Clinical applications of breath testing

Kelly M Paschke 1, Alquam Mashir 1, Raed A Dweik 1,2,
PMCID: PMC2990505  PMID: 21173863

Abstract

Breath testing has the potential to benefit the medical field as a cost-effective, non-invasive diagnostic tool for diseases of the lung and beyond. With growing evidence of clinical worth, standardization of methods, and new sensor and detection technologies the stage is set for breath testing to gain considerable attention and wider application in upcoming years.

Introduction and context

With each breath exhaled thousands of molecules are expelled, providing a window into the physiological state of the body. The utilization of breath as a medical test has been reported for centuries as demonstrated by Hippocrates in his description of fetor oris and fetor hepaticus in his treatise on breath aroma and disease [1]. Even in modern times clinicians have noted distinct changes in the breath odor of patients with specific diseases such as diabetes, renal failure, and hepatic diseases [2-4]. However, it was Linus Pauling’s milestone discovery of 250 unique substances present in exhaled breath that offered promising insight into breath testing [5]. Since this discovery, breath analysis has rapidly evolved as a new frontier in medical testing for disease states in the lung and beyond [1]. Breath analysis is now used clinically to monitor asthma, diagnose transplant organ rejection, diagnose Helicobacter pylori infection, detect blood alcohol concentration, and monitor breath gases during anesthesia, mechanical ventilation, and respiration, among numerous other applications [1,6,7].

Recent advances

Breath analysis may offer a relatively inexpensive, rapid, and non-invasive method for detecting a variety of diseases. With recent advancements in mass spectrometry (MS) and gas chromatography MS (GC-MS), it is possible to identify thousands of unique substances, such as volatile organic compounds (VOCs) and elemental gases, in the breath [8]. Improved technologies such as selected-ion flow-tube MS (SIFT-MS), multi-capillary column ion mobility MS (MCC-IMS), and proton transfer reaction MS (PTR-MS) have provided real time, precise identification of trace gases in human breath in the parts per trillion range [9-11]. On the other hand, unlike traditional quantitative breath analysis, the electronic nose is essentially trained to recognize odor patterns using an array of gas sensors. The electronic nose has shown accuracy in the detection of lung cancer, pneumonia, and asthma with specificities and sensitivities ranging from 74-98%, as well as in the discrimination between diseases such as chronic obstructive pulmonary disease and asthma [12-15]. Table 1 provides a selected list of the growing number of technologies being applied to breath testing.

Table 1. Current breath-based test technologies.

Spectrometry Gas chromatography Other
Mid-infrared absorption spectroscopy Automated thermal desorption gas chromatography mass spectrometry (ATD-GC-MS) Chemiluminescense
Multi pass cell-laser absorption spectroscopy Gas chromatography flame ionization detection (GC-FID) Electrochemical cell sensor technology
Tunable diode laser absorption spectroscopy (TDLAS) Gas chromatrography mass spectrometry (GC-MS) Gas sensor array technology
Cavity ring-down spectroscopy (CRDS) Photoionization detection gas chromatography Nanosensor technology
Cavity leak-out spectroscopy (CALOS) Solid phase microextraction gas chromatography mass spectrometry (SPME-GC-MS) Infrared and para-magnetic sensors
Cavity enhanced optical frequency comb spectroscopy Gas chromatography time of flight mass spectrometry (GC-TOF-MS) Calorimetry
Integrated cavity output spectroscopy (ICOS) Gas chromatography differential mobility spectrometry (GC-DMS) Sol-gel sensor technology
Laser magnetic resonance spectroscopy (LMRS) Gas chromatography ion mobility mass spectrometry (GC-IMS-MS) Laser based sensor technology
Laser photoacoustic spectroscopy Gas chromatography/Fourier transform ion cyclotron resonance mass spectrometry (GC/FT-ICR-MS)
Faraday-LMRS Gas-liquid partition chromatography (GLPC)
Selected ion flow tube mass spectrometry (SIFT-MS) Gas chromatography electrolyzer-powered flame ionization detector (GC-EFID)
Proton transfer reaction mass spectrometry (PTR-MS)
Faraday modulation spectroscopy
Ion trap (2D) and (3D) mass spectrometry
Time of flight mass spectrometry (TOF-MS)
Isotope ratio mass spectrometry (IR-MS)
Multi-capillary column ion mobility mass spectrometry (MCC-IMS)
High sensitivity (hs)-PTR-MS
Matrix assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF-MS)
Proton transfer reaction time of flight mass spectrometry (PTR-TOF-MS)

More recent technological advancements in breath analysis have moved beyond measuring volatiles in the gas phase into measurement of semivolatiles and compounds dissolved in aerosolized droplets in exhaled breath condensate (EBC) and in exhaled breath vapor (EBV). Aerosolized droplets in EBC can be captured by a variety of methods and analyzed for a wide range of biomarkers, such as metabolic end products, proteins, cytokines, and chemokines, with expanding possibilities [16,17]. With 3000 volatile compounds identifiable using EBC and twice the volatile metabolite concentration compared to traditional breath gas analysis, this application has the potential to provide superior information about breathprints of healthy and disease states [8,18]. EBV sampling has also yielded promising results as a new breath sampling method. EBV sampling pre-concentrates breath samples using a solid-phase microextraction fiber inserted into a modified RTube™, a common device also used in EBC sampling. This procedure provides the potential advantages of faster breath sampling and analysis, increased portability, minimal user training, use in contaminated environments, and no requirement for a power source. EBV sampling may yield additional compounds not detected in EBC and may provide greater sensitivity as a sampling method, expanding the spectrum of breath sampling [19].

Implications for clinical practice

The science of breath analysis is rapidly expanding, the technology is improving, and several new applications have been developed or are under commercial development. A major breakthrough over the past decade has been the increase in breath-based tests approved by the US Food and Drug Administration (FDA). Devices measuring common breath gases: oxygen, nitrogen, water vapor, and carbon dioxide in patient respiratory monitoring have served as a platform for technological growth in clinical breath testing applications. In particular, earlier devices, such as those providing the detection of blood alcohol concentration, H. pylori infection, lactose intolerance, and airway monitoring by end-tidal carbon dioxide, have demonstrated clinical benefits as well as diagnostic success in clinical breath testing. Table 2 provides a selected list of the breath-based tests currently approved by the FDA.

Table 2. Breath-based tests approved by the US Food and Drug Administration [33].

Molecule detected Disease/condition Trade name of analysis instrument Technology Manufacturer FDA approval date
CO2, O2, N2O Respiration Consolidated-Nier model 21-201 isotope ratio mass spectrometer Dual inlet system gas isotope ratio mass spectrometer Consolidated Electrodynamics Corporation, Inc., Pasadena, CA, USA Before 28 May 1976
CO2 Respiration Tidal Wave® Carbon Dioxide Monitor, Model 610 Sensor technology Novametrix Medical Systems, Inc., Wallingford, CT, USA 20 November 1996
H2 Lactose malabsorption Micro H2 Sensor technology MICRO DIRECT, Inc., Auburn, ME USA 24 January 1997
13C, 18O, CO2, 15N, N2, NO2 Respiration ABCA-NT System Continuous flow gas isotope ratio mass spectrometer Europa Scientific, Ltd. Concord, MA USA 16 December 1997
CO2, O2, N2O and anesthetic agents Respiration, anesthesia Datex-Ohmeda Compact Airway Module M-CAiOVX and M-COVX Infrared sensor, paramagnetic sensor Datex-Ohmeda, Inc., Tewksbury, MA, USA 23 August 2000
13CO2/12CO2 H. pylori UBiT®-IR3000 Infrared Spectrometry System Infrared (IR) spectrophotometer Otsuka Pharmaceutical Co. Ltd., Tokyo, Japan 21 December 2001
O2, CO2 Respiration BSM-4100A Sensor technology Nihon Kohden America Inc., Foothill Ranch, CA, USA 24 October 2000
13CO2/12CO2 H. pylori BreathTek™ - UBiT® UBT for Helicobacter pylori UBiT® IR spectrophotometer Meretek Diagnostics, Inc., Nashville, TN, USA 17 January 2002
O2, CO2, N2O, anesthetic agents Respiration, anesthesia Ag-920PA Sensor technology Nihon Kohden America Inc., Foothill Ranch, CA, USA 25 July 2002
O2, CO2, N2O, anesthetic agents and gases Respiration, anesthesia BSM-5130A Series Bedside Monitor Sensor technology Nihon Kohden America, Inc., Foothill Ranch, CA, USA March 04, 2003
NO Asthma, airway inflammation NIOX® Chemiluminescence Aerocrine AB, Solna, Sweden 30 April 2003
CO2 Respiration, anesthesia Datex-Ohmeda S/5 Single-Width Airway Module M-miniC MiniCO2 IR measuring sensor Datex-Ohmeda, Needham, MA, USA 23 April 2003
(C4-C20) alkanes, monomethylalkanes Grade 3 heart transplant rejection Heartsbreath Gas chromatography mass spectrometry Menssana Research, Inc., Fort Lee, NJ, USA 24 February 2004
H2 Lactose malabsorption Micro H2 Breath Monitoring Device with HYDRA Software Utility Electrochemical gas sensor Micro Medical Ltd., Kent, UK 19 May 2004
13CO2/12CO2 H. pylori POCone Infrared Spectrophotometer IR spectrophotometer Otsuka Pharmaceutical Co. Ltd., Tokyo, Japan 15 July 2004
Alcohol Breath alcohol AlcoMate CA2000 Digital Alcohol Detector Semiconductor oxide sensor KHN Solutions LLC, San Francisco, CA, USA 11 August 2004
Alcohol Breath alcohol AlcoHAWK Precision™ Digital Alcohol Detector Semiconductor oxide sensor Q3 Innovations, LLC, Eagan, MN, USA 9 February 2005
CO2 Ventilation C-CO2™ Colorimetric carbon dioxide sensor Marquest Medical Products, Inc., Englewood, CO, USA 1 March 2005
CO2 Ventilation Datex-Ohmeda S/5™ Single-width airway module, E-miniC Narrow band IR sensor GE Healthcare, Needham, MA, USA 14 October 2005
Alcohol Breath alcohol AL-6000 Breath Alcohol Tester Semiconductor oxide sensor Sentech Korea Corp., Kyeonggi-do, Korea 11 May 2006
Alcohol Breath alcohol AL-5000 Breath Alcohol Tester Semi-conductive alcohol sensor Sentech Korea Corp., Kyeonggi-do, Korea 30 October 2006
Alcohol Breath alcohol Breath Alcohol .02 Detection System Electrochemical analyzer Akers Biosciences, Inc., Thorofare, NJ, USA 18 December 2006
CO2 Respiration OLG-2800A Sensor technology Nihon Kohden America, Inc., Foothill Ranch, CA, USA 27 December 2006
Alcohol Breath alcohol BACTRACK® Breath Analyzer Semiconductor oxide sensor KHN Solutions LLC, San Francisco, CA, USA 14 September 2007
CO2 Respiration, anesthesia EMMA™ Emergency Capnometer IR gas analysis Phasein AB, Danderyd, Sweden 28 December 2007
CO CO poisoning, carboxy-haemoglobin EC50 ToxCO+ Electrochemical gas sensor technology Bedfont Scientific Ltd., Rochester, Kent, UK 21 February 2008
NO Asthma, airway inflammation NIOX MINO® Electrochemical sensor Aerocrine AB, Solna, Sweden 3 March 2008
NO Asthma, airway inflammation Apieron Insight eNO Sol-gel-heme protein sensor Apieron, Inc., Menlo Park, CA, USA 14 March 2008
Alcohol Breath alcohol AlcoHAWK® PT500 Digital Alcohol Detector Fuel cell sensor Innovations, LLC, Independence, IA, USA 25 July 2008
CO2 Ventilation Nihon Kohden TG-970P Series CO2 Sensor Kit IR absorption spectrometry Nihon Kohden America, Inc., Foothill Ranch, CA, USA 2 March 2009
Alcohol Breath alcohol BACTRACK® Select Breathalyzer Model (S30, S50, S70) Semiconductor (Si) oxide sensor KHN Solutions LLC, San Francisco, CA, USA 19 March 2009
Alcohol Breath alcohol BACTRACK® Select Breathalyzer Model S80 Fuel cell electrochemical sensor KHN Solutions LLC, San Francisco, CA, USA 24 March 2009

One recent landmark in clinical breath testing occurred in 2003 when the FDA approved the first device that measures the fraction of exhaled nitric oxide (FENO) for asthma monitoring. The desktop NIOX® (currently NIOX® FLEX) was followed by a handheld NIOX® MINO device (both by Aerocrine, Inc., Solna, Sweden) that received FDA clearance in 2008. Advantages provided by FENO monitoring devices include its non-invasive nature, ease of repeat measurements, and use in adult and child populations with severe airflow obstruction where other techniques would be difficult or impossible to perform [20]. FDA approval of these devices has largely been attributed to the standardization of clinical FENO monitoring and detection via breath analysis [21]. In order for this simple yet powerful tool to achieve its potential, we need to further understand the roles that FENO and similar biomarkers of disease play in different clinical settings and across populations, and their specific functions in disease.

A recent clinical application of breath testing has been in the diagnosis of lung cancer. Currently, clinicians rely on relatively expensive and invasive diagnostic tests, such as computed tomography exams, chest radiography, sputum analysis, and lung biopsies, which remain largely ineffective in early stage lung cancer diagnosis. Researchers have demonstrated success using trained dogs in the breath diagnosis of both early and late stage lung cancers with sensitivities and specificities approaching 99%, providing promise for future lung cancer breath tests [22]. Breath testing may provide a promising alternative diagnostic tool for lung cancer as evidenced by numerous studies with specificities and sensitivities ranging from 71-94% [14,23-28]. However, in order to be useful as an upfront screening test for high-risk populations, as a tool to evaluate pulmonary nodules, or as a diagnostic test for lung cancer, a breath test should be at least 90-95% sensitive and specific [29].

As the field of breath research has developed over the past decade, the need for standardization in sampling has grown. Attempts at sampling only critical portions of exhaled breath have proven successful by using end-tidal sampling, as evidenced by finding VOC concentrations most reflective of compounds dissolved in the blood [30]. End-tidal sampling (collecting breath only at the end of exhalation) has shown success over mixed expiratory sampling (collecting the entire exhaled breath) because samples are less likely to be diluted by mixing with dead space volume (inspired air not taking place in gas exchange) and ambient air. A useful application is buffered end-tidal on-line sampling, which measures VOC breath concentrations over a large mass range quickly and uses multiple MS technologies, such as SIFT-MS and PTR-MS, for breath analysis [31]. It is also promising because it uses on-line sampling (the sampling device is connected to the analytical device) versus less accurate off-line sampling (the sample is collected and later brought to the analytical device using reservoirs such as Tedlar® bags). Device calibration and validation have helped by accounting for exogenous VOCs and ambient air contamination in the sampling environment [19]. Since detection of many VOCs occurs at the parts per billion and parts per trillion levels, it is essential to control for exogenous sources of VOCs because ingestion of certain foods, medications, gut bacterial flora, and exposure to chemicals and pollution, amongst many other things, will alter VOCs in exhaled breath [32]. It is important for researchers to consider the change in the concentration of several VOCs in disease states as well as the utility of ranking systems for VOC predictability and new methods for accounting for ambient VOC sources, such as calculating alveolar gradients [8,28]. Despite receiving considerable attention in recent years, issues with standardization have been a major limitation of clinical breath testing. This has been evidenced by difficulties in establishing baseline VOC concentrations and the wide range of results represented in the literature for VOC concentrations in disease. Thus, it is necessary in the future to search for innovative methods for breath research.

There are numerous potential advantages for breath analysis as a clinical test. The method is non-invasive (the sample is relatively easy and painless to acquire), the sample is likely to be rich with information (a single test can scan for signatures of many abnormalities or markers of disease), it has the potential for low cost, and lends itself to easy administration. The field of breath testing has grown tremendously in recent years and with evolving technologies in sampling, sensor design, standardization, and analytical methods breath analysis has the potential to clinically benefit individuals on a global scale in the future.

Acknowledgments

KMP, AM, and RAD are supported by a Third Frontier grant from the Ohio Department of Development. RAD is supported by a National Institutes of Health (NIH) grant (NIH-HL68863).

Abbreviations

EBC

exhaled breath condensate

EBV

exhaled breath vapor

FDA

US Food and Drug Administration

FENO

fraction of exhaled nitric oxide

GC-MS

gas chromatography MS

MCC-IMS

multi-capillary column ion mobility MS

MS

mass spectrometry

PTR-MS

proton transfer reaction MS

SIFT-MS

selected-ion flow-tube MS

VOC

volatile organic compound

Competing Interests

The authors declare that they have no competing interests.

The electronic version of this article is the complete one and can be found at: http://f1000.com/reports/m/content/2/56

References

  • 1.Dweik RA, Amann A. Exhaled breath analysis: the new frontier in medical testing. J Breath Res. 2008:030301. doi: 10.1088/1752-7163/2/3/030301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Manolis A. The diagnostic potential of breath analysis. Clin Chem. 1983;29:5–15. [PubMed] [Google Scholar]
  • 3.van den Velde S, Quirynen M, van Hee P, van Steenberghe D. Halitosis associated volatiles in breath of healthy subjects. J Chromatogr B Analyt Technol Biomed Life Sci. 2007;853:54–61. doi: 10.1016/j.jchromb.2007.02.048. [DOI] [PubMed] [Google Scholar]
  • 4.Di Francesco F, Fuoco R, Trivella M, Ceccarini A. Breath analysis: trends in techniques and clinical applications. Microchem J. 2005;79:405–10. doi: 10.1016/j.microc.2004.10.008. [DOI] [Google Scholar]
  • 5.Pauling L, Robinson AB, Teranishi R, Cary P. Quantitative analysis of urine vapor and breath by gas-liquid partition chromatography. Proc Natl Acad Sci U S A. 1971;68:2374–6. doi: 10.1073/pnas.68.10.2374. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Mashir A, Dweik R. Exhaled breath analysis: the new interface between medicine and engineering. Adv Powder Technol. 2008;20:420–25. doi: 10.1016/j.apt.2009.05.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Ward K, Yealy D. End-tidal carbon dioxide monitoring in emergency medicine, Part 2: Clinical applications. Acad Emerg Med. 1998;5:637–46. doi: 10.1111/j.1553-2712.1998.tb02474.x. [DOI] [PubMed] [Google Scholar]
  • 8.Phillips M, Herrera J, Krishnan S, Zain M, Greenberg J, Cataneo RN. Variation in volatile organic compounds in the breath of normal humans. J Chromatogr B Biomed Sci Appl. 1999;729:75–88. doi: 10.1016/S0378-4347(99)00127-9. [DOI] [PubMed] [Google Scholar]
  • 9.Ross BM. Sub-parts per billion detection of trace volatile chemicals in human breath using Selected Ion Flow Tube Mass Spectrometry. BMC Res Notes. 2008;1:41. doi: 10.1186/1756-0500-1-41. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Perl T, Carstens E, Hirn A, Quintel M, Vautz W, Nolte J, Junger M. Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry. Br J Anaesth. 2009;103:822–7. doi: 10.1093/bja/aep312. [DOI] [PubMed] [Google Scholar]
  • 11.Wehinger A, Schmid A, Mechtcheriakov S, Ledochowski M, Grabmer C, Gastl GA, Amann A. Lung cancer detection by proton transfer reaction mass-spectrometric analysis of human breath gas. Int J Mass Spectrom. 2007;265:49–59. doi: 10.1016/j.ijms.2007.05.012. [DOI] [Google Scholar]
  • 12.Fens N, Zwinderman AH, van der Schee MP, de Nijs SB, Dijkers E, Roldaan AC, Cheung D, Bel EH, Sterk PJ. Exhaled breath profiling enables discrimination of chronic obstructive pulmonary disease and asthma. Am J Respir Crit Care Med. 2009;180:1076–82. doi: 10.1164/rccm.200906-0939OC. [DOI] [PubMed] [Google Scholar]; F1000 Factor 6.0 Must ReadEvaluated by Ronald Sorkness 06 Jan 2010
  • 13.Taivans I, Jurka N, Balode L, Bukovskis M, Kopeika U, Ogorodnkis V, Kleperis J, Strazda G, Šiliņš V, Martinsons A. Exhaled air analysis in patients with different lung diseases using artificial odour sensors. Proc Latvian Acad Sci. 2009;63:210–13. [Google Scholar]
  • 14.Machado RF, Laskowski D, Deffenderfer O, Burch T, Zheng S, Mazzone PJ, Mekhail T, Jennings C, Stoller JK, Pyle J, Duncan J, Dweik RA, Erzurum SC. Detection of lung cancer by sensor array analyses of exhaled breath. Am J Respir Crit Care Med. 2005;171:1286–91. doi: 10.1164/rccm.200409-1184OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Dragonieri S, Annema JT, Schot R, van der Schee MP, Spanevello A, Carratu P, Resta O, Rabe KF, Sterk PJ. An electronic nose in the discrimination of patients with non-small cell lung cancer and COPD. Lung Cancer. 2009;64:166–70. doi: 10.1016/j.lungcan.2008.08.008. [DOI] [PubMed] [Google Scholar]
  • 16.Horvath I, Hunt J, Barnes PJ, Alving K, Antczak A, Baraldi E, Becher G, van Beurden WJ, Corradi M, Dekhuijzen R, Dweik RA, Dwyer T, Effros R, Erzurum S, Gaston B, Gessner C, Greening A, Ho LP, Hohlfeld J, Jobsis Q, Laskowski D, Loukides S, Marlin D, Montuschi P, Olin AC, Redington AE, Reinhold P, van Rensen EL, Rubinstein I, Silkoff P, et al. Exhaled breath condensate: methodological recommendations and unresolved questions. Eur Respir J. 2005;26:523–48. doi: 10.1183/09031936.05.00029705. [DOI] [PubMed] [Google Scholar]
  • 17.Grob NM, Aytekin M, Dweik RA. Biomarkers in exhaled breath condensate: a review of collection, processing and analysis. J Breath Res. 2008:037004. doi: 10.1088/1752-7155/2/3/037004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Cap P, Dryahina K, Pehal F, Spanel P. Selected ion flow tube mass spectrometry of exhaled breath condensate headspace. Rapid Commun Mass Spectrom. 2008;22:2844–50. doi: 10.1002/rcm.3685. [DOI] [PubMed] [Google Scholar]
  • 19.Martin AN, Farquar GR, Jones AD, Frank M. Human breath analysis: methods for sample collection and reduction of localized background effects. Anal Bioanal Chem. 2010;396:739–50. doi: 10.1007/s00216-009-3217-7. [DOI] [PubMed] [Google Scholar]
  • 20.Ozkan M, Dweik RA. Nitric oxide and airway reactivity. Clin Pulmonary Med. 2001;8:199–206. doi: 10.1097/00045413-200107000-00001. [DOI] [Google Scholar]
  • 21.American Thoracic Society Recommendations for standardized procedures for the on-line and off-line measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide in adults and children-1999. This official statement of the American Thoracic Society was adopted by the ATS Board of Directors, July 1999. Am J Respir Crit Care Med. 1999;160:2104–17. doi: 10.1164/ajrccm.160.6.ats8-99. [DOI] [PubMed] [Google Scholar]
  • 22.McCulloch M, Jezierski T, Broffman M, Hubbard A, Turner K, Janecki T. Diagnostic accuracy of canine scent detection in early- and late-stage lung and breast cancers. Integr Cancer Ther. 2006;5:30–9. doi: 10.1177/1534735405285096. [DOI] [PubMed] [Google Scholar]
  • 23.Gordon SM, Szidon JP, Krotoszynski BK, Gibbons RD, O’Neill HJ. Volatile organic compounds in exhaled air from patients with lung cancer. Clin Chem. 1985;31:1278–82. [PubMed] [Google Scholar]
  • 24.Phillips M, Gleeson K, Hughes JM, Greenberg J, Cataneo RN, Baker L, McVay WP. Volatile organic compounds in breath as markers of lung cancer: a cross-sectional study. Lancet. 1999;353:1930–3. doi: 10.1016/S0140-6736(98)07552-7. [DOI] [PubMed] [Google Scholar]
  • 25.Di Natale C, Macagnano A, Martinelli E, Paolesse R, D'Arcangelo G, Roscioni C, Finazzi-Agro A, D'Amico A. Lung cancer identification by the analysis of breath by means of an array of non-selective gas sensors. Biosens Bioelectron. 2003;18:1209–18. doi: 10.1016/S0956-5663(03)00086-1. [DOI] [PubMed] [Google Scholar]
  • 26.Phillips M, Cataneo RN, Cummin AR, Gagliardi AJ, Gleeson K, Greenberg J, Maxfield RA, Rom WN. Detection of lung cancer with volatile markers in the breath. Chest. 2003;123:2115–23. doi: 10.1378/chest.123.6.2115. [DOI] [PubMed] [Google Scholar]
  • 27.Mazzone P, Hammel J, Dweik R, Na J, Czich C, Laskowski D, Makhail T. Diagnosis of lung cancer by the analysis of exhaled breath with a calorimetric sensor array. Thorax. 2007;62:565–68. doi: 10.1136/thx.2006.072892. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Phillips M, Altorki N, Austin JH, Cameron RB, Cataneo RN, Kloss R, Maxfield RA, Munawar MI, Pass HI, Rashid A, Rom WN, Schmitt P, Wai J. Detection of lung cancer using weighted digital analysis of breath biomarkers. Clin Chim Acta. 2008;393:76–84. doi: 10.1016/j.cca.2008.02.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.ATS/ERS recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide, 2005. Am J Respir Crit Care Med. 2005;171:912–30. doi: 10.1164/rccm.200406-710ST. [DOI] [PubMed] [Google Scholar]
  • 30.Di Francesco F, Loccioni C, Fioravanti M, Russo A, Pioggia G, Ferro M, Roehrer I, Tabucchi S, Onor M. Implementation of fowler’s method for end-tidal air sampling. J Breath Res. 2008;2:037009. doi: 10.1088/1752-7155/2/3/037009. [DOI] [PubMed] [Google Scholar]
  • 31.Herbig J, Titzmann T, Beauchamp J, Kohl I, Hansel A. Buffered end-tidal (BET) sampling - a novel method for real-time breath-gas analysis. J Breath Res. 2008;2:037008. doi: 10.1088/1752-7155/2/3/037008. [DOI] [PubMed] [Google Scholar]
  • 32.Szulejko JE, McCulloch M, Jackson J, McKee DL, Walker JC, Solouki T. Evidence for Cancer Biomarkers in Exhaled Breath. IEEE Sens J. 2010;10:185–210. doi: 10.1109/JSEN.2009.2035669. [DOI] [Google Scholar]