Abstract
Iron oxide nanoparticles (IONPs) are used in various MRI applications as negative contrast agents. A major challenge is to distinguish regions of signal void due to IONPs from those due to low signal tissues or susceptibility artifacts. To overcome this limitation, several positive contrast strategies have been proposed. Relying on IONP T1 shortening effects to generate positive contrast is a particularly appealing strategy since it should provide additional specificity when associated with the usual negative contrast from T2* effects. In this paper, Ultrashort TE (UTE) imaging is shown to be a powerful technique which can take full advantage of both contrast mechanisms. Methods of comparing T1 and T2* contrast efficiency are described and general rules that allow optimizing IONP detection sensitivity are derived. Contrary to conventional wisdom, optimizing T1 contrast is often a good strategy for imaging IONPs. Under certain conditions, subtraction of a later echo signal from the UTE signal not only improves IONP specificity by providing long T2* background suppression, but also increases detection sensitivity, as it enables a synergistic combination of usually antagonist T1 and T2* contrasts. In vitro experiments support our theory and a molecular imaging application is demonstrated using tumor-targeted IONPs in vivo.
Keywords: UTE, iron oxide nanoparticle, sequence optimization, positive contrast
Introduction
Iron oxide nanoparticles (IONPs) have been used as exogenous labels in a wide variety of MRI applications (1,2). These particles are usually regarded as negative contrast agents because their strong magnetic moment causes protons diffusing through their vicinity to lose signal coherence (3,4), resulting in signal voids with commonly used T2- and T2*-weighted pulse sequences. A major challenge is the need to distinguish regions of signal void due to IONPs from those due to low signal tissues or susceptibility artifacts, particularly when the signal-to-noise ratio (SNR) is low (2). While a pre-contrast image can be used for comparison, the slow accumulation of IONPs in tissues limits the ability to acquire post-contrast images during the same session, requiring precise co-registration of the pre- and post-contrast images, acquired hours to days later, to permit meaningful interpretation. Alternative MR acquisition strategies are desirable that are specific to IONPs and simpler to acquire.
Several investigators have recently developed original acquisition strategies to generate positive contrast from IONPs. These techniques rely on different approaches, including: 1) exploiting the frequency shift induced by IONPs (off-resonance techniques (5–8)); 2) exploiting the perturbed magnetic field surrounding IONPs (gradient-compensation techniques (9,10)); or 3) using post-processing methods to identify these magnetic field inhomogeneities (such as susceptibility-gradient mapping (11), image cross-correlation, (12) and even quantitative susceptibility-imaging (13,14)). Interestingly, these approaches share a common feature with the usual T2- and T2*-based approaches, as they all exploit the magnetic field perturbations caused by IONPs. It might therefore be expected that these techniques would share similar limitations, the most likely being poor specificity regarding other sources of magnetic field perturbations or resonance frequency shift such as large-scale bulk magnetic susceptibility (BMS) effects (e.g., due to air/tissue interfaces) or imperfect B0 shimming and/or chemical shift (e.g., from fat/lipid present in the region of interest) as demonstrated by Farrar and coworkers (6).
A method of obtaining a positive contrast from IONPs that does not rely on magnetic field perturbations, is to exploit their T1 shortening effect using appropriate imaging sequences (15–18). However, because of the short T2 and T2* induced by IONPs, even at relatively low concentrations, the positive T1 contrast signature from IONPs is usually strongly attenuated at the minimum echo time (TE) attainable using typical clinical pulse sequences. Moreover, IONP clustering and compartmentalization often occur in vivo (e.g., in liver Kupffer cells), increasing the T2* reduction as compared to in vitro homogeneous suspensions (19–21). In addition, T1 and T2 effects appear to be reduced (20–23), and may quench at high concentration due to a limited water-exchange regime (24). These phenomena typically limit the ability to obtain a T1 effect from IONPs (21–23,25). In this situation, image weighting may be uncertain because the dominant contrast mechanism varies with TE and IONP concentration as well as the particle biodistribution within the sample.
Ultrashort echo time (UTE) pulse sequences (26) have been used to probe the T1 effect generated by IONPs (27–29) including studies at high concentration. This approach has shown great promise for in vivo applications. Indeed, the technique can reduce signal loss due to rapid signal decay of short T2 or T2* components. In contrast to usual “long” echo time sequences that require an echo formation to probe the MR signal, UTE sequences acquire signal directly during the free induction decay (FID) following a short RF pulse. The center-out radial k-space trajectory combined with a ramp-sampling strategy allows signal acquisition very soon after the excitation, thus minimizing the effect of rapid signal decay. Specially tailored RF pulses (30) can be used to reduce relaxation effects that occur during the RF pulse and to avoid the need of rephasing gradient lobe. On the hardware side, a fast transmit/receive switch ensures that minimal time is lost after spin excitation, prior to data acquisition. Strong gradient capabilities also allow reduction of the signal observation time to minimize blurring effects. Using this technique, typical nominal TEs of 8–12 μs can be obtained on clinical scanners (31).
Multiecho versions of UTE sequences extend this unique T1 contrast capability by providing perfectly registered datasets acquired at latter echo times, offering complementary T2* 1 contrast and the opportunity to generate an original positive contrast that combine T1 and T2* shortening in a synergistic way. These two effects are usually opposed with conventional pulse sequences because T1 shortening enhances MR signal while T2* shortening reduces it. However, when the signal from a later TE image is subtracted from the signal obtained on the UTE image, long T2* components are reduced and the UTE detection takes advantage of the T1 effect induced by IONPs (see Fig. 1). This subtraction technique (called SubUTE here) provides long-T2* signal suppression and, more generally, reverses T2* contrast; it is thus considered to be a positive contrast technique. Note that the introduction of T2* contrast provided by SubUTE makes it prone to reduced specificity regarding undesired magnetic field perturbations, similar to that of the more usual positive contrast methods described above. SubUTE has been successfully used for IONP detection by several investigators (27,32–34), although the synergistic effect has only received brief consideration (27).
Figure 1.

a) Schematic depiction of T1 and T2* contrast. After subtracting the long TE signal from the UTE signal, long T2* components are reduced whereas short T2* components are preserved. The positive contrast that results combines both T1 and T2* effects in a synergistic way. Corresponding contrast intensities are emphasized in b). The + sign indicates positive contrast whereas the − sign stands for negative contrast.
Expected improvements in detection specificity and sensitivity toward IONPs remain open for further investigation. The present study analyzes the balance between T1 and T2* effects obtained with IONPs, and provides criteria to understand and optimize image contrast. UTE and SubUTE detection of IONPs are studied theoretically and demonstrated experimentally. A proof of concept study for in vivo molecular imaging is performed, applying these techniques to tumor-targeted IONP detection.
Theory
In the following study, the MR contrast optimization is limited to the widely used RF-spoiled gradient echo (SPGR) sequence. Assuming the RF pulse duration is short compared to T2*, the SPGR MR signal is described by the well known steady state equation, which includes T1 and T2* as well as user controlled MR parameters (flip angle θ, repetition time TR, and echo time TE):
| [1] |
S0 is a sensitivity factor accounting for proton spin density, voxel size, total scan time, RF coil sensitivity, and reception gain. This factor may be spatially heterogeneous. We assume that it does not depend on other specified parameters (i.e., θ, TR, TE, T1 and T2*). Spin density is included in this global sensitivity factor since this paper aims to consider the contrast in tissue compartments in which the spin density is assumed to be fixed.
For reasons of simplicity Eq. 1 utilizes an exponential decay to describe the complex T2*-loss mechanism. In the general case this is not correct. The magnetic field heterogeneities which lead to T2* effects can be classified according to their spatial scale (35). Microscopic effects take place at the molecular level and are hence irreversible because of random motion (i.e. they equally affect T2 and T2*). They are usually considered to decay in an exponential way.
Macroscopic effects happen at a scale much larger than the voxel size and are typically due to poor shimming or air-tissue interfaces. As shown by Yablonskiy (35) these lead to a non-exponential decay. However, if the corresponding intravoxel dephasing is low (i.e. small voxel size and low bandwidth), deviation from the exponential model is reduced.
Finally mesoscopic heterogeneities have an intermediate length scale, smaller than the voxel size, but much larger than molecular distances. The boundary between microscopic and mesoscopic scales may be formulated as a transition from the motional narrowing regime (i.e. irreversible loss) to the static dephasing regime that is assumed to be satisfied on the mesoscopic scale (35). IONP induced heterogeneities fall into both microscopic and mesoscopic scales depending on the size of the particles. In the mesoscopic contrast regime, the exponential decay model is satisfied for IONPs as shown by Yablonskiy and Haacke (36).
UTE signal is ideally modeled with TE set to zero, to describe the creation of a pure T1-weighted image with no T2* contamination. The subtraction of a later echo signal from the UTE signal, SubUTE(TE) = S(UTE) − S(TE), leads to a composite image that provides hybrid T1 and T2*contrast (Fig. 1).
The tissue relaxation time dependence on IONP concentration is described with the usual linear relaxivity approximation:
| [2] |
where T10 and T20* are the unenhanced tissue relaxation times, r1 and r2* are the corresponding IONP relaxivities expressed in s−1.mM−1, and [IONP] denotes the iron concentration in mM.
Following the approach of Buxton et al. (37), we define β = (ΔT2*/T2*)/(ΔT1/T1) as an index of T2* over T1 relaxation effect efficiency. Such a parameter allows comparison of both contrast mechanisms in the relaxation time domain. From Eq. [2], β varies with IONP concentration:
| [3] |
Interestingly for low IONP concentration β → (r2*T20*)/(r1T10), whereas β → 1 for high concentration. As a result, special attention must be paid to T10 and T20* values when comparing the relative intensity of T1 and T2* contrast mechanisms, especially at low concentration. T10 may be orders of magnitude longer than T20*, whereas r2* is typically much larger than r1, although this depends on field strength, particle size (38) and contrast regime (4,19). Figure 2 displays β as a function of [IONP] for different sets of baseline conditions using IONP relaxivities measured in 1.8% agarose gel, at 3T (see “Methods”). Although r2*/r1 is about 30, the T1 variation induced by IONPs dominates the T2* variation when T10 is much larger than T20* (T10 = 2.7 s and T20* = 25 ms correspond to our in vitro experimental conditions).
Figure 2.

Relative variations of relaxation times as a function of IONP concentration. β is defined as (ΔT2*/T2*)/(ΔT1/T1). β is given by (r2*T20*)/(r1T10) in the low concentration limit and tends to 1 at high concentration. β < 1 indicates that T1 variations are larger than T2* variations, and vice-versa when β > 1. r1 = 4.5 mM−1.s−1 and r2* = 145 mM−1.s−1 were used here, consistent with in vitro measurements. Four unenhanced relaxation time sets were considered: T10/T20* of 0.8s/80ms as a realistic case of one order of magnitude relaxation time ratio; T10/T20* of 1s/67.2ms as a realistic case leading to β ≈ 2.2 at low concentration (corresponding Bopt = 1, see Fig. 3); T10/T20* of 1s/31ms as a realistic case leading to β ≈ 1 at low concentration; and T10/T20* of 2.7s/25ms corresponding to our in vitro experimental conditions.
Contrast is defined as the signal difference induced by a certain concentration of IONPs within a given tissue (i.e., C=ΔS= S[IONP] – Sno IONP). For small concentrations, it can be calculated as:
| [4] |
where ΔT1 and ΔT2* are the relaxation time changes caused by the small quantity [IONP].
Contrast-to-noise ratio (CNR) per unit time (i.e. ) is an index of contrast efficiency. To provide a comparison of SPGR with SubUTE contrast efficiency, a factor of noise amplification is applied to SubUTE, corresponding to the addition of uncorrelated noise arising from the two signals. The time normalization is necessary when comparing sequences with different TRs; this accounts for the greater number of signal averages that would be used if TR is made shorter, thus keeping the total acquisition time constant. An interesting consequence is that TR can be arbitrarily set to a low value without significantly affecting the optimization results, as the corresponding optimum obtained for TR→0 is essentially flat. This was first described by Ernst and Anderson (39) regarding signal optimization, briefly mentioned for the T1 contrast by Edelstein et al. (40) and extended to a general contrast optimization, including T2, by Buxton et al. (37). The practical result is that as long as TR is short compared to T1, θ can be adjusted to preserve optimum signal or contrast efficiency.
Methods
Theoretical contrast optimization
We consider three different contrast regimes: 1) the usual positive T1 contrast (CNRT1) obtained with SPGR at the shortest possible TE (i.e., UTE); 2) the usual negative T2* contrast (CNRT2*), typically obtained with an SPGR sequence at long TE using a low θ to reduce concomitant T1 effects (37); 3) the positive, combined T1-T2* contrast (CNRT1–T2*) obtained with SubUTE images.
In order to compare the contrast efficiency of these three regimes, we optimize θ, TR, and TE, assuming that the IONP properties r1 and r2* and baseline tissue relaxation times T10 and T20* are known.
We further investigate the balance between T2* and T1 effects by assessing the ratio of optimal T2* contrast over optimal T1 contrast defined as Bopt. Such a parameter represents the translation of β (Eq. 3) from the relaxation time domain into the image domain, under optimal detection conditions. It is simply calculated as Bopt = |CNRT2*/CNRT1 | , where CNRT2* and CNRT1 are calculated from the SPGR signal (Eqs. 1, 2 and 4) using respective optimal MR parameters. Bopt =1 defines the boundary between dominant T2* contrast (Bopt>1) and dominant T1 contrast (Bopt <1). The CNR gain obtained with the SubUTE approach is computed as well. It is defined as the ratio of optimal CNRT1–T2* over the maximum achievable CNR that can be obtained with an SPGR sequence: this can be either CNRT2* or CNRT1 depending on the dominant mechanism, as indicated by Bopt. Theoretical calculations and parameter optimization were performed with Mathematica (Wolfram Research Inc., Champaign, IL, USA).
Experiments
All MR data presented in this paper were acquired at room temperature on a Sigma HDx 3T scanner (GE Healthcare, Milwaukee, WI, USA).
Peptide-Coated IONPs
Nanoworm IONPs (41) (about 70×30 nm2 elongated particles) were used in this study because of their superior targeting properties as compared to spherical particles. The tumor-homing peptide CR(NMe)EKA was chemically coupled on the surface of the nanoworm IONPs to produce tumor-targeting particles (42). Synthesis and coating of IONPs were performed as described previously (38, 39).
In vitro study
Nanoworm IONPs were characterized by MRI. Several concentrations were prepared (0, 0.09, 0.18, 0.9, and 1.8 mM Fe) by dilution in 1.8% agarose gel (Type I, Sigma-Aldrich Inc., St. Louis, MO, USA) and were stoked in 1 mL tuberculin syringes. Longitudinal and transverse relaxivities were calculated by measuring the T1, T2 and T2* of each sample and performing a linear regression using Eq. 2. T1, T2, and T2* were respectively measured with 1) an inversion-recovery fast spin echo (FSE) sequence (TR/TE=4s/9.3 ms, ETL=4, BW=±31.25 kHz, 5-mm slice thickness, 256×78 image matrix, and 0.47-mm in-plane resolution) using 14 inversion times (TI) ranging from 50 ms to 3.5 s; 2) a multiecho spin echo sequence (TR=1.5 s, BW=±31.25 kHz, 3-mm slice thickness, 256×128 image matrix, and 0.47-mm in-plane resolution) using eight evenly spaced TEs ranging from 8.5 ms to 68 ms; and 3) a multiecho gradient echo sequence (TR=300 ms, BW=±31.25 kHz, θ=25°, 1-mm slice thickness, 192×192 image matrix, and 0.63-mm in-plane resolution) with 16 evenly spaced TEs ranging from 3.8 ms to 57.8 ms.
On each image, signal intensities were measured by drawing ROIs in the center of each vial. T1 quantification was performed by curve-fitting the analytical function A×|1-B×exp(−TI/T1)| to the signal, with A, B, and T1 as free parameters. Using B as a free parameter allowed us to account for the not-infinitely long TR, and for actual flip angles that differed from the nominal one (a 180° flip angle ideally corresponds to B=2 when TR is infinitely long). This significantly improved the goodness of fit.
T2 and T2* were extracted by fitting [(C×exp(−TE/T2(*)))2 + D2]1/2 to the measured signal decay, with C, D, and T2(*) as free parameters. The offset term (i.e., D) was added to the usual exponential model in order to account for noise. It significantly improved the goodness of fit when the SNR was poor at the longer TEs (i.e. at high iron concentration). T1 and T2(*) nonlinear least square fitting algorithms were implemented using Matlab (The MathWorks Inc., Natick, MA, USA).
To compare the theoretical predictions derived from the previous section, further measurements were performed using the in vitro samples. Several multiecho 2D UTE sequences were acquired with varying TEs (from 8 μs to 33 ms) and θs (10, 30, 50, and 70°) with TR fixed at 300 ms. Other MR parameters were: NEX=2, BW=±62.5 kHz, 2-mm slice thickness, 256 readout data points, and 299 radial trajectories. The data were reconstructed into a 256×256 image matrix with a 7.5-cm field of view (FOV). UTE sequences were performed in multiecho mode in order to produce SubUTE images and to assess contrast variation with TE. Additionally, this allowed extraction of baseline T20* from the decay curve under the same conditions in which the UTE imaging was performed. This was important for comparing theory with results since it significantly impacts β as shown in Fig. 2.
In vivo molecular imaging study
All in vivo studies were conducted with the approval of the University’s Institutional Animal Care and Use Committee. Nude mice bearing orthotopic 22Rv-1 human prostate tumors were imaged before and 7 hours after IV injection of 5 mg iron per kg of a CR(NMe)EKA-coated nanoworm IONPs. A nontargeted form of the nanoworms was used as a control. To reduce particle uptake by the reticulo-endothelial system, each animal received an intravenous injection of nickel liposomes (0.2 mmol of Ni) 1 hr prior to the nanoworm injection (43). Mice were scanned using a custom made 3-cm-diameter bird-cage coil.
An anatomic scan consisting of a T2-weighted FSE (TR/TE=6.4 s/70 ms, ETL=32, NEX=3, BW=±15.63 kHz, 1-mm slice thickness, 18 slices, 160×160 image matrix, 3.5-cm FOV, total scan time of about 6 min) was first acquired to show the tumor location. Next, a multiecho 2D UTE sequence was run (TR=89 ms, TEs=0.008/6.2/12.4/18.6 ms, θ=50°, NEX=6, BW=±27.78 kHz, slice thickness/spacing=2/2 mm, 2 slices, 256 readout data points, and 299 radial trajectories, total scan time of 2 min, 45 s). These data were reconstructed into 256×256 image matrix with a 3.5-cm FOV. Corresponding SubUTE images were computed. For post-injection UTE scans, axial slices were carefully matched to the previous scans by measuring the height of the sections and comparing the vascular patterns and other landmarks in the images and using the FSE scan as a guide.
Results
Optimal MR contrast, theoretical results
Using the derivative approach (i.e., small concentration limit, Eq. 4), we derived optimal MR parameters as a function of tissue and IONP characteristics. Optimization rules are summarized in Table 1. The IONP positive contrast regime (CNRT1) is optimized under the same condition such that only T1 varies, as there is no interference with the T2* contrast in this case (TE→0). IONP negative contrast (CNRT2*) and synergistic T1-T2* positive contrast (CNRT1–T2*) regimes lead to specific results, different from the usual independent T2* contrast, because of the needed balance between the two relaxation effects. For those last two regimes, it is not possible to derive analytical expressions for optimal TEs and θs using simple mathematical functions because the corresponding equations are non-algebraic. However, specific methods can be used to find numerical approximations. The negative contrast to which T1 and T2* effect are opposed is typically optimized with a TE ≈ T20* (the optimal TE is actually slightly longer than T20*) and a relatively low flip angle (θT2*<θErnst) in order to reduce signal enhancement due to T1 (consistent with (37)); whereas the SubUTE contrast benefits from a slightly longer TE and a higher flip angle (θT1–T2*), which is a compromise between the two optimal angles θT1 and θT2*.
Table 1.
Optimal MR parameters derived under the small variation assumption.
| Optimal TR | Optimal TE | Optimal θ | |
|---|---|---|---|
| T1 contrast efficiencya | TR < T1 | TE ≪T2* | θ = θT1 > θErnst |
| T2* contrast efficiencya | TR < T1 | TE = T2* | θ = θErnst |
| IONP contrast efficiency | TR < T10 | TE ≪ T20* | θ = θT1 > θErnst |
| Positive contrast regime (CNRT1) | |||
| IONP contrast efficiency | TR < T10b | TE = TET2* | θ = θT2* < θErnst |
| Negative contrast regime (CNRTT2*) | |||
| IONP contrast efficiency | TR < T10 | TE = TET1–T2* | θ = θT1–T2* |
| Synergistic T1-T2* (CNRT1–T2*) | θT2* < θT1–T2* < θT1 | ||
TR < T1 means that the optimum is essentially flat, as opposed to TE ≪ T2* here.
T1 and T2* variations treated separately. See (37) for an analytical expression of θT1.
When β < 1, a low flip angle does not reduce T1 effects enough at short TR, and a longer TR (> T1) may slightly improve CNRT2*. However in such cases T1 contrast is the dominant mechanism and thus is a better strategy than T2* contrast.
Bopt and the SubUTE CNR-gain were calculated using optimal MR parameters for all three contrast mechanisms (see Table 1; TR≪T1 was used). Despite the lack of a closed form, we used numerical methods to evaluate those two quantities using a broad range of parameters covering all realistic values, and found that they vary with β but do not directly depend on T10, T20*, r1, r2*or the corresponding optimal MR parameters. Hence Figure 3 displays Bopt and the SubUTE CNR-gain as a function of β only. Such curves allow prediction of which contrast mechanism should be used for a particular application, assuming relative relaxation time variations are known. Two particular [β, Bopt] value sets are emphasized in Fig. 3-a): 1) Bopt≈0.39 for β=1; and 2) Bopt=1 for β≈2.2. The former shows that for identical relative variations of both relaxation times, optimal T2* contrast is only about 39% of optimal T1 contrast, while the latter shows that to obtain identical image contrast from both mechanisms, the relative T2* variation has to be 2.2× that of T1. This proves that T2* contrast is intrinsically less efficient than T1 contrast for SPGR sequences, consistent with the findings of Buxton et al. (37). In Fig. 3-b), a SubUTE CNR-gain > 1 indicates that the synergistic combination of T1 and T2* contrasts will increase IONP detection sensitivity. This happens when both mechanisms have comparable intensities and corresponds to 1.3< β <3.9.
Figure 3.

Theoretical comparison of optimal T1-, T2*-, and T1-T2*-contrasts as a function of relative relaxation times variations. a) Bopt is defined as CNRT2*/CNRT1 where both contrasts are calculated using respective optimal MR parameters. Bopt = 1 defines the boundary between dominant T2* contrast (Bopt > 1) and dominant T1 contrast (Bopt < 1). b) When SubUTE CNR-gain > 1, SubUTE improves detection sensitivity as compared to SPGR because T1 and T2* synergistic combination overcome the √2 noise penalty. In the worst-case scenario (β ≪ 1 or β ≫ 1), SubUTE will yield a CNR √2 times lower than the dominant contrast mechanism because the other mechanisms has a negligible contribution to SubUTE contrast.
Figure 4 displays the numerical calculation of CNR generated by IONPs as a function of TE and θ for four different baseline tissues, two IONP concentrations and measured IONP relaxivities. TR is arbitrarily set to 100 ms here as the shortest T1 is about 174 ms (for 1mM iron and a 0.8s T10). A discrete contrast definition was used for the calculation presented here (C= S[IONP] – Sno IONP instead of Eq. 4). For each concentration and tissue baseline condition, isocontours were calculated in relation to the maximal achievable contrast among the three regimes (i.e. T1, T2* or synergistic T1-T2*). The plots corresponding to 1 μM represent the low concentration limit (i.e. identical curves are obtained using the derivative approach since the first-order linearity assumption is satisfied under such small variations), whereas the 1 mM curve set display a high concentration example. These show how optimal MR contrast and parameters vary with IONP concentration. Intermediate concentrations were also studied (not shown). These show a smooth transition to the high concentration regime. We found that the same general rules as with standard T1 and T2* contrast apply when concentration varies, namely i) the shorter the T2*, the shorter the TE needed to obtain a high CNR in the T2* regime, and ii) the shorter the T1, the higher the flip angle required to optimize CNR in the T1 regime.
Figure 4.
IONP contrast-to-noise ratio as a function of TE and θ for regular SPGR (including UTE) and SubUTE, calculated for different baseline relaxation times and IONP concentrations (TR = 100 ms, r1 = 4.5 mM−1.s−1 and r2* = 145 mM−1.s−1). Isocontours are calculated in relation to the maximal achievable contrast (indicated by the 99% contour) for each particular concentration and tissue baseline conditions. SPGR exhibits the two usual contrast regimes: positive T1 contrast toward short TEs, and negative T2* contrast when longer TEs and lower θs are used. The 0% isocontours indicate the settings when T1 and T2* effects are perfectly balanced, resulting in no effective contrast from the background tissue. Conversely, SubUTE provides a single positive contrast regime: the synergistic T1-T2*-contrast as described on Fig 1.
Negative T2* contrast proved more efficient than T1 contrast only at low IONP concentration and for a moderate T10/T20* ratio (a)). The corresponding Bopt correctly predicts the ratio of optimal T1 and T2* contrast obtained for the 1 μM case, but fails to predict the correct ratio at high concentration. This is directly related to the validity of the derivative approach that is no longer correct at high concentration: in this latter case, optimal MR parameters are different from those used to calculate Bopt. However, we observe that T1 contrast is always superior to T2* contrast at high concentration, and that the relative efficiency of T2* contrast (decrease from b), d), f) to h)) follows the corresponding β values. Note that for the particular case where β =1 over the entire concentration range (e) and f)) contrast is very similar at both low and high concentration.
At high IONP concentration, the UTE approach is very important to the efficient probing of the T1 effect since the contrast falls off rapidly beyond TE=0.1ms. As expected from Fig. 3-b), when T1 or T2* contrast intensities are comparable (as in a) and c)), SubUTE provides optimal contrast because of its synergistic combination of the two relaxation effects (both positive). In the other situations, it does not improve sensitivity because of the √2 noise penalty.
Experimental results
Measured T2 and T2* relaxation times of IONP samples were nearly equal up to about 0.9 mM, showing that the motional narrowing regime (4) is satisfied for these particles in this environment (1.8% agarose gel) and concentration range. At higher concentration, T2* was significantly shorter than T2. Therefore, a unique transverse relaxivity was calculated ignoring the highest concentration (1.8mM) data point so that a consistent contrast mechanism could be described. Measured T1 and T2(*) relaxivities were r1 ≈ 4.5 mM−1.s−1 and r2(*) ≈ 145 mM−1.s−1. A large T20* difference was found between multiecho SPGR (1-mm slice and T20* ≈ 63 ms) and multiecho UTE (2-mm slice and T20* ≈ 25 ms), consistent with large-scale magnetic field inhomogeneities (44) in the slice direction.
Figure 5 displays a comparison between the multiecho UTE measurements and numerical computation for θ=10° and 50° (other angles are not shown). Calculations were performed using the measured relaxivities and baseline relaxation times (T10=2.7 s and T20*=25 ms). Comparing SPGR experimental data at 50° and TE→0 with those at 10° around TE ≈ T20*, we observed that the T1 contrast was greater than T2* contrast for the studied IONPs and “tissue” in our experimental condition.
Figure 5.
Experimental (a) and c)) versus simulated (b) and d)) MR signal as a function of TE for two different θs (other flip angle not shown). The left column displays SPGR signal (from TE = 8 μs − i.e. UTE − to TE = 33 ms) whereas the right one shows SubUTE data. Theoretical curves were calculated using TR = 300 ms, T10 = 2.7 s, T20* = 25 ms, r1 = 4.5 mM−1.s−1 and r2* = 145 mM−1.s−1, consistent with presented in vitro data. Both experimental and simulated datasets are normalized to allow clear comparison (0 mM SPGR MR signal set to 100 at TE = 0 ms and θ = 50°).
In vivo results are displayed in Fig. 6 and 7. Figure 6 compares a pre- and post-injection dataset for CR(NMe)EKA-coated IONPs and nontargeted IONPs. CR(NMe)EKA-coated IONPs accumulate in the tumor by recognizing fibrin and fibrin-associated clotted plasma proteins in tumor vessels, while nontargeted IONPs do not (39). Such particles can cause additional clotting in tumor vessels, which creates more binding sites for the peptide (42,43). This self-amplifying tumor homing effect makes the CR(NMe)EKA-coated IONP system a good candidate for MR molecular imaging as the effect provides significant accumulation of contrast media at the targeted site. Tumors appeared hyperintense relative to muscle on T2-weighted images before IONP injection and are consequently easily identifiable in row a). The signal was fairly homogeneous inside the tumor after intravenous injection of the nontargeted IONPs, suggesting that little or no iron was present in the tumor (the bright signal seen on the T2-weighted FSE, nontargeted IONPs post-injection, was probably due to a necrotic core). In line with the previous report (39), signal voids were visible on both FSE (a)) and the 6.2 ms 2nd echo UTE (c)), 7 hours after injection of the CR(NMe)EKA-coated IONPs (middle column). These areas of signal loss became hyperintense on UTE (b)) and SubUTE (d)) due to T1 and synergistic T1-T2* effects of IONPs, respectively. The combination of T1 and T2* contrast mechanisms increased the specificity for IONPs. Note that the brightness was more evident on the subtraction images (d)) due to partial background suppression (see Fig. 7 for different echo subtraction images). This improves the specificity for short T2* components. Those images illustrate the capacity of positive-contrast UTE strategies to detect IONPs in vivo.
Figure 6.
In vivo molecular imaging results. Axial images of tumor-bearing mice. Images were acquired before (left column) and after the injection of 5 mg iron per kg of targeted (middle column) or non-targeted (right column) IONPs to 2 different mice. Gd and IONP tubes were added as reference standards. 7h after injection of targeted IONPs, particles are visible as signal voids on T2 and T2* weighted images (a) and c)) and as hyperintense area on the UTE and SubUTE images (b) and d)). No particles are visible either pre injection or 7h post injection when using non-targeted IONPs.
Figure 7.
SubUTE in vivo results. Axial images of tumor bearing mouse acquired 7h after injection of 5 mg iron per kg of tumor-targeted IONPs. Different contrasts are illustrated here, obtained from the multiple echoes and corresponding subtraction image. The 2nd echo subtraction provides a better background suppression than later echoes (note the Gd signal suppression). All those images are obtained using a single acquisition, except the T2-weighted FSE, that is displayed here as a complementary contrast to distinguish the tumor.
Discussion
Using the standard theory of describing MR signal and IONP relaxivity as our starting point, we derived significant and easy-to-implement criteria for IONP contrast optimization. Our theoretical ideas were validated by in vitro experiments although there are some potential limitations.
The theoretical analysis presented here shows that, although the transverse relaxivity of IONPs is usually much higher than the corresponding longitudinal relaxivity, T1 contrast strategies may be the optimal way to detect them. This is explained in part by endogenous T10s that are often much longer than T20*s (except for fluids or highly liquid tissues), and corresponding low β values; the other reason being that T2* contrast is intrinsically less efficient than T1 contrast for SPGR sequences. T2* variations have to be 2.2 times greater than T1 variations to produce similar CNR under optimal imaging conditions (see Fig. 3). Hence we have derived the following criteria that predict optimal contrast strategies at low concentrations: when (r2*T20*)/(r1T10) > 2.2, negative T2* contrast is the preferred option, whereas when β < 2.2, T1 contrast is preferred. Indeed, in the low concentration limit, Bopt is a general indicator of contrast efficiency, as presented in Fig. 3. Note the importance of the r2*/r1 ratio as usually reported to characterize IONPs. At high IONP concentration, β→1 independent of baseline relaxation times and relaxivities, and T1 relaxation is expected to be the dominant contrast mechanism assuming a short enough TE can be employed.
This makes UTE detection particularly appealing in the high concentration range. We have shown that subtraction images (SubUTE) combine the usually-antagonist T1 and T2* contrast in a synergistic way. Although this does not always improve CNR because of a √2 noise penalty, it does increase detection sensitivity for IONPs when T1 and T2* effects make similar contributions (see Fig 3.). Moreover, because of its long T2 suppression capabilities (background suppression), significant improvement in specificity is anticipated with SubUTE (as illustrated in Fig. 7).
Despite the aforementioned advantages, IONPs are usually detected by MRI using either T2- or T2*-weighted sequences. An explanation for this may be the lack of efficient ways to detect T1 effects at high iron concentrations. The sensitivity toward T1 effects is reduced when relatively long TEs are used, and image interpretation becomes ambiguous because both positive and negative contrast can occur simultaneously. UTE and SubUTE approaches represent a new step forward in addressing this limitation.
Another explanation is the reduction of IONP T1 relaxivity and the increase in T2* relaxivity that are usually observed in vivo when particles are compartmentalized into cells (19–23). This is a major concern, especially for molecular imaging applications (2). A better understanding of the mechanisms involved in those compartmentalization-related effects is of great importance for the purpose of quantification and for distinguishing free iron from targeted iron (20). However, this is outside the scope of this paper. The major consequence to consider in our context is that r1 is observed to decrease with compartmentalization, whereas r2* increases; this typically results in a reduction of T1 contrast and an increase in T2* contrast for in vivo applications as compare to in vitro.
However, as we noticed in our experiments, large-scale magnetic field heterogeneities (due to imperfect shimming or BMS effects) may reduce T20*, and hence have a direct effect on CNR in any practical experimental setting (see Fig. 2 and 3). This is important, especially in 2D imaging, where the slice thickness is usually the largest voxel dimension and causes signal averaging. Although these effects may be corrected to improve quantification (44), they imply a reduction of T2* contrast efficiency to the benefit of the T1 effect.
In the in vivo experiment we described, we were not able to estimate β because we lacked the corresponding IONP relaxivity. However TR was set to an optimal value (TR of 89 ms for a tumor T10 of about 1s) and a 50° flip angle was used. This appeared to be a good compromise for optimizing both UTE T1-contrast and Sub-UTE synergistic contrast over a wide range of IONP concentrations, regardless of the actual β value, as illustrated by Fig. 4 for a slightly longer TR. We expect that application of our theory to realistic estimates of in vivo relaxivity and unenhanced relaxation times should prove to be a good indicator of in vivo contrast efficiency, as we have shown in the simple case of in vitro experiments where we measured corresponding relaxivity. Here we have successfully demonstrated that in vivo molecular imaging can be done with IONPs and T1 contrast using UTE. In the presented case, however, the problem of limited T1 relaxivity is unlikely to occur because CR(NMe)EKA-coated IONPs do not undergo cell internalization (42). Further studies will address compartmentalization effects using a system that causes cell internalization of the IONPs such as the iRGD peptide system (45).
Although the curves of the theoretical and experimental in vitro results do not match exactly, especially at high IONP concentrations, the agreement as shown in Fig. 5 is strong. We expect that the observed differences are due to the non-zero actual TE that was used experimentally, artifacts, or further limitations in our simple model such as nonlinear relaxivity at high concentration, non-exponential T2* decay (as seen in Fig. 5), or break down of the validity of the usual SPGR signal (Eq. 1) when T2* is of the same order as the RF pulse duration (46). In addition, some discrepancies could be attributed to RF-inhomogeneities as shown in Fig. 5-c for the case of 10° flip angle and UTE: contrast should be essentially null on such a proton density weighted image.
In contrast to the general optimization performed by Hendrick et al. (47), we did not consider the lengthening of signal readout time as TE increases (i.e., the reduction of the readout bandwidth, and thus a noise reduction). This explains why, in their case, optimal T1 CNR is obtained for TE ≠ 0. Our formulation is more appropriate for UTE imaging, where the signal intensity is essentially independent of the observation time (center of k-space sampled at the initial peak of the FID, albeit with blurring and size-dependant signal attenuation (27) from T2* decay).
Conclusions
In this study we show that contrary to conventional wisdom, optimizing T1 contrast may be a good strategy for imaging IONPs. Hence, UTE-based sequences are of considerable interest for IONP detection. T1 effects are measured at ultrashort TE (i.e., with minimal signal loss and susceptibility effects), providing a positive contrast even at relatively high IONP concentrations. Furthermore, a multiecho UTE sequence yields registered datasets and complementary types of contrasts in a single acquisition (T1, T2* and SubUTE T1-T2*). As compared to single T2* contrast strategies, such a sequence enables the additional specificity afforded by measurement of the T1 weighted UTE signal. SubUTE may increase detection sensitivity, by combining T1 and T2* contrast in a synergistic way, and specificity by offering background signal suppression. Using UTE and SubUTE, we show in a preliminary in vivo molecular imaging study the benefit of such additional specificity for voxels with both short T1 and T2*, the signature of IONPs.
Acknowledgments
The authors thank Drs. Mark Bydder and Nikolaus Szeverenyi (UCSD, Radiology Department) for helpful comments on this manuscript, as well as Jacqueline Corbeil and Salman Farshchi-Heydari (UCSD, Radiology Department) for their support in animal preparation and monitoring. This study was supported by Development Project Funds from the In-vivo Cancer and Molecular Imaging Center at UCSD (NCI-P50CA128346), and funding from Pfizer.
References
- 1.Corot C, Robert P, Idee JM, Port M. Recent advances in iron oxide nanocrystal technology for medical imaging. Adv Drug Deliver Rev. 2006;58(14):1471–1504. doi: 10.1016/j.addr.2006.09.013. [DOI] [PubMed] [Google Scholar]
- 2.Liu W, Frank JA. Detection and quantification of magnetically labeled cells by cellular MRI. European Journal of Radiology. 2009;70(2):258–264. doi: 10.1016/j.ejrad.2008.09.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Hardy PA, Henkelman RM. Transverse Relaxation Rate Enhancement Caused by Magnetic Particulates. Magn Reson Imag. 1989;7(3):265–275. doi: 10.1016/0730-725x(89)90549-3. [DOI] [PubMed] [Google Scholar]
- 4.Weisskoff RM, Zuo CS, Boxerman JL, Rosen BR. Microscopic Susceptibility Variation and Transverse Relaxation - Theory and Experiment. Magn Reson Med. 1994;31(6):601–610. doi: 10.1002/mrm.1910310605. [DOI] [PubMed] [Google Scholar]
- 5.Cunningham CH, Arai T, Yang PC, McConnell MV, Pauly JM, Conolly SM. Positive contrast magnetic resonance imaging of cells labeled with magnetic nanoparticles. Magn Reson Med. 2005;53(5):999–1005. doi: 10.1002/mrm.20477. [DOI] [PubMed] [Google Scholar]
- 6.Farrar CT, Dai GP, Novikov M, Rosenzweig A, Weissleder R, Rosen BR, Sosnovik DE. Impact of field strength and iron oxide nanoparticle concentration on the linearity and diagnostic accuracy of off-resonance imaging. NMR Biomed. 2008;21(5):453–463. doi: 10.1002/nbm.1209. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Stuber M, Gilson WD, Schar M, Kedziorek DA, Hofmann LV, Shah S, Vonken EJ, Bulte JWM, Kraitchman DL. Positive contrast visualization of iron oxide-labeled stem cells using inversion-recovery with ON-Resonant water suppression (IRON) Magn Reson Med. 2007;58(5):1072–1077. doi: 10.1002/mrm.21399. [DOI] [PubMed] [Google Scholar]
- 8.Zurkiya O, Hu XP. Off-resonance saturation as a means of generating contrast with superparamagnetic nanoparticles. Magn Reson Med. 2006;56(4):726–732. doi: 10.1002/mrm.21024. [DOI] [PubMed] [Google Scholar]
- 9.Mani V, Briley-Saebo KC, Itskovich VV, Samber DD, Fayad ZA. GRadient echo Acquisition for Superparamagnetic particles with positive contrast (GRASP): Sequence characterization in membrane and glass superparamagnetic iron oxide phantoms at 1.5T and 3T. Magn Reson Med. 2006;55(1):126–135. doi: 10.1002/mrm.20739. [DOI] [PubMed] [Google Scholar]
- 10.Seppenwoolde J-H, Viergever MA, Bakker CJG. Passive tracking exploiting local signal conservation: The white marker phenomenon. Magn Reson Med. 2003;50(4):784–790. doi: 10.1002/mrm.10574. [DOI] [PubMed] [Google Scholar]
- 11.Dahnke H, Liu W, Herzka D, Frank JA, Schaeffter T. Susceptibility gradient mapping (SGM): A new postprocessing method for positive contrast generation applied to superparamagnetic iron oxide particle (SPIO)-labeled cells. Magn Reson Med. 2008;60(3):595–603. doi: 10.1002/mrm.21478. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Mills PH, Wu YJL, Ho C, Ahrens ET. Sensitive and automated detection of iron-oxide-labeled cells using phase image cross-correlation analysis. Magn Reson Imag. 2008;26(5):618–628. doi: 10.1016/j.mri.2008.01.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.De Rochefort L, Brown R, Prince MR, Wang Y. Quantitative MR susceptibility mapping using piece-wise constant regularized inversion of the magnetic field. Magn Reson Med. 2008;60(4):1003–1009. doi: 10.1002/mrm.21710. [DOI] [PubMed] [Google Scholar]
- 14.De Rochefort L, Poirier-Quinot M, Kressler B, Wilhelm C, Gazeau F, Ginefri J-C, Darrasse L, Wang Y. SPIO Labeled Cells: Magnetic Resonance Source Quantification by Inverting the Dipole Field. Proc Int Soc Magn Reson Med. 2008;16:1691. [Google Scholar]
- 15.Guimaraes R, Clement O, Bittoun J, Carnot F, Frija G. MR Lymphography with Superparamagnetic Iron Nanoparticles in Rats: Pathologic Basis for Contrast Enhancement. Am J Roentgenol. 1994;162(1):201–207. doi: 10.2214/ajr.162.1.8273666. [DOI] [PubMed] [Google Scholar]
- 16.Grangier C, Tourniaire J, Mentha G, Schiau R, Howarth N, Chachuat A, Grossholz M, Terrier F. Enhancement of Liver Hemangiomas on T1-Weighted MR SE Images by Superparamagnetic Iron-Oxide Particles. J Comp Assist Tomogr. 1994;18(6):888–896. doi: 10.1097/00004728-199411000-00008. [DOI] [PubMed] [Google Scholar]
- 17.Chambon C, Clement O, Leblanche A, Schoumanclaeys E, Frija G. Superparamagnetic Iron-Oxides as Positive MR Contrast Agents - In vitro and In vivo Evidence. Magn Reson Imag. 1993;11(4):509–519. doi: 10.1016/0730-725x(93)90470-x. [DOI] [PubMed] [Google Scholar]
- 18.Canet E, Revel D, Forrat R, Baldyporcher C, Delorgeril M, Sebbag L, Vallee JP, Didier D, Amiel M. Superparamagnetic iron oxide particles and positive enhancement for myocardial perfusion studies assessed by subsecond T1-weighted MRI. Magn Reson Imag. 1993;11(8):1139–1145. doi: 10.1016/0730-725x(93)90241-5. [DOI] [PubMed] [Google Scholar]
- 19.Bowen CV, Zhang XW, Saab G, Gareau PJ, Rutt BK. Application of the static dephasing regime theory to superparamagnetic iron-oxide loaded cells. Magn Reson Med. 2002;48(1):52–61. doi: 10.1002/mrm.10192. [DOI] [PubMed] [Google Scholar]
- 20.Kuhlpeter R, Dahnke H, Matuszewski L, Persigehl T, von Wallbrunn A, Allkemper T, Heindel WL, Schaeffter T, Bremer C. R2 and R2*mapping for sensing cell-bound superparamagnetic nanoparticles: In vitro and murine in vivo testing. Radiology. 2007;245(2):449–457. doi: 10.1148/radiol.2451061345. [DOI] [PubMed] [Google Scholar]
- 21.Tanimoto A, Oshio K, Suematsu M, Pouliquen D, Stark DD. Relaxation effects of clustered particles. J Magn Reson Imaging. 2001;14(1):72–77. doi: 10.1002/jmri.1153. [DOI] [PubMed] [Google Scholar]
- 22.Billotey C, Wilhelm C, Devaud M, Bacri JC, Bittoun J, Gazeau F. Cell internalization of anionic maghemite nanoparticles: Quantitative effect on magnetic resonance imaging. Magn Reson Med. 2003;49(4):646–654. doi: 10.1002/mrm.10418. [DOI] [PubMed] [Google Scholar]
- 23.Simon GH, Bauer J, Saborovski O, Fu YJ, Corot C, Wendland MF, Daldrup-Link HE. T1 and T2 relaxivity of intracellular and extracellular USPIO at 1.5T and 3T clinical MR scanning. European Radiology. 2006;16(3):738–745. doi: 10.1007/s00330-005-0031-2. [DOI] [PubMed] [Google Scholar]
- 24.Bjornerud A, Johansson LO, Briley-Saebo K, Ahlstrom HK. Assessment of T1 and T2* effects in vivo and ex vivo using iron oxide nanoparticles in steady state-Dependence on blood volume and water exchange. Magn Reson Med. 2002;47(3):461–471. doi: 10.1002/mrm.10066. [DOI] [PubMed] [Google Scholar]
- 25.Oswald P, Clement O, Chambon C, SchoumanClaeys E, Frija G. Liver positive enhancement after injection of superparamagnetic nanoparticles: Respective role of circulating and uptaken particles. Magn Reson Imag. 1997;15(9):1025–1031. doi: 10.1016/s0730-725x(97)00004-0. [DOI] [PubMed] [Google Scholar]
- 26.Bergin C, Pauly JM, Macovski A. Lung parenchyma: projection reconstruction MR imaging. Radiology. 1991;179(3):777–781. doi: 10.1148/radiology.179.3.2027991. [DOI] [PubMed] [Google Scholar]
- 27.Coolen BF, Lee P, Shuter B, Golay X. Optimized MRI parameters for positive contrast detection of iron-oxide labeled cells using double-echo Ultra-short echo time (d-UTE) sequences. Proc Int Soc Magn Reson Med. 2007;15:1224. [Google Scholar]
- 28.Girard OM, Du J, Mattrey RF. Investigation of SPIO T1-signature: Positive Contrast using Ultrashort TE Imaging. Proc Int Soc Magn Reson Med. 2009;17:2769. [Google Scholar]
- 29.Diwoky C, Reinisch A, Gross D, Lehmann V, Strunk D, Stollberger R. UTE Imaging for Single Cell Detection with Positive Contrast. Proc Int Soc Magn Reson Med. 2009;17:804. [Google Scholar]
- 30.Pauly JM, Conolly SM, Nishimura DG, Macovski A. A slice-selective excitation for very short T2 species. Proc of SMRM 8th Annual Meeting; 1989. p. 28. [Google Scholar]
- 31.Du J, Hamilton G, Takahashi A, Bydder M, Chung CB. Ultrashort echo time spectroscopic imaging (UTESI) of cortical bone. Magn Reson Med. 2007;58(5):1001–1009. doi: 10.1002/mrm.21397. [DOI] [PubMed] [Google Scholar]
- 32.Boada FE, Wiener E. Ultra-short Echo Time Difference (USTED) Data Acquisition for T2* Contrast Reversal. Proc Int Soc Magn Reson Med. 2006;14:189. [Google Scholar]
- 33.Crowe LA, Ris F, Nielles-Vallespin S, Speier P, Masson S, Armanet M, Bosco D, Berney T, Vallee J-P. 3D radial UTE imaging for quantification of transplanted iron oxide labelled islet cells. Proc Int Soc Magn Reson Med. 2009;17:806. [Google Scholar]
- 34.Seevinck PR, De Leeuw H, Bos C, Bakker CJ. High positive contrast generation of a subvoxel susceptibility deviation in ultrashort TE (UTE) radial center-out imaging at 3T. Proc Int Soc Magn Reson Med. 2009;17:869. [Google Scholar]
- 35.Yablonskiy DA. Quantitation of intrinsic magnetic susceptibility-related effects in a tissue matrix. Phantom study Magn Reson Med. 1998;39(3):417–428. doi: 10.1002/mrm.1910390312. [DOI] [PubMed] [Google Scholar]
- 36.Yablonskiy DA, Haacke EM. Theory of NMR signal behavior in magnetically inhomogeneous tissues: the static dephasing regime. Magn Reson Med. 1994;32(6):749–763. doi: 10.1002/mrm.1910320610. [DOI] [PubMed] [Google Scholar]
- 37.Buxton RB, Edelman RR, Rosen BR, Wismer GL, Brady TJ. Contrast in Rapid MR Imaging - T1-Weighted and T2-Weighted Imaging. J Comp Assist Tomogr. 1987;11(1):7–16. doi: 10.1097/00004728-198701000-00003. [DOI] [PubMed] [Google Scholar]
- 38.Roch A, Muller RN, Gillis P. Theory of proton relaxation induced by superparamagnetic particles. J Chem Phys. 1999;110(11):5403–5411. [Google Scholar]
- 39.Ernst RR, Anderson WA. Application of Fourier Transform Spectroscopy to Magnetic Resonance. Review of Scientific Instruments. 1966;37(1):93–102. [Google Scholar]
- 40.Edelstein WA, Bottomley PA, Hart HR, Smith LS. Signal, Noise, and Contrast in Nuclear Magnetic-Resonance (NMR) Imaging. J Comp Assist Tomogr. 1983;7(3):391–401. doi: 10.1097/00004728-198306000-00001. [DOI] [PubMed] [Google Scholar]
- 41.Park JH, von Maltzahn G, Zhang LL, Schwartz MP, Ruoslahti E, Bhatia SN, Sailor MJ. Magnetic iron oxide nanoworms for tumor targeting and imaging. Adv Mater. 2008;20(9):1630–1635. doi: 10.1002/adma.200800004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Agemy L, Sugahara KN, Kotamraju VR, Gujraty K, Girard OM, Kono Y, Mattrey RF, Park JH, Sailor MJ, Jimenez A, Cativiela C, Zanuy D, Sayago FJ, Aleman C, Nussinov R, Ruoslahti E. Nanoparticle-induced vascular blockade in human prostate cancer. Blood. 2010 doi: 10.1182/blood-2010-03-274258. In Press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Simberg D, Duza T, Park JH, Essler M, Pilch J, Zhang LL, Derfus AM, Yang M, Hoffman RM, Bhatia S, Sailor MJ, Ruoslahti E. Biomimetic amplification of nanoparticle homing to tumors. PNAS. 2007;104(3):932–936. doi: 10.1073/pnas.0610298104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Dahnke H, Schaeffter T. Limits of detection of SPIO at 3.0 T using T-2* relaxometry. Magn Reson Med. 2005;53(5):1202–1206. doi: 10.1002/mrm.20435. [DOI] [PubMed] [Google Scholar]
- 45.Sugahara KN, Teesalu T, Karmali PP, Kotamraju VR, Agemy L, Girard OM, Hanahan D, Mattrey RF, Ruoslahti E. Tissue-Penetrating Delivery of Compounds and Nanoparticles into Tumors. Cancer Cell. 2009;16(6):510–520. doi: 10.1016/j.ccr.2009.10.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Carl M, Bydder M, Du J, Takahashi A, Han E. Optimization of RF Excitation to Maximize Signal and T2 Contrast of Tissues With Rapid Transverse Relaxation. Magn Reson Med. 2010 doi: 10.1002/mrm.22433. In Press. [DOI] [PubMed] [Google Scholar]
- 47.Hendrick RE, Kneeland JB, Stark DD. Maximizing Signal-to-Noise and Contrast-to-Noise Ratios in Flash Imaging. Magn Reson Imag. 1987;5(2):117–127. doi: 10.1016/0730-725x(87)90041-5. [DOI] [PubMed] [Google Scholar]




