Abstract
Background
The alumina-on-alumina bearing surface, which has a high wear resistance and a good biocompatibility, is widely used in THA but recently has been associated with squeaking. While various authors have reported factors associated with squeaking, they remain poorly understood.
Questions/purposes
To contribute to the debate on squeaking we therefore asked the following questions: (1) What is the incidence of squeaking in alumina-on-alumina THA? (2) What factors are associated with squeaking in alumina bearings in our practice?
Methods
We retrospectively reviewed 168 patients (173 hips) who had primary alumina-on-alumina THAs. The mean age of the patients was 53 years (range, 18 to 81 years). Minimum followup was 5.6 years (average, 7.3 years; range, 5.6–9.4 years). All patients were evaluated clinically and radiographically with attention to periprosthetic osteolysis, squeaking, and ceramic fracture. When the patient reported squeaking, we determined the onset, reproducibility, and activities associated with the squeaking. We recorded patient (gender, age, height, weight, and body mass index) and surgical factors (abduction angle of cup, size and length of ceramic head component, and diameter of cup in the implant).
Results
Eight of the 168 patients (5%) had squeaking hips. Squeaking was more common in males and in those with large ceramic heads. There were no complications or revisions in the squeaking group. One ceramic liner fracture was associated with trochanteric nonunion.
Conclusions
When recommending alumina-on-alumina bearing surfaces to patients they should be clearly informed of the possibility of squeaking. Patients with risk factors for squeaking should be followed at regular intervals.
Level of Evidence
Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.
Introduction
Periprosthetic osteolysis in total hip arthroplasty (THA) owing to wear debris is the major threat to the long-term survival of THA [3, 22]. Metal-on-polyethylene bearing surfaces have been the mainstays of THA, but wear debris from polyethylene may cause periprosthetic osteolysis and eventually component loosening may occur. To minimize wear debris, alternative bearing surfaces with improved tribologic properties have been developed and, among them, alumina-on-alumina bearing surfaces, which have been used since the 1970s [19, 20]. Alumina-on-alumina bearings reportedly decrease inflammation, osteolysis, and loosening related to polyethylene wear debris [7]. However, early-generation alumina-on-alumina bearings had relatively low mechanical strength of the ceramic material [33] with associated with ceramic fracture [9, 18, 27], accelerated ceramic wear [32], and acetabular component loosening [29]. The long-term survivorship for early-generation alumina-on-alumina bearings ranged from 45% to 68.3% at 18 years [16, 19].
Many improvements have been made in the manufacture and design of ceramic implants since then [4]. The newer generation of ceramics have been associated with a fracture rate of less than 0.004% and wear rate in vitro less than 0.001 mm per year. In addition, the short term and intermediate term studies show high levels of function [5, 12, 23, 25]. Despite these improvements, the potential for ceramic fracture still remains as one of the major concerns. More recent studies have reported ceramic fracture rates range from 0.5% to 3.2% [10, 18, 27, 37, 40]. Moreover, before 2005, the squeaking of ceramic bearings was largely ignored and interpreted as only a minor complication [39]. However, recently many authors [1, 10, 11, 17, 24, 26, 31, 34, 36, 38, 39] have reported squeaking with an incidence ranging from 0.3% to 20.9% [2, 24, 26, 30, 38]. One study [36] suggests malpositioning is a factor, while another [39] showed squeaking hips were more common in taller, heavier, and younger patients and yet another [31] reported a difference in height between patients with squeaking and nonsqueaking hips.
To contribute to the debate on squeaking we therefore asked the following two questions: (1) What is the incidence of squeaking in alumina-on-alumina THA? (2) What demographic and implant-related factors are associated with squeaking in ceramic bearings in our practice?
Patients and Methods
From March 2000 to February 2004, we performed 259 THAs of various sorts in 227 patients. The alumina-on-alumina bearing THAs were performed in patients younger than 65 years old or in patients regardless of age who wanted THA using alumina-on-alumina bearings. During the study period, 168 patients (192 hips) received cementless THA using alumina-on-alumina bearing surfaces. The mean age of the patients was 53 years (range, 18 to 81 years). Among the 168 patients, 17 patients (17 hips, 8.8%) were lost to followup because they were not seen at a clinical followup from 3 years to 5 years postoperatively and could not be reached. Two deaths occurred in 2 years and 3 years after surgery, respectively, and were unrelated to THA. Therefore, 149 patients (173 hips) had at least 5 years of followup. The study group included 53 men and 96 women. The most common reason for THA was rheumatoid arthritis (Table 1). The minimum followup was 5.6 years (average, 7.3 years; range, 5.6–9.4 years).
Table 1.
Diagnosis | Number of hips |
---|---|
Rheumatoid arthritis of the hips | 75 |
Osteoarthritis of the hips | 44 |
Osteonecrosis of the femoral head | 40 |
Ankylosing spondylitis | 8 |
Miscellaneous conditions | 6 |
We used two acetabular components: 64 Secur-Fit® hemispherical acetabular PSL cups (Stryker Ireland Ltd, Limerick, Ireland)) and 109 ABG II® cups (Stryker Ireland). The femoral components were 173 ABG II® femoral stems. The acetabular and femoral components were cementless. The ceramic insert was Biolox® Forte ceramic insert (Stryker Ireland) with a metal-backed socket. We used the ceramic V40TM femoral head (Stryker Ireland) composed of two head diameters (28 mm, 32 mm) and three modular component lengths (short [−4 mm], medium [0 mm], long [+4 mm]). The 28-mm-diameter head was used in 135 hips and the 32-mm-diameter head in 38 hips. A short-neck modular femoral head component was used in seven hips, a medium-neck modular femoral head component in 131 hips, and a long-neck modular femoral head component in 35 hips.
Surgery was performed by a single surgeon (IYC) through a lateral approach with a trochanteric osteotomy in all hips. To reattach the greater trochanter, the stainless steel wires were inserted crosswise through the femur and then inserted through holes in the greater trochanter. The tightening of wires and tying of square knots with Kirschner wire bows were performed twice. The rigidity of greater trochanter fixation was confirmed by moving the greater trochanter with the towel clip.
For postoperative rehabilitation, patients were instructed in quadriceps setting exercises immediately after surgery, and active range of motion exercise of hips and knees was encouraged as soon as possible. Patients were allowed to walk using two crutches after 3 days with partial weight bearing and full weight bearing was allowed 6 weeks postoperatively.
The clinical and radiographic evaluations were performed at 6 weeks, at 3 months, at 5 months, 1 year, and annually thereafter. Two of us (YSK, KTH) who did not take part in the surgery independently evaluated the patients clinically and radiographically. The Harris hip score [21] was checked preoperatively, 6 weeks postoperatively, and at latest followups. Thigh pain and inguinal pain were also evaluated. We retrospectively reviewed medical records and radiographs for the information. If they were too limited, we tried to get more information from outpatients by periodic followup and telephone contact.
Two of us (YSK, KTH) evaluated separately all hip AP and crosstable lateral radiographs taken postoperatively and at the latest followup. The abduction angle of the cup was measured by the angle formed by the horizontal line drawn through the teardrops and the plane of opening of the cup [8]. The stable fixation of the femoral component was defined as the absence of the reactive line adjacent to the porous-surfaced portion of the implant and the presence of spot welds of endosteal new bone contacting the porous surfaces according to the method of Engh et al. [15]. For evaluation of fixation of the acetabular component, fixation was considered stable if there was no radiolucent line surrounding three zones as described by DeLee and Charnley [14] and no positional change of acetabular component according to the method of Latimer and Lachiewicz [28]. Periprosthetic cystic or scalloped lesions with a diameter of more than 2 mm that were not present on the immediate postoperative radiograph were defined as periprosthetic osteolysis.
To determine the causes of squeaking, patients were asked: “After surgery, have you ever heard or felt any noise when moving?” If a patient said “yes”, he/she was asked: “Can you reproduce the noise?” or “Can you describe the noise?” Noises were defined as either “squeak”, “click”, “grind”, or “pop” [24]. When the patient reported squeaking, we asked further questions regarding onset, reproducibility, and activities associated with the squeaking. We additionally sought to find related factors, including gender, age, height, weight, and body mass index (BMI) among patient factors and abduction angle of the cup, size and length of the ceramic head component, diameter of cup, and head-cup diameter ratio among implant and surgical factors.
We used an independent-sample Student’s t test to compare age, height, weight, BMI, abduction angle of cup, diameter of cup, and head-cup diameter ratio between patients without and with squeaking. Differences in gender, diameter of head, and neck length of head were compared using chi square test.
Results
Incidence of noise from the hip was observed in 19 of 173 hips (11%) (Table 2). Specifically, eight hips (5%) squeaked and 11 hips (6%) clicked. There were no cases of grind and pop. The squeaking began between 8 and 17 months postoperatively among patients with squeaking hips. One patient could reproduce the squeak in the clinical setting and squeaking occurred in stair climbing. Others complained squeaking occurred incidentally when they walked. However, no patient complained of pain during the squeak.
Table 2.
Parameter | Nonsqueaking | Squeaking | P value |
---|---|---|---|
Number of hips | 165 | 8 | |
Gender | 0.015 | ||
Male | 50 | 6 | |
Female | 115 | 2 | |
Age (years)* | 52.8 | 46 | 0.232 |
Height (cm)* | 162.8 | 161.4 | 0.951 |
Weight (kg)* | 60.2 | 60.9 | 0.943 |
Body mass index (kg/m2)* | 22.8 | 22.9 | 0.956 |
Abduction angle of cup (°)* | 48.3 | 50.5 | 0.360 |
Cup size (mm)* | 52.7 | 52.4 | 0.835 |
Size of ceramic head component | 0.002 | ||
28 mm | 133 | 2 | |
32 mm | 32 | 6 | |
Head-cup diameter ratio* | 0.550 | 0.552 | 0.937 |
Neck length of ceramic head component | |||
Short | 7 | 0 | 0.690 |
Medium | 124 | 7 | |
Long | 34 | 1 |
* Values are expressed as means.
Gender was associated with squeaking (Table 2). Squeaking was higher (p = 0.015) in the hips of male patients (six males and two females in the squeaking group and 50 males and 115 females in the nonsqueaking group). Age, height, weight, and BMI were not associated with squeaking. Among the implant factors, only head size was related to squeaking. Squeaking was more frequent (p = 0.002) in hips with a 32-mm ceramic head than in those with 28-mm ceramic heads (Table 2). We found no differences in abduction angle of the acetabular component, length of ceramic head, or head-cup diameter ratio between patients with squeaking and those without.
The mean Harris hip score improved from 47 points (range, 22–75 points) preoperatively to 94 points (range, 77–98 points) at the time of the latest followup. There was thigh pain in nine hips and inguinal pain in eight hips, none of which were associated with squeaking. All of the patients experiencing pain took NSAIDs for 8 weeks and their pain was relieved. Radiographically, there was no evidence of osteolysis and loosening in the acetabular and femoral components. No subsidence of the femoral stem was observed.
There were no complications in the squeaking group and no revision was performed owing to squeaking. Specifically, there was one ceramic liner fracture at 5-year followup. The 56-year-old patient with rheumatoid arthritis had undergone a primary ceramic-on-ceramic THA with 32-mm-diameter and medium-length ceramic head. Three months after surgery, greater trochanteric nonunion was observed, but no treatment was given because the nonunion gap was less than 2 cm and there was no pain and limping when walking. Five years after surgery, she noticed a grinding sensation in her hip with instability and discomfort. The radiograph was checked and a fracture of the ceramic insert was found. Revision THA was then performed (Fig. 1). There was one case of dislocation. The patient fell 3 months after surgery and dislocation occurred. There was one case of intraoperative periprosthetic fracture treated with open reduction and internal fixation, and subsequently the fracture healed. There was one acute infection treated with intravenous first-generation cephalosporin for 6 weeks. During 6-year followup, there was no recurrence of infection. Trochanteric nonunion occurred in four hips. The migration of the greater trochanter was less than 2 cm and no impairment of walking abilities was observed.
Discussion
The alumina-on-alumina THA has been used since the 1970s and reportedly has low wear rate and low incidence of osteolysis [4, 10, 12, 13, 20, 23, 25]. It is currently used in young and active adults and the intermediate-term studies report excellent results [4, 6, 12, 40]. However, during the last 5 years, a squeaking issue has been documented, but the causes of squeaking are not understood [24, 35, 38, 39]. Patients receiving alumina-on-alumina THA at our institute began to complain of squeaking. This prompted us to observe the type of noise and the incidence of squeaking in order to determine its associated factors.
The limitations of our study are as follows. First, we followed the open question procedure in Mai et al. [31]; however, the criteria for accurately reporting a squeaking hip by the patients are unclear. We believed it preferable to the risk of underestimation or overestimation that can result from prompting. Standardization procedures would aid future research. We used open-ended questions and all patients who could reproduce a sound similar to squeaking were included in the squeaking group, which would reduce the number of neglected squeaking hips. Second, we had a small number of patients with squeaking. This limits the power of the study and increases the risk of a Type II error in presuming some factors are not associated with squeaking; that is, some of these factors might be associated with squeaking with greater numbers of patients. Third, we identified only one ceramic liner fracture precluding any generalization regarding the cause of that complication.
The incidence of squeaking in alumina-on-alumina THA in our study was 5% (eight of 173 hips). The reported prevalence of squeaking ranges from 0.3% to 20.9% (Table 3). Keurentjes et al. [26] observed a high rate of squeaking and stated that squeaking was likely to be a phenomenon indicative of potential complications that were possibly underreported, causing them to discontinue implanting this type of bearing. In contrast, Mai et al. [31], despite a high percentage of squeaking in their patients, concluded that most squeaking hips were pain-free and functioning well, and patients were generally satisfied with the outcome. They concluded that ceramic bearings are an acceptable treatment for young and active patients. In the case of our study, none of the squeakers reported pain or discomfort. As we were not able to find any clear evidence of complications in patients in squeaking group, following Mai et al. [31], we continue to use alumina-on-alumina THA. When this was explained, no patients objected to this procedure.
Table 3.
Study | Number of hips | Incidence (%) | Factors |
---|---|---|---|
Lusty et al. [30] (2007) | 301 | 0.3 | |
Walter et al. [38] (2007) | 2716 | 0.6 | Age, height, weight, position of cup |
Capello et al. [10] (2008) | 475 | 0.8 | |
Mai et al. [31] (2010) | 336 | 10 | Femoral neck geometry |
Jarrett et al. [24] (2009) | 149 | 10.7 | |
Keurentjes et al. [26] (2008) | 43 | 20.9 | Short neck length of femoral component |
Choi et al. [current study] | 173 | 5 | Gender, head size |
Our data suggest males are more likely to have a squeaking hip but we could not find a gender difference in the other studies reported previously. Walter et al. [39] demonstrated squeaking hips were more common in taller, heavier, and younger patients. Mai et al. [31] reported a difference in height between patients with squeaking and nonsqueaking hips. However, we did find differences in gender. We speculate the male predominance in squeaking could be the result of gender differences in physical activity since we assume men in the younger age group tend to be more physically active than women. The bearings in more active patients would increase mechanical demand and increase the possibility of squeaking [38].
Squeaking occurred more frequently in larger ceramic heads in our study. Mai et al. [31] reported no differences in head size in patients who squeaked and those who did not. We cannot be sure of the reasons for this discrepancy. However, larger heads could have increased ROM. Any increased ROM may lead to bony impingement at extreme ROM whereby the femur is levering against the pelvis, causing subluxation of the femoral head and edge loading, leading to stripe wear and squeak [31]. Larger heads also have a larger articulated surface which increases by the square of the head radius. Therefore, squeaking might be more likely if surface area is related to squeaking. We found no difference in abduction angle and size of cup between the squeaking and nonsqueaking groups, which supports the findings of Keurentjes et al. [26] and Restrepo et al. [36] who reported no difference in component orientation between squeaking and nonsqueaking groups.
In conclusion, we found that only male hips and large head size were more likely to be associated with squeaking. However, there was no evidence that squeaking led to complications. As a result, we continue to use an alumina-on-alumina bearing surface THA, taking care to explain the possibility of squeaking to patients. The possibility of squeaking should be clearly explained to patients who are undergoing THA using alumina-on-alumina bearing surfaces. The patient with related risk factors for squeaking should be followed up carefully at regular intervals.
Acknowledgments
We thank Professor Adam Turner, Hanyang University, for his help in revising our manuscript.
Footnotes
Each author certifies that he or she has no commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc) that might pose a conflict of interest in connection with the submitted article.
Each author certifies that his or her institution approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.
This work was performed at Hanyang University Hospital, Seoul, South Korea.
References
- 1.Affatato S, Traina F, Mazzega-Fabbro C, Sergo V, Viceconti M. Is ceramic-on-ceramic squeaking phenomenon reproducible in vitro? A long-term simulator study under severe conditions. J Biomed Mater Res B Appl Biomater. 2009;91:264–271. doi: 10.1002/jbm.b.31398. [DOI] [PubMed] [Google Scholar]
- 2.Baek SH, Kim SY. Cementless total hip arthroplasty with alumina bearings in patients younger than fifty with femoral head osteonecrosis. J Bone Joint Surg Am. 2008;90:1314–1320. doi: 10.2106/JBJS.G.00755. [DOI] [PubMed] [Google Scholar]
- 3.Berry DJ. Management of osteolysis around total hip arthroplasty. Orthopedics. 1999;22:805–808. doi: 10.3928/0147-7447-19990901-06. [DOI] [PubMed] [Google Scholar]
- 4.Bierbaum BE, Nairus J, Kuesis D, Morrison JC, Ward D. Ceramic-on-ceramic bearings in total hip arthroplasty. Clin Orthop Relat Res. 2002;405:158–163. doi: 10.1097/00003086-200212000-00019. [DOI] [PubMed] [Google Scholar]
- 5.Bizot P, Hannouche D, Nizard R, Witvoet J, Sedel L. Hybrid alumina total hip arthroplasty using a press-fit metal-backed socket in patients younger than 55 years. A six- to 11-year evaluation. J Bone Joint Surg Br. 2004;86:190–194. doi: 10.1302/0301-620X.86B2.14026. [DOI] [PubMed] [Google Scholar]
- 6.Bizot P, Larrouy M, Witvoet J, Sedel L, Nizard R. Press-fit metal-backed alumina sockets: a minimum 5-year followup study. Clin Orthop Relat Res. 2000:134–142. [DOI] [PubMed]
- 7.Bizot P, Nizard R, Hamadouche M, Hannouche D, Sedel L. Prevention of wear and osteolysis: alumina-on-alumina bearing. Clin Orthop Relat Res. 2001;393:85–93. doi: 10.1097/00003086-200112000-00010. [DOI] [PubMed] [Google Scholar]
- 8.Callaghan JJ, Salvati EA, Pellicci PM, Wilson PD, Ranawat CS. Results of revision for mechanical failure after cemented total hip-replacement, 1979 to 1982—a 2 to 5-year follow-up. J Bone Joint Surg Am. 1985;67:1074–1085. [PubMed] [Google Scholar]
- 9.Callaway GH, Flynn W, Ranawat CS, Sculco TP. Fracture of the femoral head after ceramic-on-polyethylene total hip arthroplasty. J Arthroplasty. 1995;10:855–859. doi: 10.1016/S0883-5403(05)80087-7. [DOI] [PubMed] [Google Scholar]
- 10.Capello WN, D’Antonio JA, Feinberg JR, Manley MT, Naughton M. Ceramic-on-ceramic total hip arthroplasty: update. J Arthroplasty. 2008;23:39–43. doi: 10.1016/j.arth.2008.06.003. [DOI] [PubMed] [Google Scholar]
- 11.Chevillotte C, Trousdale RT, Chen Q, Guyen O, An KN. The 2009 Frank Stinchfield Award: “Hip squeaking”: A biomechanical study of ceramic-on-ceramic bearing surfaces. Clin Orthop Relat Res. 2010;468:345–350. doi: 10.1007/s11999-009-0911-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.D’Antonio J, Capello W, Manley M, Bierbaum B. New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty. J Arthroplasty. 2002;17:390–397. doi: 10.1054/arth.2002.32183. [DOI] [PubMed] [Google Scholar]
- 13.D’Antonio J, Capello W, Manley M, Naughton M, Sutton K. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436:164–171. doi: 10.1097/01.blo.0000162995.50971.39. [DOI] [PubMed] [Google Scholar]
- 14.DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32. [PubMed] [Google Scholar]
- 15.Engh CA, Massin P, Suthers KE. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res. 1990;257:107–128. [PubMed] [Google Scholar]
- 16.Fenollosa J, Seminario P, Montijano C. Ceramic hip prostheses in young patients: a retrospective study of 74 patients. Clin Orthop Relat Res. 2000;379:55–67. doi: 10.1097/00003086-200010000-00009. [DOI] [PubMed] [Google Scholar]
- 17.Glaser D, Komistek RD, Cates HE, Mahfouz MR. Clicking and squeaking: in vivo correlation of sound and separation for different bearing surfaces. J Bone Joint Surg Am. 2008;90(Suppl 4):112–120. doi: 10.2106/JBJS.H.00627. [DOI] [PubMed] [Google Scholar]
- 18.Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH. Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res. 2007;458:106–110. doi: 10.1097/BLO.0b013e3180303e87. [DOI] [PubMed] [Google Scholar]
- 19.Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg Am. 2002;84:69–77. [PubMed] [Google Scholar]
- 20.Hannouche D, Hamadouche M, Nizard R, Bizot P, Meunier A, Sedel L. Ceramics in total hip replacement. Clin Orthop Relat Res. 2005;430:62–71. doi: 10.1097/01.blo.0000149996.91974.83. [DOI] [PubMed] [Google Scholar]
- 21.Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969;51:737–755. [PubMed] [Google Scholar]
- 22.Harris WH. Wear and periprosthetic osteolysis: the problem. Clin Orthop Relat Res. 2001;393:66–70. doi: 10.1097/00003086-200112000-00007. [DOI] [PubMed] [Google Scholar]
- 23.Hernigou P, Zilber S, Filippini P, Poignard A. Ceramic-ceramic bearing decreases osteolysis: a 20-year study versus ceramic-polyethylene on the contralateral hip. Clin Orthop Relat Res. 2009;467:2274–2280. doi: 10.1007/s11999-009-0773-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Jarrett CA, Ranawat AS, Bruzzone M, Blum YC, Rodriguez JA, Ranawat CS. The squeaking hip: a phenomenon of ceramic-on-ceramic total hip arthroplasty. J Bone Joint Surg Am. 2009;91:1344–1349. doi: 10.2106/JBJS.F.00970. [DOI] [PubMed] [Google Scholar]
- 25.Jazrawi LM, Bogner E, Della Valle CJ, Chen FS, Pak KI, Stuchin SA, Frankel VH, Di Cesare PE. Wear rates of ceramic-on-ceramic bearing surfaces in total hip implants: a 12-year follow-up study. J Arthroplasty. 1999;14:781–787. doi: 10.1016/S0883-5403(99)90025-6. [DOI] [PubMed] [Google Scholar]
- 26.Keurentjes JC, Kuipers RM, Wever DJ, Schreurs BW. High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings. Clin Orthop Relat Res. 2008;466:1438–1443. doi: 10.1007/s11999-008-0177-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Koo KH, Ha YC, Jung WH, Kim SR, Yoo JJ, Kim HJ. Isolated fracture of the ceramic head after third-generation alumina-on-alumina total hip arthroplasty. J Bone Joint Surg Am. 2008;90:329–336. doi: 10.2106/JBJS.F.01489. [DOI] [PubMed] [Google Scholar]
- 28.Latimer HA, Lachiewicz PF. Porous-coated acetabular components with screw fixation. Five to ten-year results. J Bone Joint Surg Am. 1996;78:975–981. doi: 10.2106/00004623-199607000-00001. [DOI] [PubMed] [Google Scholar]
- 29.Lerouge S, Huk O, Yahia L, Witvoet J, Sedel L. Ceramic-ceramic and metal-polyethylene total hip replacements: comparison of pseudomembranes after loosening. J Bone Joint Surg Br. 1997;79:135–139. doi: 10.1302/0301-620X.79B1.6621. [DOI] [PubMed] [Google Scholar]
- 30.Lusty PJ, Tai CC, Sew-Hoy RP, Walter WL, Walter WK, Zicat BA. Third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2676–2683. doi: 10.2106/JBJS.F.01466. [DOI] [PubMed] [Google Scholar]
- 31.Mai K, Verioti C, Ezzet KA, Copp SN, Walker RH, Colwell CW. Incidence of ‘squeaking’ after ceramic-on-ceramic total hip arthroplasty. Clin Orthop Relat Res. 2010;468:413–417. doi: 10.1007/s11999-009-1083-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Morlock M, Nassutt R, Janssen R, Willmann G, Honl M. Mismatched wear couple zirconium oxide and aluminum oxide in total hip arthroplasty. J Arthroplasty. 2001;16:1071–1074. doi: 10.1054/arth.2001.27233. [DOI] [PubMed] [Google Scholar]
- 33.Nich C, Sariali el H, Hannouche D, Nizard R, Witvoet J, Sedel L, Bizot P. Long-term results of alumina-on-alumina hip arthroplasty for osteonecrosis. Clin Orthop Relat Res. 2003;417:102–111. doi: 10.1097/01.blo.0000096820.67494.bf. [DOI] [PubMed] [Google Scholar]
- 34.Ranawat AS, Ranawat CS. The squeaking hip: a cause for concern-agrees. Orthopedics. 2007;30:738–743. doi: 10.3928/01477447-20070901-32. [DOI] [PubMed] [Google Scholar]
- 35.Restrepo C, Lettich T, Roberts N, Parvizi J, Hozack WJ. Uncemented total hip arthroplasty in patients less than twenty-years. Acta Orthop Belg. 2008;74:615–622. [PubMed] [Google Scholar]
- 36.Restrepo C, Parvizi J, Kurtz SM, Sharkey PF, Hozack WJ, Rothman RH. The noisy ceramic hip: is component malpositioning the cause? J Arthroplasty. 2008;23:643–649. doi: 10.1016/j.arth.2008.04.001. [DOI] [PubMed] [Google Scholar]
- 37.Toni A, Traina F, Stea S, Sudanese A, Visentin M, Bordini B, Squarzoni S. Early diagnosis of ceramic liner fracture. Guidelines based on a twelve-year clinical experience. J Bone Joint Surg Am. 2006;88(Suppl 4):55–63. doi: 10.2106/JBJS.F.00587. [DOI] [PubMed] [Google Scholar]
- 38.Walter WL, O’Toole GC, Walter WK, Ellis A, Zicat BA. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22:496–503. doi: 10.1016/j.arth.2006.06.018. [DOI] [PubMed] [Google Scholar]
- 39.Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, Hozack WJ, Tuke MA. Squeaking hips. J Bone Joint Surg Am. 2008;90(Suppl 4):102–111. doi: 10.2106/JBJS.H.00867. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ. Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study. J Bone Joint Surg Am. 2005;87:530–535. doi: 10.2106/JBJS.D.01753. [DOI] [PubMed] [Google Scholar]