Evaluation of CLSI Agar Dilution Method and Trek Sensititre Broth Microdilution Panel for Determining Antimicrobial Susceptibility of Streptococcus pneumoniae

Sean X Zhang 1,2,*, Prasad Rawte 1, Shirley Brown 1, Steven Lo 1, Heather Siebert 1, Sylvia Pong-Porter 3, Donald E Low 1,2,3, Frances B Jamieson 1,2
PMCID: PMC3043519  PMID: 21123533

Abstract

Both the CLSI agar dilution method and Trek Sensititre broth microdilution panel for Streptococcus pneumoniae antimicrobial susceptibility testing were evaluated against the reference CLSI broth microdilution method using the most recently published CLSI breakpoints. While agar dilution was not an optimal method, the commercial panel appeared to be an acceptable method, with minor errors encountered for ceftriaxone, penicillin, and meropenem.


Broth dilution is the reference method of choice recommended by the CLSI (Clinical and Laboratory Standards Institute) for Streptococcus pneumoniae antimicrobial susceptibility testing (AST) (3). Agar dilution is considered an acceptable method for S. pneumoniae AST and has been widely used by large-scale laboratories due to high-volume suitability and cost effectiveness; however, the performance of agar dilution using the CLSI method for S. pneumoniae AST has not been previously reported in comparison to that of the broth dilution method. A commercial broth microdilution panel (part no. STP5F, Sensititre; Trek Diagnostic Systems, Cleveland, OH) is available for S. pneumoniae AST. Recently, the CLSI published new penicillin breakpoints and interpretive criteria for S. pneumoniae for the following: oral penicillin, parenteral penicillin (nonmeningitis), and parenteral penicillin (meningitis) (3). The purpose of this study was to evaluate the CLSI agar dilution method and the Trek Sensititre broth microdilution panel for S. pneumoniae AST with the most recent CLSI interpretive breakpoints.

A total of 132 S. pneumoniae clinical isolates, consisting of a selection of isolates with MICs that span the CLSI criteria, were collected for the evaluation. The CLSI broth microdilution method was chosen as the reference method for this study. The broth microdilution panels were made by following the CLSI broth dilution protocol (2) and were tested against CLSI quality control (QC) strains (S. pneumoniae ATCC 49619, Staphylococcus aureus ATCC 29213, and Escherichia coli ATCC 25922) before they were used in the study. Their MIC results were all within the acceptable CLSI QC ranges (3). During the study, S. pneumoniae ATCC 49619 was used as a QC strain, and the MIC results of the broth microdilution plates were acceptable if the QC results were within the acceptable ranges. The CLSI agar dilution method was performed by following the CLSI agar dilution procedure, using Mueller-Hinton agar with 5% sheep blood (2). The method for the Trek Sensititre broth microdilution panel was followed according to the manufacturer's instructions. The Sensititre panels were read by a Sensititre SensiTouch reader (Trek Diagnostic Systems, Cleveland, OH). Both the agar dilution plates and the Trek Sensititre panels were run in parallel with the CLSI broth microdilution panels. The MIC was interpreted by the recently published CLSI breakpoints (3). The performance of the two testing methods (agar dilution and the Trek Sensititre panel) was determined using essential agreement (the MIC result of the test method is within ±1 doubling dilution of the MIC result of the reference method) and category agreement (agreement of interpretative results between the test method and the reference method). Error assessment was categorized as “very major error” (VME) (the test method indicates a susceptible response while the reference method indicates a resistant response), “major error” (ME) (the test method indicates a resistant response while the reference method indicates a susceptible response), and “minor error” (MiE) (the test method indicates an intermediate response while the reference method indicates a susceptible or resistant response or the test method indicates a susceptible or resistant response while the reference method indicates an intermediate response). Acceptable performance rates were measured as follows: ≥90% for essential agreement or category agreement, ≤3% for very major error or major error, and ≤7% for major error plus minor error (1, 5).

Eleven antimicrobial agents were evaluated using the agar dilution method, of which five had unacceptable results compared to those of the reference method (Table 1). For ceftriaxone, a very low level of category agreement (53.8 to 68.2%) was observed as a result of 6 VMEs (13.6%) and 42 to 55 MiEs (31.8 to 41.7%); the essential agreement rate (86.6 to 88.9%) was also lower than the acceptable rate (≥90%). MiE rates for penicillin (12.2 to 14.5%), meropenem (16%), and telithromycin (15.2%) were unacceptable. Three VMEs (3.2%) for parenteral penicillin (meningitis) and three MEs (3.7%) for clindamycin also occurred. The agar dilution method was unable to reliably determine the susceptibility of the five antimicrobial agents tested for S. pneumoniae, particularly ceftriaxone and penicillin, considered first-line agents for the treatment of S. pneumoniae infection. In contrast to our study, a British study reported that the agar dilution method using BSAC (British Society for Antimicrobial Chemotherapy) protocols was found to have good agreement with the CLSI (formerly NCCLS) broth microdilution method (5a), although the antimicrobial agents tested in the British study were different from those tested in our study, e.g., ceftriaxone was not assessed in this study, and testing parameters were different (6).

Table 1.

Evaluation of the performance of the CLSI agar dilution method for S. pneumoniae ASTc

Drug No. of isolates
% EA (no. of isolates that agree/total no. of isolates)a % CA (no. of isolates that agree) % error rate (no. of isolates that disagree)
Total S I R VME ME MiE
Ceftriaxone
    Meningitis 132 60 28 44 88.9 (72/81) 53.8 (71) 13.6 (6) 0 41.7 (55)
    Nonmeningitis 132 87 39 6 86.6 (71/82) 68.2 (90) 0 0 31.8 (42)
Clindamycin 131 82 2 47 N/Ab 93.1 (122) 2.1 (1) 3.7 (3) 3.8 (5)
Erythromycin 131 48 1 82 N/A 97.7 (128) 2.4 (2) 0 0.8 (1)
Levofloxacin 131 104 6 21 N/A 98.5 (129) 0 0 1.5 (2)
Meropenem 131 60 26 45 N/A 84.0 (110) 0 0 16.0 (21)
Penicillin
    Oral penicillin V 131 38 26 67 98.0 (99/101) 87.8 (115) 0 0 12.2 (16)
    Parenteral
        Meningitis 131 38 N/A 93 98.0 (99/101) 96.9 (127) 3.2 (3) 2.6 (1) N/A
        Nonmeningitis 131 98 32 1 98.0 (99/101) 85.5 (112) 0 0 14.5 (19)
Chloramphenicol 131 120 N/A 11 90.7 (88/97) 99.2 (130) 0 0.8 (1) N/A
Tetracycline 132 62 2 68 N/A 97.7 (129) 0 0 2.3 (3)
Telithromycin 132 131 1 0 N/A 84.8 (112) N/A 0 15.2 (20)
TMP-SMX 129 30 20 79 N/A 97.7 (126) 0 0 2.3 (3)
Vancomycin 132 132 N/A N/A N/A 100 (132) N/A 0 N/A
a

The denominators of the essential agreement (EA) were less than the total number of the isolates tested because MIC results of some isolates were out of the drug concentration ranges, and thus, the EA cannot be determined.

b

The EA cannot be determined because the MIC results were out of the drug concentration ranges.

c

S, susceptible; I, intermediate, R, resistant; EA, essential agreement, CA, category agreement; VME, very major error, ME, major error, MiE, minor error; N/A, not available; TMP-SMX, trimethoprim-sulfamethoxazole.

Overall, the CLSI agar dilution method tended to result in lower MICs than the reference method (possibly due to the stability of some drugs, such as ceftriaxone in agar media, and some other factors), resulting in six VMEs (falsely susceptible) for ceftriaxone and three VMEs for penicillin. Moreover, growth of S. pneumoniae on agar media may not be optimal due to its fastidious nature (2). This assumption may be supported by the notion that special formulations and supplements are needed for testing fastidious organisms using agar media; for example, GC agar base and 1% defined growth supplement are required for testing Neisseria gonorrhoeae (2).

Fourteen antimicrobial agents were evaluated using the Trek Sensititre broth microdilution panel. Three antimicrobial agents had MiEs compared to those evaluated by the reference method (Table 2): ceftriaxone (11.5 to 13%), penicillin (10 to 17.7%), and meropenem (9.2%). Two MEs (5.3%) occurred with parenteral penicillin (meningitis) breakpoints. Almost all the MICs accounting for the discrepant results were clustered at the breakpoints. For example, all MICs causing MiEs for ceftriaxone were at 0.5, 1, and 2 μg/ml, the interpretive breakpoints for S. pneumoniae susceptibility. The two parenteral penicillin (meningitis) MICs that were MEs were at 0.12 μg/ml (resistant breakpoint), and the reference MICs were at 0.06 μg/ml (susceptible breakpoint). Due to the lack of an intermediate interpretation category, a one-dilution difference resulted in two MEs (falsely resistant).

Table 2.

Evaluation of the performance of the Trek Sensititre broth microdilution panel for S. pneumoniae ASTc

Drug No. of isolates
% EA (no. of isolates that agree/total no. of isolates)a % CA (no. of isolates that agree) % error rate (no. of isolates that disagree)
Total S I R VME ME MiE
Ceftriaxone
    Meningitis 131 59 28 44 98.8 (81/82) 88.5 (116) 0 0 11.5 (15)
    Nonmeningitis 131 86 39 6 98.8 (80/81) 87.0 (114) 0 0 13.0 (17)
Clindamycin 130 81 2 47 N/Ab 98.5 (128) 0 0 1.5 (2)
Erythromycin 129 48 0 81 N/A 99.2 (128) 0 0 0.8 (1)
Levofloxacin 131 104 6 21 99.0 (98/99) 98.5 (129) 0 0 1.5 (2)
Meropenem 131 60 26 45 100 (71/71) 90.8 (119) 0 0 9.2 (12)
Penicillin
    Oral penicillin V 130 38 25 67 98.9 (90/91) 90.0 (117) 0 0 10.0 (13)
    Parenteral
        Meningitis 130 38 N/A 92 98.9 (90/91) 98.5 (128) 0 5.3 (2) N/A
        Nonmeningitis 130 97 32 1 98.9 (90/91) 82.3 (107) 0 0 17.7 (23)
Chloramphenicol 131 120 N/A 11 94.4 (117/124) 99.2 (130) 0 0.8 (1) N/A
Tetracycline 130 61 2 67 N/A 97.7 (127) 0 0 2.3 (3)
Telithromycin 131 130 1 0 N/A 99.2 (130) N/A 0 0.8 (1)
TMP-SMX 131 32 20 79 91.7 (22/24) 97.7 (128) 0 0 2.3 (3)
Vancomycin 131 131 N/A N/A N/A 100 (131) N/A 0 N/A
Linezolid 132 131 N/A N/A 92.2 (118/128) 100 (131) N/A 0 N/A
Cefuroxime (sodium) 130 49 3 78 N/A 97.7 (127) 0 0 2.3 (3)
Moxifloxacin 131 111 5 15 N/A 96.2 (126) 0 0 3.8 (5)
a

The denominators of the essential agreement (EA) were less than the total number of the isolates tested because MIC results of some isolates were out of the drug concentration ranges, and thus, the EA cannot be determined.

b

The EA cannot be determined because the MIC results were out of the drug concentration ranges.

c

S, susceptible; I, intermediate, R, resistant; EA, essential agreement, CA, category agreement; VME, very major error, ME, major error, MiE, minor error; N/A, not available; TMP-SMX, trimethoprim-sulfamethoxazole.

We also conducted a 3-month prospective study to test the reproducibility of the commercial panel and to rule out any technical error attributable to the above-described discrepancies. The prospective study was done using the new isolates prospectively collected during the 3-month period and with the same batch of the Sensititre plates. Any isolate with a MIC at the breakpoint for ceftriaxone, penicillin, and/or meropenem had repeat AST performed. A total of 159 isolates required repeat testing, of which 20 showed MICs different from those found when the first AST was performed, but all were within ±1 doubling dilution except 1 isolate. Only four repeats (two for ceftriaxone and two for meropenem) resulted in a change in the interpretation from intermediate to resistant and vice versa. Of note, the one repeat (ceftriaxone) with a two-doubling-dilution difference did not result in a change in the interpretation. These results demonstrated that the Trek Sensititre panel was not only found to have acceptable reproducibility (only 1 out of 159 repeat tests [≤1%] had a MIC outside the range by one doubling dilution) but also gave an acceptable level of accuracy because only 2.5% (4/159) of the repeat tests resulted in a different interpretive category.

Overall, the Trek Sensititre panel was not found to produce significant VME (falsely susceptible) and ME (falsely resistant) rates, although 9.2 to 17.7% MiE rates for penicillin, ceftriaxone, and meropenem were encountered. Other investigators have also observed minor errors for ceftriaxone and penicillin when they used commercial broth microdilution panels for S. pneumoniae AST (4, 7). In our study, most of the MiEs occurred when the reference method classified an isolate as susceptible or intermediate but the Trek Sensititre panel classified it as intermediate or resistant; thus, the Trek Sensititre panel appears to “overcall” resistance, resulting in no resistant isolates being falsely identified as susceptible. Therefore, although the Trek Sensititre panel showed relatively low error rates for the three key antimicrobial agents, it is an acceptable method for S. pneumoniae susceptibility testing, in conjunction with its ease of use and reagent storage conditions (at room temperature).

In summary, we found that agar dilution was not an optimal method for S. pneumoniae AST. In contrast, the Trek Sensititre broth microdilution panel appears to be an acceptable commercial method for routine S. pneumoniae AST, although minor errors may occur for ceftriaxone, penicillin, and meropenem.

Footnotes

Published ahead of print on 1 December 2010.

REFERENCES

  • 1. Clark R. B., Lewinski M. A., Loeffelholz M. J., Tibbetts R. J. 2009. Cumitech 31A: verification and validation of procedures in the clinical microbiology laboratory. ASM Press, Washington, DC [Google Scholar]
  • 2. CLSI 2009. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. M7-A8 CLSI, Wayne, PA [Google Scholar]
  • 3. CLSI 2009. Performance standards for antimicrobial susceptibility testing. M100-S19 CLSI, Wayne, PA [Google Scholar]
  • 4. Guthrie L. L., Banks S., Setiawan W., Waites K. B. 1999. Comparison of MicroScan MICroSTREP, PASCO, and Sensititre MIC panels for determining antimicrobial susceptibilities of Streptococcus pneumoniae. Diagn. Microbiol. Infect. Dis. 33:267–273 [DOI] [PubMed] [Google Scholar]
  • 5. International Organization for Standardization 2007. ISO 20776-2:2007(E). Clinical laboratory testing and in vitro diagnostic test systems. Susceptibility testing of infectious agents and evaluation of performance of antimicrobial susceptibility test devices. Part 2: evaluation of performance of antimicrobial susceptibility test devices. International Organization for Standardization, Geneva, Switzerland [Google Scholar]
  • 5a. NCCLS 2000. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically—fifth ed.: approved standard M7-A5. NCCLS, Wayne, PA [Google Scholar]
  • 6. Reynolds R., Shackcloth J., Felmingham D., MacGowan A. 2003. Comparison of BSAC agar dilution and NCCLS broth microdilution MIC methods for in vitro susceptibility testing of Streptococcus pneumoniae, Haemophilus influenzae and Moraxella catarrhalis: the BSAC Respiratory Resistance Surveillance Programme. J. Antimicrob. Chemother. 52:925–930 [DOI] [PubMed] [Google Scholar]
  • 7. Tenover F. C., Baker C. N., Swenson J. M. 1996. Evaluation of commercial methods for determining antimicrobial susceptibility of Streptococcus pneumoniae. J. Clin. Microbiol. 34:10–14 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES