Optimization of Magnetic Field Sweep and Field Modulation Amplitude for Continuous-Wave EPR Oximetry

J Palmer a, LC Potter a, R Ahmad b,*
PMCID: PMC3086786  NIHMSID: NIHMS276438  PMID: 21334232

Abstract

For continuous-wave electron paramagnetic resonance spectroscopy, what settings of magnetic field sweep width and field modulation amplitude yield the best accuracy in estimated linewidth? Statistical bounds on estimation error presented in this work provide practical guidance: set the sweep width and modulation amplitude to 8 and 4 times the half-width half-maximum linewidth, Γ, respectively. For unknown linewidths in the range [Γmin, Γmax] the worst-case estimation error is minimized by using settings designed for Γmax. The analysis assumes a Lorentzian lineshape and a constant modulation amplitude across the extent of the irradiated paramagnetic probe. The analytical guidelines are validated using L-band spectroscopy with a particulate LiNc-BuO probe.

Keywords: EPR, Spectroscopy, Oximetry, Overmodulation, Cramér-Rao lower bound

1. Introduction

Electron paramagnetic resonance (EPR) is a spectroscopic method capable of detecting free radicals. Over the past several decades, EPR has found numerous applications in biology, chemistry, physics, and medicine [1]. For biological applications, such as in vivo oximetry [2, 3, 4], there exists a pressing need to accelerate data acquisition for EPR spectroscopy and imaging [5].

For continuous-wave (CW) EPR, it is a universal practice to apply magnetic field modulation and phase sensitive detection [6]. To improve signal-to-noise ratio (SNR), the modulation amplitude, Bm, is generally increased to a level where it approaches or exceeds the intrinsic half-width half-maximum (HWHM) linewidth, Γ, of the EPR probe. Such Bm values, however, introduce lineshape distortion. Although modulation-induced distortion itself is well characterized [7, 8, 9], a theoretical basis for selecting an optimal value of Bm has been missing from the EPR literature.

EPR oximetry [4] entails quantifying oxygen levels by measuring the oxygen-induced linewidth broadening of oxygen-sensitive EPR probes. For EPR oximetry, an allowable scan time is limited by application. Here, we ask the question, “How are magnetic field sweep width, ΔB, and field modulation amplitude, Bm, set for best sensitivity in inferring linewidth?” We use the Cramér-Rao lower bound (CRLB) to predict the standard deviation of linewidth estimation error as a function of Γ, ΔB, Bm, signal intensity d, and noise standard deviation σ. The CRLB sets a lower bound on the error standard deviation for any unbiased estimator. The experiment parameters, ΔB and Bm, are chosen to minimize this bound and hence provide the most accurate linewidth estimation for a given scan time. Our analysis is valid for a Lorentzian lineshape and a spatially-invariant field modulation amplitude across all the irradiated spins.

2. Theory and methodology

2.1. Assumptions and models

The following assumptions are employed in our analysis: (i) the value of Γ is unknown but resides in a known range, Γmin to Γmax; (ii) the center field of the lineshape is known; (iii) the signal intensity d is unknown and proportional to spin density; (iv) the unknown value of Bm is spatially uniform across all irradiated spins; (v) the field sweep rate is sufficiently slow to not fall into the rapid-scan regime [10]; (vi) and the lineshape is a modulation-distorted first-harmonic Lorentzian.

When the ratio of field modulation frequency and gyromagnetic ratio is negligible compared to Γ, the resulting modulation-distorted lineshape can be expressed using a simplified version of the model reported by Robinson et al. [8].

f(B;d,Γ,Bm)=Im(dBmα) (1)

where,

α=a22(1+1(Bm2a)2)Bm28,witha=B+jΓ

.

Here, f is the modulation-distorted first-harmonic Lorentzian lineshape, B is the applied magnetic field, j denotes the imaginary unit, and Im(·) represents the imaginary part. Figure 1 shows simulated f for various levels of modulation.

Figure 1.

Figure 1

Simulated first-harmonic Lorentzian lineshape (Eq. 1) in the presence of various levels of field modulation. Here, Γ = 1, center field is 458.25 G, and ΔB = 12 G. The peak signal intensity is maximum at Bm = 4.0Γ.

A scan is composed of M successive data points, called measurements, taken at regular intervals across the field sweep. The measured data are given by,

Yi=f(Bi;d,Γ,Bm)+Nifi+Ni, (2)

where

Bi=iΔBM1ΔB2, (3)

i ∈ {0, …, M − 1}, and Ni is additive white Gaussian noise [11]. The noise standard deviation, σ, is proportional to T, with T being the scan time. When comparing two measurements with different Bm and ΔB, the values of both M and T are fixed.

2.2. CRLB analysis

The CRLB is computed from the inverse of the Fisher information matrix, I [12]. Let p(Y⃗(x02223)θ⃗) be the probability density function of the data Y⃗ = [Y0, ⋯,YM−1] conditioned on parameters θ⃗, with θ⃗ = [d, Γ, Bm], and let EY⃗ [· ] denote the expectation over Y⃗. Then the elements of the Fisher information matrix according to Eq. 2 are as follows:

(I(θ))k,l=EY[2θkθllog(p(Y|θ))]=12σ2EY[iθkθl(Yifi)2]=1σ2i(θkfi)(θlfi). (4)

Equation 4 is valid for any signal model f and can also be extended, if required, to include other measurement parameters such as modulation frequency, unknown center field, and parametric baseline distortion. The CRLB on the standard deviation of the estimated linewidth Γ̂ is then the square-root of the second element along the diagonal of the inverse of I:

CRLBΓ^(d,Γ,Bm,ΔB,σ)=(I1)2,2. (5)

For brevity, the left side of Eq. 5 will be referred to as CRLBΓ̂. A combination of Bm and ΔB that yields minimum CRLBΓ̂ is selected as optimal. Noisy measurements, Y⃗, are then collected using these optimal values. The unknown linewidth is estimated via a least-squares curve fit of Y⃗ to the model in Eq. 1. The precise value of modulation amplitude, which is generally not known accurately, is jointly estimated along with the other unknown parameters, yielding,

θ^=argmini=0M1(fiYi)2, (6)

where, θ^=[d^,Γ^,B^m],, with , Γ̂, and m being the estimated values of signal intensity, HWHM linewidth, and modulation amplitude, respectively.

2.3. Experimental setup

For verification, findings from the CRLB analysis were compared to experimentally observed linewidth estimation errors. A small single crystal of LiNc-BuO [13], under anoxic conditions, was used to collect data on a CW L-band (1.28 GHz) system using a volume loop-gap resonator. The anoxic linewidth of the probe (0.315 G) was known in advance. Other system parameters were as follows: 2 mW radio frequency power; 0.16 G/W resonator efficiency [14]; 100 kHz field modulation frequency; 3.9 s scan time; and 1024 samples per scan.

3. Results

3.1. Theoretical and simulated

The value of CRLBΓ̂, computed from Eq. 5, varies as a function of d, Γ, Bm, ΔB, and σ. However, the location of the minimum does not vary with Bm/Γ and ΔB, at least not over the range of Γ experienced in EPR oximetry. Figure 2 shows how CRLBΓ̂ varies with Bm/Γ and ΔB. The minimum CRLBΓ̂ occurs at approximately

Figure 2.

Figure 2

Theoretical CRLB results. CRLBΓ̂ vs. ΔB/Γ and Bm/Γ, revealing the location of the minimum. The overall shape of the plot does not vary with Γ; that is, the minimum stays in the same location. Also, the location of the minimum does not vary with d and σ. Parameters used to create the plot are as follows: Γ = 1 G, d = 10, M = 1024, σ = 1.

ΔB=8Γ,andBm=4Γ. (7)

Because CRLBΓ̂ varies with the true Γ, which we only know to be in a given range (Γmin to Γmax), we desire to select ΔB and Bm such that the standard deviation of Γ̂ across the range is minimized. Figure 3 shows that, for fixed values of d and σ, estimator performance is better at lower values of Γ, and suggests that optimizing the parameters for Γmax will minimize the worst-case estimation error across [Γmin, Γmax].

Figure 3.

Figure 3

Best CRLBΓ̂ possible as a function of Γ, showing Γ from 0.18 G to 1.36 G, which corresponds to the range 0–160 mmHg when using a LiNc-BuO probe [15]. Note that the scale on the vertical axis is determined by d and σ; values are shown here for d = 10andσ = 1.

3.2. L-band spectroscopy

The intrinsic linewidth in our experiment was Γ = 0.315 G. A variety of scans were taken with ΔB = 2.5 G ≈ 8Γ while varying Bm; likewise, a variety of scans were taken with Bm ≈ 1.24 G ≈ 4Γ while varying ΔB. As mentioned earlier, both the scan time (T = 3.9 s) and the number of samples per scan (M = 1024) were kept fixed. Twelve scans were performed at each combination of ΔB and Bm.

Joint estimates of d, Γ, and Bm were calculated by curve fitting using Eq. 6. As expected, the estimated values m were different from the nominal Bm values set by the user but the difference was within 3%. Figures 4 and 5 show the results, along with comparison to the theoretical bound. The experimental results are consistent with the theory. For illustration, Figure 6 compares standard deviation in the estimation of Γ for three selected combinations of Bm and ΔB.

Figure 4.

Figure 4

Comparison of experimental and theoretical results, with Bm 1.24 G 4Γ. Top panel: standard deviation results; bottom panel: experimental results only, means and standard deviations.

Figure 5.

Figure 5

Comparison of experimental and theoretical results, with ΔB = 2.5 G 8Γ. Top panel: standard deviation results; bottom panel: experimental results only, means and standard deviations.

Figure 6.

Figure 6

Standard deviation of Γ̂ for three different combinations of Bm and ΔB. As predicted by theory, Bm = 4Γ and ΔB= 8Γ provide the lowest standard deviation of Γ̂. The relative difference in the scaling of experimental and CRLB values is due to the different values of d and σ associated with experimental and theoretical results.

4. Discussion

Previous discussions of modulation amplitude have focused on increasing traditional figures of merit such as SNR and peak-to-peak signal strength [16]. Figure 7 illustrates how using these figures of merit may not lead to optimal estimator performance. Selecting Bm to maximize signal power (sum-of-squares) does not lead to optimal performance, and selecting Bm to maximize peak signal strength may lead to optimal performance, but only if used in conjunction with the correct value of ΔB, i.e., ΔB = 8Γ.

Figure 7.

Figure 7

Comparison of several figures of merit as functions of Bm/Γ. The vertical scale on the left belongs to independently normalized signal power and peak signal amplitude, while the vertical scale on the right belongs to CRLBΓ̂. To compute CRLBΓ̂, we selected σ = 1, d = 10, M = 1024, and Γ = 1 G. Note that maximizing peak signal amplitude is a suitable surrogate metric for estimation performance only when sweep width is optimized, i.e., ΔB = 8Γ.

The presented CRLB sensitivity analysis determines values of Bm and ΔB that generate optimal results in terms of standard deviation of the estimated linewidth, which is achieved by selecting a combination of Bm and ΔB that minimizes CRLBΓ̂. The results (Eq. 7) were computed with M = 1024; the optimal parameters do not vary with M as long as the step-size of the field sweep is considerably smaller than ΔB. Also, the optimal parameter values of Bm and ΔB are independent of both signal amplitude d and noise standard deviation σ, which are positive scalar multipliers of CRLBΓ̂ that affect the scaling but not the location of the minimum (Figure 2). Because the scan time T does not alter the lineshape f but only changes the normalized noise standard deviation, i.e., σd1T, the optimal parameter values are also independent of the scan time, provided that the sweep rate does not fall into the rapid-scan regime.

Although we have used CRLBΓ̂ as a figure of merit, equivalent acquisition times can also be computed for comparing various selections of Bm and ΔB. For Bm = 4Γ, ΔB = 16Γ, for example, the value of 1/CRLBΓ̂ is 78.4% of the 1/CRLBΓ̂ value associated with Bm = 4Γ, ΔB = 8Γ (Figure 7). Since CRLBΓ^σd and σd1T, a 16Γ scan would require 62.4% longer acquisition time as compared to an 8Γ scan for equivalent accuracy in estimated linewidth. For reference, Figure 8 shows an experimental EPR spectrum, modulated with the optimal Bm, with the optimal sweep window denoted.

Figure 8.

Figure 8

An EPR spectrum collected on a CW L-band spectrometer with Γ ≈ 0.315 G, optimally modulated with Bm ≈ 1.24 G ≈ 4Γ. The highlight denotes the optimal sweep width ΔB = 2.5 G ≈ 8Γ.

As noted in Section 2.1, the actual value of Bm, in general, is not precisely known and varies slightly but unpredictably from the nominal value set the user. Therefore, setting the Bm = 4Γ might not be possible, but the smooth nature of CRLBΓ̂ vs. Bm curve ensures that a Bm ≈ 4Γ value would only result in a marginal drop in the performance. For applications in which Bm might be precisely controlled, the resulting optimal values of Bm and ΔB are significantly different from the ones mentioned above. Specifically, sweep width and field modulation amplitude should be set to 4.7Γ and 1.6Γ, respectively, when the value of Bm is precisely known and need not be estimated. The CRLB analysis for known Bm is similar to the one presented here and is omitted for brevity. The results for both the cases, i.e., known and unknown Bm, are summarized in Table 1.

Table 1.

Optimal parameters depending on whether Bm is estimated or known in advance.

Estimating d, Γ, and Bm Estimating d and Γ; Bm is known
ΔB 8Γ 4.73Γ
Bm 4Γ 1.57Γ

Although the reported results are valid only for the listed set of assumptions, a similar analysis can be performed for a different set of assumptions. For example, effects of modulation frequency, when not negligible, can be included by modifying f (Eq. 1) accordingly. Likewise, a similar analysis can be conducted when multiple modulation harmonics are simultaneously considered [17].

5. Conclusions

Experimental settings for field modulation amplitude, Bm, and sweep width, ΔB, should be set jointly for the best performance in CW EPR. We have used standard deviation of linewidth estimation error, as predicted by the Cramér-Rao lower bound, to select Bm and ΔB. For cases where Bm cannot be controlled precisely, the optimal parameters for a modulation-distorted first-harmonic Lorentzian are found to be ΔB = 8Γ and Bm = 4Γ, with Γ being the HWHM linewidth. We have also shown that, given Γ in a range from Γmin to Γmax, the standard deviation of the worst-case estimation of linewidth is minimized by designing for Γmax: ΔB = 8Γmax and Bm = 4Γmax. Further, for cases where Bm is precisely known, the preferred experimental settings result in a much narrower sweep width: ΔB = 4.7Γ and Bm = 1.6Γ

Acknowledgments

We thank Dr. Periannan Kuppusamy for use of laboratory equipment at the Davis Heart and Lung Research Institute. This work was supported by NIH grant EB008836.

This work was supported by NIH grant EB008836

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Eaton S, Eaton G, Berliner L, editors. Biomedical EPR, Part A: Free Radicals, Metals, Medicine and Physiology. Kluwer Academic; New York: 2004. [Google Scholar]
  • 2.Swartz H. In Vivo EPR (ESR): Theory and Applications. Ch. 15. Kluwer Academic; New York: 2003. Measurement of oxygen in vivo using EPR techniques; pp. 404–440. [Google Scholar]
  • 3.Elas M, Bell R, Hleihel D, Barth E, McFaul C, Bielanska CHJ, Pustelny K, Ahn K, Pelizzari C, Kocherginsky M, Halpern H. Electron paramagnetic resonance oxygen image hypoxic fraction plus radiation dose strongly correlates with tumor cure in FSa fibrosarcomas. Int J Radiat Oncol, Biol, Phys. 2008;71:542–549. doi: 10.1016/j.ijrobp.2008.02.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ahmad R, Kuppusamy P. Theory, instrumentation, and applications of electron paramagnetic resonance oximetry. Chem Rev. 2010:3212–3236. doi: 10.1021/cr900396q. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Subramanian S, Krishna M. Dancing with the electrons: time-domain and CW in vivo EPR imaging. Magn Reson Insights. 2008;2:43–74. doi: 10.4137/mri.s1131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Poole CP., Jr . Electron Spin Resonance: A Comprehensive Treatise on Experimental Techniques. Ch. 7. Dover; Mineola: 1997. pp. 259–288. [Google Scholar]
  • 7.Wahlquist H. Modulation broadening of unsaturated Lorentzian lines. J Chem Phys. 1961;35:1708–1710. [Google Scholar]
  • 8.Robinson B, Mailer C, Reese A. Linewidth analysis of spin labels in liquids: I. Theory and data analysis. J Magn Reson. 1999;138:199–209. doi: 10.1006/jmre.1999.1737. [DOI] [PubMed] [Google Scholar]
  • 9.Mailer C, Robinson B, Williams B, Halpern H. Spectral fitting: the extraction of crucial information from a spectrum and a spectral image. Magn Reson Med. 2003;49:1175–1180. doi: 10.1002/mrm.10474. [DOI] [PubMed] [Google Scholar]
  • 10.Joshi J, Ballard J, Rinard G, Quine R, Eaton S, Eaton G. Rapid-scan EPR with triangular scans and Fourier deconvolution to recover the slow-scan spectrum. J Magn Reson. 2005;175:44–51. doi: 10.1016/j.jmr.2005.03.013. [DOI] [PubMed] [Google Scholar]
  • 11.Som S, Potter L, Ahmad R, Kuppusamy P. A parametric approach to spectral-spatial EPR imaging. J Magn Reson. 2007;186:1–10. doi: 10.1016/j.jmr.2006.12.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Stoica P, Moses R. Spectral Analysis of Signals. Prentice Hall: Upper Saddle River; 2005. [Google Scholar]
  • 13.Pandian R, Parinandi N, Ilangovan G, Zweier J, Kuppusamy P. Novel particulate spin probe for targeted determination of oxygen in cells and tissues. Free Radical Biol Med. 2003;35:1138–1148. doi: 10.1016/s0891-5849(03)00496-9. [DOI] [PubMed] [Google Scholar]
  • 14.Freed J, Leniart D, Hyde J. Theory of Saturation and Double Resonance Effects in ESR Spectra. III. RF Coherence and Line Shapes. J Chem Phys. 1967;47:2762–2773. [Google Scholar]
  • 15.Pandian R, Dang V, Manoharan P, Zweier J, Kuppusamy P. Effect of nitrogen dioxide on the EPR property of lithium octa-n-butoxy 2,3-naphthalocyanine (LiNc-BuO) microcrystals. J Magn Reson. 2006;181:154–161. doi: 10.1016/j.jmr.2006.04.004. [DOI] [PubMed] [Google Scholar]
  • 16.Nielsen R, Robinson B. The effect of field modulation on a simple resonance line shape. Concepts Magn Reson. 2004;Part A 23A:38–48. [Google Scholar]
  • 17.Ahmad R, Som S, Kesselring E, Kuppusamy P, Zweier J, Potter L. Digital Detection and Processing of Multiple Quadrature Harmonics for EPR Spectroscopy. J Magn Reson. 2010;207:322–331. doi: 10.1016/j.jmr.2010.09.016. [DOI] [PMC free article] [PubMed] [Google Scholar]