Skip to main content
Journal of Diabetes Science and Technology logoLink to Journal of Diabetes Science and Technology
. 2010 Sep 1;4(5):1269–1275. doi: 10.1177/193229681000400529

Glucose Information for Tight Glycemic Control: Different Methods with Different Challenges

Christian Weber 1, Kurt Neeser 1
PMCID: PMC2956809  PMID: 20920450

Abstract

Rigorous glucose control is essential for prevention of diabetes-related complications in diabetes patients. Even without diabetes, tight glucose control is beneficial in hospitalized, critically ill patients.

Actually, three different glucose measurement methods are used: (1) hand held devices, (2) blood-gas analyzers, and (3) laboratory analyzers in core laboratories. Each method is subject to specific challenges and limitations that can affect the overall system performance.

In this article, we aim to demonstrate that even glucose measurement results from core laboratories (professional laboratory systems) do not necessarily reflect the absolute “true” glucose level of a patient.

Keywords: accuracy, glucose measurement, glycemic control, precision, standards

Introduction

Diabetes mellitus is a group of metabolic diseases characterized by hyperglycemia or hypoglycemia, resulting from defects in insulin secretion, insulin action, or both.

Since the publication of the seminal studies of the Diabetes Control and Complications Trial (DCCT; for type 1 diabetes) and the United Kingdom Prospective Diabetes Study (UKPDS; for type 2 diabetes), there is consensus that tight glycemic control represents a cornerstone in the prevention of diabetes-related complications in outpatient care.

Hospitalized patients (with and without diabetes as an underlying condition) and especially those treated in intensive care units (ICUs) are even more prone to dys- regulations of the glucose metabolism, with a considerable consequent risk for higher mortality rates.

Measurement of glycemia in plasma or blood is therefore inalienable in the management of diabetes mellitus or otherwise altered glucose metabolism.

Three different methods are available for the measure-ment of glucose levels: (1) small handheld devices [self-monitoring of blood glucose (SMBG)] for the measurement of capillary blood samples, principally designed for self-measurement of glucose values by the patient itself but sometimes also used in hospital wards (“bedside testing”); (2) more sophisticated devices, usually able to determine a couple more laboratory values (e.g., blood-gas analysis and electrolytes) and principally designed for use in intensive or intermediate care units [i.e., blood-gas analyzer (BGA)], in Europe often referred to as “point-of-care testing;” and (3) central laboratories that are spatially separated from the care units.

The aim of this article is to highlight the accuracy, differences, and limitations of these three methods and the implications for therapy.

Chemistry Principles

Three different methods are predominantly used to determine glucose concentrations:

  1. Glucose oxidase (GOx) method: The enzyme GOx catalyzes the oxidation of glucose into gluconolacton and H2O2 (hydrogen peroxide). In a subsequent peroxidase-mediated indicator reaction, H2O2 oxides a reduced chromogen (a leuco dye) under development of color, which can be measured by photometry. The color intensity is proportional to the glucose concentration.

  2. Hexokinase method: Glucose is phosphorylated in the presence of hexokinase and adenosine triphosphate to glucose-6-phosphate. This compound reacts with nicotinamide adenine dinucleotide (NAD) phosphate, creating 6-phosphogluconate and reduced nicotinamide adenine dinucleotide phosphate (NADPH), catalyzed by glucose-6-phosphate-dehydrogenase. The rise of NADPH until the end of the reaction is measured, and the rising optical extinction is proportional to the glucose concentration in the sample.

  3. Glucose dehydrogenase (GDH) method: Glucose is oxidized by GDH to gluconolacton. The released hydrogen is transferred to NAD, creating NADPH. The rise of the latter until the end of the reaction is measured, the rising extinction being proportional to the glucose concentration in the sample.

SMBG systems employ very diverse biochemical reactions. Coenzymes flavine adenine dinucleotide (FAD), pyrrolo quinoline quinine (PQQ), and NAD can be found. Glucose oxidases or dehydrogenases catalyze the oxidation of glucose, but nearly every product is unique with respect to many important features. Simple rules such as “PQQ is bad for specificity and FAD is good” are misleading. Specificity is determined by the enzyme protein and not by the coenzyme. A review of these methods and of their advantages and disadvantages is far beyond the scope of this article.

The general chemistry principles are essentially the same in BGA devices and in central laboratory analyzers, with the main difference that a “wet chemistry principle“ and different detection methods (e.g., direct electrodes) are used. This has an implication for the samples used for analysis. The SMBG devices are only designed for testing capillary blood samples; most of these devices provide a plasma equivalent value (as recommended by the International Federation of Clinical Chemistry and Laboratory Medicine [IFCC]). Even when labeled correctly, this may lead to confusion or possible errors. The “wet chemistry” procedure allows an analysis of whole blood, hemolyzed whole blood, or plasma from venous, arterial, or capillary sources.

The identification and documentation of the type of the sample is crucial in this case, because fasting glucose values in venous blood are 5–10% lower than in arterial samples, as capillary samples are showing 5–15% higher values compared to venous blood samples. Similarly, venous or capillary plasma samples are showing 10–15% higher values compared to whole blood hemolysate; deproteinization of samples have a similar effect due to the “volume displacement effect.”1

Regulatory Standards

SMBG systems are medical devices and belong to the group of in vitro diagnostics. In Europe, SMBG systems have to comply with the directive 98/79/EC on in vitro diagnostic medical devices, which specifies general requirements to ensure overall safety and quality.2 The standard DIN EN ISO 15197:2003 specifies require-ments for SMBG systems, e.g., with regard to system performance, accuracy, and precision.3 The minimum acceptable system accuracy requirements are based on the medical requirements for glucose monitoring: “≥95% of the individual glucose results shall fall within ±15 mg/dl of the results of the manufacturer's measurement procedure at glucose concentrations <75 mg/dl and within ±20% at glucose concentrations ≥75 mg/dl.”3 Manufacturers of SMBG devices have to provide evidence of conformity with the standard DIN EN ISO 15197:2003 in order to get the Conformité Européenne label for their products.

Stricter accuracy requirements for SMBG systems are under consideration for implementation by both the Food and Drug Administration and the International Organization for Standardization (ISO).47 Obviously, even a smaller number of SMBG systems currently on the market will fulfill these stricter requirements.8

The standards for BGAs and central laboratory systems follow—beyond ISO norms—mainly recommendations of the IFCC, the Clinical and Laboratory Standards Institute, and the College of American Pathologists (with a special focus on the United States). Accuracy and precision are assured by comparing measurements internally (e.g., lot-to-lot variation and between-cartridge variation) and externally (between-device variation by comparing standard test samples in different laboratories, so-called “proficiency testing”). Results of these quality tests often serve as reference value in the assessment of the accuracy and precision of SMBG devices.

Confounding Factors

A number of possible confounding factors, independent from the methodology used, may have to be taken into consideration before translating the measurement results into clinical action.

Limitations of the systems include substance interferences and performance parameters. Substance interferences refers to interference of the system with any substance present in the blood at increased or abnormal concentration that affects the reliability of the blood glucose (BG) results. Substances interfering with the determination method can either originate from drugs taken by the patient or from therapies initiated by the medical staff or are naturally occurring in the blood. Examples are acetaminophen, ascorbic acid, dopamine, icodextrin, bilirubin, triglycerides, and paraproteins.9

Performance parameters refers to limitations due to the specifications and technology of a system, e.g., hematocrit values may play a very important role in blood samples taken in an inpatient context. Hematocrit values of 25% may result in artificially high glucose results by 25% and inversely for values above 65%. Table 1 shows the role of important performance parameters, again of different SMBG systems.

Table 1.

Performance Parameters of Different Self-Monitoring Of Blood Glucose Systemsa

Parameterb Accu-Chek® Aviva Accu-Chek® Compact Plus OneTouch Ultra2® Ascensia Contour® II FreeStyle Lite®/FreeStyle Freedom® Precision Xtra®
BG measurement range (mmol/liter) 0.6–33.3 0.6–33.3 1.1–33.3 0.6–33.3 1.1–27.8 1.1–27.8
Hematocrit (%) 20–70 25–65 30–55 0–70 15–65 30–60
Humidity (%) 10–90 20–85 10–90 10–93 5–90 10–90
BG test strip stability EXP on BG test strip vial (date of manufacture + 18 months) EXP on BG test strip vial (date of manufacture + 18 months) or PAO 3 months PAO 3 months EXP on BG test strip vial or PAO 6 months EXP on BG test strip vial EXP on BG test strip vial
Operating temperature (°C) +6 to +44 +10 to +40 +6 to +44 +5 to +45c +5 to +40 +10 to +50
Underdose protection yes yes no yes yes yes
Double dosing yes yes no yes yes yes
Altitude (m) <3094 <4000 <3000 <3000 <3000 <2200
Total cholesterol (mmol/liter) <13.0 n/a <18.1 <13.0 <13.0 <13.0
Triglycerides (mmol/liter) <54.7 <57.0 <34.2 <34.2 <34.2 <11.4
a

EXP, expiry date; PAO, period after opening

b

Taken from product user labeling of the SMBG systems.

c

When moving the BG meter Ascensia Contour from one location to another with a different temperature, Bayer recommends to allow 20 minutes for temperature adjustment; temperature variations are given in the BG test strip package.

But as there are only three general chemistry principles underlying the three measurement methodologies, it is important to point out that confounding factors are, in principle, germane to any of them, whereas the hexo-kinase method is the least prone to analytical interactions.10

Additional potential interferences should be considered. Blood-gas analyzer and central laboratory results are more prone to pre-analytical errors, e.g., the choice of wrong test tubes, erroneous or missing labeling of vials, partial filling of vials, or long time gaps between collection and analysis of the samples, leading to a degeneration of glucose.

Operator influence is probably more pronounced in the SMBG method, as it is often used by laypeople.

Table 2 provides a qualitative overview on confounding factors and potential interferences for the three methods.

Table 2.

Influences on System Performance and Consequent Risk of Errora

Influence Handheld devices BGAs Core Laboratory
Factor Risk of error Factor Risk of error Factor Risk of error
System accuracy + / ++ + / ++ ++ + ++ / +++ +
Operator influence PwD /layperson HCP ++++ HCP ++1 HCP +b
Blood sample type capillary None to low Arterial, venous, mixed, or capillary (whole blood) Lowc Arterial, venous, mixed, or capillary (varying sample presentations) Potentially highd
Blood sample transfer None n/a Short Low Long Potentially highe
Blood sample stabilization None n/a Nonec Low e.g., EDTA, heparin, NaF Low to mediumf
Blood sample storage None None Short-term Low Long-term Low to mediume
Environmental conditions Varying environmental conditions Medium to high Mostly controlled conditions Low Controlled laboratory conditions Very low
Blood sample matrix Whole blood Whole blood Whole blood Low Blood plasma Low
a

Dilution effects if sample taken from arterial or central venous lines. PwD, person with diabetes; HCP, health care professional; EDTA, ethylenediaminetetraacetic acid

b

Mainly automated.

c

Sometimes preheparinezed vials used.

d

Erroneous or missing labeling of the sample.

e

Nonrespect of required transport conditions.

f

Erroneous use of nonappropriate sample tubes/vials.

Accuracy and Precision

Generally, accuracy is defined as the closeness of agreement between a test result and the accepted reference value and precision is defined as the closeness of agreement between independent test results obtained under stipulated conditions.2

In the case of the analysis of BG, the situation is complex because (A) BG is an unstable analyte, i.e., BG is degraded gradually by glycolysis after blood sampling, and (B) there is no whole BG standard available, which is usually a prerequisite for establishing accuracy of measuring devices.7

The correct meaning of these so-called specific performance characteristics is not always clear in the laboratory medicine area, both on the side of laboratory users and manufacturers.11 Moreover, the use of terms describing specific performance characteristics, such as “accuracy,” “precision,” “trueness,” and “total error,” differs partially between locales, e.g., the United States and Europe. This shall be kept in mind if analyzing the respective literature published so far. We would recommend to adhere to the definitions given in ISO 3534-1:1993,12 ISO 5725-1:1994,13 and ISO 15197:2003.2

There is a plethora of literature commenting on the performance of portable glucose meters compared to laboratory values, and it would be beyond the scope of this article to review them in detail. In this context, we would like to point to a study by Boyd and Bruns,14 where a simulation of insulin dosing errors due to BG measurement errors was performed.

They concluded that portable BG meters, if they are meeting the current quality standards, permit at a high percentage the administration of insulin doses intended to reach the target glucose range. But the same study also indicated that large errors of insulin doses (≥4 U of insulin) occurred nevertheless in over 5% of the time when the coefficient of variation exceeded 10–15%.

An overview of Bergenstal15 and a study of Koschinsky and colleagues16 are heading in the same direction, showing that results of portable meters are, in principle, reliable and able to guide treatment options, if reasonable safety margins are added to the targeted glucose range.

Breton and Kovatchev17 investigated the impact of SMBG errors on the short- and long-term response to insulin and glucose challenges in type 1 diabetes patients, using a sophisticated computer simulator representing the human metabolic system. Their findings underline that glucose variability, the detection and the risk of hypoglycemic events, and achievement of target BG values can be improved by tightened system accuracy requirements. Such model-based findings may be very helpful as long they imply all relevant factors occurring in the real world of a clinical setting.

At least, there seems to be consensus that different patient groups (i.e., patients with type 2 diabetes and oral antidiabetic therapy, patients with type 1 diabetes and intensified insulin therapy, and patients in the ICU) deserve different levels of accuracy.18

In one study, Freckmann and associates8 evaluated 27 BG monitoring systems from 18 manufacturers for system accuracy according to DIN EN ISO 15197:2003. Twenty-four systems were compared with the GOx reaction (YSI 2300 Glucose Analyzer) and 3 systems with the hexokinase reaction (Roche Hitachi 917). Duplicate measurements of 100 blood samples with a defined distribution of BG concentrations from 20 to 600 mg/dl from ≥100 subjects were included in the evaluation. Of 27 SMBG systems, 16 fulfilled the minimum accuracy requirements of the standard ISO 15197, i.e., ≥95% of their results showed the minimum acceptable accuracy.

Astonishingly, there is a paucity of published papers dealing with the performance of BGA devices—or central laboratory results. The data available so far (i.e., from between-laboratory standard sample assessments of the College of American Pathologists19 and its counterpart in Germany,20 as well as other sources21,22) indicate a precision and accuracy of the laboratory results of ±10%, generally considered as acceptable and probably more accurate and trustful than results from SMBG devices. One has to keep in mind that these values refer to the analysis of standard samples and do not represent the (higher) total error, because it does not accommodate errors that accompany pre-analytical errors and potential interferences described earlier.

Furthermore, a clear distinction should be made between outpatient diabetes management and tight BG control in the ICU. In the latter case, patients are often in very specific conditions (e.g., jaundice, dyslipidemia, or tolerated anemia), which are usually not present in patients for which handheld devices have originally been designed, and a couple of publications point to the related problems in critically ill patients.23,24

Summarizing comment

The DCCT25 and the UKPDS26,27 could demonstrate the paramount importance of intensified therapy to prevent diabetes related complications and that there is a good correlation between glycosylated hemoglobin A1c (A1C) levels and the respective patient-relevant end points.2830 Apparently, the existing level of accuracy of SMBG systems has been sufficient to achieve major improvements in the metabolic situation of patients with diabetes.

However, there is now a growing body of evidence that not only average A1C values but also short-time glycemic peaks and nadirs (in terms of minutes or hours, glycemic variability) represent an independent risk factor for diabetic complications3133 and that short-term peaks are not necessarily reflected by an increased A1C value.3437

Postprandial and fasting glucose are without doubt the main contributors to A1C levels and intraday glycemic variability,3840 and over 15 studies published since 1998 demonstrated that elevated postprandial glycemic values contribute to an approximately three-fold risk to develop coronary heart disease or a cardiovascular event.41

On the other side the Action in Diabetes and Vascular Disease: Preterax and Diamicron Modified Release Controlled Evaluation and the Action to Control Cardiovascular Risk in Diabetes study demonstrated that too tight glycemic control comes along with a higher risk of severe hypoglycemic events and higher mortality in outpatient care.42,43

In the field of inpatient care, interest in tight glycemic control was raised after the seminal study of Van den Berghe and coworkers,44 leading to recommended glycemic targets of 80–110 mg/dl for ICU patients. Today, the targets for glycemic control in the ICU appear to be less stringent, after the Efficacy of Volume Substitution and Insulin Therapy in Severe Sepsis and Glucontrol studies45,46 demonstrated unacceptable rates of hypoglycemia in critically ill patients receiving intensive insulin therapy. Subsequent studies (especially Normoglycemia in Intensive Care Evaluation—Survival Using Glucose Algorithm Regulation47) and meta-analyses48,49 confirmed that too tight control was associated with an increased risk of severe hypo-glycemia without the benefit of significantly reduced hospital mortality.

An explanatory hypothesis could be that the care teams put too much trust in the results of the devices or methods used, considering them as absolute “true values” and ignoring that a total error is imminent to all techniques available. Blood glucose analyzers and core lab methods may have a higher analytical accuracy and precision on their own, compared to SMBG devices, but the advantage could be outweighed by the numerous possible pre-analytical errors related to the first two methods (see Table 2). Actually, the total error of core lab methods seems to be in the range of 8% to 10%.19,20

We would argue that the solution might be less in the direction of stricter standards, as a total error will remain, independent of whether SMBG, BGA, or central laboratory results are used for the implementation of a therapeutic strategy.

Practitioners, especially in the ICU, should weigh the advantages and disadvantages of the respective methods: a greater bias with potentially lower pre-analytical errors versus a somewhat lower bias but with a higher potential for pre-analytical errors.50

So, the introduction of more strict standards would not resolve the problem of the possible pre-analytical errors and/or other confounding factors. Instead, it seems advisable to consider a 10–15% safety margin at the limits of the targeted glycemic range to allow for the sources of error due to the circumstances of the measurement procedure, independent from the methodology used.

Acknowledgments

The authors thank Dr. Serge Kocher, Institute for Medical Informatics and Biostatistics AG, Switzerland, who provided medical writing services for this article.

Abbreviations

A1C

glycosylated hemoglobin A1c

BG

blood glucose

BGA

blood-gas analyzer

DCCT

Diabetes Control and Complications Trial

FAD

flavine adenine dinucleotide

GDH

glucose dehydrogenase

GOx

glucose oxidase

ICU

intensive care unit

IFCC

International Federation of Clinical Chemistry and Laboratory Medicine

ISO

International Organization for Standardization

NAD

nicotinamide adenine dinucleotide

NADPH

reduced nicotinamide adenine dinucleotide phosphate

PQQ

pyrrolo quinoline quinine

SMBG

self-monitoring of blood glucose

UKPDS

United Kingdom Prospective Diabetes Study

References

  • 1.Morrison B, Fleck A. Plasma or whole blood glucose? Clin Chim Acta. 1973;45(3):293–297. doi: 10.1016/0009-8981(73)90441-5. [DOI] [PubMed] [Google Scholar]
  • 2.Lex EU. Directive 98/79/EC on in vitro medical devices. Off J L. 1998;331:1–37. [Google Scholar]
  • 3.ISO 15197:2003. In vitro diagnostic test systems—requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. http://www.iso.org/iso/catalogue_detail.htm?csnumber=26309.
  • 4.Hamburg MA. 2009. (Answer of the FDA to a letter of Garber JR from AACE on the performance of blood glucose testing meters). Personal communication.
  • 5.Klonoff DC. The Food and Drug Administration is now preparing to establish tighter performance requirements for blood glucose monitors. J Diabetes Sci Technol. 2010;4(3):499–504. doi: 10.1177/193229681000400301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Malone B. Higher standards on the way for glucose meters? Clin Lab News. 2009;35(9):11. [Google Scholar]
  • 7.Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol. 2009;3(4):971–980. doi: 10.1177/193229680900300446. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Freckmann G, Baumstark A, Jendrike N, Zschornack E, Kocher S, Tshiananga J, Heister F, Haug C. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technol Ther. 2010;12(3):221–231. doi: 10.1089/dia.2009.0128. [DOI] [PubMed] [Google Scholar]
  • 9.Tang Z, Du X, Louie RF, Kost GJ. Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. Am J Clin Pathol. 2000;113(1):75–86. doi: 10.1309/QAW1-X5XW-BVRQ-5LKQ. [DOI] [PubMed] [Google Scholar]
  • 10.Pelletier O, Arratoon C. Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method. Clin Chem. 1987;33(8):1397–1402. [PubMed] [Google Scholar]
  • 11.Scassellati G. Specific performance characteristics of in-vitro diagnostic medical devices. Biochim Cli. 2000;24:493–497. [Google Scholar]
  • 12.ISO 3534-1:1993. Statistics—vocabulary and symbols. Part 1—probability and general statistical terms. http://www.iso.org/iso/catalogue_detail.htm?csnumber=8919.
  • 13.ISO 5725-1:1994. Accuracy (trueness and precision) of measurement method and results. Part 1: general principles and definitions. http://www.iso.org/iso/catalogue_detail.htm?csnumber=11833.
  • 14.Boyd JC, Bruns DE. Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose. Clin Chem. 2001;47(2):209–214. [PubMed] [Google Scholar]
  • 15.Bergenstal RM. Evaluating the accuracy of modern glucose meters. Insulin. 2008;3(1):5–14. [Google Scholar]
  • 16.Koschinsky T, Heckermann S, Heinemann L. Parameters affecting postprandial blood glucose: effects of blood glucose measurement errors. J Diabetes Sci Technol. 2008;2(1):58–66. doi: 10.1177/193229680800200109. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Breton MD, Kovatchev BP. Impact of blood glucose self-monitoring errors on glucose variability, risk for hypoglycemia, and average glucose control in type 1 diabetes: an in silico study. J Diabetes Sci Technol. 2010;4(3):562–570. doi: 10.1177/193229681000400309. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Klonoff DC. Regulatory controversies surround blood glucose monitoring devices. J Diabetes Sci Technol. 2010;4(2):231–235. doi: 10.1177/193229681000400201. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.CAP Today. Glucose testing variability and the need for an expert oversight. Accessed March 12, 2010. http://www.cap.org/apps/portlets/contentViewer/show.do?printFriendly=true&contentReference=cap_today/feature_stories/0505Commentary.html. [PubMed]
  • 20.Reference Institute for Bioanalytics (RfB), supported by the German Society for Clinical Chemistry and Laboratory Medicine. Accessed March 12, 2010. http://www.dgkl-rfb.de/4daction/g_send_pdfGET/results_GL101.pdf?rvID=GL101&year=2010&type=gesa&split=1&lang=e.
  • 21.Twomey PJ. Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640. J Clin Pathol. 2004;57(7):752–754. doi: 10.1136/jcp.2003.013417. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem. 2002;48(3):436–472. [PubMed] [Google Scholar]
  • 23.Hoedemaekers CW, Klein Gunnewiek JM, Prinsen MA, Willems JL, Van der Hoeven JG. Accuracy of bedside glucose measurement from three glucometers in critically ill patients. Crit Care Med. 2008;36(11):3062–3066. doi: 10.1097/CCM.0b013e318186ffe6. [DOI] [PubMed] [Google Scholar]
  • 24.Vlasselaers D, Herpe TV, Milants I, Eerdekens M, Wouters PJ, Moor BD, den Berghe GV. Blood glucose measurements in arterial blood of intensive care unit patients submitted to tight glycemic control: agreement between bedside tests. J Diabetes Sci Technol. 2008;2(6):932–938. doi: 10.1177/193229680800200603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993;329(14):977–986. doi: 10.1056/NEJM199309303291401. [DOI] [PubMed] [Google Scholar]
  • 26.UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33) Lancet. 1998;352(9131):837–853. [PubMed] [Google Scholar]
  • 27.Stratton IM, Adler AI, Neil HA, Matthews DR, Manley SE, Cull CA, Hadden D, Turner RC, Holman RR. Association of glycaemia with macrovascular and microvascular complications of type 2 diabetes (UKPDS 35): prospective observational study. BMJ. 2000;321(7258):405–412. doi: 10.1136/bmj.321.7258.405. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Nathan DM, Cleary PA, Backlund JY, Genuth SM, Lachin JM, Orchard TJ, Raskin P, Zinman B. Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Study Research Group. Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes. N Engl J Med. 2005;353(25):2643–2653. doi: 10.1056/NEJMoa052187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Holman RR, Paul SK, Bethel MA, Matthews DR, Neil HA. 10-year follow-up of intensive glucose control in type 2 diabetes. N Engl J Med. 2008;59(15):1577–1589. doi: 10.1056/NEJMoa0806470. [DOI] [PubMed] [Google Scholar]
  • 30.Gaede P, Lund-Andersen H, Parving HH, Pedersen O. Effect of a multifactorial intervention on mortality in type 2 diabetes. N Engl J Med. 2008;358(6):580–591. doi: 10.1056/NEJMoa0706245. [DOI] [PubMed] [Google Scholar]
  • 31.Ceriello A. The emerging role of post-prandial hyperglycaemic spikes in the pathogenesis of diabetic complications. Diabet Med. 1998;15(3):188–193. doi: 10.1002/(SICI)1096-9136(199803)15:3<188::AID-DIA545>3.0.CO;2-V. [DOI] [PubMed] [Google Scholar]
  • 32.Ceriello A. The possible role of postprandial hyperglycaemia in the pathogenesis of diabetic complications. Diabetologia. 2003;46(Suppl 1):M9–16. doi: 10.1007/s00125-002-0931-5. [DOI] [PubMed] [Google Scholar]
  • 33.Gallagher A, Home PD. The effect of improved post-prandial blood glucose control on post-prandial metabolism and markers of vascular risk in people with type 2 diabetes. Diabetes Res Clin Pract. 2005;67(3):196–203. doi: 10.1016/j.diabres.2004.07.010. [DOI] [PubMed] [Google Scholar]
  • 34.Hanefeld M, Koehler C, Henkel E, Fuecker K, Schaper F, Temelkova-Kurktschiev T. Post-challenge hyperglycaemia relates more strongly than fasting hyperglycaemia with carotid intima-media thickness: the RIAD Study. Risk Factors in Impaired Glucose Tolerance for Atherosclerosis and Diabetes. Diabet Med. 2000;17(12):835–840. doi: 10.1046/j.1464-5491.2000.00408.x. [DOI] [PubMed] [Google Scholar]
  • 35.Bonora E. Postprandial peaks as a risk factor for cardiovascular disease: epidemiological perspectives. Int J Clin Pract Suppl. 2002;(129):5–11. [PubMed] [Google Scholar]
  • 36.Temelkova-Kurktschiev TS, Koehler C, Henkel E, Leonhardt W, Fuecker K, Hanefeld M. Postchallenge plasma glucose and glycemic spikes are more strongly associated with atherosclerosis than fasting glucose or HbA1c level. Diabetes Care. 2000;23(12):1830–1834. doi: 10.2337/diacare.23.12.1830. [DOI] [PubMed] [Google Scholar]
  • 37.Derr R, Garrett E, Stacy GA, Saudek CD. Is HbA(1c) affected by glycemic instability? Diabetes Care. 2003;26(10):2728–2733. doi: 10.2337/diacare.26.10.2728. [DOI] [PubMed] [Google Scholar]
  • 38.Monnier L, Colette C, Dunseath GJ, Owens DR. The loss of postprandial glycemic control precedes stepwise deterioration of fasting with worsening diabetes. Diabetes Care. 2007;30(2):263–269. doi: 10.2337/dc06-1612. [DOI] [PubMed] [Google Scholar]
  • 39.Landgraf R. The relationship of postprandial glucose to HbA1c. Diab Metab Res Rev. 2004;20(Suppl 2):S9–12. doi: 10.1002/dmrr.517. [DOI] [PubMed] [Google Scholar]
  • 40.Leiter LA, Ceriello A, Davidson JA, Hanefeld M, Monnier L, Owens DR, Tajima N, Tuomilehto J. International Prandial Glucose Regulation Study Group. Postprandial glucose regulation: new data and new implications. Clin Ther. 2005;27(Suppl B):S42–56. doi: 10.1016/j.clinthera.2005.11.020. [DOI] [PubMed] [Google Scholar]
  • 41.Weber C, Schnell O. The assessment of glycemic variability and its impact on diabetes-related complications: an overview. Diabetes Technol Ther. 2009;11(10):623–633. doi: 10.1089/dia.2009.0043. [DOI] [PubMed] [Google Scholar]
  • 42.Action to Control Cardiovascular Risk in Diabetes Study Group. Gerstein HC, Miller ME, Byington RP, Goff DC, Jr, Bigger JT, Buse JB, Cushman WC, Genuth S, Ismail-Beigi F, Grimm RH, Jr, Probstfield JL, Simons-Morton DG, Friedewald WT. Effects of intensive glucose lowering in type 2 diabetes. N Engl J Med. 2008;358(24):2545–2559. doi: 10.1056/NEJMoa0802743. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.ADVANCE Collaborative Group. Patel A, MacMahon S, Chalmers J, Neal B, Billot L, Woodward M, Marre M, Cooper M, Glasziou P, Grobbee D, Hamet P, Harrap S, Heller S, Liu L, Mancia G, Mogensen CE, Pan C, Poulter N, Rodgers A, Williams B, Bompoint S, de Galan BE, Joshi R, Travert F. Intensive blood glucose control and vascular outcomes in patients with type 2 diabetes. N Engl J Med. 2008;358(24):2560–2572. doi: 10.1056/NEJMoa0802987. [DOI] [PubMed] [Google Scholar]
  • 44.Van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, Vlasselaers D, Ferdinande P, Lauwers P, Bouillon R. Intensive insulin therapy in the critically ill patients. N Engl J Med. 2001;345(19):1359–1367. doi: 10.1056/NEJMoa011300. [DOI] [PubMed] [Google Scholar]
  • 45.Brunkhorst FM, Engel C, Bloos F, Meier-Hellmann A, Ragaller M, Weiler N, Moerer O, Gruendling M, Oppert M, Grond S, Olthoff D, Jaschinski U, John S, Rossaint R, Welte T, Schaefer M, Kern P, Kuhnt E, Kiehntopf M, Hartog C, Natanson C, Loeffler M, Reinhart K. Intensive insulin therapy and pentastarch resuscitation in severe sepsis. N Engl J Med. 2008;358(2):125–139. doi: 10.1056/NEJMoa070716. German Competence Network Sepsis (SepNet) [DOI] [PubMed] [Google Scholar]
  • 46.Preiser JC, Devos P. Clinical experience with tight glucose control by intensive insulin therapy. Crit Care Med. 2007;35(9 Suppl):S503–507. doi: 10.1097/01.CCM.0000278046.24345.C7. [DOI] [PubMed] [Google Scholar]
  • 47.NICE-SUGAR Study Investigators. Finfer S, Chittock DR, Su SY, Blair D, Foster D, Dhingra V, Bellomo R, Cook D, Dodek P, Henderson WR, Hébert PC, Heritier S, Heyland DK, McArthur C, McDonald E, Mitchell I, Myburgh JA, Norton R, Potter J, Robinson BG, Ronco JJ. Intensive versus conventional glucose control in critically ill patients. N Engl J Med. 2009;360(13):1283–1297. doi: 10.1056/NEJMoa0810625. [DOI] [PubMed] [Google Scholar]
  • 48.Wiener RS, Wiener DC, Larson RJ. Benefits and risks of tight glucose control in critically ill adults: a meta-analysis. JAMA. 2008;300(8):933–944. doi: 10.1001/jama.300.8.933. [DOI] [PubMed] [Google Scholar]
  • 49.Griesdale DE, de Souza RJ, van Dam RM, Heyland DK, Cook DJ, Malhotra A, Dhaliwal R, Henderson WR, Chittock DR, Finfer S, Talmor D. Intensive insulin therapy and mortality among critically ill patients: a meta-analysis including NICE-SUGAR study data. CMAJ. 2009;180(8):821–827. doi: 10.1503/cmaj.090206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Scott MG, Bruns DE, Boyd JC, Sacks DB. Tight glucose control in the intensive care unit: are glucose meters up to the task? Clin Chem. 2009;55(1):18–20. doi: 10.1373/clinchem.2008.117291. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES