl. Physiol. 104(3), 795–802 (2008). 10.1152/japplphysiol.00959.2007 [DOI] [PubMed] [Google Scholar]
- Cavanaugh D. et al. , “Quantification of bleomycin-induced murine lung damage in vivo with micro-computed tomography,” Acad. Radiol. 13(12), 1505–1512 (2006). 10.1016/j.acra.2006.08.011 [DOI] [PubMed] [Google Scholar]
- Badea C. T. et al. , “In vivo small-animal imaging using micro-CT and digital subtraction angiography,” Phys. Med. Biol. 53(19), R319–R350 (2008). 10.1088/0031-9155/53/19/R01 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ford N. L. et al. , “In vivo characterization of lung morphology and function in anesthetized free-breathing mice using micro-computed tomography,” J. Appl. Physiol. 102(5), 2046–2055 (2007). 10.1152/japplphysiol.00629.2006 [DOI] [PubMed] [Google Scholar]
- Badea C., Hedlund L. W., and Johnson G. A., “Micro-CT with respiratory and cardiac gating,” Med. Phys. 31(12), 3324–3329 (2004). 10.1118/1.1812604 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ford N. L. et al. , “Prospective respiratory-gated micro-CT of free breathing rodents,” Med. Phys. 32(9), 2888–2898 (2005). 10.1118/1.2013007 [DOI] [PubMed] [Google Scholar]
- Johnson E. M. et al. , “A new method for respiratory gating during microcomputed tomography of lung in mice,” J. Am. Assoc. Lab. Anim. Sci. 47(4), 46–56 (2008). [PMC free article] [PubMed] [Google Scholar]
- Sera T. et al. , “Development of high-resolution 4D in vivo-CT for visualization of cardiac and respiratory deformations of small animals,” Phys. Med. Biol. 53(16), 4285–4301 (2008). 10.1088/0031-9155/53/16/005 [DOI] [PubMed] [Google Scholar]
- Hu J. et al. , “Dynamic small animal lung imaging via a postacquisition respiratory gating technique using micro-cone beam computed tomography,” Acad. Radiol. 11(9), 961–970 (2004). 10.1016/j.acra.2004.05.019 [DOI] [PubMed] [Google Scholar]
- Drangova M. et al. , “Fast retrospectively gated quantitative four-dimensional (4D) cardiac micro computed tomography imaging of free-breathing mice,” Invest. Radiol. 42(2), 85–94 (2007). 10.1097/01.rli.0000251572.56139.a3 [DOI] [PubMed] [Google Scholar]
- Ford N. L., Wheatley A. R., and Holdsworth D. W., “Optimization of a retrospective technique for respiratory-gated high speed micro-CT of free-breathing rodents,” Phys. Med. Biol. 51, 5749–5769 (2007). 10.1088/0031-9155/52/19/002 [DOI] [PubMed] [Google Scholar]
- Bartling S. H. et al. , “Intrinsic respiratory gating in small-animal CT,” Eur. Radiol. 18(7), 1375–1384 (2008). 10.1007/s00330-008-0903-3 [DOI] [PubMed] [Google Scholar]
- Ertel D. et al. , “Respiratory phase-correlated micro-CT imaging of free-breathing rodents,” Phys. Med. Biol. 54(12), 3837–3846 (2009). 10.1088/0031-9155/54/12/015 [DOI] [PubMed] [Google Scholar]
- Namati E. et al. , “In vivo micro-CT lung imaging via a computer-controlled intermittent iso-pressure breath hold (IIBH) technique,” Phys. Med. Biol. 51(23), 6061–6075 (2006). 10.1088/0031-9155/51/23/008 [DOI] [PubMed] [Google Scholar]
- Artaechevarria X. et al. , “Airway segmentation and analysis for the study of mouse models of lung disease using micro-CT,” Phys. Med. Biol. 54(22), 7009–7024 (2009). 10.1088/0031-9155/54/22/017 [DOI] [PubMed] [Google Scholar]
- Field K. J. and Lang C. M., “Hazards of urethane (ethyl carbamate): A review of the literature,” Lab Anim. 22(3), 255–262 (1988). 10.1258/002367788780746331 [DOI] [PubMed] [Google Scholar]
- Dahl G. A., Miller E. C., and Miller J. A., “Comparative carcinogenicities and mutagenicities of vinyl carbamate, ethyl carbamate, and ethyl N-hydroxycarbamate,” Cancer Res. 40(4), 1194–1203 (1980). [PubMed] [Google Scholar]
- Dahl G. A., Miller J. A., and Miller E. C., “Vinyl carbamate as a promutagen and a more carcinogenic analog of ethyl carbamate,” Cancer Res. 38(11), 3793–3804 (1978). [PubMed] [Google Scholar]
- Fernando R. C. et al. , “Detection of l,N6-ethenodeoxyadenosine and 3,N4-ethenodeoxycytidine by immunoaffinity/32P-postlabelling in liver and lung DNA of mice treated with ethyl carbamate (urethane) or its metabolites,” Carcinogenesis 17(8), 1711–1718 (1996). 10.1093/carcin/17.8.1711 [DOI] [PubMed] [Google Scholar]
- Guengerich F. P. and Kim D. H., “Enzymatic oxidation of ethyl carbamate to vinyl carbamate and its role as an intermediate in the formation ofl,N6-ethenoadenosine,” Chem. Res. Toxicol. 4(4), 413–421 (1991). 10.1021/tx00022a003 [DOI] [PubMed] [Google Scholar]
- Rivera B. et al. , “A novel method for endotracheal intubation of mice and rats used in imaging studies,” Contemp. Top. Lab. Anim. Sci. 44(2), 52–55 (2005). [PubMed] [Google Scholar]
- Macdonald K. D., Chang H. Y., and Mitzner W., “An improved simple method of mouse lung intubation,” J. Appl. Physiol. 106(3), 984–987 (2009). 10.1152/japplphysiol.91376.2008 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rivers M., “Tutorial introduction to x-ray computed microtomography data processing,” 1998. Available from http://www.mcs.anl.gov/research/projects/X-rav-cmt/rivers/tutorial.html.
- Boin M. and Haibel A., “Compensation of ring artefacts in synchrotron tomographic images,” Opt. Express 14(25), 12071–12075 (2006). 10.1364/OE.14.012071 [DOI] [PubMed] [Google Scholar]
- Sijbers J. and Postnov A., “Reduction of ring artefacts in high resolution micro-CT reconstructions,” Phys. Med. Biol. 49(14), N247–N253 (2004). 10.1088/0031-9155/49/14/N06 [DOI] [PubMed] [Google Scholar]
- Prell D., Kyriakou Y., and Kalender W. A., “Comparison of ring artifact correction methods for flat-detector CT,” Phys. Med. Biol. 54(12), 3881–3895 (2009). 10.1088/0031-9155/54/12/018 [DOI] [PubMed] [Google Scholar]
- Kyriakou Y., Prell D., and Kalender W. A., “Ring artifact correction for high-resolution micro CT,” Phys. Med. Biol. 54(17), N385–N391 (2009). 10.1088/0031-9155/54/17/N02 [DOI] [PubMed] [Google Scholar]
- Herman G. T., “Correction for beam hardening in computed tomography,” Phys. Med. Biol. 24(1), 81–106 (1979). 10.1088/0031-9155/24/1/008 [DOI] [PubMed] [Google Scholar]
- Sieren J. C. et al. , “A process model for direct correlation between computed tomography and histopathology: Application in lung cancer,” Acad. Radiol. 17, 169–180 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Grady L., “Random walks for image segmentation,” IEEE Trans. Pattern Anal. Mach. Intell. 28(11), 1768–1783 (2006). 10.1109/TPAMI.2006.233 [DOI] [PubMed] [Google Scholar]
- Grady L.et al. , “Random walks for interactive organ segmentation in two and three dimensions: Implementation and validation,” in Proc. Internation Conference on Medical Image Computing and Computer-Assisted Intervention (MICCAI 2005), Vol. 8, pp. 773–780 (2005). [DOI] [PubMed]
- Zuurbier C. J., Emons V. M., and Ince C., “Hemodynamics of anesthetized ventilated mouse models: Aspects of anesthetics, fluid support, and strain,” Am. J. Physiol. Heart Circ. Physiol. 282(6), H2099–H2105 (2002). [DOI] [PubMed] [Google Scholar]
- Leal T. et al. , “Successful protocol of anaesthesia for measuring transepithelial nasal potential difference in spontaneously breathing mice,” Lab Anim. 40(1), 43–52 (2006). 10.1258/002367706775404480 [DOI] [PubMed] [Google Scholar]
- Cruz J. I., Loste J. M., and Burzaco O. H., “Observations on the use of medetomidine/ketamine and its reversal with atipamezole for chemical restraint in the mouse,” Lab Anim. 32(1), 18–22 (1998). 10.1258/002367798780559383 [DOI] [PubMed] [Google Scholar]
- Daly M. J. et al. , “Geometric calibration of a mobile C-arm for intraoperative cone-beam CT,” Med. Phys. 35(5), 2124–2136 (2008). 10.1118/1.2907563 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cho Y. et al. , “Accurate technique for complete geometric calibration of cone-beam computed tomography systems,” Med. Phys. 32(4), 968–983 (2005). 10.1118/1.1869652 [DOI] [PubMed] [Google Scholar]
- Holland E. C., Mouse Models of Human Cancer (Wiley-Liss, Hoboken, NJ, 2004). 10.1002/0471675067 [DOI] [Google Scholar]
- Malkinson A. M., “Primary lung tumors in mice as an aid for understanding, preventing, and treating human adenocarcinoma of the lung,” Lung Cancer 32(3), 265–79 (2001). 10.1016/S0169-5002(00)00232-4 [DOI] [PubMed] [Google Scholar]
- De Clerck N. M. et al. , “High-resolution x-ray microtomography for the detection of lung tumors in living mice,” Neoplasia 6(4), 374–379 (2004). 10.1593/neo.03481 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Malkinson A. M., “Primary lung tumors in mice: An experimentally manipulate model of human adenocarcinoma,” Cancer Res. 52(9), 2670s–2676s (1992). [PubMed] [Google Scholar]
- Mole R. H., “Quantitative observations on recovery from whole body irradiation in mice. II. Recovery during and after daily irradiation,” Br. J. Radiol. 30(349), 40–46 (1957). 10.1259/0007-1285-30-349-40 [DOI] [PubMed] [Google Scholar]
- Mole R. H., “Quantitative observations on recovery from whole body irradiation in mice. I. Recovery after single large doses of radiation,” Br. J. Radiol. 29(346), 563–569 (1956). 10.1259/0007-1285-29-346-563 [DOI] [PubMed] [Google Scholar]
- Safwat A. et al. , “Effect of radiation dose rate and cyclophosphamide on pulmonary toxicity after total body irradiation in a mouse model,” Int. J. Radiat. Oncol., Biol., Phys. 34(1), 85–91 (1996). 10.1016/0360-3016(95)02078-0 [DOI] [PubMed] [Google Scholar]
- Camphausen K. et al. , “Radiation therapy to a primary tumor accelerates metastatic growth in mice,” Cancer Res. 61(5), 2207–2211 (2001). [PubMed] [Google Scholar]