Abstract
Background
The present study examined a cardiac passive restraint device which applies epicardial pressure (HeartNetTM Implant) in a clinically relevant model of dilated cardiomyopathy (DCM) to determine effects on hemodynamic and myocardial blood flow patterns.
Methods
DCM was established in 10 pigs (3 weeks atrial pacing, 240 beats per minute). Hemodynamic parameters and regional left ventricle (LV) blood flow were measured under baseline conditions and following acute HeartNet (Paracor Medical Inc, Sunnyvale, CA) placement. Measurements were repeated following adenosine infusion, allowing maximal coronary vasodilation and coronary flow reserve determination.
Results
LV dilation and systolic dysfunction occurred relative to baseline as measured by echocardiography. LV end diastolic dimension increased and LV fractional shortening decreased (3.8±0.1 vs 6.1±0.2cm and 31.6±0.5 vs 16.2±2.1%, both p<0.05 respectively) consistent with the DCM phenotype. The HeartNet was successfully deployed without arrhythmias and a computed median mid-LV epicardial pressure of 1.4 mmHg was applied by the HeartNet throughout the cardiac cycle. Acute HeartNet placement did not adversely affect steady state hemodynamics. With the HeartNet in place, coronary reserve was significantly blunted.
Conclusions
In a large animal model of DCM, the cardiac passive restraint device did not appear to adversely affect basal resting myocardial blood flow. However, following acute HeartNet placement, LV maximal coronary reserve was blunted. These unique results suggest that cardiac passive restraint devices which apply epicardial transmural pressure can alter myocardial blood flow patterns in a DCM model. Whether this blunting of coronary reserve holds clinical relevance with chronic passive restraint device placement remains unestablished.
INTRODUCTION
In the past decade the number of patients discharged from the hospital with a diagnosis of congestive heart failure (CHF) has doubled and CHF is now the most common reason for hospital admission in the elderly patient population. As the population ages, the incidence of CHF will increase and continue to be a major health care issue. The progression of CHF is characterized by left ventricular (LV) dilation and subsequent pump dysfunction. Current pharmacological therapies for CHF include modulation of critical neurohormonal pathways which become activated during progression of the disease. These pharmacological treatment strategies may provide symptomatic relief, but the LV remodeling process can continue unabated. Surgical treatment for LV dilation and systolic dysfunction has also been pursued, including the use of cardiac support devices that are postulated to attenuate progressive dilation and stabilize systolic function.(1–7) For example, placement of passive cardiac support devices in patients with dilated cardiomyopathy (DCM), a classical phenotypic presentation of LV dilation and CHF, has been performed with an aim of altering the course of CHF.(1–3, 5, 6) However, a number of issues surrounding the deployment of these devices in the context of DCM remain. Specifically, the acute hemodynamic consequences, and more importantly the direct effect on myocardial blood flow patterns with acute placement of a cardiac support device, remain poorly understood. A particular passive support device which has advanced to clinical trials is the HeartNetTM Implant (Model P140, Paracor Medical, Inc., Sunnyvale, CA), which consists of silicone-insulated nitinol wireforms. The HeartNet Implant provides resistance to circumferential dilation via transmural myocardial pressure.(5–7) In the context of DCM, a shift in the volume to wall thickness ratio occurs, whereby changes in epicardial pressure may directly affect intrinsic myocardial resistance to blood flow. These changes in pressure may be of particular relevance considering the HeartNet Implant may exert additive transmural compressive forces during diastole, the peak of myocardial blood flow delivery. However, whether this theoretical concern can be demonstrated under physiological conditions relevant to DCM remains unknown. Accordingly, the present study utilized a large animal model which recapitulates the phenotype of DCM (8) in order to measure hemodynamics and myocardial blood flow patterns at rest and after maximal coronary artery dilatation in the context of HeartNet Implant placement.
METHODS
The study used a DCM model which has been previously characterized by this laboratory.(8–15) Following the development of DCM, a terminal study was performed at which time baseline systemic hemodynamics and myocardial blood flow were measured with and without the HeartNet Implant in place. Two models of the HeartNet System were used: the Model P140 System, which is currently undergoing study in the PEERLESS-HF pivotal trial in the United States and Canada, and the Model P150 System, which is a next-generation HeartNet System. All animals were treated and cared for in accordance with the National Institutes of Health’s Guide for the Care and Use of Laboratory Animals (National Institutes of Health, 1996).
DCM Induction and Echocardiography
Yorkshire pigs (n=10, 25 kg castrated males) underwent pacemaker implantation for the purposes of establishing pacing-induced DCM. Briefly, a sterile left thoracotomy and a small pericardiotomy were performed exposing the left atrium. A shielded pacemaker lead (CapSure 4965, Medtronic, Minneapolis, MN) was sutured onto the left atrium and connected to a pacemaker (Telectronics Model 1256, St. Jude Medical, Sylmar, CA), which was buried in a subcutaneous pocket. The pericardium was irrigated and approximated, the thoracotomy closed, and the pleural space evacuated of air. On post-operative day 5, baseline echocardiography was performed (Agilent Sonos 5500, 3.5 MHz probe) and the pacemaker was then activated (240 beats per minute). A 1:1 rate of capture was verified via ECG recordings and echocardiography. Cardiac auscultation and ECG recordings were performed frequently during the pacing protocol to ensure continuous capture. At 18–21 days of pacing, the animal was returned to the laboratory for pacemaker deactivation, and a second echocardiograph was performed.
Hemodynamic and Myocardial Blood Flow Measurements
For the terminal studies, all animals were instrumented in identical fashion. Anesthetic induction commenced using isoflurane (3%) and nitrous oxide (33%) delivered via face mask. A 2 μg/kg dose of sufentanyl (ESI-Elkins-Sinn Inc, NJ), a 0.1 mg/kg dose of etomidate (Amidate, ESI, NJ), and a 10 mg dose of vecuronium (Bedford Laboratories, OH) were then administered intravenously, and an endotracheal tube was placed. Anesthesia was maintained with the delivery of inhaled isoflurane (0.5%) and intravenous morphine (ESI) at 3 mg/kg/hr. A bolus of amiodarone (5 mg/kg) (APP Pharmaceuticals LLC, IL) was administered, an intravenous infusion of lidocaine (Hospira Inc., IL) was initiated at 4 mg/kg/hr, and magnesium chloride (1 gram) (Luitpold Pharmaceuticals Inc., NY) was delivered. Following stabilization of the surgical anesthetic plane, an intravenous infusion of vecuronium was initiated at 15 mg/kg/hr.
A multi-lumened thermodilution catheter (7.5F, Baxter Healthcare, Irvine, CA) was positioned in the pulmonary artery via the left external jugular vein. An 8F introducer with side-arm was placed in the right carotid artery for blood pressure measurements and subsequent placement of the LV pressure/infusion catheter. A previously calibrated microtipped transducer catheter with a sideport for infusion and a pigtail catheter (7.5 F, Millar Instruments Inc, Houston, TX) were advanced through the hemostatic sheath and positioned in the LV. The left carotid artery was cannulated with an 8F hemostatic introducer with a side-arm for arterial blood pressure measurements and withdrawal of reference blood samples. A sternotomy was performed and the pericardium was opened and supported by pericardial sutures prior to baseline measurements. Following instrumentation and a 30 min stabilization period the experimental protocol was begun.
Experimental Protocol
Steady state hemodynamics, specifically heart rate, oxygen saturation, arterial blood pressure, pulmonary artery pressure, pulmonary capillary wedge pressure, LV pressure, and cardiac output, as well as myocardial blood flow were recorded with the ventilator temporarily suspended to prevent respiratory artifact. LV peak positive dP/dt was determined from the digitized LV pressure signal. Following completion of the steady-state hemodynamic measurements, a set of fluorescent microspheres was injected for the determination of myocardial blood flow.14–17 Briefly, blue-green (excitation 430, emission 460), yellow (excitation 490, emission 520), red-orange (excitation 530, emission 570), or scarlet (excitation 645, emission 680) fluorescent microspheres (3×106, Molecular Probes, Eugene, OR) were injected into the LV via the pigtail catheter. A reference aortic blood sample was withdrawn from the contra-lateral carotid artery at a rate of 7 mL/min, which was initiated 5 seconds prior to injection and continued for 120 seconds following injection. Next, adenosine (1.5 μmoL/kg/min) was infused through the venous access port and following 15 minutes of infusion, hemodynamics and LV pressures were recorded and fluorescent microspheres delivered. The adenosine infusion allowed for maximal coronary vasodilation and determination of maximal coronary blood flow reserve.(8, 10) Following microsphere delivery, the adenosine infusion was terminated, resulting in a rapid normalization of hemodynamics.
HeartNet Implant Placement and Measurements
Following steady state baseline measurements, the HeartNet Implant was placed circumferentially around the LV and right ventricle (RV). Two versions of the HeartNet Implant were modified for porcine sizing and utilized for the studies described herein; the P140 HeartNet and the P150 HeartNet Implant. Both the P140 and P150 HeartNet Implants consist of a series of circumferentially-oriented silicone-jacketed nitinol wireform rings that provide elastic support to the ventricular wall (approximately 0.5 – 2.5 mmHg throughout the cardiac cycle). The modified Implants consisted of 3 rows and were designed to fit a heart with a maximum circumference measuring approximately 40 cm in circumference. The animals received either P140 Implants (n=5) or P150 Implants (n=5) as determined by a pre-assigned randomization table. A demonstration of the intra-operative P140 Implant placement, assuring proper circumferential deployment, is shown in Figure 1A. Furthermore, following visual confirmation of appropriate deployment, adequate posterior region placement was confirmed using fluoroscopic imaging (Phillips BV 300). A representative fluoroscopic image of the LV with the P140 Implant in place is shown in Figure 1B. The diameters of the cardiac silhouette at the top circumferential ring of the HeartNet device changed from a baseline measurement of 244±4 pixels to 240±4 pixels (ratio of silhouette diameter after HeartNet placement to baseline diameter: 0.99±0.01, p=0.82). Once the HeartNet Implant had been placed, a second 30 min stabilization period was observed, and then steady-state hemodynamic measurements, LV pressure measurements and fluorescent microsphere delivery were performed as described in the preceding section. Following these measurements, an adenosine infusion was again instituted and the hemodynamic and coronary flow measurements were repeated.
Figure 1.
A. Intraoperative placement of the P140 Implant onto a DCM pig heart. B. Fluoroscopic image of the P140 Implant confirming circumferential deployment in the posterior region of the DCM heart.
Tissue harvest
Isoflurane anesthesia was increased to 5% and maintaining a full surgical plane of anesthesia, the heart was harvested. The LV and RV were dissected free, weighed, and subsequently fixed in 10% formaldehyde to facilitate slicing. The myocardium was later dissected into 2 RV regions and 18 LV regions, including epicardial and endocardial regions. Samples of approximately 3 g each were collected from each region and digested in a potassium hydroxide solution as previously described.(8–10, 13, 16, 17) The aortic blood samples were extracted using an identical digest solution. The fluorescence of each extracted sample was subjected to spectrofluorimetry.
Computations and Data Analysis
Echocardiographic indices of LV end diastolic dimension and fractional shortening were determined and compared between baseline pre-pacing values and DCM values using a t-test. From the hemodynamic data, pulmonary and systemic vascular resistances were computed from the thermodilution cardiac output and pressure measurements using standard formulae. The pulmonary artery waveforms recorded during determination of pulmonary capillary wedge pressure were analyzed to determine the amplitude of the a-wave (between the P-wave and the QRS complex on the ECG) and v-wave (coincident with the termination of the T-wave on the ECG) tracing before and after placement of the HeartNet devices. Regional blood flow computations were determined by using the standard formula: Qm=(Ar x Am)/Qr, where Qm is the blood flow in milliliters per minute, Ar is the fluorescence of the aortic reference sample, Am is the fluorescence of the tissue sample, and Qr is the withdrawal rate of the reference sample.20 Final blood flow values were normalized to tissue sample weights and expressed as milliliter per minute per gram. Comparisons of hemodynamics and myocardial blood flow response to adenosine, both with and without the HeartNet Implant placed, were made by repeated measures mixed model ANOVA on each outcome variable, with grouping (baseline or HeartNet Implant) as the between-subjects variable and drug status (steady state vs. adenosine) as the within-subjects variable. Finally, the relative change from absolute baseline steady state values to values following HeartNet Implant placement and adenosine infusion were calculated and log transformed, as they did not conform to a normal distribution. All pairwise comparisons were performed by Tukey adjusted p values. In addition, potential HeartNet Implant/adenosine interactions were examined by ANOVA. All statistical analyses were performed using the SAS statistical software (version 9.2) and statistical significance was considered at α = 0.05 level. All data are presented as mean ± SEM. All authors had access to the data and its interpretation and take full responsibility for its integrity. The funding agency did not influence the interpretation of the data.
RESULTS
Echocardiographic measures confirmed that the DCM phenotype was established by rapid atrial pacing and produced increased LV end diastolic dimension (3.8±0.1cm vs 6.1±0.2cm), decreased LV end diastolic wall thickness (0.8±0.1cm vs 0.6±0.1cm), and decreased LV fractional shortening (31.6±0.5% vs 16.2±2.1%), as compared to baseline values, consistent with prior reports from this laboratory and others.(8–14, 18–20) The amplitudes of the a- and v-waves on the pulmonary capillary wedge pressure (PCWP) traces were 18±3 mmHg and 22±3 mmHg, respectively, suggesting the presence of at least mild and/or moderate mitral regurgitation, consistent with the development of tachycardia-induced DCM.(18–20) The hemodynamic effects observed with adenosine infusion while the HeartNet Implant was placed on the heart are reported in Table 1. Hemodynamics were equivalent at steady state and with adenosine in the pigs randomized to P140 or P150 HeartNet Implant, and therefore, were combined for all analyses for the purposes of statistical robustness. A decrease in systemic blood pressure was observed with adenosine administration, consistent with the systemic vasodilatory effects of adenosine. Following acute placement of either the P140 or the P150 HeartNet Implant, steady state hemodynamics were unchanged from baseline steady state measurements. Following HeartNet Implant placement the amplitudes of the a-and v-waves on the PCWP tracing were similar to baseline values (a-wave: 18±3, p=0.77; v-wave: 21±3 mmHg, p=0.61, respectively), suggesting that there was minimal to no acute effect of HeartNet Implant placement on the degree of mitral regurgitation. Systemic blood pressure was reduced with adenosine infusion. Myocardial blood flow measurements at steady state and after adenosine, both with and without HeartNet Implant placement, are reported as LV endocardial, epicardial, or RV regional values (Table 2). Consistent with prior reports, myocardial blood flow was augmented by adenosine infusion in both chambers and in all regions analyzed.(8–10) Following acute HeartNet Implant placement, steady state myocardial blood flow increased from baseline steady state values and reached statistical significance in the LV base free wall epicardial region. Moreover, following adenosine infusion, myocardial blood flow did not reach similar values as compared to the baseline adenosine state. Average LV and RV myocardial blood flow changes with and without HeartNet Implant are shown in Figure 2. Prior to HeartNet Implant placement, maximal coronary blood flow response, as assessed by adenosine infusion, increased from baseline steady state values. Following acute HeartNet Implant placement, steady state LV myocardial blood flow values qualitatively increased but did not reach statistical significance. With HeartNet placement and adenosine infusion, LV myocardial blood flow did not significantly increase from either baseline steady state or HeartNet steady state. The average RV blood flow values, as shown in Figure 2, showed a similar pattern, but with a notable exception. Specifically, average RV myocardial blood flow increased from baseline steady state with adenosine infusion both prior to and following HeartNet Implant placement.
Table 1.
Hemodynamics in DCM pigs under steady state and following adenosine at baseline and with HeartNet Implant placement.∞
Baseline | HeartNet Implant | |||
---|---|---|---|---|
Steady State | Adenosine | Steady State | Adenosine | |
Heart Rate (bpm) | 101±7 | 94±6 | 108±7 | 105±8 |
LV Pressures | ||||
LV Peak (mmHg) | 107±7 | 95±5 | 118±8 | 103±8 |
LVEDP (mmHg) | 20±2 | 21±3 | 22±2 | 22±3 |
Peak positive dP/dt (mmHg/s) | 1229±113 | 1163±105 | 1461±137 | 1332±108 |
Peak negative dP/dt (mmHg/s) | 1626±120 | 1388±122 | 1540±102 | 1444±124 |
Systemic Pressures | ||||
Systolic (mmHg) | 109±6 | 96±5 | 120±8 | 103±8 |
Diastolic (mmHg) | 69±5 | 51±5* | 80±7 | 59±9* |
Mean (mmHg) | 86±6 | 68±5* | 98±8 | 76±9* |
SVR (dyn·s·cm−5 ) | 2542±273 | 1714±267* | 2767±223 | 1887±373* |
Pulmonary Pressures | ||||
Systolic (mmHg) | 32±3 | 30±3 | 35±2 | 34±3 |
Diastolic (mmHg) | 24±3 | 22±3 | 27±3 | 26±3 |
Mean (mmHg) | 28±3 | 26±3 | 31±3 | 30±3 |
PCWP (mmHg) | 19±3 | 19±3 | 21±3 | 23±3 |
PVR (dyn·s·cm−5 ) | 361±76 | 252±44 | 386±68 | 310±119 |
Cardiac Output (L/min) | 2.5±0.4 | 2.8±0.4 | 2.3±0.3 | 2.7±0.4 |
Values presented as Mean ± SEM.
p<0.05 vs respective steady state
LVEDP=left ventricular end diastolic pressure, SVR=systemic vascular resistance, PCWP=pulmonary capillary wedge pressure, PVR=pulmonary vascular resistance.
Sample size: n=10 (∞Model P140 or P150 HeartNet Implants)
Table 2.
Myocardial blood flow in DCM pigs under steady state and following adenosine at baseline and with HeartNet Implant placement. ∞
Tissue Region | Baseline | HeartNet Implant | |||
---|---|---|---|---|---|
Steady State | Adenosine | Steady State | Adenosine | ||
LV Base | |||||
Free Wall | Endocardium | 1.84±0.25 | 4.14±0.73* | 2.73±0.39 | 3.15±0.38 |
Epicardium | 1.89±0.21 | 3.79±0.51* | 3.02±0.43* | 2.92±0.34 | |
Septal Wall | Endocardium | 1.97±0.32 | 3.71±0.56* | 2.57±0.32 | 3.09±0.52 |
Epicardium | 1.94±0.39 | 3.97±0.67* | 2.64±0.41 | 3.34±0.68 | |
LV Mid | |||||
Free Wall | Endocardium | 2.12±0.26 | 4.54±0.78* | 3.06±0.43 | 3.64±0.53 |
Epicardium | 2.14±0.27 | 4.07±0.60* | 3.19±0.42 | 2.99±0.36 | |
Septal Wall | Endocardium | 2.13±0.30 | 3.84±0.68* | 2.89±0.43 | 3.27±0.64 |
Epicardium | 2.18±0.35 | 4.15±0.81* | 3.01±0.50 | 3.56±0.65 | |
LV Apex | |||||
Free Wall | Endocardium | 2.08±0.26 | 4.00±0.64* | 2.93±0.38 | 3.27±0.41 |
Epicardium | 2.03±0.28 | 3.79±0.64* | 2.86±0.42 | 2.82±0.47 | |
Septal Wall | Endocardium | 2.20±0.31 | 4.12±0.84* | 2.99±0.49 | 3.49±0.68 |
Epicardium | 1.95±0.24 | 3.80±0.67* | 2.85±0.45 | 3.33±0.65 | |
RV | |||||
Free Wall | 1.62±0.24 | 3.80±0.65* | 2.58±0.44 | 3.55±0.89* | |
Outflow Tract | 2.00±0.29 | 3.89±0.68* | 2.93±0.44 | 3.61±0.72* |
Values presented as Mean ± SEM. Blood flow reported as mL/minute/gram of tissue. Sample size: n=10 DCM pigs
p<0.05 vs baseline steady state
LV;left ventricle, RV;right ventricle
Sample size: n=10 (∞Model P140 or P150 HeartNet Implants)
Figure 2.
A. Average of left ventricular (LV) blood flow from all measured regions in DCM pigs (n=10). A significant increase in LV blood flow values was observed under steady state conditions with adenosine administration. The effect of adenosine on LV blood flow is attenuated following acute HeartNet Implant placement. B. Average of right ventricular (RV) blood flow from all measured regions in DCM pigs (n=10). Prior to and following acute HeartNet Implant placement, blood flow values increased from baseline steady state values with adenosine infusion. (+ p<0.05 vs. baseline steady state, SS=steady state, Ad=adenosine)
Through a mixed model ANOVA, a significant interaction between the HeartNet Implant and adenosine infusion effects on LV myocardial blood flow was observed (F=6.21, p=0.019). That is, the preparation with the HeartNet Implant placed showed a significant influence on adenosine mediated myocardial blood flow, as compared to baseline. In order to determine if the effect of acute HeartNet Implant placement on adenosine mediated LV blood flow was global or regional, a regional response analysis was performed (Figure 3). Prior to HeartNet Implant placement, all LV free wall regions demonstrated a consistent increase in blood flow with adenosine infusion. However, following acute HeartNet Implant placement, a selective negative myocardial blood flow response was observed in the LV free wall epicardial regions as compared to the respective endocardial regions. The LV free wall epicardial regions were more affected than the endocardial regions, with respect to HeartNet Implant placement and adenosine infusion (p=0.02).
Figure 3.
A. Average LV blood flow demonstrated by region at baseline under steady state conditions and following adenosine administration. Each region is the average blood flow from n=10 DCM pigs. All LV free wall regions show a consistent increase in myocardial blood flow with adenosine administration.
B. The same LV free wall regions, each the average of n=10 DCM pigs, at steady state and with adenosine administration following HeartNet Implant placement. The three LV epicardial regions show a decrease in blood flow as compared to the LV endocardial regions following HeartNet Implant placement (p=0.02).
The absolute change in LV and RV myocardial blood flow values as a function of steady state baseline values was computed, as shown in Figure 4. LV myocardial blood flow values following adenosine infusion and at steady state following HeartNet Implant placement were increased from steady state baseline. The LV myocardial blood flow following acute HeartNet Implant placement and adenosine infusion appeared elevated from baseline steady state but did not reach statistical significance. RV myocardial blood flow was elevated at all subsequent time points from normalized steady state baseline values.
Figure 4.
A: Change in average LV myocardial blood flow from normalized baseline steady state values. The LV myocardial blood flow values were increased under baseline adenosine and HeartNet Implant steady state conditions, yet failed to reach statistical significance under HeartNet Implant and adenosine condition. B: Change in average RV myocardial blood flow from normalized baseline steady state values. RV myocardial blood flow values were elevated under baseline adenosine, HeartNet Implant steady state, and HeartNet Implant adenosine conditions. Sample size: n=10 DCM pigs. +p<0.05 vs. normalized baseline steady state value.
DISCUSSION
The fundamental structural phenotype of DCM (dilated cardiomyopathy) involves substantial LV chamber dilation and wall thinning, with an alteration in the volume to wall thickness ratio. Because the dilated ventricle is the cornerstone of DCM, attempts have been made to alter this inexorable disease course through physical support and restraint of the heart.(1–7, 21–23) However, whether and to what degree myocardial blood flow patterns are acutely altered by HeartNet Implant placement in the context of DCM remains unknown. Accordingly, in the present study we evaluated two versions of the HeartNet Implant in a clinically relevant model of DCM to examine the effects of acute HeartNet Implant placement on hemodynamics and myocardial blood flow patterns. The pertinent findings of the study were two-fold. First, placement of the HeartNet Implant did not acutely compromise hemodynamics in this well established DCM model. Second, LV myocardial blood flow patterns were unchanged at steady state following HeartNet Implant placement. However, maximal coronary reserve was blunted, in absolute and relative terms, from steady state baseline following HeartNet Implant placement. Therefore, acute placement of the HeartNet Implant in the porcine DCM model had no negative steady state effects on hemodynamics or myocardial blood flow. However, it remains to be established whether and to what extent the acute effects on coronary reserve are operative following longer durations of deployment.
Several cardiac support devices have been developed with the goal of attenuating the progression of DCM.(1–7, 21–23) The CorCap Cardiac Support Device (Acorn Cardiovascular, Inc, St. Paul, MN) has been examined in several experimental models of heart failure and has advanced to clinical trials, demonstrating sustained reduction in LV end diastolic and systolic volumes compared to controls.(1) While the CorCap consists of a polyester-weave device which is placed around both ventricles, other devices have employed a splint approach to cardiac support. Specifically, the Myosplint System (Myocor, Maple Grove, MN) utilizes an implantable transventricular tension member fixed between epicardial pads to disperse epicardial tension.(21) The Myosplint device has fallen out of clinical use due to a lack of therapeutic effect and complications of mitral valve insufficiency.(22) The Paracor HeartNet Implant has been examined in several animal models of LV failure and dysfunction.(6, 23) For example, Magovern et al reported chronic HeartNet implantation decreased post-MI LV dilation in sheep .(23) However, no pre-clinical studies examining myocardial blood flow in a DCM model have been undertaken, making the study described herein unique.
The present study utilized a well characterized rapid pacing model which results in LV dilation and the hemodynamic phenotype of DCM.(8–14) Acute HeartNet Implant placement in this DCM model did not create hemodynamic compromise. Taken together with the finding that the pulmonary capillary wedge pressure waveforms following placement of the HeartNet devices were similar to that recorded prior to device deployment, it is unlikely that acute placement of the device altered the degree of mitral regurgitation and/or had confounding influence on the LV myocardial blood flow measurements. Therefore, the myocardial blood flow measurements were performed under an equivalent hemodynamic state. Following acute HeartNet Implant placement, LV myocardial blood flow was not compromised. In fact, there was a tendency for increased myocardial blood flow with acute HeartNet Implant placement. The reasons for this increase are likely multifactorial, including the surgical manipulation required to deploy the Implant, disruptive electrical-mechanical events, neurohormonal stimulation, and localized sympathetic stimulation from the Implant and the deployment procedure. Also, heart rate, a fundamental determinant of myocardial oxygen demand and myocardial blood flow, trended upward with HeartNet Implant placement. Whether these changes in myocardial blood flow and heart rate are maintained with long-term deployment of the HeartNet devices in the context of DCM remain to be established.
Maximal coronary flow reserve can be defined as the increase in coronary blood flow above basal flow values when the coronary vasculature is maximally dilated.(24) While there are many ways to assess myocardial flow reserve, one common method is through the intravenous infusion of adenosine.(24, 25) In the present study, adenosine infusion robustly augmented myocardial blood flow from baseline steady state values, consistent with coronary vasodilatory effects. However, maximal LV coronary blood flow reserve, as defined by adenosine infusion compared to baseline conditions, was blunted by HeartNet Implant placement. Regional blood flow analysis revealed that the area most affected with respect to adenosine mediated blood flow was the LV free wall epicardium. The reasons for the reduced adenosine mediated coronary flow reserve in the LV free wall epicardial regions are likely multifactorial and could be due to mechanical or neurohormonal factors. One possible contributing factor could have been local sympathetic stimulation related to the placement of the HeartNet Implant, which may have subsequently altered G-protein coupled responses associated with the mechanism of action for adenosine. Additionally, an adenosine hysteresis effect may have occurred, decreasing the coronary response to the second adenosine infusion. Further studies employing a crossover design would be needed to investigate this possibility. The theoretical epicardial compressive force of the HeartNet Implant was approximately 0.5 to 2.4 mmHg as determined by echocardiographic data and intra-operative HeartNet Implant measurements. This level of epicardial pressure may have influenced coronary flow reserve given the elevated LV end diastolic pressures and reduced LV wall thickness in the DCM model utilized. However, this potential mechanism remains speculative. Specifically, a recent report using finite element analysis modeling suggested that an epicardial compressive pressure of greater than 2.5 mmHg may negatively affect global LV systolic function (i.e. cardiac output).(26) However, in the present study, which utilized a physiologically and clinically relevant model and measurements, there was no compromise of LV systolic performance following acute HeartNet Implant placement. This underscores that theoretical effects must be carefully tested in relevant physiologic systems. While multiple interventions, such as adenosine infusion(9)9 or exercise,(13, 14) have been performed while measuring myocardial blood flow in this DCM model, the direct effects of ventricular elastic support have not previously been evaluated. Furthermore, an important caveat is that evaluation of maximal coronary flow reserve with adenosine infusion is more theoretical than practical,(24) therefore whether and to what extent this blunting of coronary reserve holds clinical relevance is unclear.
Consistent with the clinical phenotype, the porcine DCM model causes biventricular failure.(8, 10–14) RV failure is of particular clinical relevance, and due to the geometric and physiologic differences in this chamber, RV blood flow was measured. Steady state RV blood flow was unaffected by placement of the HeartNet Implant, and blood flow increased following adenosine infusion. Thus, RV blood flow patterns were unaffected by acute HeartNet Implant placement. It must be recognized that maximal coronary blood flow occurs at distinctly different points in the cardiac cycle for the LV and RV, with the RV maximal coronary flow during systole and the LV maximal coronary flow during diastole. Therefore, these differential findings with respect to the coronary reserve in the LV and RV suggest that wall stress at end-diastole was maximized with the HeartNet devices at maximal stretch with minimal effects on wall stresses during systole.
The present study was not without limitations. First, the HeartNet Implants were placed acutely, not chronically as in the clinical context. Second, while the rapid pacing induced DCM model used in the present study recapitulates a severe clinical phenotype, rapid pacing is an unlikely cause for DCM in patients and therefore extrapolation to the clinical context must be done with caution. Third, the present study evaluated two different HeartNet Implants (Models P140 and P150), by which the epicardial compressive forces were estimated but not directly computed using finite element analysis. Nevertheless, the unique findings from the present study demonstrated that in a porcine DCM model, the HeartNet Implant can be deployed without hemodynamic compromise. While the HeartNet Implant did not affect myocardial blood flow under steady state conditions, a region dependent decrease in LV free wall epicardial blood flow with respect to adenosine infusion was observed with acute HeartNet Implant placement. Future studies will need to address whether and to what degree this effect is present with chronic placement of the HeartNet Implant. Furthermore, the effects of HeartNet Implant placement on LV shape, size, and myocardial structure as well as load independent indices of LV function over a longer study period remain to be established.
Acknowledgments
This research was supported by a grant from Paracor Medical, Inc, as well as by NIH grants HL81692 and HL87134, and a Merit Award from the Veterans’ Affairs Health Administration.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Starling RC, Jessup M, Oh JK, Sabbah HN, Acker MA, Mann DL, et al. Sustained benefits of the CorCap Cardiac Support Device on left ventricular remodeling: three year follow-up results from the Acorn clinical trial. Ann Thorac Surg. 2007 Oct;84(4):1236–42. doi: 10.1016/j.athoracsur.2007.03.096. [DOI] [PubMed] [Google Scholar]
- 2.Mann DL, Acker MA, Jessup M, Sabbah HN, Starling RC, Kubo SH. Clinical evaluation of the CorCap Cardiac Support Device in patients with dilated cardiomyopathy. Ann Thorac Surg. 2007 Oct;84(4):1226–35. doi: 10.1016/j.athoracsur.2007.03.095. [DOI] [PubMed] [Google Scholar]
- 3.Acker MA, Bolling S, Shemin R, Kirklin J, Oh JK, Mann DL, et al. Mitral valve surgery in heart failure: insights from the Acorn Clinical Trial. J Thorac Cardiovasc Surg. 2006 Sep;132(3):568–77. 77, e1–4. doi: 10.1016/j.jtcvs.2006.02.062. [DOI] [PubMed] [Google Scholar]
- 4.Blom AS, Mukherjee R, Pilla JJ, Lowry AS, Yarbrough WM, Mingoia JT, et al. Cardiac support device modifies left ventricular geometry and myocardial structure after myocardial infarction. Circulation. 2005 Aug 30;112(9):1274–83. doi: 10.1161/CIRCULATIONAHA.104.499202. [DOI] [PubMed] [Google Scholar]
- 5.Klodell CT, Jr, Aranda JM, Jr, McGiffin DC, Rayburn BK, Sun B, Abraham WT, et al. Worldwide surgical experience with the Paracor HeartNet cardiac restraint device. J Thorac Cardiovasc Surg. 2008 Jan;135(1):188–95. doi: 10.1016/j.jtcvs.2007.09.034. [DOI] [PubMed] [Google Scholar]
- 6.Magovern JA. Experimental and clinical studies with the Paracor cardiac restraint device. Semin Thorac Cardiovasc Surg. 2005 Winter;17(4):364–8. doi: 10.1053/j.semtcvs.2005.09.006. [DOI] [PubMed] [Google Scholar]
- 7.George I, Cheng Y, Yi GH, He KL, Li X, Oz MC, et al. Effect of passive cardiac containment on ventricular synchrony and cardiac function in awake dogs. Eur J Cardiothorac Surg. 2007 Jan;31(1):55–64. doi: 10.1016/j.ejcts.2006.09.024. [DOI] [PubMed] [Google Scholar]
- 8.Spinale FG, Hendrick DA, Crawford FA, Smith AC, Hamada Y, Carabello BA. Chronic supraventricular tachycardia causes ventricular dysfunction and subendocardial injury in swine. Am J Physiol. 1990 Jul;259(1 Pt 2):H218–29. doi: 10.1152/ajpheart.1990.259.1.H218. [DOI] [PubMed] [Google Scholar]
- 9.Spinale FG, Zellner JL, Tomita M, Tempel GE, Crawford FA, Zile MR. Tachycardia-induced cardiomyopathy: effects on blood flow and capillary structure. Am J Physiol. 1991 Jul;261(1 Pt 2):H140–8. doi: 10.1152/ajpheart.1991.261.1.H140. [DOI] [PubMed] [Google Scholar]
- 10.Spinale FG, Tanaka R, Crawford FA, Zile MR. Changes in myocardial blood flow during development of and recovery from tachycardia-induced cardiomyopathy. Circulation. 1992 Feb;85(2):717–29. doi: 10.1161/01.cir.85.2.717. [DOI] [PubMed] [Google Scholar]
- 11.Kribbs SB, Merritt WM, Clair MJ, Krombach RS, Houck WV, Dodd MG, et al. Amlodipine monotherapy, angiotensin-converting enzyme inhibition, and combination therapy with pacing-induced heart failure. Hypertension. 1998 Mar;31(3):755–65. doi: 10.1161/01.hyp.31.3.755. [DOI] [PubMed] [Google Scholar]
- 12.Krombach RS, Clair MJ, Hendrick JW, Houck WV, Zellner JL, Kribbs SB, et al. Angiotensin converting enzyme inhibition, AT1 receptor inhibition, and combination therapy with pacing induced heart failure: effects on left ventricular performance and regional blood flow patterns. Cardiovasc Res. 1998 Jun;38(3):631–45. doi: 10.1016/s0008-6363(98)00050-9. [DOI] [PubMed] [Google Scholar]
- 13.Clair MJ, Krombach RS, Hendrick JW, Houck WV, Hebbar L, Kribbs SB, et al. AT1 angiotensin II receptor inhibition in pacing-induced heart failure: effects on left ventricular performance and regional blood flow patterns. J Card Fail. 1998 Dec;4(4):311–23. doi: 10.1016/s1071-9164(98)90237-8. [DOI] [PubMed] [Google Scholar]
- 14.Caparas SN, Clair MJ, Krombach RS, Hendrick JW, Houck WV, Kribbs SB, et al. Brain blood flow patterns after the development of congestive heart failure: effects of treadmill exercise. Crit Care Med. 2000 Jan;28(1):209–14. doi: 10.1097/00003246-200001000-00034. [DOI] [PubMed] [Google Scholar]
- 15.Lucas DG, Jr, Patterson T, Hendrick JW, Holder J, Sample JA, Joffs C, et al. Effects of adenosine receptor subtype A1 on ventricular and renal function. J Cardiovasc Pharmacol. 2001 Oct;38(4):618–24. doi: 10.1097/00005344-200110000-00014. [DOI] [PubMed] [Google Scholar]
- 16.Glenny RW, Bernard S, Brinkley M. Validation of fluorescent-labeled microspheres for measurement of regional organ perfusion. J Appl Physiol. 1993 May;74(5):2585–97. doi: 10.1152/jappl.1993.74.5.2585. [DOI] [PubMed] [Google Scholar]
- 17.Prinzen FW, Glenny RW. Developments in non-radioactive microsphere techniques for blood flow measurement. Cardiovasc Res. 1994 Oct;28(10):1467–75. doi: 10.1093/cvr/28.10.1467. [DOI] [PubMed] [Google Scholar]
- 18.Takagaki M, McCarthy PM, Goormastic M, Ochiai Y, Doi K, Kopcak MW, Jr, et al. Determinants of the development of mitral regurgitation in pacing-induced heart failure. Circ J. 2003 Jan;67(1):78–82. doi: 10.1253/circj.67.78. [DOI] [PubMed] [Google Scholar]
- 19.Akiyama M, Popovic ZB, Kamohara K, Cingoz F, Daimon M, Ootaki C, et al. Acute reduction of functional mitral regurgitation in canine model using an epicardial device. Ann Thorac Surg. 2008 May;85(5):1771–5. doi: 10.1016/j.athoracsur.2007.11.036. [DOI] [PubMed] [Google Scholar]
- 20.Maniu CV, Patel JB, Reuter DG, Meyer DM, Edwards WD, Rihal CS, et al. Acute and chronic reduction of functional mitral regurgitation in experimental heart failure by percutaneous mitral annuloplasty. J Am Coll Cardiol. 2004 Oct 19;44(8):1652–61. doi: 10.1016/j.jacc.2004.03.085. [DOI] [PubMed] [Google Scholar]
- 21.Schenk S, Reichenspurner H, Groezner JG, Boehm DH, Schirmer J, Scheidt WV, et al. Myosplint implantation and ventricular shape change in patients with dilative cardiomyopathy- first clinical experience. J Heart Lung Transplant. 2001 Feb;20(2):217. doi: 10.1016/s1053-2498(00)00474-5. [DOI] [PubMed] [Google Scholar]
- 22.Christiansen S, Klocke A, Autschbach R. Past, present, and future of long-term mechanical cardiac support in adults. J Card Surg. 2008 Nov-Dec;23(6):664–76. doi: 10.1111/j.1540-8191.2008.00696.x. [DOI] [PubMed] [Google Scholar]
- 23.Magovern JA, Teekell-Taylor L, Mankad S, Dasika U, McGregor W, Biederman RW, et al. Effect of a flexible ventricular restraint device on cardiac remodeling after acute myocardial infarction. ASAIO J. 2006 Mar–Apr;52(2):196–200. doi: 10.1097/01.mat.0000199751.51424.78. [DOI] [PubMed] [Google Scholar]