Multimodal coherent anti-Stokes Raman spectroscopic imaging with a fiber optical parametric oscillator

Yan-Hua Zhai 1, Christiane Goulart 1, Jay E Sharping 1,a), Huifeng Wei 2, Su Chen 2, Weijun Tong 2, Mikhail N Slipchenko 3, Delong Zhang 3, Ji-Xin Cheng 3,a)
PMCID: PMC3114562  PMID: 21677908

Abstract

We report on multimodal coherent anti-Stokes Raman scattering (CARS) imaging with a source composed of a femtosecond fiber laser and a photonic crystal fiber (PCF)-based optical parametric oscillator (FOPO). By switching between two PCFs with different zero dispersion wavelengths, a tunable signal beam from the FOPO covering the range from 840 to 930 nm was produced. By combining the femtosecond fiber laser and the FOPO output, simultaneous CARS imaging of a myelin sheath and two-photon excitation fluorescence imaging of a labeled axons in rat spinal cord have been demonstrated at the speed of 20 μs per pixel.


Coherent anti-Stokes Raman scattering (CARS) microscopy is a powerful tool for high-speed vibrational imaging of cells, tissues, and drug delivery systems.1, 2 An additional advantage of CARS microscopy is that other linear and nonlinear imaging modalities such as confocal Raman microscopy and multiphoton fluorescence microscopy can be combined with CARS on a single microscopy platform.3, 4, 5, 6 Despite its, the technical complexity of CARS microscopy and the high cost of laser systems have hindered its widespread adaptation. Developing laser sources is an integral part of the evolution of CARS microscopy toward the ultimate goal of providing a reliable, inexpensive, and user friendly microscopy system.

First-generation CARS microscopes used two temporally synchronized femtosecond (fs) or picosecond (ps) pulsed lasers.1 Second-generation CARS systems used optical parametric oscillators (OPOs) pumped with fs or ps solid state lasers to generate multiple inherently synchronized pulse trains without the need for electronic synchronization.7, 8 Fiber-based laser systems are attracting attention because they provide advantages in terms of size and complexity.9, 10, 11 Notable progress is being made on developing all-fiber laser sources for CARS, including: supercontinuum generation in photonic crystal fiber (PCF) (Ref. 10) and a two-color Er:fiber laser system.11 The average powers for these systems is limited to ∼10 mW for the Stokes beam, and a larger power from the pump beam is necessary to achieve high speed CARS imaging.

In this letter we report the application of a fiber-laser pumped, wavelength-tunable, pulsed fiber optical parametric oscillator (FOPO) (Refs. 12, 13, 14, 15) to multimodal CARS microscopy. Our FOPO resembles a state-of-the-art χ(2) OPO except that the crystal is replaced by a PCF and the FOPO operates through four-wave mixing (FWM) mediated by χ(3) nonlinearity of glass. The fiber’s dispersive properties are chosen to phase match the FWM process leading to a wide spectral bandwidth of operation. One distinct advantage of FOPOs over χ(2) OPOs is the ability to generate output at wavelengths near to that of the pump laser, which reduces the number of frequency conversion steps required for the generation of wavelengths for CARS microscopy. Furthermore, we anticipate that the fiber-based gain medium can be integrated with other fiber components, leading to a compact all-fiber FOPO system.

The optical layout is depicted in Fig. 1. The fiber laser (Uranus, PolarOnyx) produces a 50 MHz pulse train (fundamental beam) at 1032 nm with 2 W of average power and 300 fs pulse duration after a grating compressor. A half-wave plate in combination with a polarization beam splitter (PBS) controls the power coupled into the PCF which serves as the gain medium within the FOPO cavity. The PBS-reflected portion of the fundamental beam, ωp, is chosen as the Stokes beam and the FOPO short wavelength output, ωs, is used as the pump beam for the CARS microscopy.

Figure 1.

Figure 1

Schematic of a FOPO-based multimodal microscope. PBS is a polarization beam splitter cube. M1 is a 1064 nm short pass mirror. M2 is a 1064 nm long pass dichroic mirror. M3 is a protected silver cavity end mirror. M4 is a 940 nm short pass dichroic mirror. PP represents a prism pair. F1-F3 are bandpass filters. SB stands for a laser scanning box. PMT1 and PMT2 are photomultipliers. The inset shows energy diagrams for CARS and TPEF modalities.

The folded FOPO cavity, which is very similar to our previous designs,14, 15 is 3.07 m long, matching the effective cavity length of the fiber laser. Mirrors M1 and M3 are cavity end mirrors and M3 is mounted on a translation stage for fine tuning of the cavity length and output wavelength. The closely spaced prism pair (PP) is inserted inside the cavity to enhance the positive dispersion. The 1032 nm beam from the fiber laser is coupled into the PCF through an aspheric lens (C230TM-B, Thorlabs). Collimation of the output from the PCF is done using a similar aspheric lens with a C-coating. The output coupler is a long pass dichroic mirror, M2, (Long Pass 1064, Precision Photonics) which delivers the short wavelength beam, ωs, out of the cavity. A half-wave plate placed before the FOPO cavity is used to match the 1032 nm fundamental-beam polarization with the PCF’s principle axes. Another half-wave plate is placed in the ωs beam to set the beam polarization parallel to that of the 1032 nm, ωp beam. A bandpass filter F1 (850∕90 or 880∕40, Chroma) removes the weak broadband supercontinuum from the ωs beam.

The two beams, ωs and ωp, are temporally overlapped and collinearly combined using M4 (940 dcsp, Omega). The overlapped beams are then directed into a laser scanning inverted microscope (FV300∕IX71, Olympus). We use a 60×NA=1.2 water-immersed objective with improved near infrared (NIR) transmission (UPLSAPO60XWIR, Olympus) to focus the laser into the sample. The forward CARS signal is collected with an air condenser (NA=0.55) and detected by an external photomultiplier tube PMT1 (R3896, Hamamatsu) after propagating through bandpass filters (D695∕70 or 740∕140 nm, Chroma). The two-photon excitation fluorescence (TPEF) signal is collected by the objective, separated from excitation beams with a dichroic mirror (670 dcxr, Chroma) and detected with PMT2 (R7422–40, Hamamatsu) after passing the bandpass filter (520∕40 nm, Chroma). A spectrometer is mounted on the side port provides confocal Raman analysis of the objects.6

The performance of the FOPO is shown in Fig. 2. Two PCFs (Yangtze Optical Fiber and Cable Co., Ltd., China) were separately used as the gain medium. One PCF has a zero-dispersion wavelength (ZDW) calculated to be 950 nm, a length of 3.3 cm and a core diameter of ∼5 μm (ZDW950). The other PCF has a ZDW measured to be 1030 nm, a length of 4.8 cm and a core size of ∼4.6 μm (ZDW1030). The output wavelength is tuned by slightly varying the length of the cavity in the presence of cavity dispersion. Selected FOPO output spectra are shown in Figs. 2a, 2b, respectively. The tunable range is from 840 to 860 nm for ZDW950 and from 860 to 930 nm for ZDW1030. Although one obtains more tunability with higher pump power, this comes at the expense of the generation of sidebands in the output spectrum. The output of the FOPO covers a range from 840 to 930 nm using these two fibers. The spectral full-width at half-maximum (FWHM) of the FOPO output is 10 nm. The total average power ranges from 5 to 25 mW depending on the wavelength, and power fluctuations are less than 3% over 20 min. Second harmonic generation, frequency resolved optical gating (FROG) measurements are shown in Fig. 2 for ZDW1030. The measured and retrieved spectrograms are shown in Figs. 2c, 2d, respectively. Figure 2e shows the reconstructed amplitude (black) and phase (blue) in the time domain where the FWHM pulse duration is 137 fs. The quadratic dependence on phase in the time domain indicates the presence of a linear chirp in the FOPO output.

Figure 2.

Figure 2

FOPO output performance. (a) Tuning curves for ZDW950 and (b) ZDW1030. (c) Measured FROG spectrogram and (d) retrieved FROG spectrogram. (e) Reconstructed amplitude (solid) and phase (dashed) for the FOPO implemented with ZDW1030.

As a proof of concept that FOPO systems can be useful for CARS microscopy, we recorded images of live 3T3-L1 cells cultured with 50 μm of deuterated palmitic acids (see Fig. 3). Lipid droplets inside of the cells are clearly visible in the CARS images. The FOPO system, in this case, is tuned to 850 nm in order to excite the CD2 stretching modes at 2100 cm−1. A typical transmission, and the corresponding CARS, images are shown in Figs. 3a, 3b, respectively. To verify that the signal is from CARS, we perform the same measurement with the pump and Stokes beams separated in time and with the beam frequencies tuned away from the Raman resonance. Confocal Raman microspectroscopic analysis of the lipid droplets reveal a distinctive peak near 2100 cm−1 corresponding to the CD2 stretching vibration [see Fig. 3c].

Figure 3.

Figure 3

Imaging deuterated lipid droplets in 3T3-L1 cells with FOPO-based CARS. (a) Transmission image. (b) Forward detected CARS image at 2074 cm−1. (c) Raman spectrum obtained at the point indicated by the arrow in (a). C-D stretch is indicated by asterisk. The average pump and Stokes powers were 4 mW and 25 mW at the sample position, respectively. The total integration time per pixel was 20 μs.

We further demonstrated CARS-based multimodal imaging of rat spinal cord white matter using the FOPO configured with ZDW1030. The output wavelength was tuned to 880 nm to excite C=C vibrations at ∼1650 cm−1. The tissue sample was fixed, cut to 50 μm thickness, and the axons were immuno-labeled with anti-NF-160 antibody conjugated to Alexa Fluor 488. The forward-detected CARS image in Fig. 4a shows parallelly aligned myelin fibers. The epi-detected TPEF image of the labeled axons, excited using the pump beam at 1032 nm, was obtained simultaneously and is shown in Fig. 4b. The overlaid image in Fig. 4c and the magnified area in Fig. 4d show the relative position of myelin sheaths and axons. The images were acquired at a speed of 20 μs∕pixel.

Figure 4.

Figure 4

Multimodal CARS/TPEF imaging of rat spinal cord. (a) Forward detected CARS image of myelin sheath at 1673 cm−1. (b) TPEF image of axons labeled with NF160/alex488 excited using the pump at 1032 nm. (c) Overlaid CARS and TPEF images. (d) Magnified segment of the overlaid images. The average pump and Stokes powers were 3.3 mW and 15 mW at the sample position, respectively.

Despite our promising initial results, there is room for improvement in our current system. (i) In our experiment we needed to exchange fibers. The ideal system would not require such a reconfiguration. (ii) Our system exhibits pulse-to-pulse stability which is similar to that produced using supercontinuum generation. The improved cavity design should bring stability close to that of pump laser. (iii) The FOPO output has a broad spectrum in comparison with the pump, limiting CARS efficiency. The ideal system has similar bandwidths to that of pump laser. (iv) Finally, the average output power must be increased because the current implementation barely meets the requirements for this application.

In summary, we have demonstrated multimodal CARS microscopy using a fiber laser and a FOPO. In our experiment, the FOPO generates up to 25 mW output that is tunable between 840 and 930 nm. By coupling the FOPO output with that of the fiber laser, high-quality CARS and TPEF imaging were realized. The current system allows one to probe vibrational transitions between 1050 and 2220 cm−1. Further exploration is under way to understand the fundamental limitations and improve the overall performance in terms of tuning range, spectral resolution, and output power. Finally, an all-fiber OPO cavity pumped by fiber laser will be developed to ultimately eliminate the free space design, thus providing a truly compact laser source for coherent Raman microscopy.

Acknowledgments

The authors acknowledge support from the Air Force Office of Scientific Research (AFOSR; Grant No. FA9550-09-1-0483), and the National Institute of Health (NIH; Grant No. R01 EB 7243).

References

  1. Cheng J. X. and Xie X. S., J. Phys. Chem. B 108, 827 (2004). 10.1021/jp035693v [DOI] [Google Scholar]
  2. Evans C. L. and Xie X. S., Annu. Rev. Anal. Chem. 1, 883 (2008). 10.1146/annurev.anchem.1.031207.112754 [DOI] [PubMed] [Google Scholar]
  3. Wang H. -W., Le T. T., and Cheng J. -X., Opt. Commun. 281, 1813 (2008). 10.1016/j.optcom.2007.07.067 [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Chen H., Wang H., Slipchenko M. N., Jung Y. K., Shi Y. Z., Zhu J., Buhman K. K., and Cheng J. -X., Opt. Express 17, 1282 (2009). 10.1364/OE.17.001282 [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Pegoraro A. F., Ridsdale A., Moffatt D. J., Jia Y. W., Pezacki J. P., and Stolow A., Opt. Express 17, 2984 (2009). 10.1364/OE.17.002984 [DOI] [PubMed] [Google Scholar]
  6. Slipchenko M. N., Le T. T., Chen H., and Cheng J. -X., J. Phys. Chem. B 113, 7681 (2009). 10.1021/jp902231y [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Evans C. L., Potma E. O., Puoris’haag M., Cote D., Lin C. P., and Xie X. S., Proc. Natl. Acad. Sci. U.S.A. 102, 16807 (2005). 10.1073/pnas.0508282102 [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Ganikhanov F., Carrasco S., Xie X. S., Katz M., Seitz W., and Kopf D., Opt. Lett. 31, 1292 (2006). 10.1364/OL.31.001292 [DOI] [PubMed] [Google Scholar]
  9. Kieu K., Saar B. G., Holtom G. R., Xie X. S., and Wise F. W., Opt. Lett. 34, 2051 (2009). 10.1364/OL.34.002051 [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Kano H. and Hamaguchi H., Appl. Phys. Lett. 86, 121113 (2005). 10.1063/1.1883714 [DOI] [Google Scholar]
  11. Krauss G., Hanke T., Sell A., Traeutlein D., Leitenstorfer A., Selm R., Winterhalder M., and Zumbusch A., Opt. Lett. 34, 2847 (2009). 10.1364/OL.34.002847 [DOI] [PubMed] [Google Scholar]
  12. Wong G. K. L., Murdoch S. G., Leonhardt R., Harvey J. D., and Marie V., Opt. Express 15, 2947 (2007). 10.1364/OE.15.002947 [DOI] [PubMed] [Google Scholar]
  13. Sharping J. E., J. Lightwave Technol. 26, 2184 (2008). 10.1109/JLT.2008.923274 [DOI] [Google Scholar]
  14. Sharping J. E., Pailo C., Gu C., Kiani L., and Sanborn J. R., Opt. Express 18, 3911 (2010). 10.1364/OE.18.003911 [DOI] [PubMed] [Google Scholar]
  15. Sharping J. E., Sanborn J. R., Foster M. A., Broaddus D., and Gaeta A. L., Opt. Express 16, 18050 (2008). 10.1364/OE.16.018050 [DOI] [PubMed] [Google Scholar]

Articles from Applied Physics Letters are provided here courtesy of American Institute of Physics