Skip to main content
Medical Physics logoLink to Medical Physics
. 2011 Jan 28;38(2):942–947. doi: 10.1118/1.3537610

Characterization of mesothelioma and tissues present in contrast-enhanced thoracic CT scans

Neal Corson 1, William F Sensakovic 1, Christopher Straus 1, Adam Starkey 1, Samuel G Armato III 1,a)
PMCID: PMC3041808  PMID: 21452730

Abstract

Purpose: The purpose of this study was to characterize the Hounsfield unit (HU) distributions of mesothelioma and other tissues present in contrast-enhanced thoracic CT scans, to compare the HU distributions of mesothelioma, muscle, and liver by scanner and reconstruction filter∕kernel combination, and to assess interpatient HU distribution variability.

Methods: The database consisted of 28 contrast-enhanced thoracic CT scans from different patients. For each scan, regions of interest were manually outlined within each of 13 tissues, including mesothelioma. For each tissue, the empirical percentiles in HU values were calculated along with the interpatient variability. The HU distributions of select tissues were compared across three different scanner and reconstruction filter∕kernel combinations.

Results: The HU distributions of blood-containing tissues demonstrated substantial overlap, as did the HU distributions of pleural effusion, mesothelioma, muscle, and liver. The HU distribution of fat had the least overlap with the other tissues. Fat and muscle had the lowest interpatient HU variability and the narrowest HU distributions, while blood-containing tissues had the highest interpatient HU variability. A soft-tissue reconstruction filter∕kernel yielded the narrowest HU distribution, and fat with artifact had the widest HU distribution.

Conclusions: Characterization of tissues in CT scans enhances the understanding of those tissues’ HU distributions. Due to their overlapping HU distributions and close spatial proximity to one another, separating pleural effusion, mesothelioma, muscle, and liver from one another is a difficult task based on HU value thresholding alone. The results illustrate the wide distributions and large variability that exist for tissues present in clinical thoracic CT scans.

Keywords: Hounsfield units (HU), mesothelioma, tissue characterization, quantitative imaging, computer tomography (CT)

INTRODUCTION

Semiautomated and automated computer methods are proving useful in multiple medical imaging applications, such as screening, diagnosis, and therapeutic response assessment. While an understanding of the discrete information within images may not be necessary for the human visual system, computer algorithms utilize discrete information contained within images to perform their tasks. Thus, an understanding of this discrete information and its variability across different patients and imaging parameters is needed to improve computer methods. In computed tomography (CT), a CT section is made up of many pixels, each with its own discrete Hounsfield unit (HU) value. Tissue characterization captures the HU distribution of pixels for a given tissue.1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15

Many computer-aided diagnostic techniques for thoracic imaging require some level of image segmentation. The goal of the present study was to quantify the physical basis for difficulties encountered when a gray-level thresholding technique alone (which has been a common approach) is attempted for the purpose of segmenting tissues present within thoracic CT scans. Similar challenges may be encountered during the segmentation of thoracic structures for radiation therapy treatment planning. The present study sought to quantify the range (variability) of HU values for specific tissues and the extent of the overlap of these HU values for different, spatially adjacent tissues; these data are meant to serve as a guide for investigators engaged in image segmentation tasks.

The results from this study will further the understanding of HU distributions and variability contained within various tissues in thoracic CT scans. This understanding will prove beneficial when designing segmentation algorithms for thoracic CT scans.

MATERIALS AND METHODS

A database of 28 contrast-enhanced thoracic CT scans from 28 mesothelioma patients (23 male, 5 female, age 50–83 yr) was retrospectively collected and analyzed. The scans were performed at the University of Chicago Medical Center on three different CT scanners: Brilliance 16 (n=2), Brilliance 16P (n=19), and Brilliance 64 (n=7) (Philips Medical Systems, Cleveland, OH). Scans were acquired with 120 or 140 kVp and 113–330 mA s (mean: 237 mA s). All scans were infused with 120 ml of contrast agent (Omnipaque 350), according to a clinical imaging protocol that used an injection rate of 2.2 ml∕s with a 65 s delay. All scans were reconstructed axially as 512×512 pixel images and had 1 mm reconstruction interval and 1 mm section thickness. The database was collected under an Institutional Review Board-approved protocol. All applicable Health Insurance Portability and Accountability Act regulations were observed during the collection, maintenance, and use of this database.

The 13 tissue types investigated were heart, ascending aorta, descending aorta, pulmonary artery, pulmonary artery with artifact, inferior vena cava (IVC), pleural effusion, mesothelioma, muscle, spleen, liver, fat, and fat with artifact, where “artifact” refers to photon starving artifact due to contrast agent. An in-house image viewer and measurement system was used to manually construct and label the regions of interest (ROIs) within these 13 tissues. The ROIs (Fig. 1) were created away from the edges of tissues to avoid pixels found near tissue boundaries where partial volume effects occur. The ROIs were created by a single observer [NC] according to guidelines to ensure proper sampling of the sections and tissues within each section.

Figure 1.

Figure 1

ROIs selected from various tissues within the same CT section: (a) fat, (b) mesothelioma, (c) ascending aorta, (d) pulmonary artery with artifact, (e) pulmonary artery, (f) descending aorta, (g) mesothelioma contour by radiologist, and (h) muscle.

To compare the HU distributions of different tissues, distributions for each tissue were created by combining all the HU values from the ROIs within that tissue across all sections of all CT scans. The minimum, maximum, and median HU values as well as the HU value at 2.5, 25, 50, 75, and 97.5 percentiles were calculated. Overlap among the HU distributions of different tissue types was illustrated with a box-and-whiskers plot. For each tissue type, the interpatient variability (expressed as the standard deviation of median HU values across patients) was calculated.

The influence of scanner and reconstruction filter∕kernel combination on HU distributions was investigated. The ROIs of mesothelioma, muscle, and liver were compared across three different scanner and reconstruction filter∕kernel combinations: 16P B∕B, 16P D∕D, and 64 D∕D, where the “16P” and “64” refer to the number of rows in the detector. Reconstruction filter∕kernel “B∕B” indicates soft-tissue reconstruction, while “D∕D” indicates bone reconstruction.

RESULTS

The number of scans, the number of ROIs, and the average number of pixels per ROI for each tissue are presented in Table 1. Pulmonary artery with artifact, IVC, pleural effusion, and mesothelioma have fewer than 28 scans due to the limited presence of these tissues within the scans. The number of ROIs per tissue varies due to the presence of the tissue within a scan or the ability to create an ROI within the tissue. Figure 2 illustrates the overlap of the tissues’ HU distributions. The minimum, maximum, and median HU values as well as the 2.5, 25, 50, 75, 97.5, and 99.5 percentiles are presented in Table 2. The percentile differences, 25%–75% and 2.5%–97.5%, which describe the width of the distribution, and the interpatient variability are also presented in Table 2.

Table 1.

Number of scans, number of ROIs, and average number of pixels per ROI for each tissue type. AscAorta=ascending aorta, DescAorta=descending aorta, PA=pulmonary artery, PaArtifact=pulmonary artery with artifact, IVC=inferior vena cava, Effusion=pleural effusion, Meso=mesothelioma.

Tissue type No. of scans No. of ROIs Average No. of pixels per ROI
Heart 28 84 6020
AscAorta 28 84 713
DescAorta 28 84 397
PA 28 84 472
PaArtifact 26 78 331
IVC 26 76 340
Effusion 10 28 1205
Meso 25 86 664
Muscle 28 84 567
Spleen 28 84 1559
Liver 28 84 2934
Fat 28 84 890
Fat Artifact 28 82 169

Figure 2.

Figure 2

The distributions of HU values for the 13 tissue types across all 28 CT scans. Extensive overlap exists among mesothelioma, muscle, and liver.

Table 2.

Statistics for HU distributions for different tissues across all 28 CT scans. AscAorta=ascending aorta, DescAorta=descending aorta, PA=pulmonary artery, PaArtifact=pulmonary artery with artifact, IVC=inferior vena cava, Effusion=pleural effusion, Meso=mesothelioma.

Tissue type Emprical percentiles (HU) Percentile differences Interpatient variability
Min 2.5% 25% Median 75% 97.5% Max 25%–75% 2.5%–97.5% SD of medians
Heart −482 14 118 169 223 351 1908 105 337 36
AscAorta 23 147 211 246 286 381 508 75 234 47
DescAorta −142 103 197 236 278 372 598 81 269 41
PA −71 95 172 218 261 362 521 89 267 60
PaArtifact −132 97 171 216 263 396 659 92 299 56
IVC −150 14 92 129 170 267 506 78 253 30
Effusion −217 −102 −24 4 30 87 224 54 189 41
Meso −258 −52 23 56 91 181 355 68 233 26
Muscle −260 −33 37 61 85 147 393 48 180 7
Spleen −205 11 94 130 168 252 492 74 241 25
Liver −315 −31 56 92 126 205 480 70 236 23
Fat −339 −156 −112 −93 −72 −19 204 40 137 11
Fat Artifact −1000 −254 −112 −61 3 205 1545 115 459 38

For mesothelioma, muscle, and liver, the number of scans and the number of ROIs for each combination of scanner and reconstruction filter∕kernel are presented in Table 3. As illustrated in Fig. 3, reconstruction filter∕kernel B∕B has lower percentile differences than reconstruction filter∕kernel D∕D. The interpatient variability (expressed as the standard deviation of the median HU values across patients) for the different combinations of scanner and reconstruction filter∕kernel 16P B∕B, 16P D∕D, and 64 D∕D are presented in Table 4. For mesothelioma, the interpatient variability for 16P D∕D and 64 D∕D are high due to each having an outlier. When the outliers are removed, the interpatient variability for 16P D∕D and 64 D∕D are 13 HU and 13 HU, respectively.

Table 3.

Number of scans and ROIs by scanner and reconstruction filter∕kernel combination. Meso=mesothelioma.

Tissue type Filter∕kernel No. of scans No. of ROIs
Meso 16P B∕B 6 18
  16P D∕D 13 44
  64 D∕D 7 18
Muscle 16P B∕B 6 18
  16P D∕D 13 39
  64 D∕D 7 21
Liver 16P B∕B 6 18
  16P D∕D 13 39
  64 D∕D 7 21

Figure 3.

Figure 3

The distributions of HU values of mesothelioma, muscle, and liver for three different scanner and reconstruction filter∕kernel combinations.

Table 4.

HU statistics of mesothelioma, muscle, and liver by scanner and reconstruction filter∕kernel combination. Meso=mesothelioma.

Tissue type Filter∕kernel Empirical percentiles (HU) Percentile differences Interpatient variability
Min 2.5% 25% Median 75% 97.5% Max 25%–75% 2.5%–97.5% SD of medians
Meso 16P B∕B −116 −26 19 42 65 125 225 46 151 9
  16P D∕D −225 −46 29 63 99 199 355 70 245 33
  64 D∕D −200 −68 23 62 98 167 287 75 235 23
                     
Muscle 16P B∕B −160 1 40 58 76 116 186 36 115 6
  16P D∕D −260 −38 38 62 86 151 393 48 189 8
  64 D∕D −181 −48 30 60 88 153 277 58 201 5
                     
Liver 16P B∕B −92 16 70 98 125 180 300 55 164 23
  16P D∕D −223 −37 48 85 121 196 480 73 233 26
  64 D∕D −315 −50 52 93 132 223 442 80 273 18

DISCUSSION

This study characterized the HU distributions of mesothelioma and other tissues found in contrast-enhanced thoracic CT scans of patients with mesothelioma. The purpose of the study was to investigate the aggregate variability of HU values for different tissues, with secondary evaluations of interpatient variability and variability due to different scanner and reconstruction filter∕kernel combinations. The study did not evaluate the components of variability attributable to other image acquisition parameters. The results illustrate (1) large variability in the HU values within a specific tissue across patients in clinical thoracic CT scans acquired using the same imaging protocol on different scanners by the same manufacturer and (2) a large degree of overlap in the HU distributions of spatially adjacent structures.

The HU distributions of pleural effusion and mesothelioma overlap, as seen in Fig. 2. Such overlap makes separation of these structures based on HU values alone a difficult task. The HU distributions of mesothelioma, muscle, and liver overlap with the HU values of mesothelioma between pleural effusion (with lower HU values) and muscle and liver (with higher HU values). These similar distributions and similar anatomical locations make segmentation of mesothelioma, pleural effusion, liver, and muscle from one another a challenge. Segmentation algorithms that attempt to segment any of these four structures will require complex approaches beyond simple HU value thresholding. Fat, on the other hand, has the least overlap with the other distributions, as shown in Fig. 2. Thus, segmentation of fat from the other structures may be possible with simple thresholding-based segmentation techniques.16

For a given tissue, the interpatient variability (expressed as the standard deviation of the median HU values across patients) was calculated. Interpatient variability widens the HU distributions for each tissue across patients since, although the widths of tissue HU distributions for different patients were similar, the median values were shifted relative to those of other patients. A low interpatient variability demonstrates that the same tissue from different patients have similar median HU values. Muscle and fat exhibited the lowest interpatient variability.

When comparing distributions from different scanners and reconstruction filter∕kernel combinations, mesothelioma, muscle, and liver were investigated because they are difficult to segment due to their similar HU values. As expected, the smoothing effect of the soft-tissue reconstruction filter∕kernel B∕B lowers the percentile differences across all three tissues when compared to D∕D.

While the results and conclusions of this study are not surprising, a systematic investigation such as the one presented here is required to provide direct evidence of an otherwise expected finding. Many articles in the literature report algorithms that segment tissues through gray-level thresholding (simple thresholding as well as adaptive thresholding), and this study was intended to investigate the similarities of the gray levels depicted by various structures within thoracic CT scans in an effort to understand the reliability and generalizability of such approaches. Our findings suggest that segmentation of thoracic structures requires methods more complex than threshold-based segmentation alone.

We expected (1) large variability in the HU values within a specific structure across patients and scanners and (2) a large degree of overlap in the HU distributions of spatially adjacent structures. Both expectations are demonstrated in our results. These findings generate two key conclusions. First, that gray-level thresholding-based segmentation of structures in thoracic CT scans will likely prove challenging; second, that natural variability in these scans must be considered to provide segmentation algorithms with some element of generalizability. The essential finding from our study is not the absolute values of the HU distributions presented, but rather the range of HU values for specific tissues and the extent of the overlap of these HU ranges for different, spatially adjacent tissues.

Further research should investigate intrapatient variability, which would require multiple scans for the same patient. This study had a limited database of 28 patients; a similar study with a larger database would yield more robust results. While this study examined Philips scanners, future studies should include scans from multiple scanner manufacturers.

CONCLUSION

This study characterized tissues found within contrast-enhanced thoracic CT scans of mesothelioma patients. The HU values of mesothelioma tend to fall between those of pleural effusion (with lower HU values) and those of muscle and liver (with higher values). The similar HU values and close spatial proximity make computerized segmentation of these four structures based on HU values alone a challenge. Of the tissues investigated, fat has a narrow distribution, low interpatient variability, and limited overlap with other tissues, which make it an ideal candidate for threshold-based segmentation. The results illustrate the wide distributions and large variability that exist for tissues present in clinical thoracic CT scans acquired using the same imaging protocol on different scanners by the same manufacturer.

ACKNOWLEDGMENTS

The authors would like to thank Lorenzo Pesce, Ph.D., for many insightful and valuable conversations. This work was supported by USPHS under Grant No. CA102085.

References

  1. Best A. C., Lynch A. M., Bozic C. M., Miller D., Grunwald G. K., and Lynch D. A., “Quantitative CT indexes in idiopathic pulmonary fibrosis: Relationship with physiologic impairment,” Radiology 228, 407–414 (2003). 10.1148/radiol.2282020274 [DOI] [PubMed] [Google Scholar]
  2. Best A. C., Meng J., Lynch A. M., Bozic C. M., Miller D., Grunwald G. K., and Lynch D. A., “Idiopathic pulmonary fibrosis: Physiologic tests, quantitative CT indexes, and CT visual scores as predictors of mortality,” Radiology 246, 935–940 (2008). 10.1148/radiol.2463062200 [DOI] [PubMed] [Google Scholar]
  3. Gilman M. J., R. G.Laurens, Jr., Somogyi J. W., and Honig E. G., “CT attenuation values of lung density in sarcoidosis,” J. Comput. Assist. Tomogr. 7, 407–410 (1983). 10.1097/00004728-198306000-00003 [DOI] [PubMed] [Google Scholar]
  4. Hartley P. G., Galvin J. R., Hunninghake G. W., Merchant J. A., Yagla S. J., Speakman S. B., and Schwartz D. A., “High-resolution CT-derived measures of lung density are valid indexes of interstitial lung disease,” J. Appl. Physiol. 76, 271–277 (1994). [DOI] [PubMed] [Google Scholar]
  5. Sumikawa H., Johkoh T., Yamamoto S., Takahei K., Ueguchi T., Ogata Y., Matsumoto M., Fujita Y., Natsag J., Inoue A., Tsubamoto M., Mihara N., Honda O., Tomiyama N., Hamada S., and Nakamura H., “Quantitative analysis for computed tomography findings of various diffuse lung diseases using volume histogram analysis,” J. Comput. Assist. Tomogr. 30, 244–249 (2006). 10.1097/00004728-200603000-00014 [DOI] [PubMed] [Google Scholar]
  6. Sverzellati N., Zompatori M., De Luca G., Chetta A., Bna C., Ormitti F., and Cobelli R., “Evaluation of quantitative CT indexes in idiopathic interstitial pneumonitis using a low-dose technique,” Eur. J. Radiol. 56, 370–375 (2005). 10.1016/j.ejrad.2005.05.012 [DOI] [PubMed] [Google Scholar]
  7. Okada T., Iwano S., Ishigaki T., Kitasaka T., Hirano Y., Mori K., Suenaga Y., and Naganawa S., “Computer-aided diagnosis of lung cancer: Definition and detection of ground-glass opacity type of nodules by high-resolution computed tomography,” Jpn. J. Radiol. 27, 91–99 (2009). 10.1007/s11604-008-0306-z [DOI] [PubMed] [Google Scholar]
  8. Pombo F., Rodriguez E., Caruncho M. V., Villalva C., and Crespo C., “CT attenuation values and enhancing characteristics of thoracoabdominal lymphomatous adenopathies,” J. Comput. Assist. Tomogr. 18, 59–62 (1994). 10.1097/00004728-199401000-00013 [DOI] [PubMed] [Google Scholar]
  9. Park H. S., Lee J. M., Kim S. H., Jeong J. Y., Kim Y. J., Lee K. H., Choi S. H., Han J. K., and Choi B. I., “CT differentiation of cholangiocarcinoma from periductal fibrosis in patients with hepatolithiasis,” AJR, Am. J. Roentgenol. 187, 445–453 (2006). 10.2214/AJR.05.0247 [DOI] [PubMed] [Google Scholar]
  10. Sjostrom L., Kvist H., Cederblad A., and Tylen U., “Determination of total adipose tissue and body fat in women by computed tomography, 40K, and tritium,” Am. J. Physiol. 250, E736–45 (1986). [DOI] [PubMed] [Google Scholar]
  11. Ricci C., Longo R., Gioulis E., Bosco M., Pollesello P., Masutti F., Croce L. S., Paoletti S., de Bernard B., Tiribelli C., and Dalla Palma L., “Noninvasive in vivo quantitative assessment of fat content in human liver,” J. Hepatol 27, 108–113 (1997). 10.1016/S0168-8278(97)80288-7 [DOI] [PubMed] [Google Scholar]
  12. Iida T., Yagi S., Taniguchi K., Hori T., Uemoto S., Yamakado K., and Shiraishi T., “Significance of CT attenuation value in liver grafts following right lobe living-donor liver transplantation,” Am. J. Transplant. 5, 1076–1084 (2005). 10.1111/j.1600-6143.2005.00799.x [DOI] [PubMed] [Google Scholar]
  13. Davidson L. E., Kuk J. L., Church T. S., and Ross R., “Protocol for measurement of liver fat by computed tomography,” J. Appl. Physiol. 100, 864–868 (2006). 10.1152/japplphysiol.00986.2005 [DOI] [PubMed] [Google Scholar]
  14. Nandalur K. R., Hardie A. H., Bollampally S. R., Parmar J. P., and Hagspiel K. D., “Accuracy of computed tomography attenuation values in the characterization of pleural fluid: An ROC study,” Acad. Radiol. 12, 987–991 (2005). 10.1016/j.acra.2005.05.002 [DOI] [PubMed] [Google Scholar]
  15. Lee K. S., Im J. G., Choe K. O., Kim C. J., and Lee B. H., “CT findings in benign fibrous mesothelioma of the pleura: Pathologic correlation in nine patients,” AJR, Am. J. Roentgenol. 158, 983–986 (1992). [DOI] [PubMed] [Google Scholar]
  16. Yoshizumi T., Nakamura T., Yamane M., Islam A. H., Menju M., Yamasaki K., Arai T., Kotani K., Funahashi T., Yamashita S., and Matsuzawa Y., “Abdominal fat: Standardized technique for measurement at CT,” Radiology 211, 283–286 (1999). [DOI] [PubMed] [Google Scholar]

Articles from Medical Physics are provided here courtesy of American Association of Physicists in Medicine

RESOURCES