Abstract
Molecular detection of microorganisms requires microbial cell disruption to release nucleic acids. Sensitive detection of thick-walled microorganisms such as Bacillus spores and Mycobacterium cells typically necessitates mechanical disruption through bead beating or sonication, using benchtop instruments that require line power. Miniaturized, low-power, battery-operated devices are needed to facilitate mechanical pathogen disruption for nucleic acid testing at the point of care and in field settings. We assessed the lysis efficiency of a very small disposable bead blender called OmniLyse relative to the industry standard benchtop Biospec Mini-BeadBeater. The OmniLyse weighs approximately 3 g, at a size of approximately 1.1 cm3 without the battery pack. Both instruments were used to mechanically lyse Bacillus subtilis spores and Mycobacterium bovis BCG cells. The relative lysis efficiency was assessed through real-time PCR. Cycle threshold (CT) values obtained at all microbial cell concentrations were similar between the two devices, indicating that the lysis efficiencies of the OmniLyse and the BioSpec Mini-BeadBeater were comparable. As an internal control, genomic DNA from a different organism was spiked at a constant concentration into each sample upstream of lysis. The CT values for PCR amplification of lysed samples using primers specific to this internal control were comparable between the two devices, indicating negligible PCR inhibition or other secondary effects. Overall, the OmniLyse device was found to effectively lyse tough-walled organisms in a very small, disposable, battery-operated format, which is expected to facilitate sensitive point-of-care nucleic acid testing.
INTRODUCTION
Nucleic acid testing has become an important tool in infectious disease diagnosis (4, 25), biothreat detection (14, 30), and research. Point-of-care or point-of-use applications of nucleic acid testing, especially in settings with minimal infrastructure, require novel tools that can perform essential tasks in miniaturized, inexpensive formats with the same performance characteristics as currently available, expensive, laboratory-based methods (13).
Lysis of an organism to liberate its genomic material is an important step in sample preparation for nucleic acid testing. Many common pathogens can be lysed through chemical agents, such as detergents and chaotropic salts, or by enzymatic treatment (8, 31). However, lysis is a significant challenge for thick-walled microorganisms such as Bacillus anthracis spores and Mycobacterium tuberculosis cells (13, 18, 22). The multilayer structure of Bacillus spores includes an outer cortex and coat that is resistant to chemical and physical treatments (5, 23). Similarly, mycobacteria have a thick, waxy cell wall that is difficult to disrupt for the extraction of nucleic acids (9, 17). High-energy mechanical disruption methods, such as sonication and bead beating, are commonly used for lysis of thick-walled organisms, since chemical, heat, freeze-thaw, or enzymatic lysis methods alone are less effective (1, 11, 22). Lysis protocols for mycobacteria have been reported that use low-energy bead beating (2, 6) in conjunction with heat or chemical or enzymatic lytic agents, which increase process complexity and potentially introduce PCR inhibitors. We are not aware of any published or unpublished methods that can break open slow-growing mycobacteria by low-energy bead beating alone, in the absence of other lytic treatments, with the same high efficiency as the BioSpec Mini-BeadBeater.
Disruption of thick-walled organisms by sonication typically involves the exposure of a suspension containing the pathogen and beads to high frequency sound waves that are delivered by a rapidly oscillating transducer. Lysis by sonication has been attributed to cavitation, where the rapid formation and shrinkage of gas bubbles creates high pressures and temperatures (5). Lysis of thick-walled organisms by bead beating typically involves high-frequency oscillation of a closed tube containing a suspension of the target organism and beads. The mechanism of lysis by bead-beating has been attributed to high shear rates between beads and strong periodic vortical flow fields (13). The diameter of beads used during mechanical lysis is critical to lysis efficiency, with 100-μm-diameter beads being more effective than larger-diameter beads at lysing Gram-positive bacteria (11, 22).
Bead beating and sonication typically require benchtop devices with significant power demands. The BioSpec Mini-BeadBeater (Fig. 1 A) and the Sonics VibraCell Ultrasonic system are among the smallest devices on the market, at respective sizes of 3,900 cm3 for the BioSpec Mini-BeadBeater and >7,400 cm3 for the VibraCell Ultrasonic system, including the power supply (13). Larger, heavier, and more expensive bead-beating devices exist, which can process multiple samples in parallel. The BioSpec Mini-BeadBeater device has been used in previous studies to lyse bacterial spores (13, 20). Mycobacterium cells are lysed effectively using the BioSpec Mini-BeadBeater (27) and Mini-BeadBeater-8 (12, 15). The BioSpec Mini-BeadBeater has been used as a standard to benchmark relative lysis efficiencies of new devices and techniques (13).
Fig. 1.
Systems for mechanical disruption of tough-walled organisms. (A) Mini-BeadBeater (BioSpec), a typical benchtop instrument. (B) OmniLyse bead blender (Claremont BioSolutions), a miniaturized, disposable, battery-operated device.
Fully integrated sample-to-answer nucleic acid testing systems are on the market, such as the Cepheid GeneXpert, which automates sample preparation through sonication, filtration, and solid-phase extraction, followed by amplification and detection via the real-time PCR (34). Mechanical cell disruption is facilitated through a miniaturized sonication device (5). GeneXpert cartridges are available to detect B. anthracis (34), M. tuberculosis (7, 16), and Clostridium difficile (26), three pathogens that are known to be lysis resistant. In the GeneXpert, sensitive detection of these tough-walled pathogens is facilitated through mechanical cell disruption via sonication.
Alternative methods for rapid mechanical lysis of thick-walled organisms have been developed at Keck Graduate Institute and are now commercialized by Claremont BioSolutions (13). One method uses a disposable, battery-operated/low-power, light weight, miniaturized, bead blender called the OmniLyse. This device (Fig. 1B) uses four AAA 1.5-V batteries, weighs approximately 3 g, and is 1.1 cm3 in volume without a battery pack or 71 g and 44 cm3 in volume with a battery pack. The OmniLyse consists of an injection-molded chamber with inlet and outlet ports that can be operated in batch or flowthrough mode and is therefore suitable for processing large sample volumes. A micro-motor equipped with a custom-designed precision-cut impellor is installed in such a way as to dissipate a high amount of mechanical energy within the chamber. The device utilizes a proprietary method for entrapping and retaining zirconia/silica beads within the lysis chamber during activation of the motor. High-speed camera video experiments have shown that the motor can drive the impellor at speeds greater than 30,000 rpm with the chamber filled with liquid and beads. The kinetic energy imparted onto the beads in solution generates high shear forces between beads, causing disruption of cells caught within the shear flow. The system is a single-use disposable unit, rendered feasible by use of inexpensive components.
We herein report a performance evaluation, assessing the lysis efficiency of the OmniLyse device relative to the industry standard benchtop Biospec Mini-BeadBeater (Fig. 1), using B. subtilis spores as a BSL1 surrogate for B. anthracis spores, and M. bovis BCG, an attenuated vaccine strain, as a BSL2 surrogate for M. tuberculosis.
MATERIALS AND METHODS
General reagents.
Standard Taq buffer (B9014S), Taq DNA polymerase (M0273L), and deoxynucleotide triphosphate (dNTP; N0447S) were purchased from New England Biolabs. Magnesium chloride (ES25-100ml) was obtained from Amresco (Solon, OH). Uracil DNA glycosylase (catalog no. 78310) and PCR nucleotide mix with dUTP (catalog no. 77330) were acquired from USB (Cleveland, OH). Oligonucleotide primers were ordered from Integrated DNA Technologies (Coralville, IA) and Sigma-Aldrich (St. Louis, MO). SYBR green I (S7567) was acquired from Invitrogen (Carlsbad, CA). Power SYBR mix (catalog no. 4368702) was purchased from Applied Biosystems. Bovine serum albumin (BSA) was obtained from New England Biolabs (B9001S; Ipswich, MA) and Thermo Scientific (HyClone SH30574.03; Waltham, MA). TE buffer was composed of 10 mM Tris and 1 mM EDTA (pH 8.3). B. subtilis spores, strain GB03, were donated by Gustafson, LLC, now a subsidiary of Bayer Corp. Aeromonas hydrophila (strain 7966) was obtained from the American Type Culture Collection (Manassas, VA). The Kinyoun TB Stain Kit K (catalog no. 212315) was ordered from BD (Franklin Lakes, NJ). Gel electrophoresis to size double-stranded DNA following PCR was performed on the Agilent 2100 Bioanalyzer (Santa Clara, Ca) by using an Agilent DNA 1000 kit (catalog no. 5067-1504). The Mini-BeadBeater and the Mini-BeadBeater-8 were purchased from BioSpec (Bartlesville, OK).
OmniLyse device.
Studies performed herein utilized the current version of Claremont BioSolutions (Upland, CA) OmniLyse device, as shown in Fig. 1B, with an inlet luer-lock fitting, a stiff outlet tube, and a lead-free micro-motor that in the current device configuration can operate at full speed for up to 20 min. The lysis chamber contains 200 mg of 100-μm-diameter acid-washed zirconia/silica beads. The OmniLyse device was operated using four AAA alkaline batteries at a nominal voltage of 6 V with a low electrical current between 100 and 200 mA. Since the typical capacity of an alkaline AAA battery is approximately 1,200 mAh, a single battery pack should operate between 6 and 12 h, which translates to an expected 250 to 500 lysis preparations of 1.5 min each per battery pack. This disposable lysis device is fabricated using inexpensive manufacturing methods. The main body and fittings of the device are produced through inexpensive injection molding. Also, the motors are mass produced in extremely high quantities and obtained at low cost. Even in low production volumes, Claremont BioSolutions is currently able to sell the devices within the price range of other disposables used in bioscience research. As volumes increase, both the cost of goods and the unit price are anticipated to drop precipitously.
Preparation of B. subtilis spore stocks.
B. subtilis spores were suspended and washed three times in Nanopure filtered water. After the final wash and resuspension, the spores were quantified by using an iNcyto (Chungchongnam, Korea) C-chip hemocytometer (DHC-N01-2) and through colony counting using LB agar plates. A Nikon Eclipse TE2000-S microscope was used to view microbial samples. Both methods resulted in similar spore counts. The Schaeffer-Fulton staining method (32) was used to confirm that the B. subtilis microorganisms were present as spores and not as vegetative cells. The spore stocks were stored at −20°C until further use.
Lysis of B. subtilis spores.
Dilution series of B. subtilis spores ranging from 104 to 108 spores/ml in the first experiment and from 103 to 108 spores/ml in the second experiment were prepared in TE buffer, to which 106 copies of purified Escherichia coli K-12 genomic DNA/ml was added as an internal control. (E. coli K-12 was grown in nutrient broth at 37°C, and the DNA was purified and extracted by using a Qiagen [Venlo, Netherlands] DNeasy Blood & Tissue Kit [catalog no. 69504].) Each sample in this dilution series was processed in triplicate by using the OmniLyse device and the BioSpec Mini-BeadBeater. The same 100-μm-diameter zirconia/silica beads were used in both devices. For lysis using the OmniLyse device, 500-μl sample aliquots were aspirated from the sample tube through the stiff tubing into the device and into a syringe attached at the port with the luer lock and then dispensed back through the device into the sample tube for a total of three passes, with a total processing time per sample of 1.5 min. Each replicate was performed with a new device. Crude lysates were kept on ice until samples were tested using PCR. For lysis using the BioSpec Mini-BeadBeater, 625-μl sample aliquots plus 250 mg of 100-μm-diameter zirconia/silica beads were added to 2-ml conical tubes. The samples were bead beaten at maximum speed (4,800 rpm) for 1-min pulses, three times, with a 30-s rest on ice between pulses. When finished, the samples were immediately transferred to fresh tubes and kept on ice until testing by PCR.
PCR analysis of B. subtilis samples.
After processing, each sample was split and interrogated with two primer sets, one specific for B. subtilis and the other specific for E. coli DNA (Table 1). Then, 12.5 μl of sample was added to 37.5 μl of master mix. The thermocycling conditions consisted of a 10-min 94°C hot start, followed by 45 cycles of 94°C for 8 s, 58.7°C for 15 s, and 72°C for 28 s. For the second experiment, a 37°C 10-min UNG (uracil N-glycosylase) step was added before the 10-min 94°C hot start step. Each PCR mixture contained 1× standard Taq buffer, 0.25 mM concentrations of dNTPs, 0.08 U of Taq Polymerase/μl, 4.5 mM MgCl2, 0.3 μM concentrations of the forward and reverse (F/R) primers, 0.8 mg of BSA/ml, and 1× SYBR green I. For the second test dUTPs replaced the dTTPs (at the same final concentration) and 0.008 U of UNG/μl was added to the master mix. PCR was performed using the DNA Engine Opticon 2 (Bio-Rad, Hercules, CA). For the lysis control experiments, each PCR sample contained 0.3 μM concentrations of the F/R primers, 0.8 mg of BSA/ml, and 1× Power SYBR green PCR mix.
Table 1.
Sequences of forward and reverse primers used during PCR
| Target organism | Primer sequence (5′–3′) |
Product size (bp) | Reference | |
|---|---|---|---|---|
| Forward | Reverse | |||
| B. subtilis | GTTTTGTTCTTTTCCTGTGCC | GCTTCCAGCTTACTGATATCC | 128 | 28 |
| E. coli | ACGCTGCCCGATATAACAAC | GCAATGGCGTAAAAATTGGT | 236 | 21 |
| A. hydrophila | ATTGAGCCGCCTTAACAGG | AACTGTTATCCCCCTCGAC | 189 | 10 |
| M. bovis BCG | CCTGCGAGCGTAGGCGTC | CTCGTCCAGCGCCGCTTC | 123 | 3a |
The oligonucleotide primer sequence was modified from reference 3.
Cultivation of M. bovis BCG.
Mycobacterium bovis BCG strain Russia was grown in Difco Middlebrook 7H9 broth medium supplemented with 10% albumin-dextrose-catalase enrichment (BBL), 0.2% glycerol, and 0.05% Tween 80 at 37°C. Cultures were maintained in log-phase growth (optical density at 600 nm [OD600] = 0.2 to 1.0, where 1 OD600 = 4.5E8 cells/ml) before use in experiments. The cells were pelleted by centrifugation at 16,000 × g for 1 min or 3,000 × g for 30 min and washed once with 1× phosphate-buffered saline (equivalent culture volume) and resuspended in TE to the desired concentration. A Kinyoun TB Stain Kit K was used to confirm the presence of Mycobacterium cells.
Lysis of M. bovis BCG cells.
A dilution series of M. bovis BCG, in TE buffer containing 106 copies of purified Aeromonas hydrophila genomic DNA/ml, was prepared ranging from 104 to 108 CFU/ml in the first experiment and from 103 to 108 CFU/ml in the second experiment. (A. hydrophila was grown on nutrient agar or in nutrient broth at 28 to 30°C, and the DNA was purified and extracted by using a MoBio Laboratories [Carlsbad, CA] UltraClean microbial DNA isolation kit [10].) Then, 4 mg of BSA/ml was added upstream to the TE buffer in the second experiment. Triplicates of samples in the dilution series were processed using either the OmniLyse device or the BioSpec Mini-BeadBeater-8. For lysis using the OmniLyse device, portions (1 ml in the first experiment and 500 μl in the second experiment) of the samples were aspirated from the sample tube through the stiff tubing into the device and into a syringe attached at the port with the luer lock and then dispensed back through the device into the sample tube for a total of three passes, with a total processing time per sample of 1.5 min. The experiment was performed in triplicates, using a new device for each replicate. For lysis using the BioSpec Mini-BeadBeater-8 device, 1.2 ml or 625 μl in the first and second experiments, respectively, plus 250 mg of beads, was added to 2-ml conical tubes. For the BioSpec Mini-BeadBeater-8, Sigma-Aldrich 106-μm-diameter glass beads (G-4649) were used in the first experiment, and the same 100-μm-diameter zirconia/silica beads used in the OmniLyse blender were used in the second experiment. The samples were bead beaten at maximum speed (2,800 rpm) for 1-min pulses, three times, with a 30-s rest on ice between pulses (10, 33). Crude lysates were stored at −80°C after processing until tested using PCR. A control experiment was also performed in which processed samples were split and either frozen at −80°C or kept on ice until tested using PCR.
PCR analysis of M. bovis BCG samples.
After processing, each sample was split and interrogated with two primer sets, one specific for M. bovis BCG and the other specific for A. hydrophila DNA (Table 1). For PCR analyses, 5 μl of sample was added to 20 μl of master mix. Each PCR sample contained the F/R primers 0.3 μM and 1× Power SYBR green PCR mix. In experiment 1, each PCR sample further contained 0.8 mg of BSA/ml in the master mix. The thermocycling conditions consisted of a 10-min 95°C hot start, followed by 45 cycles of 95°C for 15 s, 59°C for 30 s, and 72°C for 30 s. A dissociation curve between 95 and 65°C was acquired to confirm the identity of the amplicons. PCR was performed using an Applied Biosystems 7500 Real-Time instrument (Foster City, CA).
RESULTS
We evaluated the capability of the OmniLyse device to mechanically disrupt thick-walled organisms and liberate DNA suitable for PCR amplification, relative to the BioSpec Mini-BeadBeater and Mini BeadBeater-8, two industry-standard benchtop mechanical lysis devices. Figure 2 provides an overview of the experiments performed. To the extent possible, we used similar conditions in the OmniLyse and the BioSpec devices. Unless otherwise stated, all data points represent the averaged cycle threshold (CT) values of triplicate runs, with the error bars representing the standard deviation. In all experiments, we included triplicates of a control sample that only contained a fixed concentration of the internal control DNA but none of the targeted microorganism DNA. Since the x axes in Fig. 3 to 5 represent the concentrations of the targeted microorganisms, this control sample was added as an additional point at 0 CFU/ml.
Fig. 2.
Procedure overview. (A and B) Serial dilutions of the targeted microorganism (TMO) plus a fixed amount of internal control (IC) genomic DNA were processed in triplicate using an OmniLyse device for 1.5 min or a BioSpec Mini-BeadBeater/Mini-BeadBeater-8 in three 1-min pulses, with 30-s rests on ice between pulses. (C) Processed samples were split and analyzed by PCR using primers either specific for the TMO or the IC.
Fig. 3.
Lysis of B. subtilis spores. PCR CT values are shown for solutions containing spores at different concentrations, processed using either the OmniLyse device (⧫) or the BioSpec Mini-BeadBeater (□), with primers specific to B. subtilis. The numbers for CFU per reaction (CFU/rxn) indicated in the graph are theoretical maxima, based on 12.5 μl of lysed sample added to each PCR, and assume no loss during sample processing. In both experiments A and B, the CT values of spores lysed obtained using either device were comparable, indicating similar lysis efficiency. During experiment A, the no-target control (0 CFU/ml) samples appeared earlier than expected. Gel electrophoresis results (data not shown) suggested that this was predominantly due to primer dimers.
Fig. 5.
Assessment of secondary effects on PCR amplification of internal control DNA. (A) CT values for amplification of E. coli genomic DNA spiked as an internal control at a constant concentration in B. subtilis suspensions of various concentrations, processed via the OmniLyse (⧫) or BioSpec (□) device. Based on a 12.5-μl sample added per PCR and assuming no loss during processing, each PCR should contain 12,500 copies E. coli genomic DNA. (B) CT values for amplification of A. hydrophila genomic DNA spiked as an internal control at a constant concentration into M. bovis BCG suspensions of various concentrations, processed via the OmniLyse (⧫) or BioSpec (□) device. Based on a 5-μl sample added per PCR, and assuming no loss during processing, each PCR should contain 5,000 copies A. hydrophila genomic DNA. Overall, the CT values of the internal controls are reasonably comparable for the OmniLyse device and the BioSpec Mini-BeadBeater devices, and are independent of the concentration of the targeted microorganism.
Lysis efficiency.
The targeted microorganisms were B. subtilis spores or M. bovis BCG cells. The CT values obtained from PCRs with primers specific for each organism were used as indicators of lysis efficiency. If both devices have similar lysis efficiencies, then similar CT values should be obtained at each concentration of the targeted microorganism. In general, B. subtilis spores processed through either the OmniLyse device or the BioSpec Mini-BeadBeater, quantified by using B. subtilis-specific primers, yielded comparable CT values at all concentrations tested in two independent experiments (Fig. 3). Likewise, M. bovis BCG cells processed through either the OmniLyse device or the BioSpec Mini-BeadBeater, using M. bovis BCG-specific primers, resulted in comparable CT values at all concentrations tested in two independent experiments (Fig. 4). Since the purpose of the present study is to determine a difference in CT value as function of device, independent of the targeted microorganism's concentration, the sample size per device includes all replicates for all concentrations and both experiments. This experimental design has a statistical power of >90% for predicting a significant ΔCT of 0.5 between the two devices, with a standard deviation in CT of 0.5 (signal-to-noise ratio [S/N] = 1). Based on a two-factor analysis of variance (ANOVA, see Fig. S1 in the supplemental material), there is no significant difference between the CT values (as indicators of lysis efficiency) obtained using the BioSpec versus those obtained using the OmniLyse device, for either B. subtilis spore lysis (P = 0.3225) or lysis of M. bovis BCG (P = 0.7845). This ANOVA treated the targeted microorganism concentration as factor 1 and lysis device as factor 2, using a blocked experimental design, wherein experimental sets (Fig. 3 and 4) are treated as blocks.
Fig. 4.
Lysis of M. bovis BCG cells. CT values are shown for solutions containing M. bovis BCG at different concentrations, processed using either the OmniLyse device (⧫) or the BioSpec Mini-BeadBeater (□), with primers specific to M. bovis BCG. The numbers for CFU/rxn indicated in the graph are theoretical maxima, based on 5 μl of lysed sample added to each PCR, and assume no loss during sample processing. Since mycobacteria are prone to clumping, one CFU may represent more than one cell or genome. In both experiments A and B, the CT values of BCG cells lysed using either device were comparable, indicating similar lysis efficiency. For experiment A, only one replicate was tested for the 108 CFU/ml concentration using the OmniLyse blender. In experiment A, glass beads were used in the BioSpec device, while zirconia/silica beads were used in all other experiments.
In the first experiment (Fig. 4A), glass beads were used in the BioSpec device and zirconia/silica beads were used in the OmniLyse blender. In the second experiment (Fig. 4B), zirconia/silica beads were used in both devices. The CT values obtained using the BioSpec device are almost identical in both experiments, indicating that the bead type used does not appreciably affect the lysis efficiency.
In lysis experiments of B. subtilis spores and M. bovis BCG cells, averaged CT values over the range of concentrations tested for both devices were within a standard deviation of each other. Having BSA present in the master mix was found to be important for obtaining robust and reproducible experimental results when zirconia/silica beads were used in either device, especially at lower BCG cell concentrations, as discussed in below. When glass beads were used in the Biospec device, robust and reproducible results were obtained regardless of the cell concentration or presence of BSA.
B. subtilis spores are known to carry extracellular DNA (5, 19). To rule out the possibility that PCR experiments predominantly detected extracellular DNA, we conducted a control experiment. B. subtilis spores at concentrations of 104, 106, and 108 CFU/ml were either lysed using the BioSpec system, followed by PCR analysis, or analyzed via PCR without the mechanical disruption step. At these concentrations, the unprocessed spores amplified approximately 4 to 6 CT values later than the samples that underwent mechanical lysis (Fig. 6). A similar experiment was performed for M. bovis BCG cells at concentrations of 108 and 106 CFU/ml (see Fig. S2A in the supplemental material). Again, the unprocessed bacterial cells amplified approximately 4 to 6 CT values later than the samples that underwent mechanical lysis. Therefore, mechanical lysis results in 10- to 100-fold more PCR-amplifiable target DNA, assuming a PCR efficiency of ca. 100%. The true difference is likely greater since unprocessed control samples probably underwent partial heat lysis during the 10-min, 95°C PCR hot start step. In another experiment, intact M. bovis BCG cells were removed from processed and unprocessed samples by centrifugation prior to PCR. In this case, unprocessed samples amplified ∼10 CT values later than processed samples (see Fig. S2B in the supplemental material).
Fig. 6.
Control experiment to differentiate detection of B. subtilis extracellular DNA from detection of DNA as result of B. subtilis spore lysis. Samples lysed using the BioSpec system followed by PCR analysis (⧫) amplify approximately 4 to 6 CT values earlier than samples containing the same concentration of B. subtilis spores but analyzed via PCR without the mechanical disruption step (□). Mechanical lysis results in 10- to 100-fold more PCR-amplifiable target DNA, assuming a PCR efficiency of ca. 100%.
For the M. bovis BCG study, lysed samples were frozen at −80°C prior to PCR processing. To investigate whether this freeze-thaw step artificially enhanced mycobacterial lysis, we conducted a separate experiment, wherein M. bovis BCG samples lysed using the BioSpec device were either kept on ice or frozen at −80°C prior to PCR processing. The CT values obtained using both processing methods were comparable (see Fig. S2A in the supplemental material), and averages were within a standard deviation of each other.
Internal control results.
To assess secondary effects such as DNA degradation and PCR inhibition, we added an internal control (IC) to each reaction upstream of the lysis step, consisting of a fixed amount of purified genomic DNA from a different organism. If there are no secondary effects or if unaccounted secondary effects are similar for both devices and at each concentration of the targeted microorganism, then the IC CT values should be comparable and relatively constant.
For the B. subtilis lysis experiment, E. coli genomic DNA was added as an IC and processed by PCR for the first experiment only, as shown in Fig. 5A. For the M. bovis BCG lysis experiments, A. hydrophila genomic DNA was added as an IC and processed by PCR in both experiments. The results for IC amplification were similar in both experiments; data for the second experiment are shown in Fig. 5B.
Overall, the CT values of the internal controls were comparable for the OmniLyse device and the BioSpec Mini-BeadBeaters. Cross-amplification, i.e., amplification of the DNA from the targeted microorganism using IC-specific primers and vice versa, was not observed in either case. In the B. subtilis and M. bovis BCG experiments, correct amplicon formation was confirmed via gel electrophoresis and melting-curve analysis, respectively. Amplification with B. subtilis specific primers for some of the lower B. subtilis concentrations and the no-target controls also contained primer dimers.
The presence of BSA in the master mix appeared to be critical to obtaining reproducible and robust CT values following mechanical lysis using the zirconia/silica beads in either the OmniLyse or the BioSpec Mini-BeadBeater-8. When BSA was omitted in initial experiments with M. bovis BCG, the CT values were inconsistent, with many failed amplifications, especially at the lower M. bovis BCG concentrations. The CT values of the IC (A. hydrophila genomic DNA) increased with decreasing M. bovis BCG concentration. Late amplification was also observed in the no-target control, containing A. hydrophila DNA only in buffer that was processed through the respective bead beating device. Dilution series of the purified A. hydrophila DNA not processed through a lysis device did not show this effect. Adding BSA directly to the bacterial suspension prior to lysis, or to the master mix following lysis, eliminated these effects and enabled consistent results.
DISCUSSION
The miniaturized OmniLyse device mechanically lyses B. subtilis spores and M. bovis BCG cells with efficacy similar to the industry-standard benchtop BioSpec Mini-BeadBeater, based on the comparable CT values obtained from PCRs with primers specific for these two microorganisms. Bacillus spores carry extracellular DNA that may be bound on the bacterial spore coats. This extracellular DNA is not readily removed by multiple wash steps and can be amplified during PCR even without mechanical lysis (5, 19). Although it is possible that extracellular DNA was coamplified in the experiments described here, a control experiment demonstrated that mechanical lysis significantly increased (between 10- and 100-fold) the amount of PCR-amplifiable target DNA, a finding in line with the expectation that mechanical lysis is required to achieve a low limit of detection. In addition, a previous time series lysis study of DNase-treated B. subtilis spores using the OmniLyse device clearly demonstrated that this device mechanically disrupts spores and liberates intracellular DNA (13). Likewise, other studies have confirmed the ability of the BioSpec Mini-BeadBeater to lyse Bacillus spores (20, 24).
Using the protocol described here, the experiments did not appear to be confounded by secondary effects such as DNA degradation or PCR inhibition, based on consistent internal control CT values obtained for both devices and at each concentration of the targeted microorganism. For lysis experiments using zirconia/silica beads in either device, the presence of BSA in the master mix was required for consistent, reproducible results. Although further experiments are required to fully understand this effect, it is possible that BSA relieves a PCR-inhibitory effect linked to fragments of zirconia/silica beads carried into the PCR. This hypothesis is supported by the observation that when glass beads were used in the BioSpec device in place of zirconia/silica beads, no inhibition was observed, and reproducible results were obtained at all cell concentrations with or without BSA. BSA is commonly used as a PCR facilitator (29, 36), and it may bind to fragments of zirconia/silica beads carried into the PCR. The inhibitory effect was unlikely to be caused by cellular debris because the effect was less prominent at higher concentrations of the targeted microorganism and was also observed in the no-target controls containing IC DNA only in buffer processed through a bead-beating device.
To the extent possible, we used similar conditions in the OmniLyse and the BioSpec devices, which included the use of similar or identical beads and bead-to-volume ratios. The processing time per sample for the OmniLyse device (1.5 min without rest), however, was significantly less than for the BioSpec BeadBeater devices (three 1-min pulses with 30-s rests between pulses). The BioSpec processing time used in the present study was based on standard published protocols (10, 13, 33, 35), and appears to be the time required using this system for effective lysis of Bacillus spores and mycobacterial cells. In most experiments, the same zirconia/silica beads were used in both devices. BioSpec protocols reported in the literature use either glass beads or zirconia/silica beads for lysis of mycobacteria (22, 33, 35). Based on our results (Fig. 4), the use of glass beads versus zirconia/silica beads in the BioSpec did not appreciably affect the lysis efficiency using this device and did not alter the relative lysis efficiency compared to the OmniLyse.
In summary, for both lysis-resistant bacterial strains studied, mechanical lysis using the OmniLyse blender and the BioSpec Mini-BeadBeaters resulted in similar CT values for PCR analysis of processed samples over the range of concentrations studied. Amplification of IC DNA spiked at a constant concentration into the samples upstream of lysis using either device resulted in consistent PCR CT values for all processed samples. Overall, the OmniLyse device facilitates effective lysis of tough-walled organisms in a miniaturized, disposable, battery-operated format, with an efficiency comparable to that of a widely used benchtop system. Future work will focus on combining lysis with extraction of nucleic acids within the bead blender with suitable nucleic acid amplification and detection methods in a fully integrated cartridge for handheld point-of-care nucleic acid testing in low-resource settings.
Supplementary Material
ACKNOWLEDGMENTS
We thank Donald S. Kenney (Gustafson, LLC) for graciously supplying B. subtilis spores for this study.
This study was supported by Public Health Service grant AI090831 from the National Institute of Allergy and Infectious Diseases. P.E.V. acknowledges support through a Science, Mathematics, and Research for Transformation (SMART) scholarship from the Department of Defense.
P.E.V., K.M.W., B.E., G.A.C., and A.N. declare that they have no conflicts of interest. R.D., A.N., and B.I. each own a minority interest in Claremont BioSolutions, LLC. J.S. works on a part-time basis for Claremont BioSolutions, LLC.
Footnotes
Supplemental material for this article may be found at http://jcm.asm.org/.
Published ahead of print on 4 May 2011.
REFERENCES
- 1. Amaro A., Duarte E., Amado A., Ferronha H., Botelho A. 2008. Comparison of three DNA extraction methods for Mycobacterium bovis, Mycobacterium tuberculosis, and Mycobacterium avium subsp. avium. Lett. Appl. Microbiol. 47:8–11 [DOI] [PubMed] [Google Scholar]
- 2. Anderberg R. J., Strachan J. A., Cangelosi G. A. 1995. Purification of DNA from Mycobacterium species without sonication or phenol. Biotechniques 18:217–219 [PubMed] [Google Scholar]
- 3. Balamurugan R., Venkataraman S., John K. R., Ramakrishna B. S. 2006. PCR amplification of the IS6110 insertion element of Mycobacterium tuberculosis in fecal samples from patients with intestinal tuberculosis. J. Clin. Microbiol. 44:1884–1886 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Barken K. B., Haagensen J. A. J., Tolker-Nielsen T. 2007. Advances in nucleic acid-based diagnostics of bacterial infections. Clin. Chim. Acta 384:1–11 [DOI] [PubMed] [Google Scholar]
- 5. Belgrader P., et al. 1999. A minisonicator to rapidly disrupt bacterial spores for DNA analysis. Anal. Chem. 71:4232–4236 [DOI] [PubMed] [Google Scholar]
- 6. Belisle J. T., Mahaffey S. B., Hill P. J. 2009. Isolation of Mycobacterium species genomic DNA, p. 1–12 In Parish T., Brown A. C. (ed.), Mycobacteria protocols, 2nd ed., vol. 465 Humana Press, Inc., Totowa, NJ: [DOI] [PubMed] [Google Scholar]
- 7. Boehme C. C., et al. 2010. Rapid molecular detection of tuberculosis and rifampin resistance. N. Engl. J. Med. 363:1005–1015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Boom R., et al. 1990. Rapid and simple method for purification of nucleic acids. J. Clin. Microbiol. 28:495–503 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Brennan P. J., Nikaido H. 1995. The envelope of mycobacteria. Annu. Rev. Biochem. 64:29–63 [DOI] [PubMed] [Google Scholar]
- 10. Cangelosi G. A., Weigel K. M., Lefthand-Begay C., Meschke J. S. 2010. Molecular detection of viable bacterial pathogens in water by ratiometric pre-rRNA analysis. Appl. Environ. Microbiol. 76:960–962 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. de Boer R., et al. 2010. Improved detection of microbial DNA after bead-beating before DNA isolation. J. Microbiol. Methods 80:209–211 [DOI] [PubMed] [Google Scholar]
- 12. Dhandayuthapani S., Zhang Y., Mudd M. H., Deretic V. 1996. Oxidative stress response and its role in sensitivity to isoniazid in mycobacteria: characterization and inducibility of ahpC by peroxides in Mycobacterium smegmatis and lack of expression in M. aurum and M. tuberculosis. J. Bacteriol. 178:3641–3649 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Doebler R. W., et al. 2009. Continuous-flow, rapid lysis devices for biodefense nucleic acid diagnostic systems. J. Assoc. Lab. Automation 14:119–125 [Google Scholar]
- 14. Fan J., Kraft A. J., Henrickson K. J. 2006. Current methods for the rapid diagnosis of bioterrorism-related infectious agents. Pediatr. Clin. N. Am. 53:817. [DOI] [PubMed] [Google Scholar]
- 15. George K. M., Yuan Y., Sherman D. R., Barry C. E. 1995. The biosynthesis of cyclopropanated mycolic acids in Mycobacterium tuberculosis: identification and functional analysis of CMAS-2. J. Biol. Chem. 270:27292–27298 [DOI] [PubMed] [Google Scholar]
- 16. Helb D., et al. 2010. Rapid detection of Mycobacterium tuberculosis and rifampin resistance by use of on-demand, near-patient technology. J. Clin. Microbiol. 48:229–237 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Hoffmann C., Leis A., Niederweis M., Plitzko J. M., Engelhardt H. 2008. Disclosure of the mycobacterial outer membrane: cryo-electron tomography and vitreous sections reveal the lipid bilayer structure. Proc. Natl. Acad. Sci. U. S. A. 105:3963–3967 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Hurley S. S., Splitter G. A., Welch R. A. 1987. Rapid lysis technique for mycobacterial species. J. Clin. Microbiol. 25:2227–2229 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Johns M., Harrington L., Titball R. W., Leslie D. L. 1994. Improved methods for the detection of Bacillus anthracis spores by the polymerase chain reaction. Lett. Appl. Microbiol. 18:236–238 [Google Scholar]
- 20. Jones C. A., Padula N. L., Setlow P. 2005. Effect of mechanical abrasion on the viability, disruption and germination of spores of Bacillus subtilis. J. Appl. Microbiol. 99:1484–1494 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Kamei T., et al. 2005. Microfluidic genetic analysis with an integrated a-Si:H detector. Biomed. Microdevices 7:147–152 [DOI] [PubMed] [Google Scholar]
- 22. Kaser M., Ruf M. T., Hauser J., Marsollier L., Pluschke G. 2009. Optimized method for preparation of DNA from pathogenic and environmental mycobacteria. Appl. Environ. Microbiol. 75:414–418 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Kazakov S., Bonvouloir E., Gazaryan I. 2008. Physicochemical characterization of natural ionic microreservoirs: Bacillus subtilis dormant spores. J. Phys. Chem. B 112:2233–2244 [DOI] [PubMed] [Google Scholar]
- 24. Longchamp P., Leighton T. 1999. Molecular recognition specificity of Bacillus anthracis spore antibodies. J. Appl. Microbiol. 87:246–249 [DOI] [PubMed] [Google Scholar]
- 25. Millar B. C., Xu J. R., Moore J. E. 2007. Molecular diagnostics of medically important bacterial infections. Curr. Issues Mol. Biol. 9:21–39 [PubMed] [Google Scholar]
- 26. Novak-Weekley S. M., et al. 2010. Clostridium difficile testing in the clinical laboratory by use of multiple testing algorithms. J. Clin. Microbiol. 48:889–893 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Odumeru J., Gao A., Chen S., Raymond M., Mutharia L. 2001. Use of the bead beater for preparation of Mycobacterium paratuberculosis template DNA in milk. Can. J. Vet. Res.-Rev. Can. Recherche Vet. 65:201–205 [PMC free article] [PubMed] [Google Scholar]
- 28. Pemov A., Modi H., Chandler D. P., Bavykin S. 2005. DNA analysis with multiplex microarray-enhanced PCR. Nucleic Acids Res. 33:e11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Radstrom P., Knutsson R., Wolffs P., Lovenklev M., Lofstrom C. 2004. Pre-PCR processing: strategies to generate PCR-compatible samples. Mol. Biotechnol. 26:133–146 [DOI] [PubMed] [Google Scholar]
- 30. Rao S. S., Mohan K. V., Atreya C. D. 2010. Detection technologies for Bacillus anthracis: prospects and challenges. J. Microbiol. Methods 82:1–10 [DOI] [PubMed] [Google Scholar]
- 31. Salazar O., Asenjo J. A. 2007. Enzymatic lysis of microbial cells. Biotechnol. Lett. 29:985–994 [DOI] [PubMed] [Google Scholar]
- 32. Schaeffer A. B., Fulton M. D. 1933. A simplified method of staining endospores. Science 77:194. [DOI] [PubMed] [Google Scholar]
- 33. Sherman D. R., et al. 2001. Regulation of the Mycobacterium tuberculosis hypoxic response gene encoding alpha-crystallin. Proc. Natl. Acad. Sci. U. S. A. 98:7534–7539 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Ulrich M. P., et al. 2006. Evaluation of the Cepheid GeneXpert (R) system for detecting Bacillus anthracis. J. Appl. Microbiol. 100:1011–1016 [DOI] [PubMed] [Google Scholar]
- 35. Via L. E., Falkinham J. O. 1995. Comparison of methods for isolation of Mycobacterium avium complex DNA for use in PCR and rapid fingerprinting. J. Microbiol. Methods 21:151–161 [Google Scholar]
- 36. Wilson I. G. 1997. Inhibition and facilitation of nucleic acid amplification. Appl. Environ. Microbiol. 63:3741–3751 [DOI] [PMC free article] [PubMed] [Google Scholar]