Flow Estimation by Indicator Dilution (Bolus Injection)

REDUCTION OF ERRORS DUE TO TIME-AVERAGED SAMPLING DURING UNSTEADY FLOW

James B Bassingthwaighte 1, Thomas J Knopp 1, Dennis U Anderson 1
PMCID: PMC3098034  NIHMSID: NIHMS233516  PMID: 4917088

Abstract

Indicator dilution techniques used for the estimation of flow (F), mean transit time (t̄), dispersion (σ), and mean transit time volume (V) in the circulation are subject to error when (1) flow is not steady and (2) concentrations are obtained by sampling at a constant rate (time averaging) rather than at rates proportional to the instantaneous flow past the sampling site (volume averaging).

Using a simple descriptive model for indicator transport, the effects of simulated aortic flow or of sinusoidal flow of widely variable frequency were assessed. Errors in estimates of F, t̄, σ, and V are greater with bolus injections than with constant-rate injections. Errors are roughly proportional to the amplitude of variation in flow. They are maximal when the period of flow fluctuation is similar to the passage time of dye dilution curve, which, for the human central circulation, is about the time for one respiratory cycle. With sinusoidal flow between 50% and 150% of the mean flow, errors were at worst up to 60% in F, 30% in t̄, 50% in σ, and 70% in V, with a bimodal distribution. Errors are minimal at cardiac frequencies. The troublesome lower frequencies can be avoided. Preliminary tests of a method for converting time- to volume-averaged concentrations gave encouraging results

Keywords: blood flow, dye curves, cardiac output, circulatory dispersion, mathematical model, circulation model, computer simulation, nonstationary differential operators


Flow in the circulation fluctuates continuously, but the classical approaches to the uses of indicator dilution techniques for the estimation of blood flow (F), mean transit time (t̄), and dispersion (σ) in the circulation are based on the assumption that flow is steady (1, 2). It is therefore important to estimate the magnitudes of errors involved in the calculation of F, t̄, and σ. The indicator dilution method is theoretically accurate if the flow stream can be sampled by obtaining volumes in proportion to the instantaneous flow past the sampling point. This is volume-averaged sampling (3). Ordinarily, we sample the flow stream at a constant rate or observe its average concentration in a cross section continuously; this is time-averaged sampling. Although some indication of the magnitudes of the errors resulting from time-averaged sampling has been given for constant-rate dye injections (4) and for the Fick technique (5-7), no quantitative estimates of the errors have been made for the technique of sudden bolus injection. It is the purpose of this paper to describe the sources and magnitudes of such errors and to provide some insight into mechanisms for reducing them.

Theoretical Approach

Previous studies on the transport functions of material in the arterial system of the human leg (8), in the aorta (9), and in the pulmonary circulation (10) of the dog have shown that the relative dispersion, σ/t̄ (the ratio of the standard deviation of transport function, σ, to its mean transit time, t̄, is approximately constant over a wide range of flows. It has been deduced that, even over a 10-fold range of flows with Reynolds numbers ranging from 300 to 3,000, the spatial dispersion of indicator is apparently independent of the flow and dependent solely on the distance traveled along the vascular system (8).

On this basis it appears reasonable to examine the effects on slug-injection dye curves of continuously variable flow, F(t), in a conservative, constant-volume system. To obtain a wide range of frequencies and amplitudes of variation in flow and to have injections made during different phases of cyclic variation in flow, a model system was used; in doing this a mathematical model has some advantages over a hydraulio model.

A linear differential operator with flow-proportional frequency coefficients and with other coefficients constant satisfies the requirement of constant relative dispersion, and the coefficients can be chosen so that the impulse response is similar to a dye dilution curve in the circulation.

A fairly high-order operator is required to produce an impulse response (transport function, or probability density function of transit times), h(t), with an appropriate delay and σ/t̄. A twelfth-order operator was used in this study; this order is not high enough to provide a perfect description of h(t) in the circulation but is a compromise used to avoid overly long solution times. (A continuously variable delay operator in series with a low-order operator was found to be descriptive but was numerically less reliable at high frequencies.) The model’s h(t) is a skewed unimodal density function similar in shape to others (11-13) which could have been used for this study but which cannot be computed as quickly.

The operator may be represented by the family of six differential equations, consisting of three identical pairs of second-order operators:

d2yidt2+2ζ1ω1dyidt+ω1yi2=ω1yi12d2yi+1dt2+2ζ2ω2dyi+1dt+ω2yi+12=ω2yi2} (1a)

with i = 1,3,5 in which

ω2=βω1 (1b)

and

ω1=2παF(t)V. (1c)

Here, yi is the output of the ith component of the model, y0 is the concentration-time curve of indicator at the input to the segment of the circulation, y0 is the output concentration-time curve C’(t).

The model parameters, ζ1 and ζ2, must be set so that h(t) has no negative values (that is, the response is overdamped). By varying ζ1, ζ2, and β, widely different vascular segments may be simulated with flow constant or variable at any positive rate. α is a normalizing, dimensionless constant, which has a value such that t of h(t̄) equals V/F during steady flow.

A segment of the circulation acts as a low pass filter, causing delay in transit and damping of the higher frequency fluctuations in concentration. In a linear system, when the input is sinusoidal at frequency ω, so also is the output. The cutoff frequency, ωe, is defined as the frequency of sinusoidal variation in concentration at which the ratio of output amplitude to input amplitude is 0.707, or 2−0.5. When a system has constant volume and its h(t) has constant relative dispersion then ωc, like ω1 and ω2, is proportional to F. Thus, increasing the flow enables higher frequency information to be transmitted along a segment of the circulation, while reducing the flow reduces the input frequencies discernible at the downstream or output end of the segment. When flow is continuously variable, any input signal, y0, is frequency modulated in a transient fashion by changes in flow, and the high-frequency content of y0 is filtered to an extent reciprocally related to the flow.

Methods

The equations were solved numerically on a digital computer using a simulation control system (14) providing automatic or manual means of changing parameters and allowing choice of variables for on-line display of solutions.

Specific Model Parameters

For this study, a value of 0.28 was chosen for σ/t̄. To obtain this, ζ1 was set to 0.95, ζ2 to 0.8, and β = 1.82. α was 1.32. This is still quite general because σ/t̄ has a rather narrow range in the circulation—from 0.2 in peripheral vessels to about 0.45 for the pulmonary circulation (8-10).

Input, y0(t)

For a pulse injection, the total mass of indicator injected is m. The rate of entry of indicator at the upper end of the circulatory segment is ṁ(t) or dm/dt. If the injection is made at the entrance,

y0(t)=m.(t)F(t)for0<tTinjy0(t)=0fort<0t>Tinj}, (2)

in which ṁ(t) mg/sec is the indicator injection rate and Tinj is the duration of injection. Therefore m is 0Tinjm.(t)dt. For consistency of approach, F(t) should be considered to be the instantaneous average flow at a cross section of the vessel, the injection to be flow labeling, and, therefore, the concentration, yo(t), to be an instantaneous flow-weighted average across the cross section. Similarly the output concentration, y6(t), is considered to be the average concentration in the cross section (3). The transport function or unit impulse response of the model, h(t), is y6(t) when the input function, y0(t), is an impulse or very short pulse of high amplitude and unit area. In this study, the injection rate was constant at 8 mg/sec for 0.5 seconds, the total amount of indicator injected being 4 mg.

Flow, F(t)

For the simulation the flow was sinusoidal:

F(t)=F[1+Ksin(ωt+ϕ], (3)

in which F̄ is the mean flow over one cycle, K = amplitude of sinusoidal variation, ω = frequency of oscillation in flow (radians/sec). F(t) was limited to positive values when K was greater than 1.0. ϕ was defined to be zero when the midpoint of the injection coincided with F(t) = F̄ during the part of the cycle when F(t) was increasing. Another reference flow used for the analysis was F̄u, defined as the mean flow from the time of injection until the time at which the concentration had diminished to 1% of its peak value.

Estimation of F, t̄, and σ

Estimates were obtained from the curves with unsteady flow by using the output concentration-time curve, C’(t), with the assumption of steady flow and a linear system:

Fu=m/0C(t)dt. (4a)
tu=(0tC(t)dt/0C(t)dt)tinj. (4b)
Vu=Futu. (4c)
σu=[(0C(t)(ttu)2dt/0C(t)dt)σinj2]12. (4d)

As indicated by equations 4b and 4d, the mean transit time t̄u and standard deviation σu are corrected for delay and dispersion of the input function at the injection site. No extrapolation of the downslope was necessary, since recirculation was not simulated.

Results

Adequacy of the Model

In Figure 1 are shown the output of the model, y6(t) or C’(t), superimposed on a dye dilution curve, C(t), recorded using a sampling system with a 0.1-second mean transit time from the ascending aorta of a dog after a slug injection of 1.25 mg of indocyanine green into the pulmonary artery. The signal obtained from the flowmeter (Carolina Medical Electronics, Winston-Salem, N. C.) on the ascending aorta was used as F(t) in equation 1c. In view of the fact that the animal’s system between injection and sampling points is a variable-volume system and the model is a constant-volume system, the model is considered to give a remarkably good fit to the data.

FIGURE 1.

FIGURE 1

Comparison of model output, y0(t), labeled S with an experimental dye curve from the ascending aorta, C(t), labeled E when the model was driven by the recorded flowmeter signal F(t). The parameters of the model were V = 250 ml, = 45 ml/sec, ζ1=l.l, ζ2 = 0.8, ω1/ω2 = 7.0. The input pulse to the model was 1 second in duration and was delayed 1.6 seconds compared to the actual injection.

Errors in Estimation of Flow from Dye Curves Obtained from Time-Averaged Sampling

Simulated dye dilution curves have been obtained at 250 different frequencies of sinusoidal flow (ω/ωc ranging from 0.002 to 2.0) and with six different phases of injection in order to suggest the envelope of error, and with K = 0.5. The 1,500 results therefore encompass the range of situations existing experimentally.

For convenience in making a practical interpretation of the results, we considered a working example similar to the pulmonary circulation in a dog: V = 250 ml, and F̄ = 40 ml/sec. For these values, with steady flow, t̄ = 6.25 seconds, σ = 1.73 seconds, and ωc = 0.51 cycles/sec.

Either of the two correct average flows, F̄ or F̄u, defined after equation 3, may be desired by the investigator, and therefore both will be considered. Fu is the estimate obtained by equation 4a, and Fu/F̄ is plotted versus relative frequency of flow variation in the upper panel of Figure 2. At high frequencies the error is negligible. At lower frequencies there is broadening of the envelope of error, a hypothetical pair of lines enclosing all of the lines at various phases. But at both very low and very high frequencies there is almost no error in Fu/F̄u (lower panel of Figure 2) F̄u being the average flow during the passage of the dye curve. The largest errors in Fu occur in the middle range of frequencies when the highest concentrations pass by the sampling site at a time when the flow is either fastest or slowest, causing the area under the curve to be under- or overestimated. Since the maximum error in area with various phases is almost symmetrical about the correct value, the error in Fu is skewed toward overestimation.

FIGURE 2.

FIGURE 2

Effect of frequency and phase of sinusoidal flow on ratio of calculated flow to mean flow. Actual flow is F, flow calculated by equation 4a is Fu, and F̄ and F̄u are mean flows.

The frequency at which maximal errors occur is an expression of the fundamental frequency of the dilution curve and is related to the spatial dispersion of the bolus in the flowing fluid. The maximal error occurs at ω/ωc = 0.15. This broad generalization is independent of the particular parameters of the model and indeed of the model itself, σ/t̄ and ωc have a roughly reciprocal relation, and the data of Figure 2 would be little affected by a shift in σ/t̄. Only if the form of h(t̄) were dramatically different in different segments of the circulation would this generalization not hold, but dramatic differences have not been seen.

Effect of Amplitude of Variation in Flow on Areas of Dye Curves

The envelope of error in Figure 2 would be broader if K were larger than 0.5. But the relationship between the amplitude of the fluctuation in flow and the magnitude of the error produced is not simply linear (Fig. 3). Au and Au are, respectively, the areas of the dilution curves with flow unsteady and with flow steady at F(t) = F̄u. Figure 3 presents a plane of Au/Au versus K at ω/ωc = 0.15. The difference between Au and Au usually increases with increasing amplitude, but is clearly not linear.

FIGURE 3.

FIGURE 3

Effect of amplitude of flow variation on area of dilution curve.

There is also an asymmetrical distribution of the error at different phases. The area is dependent on the length of time each portion of the bolus in the blood stream remains at the sampling site. If the portion of the bolus with the highest concentration passes the sampling site when flow is at a low point in the cycle, the area is overly large (for example, at ϕ = 2π/3) compared to that at steady flow. When the bolus is not very broad relative to the distance traveled along the stream in one cycle, it is most likely that the peak will pass when flow is fast and therefore that the resultant areas will be small (ϕ = π, 4π/3, 5π/3, or 0).

Errors in Estimation of t̄ from Dye Curves Obtained by Time-Averaged Sampling

If flow were steady, t̄ would be V/F. In the upper panel of Figure 4 is plotted the ratio t̄u/V/F̄. The errors are maximal at low frequencies, particularly when injection was made when the flow was near its minimum (ϕ = 4π/3 or ϕ = 5π/3). At respiratory frequencies (around ω/ωc = 0.2 in our working example), the errors are small relative to those at the lowest frequency. At high frequencies there is essentially no error in t̄u.

FIGURE 4.

FIGURE 4

Effect of frequency and phase of sinusoidal variation in flow on estimated mean transit time. Ordinate is scaled twice as large on lower panel as on upper panel.

When F̄u, is considered to be the correct reference flow (Fig. 4, bottom), the differences between t̄u and tu which equals V/F̄u, and is the mean transit time which would have existed had the flow been steady at F̄u) are small at very low and at high frequencies; maximal errors occur at the mid range of frequencies. However, the largest errors in tu/t̄u occur not at ω/ωc = 0.15 to 0.2 as did errors for Fu but at ω/ωc = 0.1 and are less than for Fu/F̄u.

Errors in Estimation of σ from Dye Curves Obtained by Time-Averaged Sampling

In Figure 5 is plotted the relative dispersion σu/t̄u as a ratio to that which would have been obtained at the steady flow, σs/t̄s, (which is identical to σu/t̄u). As with t̄u, maximum error occurs around ω/ωc = 0.1. The usefulness of the ratio σ/t̄ is emphasized because the error in relative dispersion is much less than that for the actual dispersion σ.

FIGURE 5.

FIGURE 5

Effect of frequency and phase of sinusoidal variation in flow on relative dispersion of dye dilution curve. In this hypothetical situation described by the model, spatial dispersion of indicator is entirely independent of flow, and therefore frequency of variation in flow has no influence on actual spread of bolus with respect to distance along segment of circulation.

Errors in Estimation of V

The ratio of Vu (calculated by equation 4c) to V is shown in Figure 6. Errors in Vu are very large when the errors in t̄ and in Fu are both in the same direction—for example, with ϕ = π at ω/ωc = 0.15. Whereas the largest errors occur between ω/ωc = 0.02 and 0.20 and are mainly overestimates, at other frequencies the errors in Fu and t̄u or in Fu/F̄ and t̄u/V/F̄ are in opposite directions, which reduces the error in Vu. This is particularly apparent at very low frequencies.

FIGURE 6.

FIGURE 6

Effects of phase of sinusoidal variation in flow on estimates of mean transit time volumes.

Reliability of Mean Values

The data on area and flow, as exemplified by Figures 2 and 3, illustrate that in the presence of unsteady flow randomization of the phase of injection results in several estimates whose mean may differ from the true mean. Figure 2 shows that when ω/ωc is between 0.02 and 0.20 the mean flow will be significantly overestimated. At ω/ωc = 0.15, the mean of the six estimates is 1.15. When ω/ωc is less than 0.02, the average Fu is F̄. Figure 3 shows that the magnitude of the biasing of the mean can only be approximately linearly related to the amplitude K. That biasing toward lower-than-correct mean estimates can also occur is seen for Fu/F̄u at ω/ωc = 0.3 (Fig. 2), where the mean of the six estimates is about 0.97.

However, the mean ratio of Fu/F̄u or Au/Au is not very informative because the distribution of errors at a given frequency is bimodal or trimodal, depending on the frequency at which the distribution is examined and on the nature of the flow. In Figure 7 are shown the distributions of error in Au/Au, Fu/F̄u, t̄u/tu, and Vu/V at ω/ωc = 0.15 with all ϕ equally likely. The peaks of the probability density function mainly occur immediately adjacent to the envelope of error and occur when the peak concentration of the bolus is passing by the sampling site at either maximal or minimal flow. This occurs most obviously at ω/ωc = 0.15, since the sampling site is one-half wavelength downstream, at which point the magnitude of the errors is maximal. A similar result was obtained by Cropp and Burton (4) for constant-rate injections. (A wavelength is considered to be the mean distance traveled by the blood and indicator during one cycle.) At ω/ωc = 0.15, the probability of obtaining estimates within ± 10% of the correct value was 13.5% for Au, 12.6% for Fu, 35% for t̄u, and 18% for Vu. The probability of overestimating the mean was 30% for A, 70% for Fu, 52% for t̄u, and 62.5% for Vu.

FIGURE 7.

FIGURE 7

Probability density functions of errors in estimating A, F, t̄, and V from dye dilution curves during unsteady flow. P is probability of occurrence per 0.01 unit, so total area of each curve is 1.0. For these curves, ω/ωc = 0.15, K = 0.5, and injections were made at all ϕ with equal probability. (Data obtained from 200 solutions with ϕ at steps of 0.01πfrom 0 to 2π.)

Discussion

General

A sinusoidal F(t) was used in this study because its general form is close to a wide variety of flow patterns and also because, even when F(t) in nature is not sinusoidal, the fundamental frequency of F(t) is the important source of error rather than the higher-frequency harmonic content.

The prominent sources of variation in flow occur with cardiac contraction and with pulmonary ventilation, and thus the frequencies of these sources of error are readily determined. At any given frequency, the importance of the flow variation depends on its amplitude. In general, cardiac frequencies are sufficiently high that even large amplitude fluctuations are of little consequence, but respiratory variations in flow, although of relatively low amplitude, are quite important because of their lower frequency.

Fluctuations at Cardiac Frequencies

The flow at the root of the aorta during the cardiac cycle varies from zero to almost five times the mean flow (15). In a peripheral artery the range is usually less (0.2 to 2.0 times the mean) but is very dependent on the physiologic state of the subject. The errors are exacerbated when there is reversal of flow as can occur in th abdominal aorta during diastole when splanchnic flow is high (16), a situation which we could not simulate using the methods described. At normal heart rates, systole is shorter than diastole, a situation which could produce larger errors than sinusoidal flow and which can be demonstrated by using aortic flow as F(t), as shown in Figure 1.

Nevertheless, in general, flow variation at cardiac frequencies produces little error; Figures 2,6 show this, for the estimates are very nearly constant when ω/ωc is greater than 1.0 or when the cycle length is less than a quarter of the dye curve passage time. Accordingly, when using dye dilution curves in two hydraulic circulation models with varying degrees of valvular regurgitation, Hoffman and Rowe (17, 18) observed no significant differences between estimates obtained during steady flow and those obtained with flow pulsating at even the rather low rates of 40 to 50/min.

Fluctuations at Low Frequencies

Respiratory fluctuations have lower amplitude but are more important because the frequencies are lower. Hoffman and co-workers (19) observed the maximal stroke volume during inspiration in awake dogs to be “sometimes 50% greater than the minimal stroke volume in expiration, but as a rule the difference was the order of 10-20%.” In humans, Vermeire and Butler (20) observed pulmonary capillary flows during inspiration as much as four times the flow during expiration. From their Table 1 we calculate that the amplitude of respiratory variation in flow (the difference between maximum and minimum, divided by two times the average) averaged 24% in their 15 subjects and ranged up to 62%. Gabe and co-workers (15) observed 25% fluctuations in stroke volume in the pulmonary artery of humans with deep respirations and 20% in the aorta.

These remarks are made to emphasize that the theoretical effects of variations in flow with the respiratory cycle are in fact seen in reality. Of equal importance in animals in poor condition or in humans in shock are the flow variations that occur with Mayer waves. Killip (21) observed up to twofold fluctuations in aortic flow over cycles lasting 15 to 40 seconds. Knowing that low cardiac output accompanies such states, we may predict with surety that these fluctuations would interfere with accurate estimation of flow, because the relative frequency, ω/ωc would be in the range 0.05 to 0.2.

Experimental Estimates of Errors with Unsteady Flow

The results of this study provide a theoretical background for considering the variation in estimates of area or flow in human and animal studies, and permit us to estimate the concentration that time-averaged sampling in the presence of unsteady flow makes to the total variability in the estimates obtained by the dye dilution method.

For example, in a previous study, dilution curves were obtained simultaneously from pairs of sampling sites in the dog (22) while the ventilation rate was 40/min. The standard deviation of the difference between the areas of the paired curves was about 6%. Injections were random with respect to respiratory phase. Mean transit times from central venous injection sites to aortic sampling sites were 4 to 10 seconds (corrected for the sampling-system transit) at different flow rates. By equation 6, using σ/t̄ = 0.45 (10), ωc is 0.50 to 1.20 cycles/sec. Therefore, ω/ωc for respiratory variation was 0.6 to 1.4, and it is reasonable to conclude that respiration produced only minor errors in the results of that study, and that the standard deviations of 5 to 7% were due mainly to spontaneous variations in flow at lower frequency. Injection upstream from the lung reduces errors due to nonmixing, and streaming at the aortic sampling site is probably of little importance when the bolus takes many heart beats to pass.

If the residual variation in a carefully controlled study in anesthetized animals was 7%, then it is not surprising that, in studies done on awake humans to compare the Fick and dye techniques, the combination of spontaneous variation in flow and errors due to time-averaged sampling should result in standard deviations of 10 to 15% (23-25). A similar standard deviation was found by Hamilton and associates (26) in a comparison of the dye technique with readings from an electromagnetic flowmeter.

The extensive observations of Opdyke (27) can probably be explained in the light of our model experiments. He observed what he considered to be an extraordinarily large variance in the ratios of the values for cardiac output calculated from dye curves sampled from the descending aorta or from the left atrium to those values calculated from dye curves obtained simultaneously from the pulmonary artery. His Figure 2 shows non-Gaussian, highly platykurtotic distributions of ratios, which are comparable to those in our Figure 7 except that the theoretical bimodality with the peaks at the extremes of the distribution is not apparent. This is not surprising in view of the presence of more than one source of variation and the fact that he was comparing flow estimates from two sampling sites rather than an estimated flow with a true mean flow. It is particularly interesting that he noted that the ratios of flow estimates obtained when the injection was made 0.1 second after the onset of inspiration (at 10 to 12/min) averaged 1.08 (LA sampling) and 1.23 (descending aorta), and when the injection was made 4.0 seconds after the onset of inspiration averaged 0.94 (LA) and 0.89 (aorta). These observations substantiate the theoretical results of our Figure 2: at ω/ωc of about 0.2 and with the injection early in the inspiratory phase (i.e., ϕ = 0, π/3 or π/2) then the cardiac output, FPA, calculated from the PA curve, would be underestimated. The corresponding estimate, Fa0, calculated from the relatively dispersed dye curve sampled simultaneously from the descending aorta, for which ω/ωc is likely less than 0.1, would have less error, i.e., from our Figure 1 the values for Fu/F̄u — 1.0 would be smaller. Therefore Opdyke’s ratio, Fa0/FPA, would be predicted to be greater than 1.0 with injection in this phase. With injection 4 seconds after the onset of inspiration (ϕ = π or 4/3 π in our Fig. 2), FPA would be markedly overestimated, resulting in the low values of Fa0/FPA that he observed.

Practical Applications of the Theory

For different segments of the circulatory system, the skewness and kurtosis vary little, but the relative dispersion, σ/t̄, differs from region to region. But the important fact is that for a given segment σ/t̄ remains almost independent of flow (8-10). This permits us to estimate ω/ωc and to devise means for reducing the errors.

On the graphs showing error in estimates, ω/ωc has been used as the abscissa. When ωc can be obtained in an experimental situation, then the possible importance of an interfering frequency, ω, can be estimated from the graphs in Figures 2 to 7. If the amplitude of flow variation can be estimated, then an approximate estimate of the probable error can be obtained and its accuracy improved by estimating ϕ. When a system has transport functions at different steady flows which have constant shape (constant σ/t̄, skewness, etc.) then ωc is reciprocally related to t̄:

ωc=Rt(radians/sec). (5)

For the model parameters used in this study (σ/t̄ = 0.28 for example), R’ is 20 radians. This type of relationship can be generalized somewhat approximately for a family of systems having right-skewed transport functions and a range of relative dispersions:

ωcR(tσ)t=Rσ, (6)

where R is about 5.6 radians. Using this, one can obtain ωc from σ above or from t̄ and σ/t̄. Equation 6 is an extrapolation from equation 5 and is applicable over the narrow range of σ/t̄’s observed in the circulation.

The t̄ to use in equation 6 is the mean transit time through the segment of the system in which flow is unsteady, that is, the t̄ of the recorded curve minus that of the sampling system. In experimental situations, other corrections may be important. For example, when sampling from the radial artery after dye injection into the pulmonary artery, the subclavian and brachial arteries may be considered as an extension of the sampling system. If the flow in these peripheral arteries is constant, then their t̄ also should be subtracted in order to find t̄ through the central circulation only. If the peripheral arterial flow is a constant fraction of aortic flow, thus providing ideal flow proportional sampling (3), then no such correction should be made. Ordinarily one cannot ascertain the steadiness of peripheral arterial flow and therefore advancing the sampling catheter to the ascending aorta eliminates the uncertainty. The principle is that, to obtain the pertinent t̄, sampling should be from a segment through which all of the injected indicator passes—or better still, from the most downstream point of the artery in which flow is proportional to aortic flow.

Reduction of Error

When it is recognized that flow is varying and that ω/ωc is 0.05 t 0.30 or when the cycle length is one-half to three times the pertinent mean transit time, a number of methods are available for reducing this source of error. These are (1) increase the interfering frequency, (2) decrease the interfering frequency, (3) reduce the amplitude at the interfering frequency, (4) move the injection and sampling sites closer together to increase ωc (effectively decreasing the interfering frequency toward ω/ωc = 0.06), (5) increase the distance between the injection and sampling sites (decreasing ωC and effectively increasing the interfering frequency toward ω/ωc = 0.6), (6) prolong the duration of injection, and (7) average several estimates.

The correct choice of maneuver is important, for a wrong choice can have drastic effects on the error. For example, if the interfering frequency were at ω/ωc = 0.30, a shortening of the distance between injection and sampling sites which doubled ωc would result in the interfering frequency occurring at ω/ωc = 0.15, the position of maximal error, and would produce a quadrupling of the magnitude of the envelope of error compared to that at ω/ωc of 0.30.

  1. Increase the frequency. Increasing respiratory rate is practical because not only does the envelope of error fall off quickly at high frequencies but also the tidal volume and the amplitude of variation in cardiac output diminish at higher rates. This improves the estimation of F̄ and also of F̄u. This, therefore, is the recommended mode of reducing the respiratory influence on the amplitude of variation in cardiac output and in the error of estimation of flow by the standard sluginjection dye dilution technique.

  2. Decrease the frequency. If the interfering frequency is respiratory, one may ask the patient to breathe more slowly or, if the patient is anesthetized, the ventilation rate may be reduced. If tidal volume were also reduced this would help, but the method may not be very effective if, in order to maintain the respiratory minute volume, the tidal volume must also be increased. In general this method is not very useful because, even though it improves the accuracy of the estimation of F̄u, it decreases the accuracy of estimation of F̄.

  3. Reduce the amplitude. This is very useful in those patients undergoing cardiac catheterization who frequently tend to have sighing respiration or to hyperventilate and who, on request, will breathe quietly and shallowly during the period of recording of the dilution curve. However, one cannot go to the extreme of asking the subject to stop breathing during the recording of the curve, for he may then unwittingly perform a Valsalva or a Mueller maneuver and change the flow. Also, if continuous sampling densitometry is used, the densitometer base line will shift as CO2 accumulates, changing erythrocyte size and blood optical density.

  4. Move injection and sampling sites closer together (reduce intravascular transit time). This has the disadvantage that, as ωc is increased to avoid respiratory fluctuations in flow, fluctuations at cardiac frequencies gain importance. Despite this, the maneuver may still be useful because, in shortening the segment length, the absolute dispersion (not the relative dispersion) is reduced and the separation of the dye passing the sampling site for the first time from that recirculating can be made with greater precision. Therefore, in some cases more accuracy may be gained than lost. However, as with method 2, the accuracy of estimation of F̄u is improved more than that of F̄.

  5. Increase the distance between injection and sampling sites (increase intravascular transit time). By shifting ωc to a lower value, the interfering respiratory and cardiac cycles both are shifted to the right toward ω/ωc = 0.6 or beyond, and the errors are reduced. The risk is that the accuracy of extrapolation to exclude recirculation may be diminished.

  6. Prolong duration of injection. When injection and sampling sites are close together and ωc is high, some reduction in errors is obtained by prolonging the injection. This is most effective in situations in which recirculation does not obscure the downslope of the curve. For example, with K = 0.5 at ω/ωc = 0.2 (the region of maximal error), the minimum to maximum range of Fu/F̄ at TinJ = t̄/16, t̄/4, t̄, and 4t̄ was 0.6 to 1.6, 0.7 to 1.5, 0.8 to 1.4, and 0.9 to 1.3, respectively. (Figures 2 through 7 were obtained with TInJ = t̄/12.5.)

  7. Averaging of estimates. One might attempt to improve the estimates of F, A, t̄, and V by making several observations with injections at different phases and averaging them. With distributions such as those in Figure 7 it is apparent that the mean will have a systematic bias and a large standard error. Biasing of the average cannot be avoided in the midrange (0.02 to 0.20) of interfering frequencies; but as Katz and Shinnar have pointed out (S. Katz and R. Shinnar, personal communication), it is minimized by averaging the variable itself rather than its components, i.e., the best estimate of V is ΣVu/N, where N is the number of observations, rather than the average (ΣFu/N) · (Σt̄u/N). Similarly, when dye curves have been obtained by sampling two or more sites simultaneously, dividing the amount injected by the average of the areas of the dilution curves is useful, since the sampling sites will be at different distances or wavelengths downstream.

Approximation to Volume-Averaged Sampling

Errors due to unsteady flow cannot be completely avoided with the techniques suggested above, and it is therefore highly desirable to obtain a volume-averaged sample, one in which the dye-blood mixture is sampled in proportion to the flow in each part of the cross section of the vessel. One method for converting time-averaged sampling to approximate volume-averaged sampling is to use a velocity signal from the flow stream near the sampling site. The equation of conservation of the injected material passing by the sampling site for the first time is

m=0F(t)C(t)dt. (7)

Dividing both sides by the mean flow, F̄u, taking the reciprocals, and multiplying by the injected dose, m, leads to

Fu=m0[F(t)Fu]C(t)dt (8)

The ratio, F(t)/F̄u, is a weighting factor for the instantaneous concentrations, C(t); it is large when F(t) is large and zero when the flow is stopped. Since this ratio is defined as having a mean value of 1.0 over the passage time of the dilution curve, it can be obtained from a linearly flow-proportional signal which need not be calibrated in absolute terms. This can be derived from the pressure difference between two orifices of a double-lumen catheter (22) or from a velocity transducer (28). Equations similar to equation 8 can be used for t̄ and σ While such a method can be expected to provide a vast improvement in accuracy, the problems of cross-sectional averaging of flow and of concentration remain to plague the perfectionist.

When this technique is applied to the dye curves in Figure 1, the estimates of F̄u, t̄, σ/t̄, and V were 99.6%, 99.0%, 99.2%, and 98.6% of those given by the classical equations. For this curve, ωc, was calculated by equation 6 to be 0.4/sec. The heart rate and respiratory rates were 2.08/sec and 0.66/sec, giving values of ω/ωc of 5.2 and 1.65 at these frequencies. At these values of ω/ωc the errors in the estimates should theoretically be small, as the above observed values suggest.

Determination of Flow-Dependent Transport Function

The assumptions that ωc is proportional to F(t) and that other shaping parameters of h(t) do not change are reasonable for fairly wide ranges of estimates of flow in large vessels and for the pulmonary and renal circulations. However, dispersion, skewness, and kurtosis of h(t) would vary whenever a change in flow results in a change of velocity profile in large vessels or in a change of the distribution of regional flows in a capillary network. The approximation can be useful in obtaining a description of the transport function when F(t)/F̄u is known. General treatments have been given by Stephenson (1) and Sherman (29), and a closer approach to the problem was provided by Zadeh (30). His treatment can be simplified considerably when the injection is an impulse, δ(t), and when concentration and flow are the same across the cross section of the vessel.

The present model can be used moderately well even if F(t)/F̄u is not completely described. If respiratory and cardiac frequencies are known or recorded along with concentration-time curves from the input and output ends of the segment, then F(t) may be considered as the sum of two periodic functions of unknown amplitude but of approximately known form. With the upstream curve used as input, y0(t), the parameters of the differential operator may be varied to obtain optimal matching of y6(t) with the curve recorded at the downstream point. Even with imperfect estimates of F(t), this should provide much more accurate estimations of the circulatory transport function than could be obtained by assuming that F(t) = F̄.

Acknowledgment

The authors appreciate the help of Miss Linda Staab and Mrs. Jean Frank in the preparation of the illustrations and of Mrs. Jane Irving in the preparation of the manuscript. Dr. D. E. Donald assisted in the animal experiments.

This investigation was supported in part by U. S. Public Health Service Grants HE-9719, HE-4664, and FR-00007 from the National Institutes of Health. Dr. Bassingthwaighte is the recipient of a Career Development Award (K3HE22649).

References

  • 1.Stephenson JL. Theory of the measurement of blood flow by the dilution of an indicator. Bull Math Biophys. 1948;10:117. doi: 10.1007/BF02477486. [DOI] [PubMed] [Google Scholar]
  • 2.Meier P, Zierler KL. On the theory of the indicator-dilution method for measurement of blood flow and volume. J Appl Physiol. 1954;6:731. doi: 10.1152/jappl.1954.6.12.731. [DOI] [PubMed] [Google Scholar]
  • 3.González-Fernández JM. Theory of the measurement of the dispersion of an indicator in indicator-dilution studies. Circ Res. 1962;10:409. doi: 10.1161/01.res.10.3.409. [DOI] [PubMed] [Google Scholar]
  • 4.Cropp GJA, Burton AC. Theoretical considerations and model experiments on the validity of indicator dilution methods for measurements of variable flow. Circ Res. 1966;18:26. doi: 10.1161/01.res.18.1.26. [DOI] [PubMed] [Google Scholar]
  • 5.Stow RW. Systematic errors in flow determinations by the Fick method. Minn Med. 1954;37:30. [PubMed] [Google Scholar]
  • 6.Visscher MB, Johnson JA. The Fick principle: Analysis of potential errors in its conventional application. J Appl Physiol. 1953;5:635. doi: 10.1152/jappl.1953.5.10.635. [DOI] [PubMed] [Google Scholar]
  • 7.Zierler KL. Theory of the use of arteriovenous concentration differences for measuring metabolism in steady and non-steady states. J Clin Invest. 1961;40:2111. doi: 10.1172/JCI104437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Bassingthwaighte JB. Plasma indicator dispersion in arteries of the human leg. Circ Res. 1966;19:332. doi: 10.1161/01.res.19.2.332. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Bassingthwaighte JB, Ackerman FH. Mathematical linearity of circulatory transport. J Appl Physiol. 1967;22:879. doi: 10.1152/jappl.1967.22.5.879. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Knopp TJ, Bassingthwaighte JB. Effect of flow on transpulmonary circulatory transport functions. J Appl Physiol. 1969;27:36. doi: 10.1152/jappl.1969.27.1.36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Bassingthwaighte JB, Ackerman FH, Wood EH. Applications of the lagged normal density curve as a model for arterial dilution curves. Circ Res. 1966;18:398. doi: 10.1161/01.res.18.4.398. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Sheppard CW. Mathematical considerations of indicator dilution techniques. Minn Med. 1954;37:93. [PubMed] [Google Scholar]
  • 13.Thompson HK, Jr., Starmer CF, Whalen RE, Mcintosh HD. Indicator transit time considered as a gamma variate. Circ Res. 1964;14:502. doi: 10.1161/01.res.14.6.502. [DOI] [PubMed] [Google Scholar]
  • 14.Knopp TJ, Anderson DU, Bassingthwaighte JB. SIMCON: Simulation control to optimize man-machine interaction. Simulation. 1970;14:81. doi: 10.1177/003754977001400205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Cabe IT, Gault JH, Ross J, Jr., Mason DT, Mills CJ, Shillingford JP, Braunwald E. Measurement of instantaneous blood flow velocity and pressure in conscious man with a catheter-tip velocity probe. Circulation. 1969;40:603. doi: 10.1161/01.cir.40.5.603. [DOI] [PubMed] [Google Scholar]
  • 16.McDonald DA. Blood Flow in Arteries. Williams & Wilkins Co.; Baltimore: 1960. p. 328. [Google Scholar]
  • 17.Hoffman JIE, Rowe GG. Some factors affecting indicator dilution curves in the presence and absence of valvular incompetence. J Clin Invest. 1959;38:138. doi: 10.1172/JCI103782. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hoffman JIE. Calculation of output, central volume and variance from indicator-dilution curves. J Appl Physiol. 1960;15:535. doi: 10.1152/jappl.1960.15.3.535. [DOI] [PubMed] [Google Scholar]
  • 19.Hoffman JIE, Guz A, Charlier AA, Wilcken DEL. Stroke volume in conscious dogs: Effect of respiration, posture and vascular occlusion. J Appl Physiol. 1965;20:865. doi: 10.1152/jappl.1965.20.5.865. [DOI] [PubMed] [Google Scholar]
  • 20.Vermeire P, Butler J. Effect of respiration on pulmonary capillary blood flow in man. Circ Res. 1968;22:299. doi: 10.1161/01.res.22.2.299. [DOI] [PubMed] [Google Scholar]
  • 21.Killip T., III Oscillation of blood flow and vascular resistance during Mayer waves. Circ Res. 1962;11:987. doi: 10.1161/01.res.11.6.987. [DOI] [PubMed] [Google Scholar]
  • 22.Bassingthwaighte JB, Edwards AWT, Wood EH. Areas of dye-dilution curves sampled simultaneously from central and peripheral sites. J Appl Physiol. 1962;17:91. doi: 10.1152/jappl.1962.17.1.91. [DOI] [PubMed] [Google Scholar]
  • 23.Hamilton WF, Riley RL, Attyah AM, Cournand A, Fowell DM, Himmelstein A, Noble RP, Remington JW, Richards DW, Jr., Wheeler NC, Witham AC. Comparison of the Fick and dye injection methods of measuring the cardiac output in man. Amer J Physiol. 1948;153:309. doi: 10.1152/ajplegacy.1948.153.2.309. [DOI] [PubMed] [Google Scholar]
  • 24.Miller DE, Gleason WL, McIntosh HD. Comparison of the cardiac output determination by the direct Fick method and the dye-dilution method using indocyanine green dye and a cuvette densitometer. J Lab Clin Med. 1962;59:345. [PubMed] [Google Scholar]
  • 25.Werkö L, Lagehlöf H, Bucht H, Wehle B, Holmgren A. Comparison of the Fick and Hamilton methods for the determination of cardiac output in man. Scand J Clin Lab Invest. 1949;1:109. [Google Scholar]
  • 26.Hamilton FN, Minzel JC, Schlobohm RM. Measurement of cardiac output by two methods in dogs. J Appl Physiol. 1967;22:362. doi: 10.1152/jappl.1967.22.2.362. [DOI] [PubMed] [Google Scholar]
  • 27.Opdyke DF. Agreement of cardiac outputs calculated from paired indicator-dilution curves. J Appl Physiol. 1965;20:9. doi: 10.1152/jappl.1965.20.1.9. [DOI] [PubMed] [Google Scholar]
  • 28.Fry DL, Griggs DM, Jr., Greenfield JC., Jr. In vivo studies of pulsatile blood flow: Relationship of the pressure gradient to the blood velocity. In: Attinger EO, editor. Pulsatile Blood Flow. McGraw-Hill Book Co., Inc.; New York: 1964. pp. 101–114. [Google Scholar]
  • 29.Sherman H. On the theory of indicator-dilution methods under varying blood-flow conditions. Bull Math Biophys. 1960;22:417. [Google Scholar]
  • 30.Zadeh LA. Frequency analysis of variable networks. Proc IRE. 1950;38:291. [Google Scholar]

RESOURCES