Abstract
Blood pressure waveform analysis may permit continuous (i.e., automated) and less invasive (i.e., safer and simpler) central hemodynamic monitoring in the intensive care unit and other clinical settings without requiring any instrumentation beyond what is already in use or available. This practical approach has been a topic of intense investigation for decades and may garner even more interest henceforth due to the evolving demographics as well as recent trends in clinical hemodynamic monitoring. Here, we review techniques that have appeared in the literature for mathematically estimating clinically significant central hemodynamic variables, such as cardiac output, from different blood pressure waveforms. We begin by providing the rationale for pursuing such techniques. We then summarize earlier techniques and thereafter overview recent techniques by our collaborators and us in greater depth while pinpointing both their strengths and weaknesses. We conclude with suggestions for future research directions in the field and a description of some potential clinical applications of the techniques.
Keywords: arterial tree, cardiac output, intensive care unit, modeling, signal processing
the proportion of the elderly population is growing as the clinical staff per capita declines (36). By 2020, a ∼30% shortfall in nurses is projected (36). These demographic changes accentuate the importance of effective, continuous (i.e., automated), and less invasive (i.e., safer and simpler) patient monitoring systems. Such systems are especially needed for the hemodynamic monitoring of cardiovascular disease.
Today, hemodynamic monitoring often entails the continuous measurement of blood pressure (BP) waveforms. Indeed, each year, minimally invasive catheters are used in at least 10 million intensive care unit (ICU) and surgery patients to measure the peripheral arterial BP (ABP) waveform from, typically, the radial artery (59a). Invasive catheters are also used, although to a lesser degree, in these and other patients to monitor BP waveforms from the right heart and pulmonary artery (66). Moreover, the Food and Drug Administration (FDA)-approved systems for noninvasive measurements of the peripheral ABP waveform are now commercially available (14a, 35a), whereas implantable devices for chronic measurements of the right ventricular pressure (RVP) waveform or other BP waveforms may follow (19a, 63, 80). These newer sensors could potentially be used in the ICU in addition to other settings. However, while these systems are continuous and afford a level of invasiveness safe enough for regular monitoring, they are ineffective in the sense that the measured BP levels are imprecise indicators of circulatory status.
One reason is that BP levels in the right heart and peripheral arteries are not as relevant as their counterparts in the left heart and central aorta. That is, the left heart is more prone to disease than the right heart, whereas central ABP truly indicates cardiac afterload and perfusion and is more predictive of patient outcome than peripheral ABP (see Fig. 1A) (95, 113).
Fig. 1.
Central hemodynamic variables are clinically significant. A: central arterial blood pressure (ABP) is a superior predictor of coronary artery disease (CAD) severity than peripheral ABP. *P < 0.05; **P < 0.01; ***P < 0.001. B: cardiac output (CO) provides an earlier indicator of a bleed than ABP. TPR, total peripheral resistance; HR, heart rate; BP, blood pressure; RAP, right atrial pressure. C: CO, left atrial pressure (LAP), and left ventricular ejection fraction (EF) permit diagnosis. D: CO helps direct therapy so as to improve postcardiac surgery outcomes. [Adapted from Refs. 13, 71, and 113.]
A second reason is that BP levels do not provide an early indicator of a hemodynamic event. For example, in the initial stages of a bleed, the cardiovascular control system maintains mean ABP at the cost of cardiac output (CO; see Fig. 1B) (13, 40). While frank hypotension may eventually occur (see Fig. 1B), it is often too late to intervene at this point (13, 40). Thus, it is CO monitoring that would give an early enough indication of the need for therapy.
A third reason is that BP levels alone do not permit diagnosis. For example, when the cardiovascular control system eventually fails so as to result in hypotension, the root cause could be sepsis, systolic or diastolic dysfunction, or hypovolemia (see Fig. 1C). To reveal the root cause, CO, left atrial pressure (LAP) [or other left ventricular (LV) filling pressure (LVFP) measures], and LV ejection fraction (EF) must also be monitored (see Fig. 1C) (55, 84). Note that EF has also proven to be a powerful predictor of outcome in heart failure patients (27).
A fourth reason is that BP levels are not sufficient for directing therapy during obvious hypotension of known etiology. Other measurements that more adequately reflect global oxygen delivery, especially CO, are also needed to implement the recently celebrated “goal-directed therapy” and thereby improve patient outcomes (see Fig. 1D) (71, 83).
Several methods are available for measuring CO, LAP, EF, and central ABP (see Table 1). The most commonly used methods for monitoring CO and LAP in clinical practice both require the insertion of a pulmonary artery catheter (66). In particular, CO has been estimated with the bolus thermodilution method, and LAP has been approximated with the pulmonary capillary wedge pressure (PCWP) method. However, these two methods require an operator to inject a fluid bolus through a catheter side port and inflate a catheter-tip balloon and are thus not continuous. Moreover, the PCWP method is not easy to implement, as exemplified by the frequent occurrence of partial wedging and balloon overinflation (59, 74). Similarly, proper implementation of the thermodilution method is hindered by variations in injectate volume, rate, and temperature, which introduce 10–15% error in the measurement (26, 48, 72, 103). Perhaps due in part to these limitations along with the moderate level of invasiveness of the pulmonary artery catheter, this longstanding ICU device has yet to reveal a clinical benefit (see, e.g., Refs. 43 and 97), and its use has declined (118). The standard clinical method for monitoring EF [i.e., the ratio of stroke volume (SV) to end-diastolic volume (EDV)] is by imaging the ventricular volume via, e.g., echocardiography or contrast angiography (93). However, noninvasive imaging methods require an expert operator and capital equipment. Alternative methods for measuring CO, LAP, and EF suffer from even more severe disadvantages that have limited or precluded their use in clinical practice. Finally, the conventional method for measuring central ABP is to place a catheter in the aorta. However, this method is rarely performed for regular monitoring due to its high level of invasiveness, carrying risks of thromboembolism (e.g., stroke). Thus, while these methods are effective in terms of monitoring vital central hemodynamic variables, they are not both continuous and less invasive, and, as a result, their application is limited today and likely to be even more so hereafter.
Table 1.
Conventional methods for measuring CO, LAP, left ventricular EF, and central ABP generally require an operator and/or a high level of invasiveness
| Measurement Method | Advantages | Disadvantages |
|---|---|---|
| Measurement of CO | ||
| Aortic flow probe (33) | Continuous (i.e., automated), accurate | Thoracotomy |
| Bolus thermodilution (33) | Relatively simple, inexpensive | Pulmonary artery catheterization, operator |
| Continuous thermodilution (33) | Continuous right ventricular EF (119, 129) | Pulmonary artery catheterization, less accurate than bolus thermodilution (130) |
| Oxygen fick (33) | Accurate, inexpensive | Two catheterizations, operator |
| Thoracic bioimpedance (33, 60) | Continuous, noninvasive | Inaccurate in the presence of lung fluids (25) |
| Ultrasound [e.g., Doppler (33, 60)] | Noninvasive | Expert operator, expensive |
| Measurement of LAP | ||
| Left atrial catheter | Continuous, accurate | Too invasive and risky |
| Pulmonary capillary wedge pressure (89, 110) | Relatively simple, inexpensive | Pulmonary artery catheterization, operator |
| Measurement of EF | ||
| Conductance catheter (12) | Continuous | Too invasive and risky, repeated calibrations (15) |
| Imaging [e.g., echocardiography (93)] | Can be noninvasive | Expert operator, expensive |
| Nonimaging nuclear monitor (31, 128) | Continuous, noninvasive | Too difficult to position |
| Sonomicrometry (94) | Continuous, can be accurate | Thoracotomy |
| Measurement of central ABP | ||
| Aortic catheter | Continuous, accurate | Too invasive and risky |
CO, cardiac output; LAP, left atrial pressure; EF, ejection fraction; ABP, arterial blood pressure.
One potential solution for overcoming the aforementioned limitations of conventional hemodynamic monitoring systems is to estimate essential central hemodynamic variables by the mathematical analysis of BP waveforms (see Fig. 2). The main advantage of this solution is that hemodynamic monitoring systems may become simultaneously effective, continuous, and less invasive without requiring any instrumentation beyond what is already in use or available.
Fig. 2.
Continuous (i.e., automated) and less invasive monitoring of central hemodynamic variables by the mathematical analysis of peripheral ABP, pulmonary artery pressure (PAP), and right ventricular pressure (RVP) waveforms. [Reproduced in part from Refs. 14a, 47a, 72a, and 72b.]
In this article, we review BP waveform analysis techniques for central hemodynamic monitoring that have appeared in the literature. We first summarize earlier techniques and then highlight recent techniques by our collaborators and us. We conclude by providing a summary, suggesting future research avenues, and describing possible applications in the ICU and beyond.
Summary of Earlier BP Waveform Analysis Techniques
Central hemodynamic monitoring by BP waveform analysis has attracted the attention of many for decades. The earlier techniques span the estimation of several cardinal central hemodynamic variables from different BP waveforms, with particular emphasis on tracking CO through ABP waveforms.
CO monitoring by BP waveform analysis.
ABP WAVEFORM ANALYSIS.
Over a century ago, Frank (37) first suggested that CO could be measured from ABP waveforms. Shortly thereafter, Erlanger et al. (34) introduced the first specific analysis technique. They observed that pulse pressure [PP; i.e., systolic pressure (SP) minus diastolic pressure] in the aorta was positively correlated to SV. Thus, their technique simply involved detecting PP from the central ABP waveform to determine SV for each beat and CO (via PP × heart rate) to within constant scale factors. In this way, the technique could be used to monitor the relative changes in these variables or their absolute values after one calibration. (Note that tracking changes is most relevant in acute settings such as the ICU.) However, this PP technique assumes that cardiac ejection occurs instantly when it actually occurs over finite time in which a portion of the ejected blood passes through the aorta.
Among the numerous ensuing studies (e.g., Refs. 61, 67, 90, 102, 112, 114, and 115), Bourgeois et al. (16, 17) performed one of the most compelling. These investigators developed a central ABP waveform analysis technique that made no cardiac ejection time assumptions. They represented the arterial tree with a Windkessel model accounting for the lumped compliance of the large arteries (AC) and the total peripheral resistance (TPR) of the small arteries (see Fig. 3A). According to this model, central ABP should decay like a pure exponential during diastole with a time constant (τ) equal to TPR × AC. By assuming constant AC, their technique involved fitting an exponential to each diastolic interval of the waveform to measure τ and then dividing mean ABP by τ to compute CO to within a 1/AC scale factor (see Fig. 3B). Their technique also yielded proportional SV for each beat via the dynamic model equation (see Fig. 3B). This Windkessel technique proved to be highly accurate (see Fig. 3C), because the diastolic intervals of the central ABP waveform indeed resemble pure exponential decays (see Fig. 3B).
Fig. 3.
Earlier technique for monitoring CO by analysis of the central ABP waveform. A: Windkessel model of the arterial tree accounting for TPR and the nearly constant arterial compliance (AC). τ, Time constant. B: estimation of relative changes in CO and beat-to-beat stroke volume (SV) via the model. pa(t), Central ABP waveform; tbe, beginning of the ejection interval; tee, end of the ejection interval. C: exemplary canine results. [Adapted from Refs. 17.]
CHALLENGE OF WAVE REFLECTIONS.
However, as discussed above, central ABP is seldom measured for monitoring. Moreover, unlike the central ABP waveform, the analysis of a readily available peripheral ABP waveform is complicated by wave reflection phenomena (85). That is, when the heart ejects blood, pressure and flow waves are initiated, which travel through the arterial tree. Whenever the waves reach a site of impedance mismatch, especially the high-resistance arteries (68, 77, 117), they are in part reflected back toward the heart. The pressure (or flow) at a given arterial site is thus equal to the sum (or difference) of the forward- and backward-traveling waves at that site (117). As a result, the ABP waveform becomes progressively distorted with increasing distance from the heart (see Fig. 4) (68). Most notably, PP and SP become more and more amplified, whereas exponential diastolic decays become less apparent. Furthermore, the magnitude and timing of wave reflections are not unvarying. For example, reflected waves, and thus PP and SP amplification, are augmented by vasoconstriction and blunted by vasodilation (117). Thus, a peripheral ABP waveform may not be well represented with the Windkessel model or other lumped parameter models, as these models ignore confounding wave reflections.
Fig. 4.
The ABP waveform becomes progressively distorted from the central aorta to the peripheral arteries due to wave reflections in the arterial tree. [Adapted from Ref. 68.]
Consequently, the Windkessel and PP techniques do not produce the same results when applied to a peripheral ABP waveform. Even so, three PP- or Windkessel-based techniques are now commercially available for continuous and minimally invasive CO monitoring (30). These techniques have shown overall accuracies good enough for the FDA. But, the techniques have not proven reliable during hemodynamic instability (23, 49, 98), when monitoring is most needed, and can even predict increases in CO during progressive hypovolemia (see Fig. 5B) (14, 82). This wrong prediction occurs because the hypovolemia-induced vasoconstriction augments the wave reflections so as to buffer the decrease in peripheral PP via SV.
Fig. 5.
Recent technique for monitoring CO by analysis of a peripheral ABP waveform. A: estimation of relative CO change by circumventing the confounding wave reflections via long-time interval analysis. Here, the cardiac contraction signal [x(t)] and the peripheral ABP waveform segment [Pp(t)] are actually 1–6 min in duration. PPj and Rj are the pulse pressure and the onset time of ABP upstroke of the jth beat. h(t), impulse response. B: group-averaged human lower body negative pressure (LBNP) results (means ± 2SE). The CO estimates correctly decreased with LBNP, whereas classic (PP × HR) estimates of CO did not. [Adapted from Refs. 76 and unpublished observations of A. T. Reisner, D. Xu, K. L. Ryan, V. A. Convertino, C. A. Rickards, and R. Mukkamala.]
ANALYSIS OF OTHER BP WAVEFORMS.
A few teams have applied techniques designed for ABP waveforms to the pulmonary artery pressure (PAP) waveform to try to improve CO estimation reliability (but at the cost of increased invasiveness) (22, 32, 111, 126, 127). Bennett and coworkers (50) have more recently tailored one of these techniques to the RVP waveform to track CO with their implantable RVP monitor for heart failure patients (63, 80). However, these PAP and RVP waveform analysis techniques may not yield greater accuracy (22, 111), as the waveforms are similarly confounded by wave reflections in addition to inertial effects in the low-resistance pulmonary circulation [see, e.g., complex PAP diastolic decays in Fig. 2 (111)].
LESSONS LEARNED.
Much has been gained from the earlier studies on CO monitoring by BP waveform analysis (often called “pulse contour analysis”). For example, two important insights can be drawn from the studies of Bourgeois et al. (16, 17).
First, these studies demonstrate that, from the perspective of the aorta, the arterial tree can be regarded simply as a single reservoir (or Windkessel) rather than spatially dispersed, infinitesimal reservoirs of different pressures. Indeed, the long and varying distances between the aorta and the main reflection sites at the high-resistance arteries result in forward and backward waves in the aorta with large phasic differences (77). So, waves with relatively short wavelengths, which correspond to high-frequency ABP changes, destructively interfere so as to lessen their buildup in the aorta. On the other hand, waves with longer wavelengths, which correspond to low-frequency ABP changes, constructively add and are felt by the aorta. The key point is that, as the wavelengths of the propagating waves increase, the pressures at the various arterial sites converge to the same level such that the arterial tree behaves more like a single reservoir (77). [Note that this point also implies that low-frequency ABP variability can be well represented with the Windkessel model regardless of the measurement site (73, 77, 87, 88).]
Second, the studies indicate that AC is approximately constant. Indeed, AC is mainly due to the aorta, which is relatively sparse in smooth muscle (19, 69). So, changes in vasomotor tone should only have a small effect on AC. Furthermore, AC is not very sensitive to ABP changes, as wave travel delay time, which is determined by AC, varies only modestly with large ABP changes (86). Thus, the well-known inverse relationship between AC and ABP (41) may only become a major factor during extreme ABP perturbations. However, AC does significantly decrease with aging (41). Consequently, and more precisely, AC is approximately constant over a wide ABP range and on the time scale of months to years. Nevertheless, correcting CO estimates for AC change would be useful, particularly at the limits of the ABP range. Wesseling and colleagues have developed an accurate curve relating AC to ABP (57) and have applied it for ABP-dependent calibration of their Windkessel-based CO-monitoring technique (115).
LAP monitoring by BP waveform analysis.
Far fewer techniques have been introduced for estimating LAP from BP waveforms. Over half a century ago, Cournand (24) helped establish the classic technique of monitoring LAP through end-diastolic PAP (42) that is sometimes used in clinical practice (66). Bennett and coworkers (21, 79, 92) recently showed that end-diastolic PAP may be well estimated from the value of RVP at the time of its maximal derivative and may therefore be tracked with their implantable RVP monitor. However, end-diastolic PAP is not as accurate as PCWP (47) and is contraindicated for following LAP during pulmonary hypertension (46, 66). Within the past two decades, researchers have proposed techniques to predict LAP from a BP waveform via formulas derived from a training data set of LVFP measures and BP waveforms. In particular, McIntyre and colleagues (70, 100) led the development of a technique to monitor LV end-diastolic pressure from the peripheral ABP response to a Valsalva maneuver. This technique is now available on the market with FDA clearance (27a). Marik and others (65, 125) formulated a technique to monitor PCWP from peripheral SP variations induced by mechanical ventilation. Most recently, deBoisblanc et al. (29) conceived a technique to predict PCWP from the PAP waveform specifically using a neural network. The former two techniques, which extract LVFP information from the induced intrathoracic pressure variations, provide an attractive minimally invasive or noninvasive means to sensitively monitor LAP but do require a hemodynamic perturbation. On the other hand, the latter, more invasive technique is continuous and may be more specific, as PAP is directly determined by LAP, whereas ABP is not. For all of these techniques, accuracy is reliant on the availability of comprehensive training data sets, which may be very difficult to obtain.
EF monitoring based on BP waveform analysis.
Most of the earlier BP waveform analysis techniques do not explicitly model the ventricle and therefore cannot estimate EF. However, Guarini et al. (39) and Xiao et al. (121) have proposed techniques based on ABP waveform analysis that account for the LV and could be extended to monitor EF. These techniques specifically estimate the ventricular parameters needed for EF computation. However, the techniques also require a CO measurement for the estimation.
Central ABP monitoring by peripheral ABP waveform analysis.
Starting with the work of O'Rourke and colleagues (53) over a decade and a half ago, there has been considerable interest in techniques for estimating the central ABP waveform from a peripheral ABP waveform. Nearly all of the techniques that have been introduced involve developing an average transfer function from simultaneous peripheral ABP and central ABP measurements in a group of subjects and then applying this transfer function to a peripheral ABP waveform measured from a new subject to estimate the central ABP waveform (e.g., Refs. 20, 35, 51, and 52). One such “generalized transfer function” technique is commercially available with FDA clearance (11a). While the accuracy of this technique has not been proven in a diverse patient population, it has been shown to determine some clinical outcomes better than the raw peripheral ABP waveform (120). Even so, the transfer function relating peripheral ABP to central ABP is not invariant, as wave reflection phenomena are known to change over time due to, e.g., neurohumoral modulation of TPR, and vary among subjects due to, e.g., age-related AC differences (see above). With this recognition, Sugimachi et al. (105) and Westerhof et al. (116) have recently proposed a technique to adapt the transfer function by defining it through an arterial tube model with a personalized value for a model parameter reflecting the wave travel delay time. However, the technique uses population averages for the remaining parameters.
Overview of Our Recent BP Waveform Analysis Techniques
The earlier studies have laid the foundation for further research in the field. In a series of recent investigations, we, along with our collaborators, have attempted to build on this seminal body of work by accounting for additional facets of physiology.
CO monitoring by long-time interval analysis of a peripheral ABP waveform.
IDEA.
We (76) developed a technique for continuous and minimally invasive or noninvasive monitoring of CO by analysis of a peripheral ABP waveform. The idea is to circumvent the confounding arterial wave reflections by exploiting the fact that they diminish with decreasing frequency or increasing time scale (see above). Thus, the Windkessel model (see Fig. 3A) becomes a more valid representation of the arterial tree with longer time scales (73, 77, 87, 88). So, for example, if the heart suddenly stopped beating, then peripheral ABP would eventually decay like a pure exponential once the faster wave reflections vanished.
The technique therefore analyzes a peripheral ABP waveform over long time intervals to determine the pure exponential decay that would eventually result if the heart abruptly ceased to beat. More specifically, the ABP response to one heart beat is estimated from a 1- to 6-min segment of the waveform [see h(t) in Fig. 5A]. Then, τ is determined by fitting an exponential to the tail end of this single-contraction ABP response once the faster wave reflections have dissipated (see Fig. 5A). Finally, proportional CO is computed (see Fig. 5A).
STEPS.
The single-contraction ABP response is determined in two steps (see Fig. 5A). First, a cardiac contraction signal [x(t)] is constructed by forming an impulse train in which each impulse is located at the onset of upstroke of an ABP pulse and is scaled by the ensuing PP. An impulse response [h(t)] is then estimated, which, when convolved with x(t), optimally fits the ABP waveform [pp(t)] segment. By definition, h(t) represents the (scaled) ABP response to one heart beat. In principle, reliable determination of τ is then achieved by virtue of h(t) accurately coupling the beat-to-beat variations in x(t) to pp(t).
EVALUATION.
We evaluated this long-time interval analysis technique against independent CO measurements from swine and humans over a wide range of physiological conditions (44, 62, 75, 76). The overall relative CO error was ∼15% after a single calibration in each subject. For comparison, the Windkessel technique (see Fig. 3B) yielded a corresponding error that was, on average, ∼50% higher. We also assessed the technique as applied to a noninvasive peripheral ABP waveform from humans using lower body negative pressure (LBNP), a popular hemorrhage model (A. T. Reisner, D. Xu, K. L. Ryan, V. A. Convertino, C. A. Rickards, and R. Mukkamala; unpublished observations). We specifically tested the technique in terms of its ability to predict reductions in CO that occur with LBNP. For comparison, we likewise evaluated the PP technique, with thoracic bioimpedance (see Table 1) as a reference. Consistent with early bleeding, mean ABP was maintained, whereas the reference CO decreased, with increasing LBNP (see Fig. 5B). The CO estimates of our technique correctly decreased with escalating LBNP and then returned to baseline after recovery from LBNP (see Fig. 5B). In contrast, for the reasons mentioned above, the CO estimates of the classic technique erroneously increased during LBNP (see Fig. 5B).
LIMITATIONS.
Unlike earlier techniques, this technique cannot detect rapid CO changes. The technique also does not account for any AC changes in its present form and, as with the Windkessel technique, would monitor total rather than forward flow during aortic regurgitation.
CO and LAP monitoring by long-time interval analysis of the PAP or RVP waveform.
IDEA.
We extended the preceding technique to PAP and RVP waveforms so as to monitor LAP in addition to CO and thereby potentially realize continuous monitoring with the pulmonary artery catheter and chronic monitoring with the aforesaid implantable device (122, 123). The ideas are to include a constant term in the analysis to account for the major contribution of LAP to PAP in the low-resistance pulmonary circulation and to invoke the fact that RVP is normally close to PAP during the ejection intervals.
STEPS.
The extended PAP waveform analysis technique is performed in four steps (see Fig. 6A). First, a cardiac contractions signal [x(t)] is constructed from a PAP waveform segment [ppa(t)] as described above. An impulse response [h(t)] and an additive constant term are then estimated so as to optimally couple x(t) to ppa(t). The constant term represents PAP attained in the absence of cardiac contractions and is thus indicative of average LAP, whereas h(t) represents the PAP-LAP response to a single heart beat. Next, the Windkessel time constant of the pulmonary arterial tree (τ = PAR × PAC, where PAR is pulmonary arterial resistance and PAC is pulmonary artery compliance) is determined by fitting an exponential to the tail end of h(t) once the faster wave reflections and inertial effects vanish. Finally, since PAC may also be approximately constant over a wide PAP range (56, 101), proportional CO is computed by dividing mean PAP minus LAP with τ and readily calibrated with a single thermodilution measurement. (This latter measurement can be made more accurate by averaging a number of carefully repeated measurements, which may be feasible for a one-time calibration in the ICU.)
Fig. 6.
Recent extended techniques for monitoring CO and LAP by analysis of the PAP or RVP waveform. A: estimation of LAP in addition to the relative CO change from the PAP waveform by including a constant term in the long-time interval analysis. ppa(t), the PAP waveform segment. B: estimation of the relative CO change and LAP from the RVP waveform by further invoking the fact that RVP is nearly equal to PAP during ejection intervals. pRV(t), the RVP waveform segment; z(t), complete PAP waveform segment; tMPDj, the time of the maximal derivative of the jth RVP beat. C: exemplary canine results. [Adapted from Refs. 122 and 123.]
The extended RVP waveform analysis technique is executed as follows (see Fig. 6B). First, each ejection interval of an RVP waveform segment [prv(t)] is detected from the time of the maximal derivative (see above) to shortly after the peak value. The technique then proceeds similarly to the PAP waveform analysis technique with the formation of a cardiac contraction signal [x(t)] and the subsequent estimation of average LAP (constant term) and the single-contraction PAP-LAP response [impulse response h(t)] by optimally coupling x(t) to only the detected ejection intervals of prv(t) (i.e., an incomplete PAP waveform segment specifically lacking diastolic intervals). Next, after τ is determined from h(t), the complete PAP waveform segment [z(t)] is constructed by adding average LAP to the convolution between x(t) and h(t). Finally, proportional CO is determined by dividing mean z(t) minus LAP with τ.
EVALUATION.
We evaluated these extended long-time interval analysis techniques against gold standard aortic flow probe CO and LAP catheter measurements from dogs during drug and volume interventions (122, 123). The overall relative calibrated CO error and absolute LAP error from the PAP waveforms were 15.2% and 1.7 mmHg, whereas the corresponding errors from the RVP waveforms were 16.0% and 2.0 mmHg (excluding one outlier; see Fig. 6C). For comparison, classic end-diastolic PAP estimates resulted in LAP errors that were almost three times larger. These initial results suggest that continuous monitoring may be achieved with right ventricular catheterization rather than the more risky pulmonary artery catheterization.
LIMITATIONS.
These techniques carry weaknesses similar to the initial long-time interval analysis technique. The extended techniques also do not account for intrathoracic pressure changes that are directly transmitted to PAP and the vascular waterfall effect (13, 40). Furthermore, the RVP waveform analysis technique is contraindicated during pulmonic valve disease.
EF monitoring by lumped parameter model-based analysis of the central ABP waveform.
IDEA.
We (106) developed a technique for the continuous monitoring of beat-to-beat EF (along with CO and beat-to-beat SV) by analysis of only the central ABP waveform. The idea is to extend the Windkessel technique (17) by modeling the LV so as to split a proportional SV estimate into likewise scaled end-systolic volume and EDV components. Absolute EF is then computed by cancellation of the common scale factor.
STEPS.
The technique is performed in three steps. First, the entire central ABP waveform is represented with a lumped parameter model of the LV and arterial tree (see Fig. 7A). In particular, the LV is characterized by the well-known variable elastance (reciprocal of compliance) model in which the ventricular elastance (VE) oscillates over time [E(t)] so as to drive blood flow and the ventricular unstressed volume (VUV) is nonzero (96, 104), the aortic valve is modeled with an ideal diode, and the arterial tree is embodied by the Windkessel model.
Fig. 7.
Recent technique for monitoring EF by analysis of only the central ABP waveform. A: lumped parameter model of the left ventricle and arterial tree comprising the time-varying ventricular elastance (VE) model and Windkessel model. B: estimation of the model parameters so as to compute beat-to-beat EF and relative changes in CO and beat-to-beat SV via Eqs. 1 and 2. Emax, maximum elastance; E(t), VE over time; Ts, the time duration to reach Emax from minimum elastance. C: exemplary canine results. [Adapted from Ref. 106.]
The lumped model parameters are then estimated to within AC scale factors from each beat of the central ABP waveform [pa(t)] via the following discretized equation governing model behavior during the ejection interval:
| (1) |
where T is the sampling period and tbe and tee are the beginning and end times of the ejection interval. Both T and pa(t) for tbe ≤ t ≤ tee are known, whereas the proportional model parameters, τ = TPR × AC and AC × E(t) for tbe ≤ t ≤ tee, are unknown. Thus, Eq. 1 represents a set of N equations with N + 2 unknowns (where, e.g., N ≈ 80 for T = 4 ms). To arrive at a set of equations with a unique solution, a parametric raised cosine function is used to succinctly characterize the variable VE as justified by previous studies (45, 99) [see E(t) in Fig. 7B]. Substituting this parametric function into Eq. 1 reduces the number of unknowns to τ, AC × maximum elastance (Emax), AC × E(tbe), and Ts where Ts is the time duration to reach Emax from minimum elastance (see Fig. 7B). Initially, τ is estimated by fitting an exponential to the diastolic interval of pa(t) (see Fig. 7B). This τ is then substituted into Eq. 1, and the remaining three parameters are estimated from pa(t) over the ejection interval by optimally matching both sides of this equation.
Finally, AC × E(t) for tbe ≤ t ≤ tee and pa(t) are used to compute beat-to-beat absolute EF (by cancellation of the AC scale factor) as well as beat-to-beat proportional SV (and thus proportional CO) via the following variable VE model equation:
| (2) |
Note that VUV/AC here is not estimated. Normally, however, this parameter is relatively small (28, 54) and may thus be neglected or set to a nominal value. Alternatively, especially for stiff, dilated hearts, VUV/AC may be periodically determined from imaging measurements of EF.
EVALUATION.
We evaluated this lumped parameter model-based analysis technique against standard transthoracic echocardiography EF measurements from dogs during drug and volume interventions (106). The overall uncalibrated absolute EF error was 5.6%. Furthermore, unlike the reference measurements, these estimates offer continuous monitoring (see Fig. 7C).
LIMITATIONS.
This technique ignores the arterial characteristic impedance, which may be important for fitting the ejection interval (77). The technique as described is also contraindicated during aortic valve disease. More significantly, the central ABP waveform is rarely available.
Central ABP monitoring by analysis of multiple peripheral ABP waveforms via adaptive multichannel blind system identification.
IDEA.
We (107, 108) developed a technique for minimally invasive or noninvasive monitoring of the central ABP waveform by analysis of multiple peripheral ABP waveforms. The idea is to derive central ABP in a fully adaptive manner by finding the commonality in two or more peripheral ABP waveforms via multichannel blind system identification (1).
STEPS.
The technique is executed as follows. First, m (>1) peripheral ABP waveforms [ppi(t), 1 ≤ i ≤ m] are modeled as outputs of m unknown systems driven by the common unknown central ABP waveform [pa(t)] input (see Fig. 8A). Each of the systems relating the common input to the distinct outputs characterizes wave reflection phenomena in different arterial paths. These systems are represented with finite impulse responses [hi(t), 1 ≤ i ≤ m] that are different from each other (or, more precisely, coprime) (124). In this way, all of the similarity in the outputs may be attributed to only the input. Then, hi(t) (1 ≤ i ≤ m) values are estimated from ppi(t) (1 ≤ i ≤ m) by optimally matching the cross relations [e.g., pp1(t) ⊗ h2(t) = pp2(t) ⊗ h1(t) for m = 2, where ⊗ is convolution] (124). Note that the scale factor of hi(t) (1 ≤ i ≤ m) is determined by invoking the fact that mean ABP is similar throughout the arterial tree due to Poiseuille's law (66, 77, 81). Finally, pa(t) is determined by optimally deconvolving the estimated hi(t) (1 ≤ i ≤ m) from ppi(t) (1 ≤ i ≤ m) (1).
Fig. 8.
Recent technique for monitoring central ABP by analysis of multiple peripheral ABP waveforms. A: fully adaptive estimation of central ABP by identifying the commonality in two or more peripheral ABP waveforms via multichannel blind system identification. m, Number of arterial paths; hi(t), impulse response of the system corresponding to the ith arterial path. B: exemplary swine results obtained from two peripheral ABP waveforms. [Adapted from Ref. 108.]
EVALUATION.
We (108) evaluated this adaptive multichannel blind system identification technique as applied to two peripheral ABP waveforms from the radial and femoral arteries against central ABP catheter measurements from swine during several interventions. The overall absolute central ABP sample-to-sample and PP errors were 3.5 and 5.6 mmHg (see Fig. 8B). For comparison, the corresponding errors between the raw peripheral ABP waveforms and time-aligned central ABP waveforms were, on average, 8.6 and 14.3 mmHg (see Fig. 8B).
ADDITIONAL IDEA.
The preceding lumped parameter model-based analysis technique may then be applied to the derived central ABP waveform, which is less complicated by wave reflections than peripheral ABP waveforms (see Fig. 9A). The idea is to attenuate (rather than circumvent) the confounding wave reflections so as to estimate beat-to-beat EF and proportional SV in addition to proportional CO.
Fig. 9.
Recent technique for monitoring EF along with CO and SV by analysis of peripheral ABP waveforms. A: estimation of beat-to-beat EF and relative changes in CO and beat-to-beat SV by first attenuating the confounding wave reflections via fully adaptive derivation of the central ABP waveform and then applying the lumped parameter model-based analysis to the derived waveform. B: exemplary swine results obtained from the central ABP waveform derived via adaptive multichannel blind system identification. [Adapted from Ref. 108.]
EVALUATION.
We (108) evaluated the preceding technique as applied to the aforesaid derived central ABP waveforms against gold standard aortic flow probe SV measurements. The overall relative calibrated beat-to-beat SV error was 12.9% (see Fig. 9B).
LIMITATIONS.
The main concern for this technique is that any wave reflection dynamics shared by the different arterial paths will be erroneously attributed to the central ABP waveform. In addition, only one peripheral ABP waveform is commonly measured in clinical practice.
Central ABP monitoring by analysis of one peripheral ABP waveform via an adaptive transfer function.
IDEA.
We (109) developed another technique for minimally invasive or noninvasive monitoring of the central ABP waveform, but this time by analysis of one peripheral ABP waveform. The idea is to derive central ABP in a fully adaptive manner by defining the transfer function relating peripheral ABP to central ABP through a distributed model of arterial pressure and flow to account for the confounding wave reflections (105, 116) and then estimating all model parameters by exploiting the preknowledge of aortic flow.
STEPS.
The technique is implemented as follows. First, the arterial tree is modeled as m parallel, uniform tubes terminated by lumped parameter loads (see Fig. 10A). The ith tube represents the path between the aorta and the ith peripheral artery. Consistent with Poiseuille's law, each tube is frictionless and thus has constant characteristic impedance , where li and ci are the large artery inertance and compliance] and allows waves to travel with constant time delay . The ith terminal load represents the arterial bed distal to the ith peripheral artery. Each terminal load has a frequency-dependent impedance [Zi(ω)] characterized by two parameters that are dependent on the peripheral resistance and compliance (Ai and Bi) and Zci (105, 116). Waves traveling along each tube are reflected at the terminal load in a frequency-dependent manner so as to mimic the progressive distortion that actual arterial pressure and flow waveforms undergo with increasing distance from the heart (see Fig. 4).
Fig. 10.
Recent technique for monitoring central ABP by analysis of one peripheral ABP waveform. A: distributed model of the arterial tree accounting for confounding wave reflections. qai(t), Central arterial flow; Tdi, wave travel delay time; Zi, impedance; ω, angular frequency; Ai, and Bi, unknown parameters characterizing the ith terminal load. B: fully adaptive estimation of central ABP by defining the transfer function relating peripheral ABP to central ABP through the model and estimating its parameters by assuming negligible aortic flow during diastole. C: exemplary canine results of the adaptive transfer function and conventional generalized transfer function. [Adapted from Ref. 109.]
According to this model, a peripheral ABP waveform [ppi(t)] is related to the central ABP waveform [pa(t)] through a transfer function in terms of Tdi, Ai, and Bi (see the pressure → pressure transfer function in Fig. 10B). These three unknown parameters are estimated as follows.
First, the wave travel delay time Tdi is first measured noninvasively via any of the known methods (see, e.g., Refs. 105 and 116). Only one such measurement is made for a subject over a period of months to years, as Tdi may not greatly vary over this time period (see above).
All three parameters are then estimated from a 15- to 60-s segment of ppi(t) and the initial Tdi value by exploiting the fact that aortic flow is negligible during diastole. Thus, the flow at each tube entrance in the model (central arterial flow) may likewise be small during this time interval (as indicated in Fig. 10A). In particular, according to this model, ppi(t) is related to the central arterial flow waveform [qai(t)] through a transfer function also in terms of Tdi, Ai, and Bi (see the pressure → flow transfer function in Fig. 10B). These common parameters are estimated by finding the pressure → flow transfer function that, when applied to ppi(t), optimally fits qai(t) (scaled by Zci) to zero over its diastolic intervals [as approximated via heart rate (64); see Fig. 10B].
Finally, the estimated parameters are substituted into the pressure → pressure transfer function. This transfer function is then applied to ppi(t) so as to derive pa(t).
EVALUATION.
We (109) evaluated this adaptive transfer function technique against central ABP catheter measurements from dogs during various interventions. The overall absolute central ABP sample-to-sample and PP errors were 3.7 and 3.4 mmHg, whereas the parallel errors of the time-aligned peripheral ABP waveforms were 8.6 and 20.3 mmHg (see Fig. 10C). For comparison, the analogous errors of totally or partially generalized transfer function techniques (35, 105, 116) were, on average, 63% higher than the technique (see Fig. 10C), even though they had the advantage of being developed with a subset of the reference central ABP waveforms.
LIMITATIONS.
This technique disregards wave reflections caused by arterial tapering, stiffening, and branching. The technique is also contraindicated during aortic regurgitation.
Conclusions
Summary.
BP waveform analysis provides a practical approach for achieving the much-needed continuous and less invasive monitoring of vital central hemodynamic variables. As a result, investigation of this approach has been longstanding. Various techniques for estimating CO, LAP, or central ABP from more readily available ABP, PAP, or RVP waveforms were developed earlier. Some of these techniques are now commercially available with FDA clearance. However, the techniques generally neglect key facets of physiology, which has left room for further research. These efforts allowed us to recently conceive ideas to 1) overcome the challenge of wave reflections by circumventing their confounding effects (via long-time interval analysis) or attenuating these effects upfront, 2) estimate LAP without training data by identifying PAP attained in the absence of cardiac contractions, 3) derive EF by modeling the LV, and 4) estimate central ABP in a fully adaptive manner by exploiting the commonality in multiple peripheral ABP waveforms or preknowledge of aortic flow.
Future directions.
Although much progress has been made in the field, future investigations are warranted. First, there will always be a need for the development of more accurate and simpler techniques. Such development efforts could, for example, help overcome the limitations of our work revealed herein. In this regard, we believe that the development of an effective and convenient AC correction factor for CO estimation is particularly worthwhile. We also advocate subsequent focus on realizing minimally invasive or noninvasive monitoring of specific markers of LVFP and extending the analysis to even more readily available pulse oximeter waveforms. Second, compelling verification of both existing and new techniques is a must for widespread clinical adoption. Such validation efforts should show that the techniques are generally accurate enough, which means that they should yield overall errors that are within the 10–15% relative and absolute errors of thermodilution and transthoracic echocardiography measurements (11, 18, 26, 38, 48, 103), the 1- to 2-mmHg error of valid PCWP measurements (47, 58), and the 5 ± 8-mmHg FDA limits for ABP measurement (78) in a broad patient population. The validation should also indicate that the techniques do not have critical failure modes such as hemodynamic instability. Furthermore, demonstrations that the techniques, when coupled with defined therapeutic protocols, can improve patient outcomes would be of tremendous value.
Potential applications.
Established BP waveform analysis techniques have a number of potential clinical applications. First and foremost, the techniques could be readily used in the ICU and other settings where invasive catheters are broadly in use. The continuous and/or less invasive central hemodynamic monitoring capabilities of the techniques would offer several advantages over the methods that have been used in these settings. These advantages include 1) reducing patient risk, 2) avoiding the technical problems associated with the implementation of PCWP and thermodilution methods, 3) saving precious time in the busy clinical environment, 4) obtaining an early indication of a hemodynamic event, 5) guiding therapy more effectively, and 6) permitting remote monitoring. In addition, the techniques could be used, for example, with noninvasive systems for 1) early detection of circulatory shock in patients entering the emergency room and during transport of combat and civilian trauma casualties, 2) identification of patients who become unstable during convalescence, 3) risk stratification of outpatients, and 4) home healthcare; the techniques could also be used in implantable devices for chronic heart failure management.
GRANTS
This work was supported in part by awards from Michigan State University, the American Heart Association, National Heart, Lung, and Blood Institute Grant HL-080568, and National Science Foundation CAREER Grant 0643477.
DISCLOSURES
No conflicts of interest, financial or otherwise, are declared by the author(s).
ACKNOWLEDGMENTS
The authors thank all of the collaborators referenced herein.
REFERENCES
- 1.Abed-Meraim K, Qiu W, Hua Y. Blind system identification. Proc IEEE 85: 1310–1332, 1997 [Google Scholar]
- 11.Apfel HD, Shen Z, Gopal AS, Vangi V, Solowiejczyk D, Altmann K, Barst RJ, Boxt LM, Allan LD, King DL. Quantitative three dimensional echocardiography in patients with pulmonary hypertension and compressed left ventricles: comparison with cross sectional echocardiography and magnetic resonance imaging. Heart 76: 350–354, 1996 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11a.AtCor Medical. AtCor Medical Homepage (online). http://www.atcormedical.com[12 July 2010]
- 12.Baan J, van der Velde ET, de Bruin HG, Smeenk GJ, Koops J, van Dijk AD, Temmerman D, Senden J, Buis B. Continuous measurement of left ventricular volume in animals and humans by conductance catheter. Circulation 70: 812–823, 1984 [DOI] [PubMed] [Google Scholar]
- 13.Barcroft H, Edholm OG, McMichael J, Sharpey-Schafer EP. Posthaemorrhagic fainting: study by cardiac output and forearm flow. Lancet 1: 489–491, 1944 [Google Scholar]
- 14.Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, Scholz J, Doerges V. The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg 105: 107–113, 2007 [DOI] [PubMed] [Google Scholar]
- 14a.BMEYE. BMEYE Homepage (online). http://www.bmeye.com[12 July 2010]
- 15.Boltwood CMJ, Appleyard RF, Glantz SA. Left ventricular volume measurement by conductance catheter in intact dogs. Parallel conductance volume depends on left ventricular size. Circulation 80: 1360–1377, 1989 [DOI] [PubMed] [Google Scholar]
- 16.Bourgeois MJ, Gilbert BK, Donald DE, Wood EH. Characteristics of aortic diastolic pressure decay with application to the continuous monitoring of changes in peripheral vascular resistance. Circ Res 35: 56–66, 1974 [DOI] [PubMed] [Google Scholar]
- 17.Bourgeois MJ, Gilbert BK, Von Bernuth G, Wood EH. Continuous determination of beat to beat stroke volume from aortic pressure pulses in the dog. Circ Res 39: 15–24, 1976 [DOI] [PubMed] [Google Scholar]
- 18.Buck T, Hunold P, Wentz KU, Tkalec W, Nesser HJ, Erbel R. Tomographic three-dimensional echocardiographic determination of chamber size and systolic function in patients with left ventricular aneurysm: comparison to magnetic resonance imaging, cineventriculography, and two-dimensional echocardiography. Circulation 96: 4286–4297, 1997 [DOI] [PubMed] [Google Scholar]
- 19.Burton AC. Relation of structure to function of the tissues of the wall of blood vessels. Physiol Rev 34: 619–642, 1954 [DOI] [PubMed] [Google Scholar]
- 19a.CardioMEMS. CardioMEMS Homepage (online). http://www.cardiomems.com[12 July 2010]
- 20.Chen CH, Nevo E, Fetics B, Pak PH, Yin FC, Maughan WL, Kass DA. Estimation of central aortic pressure waveform by mathematical transformation of radial tonometry pressure. Validation of generalized transfer function. Circulation 95: 1827–1836, 1997 [DOI] [PubMed] [Google Scholar]
- 21.Chuang PP, Wilson RF, Homans DC, Stone K, Bergman T, Bennett TD, Kubo SH. Measurement of pulmonary artery diastolic pressure from a right ventricular pressure transducer in patients with heart failure. J Card Fail 2: 41–46, 1996 [DOI] [PubMed] [Google Scholar]
- 22.Cibulski AA, Lehan PH, Hellems HK. Pressure methods for estimating right and left ventricular stroke volumes. Am J Physiol 225: 1460–1466, 1973 [DOI] [PubMed] [Google Scholar]
- 23.Cooper ES, Muir WW. Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs. Crit Care Med 35: 1724–1729, 2007 [DOI] [PubMed] [Google Scholar]
- 24.Cournand A. Some aspects of the pulmonary circulation in normal man and in chronic cardiopulmonary diseases. Circulation 2: 641–657, 1950 [DOI] [PubMed] [Google Scholar]
- 25.Critchley LA. Impedance cardiography. The impact of new technology. Anaesthesia 53: 677–684, 1998 [DOI] [PubMed] [Google Scholar]
- 26.Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15: 85–91, 1999 [DOI] [PubMed] [Google Scholar]
- 27.Curtis JP, Sokol SI, Wang Y, Rathore SS, Ko DT, Jadbabaie F, Portnay EL, Marshalko SJ, Radford MJ, Krumholz HM. The association of left ventricular ejection fraction, mortality, and cause of death in stable outpatients with heart failure. J Am Coll Cardiol 42: 736–742, 2003 [DOI] [PubMed] [Google Scholar]
- 27a.CVP Diagnostics. Vericor Monitor (online). http://cvpdiagnostics.com [12 July 2010]
- 28.Davis TL. Teaching Physiology Through Interactive Simulation of Hemodynamics Cambridge, MA: Massachusetts Institute of Technology, 1991 [Google Scholar]
- 29.deBoisblanc BP, Pellett A, Johnson R, Champagne M, McClarty E, Dhillon G, Levitzky M. Estimation of pulmonary artery occlusion pressure by an artificial neural network. Crit Care Med 30: 261–266, 2003 [DOI] [PubMed] [Google Scholar]
- 30.Della Rocca G, Cecconi M, Costa MG. Mini invasive hemodynamic monitoring: from arterial pressure to cardiac output. Signa Vitae 3: S7–S9, 2008 [Google Scholar]
- 31.Dellegrottaglie S, Filardi PP, Pace L, Ponticelli MP, Corrado L, Cafiero M, Polimeno M, Camerino R, Chiariello M. Non-imaging nuclear monitoring of left ventricular function: twenty-five years of technical development and clinical experience. Ital Heart J 3: 300–307, 2002 [PubMed] [Google Scholar]
- 32.DeLoskey AF, Nichols WW, Conti CR, Pepine CJ. Estimation of beat-to-beat stroke volume from the pulmonary arterial pressure contour in man. Med Biol Eng Comput 16: 707–714, 1978 [DOI] [PubMed] [Google Scholar]
- 33.Ehlers KC, Mylrea KC, Waterson CK, Calkins JM. Cardiac output measurements. A review of current techniques and research. Ann Biomed Eng 14: 219–239, 1986 [DOI] [PubMed] [Google Scholar]
- 34.Erlanger J, Hooker DR. An experimental study of blood-pressure and of pulse-pressure in man. Bull Johns Hopkins Hosp 12: 145–378, 1904 [Google Scholar]
- 35.Fetics B, Nevo E, Chen CH, Kass DA. Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry. IEEE Trans Biomed Eng 46: 698–706, 1999 [DOI] [PubMed] [Google Scholar]
- 35a.Finapres Medical Systems. Finapress Medical Systems Homepage (online). http://www.finapres.com/site/[12 July 2010]
- 36.Fleming KC, Evans JM, Chutka DS. Caregiver and clinician shortages in an aging nation. Mayo Clin Proc 78: 1026–1040, 2003 [DOI] [PubMed] [Google Scholar]
- 37.Frank O. Die Grundform des arteriellen Pulses. Erste Abhandlung Mathematische Analyse. Z Biol 37: 483–526, 1899 [Google Scholar]
- 38.Gopal AS, Shen Z, Sapin PM, Keller AM, Schnellbaecher MJ, Leibowitz DW, Akinboboye OO, Rodney RA, Blood DK, King DL. Assessment of cardiac function by three-dimensional echocardiography compared with conventional noninvasive methods. Circulation 92: 842–853, 1995 [DOI] [PubMed] [Google Scholar]
- 39.Guarini M, Urzúa J, Cipriano A, González W. Estimation of cardiac function from computer analysis of the arterial pressure waveform. IEEE Trans Biomed Eng 45: 1420–1428, 1998 [DOI] [PubMed] [Google Scholar]
- 40.Guyton AC, Hall JE. Textbook of Medical Physiology. Philadelphia, PA: Saunders, 1996 [Google Scholar]
- 41.Hallock P, Benson IC. Studies on the elastic properties of human isolated aorta. J Clin Invest 16: 595–602, 1937 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Harvey RM, Ferrer MI. Some effects of digoxin upon the heart and circulation in man; digoxin in left ventricular failure. Am J Med 7: 439–453, 1949 [DOI] [PubMed] [Google Scholar]
- 43.Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K, PAC-Man Study Collaboration Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 366: 472–477, 2005 [DOI] [PubMed] [Google Scholar]
- 44.Heldt T. Continuous blood pressure-derived cardiac output–should we be thinking long term? J Appl Physiol 101: 373–374, 2006 [DOI] [PubMed] [Google Scholar]
- 45.Heldt T, Shim EB, Kamm RD, Mark RG. Computational modeling of cardiovascular response to orthostatic stress. J Appl Physiol 92: 1239–1254, 2002 [DOI] [PubMed] [Google Scholar]
- 46.Herbert WH. Limitations of pulmonary artery end-diastolic pressure as the reflection left ventricular end-diastolic pressure. NY State J Med 72: 229–232, 1972 [PubMed] [Google Scholar]
- 47.Humphrey CB, Virgilio RW, Folkerth TL, Fosburg RG. An analysis of direct and indirect measurements of left atrial pressure. J Thorac Cardiovasc Surg 71: 643–647, 1976 [PubMed] [Google Scholar]
- 47a.ICU-USA. Arterial Line (online). http://www.icu-usa.com/tour/equipment/aline.htm[12 July 2010]
- 48.Jansen JR, Schreuder JJ, Mulier JP, Smith NT, Settels JJ, Wesseling KH. A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients. Br J Anaesth 87: 212–222, 2001 [DOI] [PubMed] [Google Scholar]
- 49.Johansson A, Chew M. Reliability of continuous pulse contour cardiac output measurement during hemodynamic instability. J Clin Monit Comput 21: 237–242, 2007 [DOI] [PubMed] [Google Scholar]
- 50.Karamanoglu M, Bennett TD. A right ventricular pressure waveform based pulse contour cardiac output algorithm in canines. Cardiovasc Eng 6: 83–92, 2006 [DOI] [PubMed] [Google Scholar]
- 51.Karamanoglu M, Feneley MP. Derivation of the ascending aortic-carotid pressure transfer function with an arterial model. Am J Physiol Heart Circ Physiol 271: H2399–H2404, 1996 [DOI] [PubMed] [Google Scholar]
- 52.Karamanoglu M, Feneley MP. On-line synthesis of the human ascending aortic pressure pulse from the finger pulse. Hypertension 30: 1416–1424, 1997 [DOI] [PubMed] [Google Scholar]
- 53.Karamanoglu M, O'Rourke MF, Avolio AP, Kelly RP. An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man. Eur Heart J 14: 160–167, 1993 [DOI] [PubMed] [Google Scholar]
- 54.Kass DA, Midei M, Graves W, Brinker JA, Maughan WL. Use of a conductance (volume) catheter and transient vena caval occlusion for rapid determination of pressure-volume relationships in man. Cath Cardiovasc Diagn 15: 192–202, 1988 [DOI] [PubMed] [Google Scholar]
- 55.Katz AM. Physiology of the Heart. New York: Raven, 1992 [Google Scholar]
- 56.Kobayashi K, Washizu M, Kondo M, Matsukura Y, Motoyoshi S, Miyasaka K, Takata M. Longitudinal distribution of pulmonary vascular compliance in dogs. J Vet Med Sci 58: 41–46, 1996 [DOI] [PubMed] [Google Scholar]
- 57.Langewouters GJ, Wesseling KH, Goedhard WJ. The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new model. J Biomech 17: 425–435, 1984 [DOI] [PubMed] [Google Scholar]
- 58.Lappas D, Lell WA, Gabel JC, Civetta JM, Lowenstein E. Indirect measurement of left-atrial pressure in surgical patients. Pulmonary-capillary wedge and pulmonary-artery diastolic pressure compared with left-atrial pressure. Anesthesiology 38: 394–397, 1973 [DOI] [PubMed] [Google Scholar]
- 59.Leatherman JW, Shapiro RS. Overestimation of pulmonary artery occlusion pressure in pulmonary hypertension due to partial occlusion. Crit Care Med 31: 93–97, 2003 [DOI] [PubMed] [Google Scholar]
- 59a.LiDCO. Cardiovascular Monitoring (online). http://www.lidco.com/html/strategy/cardiovascular.asp[12 July 2010]
- 60.Linton DM, Gilon D. Advances in noninvasive cardiac output monitoring. Ann Card Anaesth 5: 141–148, 2002 [PubMed] [Google Scholar]
- 61.Linton NW, Linton RA. Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb. Br J Anaesth 86: 486–496, 2001 [DOI] [PubMed] [Google Scholar]
- 62.Lu Z, Mukkamala R. Continuous cardiac output monitoring in humans by invasive and noninvasive peripheral blood pressure waveform analysis. J Appl Physiol 101: 598–608, 2006 [DOI] [PubMed] [Google Scholar]
- 63.Magalski A, Adamson P, Gadler F, Boehm M, Steinhaus D, Reynolds D, Vlach K, Linde C, Cremers B, Sparks B, Bennett T. Continuous ambulatory right heart pressure measurements with an implantable hemodynamic monitor: a multicenter, 12-month follow-up study of patients with chronic heart failure. J Card Fail 8: 63–70, 2002 [DOI] [PubMed] [Google Scholar]
- 64.Malik M. If Dr. Bazett had had a computer… Pacing Clin Electrophysiol 19: 1635–1639, 1996 [DOI] [PubMed] [Google Scholar]
- 65.Marik PE. The systolic blood pressure variation as an indicator of pulmonary capillary wedge pressure in ventilated patients. Anaesth Intensive Care 21: 405–408, 1993 [DOI] [PubMed] [Google Scholar]
- 66.Marino PL, Sutin KM. The ICU Book. Baltimore, MD: Lippincott, Williams & Wilkins, 2006 [Google Scholar]
- 67.Martin JF, Volfson LB, Kirzon-Zolin VV, Schukin VG. Application of pattern recognition and image classification techniques to determine continuous cardiac output from the arterial pressure waveform. IEEE Trans Biomed Eng 41: 913–920, 1994 [DOI] [PubMed] [Google Scholar]
- 68.McDonald DA. Blood Flow in Arteries. London: Arnold, 1974 [Google Scholar]
- 69.McDonald DA. Regional pulse-wave velocity in the arterial tree. J Appl Physiol 24: 73–78, 1968 [DOI] [PubMed] [Google Scholar]
- 70.McIntyre KM, Vita JA, Lambrew CT, Freeman J, Loscalzo J. A noninvasive method of predicting pulmonary-capillary wedge pressure. N Engl J Med 327: 1715–1720, 1992 [DOI] [PubMed] [Google Scholar]
- 71.McKendry M, McGloin H, Saberi D, Caudwell L, Brady AR, Singer M. Randomised controlled trial assessing the impact of a nurse delivered, flow monitored protocol for optimisation of circulatory status after cardiac surgery. Br Med J 329: 258–261, 2004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Medin DL, Brown DT, Wesley R, Cunnion RE, Ognibene FP. Validation of continuous thermodilution cardiac output in critically ill patients with analysis of systematic errors. J Crit Care 13: 184–189, 1998 [DOI] [PubMed] [Google Scholar]
- 72a.Medline Plus. Swan Ganz Catheterization (online). http://www.nlm.nih.gov/medlineplus/ency/imagepages/18087.htm[12 July 2010]
- 72b.Medtronic. Medtronic Homepage (online). http://www.medtronic.com [12 July 2010] [Google Scholar]
- 73.Mohiuddin MW, Laine GA, Quick CM. Increase in pulse wavelength causes the systemic arterial tree to degenerate into a classical windkessel. Am J Physiol Heart Circ Physiol 293: H1164–H1171, 2007 [DOI] [PubMed] [Google Scholar]
- 74.Morris AH, Chapman RH, Gardner RM. Frequency of technical problems encountered in the measurement of pulmonary artery wedge pressure. Crit Care Med 12: 164–170, 1984 [DOI] [PubMed] [Google Scholar]
- 75.Mukkamala R, Reisner AT. Reply to van Lieshout and Jansen. J Appl Physiol 102: 827, 2007 [DOI] [PubMed] [Google Scholar]
- 76.Mukkamala R, Reisner AT, Hojman HM, Mark RG, Cohen RJ. Continuous cardiac output monitoring by peripheral blood pressure waveform analysis. IEEE Trans Biomed Eng 53: 459–467, 2006 [DOI] [PubMed] [Google Scholar]
- 77.Noordergraaf A. Circulatory System Dynamics New York: Academic, 1978 [Google Scholar]
- 78.O'Rourke MF, Pauca AL. Augmentation of the aortic and central arterial pressure waveform. Blood Press Monit 9: 179–185, 2004 [DOI] [PubMed] [Google Scholar]
- 79.Ohlsson A, Bennett T, Nordlander R, Ryden J, Astrom H, Ryden L. Monitoring of pulmonary arterial diastolic pressure through a right ventricular pressure transducer. J Card Fail 1: 161–168, 1995 [DOI] [PubMed] [Google Scholar]
- 80.Ohlsson A, Kubo SH, Steinhaus D, Connelly DT, Adler S, Bitkover C, Nordlander R, Ryden L, Bennett T. Continuous ambulatory monitoring of absolute right ventricular pressure and mixed venous oxygen saturation in patients with heart failure using an implantable haemodynamic monitor: results of a 1 year multicentre feasibility study. Eur Heart J 22: 942–954, 2001 [DOI] [PubMed] [Google Scholar]
- 81.Pauca AL, Wallenhaupt SL, Kon ND, Tucker WY. Does radial artery pressure accurately reflect aortic pressure? Chest 102: 1193–1198, 1992 [DOI] [PubMed] [Google Scholar]
- 82.Piehl MD, Manning JE, McCurdy SL, Rhue TS, Kocis KC, Cairns CB, Cairns BA. Pulse contour cardiac output analysis in a piglet model of sever hemorrhagic shock. Crit Care Med 36: 1189–1195, 2008 [DOI] [PubMed] [Google Scholar]
- 83.Pinsky MR. Hemodynamic evaluation and monitoring in the ICU. Chest 132: 2020–2029, 2007 [DOI] [PubMed] [Google Scholar]
- 84.Pinsky MR, Vincent J. Let us use the pulmonary artery catheter correctly and only when we need it. Crit Care Med 33: 1119–1122, 2005 [DOI] [PubMed] [Google Scholar]
- 85.Porje EG. Studies of the arterial pulse wave, particularly in the aorta. Acta Physiol Scand Suppl 13: 1–68, 1946 [Google Scholar]
- 86.Pruett JD, Bourland JD, Geddes LA. Measurement of pulse-wave velocity using a beat-sampling technique. Ann Biomed Eng 16: 341–347, 1988 [DOI] [PubMed] [Google Scholar]
- 87.Quick CM, Berger DS, Hettrick DA, Noordergraaf A. True arterial system compliance estimated from apparent arterial compliance. Ann Biomed Eng 28: 291–301, 2000 [DOI] [PubMed] [Google Scholar]
- 88.Quick CM, Berger DS, Noordergraaf A. Apparent arterial compliance. Am J Physiol Heart Circ Physiol 274: H1393–H1403, 1998 [DOI] [PubMed] [Google Scholar]
- 89.Rapaport E, Dexter L. Pulmonary “capillary” pressure. Methods Med Res 7: 85–93, 1958 [PubMed] [Google Scholar]
- 90.Redling JD, Akay M. Noninvasive cardiac output estimation: a preliminary study. Biol Cybern 77: 111–122, 1997 [DOI] [PubMed] [Google Scholar]
- 92.Reynolds D, Bartelt N, Taepke R, Bennett T. Measurement of pulmonary artery diastolic pressure from the right ventricle. J Am Coll Cardiol 25: 1176–1182, 1995 [DOI] [PubMed] [Google Scholar]
- 93.Rumberger JA, Behrenbeck T, Bell MR, Breen JF, Johnston DL, Holmes DR, Enriquez-Sarano M. Determination of ventricular ejection fraction: a comparison of available imaging methods. Mayo Clin Proc 72: 860–870, 1997 [DOI] [PubMed] [Google Scholar]
- 94.Rushmer RF, Franklin DL, Ellis RM. Left ventricular dimensions recorded by sonocardiometry. Circ Res 4: 684–688, 1956 [DOI] [PubMed] [Google Scholar]
- 95.Safar ME, Blacher J, Pannier B, Guerin AP, Marchais SJ, Guyonvarc'h P, London GM. Central pulse pressure and mortality in end-stage renal disease. Hypertension 39: 735–738, 2002 [DOI] [PubMed] [Google Scholar]
- 96.Sagawa K, Suga H, Shoukas AA, Bakalar KM. End-systolic pressure/volume ratio: a new index of ventricular contractility. Am J Cardiol 40: 748–753, 1977 [DOI] [PubMed] [Google Scholar]
- 97.Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 348: 5–14, 2003 [DOI] [PubMed] [Google Scholar]
- 98.Schuerholz T, Cobas Meyer M, Friedrich M, Przemeck M, Sumpelmann R, Marx G. Reliability of continuous cardiac output determination by pulse-contour analysis in porcine septic shock. Acta Anaesthesiol Scand 50: 407–413, 2006 [DOI] [PubMed] [Google Scholar]
- 99.Senzaki H, Chen CH, Kass DA. Single-beat estimation of end-systolic pressure-volume relation in humans. A new method with the potential for noninvasive application. Circulation 94: 2497–2506, 1996 [DOI] [PubMed] [Google Scholar]
- 100.Sharma GV, Woods PA, Lambrew CT, Berg CM, Pietro DA, Rocco TP, Welt FW, Sacchetti P, McIntyre KM. Evaluation of a noninvasive system for determining left ventricular filling pressure. Arch Intern Med 162: 2084–2088, 2002 [DOI] [PubMed] [Google Scholar]
- 101.Shoukas AA. Pressure-flow and pressure-volume relations in the entire pulmonary vascular bed of the dog determined by two-port analysis. Circ Res 37: 809–818, 1975 [DOI] [PubMed] [Google Scholar]
- 102.Starmer CF, McHale PA, Cobb FR, Greenfield JCJ. Evaluation of several methods for computing stroke volume from central aortic pressure. Circ Res 33: 139–148, 1973 [DOI] [PubMed] [Google Scholar]
- 103.Stetz CW, Miller RG, Kelly GE, Raffin TA. Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respirat Dis 126: 1001–1004, 1982 [DOI] [PubMed] [Google Scholar]
- 104.Suga H, Sagawa K. Instantaneous pressure-volume relationships and their ratio in the excised, supported canine left ventricle. Circ Res 35: 117–126, 1974 [DOI] [PubMed] [Google Scholar]
- 105.Sugimachi M, Shishido T, Miyatake K, Sunagawa K. A new model-based method of reconstructing central aortic pressure from peripheral arterial pressure. Jpn J Physiol 51: 217–222, 2001 [DOI] [PubMed] [Google Scholar]
- 106.Swamy G, Kuiper J, Gudur MSR, Olivier NB, Mukkamala R. Continuous left ventricular ejection fraction monitoring by aortic pressure waveform analysis. Ann Biomed Eng 37: 1055–1068, 2009 [DOI] [PubMed] [Google Scholar]
- 107.Swamy G, Ling Q, Li T, Mukkamala R. Blind identification of the aortic pressure waveform from multiple peripheral artery pressure waveforms. Am J Physiol Heart Circ Physiol 292: H2257–H2264, 2007 [DOI] [PubMed] [Google Scholar]
- 108.Swamy G, Mukkamala R. Estimation of the aortic pressure waveform and beat-to-beat relative cardiac output changes from multiple peripheral artery pressure waveforms. IEEE Trans Biomed Eng 55: 1521–1529, 2008 [DOI] [PubMed] [Google Scholar]
- 109.Swamy G, Xu D, Olivier NB, Mukkamala R. An adaptive transfer function for deriving the aortic pressure waveform from a peripheral artery pressure waveform. Am J Physiol Heart Circ Physiol 297: H1956–H1963, 2009 [DOI] [PubMed] [Google Scholar]
- 110.Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D. Catheterization of the heart in man with the use of a flow-directed balloon-tipped catheter. N Engl J Med 283: 447–451, 1970 [DOI] [PubMed] [Google Scholar]
- 111.Tajimi T, Sunagawa K, Yamada A, Nose Y, Takeshita A, Kikuchi Y, Nakamura M. Evaluation of pulse contour methods in calculating stroke volume from pulmonary artery pressure curve (comparison with aortic pressure curve). Eur Heart J 4: 502–511, 1983 [DOI] [PubMed] [Google Scholar]
- 112.Verdouw PD, Beaune J, Roelandt J, Hugenholtz PG. Stroke volume from central aortic pressure? A critical assessment of the various formulae as to their clinical value. Basic Res Cardiol 70: 377–389, 1975 [DOI] [PubMed] [Google Scholar]
- 113.Waddell TK, Dart AM, Medley TL, Cameron JD, Kingwell BA. Carotid pressure is a better predictor of coronary artery disease severity than brachial pressure. Hypertension 38: 927–931, 2001 [DOI] [PubMed] [Google Scholar]
- 114.Welkowitz W, Cui Q, Qi Y, Kostis JB. Noninvasive estimation of cardiac output. IEEE Trans Biomed Eng 38: 1100–1105, 1991 [DOI] [PubMed] [Google Scholar]
- 115.Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 74: 2566–2573, 1993 [DOI] [PubMed] [Google Scholar]
- 116.Westerhof BE, Guelen I, Stok WJ, Wesseling KH, Spaan JA, Westerhof N, Bos WJ, Stergiopulos N. Arterial pressure transfer characteristics: effects of travel time. Am J Physiol Heart Circ Physiol 292: H800–H807, 2007 [DOI] [PubMed] [Google Scholar]
- 117.Westerhof N, Sipkema P, Van Den Bos GC, Elzinga G. Forward and backward waves in the arterial system. Cardiovasc Res 6: 648–656, 1972 [DOI] [PubMed] [Google Scholar]
- 118.Wiener RS, Welch HG. Trends in the use of the pulmonary artery catheter in the United States, 1993–2004. JAMA 298: 423–429, 2007 [DOI] [PubMed] [Google Scholar]
- 119.Wiesenack C, Fiegl C, Keyser A, Laule S, Prasser C, Keyl C. Continuously assessed right ventricular end-diastolic volume as a marker of cardiac preload and fluid responsiveness in mechanically ventilated cardiac surgical patients. Crit Care 9: R226–R233, 2005 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.Williams B, Lacy PS, Thom SM, Cruickshank K, Stanton A, Collier D, Hughes AD, Thurston H, O'Rourke MF. Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes: principal results of the conduit artery function evaluation (CAFE) study. Circulation 113: 1213–1225, 2006 [DOI] [PubMed] [Google Scholar]
- 121.Xiao X, Ozawa ET, Huang Y, Kamm RD. Model-based assessment of cardiovascular health from noninvasive measurements. Ann Biomed Eng 30: 612–623, 2002 [DOI] [PubMed] [Google Scholar]
- 122.Xu D, Olivier NB, Mukkamala R. Cardiac output and left atrial pressure monitoring by right ventricular pressure waveform analysis for potential implantable device application. IEEE Trans Biomed Eng 56: 2335–2339, 2009 [DOI] [PubMed] [Google Scholar]
- 123.Xu D, Olivier NB, Mukkamala R. Continuous cardiac output and left atrial pressure monitoring by long time interval analysis of the pulmonary artery pressure waveform: proof of concept in dogs. J Appl Physiol 106: 651–661, 2009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 124.Xu G, Liu H, Tong L, Kailath T. A least-squares approach to blind channel identification. IEEE Trans Image Process 43: 2982–2993, 1995 [Google Scholar]
- 125.Xu H, Zhou S, Ma W, Yu B. Prediction of pulmonary arterial wedge pressure from arterial pressure or pulse oximetry plethysmographic waveform. Chin Med J 115: 1372–1375, 2002 [PubMed] [Google Scholar]
- 126.Zacharoulis AA, Evans TR, Ziady GM, Coltart DJ, Shillingford JP. Measurement of stroke volume from pulmonary artery pressure record in man. Br Heart J 37: 20–25, 1975 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 127.Zacharoulis AA, Mills CJ, Gabe IT, Shillingford JP. Estimation of stroke volume from the pulmonary artery pressure record. Cardiovasc Res 8: 506–516, 1974 [DOI] [PubMed] [Google Scholar]
- 128.Zaret BL, Wackers FJ. Cardiology–second of two parts. N Engl J Med 329: 855–863, 1993 [DOI] [PubMed] [Google Scholar]
- 129.Zink W, Nöll J, Rauch H, Bauer H, Desimone R, Martin E, Böttiger BW. Continuous assessment of right ventricular ejection fraction: new pulmonary artery catheter versus transoesophageal echocardiography. Anaesthesia 59: 1126–1132, 2004 [DOI] [PubMed] [Google Scholar]
- 130.Zollner C, Goetz AE, Weis M, Morstedt K, Pichler B, Lamm P, Kilger E, Haller M. Continuous cardiac output measurements do not agree with conventional bolus thermodilution cardiac output determination. Can J Anaesth 48: 1143–1147, 2001 [DOI] [PubMed] [Google Scholar]