Abstract
Since the invention of pulse oximetry by Takuo Aoyagi in the early 1970s, its use has expanded beyond the perioperative care into neonatal, paediatric and adult intensive care units (ICUs). Pulse oximetry is one of the most important advances in respiratory monitoring as its readings (SpO2) are used clinically as an indirect estimation of arterial oxygen saturation (SaO2). Sensors were placed frequently on the sole, palm, ear lobe or toes in addition to finger. On performing an extensive Medline search using the terms “accuracy of pulse oximetry” and “precision of pulse oximetry”, limited data were found in congenital heart disease patients in the immediate post-corrective stage. Also, there are no reports and comparative data of the reliability and precision of pulse oximetry when readings from five different sensor locations (viz. finger, palm, toe, sole and ear) are analysed simultaneously. To fill these lacunae of knowledge, we undertook the present study in 50 infants and children with cyanotic heart disease in the immediate post-corrective stage.
Keywords: Accuracy, bias, precision, co-oximeter, pulse oximetry, SpO2, SaO2
INTRODUCTION
To examine the accuracy and precision of pulse oximetry in cyanotic heart disease patients in the early post-operative period:
When sensor is placed at five different locations (viz. finger, palm, toe, sole and ear).
At low saturation states (SaO2<90).
To identify the best sensor location.
METHODS
After ethics committee approval, 50 children in the age group of 1 month to 7 years after various corrective surgeries for cyanotic congenital heart diseases in the early post-operative period were selected for this observational study.
Exclusion criteria
Core body temperature <35°C
Children with diseases or conditions known to affect pulse oximeter accuracy, e.g. sickle cell disease, congenital methaemoglobinaemia etc.
Lactate level >2 mmol/L
High inotropic support (dopamine/dobutamine ≥5 μ/kg/min).
All the patients were on ventilatory support with an FiO2 of 0.6 and invasive arterial line in place. Measurements were taken within 3h of arrival in the intensive care unit (ICU) after corrective surgery.
The pulse oximeter used was a Philips M1020A pulse oximetry module (Philips Medical Systems, Eindhoven, Netherlands). An appropriate-sized pulse oximeter sensor (Phillips M1020A/ M1192A/ M1194A/ M1195A) was applied to the finger and palm of the same upper limb, on the toe and sole of the same lower limb along with the ear lobe. At the same time, an arterial blood sample was drawn. A single reading of SpO2 is taken in individual sensor locations after the pulse oximeters had achieved optimal plethysmographic signals and heart rates matching with the electrocardiogram monitor (CMS; Philips Medical Systems). The sensors were covered with black carbon paper, the overhead light was dimmed and ambient lights were reduced by screens to prevent interference. Simultaneously, an arterial blood gas (ABG) analysis was done for co-oximetric measurement of SaO2 using an ABL 800 FLEX ABG machine (Radiometer America Inc., Ohio, USA). Core temperature, arterial pressure, heart rate and ABG value of haemoglobin were recorded simultaneously.
The study population was further subdivided into Group A (SaO2<90%) and Group B (SaO2≥90%) to analyse changes in accuracy and precision at low oxygen saturation.
Comparison of pulse oximetry reading (SpO2) with arterial oxygen saturation (SaO2) is reported in terms of bias and precision as described by Bland and Altman.[1] Bias is the difference between the specific body location SpO2 (finger, palm, toe, sole or ear) and SaO2 (i.e., SpO2-SaO2) and precision is the ±1 standard deviation of the difference.[1] A low bias in a sensor site implies that the pulse oximeter sensor gives a more accurate reading at that site and vice versa. Precision implies the reproducibility of the measurement [Figure 1]. Software SPSS 12.0 version was used for statistical analysis. One-way analysis of variance (ANOVA) test was used to compare the bias of two sensor sites statistically. The mean bias values of the two groups (Normoxaemic and Hypoxaemic) were compared using the unpaired t-test. P<0.05 was considered as significant.
RESULTS
A total of 50 SaO2 measurements were obtained in 50 patients. The mean SaO2 was 96.1%±4.6%, with a range of 83–99.8%. Eight patients had SaO2 measurements <90% (Group A). The preoperative diagnosis of the study population is shown in Table 1.
Table 1.
Predominant preoperative finding | Number of patients |
---|---|
Double outlet right ventricle | 8 |
Transposition of great arteries | 18 |
Tetralogy of Falot | 10 |
Miscellaneous | 14 |
Ear probes consistently showed the highest SpO2 results along with or without other sensor sites in 45 patients.
When the bias (i.e., SpO2- SaO2) is calculated in all the patients, it was observed that the bias is lowest with the sole sensor (-0.088) and highest in the ear sensor (1.572) (P=0.0049, using one-way ANOVA). Thus, in terms of accuracy (i.e., inverse of bias), we can say that the sole sensor is most accurate among the five sites of sensor location.
In Group A also, the sole sensor was found to have the least bias [Figure 2] and hence the most accuracy. Statistical correlation between the sole and the ear sensor was found to be highly significant (P<0.001) using the one-way ANOVA test.
The same trend was maintained by the sole sensor even in Group B [Figure 3]. A statistically significant correlation was found between the sole and the ear sensors (P<0.0001) using one-way ANOVA.
It is noted that in Group A, the value of mean SpO2 (mean of all the five SpO2 readings) is always higher than the SaO2 [Table 2]. Thus, we can say that at low saturation states, SpO2 overestimates SaO2.
Table 2.
Mean SpO2 | SaO2 | Bias |
---|---|---|
89.4 | 85.8 | 3.6 |
91.6 | 89.6 | 2 |
85.6 | 83 | 2.6 |
90.8 | 89 | 1.8 |
93.8 | 89 | 4.8 |
86.6 | 84 | 2.6 |
91.4 | 89.5 | 1.9 |
87.6 | 85 | 2.6 |
Mean Bias 2.72, SD of Mean Bias 1.013
When the SaO2 value is deducted from this mean SpO2 value of a patient, we get the bias for that particular patient. When we consider the average of all the bias values in a particular group, we get the “mean bias” value for the group. The mean group bias in our study was found to be 0.631 and 2.74 in Group B and Group A, respectively [Figure 4] (P=0.0003 using unpaired t-test).
DISCUSSION
Pulse oximetry in the words of Severinghaus and Astrup is “arguably the most significant technologic advance ever made in monitoring the well being of patients during anaesthesia, recovery and critical care”.[2] Pulse oximetry estimates arterial oxygen saturation by measuring the absorption of light of two wavelengths (approximately 660 nm and 940 nm) in human tissue beds. The amount of light absorption varies with the amount of blood in the tissue bed and the relative amounts of oxygenated and deoxygenated haemoglobin.[3] The accuracy of commercially available oximeters differ widely, probably due to the algorithm differences in signal processing.[4]
The aim of our study is to evaluate the accuracy and precision of pulse oximeter at five different sensor locations and to report any significant decline with hypoxaemia (SaO2<90%). Various studies reported a decline in accuracy and precision as the SaO2 decreases below various cut-off values (<90% or <80%)[5,6] The Philips M1020A module was selected for clinical use in our patients because of the accuracy and precision reported by Carter and colleagues.[7] They showed that the Philips M1020A was not affected by foetal haemoglobin (HbF). Different sensors also affect the accuracy of SpO2 measurements. Clayton and others found that the overall rankings were much better for the finger sensors in patients with poor peripheral perfusion.[8] Bell and others,[9] while comparing the traditional band-wrap disposable pulse oximeter sensor with the reusable clip-type sensor, found that the type of sensor selected has little effect on the accuracy of pulse oximetry in children.
Ambient light, skin pigmentation, dyshaemoglobinae-mia, low peripheral perfusion states and motion artefact can affect the performance of pulse oximeters.[10,11] The interference of ambient light can be overcome by simply wrapping the oximeter sensor in opaque material. Villanueva et al.[12] found that age, weight, core and skin temperature, haemoglobin concentration, pulse pressure and percent flow have little effect on the accuracy of pulse oximetry in children. At low levels of saturation (SaO2 below 80%), pulse oximetry is not as accurate as at higher saturations and overestimates the true value.[6] Although the exact mechanism is not known, various investigators have found that SpO2 overestimates SaO2 in polycythemia and underestimates SaO2 with anaemia.[13,14] This might explain the overestimation of SaO2 in the hypoxaemic group observed in the aforementioned studies as well as in our study. Sedaghat-Yazdi and others[15] while studying the effect of sensor location on pulse oximeter accuracy and precision in cyanotic children found that there are no significant differences in bias and precision between finger and toe sensors regardless of SaO2 values. They also found that sensor locations with the worst accuracy and precision were the sole and palm when SaO2 was <90%. This is contrary to the results of our study. Although the reason of our finding is not explainable clearly, factors like increased viscosity of blood, vascular and tissue changes due to clubbing in fingers and toes, etc. might play some yet unproven role. Better tissue perfusion in sole as compared to more peripheral sites like toe, finger and ear lobes might cause the sole sensor to perform better in terms of accuracy. In healthy volunteers, oximeters commonly have a mean difference (bias) of <2% and a standard deviation (precision) of <3% when SaO2 is 90% or above.[16,17] Comparable results have also been obtained in critically ill patients with good arterial perfusion.[18] However, the accuracy deteriorates when SaO2 falls to 80% or less (bias varies from –15.0 to 13.1 while the precision ranges from 1.0 to 16.0).[16]
CONCLUSION
Cyanotic heart disease patients pose a unique dilemma in terms of the reliability and precision of the pulse oximetry readings and the determination of the best location of the sensor, especially in infants. An understanding of the bias and precision of the pulse oximetry at various sensor sites would go a long way in the effective management of patients with cyanotic heart disease in the perioperative period. We strongly recommend that clinicians should verify the measurements by a co-oximeter and evaluate the pulse oximetry sensor used with the particular body site reliability indices to avoid any unacceptable over- or underestimation of the SaO2. This becomes even more relevant in hypoxaemic patients with low SaO2 readings as the margin of safety is very small. We found that sole is the most accurate site of sensor location in cyanotic heart disease paediatric patients. We could also re-establish the finding that at low saturation states, pulse oximetry accuracy deteriorates and tends to overestimate the SaO2. In terms of reproducibility, the best sensor site could not be determined definitely and consistently in our study.
Footnotes
Source of Support: Nil
Conflict of Interest: None declared
REFERENCES
- 1.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10. [PubMed] [Google Scholar]
- 2.Severnghaus JW, Astrup PB. History of blood gas analysis VI. Oximetry. J Clin Monit. 1986;2:270–88. doi: 10.1007/BF02851177. [DOI] [PubMed] [Google Scholar]
- 3.Hess DR, Branson RD. Noninvasive respiratory monitoring equipment. In: Branson RD, Hess DR, Chatburn RL, editors. Respiratory care equipment. Philadelphia: JP Lippincott; 1995. p. 193. [Google Scholar]
- 4.Jubran A. Pulse oximetry. In: Tobin MJ, editor. Principles and Practice of Intensive Care Monitoring. New York: McGraw Hill, Inc; 1998. pp. 261–87. [Google Scholar]
- 5.Torres A, Jr, Skender KM, Wohrley JD, Aldag JC, Raff GW, Bysani GK, et al. Pulse oximetry in children with congenital heart disease: Effects of cardiopulmonary bypass and cyanosis. J Intensive Care Med. 2004;19:229–34. doi: 10.1177/0885066604263819. [DOI] [PubMed] [Google Scholar]
- 6.Schmitt HJ, Schuetz WH, Proeschel PA, Jaklin C. Accuracy of pulse oximetry in children with cyanotic congenital heart disease. J Cardiothorac Vasc Anesth. 1993;7:61–5. doi: 10.1016/1053-0770(93)90120-a. [DOI] [PubMed] [Google Scholar]
- 7.Carter BG, Carlin JB, Tibballs J, Mead H, Hochmann M, Osborne A. Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation. Crit Care Med. 1998;26:1128–133. doi: 10.1097/00003246-199806000-00040. [DOI] [PubMed] [Google Scholar]
- 8.Clayton D, Webb RK, Ralston AC, Duthie D, Runciman WB. Pulse oximeter sensors: a comparison between finger, nose, ear, and forehead sensors under conditions of poor perfusion. Anaesthesia. 1991;46:260–5. doi: 10.1111/j.1365-2044.1991.tb11492.x. [DOI] [PubMed] [Google Scholar]
- 9.Bell C, Luther MA, Nicholson J, Fox CJ, Hirsh JL. Effect of sensor design on accuracy and reliability of pulse oximetry in pediatric patients. J Clin Anesthnone. 1999;11:323–7. doi: 10.1016/s0952-8180(99)00053-7. [DOI] [PubMed] [Google Scholar]
- 10.Trivedi NS, Ghouri AF, Shah NK, Lai E, Barker SJ. Effects of motion, ambient light, and hypoperfusion on pulse oximeter function. J Clin Anesth. 1997;9:179–83. doi: 10.1016/s0952-8180(97)00039-1. [DOI] [PubMed] [Google Scholar]
- 11.Zeballos RJ, Weisman IM. Reliability of noninvasive oximetry in black subjects during exercise and hypoxia. Am Rev Respir Dis. 1991;144:1240–4. doi: 10.1164/ajrccm/144.6.1240. [DOI] [PubMed] [Google Scholar]
- 12.Villanueva R, Bell C, Kain Z, Colingo KA. Effect of peripheral perfusion on accuracy of pulse oximetry in children. J Clin Anest h. 1999;11:317–22. doi: 10.1016/s0952-8180(99)00054-9. [DOI] [PubMed] [Google Scholar]
- 13.Volgyesi GA, Kolesar R, Lerman J. An in vitro model for evaluating the accuracy of pulse oximeters. Can J Anaesth. 1990;37:S67. [PubMed] [Google Scholar]
- 14.Kolesar R, Volgyesi G, Lerman J. Effect of haemoglobin concentration on the accuracy of pulse oximetry. Can J Anaesth. 1990;37:S88. [PubMed] [Google Scholar]
- 15.Sedaghat-Yazdi F, Torres A, Jr, Fortuna R, Geiss DM. Pulse oximeter accuracy and precision affected by sensor location in cyanotic children. Pediatr Crit Care Med. 2008;9:393–7. doi: 10.1097/PCC.0b013e3181727967. [DOI] [PubMed] [Google Scholar]
- 16.Nickerson BG, Sarkisian C, Tremper KK. Bias and precision of pulse oximeters and arterial oximeters. Chest. 1988;93:515–7. doi: 10.1378/chest.93.3.515. [DOI] [PubMed] [Google Scholar]
- 17.Morris RW, Nairn M, Torda TA. A comparison of fifteen pulse oximeters. Part I: a clinical comparison; Part II: a test of performance under conditions of poor perfusion. Anaesth Intensive Care. 1989;17:62–82. doi: 10.1177/0310057X8901700113. [DOI] [PubMed] [Google Scholar]
- 18.Emergency Care Research Institute: Pulse oximeters. Health Devices. 1989;18:185–230. [PubMed] [Google Scholar]