Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2011 Jul 5.
Published in final edited form as: J Nucl Med. 2011 Jun 16;52(7):1119–1124. doi: 10.2967/jnumed.111.088278

Hepatic Blood Perfusion Measured by 3-min Dynamic 18F-FDG PET in Pigs

Michael Winterdahl 1, Ole Lajord Munk 1, Michael Sørensen 1,3, Frank Viborg Mortensen 2, Susanne Keiding 1,3
PMCID: PMC3129486  NIHMSID: NIHMS303245  PMID: 21680685

Abstract

There is an unmet clinical need for an imaging method for quantification of hepatic blood perfusion. The purpose of the present study was to develop and validate a PET method using blood-to-cell clearance (K1) of 2-[18F]fluoro-2-deoxy-D-glucose (18F-FDG), 3-O-[11C]-methylglucose (11C-MG) or 2-[18F]fluoro-2-deoxy-D-galactose (18F-FDGal) as a measure of hepatic blood perfusion without the need of portal venous blood samples. We aimed to make the method as simple as possible with the perspective of future application to clinical studies. For this purpose, we examined the possibility of using 3-min data acquisition and a model-derived dual-input calculated from measurements of radioactivity concentrations in a peripheral artery.

Methods

Pigs (40 kg) underwent dynamic PET of the liver with 18F-FDG, 11C-MG or 18F-FDGal with simultaneous measurements of time-activity curves (TAC) in blood sampled from a femoral artery and the portal vein (PV); blood flow rates were measured in the hepatic artery (HA) and portal vein (PV) by transit-time flow-meters. Two input functions were compared: A measured dual-input and a model-derived dual-input, the latter with the PV TAC estimated from the measured arterial TAC and a previously validated one-parametric PV-model. K1 was estimated for each tracer by fitting compartmental models to the data comparing 3-min and 60-min data acquisitions and the two dual-input TACs.

Results

Agreement between K1 estimated using the measured and the model-derived dual-input was good for all 3 tracers. For 18F-FDG and 11C-MG, K1 (3-min data acquisition, model-derived dual-input and one-tissue compartmental model) correlated to the measured blood perfusion (P = 0.01 and P = 0.07, respectively). For 18F-FDGal the correlation was not significant.

Conclusion

A simplified method for quantification of hepatic blood perfusion using 3-min dynamic 18F-FDG PET or 11C-MG PET with blood sampling from only a peripheral artery was developed. Parametric K1 images were constructed and showed homogeneous blood perfusion in these normal livers.

Keywords: PET kinetics, molecular imaging, pharmacokinetics, liver PET, liver hemodynamics


Total hepatic blood flow can be measured in terms of volume pr. time unit (e.g. mL blood/min) by various methods for example using infusion of indocyanine green (1) or direct flow-meter measurements (2). Studies have shown that the hepatic blood flow tends to be increased in patients with liver cirrhosis (3) but it is not known to what extent the large heterogeneity in liver tissue histopathology (4,5) and intrahepatic portal vein pressure (6) may be reflected in intrahepatic variations in the liver tissue blood perfusion (mL blood/mL liver tissue/min). This is relevant for understanding liver pathophysiology and when planning local treatment of patients with liver diseases, inclusive cancer in the liver. Compared to other organs, in vivo measurements of the hepatic blood perfusion is, however, challenging due to the special blood supply to the liver with the major proportion coming from the portal vein (PV) which is inaccessible to blood sampling in humans and only a minor part coming from the hepatic artery (HA). A simplified method for quantification of regional variations in the hepatic blood perfusion is thus needed. Positron emission tomography (PET) is an attractive technique for this purpose since it can provide high resolution 3D-images of kinetic parameters and we can build on experience from previous studies of the dual-input of tracer to the liver (713).

We and others previously found that PET-estimates of hepatic blood-to-cell clearance (K1) of 2-[18F]fluoro-2-deoxy-D-glucose (18F-FDG) (7,8), 3-O-[11C]-methylglucose (11C-MG) (7) and 2-[18F]fluoro-2-deoxy-D-galactose (18F-FDGal) (9) are comparable with the hepatic blood perfusion measured simultaneously by surgically placed ultrasound transit-time flow-meters in pigs. Since the estimation of K1 is determined by the initial time-activity curves (TACs) in tissue and input (7) this requires blood sampling from an artery and the PV in order to calculate the correct dual-input TAC to the liver. Because the PV is inaccessible to blood sampling in humans various approaches have attempted to overcome this problem. Using dynamic H215O PET (10) and 18F-FDG PET (11) Kudomi et al. developed a non-invasive method to extract tissue-derived hepatic input functions in pigs. The authors modeled the arterial input function by assuming a rectangular tracer administration and a three-compartmental whole-body model. The PV input function was described by a one-tissue compartmental model and hepatic tracer kinetics was analyzed by compartmental models. The combined set of models was fitted to multiple liver tissue TACs using several constrained and fixed parameters. The estimated input functions were shown to compare well with measured dual-input TACs of H215O and 18F-FDG from Iozzo et al. (10).

The purpose of the present study was to develop a simplified method using a model-derived dual-input TAC calculated from arterial TAC with only two fixed parameters, namely β (min), which reflects the tracer-specific mean splanchnic transit time (12), and the mean hepatic arterial flow fraction (13). We have previously validated the replacement of the measured dual-input function with the model-derived dual-input using simulated tissue TACs (13). In the present study we estimated K1 of 18F-FDG, 11C-MG and 18F-FDGal from liver PET data of anaesthetized pigs (TACs) using measured and model-derived dual-input. We compared the K1 estimates with direct measurements of the hepatic blood perfusion and evaluated the use of 3-min data acquisition instead of the traditional 60-min data acquisition since K1 is determined by the initial data.

MATERIALS AND METHODS

The material comprised raw data from previous PET studies of six pigs that underwent both 11C-MG and 18F-FDG PET (9) and nine pigs that underwent 18F-FDGal PET (11). Data included dynamic PET recordings of the liver and measurements of tracer concentrations in blood samples collected manually from a femoral artery and the PV. The blood flow rates in the HA and PV were continuously recorded by surgically placed ultrasound transit-time flow-meters. In (9) 11C-MG PET was performed before 18F-FDG PET due to the radioactive half-lives of the two tracers. All studies were performed using an ECAT EXACT HR-47 PET camera (CTI, Knoxville,TN; Siemens Medical Systems, Inc., Hoffman Estates, IL).

Liver tissue blood perfusion (Q; mL blood/mL liver tissue/min) was calculated as the sum of the average flow-meter-measured HA and PV blood flow rates during the initial 3 minutes after tracer administration (mL blood/min) divided by the wet liver weight (g) corrected for liver tissue density (1.07 g/mL tissue).

Kinetic Analysis

PET emission data were reconstructed using filtered back projection resulting in images with 128 × 128 × 47 voxels of 2.4 mm × 2.4 mm × 3.1 mm and a central spatial resolution of 6.7 mm full width at half maximum (FWHM). The dynamic recording included 35 time frames: 18 × 10 s, 4 × 30 s, 5 × 1 min, 6 × 5 min, and 2 × 10 min (total 60 min). A volume-of-interest (VOI) (107–143 mL) was drawn in the liver keeping a distance of 1.5–2 cm to the edge of the liver and avoiding the central area where the large vessels are present and used to generate a liver tissue TAC (Cliver(t)), corrected for radioactive decay back to start of the PET recording.

The measured dual-input TAC, Cdual(t), was calculated as the flow-weighted mean of the TACs measured in arterial blood, CA(t) and PV blood, CPV(t), corrected for radioactive decay back to start of the PET recording, and the measured blood flow rates in the hepatic artery, FHA, and in the PV, FPV (7):

Cdual(t)=fHACA(t)+(1fHA)CPV(t);fHA=FHA/(FHA+FPV) Eq. 1

Since the blood sampling sites were close to the liver, no time delay was needed. The model-derived PV TAC (PV(t)) was calculated from the arterial TAC (CA(t)) by applying a PV-model for the transfer of tracer from the arterial system to the PV (12) and using population means of a tracer-specific parameter for the transfer, β̅, being 0.50 minutes for 18F-FDG, 0.57 minutes for 11C-MG and 0.82 minutes for 18F-FDGal (13):

C˜PV(t)=0th(tτ)CA(τ)dτ;h(t)=β¯/(t+β¯)2 Eq. 2

Using PV(t) and a mean HA blood flow fraction fHA of 0.25 (13), a model-derived dual-input TAC (dual(t)) was calculated as (13):

C˜dual(t)=fHA¯CA(t)+(1fHA¯)C˜PV(t) Eq. 3

The kinetic analysis was performed as VOI-based kinetic analysis followed by construction of parametric images of K1 of 18F-FDG and 11C-MG. The VOI-based data analysis focused on the estimation of K1 and its correlation with Q, measured by the flow-meters, i.e. verifying that K1 can be used as a measure of Q. For 18F-FDG and 18F-FDGal, we used a two-tissue compartmental model of tracer distribution and metabolism for 60-min data acquisition with the parameters K1 (mL blood/mL liver tissue/min), rate constant for backflux of tracer from cell-to-blood (k2; /min), rate constant for intracellular phosphorylation of the tracer (k3; /min), rate constant for intracellular dephosphorylation (k4; /min), and a vascular distribution volume (V0; mL blood/mL liver tissue). Because 11C-MG is not metabolized in the liver (14), we used a one-tissue compartmental model of tracer distribution for 60-min data acquisition for this tracer with the parameters K1, k2, and V0 (7). Next, a one-tissue compartmental model with the parameters K1, k2, and V0 was fitted to all three tracer using data from the initial 3-min of the PET recordings. The kinetic parameters were estimated by non-linear fitting of the respective model to the VOI-based liver tissue TAC (Cliver(t)) with the use of Cdual(t) and dual(t).

For the construction of parametric images of hepatic blood perfusion in terms of K1 for 18F-FDG and 11C-MG, noise was reduced by applying an additional Gaussian filter (8 mm FWHM) to the images thereby reducing the central spatial resolution to 10.5 mm FWHM corresponding to a volumetric resolution of around 1 mL. The image of K1 was generated by fitting a linearized one-tissue compartmental model (15) with the parameters K1, k2, and V0 (16) to the initial 3 minutes of the 18F-FDG and 11C-MG PET recordings using dual(t) as input. V0 was fixed to 0.40 mL blood/mL liver tissue (Winterdahl, unpublished observations, similar to values reported by Iozzo et al. (8)).

Statistical analysis

Data are expressed in terms of the mean ± standard error of the mean (SEM). Differences between K1 values estimated using dual(t) or Cdual(t) were tested by a paired two-sample t-test; differences with a P value < 0.05 were considered statistically significant. Correlation between K1 and Q was tested by the Pearson product-moment correlation coefficient, r; P < 0.05 was considered to indicate statistically significant correlation. Mean relative deviations of K1 from Q, i.e. (K1Q)/ Q, were tested by a one-sample t-test; deviations with a P value < 0.05 were considered statistically significant. The mean relative deviation was interpreted as the accuracy of the PET-method and SEM of the deviation as the precision. Differences between variables between animals versus within animals were assessed by ANOVA.

RESULTS

The agreement between K1 estimated using dual(t) or Cdual(t) was good for all combinations of models, tracers and data acquisition periods using the VOI-based liver tissue TACs (Cliver(t)) (paired t-tests, all P > 0.1).

The correlations between K1 for 18F-FDG and Q and between K1 for 11C-MG and Q were not statistically significant when using the 60-min analyses and Cdual(t) (Table 1). The correlation coefficient for 60-min 18F-FDG was similar to a correlation coefficient of 0.57 (P = 0.03) found by Iozzo et al. using 180 min acquisition (8). The use of 3-min acquisition period, a one-tissue compartmental model and Cdual(t) resulted in better correlation for both tracers; this was further improved by the use of dual(t), and the correlation was statistically significant for 18F-FDG using dual(t) (Table 1). The mean relative deviation of K1 from Q for both 18F-FDG and 11C-MG, i.e. (K1Q)/ Q, was positive in all cases but not statistically significantly different from zero except for 11C-MG using 3-min acquisition and Cdual(t) (Table 1). Fig. 1 shows the significant correlations between Q and K1 for 18F-FDG as well as between Q and K1 for 11C-MG (3-min, one-tissue compartmental model, dual(t)). The tendency for K1 to overestimate Q for both tracers (Fig. 1) was not statistically significant in agreement with the accuracy estimates (Table 1).

TABLE 1.

Blood-to-cell Clearance of 18F-FDG and 11C-MG as a Measure of Hepatic Blood Perfusion in Pigs

18F-FDG
(n = 6)
11C-MG
(n = 6)
Correlation
between K1
and Q, r
Relative
deviation
of K1 from Q
Correlation
between K1
and Q, r
Relative
deviation
of K1 from Q
Measured Cdual(t), 60-min; two-tissue compartmental model for 18F-FDG, one-tissue compartmental model for 11C-MG 0.60
(P = 0.14)
0.08 ± 0.09
(P = 0.43)
0.26
(P = 0.61)
0.11 ± 0.10
(P = 0.28)
Measured Cdual(t), 3-min; one-tissue compartmental model for both 18F-FDG and 11C-MG 0.88
(P = 0.02)
0.10 ± 0.04
(P = 0.08)
0.58
(P = 0.23)
0.32 ± 0.08
(P = 0.01)
Model-derived dual(t), 3-min; one-tissue compartmental model for both 18F-FDG and 11C-MG 0.94
(P = 0.01)
0.19 ± 0.11
(P = 0.13)
0.77
(P = 0.07)
0.27 ± 0.12
(P = 0.07)

K1, PET-estimated blood-to-cell clearance of tracer (mL blood/mL liver tissue/min). Q, blood perfusion of liver tissue measured by ultrasound transit-time flow-meters (mL blood/mL liver tissue/min). Cdual(t), measured dual-input TAC (kBq/mL blood). dual(t), estimated dual-input TAC (kBq/mL blood). r, Pearson’s correlation coefficient for correlation between individual pairs of K1 and Q. P < 0.05 was considered to indicate statistically significant correlation. Relative deviation of K1 from Q is presented as mean ± SEM (mean deviations with P < 0.05 were considered statistically significantly different from zero; one-sample t-test).

FIGURE 1.

FIGURE 1

Relationship between VOI-based blood-to-cell clearance, K1, for 18F-FDG (●) and 11C-MG (○) and hepatic blood perfusion, Q, measured simultaneously by ultrasound transit-time flow-meters in six 40-kg pigs. K1 was estimated from dynamic PET of the liver using a model-derived dual-input, dual(t) (see text), 3-min data acquisition period and a one-tissue compartmental model. Measurements from same animal are connected by a straight line.

For 18F-FDGal, there was very little or no correlation between K1 and Q for any of the analyses and we did not proceed with examining the use of 18F-FDGal as a flow tracer.

For both 18F-FDG and 11C-MG, the model fit underestimated the initial rise of Cliver(t) when using the compartmental 60-min models but not when using the 3-min model. In agreement with this, the estimates of K1 for 18F-FDG were on average 5% (range −3% to +23%) higher for the 3-min than for the 60-min data acquisition for both Cdual(t) and dual(t) and for 11C-MG, K1 was on average 20% (range +11% to +29%) higher for 3-min than for 60-min data acquisition, for both Cdual(t) and dual(t).

For the six animals that underwent both 18F-FDG and 11C-MG PET, 2.4% of the variance of the relative deviation of K1 from Q was due to variation within individual animals; 72% of the variance was due to inter-individual variations. This high reproducibility of the K1 estimates allowed us to regard the 18F-FDG and 11C-MG studies as double determinations within the same animal. In three animals normal physiological variations in the hepatic blood perfusion gave Q values that were higher during the 11C-MG PET than during the 18F-FDG PET and opposite in the other three animals. The correlation between changes in Q and changes in K1 was highly significant (r = 0.924; P = 0.008). K1 was accordingly able to detect the direction of the changes in the perfusion within each animal (Fig. 1). These results are noticeable because changes in Q were only 1–36%.

Examples of parametric images of K1 for 18F-FDG and 11C-MG are shown in Fig. 2. As the dual-input applies only for the liver, it should be noted that the estimates of K1 are only valid for the liver, not for other organs or structures seen in the image. The figure illustrates how the two tracers gave very similar images of the liver tissue blood perfusion and that the blood perfusion in the normal pig liver is homogeneous. The coefficient of variation of the mean perfusion estimate was 13% for 18F-FDG and 14% for 11C-MG.

FIGURE 2.

FIGURE 2

Hepatic blood perfusion (mL blood/mL liver tissue/min) in a 40-kg pig estimated as blood-to-cell clearances, K1 of 18F-FDG (upper panel) and 11C-MG (lower panel) from dynamic PET using a model-derived dual-input, dual(t), calculated from the arterial TAC, 3-min data acquisition period and a linearized one-tissue compartmental model. R, right side of the pig; L, left side; A, anterior; P, posterior.

DISCUSSION

The main results of the present pig study are that the measured dual-input TAC could successfully be replaced by the model-derived input TAC, based on a measured arterial TAC, for both 18F-FDG and 11C-MG. The use of 3-min data acquisition gave K1 values which correlated better with the measured blood perfusion than the use of 60-min acquisitions for both tracers.

The physiological basis for using K1 as an estimate of hepatic blood perfusion is founded in a fast blood-to-cell transfer of glucose and 18F-FDG and 11C-MG. In contrast to capillaries in other organs, the liver sinusoids are lined by highly fenestrated endothelial cells. This means that most blood-borne substances come in close contact to the hepatocyte plasma-membrane which has a huge surface area and contains facilitative glucose transporters (17,18) yielding a large permeability-surface area product (PS-product) (18,19). A reason for the poor correlation between K1 for 18F-FDGal and Q in spite of equally high permeability for 18F-FDGal (9) could be that the rate constant for hepatocellular metabolism of 18F-FDGal, k3 (9) is much higher that for the metabolism of 18F-FDG which may influence the estimation of the K1 but the present data do not allow us to draw any conclusions on this.

In all fitting procedures we used a uniform weighting according to Yaqub et al. (20) and this could explain the tendency to underestimation of the initial rise of the fitted liver TAC for 18F-FDG and 11C-MG when using 60-min data acquisition compared to 3-min data acquisition. The accuracy of the K1 estimates is probably more affected by the weighting schemes when analyzing data with long acquisition periods than short acquisition periods because the determination of K1 is particularly sensitive to the initial measurements following tracer injection (7). In agreement with this, for both 18F-FDG and 11C-MG, the use of a one-tissue compartmental model of distribution and 3-min data acquisitions yielded robust estimation of K1.

It is unlikely that the tendency for K1 to overestimate Q for both 18F-FDG and 11C-MG (3-min, dual(t)) (Fig. 1) is due to underestimation of Q by the flow-meter measurements since these are highly accurate and precise (21,22). Another explanation is that intrahepatic blood vessels are unavoidably included in the drawing of the VOI which may influence the VOI-based analysis. Combination of dynamic PET with visualization of intrahepatic vessels by contrast-enhanced CT is likely to minimize this problem but we did not have access to PET/CT at the time of the experiments.

When regarding 18F-FDG and 11C-MG as double determinations in the six pigs that underwent PET with both tracers, we found a highly significant correlation between changes in the hepatic blood perfusion, Q, and changes in K1 (Fig. 1). In each case, the change in blood perfusion between the two PET studies resulted in a corresponding change in the VOI-based K1. Even minor physiological variations in the blood perfusion (1–36%) were thus detectable by the simplified PET method and this may have potential for detection of flow changes in small volumes (1-mL) using parametric images, which could for example be important for early tumor detection.

Compared to the VOI-based analysis, the parametric images tended to give higher estimates of K1. Despite this, the parametric images yielded information about regional variations in the hepatic blood perfusion and is thus a useful supplement to the VOI-based analysis. In general, parametric images are often noisy and biased as a result of large statistical uncertainties of kinetic parameters in individual voxels. In the present study we reduced the noise in the K1 images by applying a filter that reduced the spatial resolution. Future applications may improve the statistical properties using direct parametric image reconstruction techniques that allow for the kinetic model to be applied as part of the image reconstruction including an appropriate weighting scheme (23).

A crucial point for the transfer of this simplified PET method to humans is the estimation of the model-derived dual-input TAC, dual(t), from measured arterial TAC. This includes estimation of the model-derived PV TAC, PV(t), based on a population mean tracer-specific parameter β̅, which reflects the tracer-specific mean splanchnic transit time (12), and a mean HA blood flow fraction of 0.25 (13). In terms of mean HA flow fraction and mean transit time for the PET tracer 15O-carbon monoxide through the splanchnic circulation, human splanchnic circulation (2) is very similar to that in pigs. We have previously shown that the estimation of K1 for 18F-FDG and 11C-MG are insensitive to minor variations in β̅ and the HA blood flow fraction and it is therefore likely that the method quite readily can be transferred to humans even in spite of minor differences in the two key parameters (13). This is important for the potential clinical use of the method in patients in whom the liver structure is partially altered by cirrhosis, cancer, steatosis, or other disease as it indicates that only in cases with profoundly changed model parameters (e.g. end-stage cirrhosis), the use of population means may not be valid, but this needs to be quantified and validated in detail in future prospective clinical studies.

An interesting clinical perspective of the present simplified method is that an 18F-FDG bolus injection can be utilized to combine measurements of hepatic blood perfusion (in terms of K1 from the initial 3-min data) using dual(t) and studies of hepatic glucose metabolism if the PET recording is continued for 45–60 minutes. Combination of parametric images of the hepatic blood perfusion and quasi-steady state metabolism of 18F-FDG using single arterial input (24,25) allows for simultaneous assessment of regional variations in liver blood perfusion and glucose metabolism. This may be clinically useful for early distinction between residual tumor tissue and changes of the parenchymal perfusion and metabolism after local treatment of tumors such as radiofrequency ablation of colorectal liver metastases. For 11C-MG, the short radioactive half-life of 20 min enables repeated measurements for example before and after administration of a drugs with effects on regional hepatic hemodynamics.

Liver tumors are characterized by a predominantly arterial blood perfusion, viz. increased fHA. If nevertheless 18F-FDG PET data are analyzed for the whole liver using the dual-input TAC, the use of this incorrect and less dynamic input to the tumor will give rise to an artificially too high K1 in the tumor as compared with the correctly estimated blood perfusion in the surrounding non-malignant tissue. The tumor-to-background on a parametric image of K1 will be enhanced. This has to be validated in prospective studies once the present method has been transferred to humans.

CONCLUSION

We have developed and validated a simplified method to quantify and image blood perfusion of normal pig livers using 3-min dynamic PET recording after intravenous injection of 11C-MG or 18F-FDG, the latter being a commonly available PET tracer. The method does not require sampling of PV blood or measurements of blood flow in the HA or PV, it only requires measurements of arterial TAC and calculation of the model-derived dual-input. Despite the simplicity of the method, the estimated hepatic blood perfusion was highly correlated with independently measured hepatic blood perfusion. Parametric images were constructed allowing for the assessment of regional variations in the hepatic blood perfusion which was homogeneous in normal pig livers. Similarities between the splanchnic circulation in humans and pigs are promising for the transfer to humans and moreover the 3-min perfusion measurement can be an integrated part of a routine clinical 18FDG PET/CT examination without additional tracer administration and radiation burden to the patient.

ACKNOWLEDGEMENTS

Grant support: The study was supported by the National Institutes of Health (R01-DK074419), the Danish Medical Research Council (09-073658), the Novo Nordic Foundation (A10313), Aase and Ejnar Danielsen’s Foundation (106309), the A. P. Møller Foundation for the Advancement of Medical Science (070118) and Helga and Peter Korning’s Foundation.

List of symbols and their units

18F-FDG

2-[18F]Fluoro-2-deoxy-D-glucose

11C-MG

3-O-[11C]-Methylglucose

18F-FDGal

2-[18F]Fluoro-2-deoxy-D-galactose

HA

Hepatic artery

PV

Portal vein

VOI

Volume-of-interest (mL liver tissue)

TAC

Time-activity curve (kBq/mL tissue or blood versus time)

β

Tracer-specific parameter of the PV model (min)

β̅

Population mean of tracer-specific β (min)

fHA

HA blood flow fraction

fHA

Mean HA blood flow fraction

CA(t)

Arterial TAC (kBq/mL blood)

CPV(t)

Portal vein TAC (kBq/mL blood)

Cdual(t)

Measured dual-input TAC (kBq/mL blood)

dual(t)

Model-derived dual-input TAC (kBq/mL blood)

Cliver(t)

Liver tissue TAC (kBq/mL liver tissue)

K1

Blood-to-cell clearance of tracer (mL blood/mL liver tissue/min)

Q

Blood perfusion of liver tissue (mL blood/mL liver tissue/min)

FPV

Portal venous blood flow (mL blood/min)

FHA

Hepatic arterial blood flow (mL blood/min)

Footnotes

Disclosures: The authors disclose no conflicts of interest.

REFERENCES

  • 1.Skak C, Keiding S. Methodological problems in the use of indocyanine green to estimate hepatic blood flow and ICG clearance in man. Liver. 1987;7:155–162. doi: 10.1111/j.1600-0676.1987.tb00336.x. [DOI] [PubMed] [Google Scholar]
  • 2.Doi R, Inoue K, Kogire M, et al. Simultaneous measurements of hepatic arterial and portal venous flows by transit time ultrasound volume flowmetry. Surg Gynecol Obstet. 1988;167:65–69. [PubMed] [Google Scholar]
  • 3.Rodríguez-Vilarrupla A, Fernández M, Bosch J, García-Pagán JC. Current concepts on the pathophysiology of portal hypertension. Ann Hepatol. 2007;6:28–36. [PubMed] [Google Scholar]
  • 4.Maharaj B, Maharaj RJ, Leary WP, et al. Sampling variability and its influence on the diagnostic yield of percutaneous needle biopsy of the liver. Lancet. 1986;i:523–525. doi: 10.1016/s0140-6736(86)90883-4. [DOI] [PubMed] [Google Scholar]
  • 5.Hølund B, Poulsen H, Schlichting P. Reproducibility of liver biopsy diagnosis in relation to the size of the specimen. Scand J Gastroenterol. 1980;15:329–335. doi: 10.3109/00365528009181479. [DOI] [PubMed] [Google Scholar]
  • 6.Keiding S, Vilstrup H. Intrahepatic heterogeneity of hepatic venous pressure gradient in human cirrhosis. Scand J Gastroenterol. 2002;37:960–964. doi: 10.1080/003655202760230946. [DOI] [PubMed] [Google Scholar]
  • 7.Munk OL, Bass L, Roelsgaard K, Bender D, Hansen SB, Keiding S. Liver kinetics of glucose analogs measured in pigs by PET: importance of dual-input blood sampling. J Nucl Med. 2001;42:795–801. [PubMed] [Google Scholar]
  • 8.Iozzo P, Jarvisalo MJ, Kiss J, et al. Quantification of liver glucose metabolism by positron emission tomography: validation study in pigs. Gastroenterology. 2007;132:531–542. doi: 10.1053/j.gastro.2006.12.040. [DOI] [PubMed] [Google Scholar]
  • 9.Sørensen M, Munk OL, Mortensen FV, et al. Hepatic uptake and metabolism of galactose can be quantified in vivo by 2-[18F]fluoro-2-deoxy-galactose positron emission tomography. Am J Physiol Gastrointest Liver Physiol. 2008;295:G27–G36. doi: 10.1152/ajpgi.00004.2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kudomi N, Slimani L, Järvisalo MJ, et al. Non-invasive estimation of hepatic blood perfusion from H215O PET images using tissue-derived arterial and portal input functions. Eur J Nucl Med Mol Imaging. 2008;35:1899–1911. doi: 10.1007/s00259-008-0796-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kudomi N, Järvisalo MJ, Kiss J, et al. Non-invasive estimation of hepatic glucose uptake from [18F]FDG PET images using tissue-derived input functions. Eur J Nucl Med Mol Imaging. 2009;36:2014–2026. doi: 10.1007/s00259-009-1140-y. [DOI] [PubMed] [Google Scholar]
  • 12.Munk OL, Keiding S, Bass L. Impulse-response function of splanchnic circulation with model-independent constraints: theory and experimental validation. Am J Physiol Gastrointest Liver Physiol. 2003;285:G671–G680. doi: 10.1152/ajpgi.00054.2003. [DOI] [PubMed] [Google Scholar]
  • 13.Winterdahl M, Keiding S, Sørensen M, et al. Tracer input for kinetic modelling of liver physiology determined without sampling portal venous blood in pigs. Eur J Nucl Med Mol Imaging. 2010;38:263–270. doi: 10.1007/s00259-010-1620-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Bender D, Munk OL, Feng H, Keiding S. Metabolites of 18F-FDG and 3-O-11C-methylglucose in pig liver. J Nucl Med. 2001;42:1673–1678. [PubMed] [Google Scholar]
  • 15.Lawson CL, Hanson RJ. Solving least squares problems. Prentice-Hall; 1974. [Google Scholar]
  • 16.Blomqvist G. On the construction of functional maps in positron emission tomography. J Cereb Blood Flow Metab. 1984;4:629–632. doi: 10.1038/jcbfm.1984.89. [DOI] [PubMed] [Google Scholar]
  • 17.Thorens B. Glucose transporters in the regulation of intestinal, renal, and liver glucose fluxes. Am J Physiol. 1996;270:G541–G553. doi: 10.1152/ajpgi.1996.270.4.G541. [DOI] [PubMed] [Google Scholar]
  • 18.Goresky CA, Nadeau BE. Uptake of materials by the intact liver. The exchange of glucose across the cell membrane. J Clin Invest. 1974;53:634–646. doi: 10.1172/JCI107598. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Keiding S, Sørensen M. Hepatic removal kinetics: importance for quantitative measurements of liver function. In: Rodes J, Benhamou J-P, Dufour JF, Blei A, Reichen J, Rizzetto M, editors. Textbook of Hepatology: From Basic Science to Clinical Practice. 3. ed. Oxford: Blackwell; 2007. pp. 468–478. [Google Scholar]
  • 20.Yaqub M, Boellaard R, Kropholler MA, Lammertsma AA. Optimization algorithms and weighting factors for analysis of dynamic PET studies. Phys Med Biol. 2006;51:4217–4232. doi: 10.1088/0031-9155/51/17/007. [DOI] [PubMed] [Google Scholar]
  • 21.Laustsen J, Pedersen EM, Terp K, et al. Validation of a new transit time ultrasound flow-meter in man. Eur J Vasc Endovasc Surg. 1996;12:91–96. doi: 10.1016/s1078-5884(96)80282-6. [DOI] [PubMed] [Google Scholar]
  • 22.Mortensen FV, Rasmussen JS, Viborg O, Laurberg S, Pedersen EM. Validation of a new transit time ultrasound flow-meter for measuring blood flow in colonic mesenteric arteries. Eur J Surg. 1998;164:599–604. doi: 10.1080/110241598750005705. [DOI] [PubMed] [Google Scholar]
  • 23.Wang G, Qi J. Generalized algorithms for direct reconstruction of parametric images from dynamic PET data. IEEE Transactions Medical Imaging. 2009;28:1717–1725. doi: 10.1109/TMI.2009.2021851. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Messa C, Choi Y, Hoh CK, et al. Quantification of glucose utilization in liver metastases: parametric imaging of 18F-FDG uptake with PET. J Comput Assist Tomogr. 1992;16:684–689. doi: 10.1097/00004728-199209000-00003. [DOI] [PubMed] [Google Scholar]
  • 25.Slimani L, Kudomi N, Oikonen V, et al. Quantification of liver perfusion with [15O]H2O-PET and its relationship with glucose metabolism and substrate levels. J Hepatol. 2008;48:974–982. doi: 10.1016/j.jhep.2008.01.029. [DOI] [PubMed] [Google Scholar]

RESOURCES