Abstract
Photoacoustic tomography (PAT) is a fast-developing biomedical imaging technology suitable for in vivo imaging. PAT in spherical or circular geometry gives good image resolution yet is slow or expensive in signal acquisition and image formation. Reducing the number of detection angles can ameliorate such issues usually at the expense of image quality. This paper introduces a deconvolution-based algorithm that models the imaging process as a linear and shift-invariant system. As demonstrated by the in vivo experiment, this algorithm not only runs much faster than the back-projection algorithm, but also shows stronger robustness in that it provides better image quality when detection angles are sparse. Therefore, this algorithm promises to enable real-time PAT in circular geometry.
Keywords: photoacoustic tomography, deconvolution reconstruction, sparse detection angles, real-time imaging, circular geometry
1 Introduction
In the past decade, photoacoustic tomography (PAT) has emerged as a fast-developing biomedical imaging technology [1], [2]. Based on the photoacoustic (PA) effect [3], PAT utilizes a laser source to illuminate tissue and then detects the generated acoustic signals for imaging. PAT offers advantages in many aspects. First, PAT uses nonionizing illumination that poses no health hazard and is noninvasive. Second, as a hybrid of optical and ultrasound imaging techniques, PAT combines the merits of both. It provides high optical contrast and good ultrasonic resolution, breaks through the depth limit at the optical transport mean free path (~1 mm), and does not suffer from ultrasound speckles. Third, since optical absorption is strongly related to physiological status, such as oxygen saturation and concentration of hemoglobin, PA signals contain functional information. Accordingly, in recent years PAT has been successfully applied to in vivo blood vessel imaging [4], [5] and brain structural and functional imaging [6]-[8] of small animals. Thermoacoustic tomography (TAT) is another important PA-based imaging technology. Utilizing a microwave illumination source, TAT can image deeper than PAT, and is especially suitable for breast tumor imaging [9]-[11].
To enable the reconstruction of three-dimensional (3-D) tissue structures in PAT, spherical, planar, and cylindrical geometries are normally used in PA signal acquisition. In most applications, they are reduced to circular and linear geometries to image a two-dimensional (2-D) tissue cross section. Linear detection geometry normally provides poorer image quality than circular geometry, because of its limited view of detection. Yet linear geometry is widely used because it is easily applicable to various tissue shapes and can lead to fast, or even real-time, imaging [4], [12] by using a linear transducer array and corresponding fast algorithms [13], [14]. Circular geometry offers full-view detection and hence good image resolution. However, imaging in such geometry is usually very slow for two reasons: the long signal acquisition time (introduced by rotating a single-element transducer around the tissue) and the large time cost of the reconstruction algorithms (either the commonly used exact back-projection algorithm [15] or the simple delay-and-sum algorithm [16]). The former can be overcome by using a ring-shaped transducer array [17]. But the latter, to the best of our knowledge, still remains a problem. For example, the 512-element-array-based PAT system reported in [17] can achieve real-time signal acquisition, but not real-time image formation.
A possible solution is to reduce the number of detection angles in circular geometry, which will decrease not only image reconstruction time, but also signal acquisition time or equipment cost. Nevertheless, for most algorithms the quality of the reconstructed image is strongly related to the number of detection angles. In many cases, 128 angles or fewer may result in strong artifacts in the reconstructed image. The compressed sensing algorithm can be utilized to improve image quality [18]. But this algorithm is in essence an optimization approach that requires iterative calculation, and therefore is not suitable for fast or real-time imaging.
In this paper, we introduce the deconvolution reconstruction (DR) algorithm for PAT in 3-D spherical geometry and 2-D circular geometry. The DR algorithm is very fast since its key step is the Fourier-based deconvolution. Moreover, the DR algorithm renders better image quality than other popular algorithms when using a small number of detection angles. Therefore, the DR algorithm promises to enable real-time imaging in circular geometry. We have previously published basic formulas and simulation results of the DR algorithm to show that it has good precision and is insensitive to data noise [19]. Here, the DR algorithm is systematically described and then verified by experimental data, with an emphasis on the case of sparse detection angles.
The paper is organized as follows. First, a brief description of the PA forward problem and solution is provided. Then the DR algorithm is derived mathematically and explained physically. In vivo experimental results on the mouse cerebral cortex are then presented to validate this algorithm.
2 Forward Solution
The PA theory, as summarized in [20], is briefly described here. The relation between the illuminating light and the excited acoustic wave in an inviscid medium obeys the following wave equation:
(1) |
where A(r) is the absorbed energy density, p(r, t) is the excited acoustic pressure, r is the 3-D position vector, t is the time, I(t) is the illumination pulse function, c is the sound speed, β is the coefficient of volumetric thermal expansion, and Cp is the specific heat.
I(t) is usually assumed to be a delta function, and c is assumed to be constant. PA signals are detected along a sphere, defined by rd, whose radius is rd and whose center is the origin. Then, by using Green’s functions, the detected PA signal p(rd, t) can be expressed as
![]() |
(2) |
In practical experiments, each detected signal is convolved by both the illumination pulse and the transducer impulse response. We can get a better estimation of p(rd, t) by deconvolving the signal if the impulse response is known and the signal-to-noise ratio is sufficiently high. Then image reconstruction is an inverse problem of calculating A(r) from p(rd, t) , solved algorithmically.
3 DR Algorithm
The basic idea of the DR algorithm is to construct a 3-D system. When the system input is A(r) , the output will be a 3-D space function that is related to p(rd, t) . By reasonable design, this system can be constructed to be linear and shift-invariant. Then A(r) can be calculated by deconvolution. In this section, we will give the 3-D formulation, physical meaning, and 2-D discrete implementation (with time complexity analysis) of the DR algorithm.
3.1 Formulation
Fundamental formulas of the DR algorithm are presented here. Readers can refer to [19] for more details.
Let us define
(3) |
Substituting (2) into (3) leads to
![]() |
(4) |
where η = β/(4πCp) .
In order to construct the linear and shift-invariant system, we start by constructing the system output from p(rd, t) when the input is A(r) . The output function C(r) is constructed as
(5) |
where tmax is an adjustable parameter. PA signals during time 0−tmax are used to construct C(r) . In order to cover the major information in PA signals, normally it should satisfy that ctmax ≥ 2rd . C(r) is zero when |r| > ctmax.
Substituting (4) into (5) leads to
![]() |
(6) |
where the integral surface Ω can be expressed as
(7) |
Considering that the detected tissue is contained by the sphere rd, A(r’) has nonzero values only if |r’| < rd (which means A(r’) needs to be integrated only over Ω within this region). Further, 2 r’ < ctmax − rd if ctmax is greater than 3rd or the tissue volume is relatively small. If each ultrasonic detector receives photoacoustic signals within a sufficiently small solid angle, (7) can be approximated to the following surface by using the Taylor series expanded to the first order [19]:
(8) |
The smaller the maximum of |r’| is, the smaller the ignored higher-order Taylor polynomials. In other words, a smaller tissue volume in comparison to the detection radius results in less error. Thus, (6) can be approximated as
![]() |
(9) |
Now the relationship between A(r) and C(r) is clear. The constructed system as described by (9) can be easily shown to be linear and shift-invariant. In other words, if A(r’) is shifted spatially by Δr , the response C(r) will be shifted by the same distance in the same direction because |r’ − r| = ctmax − rd must hold. Therefore, (9) can be rewritten as
(10) |
where h(r) is the system impulse response—the response of the constructed system to a point photoacoustic source located at the origin—and * represents 3-D convolution. When the system input is the 3-D delta function, h(r) is the system output:
![]() |
(11) |
or
(12) |
Based on (10), A(r) can be calculated by a simplified Wiener deconvolution method [21]:
(13) |
where , , and are the 3-D Fourier transforms of A(r) , h(r) , and C(r) , respectively; and λ is a constant (adjustable according to applications).
To summarize, the DR algorithm consists of two steps: first, construct C(r) based on (3) and (5); second, calculate A(r) by (13). It is worth noting again that DR is an approximate algorithm because (9) is approximated from (6). Smaller tissue volumes in comparison to the detection radius result in less error. It has also been shown that the optimal value of tmax is 2rd/c (as used in the following experiments), where DR provides the fastest calculation while maintaining good image quality [19].
3.2 Physical meaning
The physical meanings of the key formulas in this DR algorithm are explained here. The function S(rd, t) , defined in (3), can be understood as the processed time-domain PA signal, which is proportional to the velocity potential multiplied by the time of arrival [20]. It is due to the integration of A(r) over a spherical shell, as shown in (4). Then S(rd, t) is transformed to C(r) by (5). The value of C(r) at position r corresponds to the signal S processed from the pressure p received by the r -direction transducer, located at position r · rd/|r|, at time (ctmax − |r|)/c . In other words, the value of C(r) at position r corresponds to the tissue lying on a spherical surface whose center is r · rd/|r| and whose radius is ctmax − |r| , as shown in Fig. 1. This transformation allows us to merge two variables (rd, t) into a single variable r .
Figure 1.
Illustration of detection geometry and photoacoustic signal integration.
According to (10), C(r) approximates to the convolution of A(r) and h(r) when each acoustic detector receives PA signals within a narrow cone. As shown in (12), h(r) approaches infinity on the sphere whose center is the origin and whose radius is ctmax − rd, and equals zero at other places. Based on the characteristics of the delta function and the origin-symmetric shape of h(r) , it is easy to derive that A(r) * h(r) at position r equals to the integration of A(r) over a sphere whose center is r and radius is ctmax− rd, as shown in Fig. 1.
In short, C(r) and A(r) * h(r) correspond to integrations on two spherical surfaces that do not perfectly overlap, and thereby only approximately equal. This misalignment is the physical origin of the error in the DR algorithm. It is also worth mentioning that these two spheres are tangential on the acoustic axis of each acoustic detector in the direction of −r (or r ). If the tissue volume is relatively small, the mismatch between these two spheres within the tissue will also be small, and so will be the error of this DR algorithm. This explanation is consistent with the conclusion of the previous mathematical derivation.
While the integration spherical surface of C(r) has a radius of curvature dependent on variable r , the integration spherical surface of A(r) * h(r) has a constant radius of curvature once tmax is set. We can exactly match the approximate spherical surface with the original spherical surface at one position. If the tissue volume is centered within the detection surface, it is reasonable to set the matching position at the origin by choosing ctmax = 2rd.
One can further approximate the integration spherical shells to planes orthogonal to the acoustic axis [22], which is tantamount to a zero-order Taylor expansion of the cosine function of the acoustic detection angle. Although such an approximation reduces the spherical Radon transform to the planar counterpart, errors in the reconstructed images can be severe unless the tissue volume is extremely small.
3.3 Discrete implementation
Here we describe how to implement the DR algorithm in the most common 2-D applications, where PA signals are detected along a circle. The transducer used here should receive signals from the focal plane and reject out-of-plane signals because the acoustic detection is cylindrically focused. In this case, the convolution in (10) will approximately hold true in the 2-D case ( r is reduced to a 2-D position vector).
During detection, the generated acoustic signal p(rd, t) , in which rd is reduced to follow a circle, is sampled in both the space and time domains. Thus, the detected p(rd, t) can be represented by an Na × Nt matrix, where Na is the number of detection angles and Nt is the number of sampled time points at each detection angle. Also, S(rd, t) is represented by an Na × Nt matrix.
Suppose we want to reconstruct an image with the resolution of N × N (stored as an A(r) matrix), which represents a spatial size of a × a (located at the center of the detection circle). h(r) can be represented by a matrix corresponding to the spatial size of 2rd × 2rd (here, ctmax = 2rd). Since the A(r) matrix and h(r) matrix will be convolved, each of their elements should represent the same discrete space interval. So h(r) should be constructed as a (N · (2rd / a)×(N · 2rd / a) matrix, and C(r) , the convolution of A(r) and h(r) , should be constructed as a [N · (2rd / a +1)−1] × [N · (2rd / a +1)−1] matrix. While calculating an element in the C(r) matrix, the spatial coordinate of this element is transformed to the corresponding coordinate in the S(rd, t) matrix, and then the value of the nearest S(rd, t) element is chosen (bilinear interpolation can be used for better accuracy at the cost of computation time). After all the elements of the C(r) matrix are calculated, the A(r) matrix, namely, the to-be-reconstructed image, can be obtained by using deconvolution.
The main steps in the DR algorithm are constructing C(r) and deconvolution. To speed up the algorithm, h(r) and the relationship between the variables of C(r) and S(rd, t) can be precalculated. The size of the C(r) matrix is on the order of N2, so constructing C(r) has a time complexity of Θ(N2) . The deconvolution step involves fast Fourier transformation (FFT), division in the frequency domain, and inverse fast Fourier transformation (IFFT). The time complexity of the division is Θ(N2) , and that of both FFT and IFFT is Θ(N2 log N) . Thus, in total, the DR algorithm has a time complexity of Θ(N2 log N) .
In comparison, back-projection algorithms and the delay-and-sum algorithm, two kinds of commonly used time-domain algorithms, add up all the Na detected signals (with or without processing in the time domain) at each pixel of reconstructed image. So, they have a time complexity of Θ(NaN2) . Considering that Na is normally of the same order as N for good image quality, the time complexity of these algorithms can be written as Θ(N3) . Therefore, the DR algorithm is much faster than back-projection algorithms and the delay-and-sum algorithm, especially when reconstructing high-resolution images. Moreover, as far as we know, the available frequency-domain algorithms for circular geometry either use a similar projection strategy [15] as these time-domain algorithms or calculate based on Bessel and Hankel functions [23]. They are even more time-consuming and in fact rarely used in practice.
4 Results and Discussion
The DR algorithm was applied to in vivo mouse brain imaging. The head of a mouse was depilated and then imaged with intact skull and skin. The experimental setup, as shown in Fig. 2, was similar to that of [17]. The Nd:YAG laser (LOTIS II LS-2137/2) generated pulses with a wavelength of 532 nm, a width of 16 ns and a repetition rate of 10 Hz. PA signals were received by a 512-element circular transducer array whose diameter was 5 cm. The center frequency of the transducer was 5 MHz, and the bandwidth was greater than 80%. The signals were amplified, sampled at 40 MHz, and transferred to a computer for processing by the signal acquisition system.
Figure 2.
Experimental setup.
The received PA signals (after deconvolving the transducer impulse response) are shown in Fig. 3(a). The two axes denote time and detection angle; the grayscale denotes normalized signal amplitude. The constructed space function C(r) (also normalized) in the DR algorithm is shown in Fig. 3(b). Here the parameter tmax is set to 2rd/c , and only the signals during 10–33 μs are used.
Figure 3.
Received photoacoustic signals (a) and constructed space function C(r) (b).
Figures 4(a) and 4(b) show the images reconstructed by the exact back-projection algorithm [15] and the DR algorithm, respectively, when signals from all 512 detection angles are used. Then we tested the robustness of these two algorithms by reducing the number of angles uniformly around the detection circle. Figures 4(c) and 4(d) show the images reconstructed over 128 angles (selected every fourth one from the 512 angles). Figures 4(e) and 4(f) show the images reconstructed over 64 angles. For comparison, we opened the skin of the mouse brain after PA imaging and took a photograph of the cerebral cortex [Fig. 4(g)].
Figure 4.
In vivo and noninvasive reconstructed images using (a) back-projection over 512 angles, (b) deconvolution reconstruction over 512 angles, (c) back-projection over 128 angles, (d) deconvolution reconstruction over 128 angles, (e) back-projection over 64 angles, (f) deconvolution reconstruction over 64 angles. (g) Photograph of the mouse cerebral cortex taken after imaging.
When reconstructed over 512 angles, the results of both algorithms are nearly identical and agree well with the photograph. The DR algorithm outperforms the back-projection algorithm when using signals from fewer angles. It can be seen that the reconstructed image of the DR algorithm over 128 angles is as good as the image reconstructed over 512 angles, while the back-projection algorithm’s result suffers from many vessel-like artifacts. When reconstructed over 64 angles, the image from the back-projection algorithm has very strong vessel-like artifacts that obscure real vessels. The image from the DR algorithm is blurred but generally better.
Here we explain why the DR algorithm performs better. The exact back-projection algorithm uses an inverse formula, which ideally requires knowledge of the detected signals at every position of the detection circle. But in practice one can detect signals from only a limited number of angles. The back-projection algorithm does not project the undetected signals by assuming them to be zero. This assumption is in fact not true, and it introduces large errors if detection angles are sparse. In contrast, the DR algorithm is based on a valid forward model given by (10). The effectiveness of this forward model will not be affected by the number of detection angles. In fact, in C(r) only the elements perfectly matching with the measurement points are exactly known. If the detection angles are too sparse, this model will be ill-conditioned and lead to unstable solutions. But it is a better starting point than an invalid inverse formula. Moreover, the DR algorithm’s forward model is based on convolution, which can be directly and quickly solved using a deconvolution method without iterative calculations normally included in most forward-model-based methods. In our deconvolution, the unmeasured elements in C(r) are estimated by the detected signals. The solution is not a least-squares one, but is close to.
The DR algorithm assumes, as mentioned above, that the object dimension is relatively small compared to the detection circle. In our experiment, the cross section of the mouse head is approximately elliptically shaped with a major axis of 2 cm, which is not much less than the detection diameter (5 cm). However, most vessels are within a 0.6 × 0.8 cm2 region, roughly centered within the detection circle, and are clearly imaged. In practice, if an object is so large that the DR algorithm cannot reconstruct the full cross section clearly, we can limit the region of interest at the detection center for better image quality.
To compare the speeds of the back-projection and DR algorithms, we reconstructed images of a 2 × 2 cm2 region (at the center of the transducer array) with different numbers of pixels. In practical applications, we reconstruct images with more pixels if we can achieve better spatial resolution, otherwise adding pixels is meaningless because the added pixels contain little new spatial information. Moreover, more detection angles are normally required for better spatial resolution. Therefore, in order to investigate the calculation efficiencies of these two algorithms with different requirements of spatial resolution, the number of detection angles and the pixel resolution of reconstructed images in our calculation were varied simultaneously. In other words, we used signals from N detection angles to reconstruct an N × N image and recorded the time cost of each algorithm (Intel Core2 Duo CPU @ 3.00GHz, Matlab R2009a), where N was chosen to vary from 64, 128, 256, to 512. The results are shown in Fig. 5.
Figure 5.
Time costs of the back-projection algorithm and deconvolution reconstruction algorithm when reconstructing N × N images using signals from N detection angles.
It can be seen that the time cost of the DR algorithm is much smaller than that of the back-projection algorithm. For each algorithm, the trend of increased time cost with N is consistent with the time complexity analysis in section 3.3. When N = 512 , the DR algorithm is about ten times faster than the back-projection algorithm. It can be predicted that the speed advantage of the DR algorithm is greater when N > 512 . Moreover, according to these data, real-time imaging is possible only if N ≤ 256 (in our system) if the DR algorithm is used. Since the DR algorithm presents stronger robustness for sparse detection angles, it should be preferably considered for real-time PA imaging in circular geometry.
5 Conclusions
Based on previous simulations, this work supplements the DR algorithm theoretically and validates it experimentally. The DR algorithm uses a convolution-based forward model, which is more effective than the inverse formula for sparse detection angles. Moreover, this model can be quickly solved by using a deconvolution-based method. Experimental results show that the DR algorithm provides image quality nearly identical to that of exact back-projection algorithm, when using signals detected from 512 angles. When using 128 and 64 angles, the DR algorithm provides better images without vessel-like artifacts. It is also shown that the DR algorithm runs faster, agreeing well with the time complexity analysis. When designing a real-time PAT system, the number of detection angles cannot be very large, and the image reconstruction algorithm should be fast, so the DR algorithm becomes a good choice. In addition, by utilizing the DR algorithm in PAT systems, we can use fewer measurements for given image quality, thereby reducing experimental cost.
Acknowledgements
This work was sponsored in part by National Institutes of Health grants R01 EB008085, U54 CA136398, and R01 CA113453901. L.W. has a financial interest in Microphotoacoustics, Inc. and Endra, Inc., which, however, did not support this work.
References
- [1].Wang LV. Multiscale photoacoustic microscopy and computed tomography. Nat. Photonics. 2009 Sep;3:503–509. doi: 10.1038/nphoton.2009.157. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Xu M, Wang LV. Photoacoustic imaging in biomedicine. Rev. Sci. Instrum. 2006 Apr.77(4):41101–41122. [Google Scholar]
- [3].Bell AG. On the production of sound by light. Am. J. Sci. 20:305, 1880. [Google Scholar]
- [4].Niederhauser JJ, Jaeger M, Lemor R, Weber P, Frenz M. Combined ultrasound and optoacoustic system for real-time high-contrast vascular imaging in vivo. IEEE Trans. Med. Imaging. 2005 Apr.24(4):436–440. doi: 10.1109/tmi.2004.843199. [DOI] [PubMed] [Google Scholar]
- [5].Zhang HF, Maslov K, Stoica G, Wang LV. Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging. Nat. Biotechno. 2006 Jul.24(7):848–851. doi: 10.1038/nbt1220. [DOI] [PubMed] [Google Scholar]
- [6].Wang X, Pang Y, Ku G, Xie X, Stoica G, Wang LV. Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain. Nat. Biotechno. 2003 Jul.21(7):803–806. doi: 10.1038/nbt839. [DOI] [PubMed] [Google Scholar]
- [7].Li C, Wang LV. Photoacoustic tomography of the mouse cerebral cortex with a high-numerical-aperture-based virtual point detector. J. Biomed. Opt. 2009 Apr.14(2):024047. doi: 10.1117/1.3122365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Hu S, Maslov K, Tsytsarev V, Wang LV. Functional transcranial brain imaging by optical-resolution photoacoustic microscopy. J. Biomed. Opt. 2009 Aug.14(4):040503. doi: 10.1117/1.3194136. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Wang LV, Zhao XM, Sun HT, Ku G. Microwave-induced acoustic imaging of biological tissues. Rev. Sci. Instrum. 1999 Sep.70(9):3744–3748. [Google Scholar]
- [10].Kruger RA, Miller KD, Reynolds HE, Kiser WL, Reinecke DR, Kruger GA. Breast cancer in vivo: contrast enhancement with thermoacoustic CT at 434 MHz – feasibility study. Radiology. 2000 Jul.216:279–283. doi: 10.1148/radiology.216.1.r00jl30279. [DOI] [PubMed] [Google Scholar]
- [11].Pramanik M, Ku G, Li CH, Wang LV. Design and evaluation of a novel breast cancer detection system combining both thermoacoustic (TA) and photoacoustic (PA) tomography. Med. Phys. 2008 Jun.35(6):2218–2223. doi: 10.1118/1.2911157. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Kolkman RGM, Brands PJ, Steenbergen W, van Leeuwen TG. Real-time in vivo photoacoustic and ultrasound imaging. J. Biomed. Opt. 2008 Oct.13(5):050510. doi: 10.1117/1.3005421. [DOI] [PubMed] [Google Scholar]
- [13].Xu Y, Feng D, Wang LV. Exact frequency-domain reconstruction for thermoacoustic tomography – I: planar geometry. IEEE Trans. Med. Imaging. 2002 Jul.21(7):823–828. doi: 10.1109/tmi.2002.801172. [DOI] [PubMed] [Google Scholar]
- [14].Kostli KP, Beard PC. Two-dimensional photoacoustic imaging by use of Fourier-transform image reconstruction and a detector with an anisotropic response. Appl. Opt. 2003 Apr.42(10):1899–1908. doi: 10.1364/ao.42.001899. [DOI] [PubMed] [Google Scholar]
- [15].Xu M, Wang LV. Universal back-projection algorithm for photoacoustic computed tomography. Phys. Rev. E. 2005 Jan.71(1):016706. doi: 10.1103/PhysRevE.71.016706. [DOI] [PubMed] [Google Scholar]
- [16].Hoelen CGA, Mul FFM. Image reconstruction for photoacoustic scanning of tissue structures. Appl. Opt. 2000 Nov.39(31):5872–5883. doi: 10.1364/ao.39.005872. [DOI] [PubMed] [Google Scholar]
- [17].Gamelin J, Maurudis A, Aguirre A, Huang F, Guo P, Wang LV, Zhu Q. A real-time photoacoustic tomography system for small animals. Opt. Express. 2009 Jun.17(13):10489–10498. doi: 10.1364/oe.17.010489. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Provost J, Lesage F. The application of compressed sensing for photo-acoustic tomography. IEEE Trans. Med. Imaging. 2009 Apr.28(4):585–594. doi: 10.1109/TMI.2008.2007825. [DOI] [PubMed] [Google Scholar]
- [19].Zhang C, Wang Y. Deconvolution reconstruction of full-view and limited-view photoacoustic tomography: a simulation study. J. Opt. Soc. Am. A. 2008 Oct.25(10):2436–2443. doi: 10.1364/josaa.25.002436. [DOI] [PubMed] [Google Scholar]
- [20].Wang LV, Wu H-I. Biomedical Optics: Principles and Imaging. Wiley; Hoboken, N.J.: 2007. [Google Scholar]
- [21].Riad SM. The deconvolution problem: an overview. Proc. IEEE. 1986 Jan.74(1):82–85. [Google Scholar]
- [22].Kruger RA, Liu P, Fang Y, Appledorn CR. Photoacoustic ultrasound (PAUS) – reconstruction tomography. Med. Phys. 1995 Oct.22(10):1605–1609. doi: 10.1118/1.597429. [DOI] [PubMed] [Google Scholar]
- [23].Xu Y, Xu M, Wang LV. Exact frequency-domain reconstruction for thermoacoustic tomography – II: cylindrical geometry. IEEE Trans. Med. Imaging. 2002 Jul.21(7):829–833. doi: 10.1109/TMI.2002.801171. [DOI] [PubMed] [Google Scholar]