Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2012 Apr 1.
Published in final edited form as: Nucl Med Biol. 2010 Dec 3;38(3):353–362. doi: 10.1016/j.nucmedbio.2010.10.002

Study of an image-derived SUV and a modified SUV using mouse FDG-PET

Xiujuan Zheng 1, Chin-Lung Yu 2, Wei Sha 3, Caius Radu 4, Sung-Cheng Huang 5, Dagan Feng 6
PMCID: PMC3078337  NIHMSID: NIHMS245882  PMID: 21492784

Abstract

Introduction

Standard uptake value (SUV) is calculated without consideration of the differences in plasma FDG clearance. Its variability can be affected by changes of the amount of excreted FDG by renal function. Moreover, the estimation of SUV is quite sensitive to errors in the measurements of body weight and injected dose. This study aims to develop an image-based method to obtain an image-derived SUV (iSUV) and a modified SUV (mSUV) to overcome these problems.

Methods

31 tumor-planted SCID mice were scanned in microPET at ~60min post FDG injection, and then scanned in microCT. Using image-based method, the body weight and injected dose were derived from the microPET/CT images to calculate iSUV. The volumes and the total activities of FDG within the bladder and the whole-body were also obtained to calculate mSUV. For the selected targets, the iSUVs and mSUVs were compared against their corresponding SUVs.

Results

Compared with SUV factor (injected dose/body weight), iSUV factor had an average percentage error of −0.7%. The linear regressions between SUV and iSUV had a slope of 0.99 with correlation coefficient of 0.95. Compared with SUV and iSUV, CV% of mSUV decreased while the tumor-to-background separation of mSUV increased.

Conclusions

Using this image-based method, the iSUV can replace SUV when the actual measurements were missing or unreliable. The mSUV can reduce the inter-subject variability and enhance the tumor-to-background separation in mouse FDG-PET studies.

Keywords: FDG-PET, standard uptake value, image-based method, tumor-to-background separation

Introduction

Positron emission tomography (PET) with 2-deoxy-2-[18F]fluoro-D-glucose (FDG) is a well established oncologic imaging technique which has beneficial effects on clinical routines and preclinical studies [1, 2]. In the quantitative analysis for FDG-PET image, standard uptake value (SUV) is frequently used to measure the tracer uptake as a means to differentiate malignancy [3]. However, SUV can be influenced by many factors including the length of uptake period, plasma glucose level, recovery coefficient, partial volume effects [4, 5] and the subject preparation [6]. In addition, the body weight may not be the best parameter in SUV calculation due to the lower FDG concentration in body fat. The lean body weight or surface area has been proposed as a substitute [7, 8]. Recently the impact of renal function variation has been studied in FDG-PET [9, 10], since SUV is regularly defined as the local target concentration normalized by injected dose per unit body weight without regard to the differences in plasma FDG clearance [11]. As a result, changing the amount of excreted FDG by renal function could affect the estimation of SUV, even if the glucose utilization in the rest of the body stays unchanged.

On the other hand, the subject’s body weight and injected dose, which are needed in the calculation of SUV, are measured separately from imaging. Generally, the body weight is measured by a scale, and the injected dose is measured by a dosimeter. Because it is difficult to redo these physic measurements sometimes, any missing records would lead to the failure in the calculation of SUV. This is a notable issue in the data reutilization especially for multidisciplinary and inter-institution sharing in education and scientific research, when the data were stored in a shared database such as Mouse Quantitation Program (MQP) [12]. The tumor-to-background ratio (TBR), which is calculated as the ratio of the tracer concentration in tumor over that in normal tissue, is sometimes applied as an index to quantify the tracer uptake without measurements of body weight and injected dose [13, 14]. However, to find a suitable normal tissue as background may be difficult for some oncologic imaging [15].

Hence, this study aims to explore a purely image-based method to estimate SUV and, further, to find a way to improve SUV variability that will eventually help the use of SUV for differentiation of malignant from benign lesions. This study is evaluated in a set of mouse FDG-PET studies.

Materials and Methods

1.1 Animals and Image Acquisitions

All animal experiments were conducted in compliance with the Animal Care and Use Program established by the Chancellor’s Animal Research Committee of UCLA. Thirty-one SCID mice (20-33g) were implanted with U251 tumor in the left flank at least one week before the investigation. These mice would undergo the microPET/CT scans after tumor volumes had reached 100 mm3. All the mice were fasted overnight (about 15 hours) before microPET and microCT studies. For each mouse, the body weight was measured using a standard lab weighting scale. The injected dose was measured by a dosimeter and corrected for the residue of dose remained in the syringe. At 60 minutes post intraperitoneal injection of FDG (4.81-5.92 MBq), a 10 minutes static PET scan was performed on a microPET Focus 220 tomography (Siemens Preclinical Solutions, Knoxville, TN), whose imaging field of view (FOV) is 51.2 mm diameter in transverse by 75.6 mm in axial. The spatial resolution at the center of FOV was 1.7 mm full width at half maximum (FWHM). The microPET images were reconstructed using 2D filtered back-projection with a voxel size of 0.4×0.4×0.8 mm3 in a 128×128×95 matrix. The PET image was corrected for scatter radiation, random coincidence, dead time and physical decay. After mircoPET imaging, a 7-10 minutes microCT scan was acquired on a MicroCAT II tomography (Siemens Preclinical Solutions, Knoxville, TN), which has an imaging field of 51.2 mm in diameter in transverse by 99.2 mm in the axial direction. The microCT image had a voxel size of 0.2×0.2×0.2 mm3 in 256×256×496 matrix. The animals were anesthetized (~2% isoflurane) and placed in an imaging chamber with a heating bed during the scans [16]. The imaging chamber was compatible with the microPET and microCT systems [17, 18]. The microCT image was aligned to the mircoPET image using a predetermined, geometric transformation matrix [19]. The aligned microCT image was used in PET reconstruction for attenuation correction [20].

In the image analysis, AMIDE software [21] was used to manually remove the imaging chamber and nose cone in the microCT image. In every mouse study, ellipsoidal ROIs were drawn for major organs (brain, heart, lung, liver, kidney and muscle) in the microPET image with the guide of the aligned microCT image. To reduce the effects of ROI delineation, the ellipsoidal ROIs for the same major organ were of the same size and were placed in the core part of the organ for all mouse studies. The tumor volume for each mouse was large enough for manually delineating the ellipsoidal ROI on the left flank where the implanted tumor was. The size of the tumor ROI depended on the shape of the lesions in the microPET/CT images. The average and maximum values of tracer concentrations in these ROIs were calculated for further quantitative analysis. The bladder ROIs were also manually delineated with the visual guide from microCT and microPET images by an experienced operator. The volumes of the manually drawn bladder ROIs were used as the standard for validation of the bladder volumes derived from the automatic image-based method described as follows.

1.2 Image-Based Methods for Calculating iSUV and mSUV

An automatic image-based method was explored to obtain the whole-body mask (composed of the bone mask, body mask and lung mask) and bladder mask using the microCT and microPET images. These masks were used to locate the corresponding targets in the microPET and microCT images, and then to get the parameters for calculating iSUV and mSUV. Figure 1 shows the flow chart of this method. The entire process was described as follows.

Figure 1.

Figure 1

The flow chart of image-based method for the estimation of iSUV and mSUV.

In the reconstructed microCT image (in Hounsfield units, HU), the threshold method was firstly utilized to segment the mouse body from background using an empirical threshold window of [−250, 4000] HU. That is, the voxels with intensity values higher than −250 HU and lower than 4000 HU were considered as belonging to the rough body mask. Within the rough body mask, the bone tissue was further segmented using another empirical threshold window of [400, 4000] HU. As a result, the voxels with intensity values higher than 400 HU and lower than 4000 HU were set as the rough bone mask. These two rough masks were saved as two binary images in the same matrix dimension as the microCT image.

Secondly, the “morphological opening” operation (Matlab Image Processing Toolbox™) was applied to smooth the boundary of the rough body mask and bone mask. The voxels within the body part that have low density (i.e., lung, trachea and the gas in the alimentary system) were excluded in the rough body mask at the first step. The “morphological fill” operation (Matlab Image Processing Toolbox™) was then applied to fill the holes in the body mask to obtain a whole-body mask that included all the voxles within the body. The lung mask was then derived by subtracting the whole-body mask from the rough whole-body mask in the upper body. The lung mask was then refined by using the “morphological opening” operation to remove the artifacts of the trachea. Using these masks, the microCT image of the whole mouse was separated into three regions, namely soft tissue, lung and bone.

The total voxel numbers of the body, bone and lung regions were counted, and the volumes were calculated by multiplying the voxel number with the voxel volume (0.008mm3). The volumes of these different regions derived from microCT image were used to estimate the subjects’ body weight. The whole-body mask was re-sampled into the image voxel matrix of the aligned microPET to define the whole-body region on the microPET image. The total radioactivity in the whole-body was calculated as the sum of the values of the voxels within the whole body mask in the microPET image times the microPET voxel volume.

In the static microPET image, the bladder had a high radioactivity concentration, which was different from the surrounding tissues, so the max gradient searching method was applied to derive the bladder mask automatically. This method consisted of four steps.

Step 1: Define the bottom region of the microPET image which contained bladder;

Step 2: Calculate the gradient at each voxel within the predefined bottom region as:

G(x,y,z)=(C(x,y,z)x)2+(C(x,y,z)y)2+(C(x,y,z)z)2 (1)

where C(x, y, z) is the value of radioactivity at voxel (x, y, z).

Step 3: Any voxel with gradient greater than 90% of the maximal gradient within the bottom region would be marked as the bladder boundary. All voxels within the bottom region with image values above the minimal value of bladder boundary voxels were used to form the bladder mask image.

Step 4: Apply the “morphological opening” and “morphological fill” operations to the bladder mask image to remove any residual artifact.

The total bladder volume was calculated by multiplying the voxel number of the bladder mask by the microPET voxel volume (0.4×0.4×0.8 mm3). The total radioactivity in the bladder was calculated as the sum of the radioactivity concentrations for all voxels located within the bladder mask in the microPET image multiplied with the microPET voxel volume.

1.3 Image-derived SUV and Modified SUV

A. Image-derived SUV (iSUV)

iSUV was calculated using a similar formula as SUV, except that the body weight (BW) and injected dose (ID) were estimated from the microCT and microPET images by the proposed imaged-based method described in 1.2. The equation is shown in (2).

iSUV=CiIDθBWθ (2)

where IDθ and BWθ are the estimated injected dose and body weight obtained by the imaged-based method. The term in the denominator is referred to as the iSUV factor; Ci is the average or maximum value of tracer concentration within the target ROI.

The iSUV calculated by average tracer concentration within an ROI was noted as iSUVmean. Similarly, the iSUV estimated by the maximum ROI value was recorded as iSUVmax.

The estimated body weight was calculated as the sum of weights of the soft tissue, bone and lung regions as given in (3). The specific densities of soft tissue, bone and lung regions were assumed as 1, 1.14 and 0.28 g/ml, respectively [22, 23].

BWθ=Vnb×Dnb+Vb×Db+Vl×Dl (3)

where BWθ is the estimated body weight; Vnb is the soft tissue volume; Vb is the bone volume; Vb is the lung volume; Dnb, Db and Dl are the specific densities of soft tissue, bone and lung, respectively.

The estimated injected dose was calculated as the sum of the radioactivity within the whole body. Since the FOV of microPET is smaller than that of the microCT, some parts of the animal were not imaged by the microPET. The average voxel radioactivity of the part outside the microPET image was assumed to be equal to the average voxel radioactivity of the part inside the microPET image. Thus, the injected dose was calculated by (4).

IDθ=(Ain+Aout)×CF=(AinVwb_in×Vwb)×CF (4)

where IDθ denotes the estimated inject dose,Ain is the total radioactivity of the body within the microPET image, and Aout is the radioactivity of the part outside the microPET image (but inside the corresponding microCT image). Vwb_in is the body volume within microPET image. Vwb is whole-body volume in microCT image, and CF is the calibration factor that converts radioactivity unit from PET units to MBq/ml.

B. Modified SUV (mSUV)

Because the amount of the extracted tracer by renal function varied among subjects, the variability of SUV can be influenced by the tracer excretion. In order to reduce the variability of SUV, mSUV was calculated based on the target concentration normalized by the total uptake of the whole-body, excluding the amount excreted into the bladder. If the average body density was assumed to be 1 g/cm3, the body volume could be used instead of body weight in the calculation. Thus, mSUV can be calculated by (5). The term in the denominator of (5) will be referred to as the mSUV factor. mSUV calculated by average value of tracer concentration was noted as mSUVmean. Similarly, the mSUV estimated by the maximum value was recorded as mSUVmax.

mSUV=Ci(AwbAbl)(VwbVbl) (5)

where Ci is the average or maximum value of the tracer concentration within the target ROI (in PET units); Awb and Abl are the total radioactivity in the whole-body and the bladder; Vwb and Vbl are the volumes of the whole body and the bladder, respectively.

1.4 Statistical Analysis

A. iSUV Validation

The estimated body weight and injected dose were compared with the actual measurements of body weight and injected dose. The percent errors of the estimated body weight and injected dose compared with the actual measurements were calculated. The iSUV factor, which is defined as the ratio between the estimated body weight and injected dose, was compared with the SUV factor calculated by the actual measurements. The percent errors of the 1/(iSUV factor) to 1/(SUV factor) were also calculated. The linear regression between iSUV factor and SUV factor were performed to evaluate their correspondence. The significant difference between the iSUV factor and SUV factor was evaluated by paired t-test. A p-value <0.05 was considered statistically significant.

Moreover, iSUVmean was validated by the SUV calculated by the average value of tracer concentration within the ROI (SUVmean), and iSUVmax was validated by the SUV calculated by the maximum value within the ROI (SUVmax). The average and standard deviation (SD) of these two sets of SUV and iSUV were compared, and linear regressions were performed both for iSUVmean vs. SUVmean and iSUVmax vs. SUVmax for all target ROIs (major organs and tumor). Paired t-test was also performed to detect the significant difference between the iSUV and the SUV for each target ROI.

B. mSUV Evaluation

mSUVmean was compared with iSUVmean and SUVmean, while mSUVmax was compared with iSUVmax and SUVmax. The average and standard deviation (SD) of the two sets of mSUV, iSUV and SUV were calculated for all target ROIs. The coefficient of variation (CV%) was used as a measure to evaluate the inter-subject variability. The tumor-to-background separations for these two sets of mSUV, iSUV and SUV were quantified by the Mahalanobis distance that is defined in mathematical classification theory as given by (6).

D=meantmeanbSDt2+SDb2 (6)

where D is the Mahalanobis distance; mean is the population average value of SUV/iSUV/mSUV for tumor or background; SD is the standard deviation of SUV/iSUV/mSUV for tumor or background among the animals.

Experimental Results

1.5 iSUV vs. SUV

In the statistical analysis, average percent errors of the estimated body weight and injected dose to their actual measurements were respectively −9.5±7.3% and −8.5±11.5%, which are shown in Figure 2 by box-and-whisker diagrams. Considering the reciprocal of iSUV factor, its average percent error to the reciprocal of SUV factor was about −0.7±14.3%. Figure 3 shows the linear regression for 31 pairs of iSUV factor and SUV factor. The corresponding regression equation was y=0.83x+0.037 with R2=0.44. Paired t-test did not show significant difference between SUV factor and iSUV factor.

Figure 2.

Figure 2

Box-and whisker diagrams of estimated error percentage of body weight (BW), injected dose (ID) and 1/iSUVfactor. Box height shows inter-quartile range (IQR). The line in the box is for the median. Whiskers indicate the largest observation (minimum to maximum).

Figure 3.

Figure 3

The linear regression performed for all 31 pairs of SUV factor and iSUV factor. The open circle symbols note the scatter plot of the 31 pairs of iSUV factor and SUV factor. The red solid line is the regression line.

The iSUVmean and iSUVmax for each major organ were validated by their corresponding SUVmean and SUVmax. Figure 4 shows the comparisons using box-and-whisker diagrams. The average and standard deviation of SUVmean and iSUVmean for the defined ROIs (major organ and tumor) are listed in Table 1, and those for SUVmax and iSUVmax are listed in Table 2. Figure 5 (a) shows the linear regression between SUVmean and iSUVmean of all major organs for all mouse studies. The linear regression line was y=0.99x+0.018 with R2 =0.95. Similarly, Figure 5 (b) shows the plot of SUVmax vs. iSUVmax. The linear regression line was y=0.99x+0.027 with R2 =0.95. For each major organ, the paired t-test did not detect any significant difference between SUVmean and iSUVmean. The same results were found between SUVmax and iSUVmax.

Figure 4.

Figure 4

The comparisons between two sets of SUV and iSUV, SUVmean vs. iSUVmean and SUVmax vs. iSUVmax, for brain (a), heart (b), lung (c), liver (d), kidney (e), muscle (f), and tumor (g) shown by box-and-whisker diagram. In each sub-figure, box height shows inter-quartile range (IQR). The line in the box is for the median. Whiskers indicate the largest observation (minimum to maximum).

Table 1.

The comparison of SUVmean, iSUVmean and mSUVmean by mean, standard deviation and coefficient of variance

N=31 SUVmean iSUVmean mSUVmean
Mean SD CV% Mean SD CV% Mean SD CV%
brain 4.14 1.00 24% 4.14 1.09 26% 5.14 1.11 22%
heart 0.71 0.26 37% 0.70 0.24 35% 0.88 0.32 36%
lung 0.41 0.06 15% 0.40 0.06 14% 0.50 0.05 10%
liver 0.51 0.08 16% 0.50 0.08 15% 0.63 0.09 15%
kidney 1.76 0.42 24% 1.77 0.51 29% 2.21 0.58 26%
muscle 0.29 0.08 27% 0.29 0.09 31% 0.36 0.10 27%
tumor 2.05 0.55 27% 2.06 0.63 30% 2.55 0.65 25%
*

For each ROI, the lowest CV% among SUV, iSUV, and mSUV for each ROI is shaded.

Table 2.

The comparison of SUVmax, iSUVmax and mSUVmax by mean, standard deviation and coefficient of variance

N=31 SUVmax iSUVmax mSUVmax
Mean SD CV% Mean SD CV% Mean SD CV%
brain 5.67 1.31 23% 5.67 1.44 25% 6.90 1.41 20%
heart 1.10 0.42 39% 1.08 0.38 36% 1.32 0.48 36%
lung 0.58 0.09 15% 0.58 0.08 14% 0.71 0.08 11%
liver 0.67 0.13 19% 0.67 0.14 21% 0.82 0.15 18%
kidney 2.35 0.62 27% 2.37 0.73 31% 2.90 0.83 29%
muscle 0.70 0.18 26% 0.70 0.18 25% 0.85 0.21 24%
tumor 2.48 0.64 26% 2.48 0.68 27% 3.02 0.70 23%
*

For each ROI, the lowest CV% among SUV, iSUV, and mSUV for each ROI is shaded.

Figure 5.

Figure 5

The linear regression plots between SUV and iSUV. (a) Linear regression is performed for 217 (7×31, 7 major organs for 31 mice studies) pairs of iSUVmean and SUVmean. (b) Linear regression is performed for 217 (7×31, 7 major organs for 31 mice studies) pairs of iSUVmax and SUVmax. The black dot symbols note the scatter plot of pairs of iSUV and SUV. The red solid line is the regression line.

1.6 mSUV vs. SUV

The bladder volume estimated by the automatic image-based method was compared with that obtained from manual definition. Figure 6 shows the linear regression between the two estimates of bladder volume. The linear regression line was expressed as y=1.10×−4.25. The regression slope was close to 1 with R2=0.97.

Figure 6.

Figure 6

The validation of bladder volume. The estimated bladder volume obtained by the automatic image-based method is compared to the value obtained by manual method with visual support from microCT image. The open circle symbols denote the scatter plot of the pairs of automatic and manual results. The red solid line is the linear regression line.

The averages, SDs and CV% for SUVmean, iSUVmean and mSUVmean in each defined major organs ROI are shown in Table 1. Those metrics for SUVmax, iSUVmax and mSUVmax are given in Table 2. From these results, the CV% of mSUVmean decreased by about 1-5% in the brain, heart, lung and liver compared to those of SUVmean, while the CV% of mSUVmax decreased by about 1-4% in the comparison with SUVmax. For the tumor ROI, the CV% of mSUVmean was 2% lower than that of SUVmean and the CV% of mSUVmax was 3% lower than that of SUVmax. Compared with iSUVmean, the CV% of mSUVmean was decreased by about 1-5% in major organs and tumor, except in the heart the CV% of mSUVmean was increased by about 1%. Similar results were observed in the comparisons between mSUVmax and iSUVmax..

Tumor-to-background separation was increased by using mSUV because of the decrease in the variances of the mSUV values in the tumor and background regions. In this study, lung, liver and muscle were chosen as the background. The Mahalanobis distance (D) was used to quantify the tumor-to-background separation. From the D values given in Table 3, among SUVmean, iSUVmean, and mSUVmean the highest D values were obtained from mSUVmean regardless of the background used. Figure 7 (a) shows the tumor-to-background separation distribution of SUVmean, iSUVmean and mSUVmean. The result indicated that mSUVmean had the largest separation between the tumor and background, which was about 5% and 20% higher in D values than SUVmean and iSUVmean. Similar results were derived by comparing the tumor-to-background separations among SUVmax, iSUVmax and mSUVmax (shown in Table 3). It is clear that mSUVmaxhad the largest separation which is about 8% and 17% higher in D values than SUVmax and iSUVmax. Figure 7 (b) shows the tumor-to-background separation distribution of SUVmax, iSUVmaxand mSUVmax. In this figure, a threshold of 1 (gray dash line) is given as an example to separate tumor form background. As a result, using SUVmax and iSUVmax could both yield some false positives, while mSUVmax could separate tumor from background without error.

Table 3.

Comparison of Mahalanobis distance to evaluate SUV, iSUV and mSUV for tumor-to-background separation

Mahalanobis distance* (D)
tumor to lung tumor to liver tumor to muscle
SUVmean 2.99 2.80 3.18
iSUVmean 2.61 2.44 2.77
mSUVmean 3.15 2.94 3.34
SUVmax 2.95 2.78 2.69
iSUVmax 2.77 2.60 2.53
mSUVmax 3.27 3.06 2.96
*

The higher D value means more separation between tumor and background. The highest D values for each background reference are shaded.

Figure 7.

Figure 7

Plot of tumor and background (lung, liver and muscle) SUVs, iSUVs and mSUVs. The values for tumor (square), lung (circle), liver (triangle) and muscle (cross) are grouped for SUV, iSUV and mSUV to compare the tumor-to-background separations (please see the Table 3 for quantitative measurements of the separation). (a) This figure shows that the tumor-to-background separation is increased using mSUVmean compared with using SUVmean and iSUVmean. A threshold of 1 (gray dash line) is given as an example to separate tumor from background in this figure. (b) This figure shows that the tumor-to-background separation is increased using mSUVmax compared with using SUVmax and iSUVmax. A threshold of 1 (gray dash line) is given as an example to separate tumor from background in this figure.

The automatic image-based method was run in Matlab 7.0. It took ~38 seconds to do all the calculations for one animal using a Mac with OS X 10.5.5 (2.4GHz Intel Core 2 Duo, 4GB SDRAM).

Discussion

This study developed a robust and automatic image-based method for calculating iSUV which can be used as a substitute of the regular SUV in mouse FDG PET studies, when the body weight or injected dose is not available or unreliable. Based on this image-based method, a modification of SUV was applied to reduce the inter-subject variability with regards of the tracer excretion variability among subjects. The microPET/CT images used in the image-based method were obtained in separate scanners. A hardware registration method was applied to minimize the error in the alignment of two modalities images [19]. Moreover, the scans were performed when the implanted tumors had relatively large volumes (>100 mm3). This also allowed the reliable PET-to-CT alignment for the tumor lesions.

The results of the mouse studies indicated that the average percent error of estimated body weight was about −9.5% and that of estimated injected dose was about −8.5%. In the worst case, the percentage errors of the estimated body weight and injected dose were −22.5% and −35.1%, respectively. The underestimations of the body weight and the injected dose, however, occurred together, thus resulted in much less error in the ratio (e.g., the iSUV factor). There are several error sources that might affect the accuracy of the estimations using the image-based method. First is the incomplete coverage of the whole body in the microCT and the microPET images, as shown in Figure 8 (a) and (b). In this case, part of the tail and the hind legs were missing because of the limited field of view in microCT. The missing parts are manually sketched out in Figure 8 (c). In addition, the radioactivity outside the field of view of the microPET would introduce out- of-field scatter and influenced the quantitative accuracy of the image values [24]. The accuracy of the injected dose estimate also relied on that of the PET data correction, including the normalization, attenuation correction, scatters correction, random correction, dead-time correction, and the calibration. Urine leakage from the bladder during the microPET scan would also impact the injected dose estimation.

Figure 8.

Figure 8

The microCT and microPET images and the missing body parts in the images. (a) the microCT image has a field of view of 51.2mm×51.2mm×99.2mm. The microCT image voxel size is 0.2mm×0.2mm×0.2mm in a 256×256×496 matrix. (b) The microPET image has a field of view of 51.2mm×51.2mm×75.6mm. The microPET image voxel size is 0.4mm×0.4mm×0.796mm in a 128×128×95 matrix. (c) The fused mircoPET-CT image. In this figure, the contour of the whole body, included the missing parts of the mouse body, is sketched out by red solid line.

Another source of error is in the assumed values of specific gravity (i.e., density) for various tissues. The densities used in the body weight estimation were assumed uniform in each type of tissue, which were grouped into soft tissue, bone and lung. The lung has only a small impact on the body weight estimation, and its density could be assumed to be zero without great effect on the estimated body weight.

We expect that the major improvement in the accuracy of estimation is to position more consistently the mouse in the imaging chamber, and to fit the entire body in the field of view. This may reduce the influence of the missing regions in the estimation of body weight and injected dose by the image-based method. The studies using the newer microPET scanners (e.g., Inveon Dedicated PET System), which have larger FOVs, are expected to have less variability by covering the entire animal within the FOV. Also, the conversion of the attenuation coefficient from microCT images to that of microPET for attenuation correction could also be improved to give a better estimate of the injected dose. However, the correlated underestimations of body weight and injected dose did not introduce a large estimation error to the iSUV factor, of which the percent error was about −0.7% in average compared with SUV factor. Including all 217 pairs of iSUV and SUV (7 major organs for 31 mice studies) in linear regression, the results in Figure 5 (a) and (b) indicated that iSUV closely match the SUV (R2=0.95), either for iSUVmean vs. SUVmean or iSUVmax vs. SUVmax. These data demonstrated the validity of the iSUV from the automatic image-based method in this study.

In the future, the body surface area could be obtained directly from the microCT image (e.g., by a marching cubes algorithm [25]), and it could be used as the normalization factor instead of body weight. To use of body surface area has been reported in some studies to be preferable in SUV calculation [7], e.g. for obese subjects with increased fraction of body fat.

On the other hand, mSUV was designed to improve SUV reliability for accurate diagnosis of tumor malignancy by removing the variability of tracer excretions among subjects. To a certain extent, mSUV could be considered as the SUV with FDG excretion correction. Bladder voiding performed immediately before a FDG PET scan as part of the animal preparation procedure can achieve the same objective, and can be considered to be equivalent to replacing the image processing procedures for calculating the accumulated bladder radioactivity. However, the renal pelvis beside bladder usually also contained some urine. This could influence the mSUV in the kidney ROI. As a result, CV% of mSUV in kidney was found to be 2% higher than that of SUV both for mSUVmean vs. SUVmean and mSUVmax vs. SUVmax. It is assumed in this study that the amount of FDG in renal pelvis is much less than that in bladder, andmSUV mainly considered the influence of the FDG excreted into bladder. In the future, the influence from the renal pelvis could be removed in the calculation of mSUV by defining kidney ROIs to exclude the renal pelvis.

In common practical application, a predefined threshold of SUV is usually set for determining malignancy. That is, the lesion would be regarded as malignant when its SUV is higher than the threshold. Therefore, to enlarge the separation between the tumor and background is expected to be helpful for defining a threshold to detect the malignancy and improving the accuracy of diagnosis. mSUV is shown to increase the separation between the tumor target and background in Figure 7. Especially, mSUVmax showed its advantage in eliminating the false positive compared with SUVmax and iSUVmax. The Mahalanobis distance was adopted to measure the separation. The results indicated that the tumor to background (liver, lung or muscle) separation all increased using mSUV compared with SUV in light of the higher D values.

The amount of reduction in CV% of tumor mSUV shown in this study is not very large, since all the animals studied were quite uniform in size, in their preparations, and in the study procedure. When the physiological condition of the animals is more variable and the study condition is less uniform, the amount of improvement in mSUV is expected to be increased. Also, other body tissues that have large population variability in FDG extraction could also be excluded, just like the bladder that is addressed in this study. That is, the radioactivity taken up in these tissues and the corresponding volume/weight could be taken out of the injected dose and the body weight, respectively, in the calculation of mSUV. The reliability of the resulted mSUV is expected to be further improved, though further work is needed to demonstrate the amount of improvement in practical situations.

The present study is focused on mouse FDG PET study to improve the reliability of SUV. The underlying concept of mSUV can be extended to whole-body human FDG PET studies. Although the effectiveness of the approach for improving SUV reliability in these human studies needs many future investigations, results from the present study clearly indicated its feasibility and potentials.

Conclusion

This investigation demonstrated that, for mouse FDG PET studies, it is feasible to use microCT and microPET images to calculate iSUV, which could be used as a semi-quantitative index to substitute for SUV when directly measured body weight and/or injected dose are missing or unreliable.

The mSUV is shown to reduce the inter-subject variability of SUV and increase the tumor-to-background separation. Therefore, with further clinical investigations mSUV has the potential to be useful for increasing the accuracy to separate malignancy from benign lesions in FDG PET studies.

Acknowledgement

We thank David Truong, David Vu and Weber Shao at Department of Molecular & Medical Pharmacology, UCLA for their IT supports and the staffs at the Crump Institute for Molecular Imaging for their technical assistances in small animal studies. This research was partially supported by NIH R01-EB001943, P50 CA086306, DOE contract DE-FG02-06ER64249, PolyU and ARC grants.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Contributor Information

Xiujuan Zheng, Department of Electronic Information Engineering, the Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong.

Chin-Lung Yu, Department of Molecular and Medical Pharmacology, David Geffen School of Medicine at UCLA, University of California Los Angeles, CA, USA.

Wei Sha, Department of Molecular and Medical Pharmacology, David Geffen School of Medicine at UCLA, University of California Los Angeles, CA, USA.

Caius Radu, Department of Molecular and Medical Pharmacology, David Geffen School of Medicine at UCLA, University of California Los Angeles, CA, USA.

Sung-Cheng Huang, Department of Molecular and Medical Pharmacology, David Geffen School of Medicine at UCLA, University of California Los Angeles, CA, USA.

Dagan Feng, Department of Electronic Information Engineering, the Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong; School of Information Technologies, the University of Sydney, Sydney, Australia.

References

  • [1].Kubota K. From tumor biology to clinical PET: a review of positron emission tomography (PET) in oncology. Ann Nucl Med. 2001 Dec;15(6):471–486. doi: 10.1007/BF02988499. [DOI] [PubMed] [Google Scholar]
  • [2].Wood KA, Hoskin PJ, Saunders MI. Positron emission tomography in oncology: a review. Clin Oncol (R Coll Radiol) 2007 May;19(4):237–255. doi: 10.1016/j.clon.2007.02.001. [DOI] [PubMed] [Google Scholar]
  • [3].Huang SC. Anatomy of SUV. Standardized uptake value. Nucl Med Biol. 2000 Oct;27(7):643–646. doi: 10.1016/s0969-8051(00)00155-4. [DOI] [PubMed] [Google Scholar]
  • [4].Hamberg LM, Hunter GJ, Alpert NM, Choi NC, Babich JW, Fischman AJ. The Dose Uptake Ratio as an Index of Glucose Metabolism: Useful Parameter or Oversimplification? J Nucl Med. 1994 Aug 1;35(8):1308–1312. [PubMed] [Google Scholar]
  • [5].Keyes JW., Jr. SUV: Standard Uptake or Silly Useless Value? J Nucl Med. 1995 Oct 1;36(10):1836–1839. [PubMed] [Google Scholar]
  • [6].Fueger BJ, Czernin J, Hildebrandt I, Tran C, Halpern BS, Stout D, Phelps ME, Weber WA. Impact of Animal Handling on the Results of 18F-FDG PET Studies in Mice. J Nucl Med. 2006 Jun 1;47(6):999–1006. [PubMed] [Google Scholar]
  • [7].Kim CK, Gupta NC, Chandramouli B, Alavi A. Standardized Uptake Values of FDG: Body Surface Area Correction is Preferable to Body Weight Correction. J Nucl Med. 1994 Jan 1;35(1):164–167. [PubMed] [Google Scholar]
  • [8].Thie JA, Hubner KF, Isidoro FP, Smith GT. A weight index for the standardized uptake value in 2-deoxy-2-[F-18]fluoro-D-glucose-positron emission tomography. Mol Imaging Biol. 2007 Mar-Apr;9(2):91–98. doi: 10.1007/s11307-006-0068-x. [DOI] [PubMed] [Google Scholar]
  • [9].Minamimoto R, Takahashi N, Inoue T. FDG-PET of patients with suspected renal failure: standardized uptake values in normal tissues. Ann Nucl Med. 2007 Jun;21(4):217–222. doi: 10.1007/s12149-007-0012-4. [DOI] [PubMed] [Google Scholar]
  • [10].Laffon E, Cazeau AL, Monet A, de Clermont H, Fernandez P, Marthan R, Ducassou D. The effect of renal failure on 18F-FDG uptake: a theoretic assessment. J Nucl Med Technol. 2008 Dec;36(4):200–202. doi: 10.2967/jnmt.107.049627. [DOI] [PubMed] [Google Scholar]
  • [11].Allen-Auerbach M, Weber WA. Measuring response with FDG-PET: methodological aspects. Oncologist. 2009 Apr;14(4):369–377. doi: 10.1634/theoncologist.2008-0119. [DOI] [PubMed] [Google Scholar]
  • [12].Huang SC, Wu HM, Truong D, Zhang X, Prins M, Stout DB, Chatziioannou AF, Schelbert HR. A Public Domain Dynamic Mouse FDG MicroPET Image Data Set for Evaluation and Validation of Input Function Derivation Methods; Proc. Nuclear Science Symposium Conference Record, 2006; IEEE. 2006; Pages. [Google Scholar]
  • [13].Benz MR, Evilevitch V, Allen-Auerbach MS, Eilber FC, Phelps ME, Czernin J, Weber WA. Treatment Monitoring by 18F-FDG PET/CT in Patients with Sarcomas: Interobserver Variability of Quantitative Parameters in Treatment-Induced Changes in Histopathologically Responding and Nonresponding Tumors. J Nucl Med. 2008 Jun 13;:1038–1046. doi: 10.2967/jnumed.107.050187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Dirisamer A, Halpern BS, Schima W, Heinisch M, Wolf F, Beheshti M, Dirisamer F, Weber M, Langsteger W. Dual-time-point FDG-PET/CT for the detection of hepatic metastases. Mol Imaging Biol. 2008 Nov-Dec;10(6):335–40. doi: 10.1007/s11307-008-0159-y. [DOI] [PubMed] [Google Scholar]
  • [15].Mankoff DA, Muzi M, Zaidi H. Quantitative Analysis in Nuclear Oncologic Imaging. In: Zaidi H, editor. Quantitative Analysis in Nuclear Medicine Imaging. Springer Science+Business Media, Inc.; New York: 2006. pp. 494–536. [Google Scholar]
  • [16].Suckow C, Kuntner C, Chow P, Silverman R, Chatziioannou A, Stout D. Multimodality rodent imaging chambers for use under barrier conditions with gas anesthesia. Mol Imaging Biol. 2009 Mar-Apr;11(2):100–106. doi: 10.1007/s11307-008-0165-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [17].Stout DB, Chatziioannou AF, Lawson TP, Silverman RW, Gambhir SS, Phelps ME. Small animal imaging center design: the facility at the UCLA Crump Institute for Molecular Imaging. Mol Imaging Biol. 2005 Nov-Dec;7(6):393–402. doi: 10.1007/s11307-005-0015-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Chatziioannou A-X, Stout D, Silverman R. Book Method and apparatus for animal positioning in imaging systems, vol. WO/2006/020896, Series Method and apparatus for animal positioning in imaging systems. 2006. Method and apparatus for animal positioning in imaging systems. B. W. A01K 15/04 (2006.01), A47B 13/00 (2006.01), A61B 6/04 (2006.01) ed., City. [Google Scholar]
  • [19].Chow PL, Stout DB, Komisopoulou E, Chatziioannou AF. A method of image registration for small animal, multi-modality imaging. Phys Med Biol. 2006 Jan 21;51(2):379–390. doi: 10.1088/0031-9155/51/2/013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Chow PL, Rannou FR, Chatziioannou AF. Attenuation correction for small animal PET tomographs. Phys Med Biol. 2005 Apr 21;50(8):1837–1850. doi: 10.1088/0031-9155/50/8/014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Loening AM, Gambhir SS. AMIDE: a free software tool for multimodality medical image analysis. Mol Imaging. 2003 Jul;2(3):131–137. doi: 10.1162/15353500200303133. [DOI] [PubMed] [Google Scholar]
  • [22].Irwin R, Lin HV, Motyl KJ, McCabe LR. Normal Bone Density Obtained in the Absence of Insulin Receptor Expression in Bone. Endocrinology. 2006 Dec 1;147(12):5760–5767. doi: 10.1210/en.2006-0700. [DOI] [PubMed] [Google Scholar]
  • [23].Saw CB, Loper A, Komanduri K, Combine T, Huq S, Scicutella C. Determination of CT-to-density conversion relationship for image-based treatment planning systems. Medical Dosimetry. 2005;30(3):145–148. doi: 10.1016/j.meddos.2005.05.001. [DOI] [PubMed] [Google Scholar]
  • [24].Spinks TJ, Miller MP, Bailey DL, Bloomfield PM, Livieratos L, Jones T. The effect of activity outside the direct field of view in a 3D-only whole-body positron tomograph. Phys Med Biol. 1998 Apr;43(4):895–904. doi: 10.1088/0031-9155/43/4/017. [DOI] [PubMed] [Google Scholar]
  • [25].William EL, Harvey EC. Marching cubes: A high resolution 3D surface construction algorithm. Comput Graph. 1987;21(4):163–169. [Google Scholar]

RESOURCES