Use of polyphase continuous excitation based on the Frank sequence in EPR.

Mark Tseitlin a, Richard W Quine b, Sandra S Eaton a, Gareth R Eaton a
PMCID: PMC3148075  NIHMSID: NIHMS310424  PMID: 21737326

Abstract

Polyphase continuous excitation based on the Frank sequence is suggested as an alternative to single pulse excitation in EPR. The method allows reduction of the source power, while preserving the excitation bandwidth of a single pulse. For practical EPR implementation the use of a cross-loop resonator is essential to provide isolation between the spin system and the resonator responses to the excitation. Provided that a line broadening of about 5% is acceptable, the cumulative turning angle of the magnetization vector generated by the excitation sequence can be quite large and can produce signal amplitudes that are comparable to that achieved with a higher power 90° pulse.

Keywords: EPR, Frank sequence, polyphase continuous excitation

1. Introduction

In typical pulse NMR and EPR experiments the spins are excited with short high-power pulses. High-power amplifiers are expensive and for in vivo applications it is desirable to decrease power deposition to meet restrictions on the specific absorption rate. One approach that has been demonstrated to decrease power requirements is stochastic excitation [1-4]. It has also been shown that substitution of a short 90° pulse by a series of lower intensity pulses with phases selected with the Frank sequence allowed substantial reduction in peak excitation power in NMR spectroscopy, while maintaining the same excitation bandwidth [5].

In a magnetic resonance experiment with a single on-resonance pulse of length tp, the rf (microwave) magnetic field B1 turns the net magnetization vector by an angle:

α=γB1tp, (1)

where γ is the electron or nuclear gyromagnetic ratio and α is in radians. The power required to achieve turning angle α is given by Eq. (2):

power=Dα2tp2, (2)

where D is a parameter that can be measured for a resonator/sample combination with a particular coupling [6]. This expression shows that the power requirement is proportional to α2, which makes the small turning angles of the Frank sequence advantageous. Typical electron spin relaxation times are orders of magnitude shorter than proton NMR relaxation times, which requires that pulsed EPR experiments are performed with tp orders of magnitude shorter than for NMR. Since power increases inversely proportional to tp2, the peak power required for EPR experiments is substantially greater than for NMR, and reducing the required power is even more advantageous than in NMR. Depending on the efficiency of the resonator, the coupling of the resonator and the lengths of the pulses, the peak power for pulse EPR experiments may be in the kilowatts.

To explore whether the Frank sequence of polyphase excitation could be used in EPR, an experiment was conducted with 256 small turning-angle pulses for an aqueous sample of a triarylmethyl radical [7]. The present paper reports calculations that were performed to examine the following aspects of polyphase excitation and detection: (i) optimization of excitation power; (ii) comparison of (a) data analysis by cross-correlation with the Fourier transform of the excitation phase sequence and (b) deconvolution with the Fourier transform of the excitation waveform including the pulse profile, (iii) spectral distortion due to deviation of the spin response from linearity, and (iv) comparison with single-pulse excitation. Experiments were done at 250 MHz for triarylmethyl radicals and at 1 GHz for lithium phthalocyanine (LiPc) to test the predictions concerning power and deviation from linearity. A block diagram for the experiments and data analysis is shown in Fig. 1. Details are provided in the following paragraphs.

Fig. 1.

Fig. 1

Block diagram of the polyphase continuous excitation data acquisition and analysis.

2. Samples and Spectroscopy

At 250 MHz the samples were 0.2 mM aqueous solutions of the trityl radical OX63 (methyl tris(8-carboxy-2,2,6,6-tetra(hydroxyethyl)-benzo[1,2-d:4,5-d’]bis(1,3)-dithiol-4-yl)-tripotassium salt) [8] in a 16 mm o.d. tube or the deuterated analog OX63-d24 in a 25 mm o.d. tube. Both samples were purged with N2 and flame-sealed. Lithium phthalocyanine (LiPc) prepared electrochemically following procedures in the literature [9,10] was provided by Prof. Swartz, Dartmouth University. For the 1 GHz experiments multiple small crystals of LiPc were placed in a 3-mm OD tube. The tube was extensively evacuated and then flame sealed.

CW experiments at 250 MHz were performed with a reflection resonator that has been described previously [11]. To provide isolation between the spin excitation and detected spin response [2,12] in the polyphase continuous excitation experiments, cross-loop resonators (CLR) were used. The 250 MHz resonator has been described [13] and the 1 GHz resonator is analogous, but with modified dimensions, to a previously described L-band resonator [14]. For the 250 MHz CLR containing the 16 mm aqueous sample the quality factor for the excitation resonator was Q1 = 6, and for the detection resonator Q2 = 13. For the L-band experiments the excitation resonator had Q1 = 250, and the detection resonator had Q2 = 250. At both frequencies the isolation between the two resonators was about 60 dB.

To facilitate creation of the excitation waveforms, a Matlab program was written to produce files in a format that could be read by an arbitrary waveform generator (AWG). The 8-bit signal channel of the AWG was used to create the waveform with constant frequency and varying phase, determined using the Frank sequence. The length of the excitation sequences were selected to be 5 to 6 T2. The spins are excited by a time-varying voltage from the AWG. At 250 MHz a Tektronix AWG2041 with maximum clock frequency of 1024 MHz gave 4 points per cycle. The excitation waveform had 1024 rf phases during a time of 29 μs. The AWG trigger synchronized the waveform with the Bruker SpecJet II digitizer that was run at the 1 GS/s sampling rate. The signal was amplified by 44 dB prior to the digitizer. Data analysis used 29,000 points which encompassed one full Frank sequence. At 1 GHz a Tektronix AWG7122C operating with a clock frequency ten times the resonance frequency, gave 10 points per cycle. The excitation waveform had 256 rf phases during a time of 15 μs. An Acqiris U1084A digitizer was used with a sampling rate of 4 GS/s. The excitation waveform was amplified by 20 dB prior to the resonator and the signal was amplified by 30 dB prior to the digitizer. At both frequencies the detected signals were processed by digital phase sensitive detection [15-19]. Direct detection at the carrier frequency followed by digital phase sensitive detection has two major advantages: (i) the noise is defined by the thermal noise and the first stage high frequency amplifier; and (ii) the quadrature signals are perfectly orthogonal, which is important for analysis of polyphase excitation data.

3. Optimization of polyphase excitation

Analysis of data obtained by polyphase excitation is based on the assumption that the spin system is linear. A linear system is characterized by its impulse response function h(t), which is called the free induction decay (FID) in magnetic resonance. Excitation with an arbitrary shape x(t) produces a response y(t) that is equal to the convolution of x(t) and h(t). In the Fourier conjugate frequency domain convolution becomes multiplication:

Y(ω)=H(ω)X(ω) (3)

where X(ω), Y(ω), H(ω) are Fourier transforms of x(t), y(t), and h(t). For a perfect noise-free measurement, H(ω) can be accurately restored by Fourier deconvolution: H(ω)=Y(ω)/X(ω). For data containing noise it is important to select x(t) to avoid small values in the denominator X(ω) that would amplify noise. The best choice would be a function x(t) for which

X(ω)X(ω)=C2, (4)

where C is an arbitrary real constant. The periodic polyphase sequences that satisfy this criterion are called ‘perfect’. The autocorrelation of a perfect polyphase sequence is a delta-function and the absolute value of its Fourier transform is constant, | X(ω) | = | C |.

3.1 Design of polyphase excitation sequence to optimize use of power

The first implementations of the Frank sequence for NMR [5] and EPR [7] used the excitation waveform shown in Fig. 2. There were N pulses of equal duration tp separated by gaps of ts-tp. Pulses had equal intensity and the same carrier frequency. The average power for the sequence of pulses shown in Fig. 2 can be found from Eq. (5), which differs from the expression for peak power given in Eq. (2).

<power>=DNα2tpT (5)

Fig. 2.

Fig. 2

Schematic of the polyphase excitation sequence. The duration of each pulse is tp, and the entire sequence is repeated cyclically with period T. The time between the start of two successive pulses is ts, so ts N = T. Phases p1 .. pN are the elements of the Frank sequence.

It follows from Eq. (5) that to minimize average power for a given number of pulses N and sequence duration T, tp should be maximized.

The minimum power requirement is achieved when there are no gaps and ts = tp = T/N. The end of each pulse is the beginning of the next pulse, so there are no longer discrete pulses, and the waveform becomes polyphase continuous excitation. In this case average power is equal to peak power. In the original NMR paper [5] the gaps were needed to permit collection of data after the so-called ‘dead time’, the time for the resonator to dissipate energy from the excitation pulse. Since nuclear spin relaxation times are much longer than the ring-down time of the resonator, data acquisition after a dead time allows separation of the spin system response from the resonator response. Electron spin relaxation lasts for about the same time as the resonator ring-down, so data are collected in the presence of the resonator ring-down. For EPR, having gaps between pulses exacerbates the background problem, because it requires more power for a similar turning angle, thus producing a more intense resonator ring-down signal. Cross-loop resonators isolate excitation from detection by 40-60 db [13,14], which decreases the resonator background signal by the amount of the isolation.

The power for Frank sequence EPR with ts = tp can be compared with the power required for a single π/2 pulse, based on Eqs. (1, 2, 5). The reduction is [π/(2α)]2 or [π/(2α)]2/N for the peak or average power, respectively.

3.2 Data Analysis

In the first implementation of the Frank sequence for NMR [5], one data point was acquired shortly before each successive pulse, forming an array of data points with length N (Fig. 2). In the initial EPR implementation [7] a few points were averaged between successive pulses with the goal of improving the signal-to-noise ratio, SNR. The number of data points in the resultant array again was equal to N. Cross-correlation of the data arrays and the Frank sequence produced NMR and EPR FID with N elements. Although the outcomes were demonstrated, the mathematical basis has not been discussed. A fundamental issue is that the Frank sequence is a discrete series of numbers, whereas the excitation waveform is a continuous function in which the phase changes are encoded by the sequence. The following derivation examines the implications for data analysis.

An analytical expression for the Fourier transform of the periodic continuous excitation function x(t) with ts = tp is needed. Since the function has period T, it can be represented as a Fourier series:

x(t)=x(t+T)=k=Xkej2πf0t,f0=1T (6)

where the discrete function Xk is defined for a set of angular frequencies that are integer (k) multiples of 2πf0:

Xk=1T0Tx(t)ejk2πf0tdt (7)

Using the fact that the function x(t) is composed of N phases (Fig. 2), Eq. (7) can be rewritten as a sum of N integrals corresponding to the N phases:

Xk=ejp0T0tsf(t)ejk2πf0tdt+ejp1Tts2tsf(tts)ejk2πf0tdt++ejpN1T(N1)tsNtsf(t(N1)ts)ejk2πf0tdt=1Tn=0N1nts(n+1)tsejpnf(tnts)ejk2πf0tdt. (8)

Here f(t) describes a square pulse with duration ts, and pn are the phases defined by the elements of the Frank sequence. Eq. (8) can be simplified by changing variables to τ = t-nts , where n is the number of the pulse:

Xk=1T0tsf(τ)ejk2πf0τdτ×n=0N1ejpnejn2πf0kts. (9)

Eq. (9) can be represented as multiplication of two discrete functions:

Xk=FkPk (10)

The first term, Fk, is the Fourier series for a square pulse with duration ts. For a pulse with amplitude B1, f(τ) can be set equal to B1 in the interval 0 to ts, and the expression for Fk can be found by integration:

Fk=B1Tejk2πf0ts1jk2πf0=tpB1Tejk2πf0ts2sinc(kπf0ts), (11)

where sinc(x)=sin(x)/x. The complex coefficient of the sinc function appears in Eq. (11) as a result of the asymmetry of the pulse with respect to t = 0. The second term in Eq. (10) is Pk,

Pk=n=0N1ejpnej2πnkN (12)

which is the discrete Fourier transform of the Frank sequence. It is periodic in the frequency domain with period equal to 2πf0N (Eq. (9)). Examination of Eq. (11) and (12) shows that function Xk is the product of periodic function Pk and function Fk, which is not periodic, where both of these are discrete. Substitution of Eq. (10) into Eq. (3) gives:

Yk=FkPkHk (13)

where Yk and Hk are discrete functions that are defined for the same angular frequencies as Pk and Fk.

Eq. (13) provides a basis for comparison of single pulse FID and multiphase Frank sequence experiments. For a single pulse experiment the summation in Eq. (8) is reduced to only the first term, which has p0 = 0, giving Eq. (14)

Xk(single pulse)=1T0tsf(t)ejk2πf0tdt=Fk (14)

and Yk (single pulse) becomes

Yk(single pulse)=FkHk(single pulse) (15)

As shown in Eq. (11), Fk is proportional to B1. Since |Pk| = √N, comparison of Eq. (13) and Eq. (15) indicates that if the same B1 (peak power) is used, the signal obtained by polyphase continuous excitation in the linear response regime is √N times larger than that obtained with a single pulse. The excitation bandwidth is the same for the two experiments.

Eq. (13) also is the basis for the cross-correlation method that was used in analyzing prior data (Fig. 1) [5,7]. In the frequency domain cross-correlation is performed by multiplication of Yk by Inline graphic, so Eq. (13) becomes Eq. (16) because Inline graphic = N. The result

HkFk=YkPkN (16)

is proportional to the product of EPR spectrum Hk and the Fourier transform of the excitation pulse profile (Fk). This spectrum is the same as the Fourier transform of the FID that would have been obtained with a single pulse of duration ts. A pulse provides an excitation bandwidth of about 1.2/ts (full width at half height) [20]. If the spectral width is small with respect to that bandwidth, Fk is flat in the region of interest, and cross-correlation produces an undistorted spectrum. Otherwise, the excitation profile of the pulse has to be taken into account to accurately restore the lineshape.

An alternate way to analyze the data (Fig. 1) is deconvolution where Hk = Yk/Xk (Eq. (3) but now written for discrete functions), and Xk = Fk Pk (Eq. 10) is the Fourier transform of the excitation waveform (Eq. (7)). The resulting spectrum, Hk, is unaffected by the shape of the pulse (Eq. (17)). Thus, if the excitation profile is not uniform over the spectrum, the relative amplitudes and phases of lines in the spectrum will be more accurate if data are analyzed using Eq. (17) instead of Eq. (16).

Hk=Yk(PkFk). (17)

3.3 Resonator Q, background signal, and spectral window

3.3.a Resonator Q. In a pulse experiment the resonator acts as a band pass filter

When a cross-loop resonator is used both the filtering of the excitation by resonator 1 (Q1) and the filtering of the detected signal by resonator 2 (Q2) must be considered. If Q for both resonators is sufficiently low, the impact on the spectrum is negligible. However, lowering Q decreases signal amplitude which decreases SNR, and decreases the resonator efficiency, which requires higher power to achieve the same turning angle (Eq. (2)). Thus the selection of Q is a compromise between these factors.

3.3.b Background signal

The dominant contribution to the background signal is direct ‘leakage’ from the excitation resonator to the detection resonator, which has the shape λ Fk Pk, and is the same as Eq. (13) except for the spin response. Here λ is the voltage attenuation coefficient describing isolation between the excitation and detection resonators. Depending on the CLR design, λ varies from 100 to 1000 [13,14]. The majority of this contribution is removed by subtraction of an off-resonance signal (Fig. 1), although this decreases SNR by √2. For a weak EPR signal there often is a residual background contribution to Hk, which is difficult to separate from signal in the frequency domain. In these cases Hk was inverse Fourier transformed to h(t). In the time domain the decay of the background signal depends on Q1 and Q2, analogous to the ring-down after a single pulse. If resonator Q is sufficiently low, the background signal decays faster than the spin response so it is greater for the first few points of h(t). Elimination of these points, which is similar to delaying data collection until after the ‘dead time’ in a one-pulse FID experiment, decreases SNR but improves lineshape. After elimination of the first few points, h(t) was Fourier transformed to recover Hk with improved lineshape (Fig. 1).

If isolation is not perfect, a weak EPR signal is superimposed on a large background signal. This large signal may saturate the first stage amplifier unless it has a very large dynamic range. If the background saturates the amplifier it cannot be accurately subtracted. In addition, when the background signal is very large, the superimposed EPR signal may be less than one bit of the digitizer. The effective number of bits in the digitizer can be increased by signal averaging, but excessive averaging usually results in leakage of coherent clock signals into the digitized data. The constraints imposed by amplifier saturation and limited bits of the digitizer make minimization of the background by maximizing the isolation between resonators very important.

3.3.c. Spectral window

In previous implementations of the Frank sequence for NMR [5] and EPR [7], signal sampled with the rate fs = 1/ts was used for cross-correlation. This limits the spectral window of Hk to the interval [-fN, fN], where fN = fs/2 is the Nyquist frequency. This sampling could increase noise in the spectrum because of the aliasing of frequencies higher than fN. Function Xk in Eq. (7) and Eq. (13) is defined from –infinity to +infinity, so the sampling frequency can be selected, independent of ts, to define a desired signal bandwidth and avoid aliasing. There is still a limitation on the frequencies for which the EPR spectrum can be restored, which comes from the multiplication by the sinc function in Eq. (11) that has zeros at frequencies f = m fs, where m is an arbitrary integer > 0.

3.4 Linearity of the spin system

For a single pulse, the on-resonance magnetic field B1 turns magnetization by an angle defined in Eq. (1). The signal measured in an experiment is proportional to the nonlinear function sin(α). Increasing B1 by a factor of two, increases the signal amplitude by a factor of sin(2α)/ sin(α), which is not a factor of 2. This shows that the spin system is non-linear. However, for very small turning angles sin(α) ≈ α, and the spin system approximates linear behavior. In this case Eq. (3) can be used to recover the EPR spectrum obtained by polyphase excitation with either Eq. (16) or (17). For practical application of polyphase excitation (Fig. 1) it is important to estimate the threshold turning angle for which the deviation from linearity is so large that the use of Eq. (3) results in significant distortion of the lineshape. This limit was determined by numerical simulation with parameters based on the relaxation times for LiPc which was used in the experiments, T2 = T1 = 2.32 μs. LiPc has a Lorentzian lineshape [21]. The sequence duration T was set to ~6T2 (T = 15 μs). Six T2 instead of the 5T2 originally suggested [5], was used to ensure decay to baseline, provide leeway if there is a distribution of values of T2, and provide opportunity for zero-filling, which increases spectral resolution (Eq. (6)). If there is a distribution of T2 values, T should be selected to be at least five times the longest T2. N = 256 was selected to ensure adequate signal bandwidth. Simulations were performed with the same T and N as the experiments. In the simulations it was assumed that resonator Q was low enough that its impact on lineshape was negligible.

In the simulations the Bloch equations were solved numerically for an EPR line on resonance at a series of B1 amplitudes. Calculated Mx and My components of the magnetization for a full cycle were combined to form a complex function y(t) that was Fourier transformed and deconvolved using Eq. (17) to obtain the EPR spectrum. In the linear response regime h(t) decays exponentially to approximately baseline within the time 5T2. However, in the nonlinear regime artifacts extend beyond 5T2. To minimize the impacts of the artifacts, an apodization filter with cutoff at 5T2 was applied to both the simulated and experimental h(t). Multiplication by that function does not noticeably broaden the signal, but reduces high frequency oscillations. Calculated Hk were fit with the Matlab nonlinear least-squares fitting routine lsqnonlin to obtain three characteristics: (i) amplitude, (ii) Lorentzian linewidth, and (iii) position. For the x-axis of Fig. 3, α was calculated from B1 using Eq. (1) with tp = ts and converted to degrees. The vertical axis for signal amplitude is scaled to that at α=90°/N (Fig. 3, left-axis). As B1 (and turning angle) increases, the signal amplitudes goes through a maximum and the linewidth increases, as is typical for power saturation [1]. The maximum signal amplitude occurs at α = 5.6° which is much larger than the α = 90°/N = 0.35°, initially proposed as optimum for the Frank sequence [5]. At the maximum in the power saturation curve the line is broadened by about 5% and the signal intensity is about 10 times greater than at α = 0.35°. These calculations show that the amplitude of the EPR signal can be increased by an order of magnitude, relative to that observed with the small turning angle initially proposed for the Frank sequence [5], if small lineshape distortion due to non-linearity of the spin system is acceptable. For α greater than the peak in the power saturation curve, the lineshapes were sufficiently distorted that fitting to a Lorentzian was not meaningful.

Fig. 3.

Fig. 3

Relative amplitudes (blue) and linewidths (green) as a function of turning angle α calculated for T = 15 μs, N = 256, and T2 = T1 = 2.32 μs. Simulated amplitudes (solid line) were normalized to 1.0 corresponding to amplitude at α=90°/N. Simulated linewidths are represented by a dashed line. Experimental values for LiPc are shown for relative amplitudes (blue diamonds) and FWHM linewidths (green circles).

To test the simulations, spectra of LiPc were obtained by polyphase excitation at υ0 = 1037.8 MHz. The maximum B1 in the resonator with this sample was measured at the full voltage output of the AWG, amplified to 4.3 Vpp. The on-resonance nutation frequency [22] gave B1 = 0.40 ± 0.05 G and the length of the pulse that maximized the amplitude of an FID gave B1 = 0.47 ± 0.05 G. B1 was varied by using different combinations of 3, 6, or 10 db attenuators along with changing the AWG output voltage. Comparison with the calculated saturation curve (blue line, Fig. 3) showed that maximum B1 = 0.46 G gave the best match. This value was used to calculate the turning angles in Fig. 3. Experimental points for relative amplitudes are in a good agreement with the calculated curve. The experimental linewidths are 2 - 3 mG larger than that expected from the independently measured value of T2, which is attributed to uncertainties in the measurements.

Fig. 4 shows experimental and simulated spectra corresponding to α = 5°, which is near the peak of the power saturation curve. Since phasing of the original experimental data is arbitrary, the phase was adjusted to make the line symmetrical. The phase is accurately defined by the simulations, but there are lineshape distortions at larger values of α that contribute to the small difference between the simulated and experimental lineshapes at α = 5 °.

Fig. 4.

Fig. 4

Comparison of simulated (dashed green) and experimental (solid blue) LiPc spectra for turning angle α = 5° which is near the peak of the power saturation curve (a) and for α = 8.1° (b).

Parameter α is the angle by which spins at resonance are turned in an individual segment of length ts in the polyphase sequence Eq. (1). The spin system is excited continuously with B1 of constant amplitude, but varying phase. To show the time evolution of the spin system the cumulative phase angle,αc, was examined:

αc=180πsin1(Mx2+My2M0) (18)

where Mx, My are the x, y components of the magnetization vector, M0 is the magnetization at thermal equilibrium, and αc is in degrees. Parameter αc is the net result of many phase changes and the relaxation processes. It reflects the overall intensity of the recorded signal and provides a basis for comparison of polyphase excitation and single pulse excitation with αc = α = 90°. Fig. 5 shows the time evolution of αc calculated with the Bloch equations, for a polyphase cycle after the spins have come to dynamic equilibrium in several prior cycles. Calculations were performed for α = 5°. Note that at t = 0 for this cycle αc is not zero because of the cumulative effect of the prior cycles.

Fig. 5.

Fig. 5

Time dependence of the cumulative turning angle αc during an excitation sequence of 256 phase changes and T = 15 μs, calculated using the Bloch equations for α = 5° and relaxation times characteristic of LiPc.

For a sample with the relaxation times characteristic of LiPc, the signal amplitude obtained with α = 5° is 10 times larger than with α = 90°/N = 0.35° (Fig. 3). Calculations were performed to compare these amplitudes with the results expected for a single pulse with three different turning angles (Table 1). As discussed above, at α = 0.35° the single pulse gives a signal that is weaker by a factor of 1/√N. At α = 5° the signal amplitude is less than 16 times the amplitude at α = 0.35° because of saturation. The amplitude obtained with a single 90° pulse is about the same as for the polyphase continuous excitation at 5°. The ratios are theoretical predictions. In practice the large background signal makes it difficult to achieve the theoretical advantages of polyphase excitation.

Table 1.

Calculated relative EPR amplitudes compared for constant tsa

Turning angle, α 0.35° 90°
Polyphase continuous 1.0 10 -
Single pulse 1/16 0.88 10
a

Calculated for T1 = T2 = 2.32 μs, T = 15 μs, N = 256, assuming no deadtime in detection. Amplitudes were scaled to that obtained with polyphase excitation at α = 0.35°.

Table 2 highlights the decreased power requirements for polyphase continuous excitation. Relative to a power of 1.0 for a single 90° pulsethe peak power required for α = 5° or α = 0.35° are only 0.003 and 1.3×10-5, respectively. The reduction in the average power over the time of the sequence is smaller, because the power is on continuously.

Table 2.

Powera for polyphase continuous excitation relative to single pulse, constant ts and N = 256.

Turning angle, α 0.35° 90°a
Peak power 1 / (257)2 1 / (18)2 1.0
Average power 1 /258 1 / 1.3 -
a

Normalized to 1.0 for the peak power required for a 90° pulse.

The results shown in Fig. 4 and 5 and Tables 1 and 2 were calculated for T1 = T2 = 2.32 μs, N = 256, and T = 15 μs. The same results would be obtained for other cases with T1 = T2 provided that T and 1/B1 are scaled proportionally.

4. Experimental results with trityl OX63

To test the generality of the method, polyphase continuous excitation was applied to deoxygenated aqueous solutions of the OX63 trityl radical at 250 MHz. The lineshapes for these spectra, which have substantial contributions from inhomogeneous broadening and resolved 13C hyperfine lines, test the performance of the sequence for multiline spectra. The experiments were performed with N = 1024, in the low-power linear regime.

Fig. 6 compares the spectra obtained with polyphase continuous excitation with a CW spectrum acquired with phase sensitive detection in a different resonator. The signal at 250 MHz obtained by polyphase excitation was digitized with a sampling interval of 1 ns. The resulting EPR signal was differentiated for comparison with the CW spectrum. The CW spectrum was recorded with 5 kHz modulation frequency at low modulation amplitude. The peak-to-peak linewidths of 160 mG are in good agreement for the spectra obtained by the two methods. The 13C hyperfine lines with splittings of 2.44 and 1.27 G [23] are clearly observed in both spectra.

Fig. 6.

Fig. 6

Comparison of EPR spectra of OX63 at 250 MHz obtained by CW (solid blue) and by polyphase continuous excitation (red dashed) with 1024 pulses. The Q of the CLR was lowered to 6 for the excitation resonator and to 10 for the detection resonator to minimize spectral distortion and to provide large enough bandwidth to observe the 13C-hyperfine lines that are marked with *.

The spectrum of OX63-d24 measured with polyphase continuous excitation in a homogeneous magnetic field (Fig. 7a) had FWHM = 114 mG, which is in good agreement with the value observed by CW EPR. To demonstrate the possibility of polyphase continuous excitation for EPR imaging, a 1D projection was measured in the presence of a constant magnetic field gradient of 0.44 G/cm (Fig. 7b). The experimental spectrum was compared with one that was calculated by convolution of the homogenous spectrum Fig. 7a with a 1D projection of the 25mm tube diameter.

Fig. 7.

Fig. 7

EPR spectra of OX63-d24 at 250 MHz obtained with polyphase continuous excitation with 1024 pulses. (a) Spectrum in homogeneous magnetic field, green line corresponds to absorption and blue to dispersion signals; (b) spectrum measured in magnetic field gradient of 0.44 G/cm. The green dashed line shows the absorption spectrum, and the blue solid line is computed based on the known sample geometry and homogeneous lineshape (a) profile.

5. Conclusions

The energy introduced into a resonator after a single 90° pulse is many orders of magnitude larger than that absorbed by a spin system, so the spin system response is weak with respect to the resonator response. In a standard pulsed NMR or EPR experiment data collection starts after a dead time during which the resonator ring down decays to a level that is small enough to permit detection of the weak spin response. Substitution of the single 90° pulse by polyphase continuous excitation significantly reduces the amount of energy that is used for excitation. The background signal becomes much smaller, but is present all of the time. Separation of the EPR and background signals based on the concept of ‘dead time’ is now not applicable during data collection. Instead, the dead time is taken into account by excluding the first few points in the time-domain data array after deconvolution or cross-correlation of the polyphase excitation data.

Mathematical analysis of the polyphase continuous excitation shows that EPR spectra can be restored by two different but closely related methods: by cross-correlation with the Fourier transform of the excitation phase sequence (Eq. (16) or deconvolution with the Fourier transform of the excitation waveform including the pulse profile (Eq. (17)), (Fig. 1). In the case of deconvolution the limitation of the sampling rate can be overcome. The deconvolution method might be more useful since it eliminates the effect of the pulse shape on EPR amplitudes and phases, which is important for EPR imaging.

If the pulse length for a single pulse is equal to ts for the polyphase continuous excitation, the excitation profiles are the same for the two experiments. If B1 is the same, the signal obtained with a single pulse is weaker than with polyphase excitation. Numerical solution of the Bloch equations followed by deconvolution and line fitting demonstrated that the EPR lineshape obtained by polyphase excitation is distorted and broadened as the excitation power increases. However, the amplitude of the spectrum increases faster than the linewidth in percentage terms. If a small (~5%) broadening is acceptable, the EPR signal amplitude can be increased by an order of magnitude relative to what would be obtained at the small turning angles initially proposed for Frank sequence excitation [5].

When available power is limited, as may occur when the resonator efficiency is low, polyphase excitation based on the Frank sequence excitation may be favorable. The major problem with continuous polyphase excitation is that, even with a cross loop resonator, the spin system response is still much weaker than the one induced in the resonator by the excitation. The large background signals are difficult to amplify and digitize within the linear region of the equipment, and complicate detection of the weak spin signals in the presence of the large background.

For spectral fidelity the resonator and excitation bandwidth should be large enough to encompass the entire spectrum. The power should be optimized to ensure that the spin system approximates a linear system. If additional power is available, B1 can be increased and ts decreased, producing a flatter profile over the region of interest. This would not contribute significantly to additional heating of the sample, because the bandwidth of resonator 1 limits the range of frequencies incident on the sample. Sequence duration T should be chosen such that the FID with the longest relaxation time decays to baseline. Since there is often some uncertainty in estimation of the longest T2 in the sample, T = 6 T2 is recommended. The T value can be corrected after performing the first measurement. Increasing the sequence duration beyond 6T2 does not impact the lineshape but would decrease the excitation bandwidth as a result of increased ts. Experimental testing of the polyphase continuous excitation method at 250 MHz and 1 GHz showed that it is applicable to samples with inhomogeneous line broadening and multi-line spectra. Measurement of a 1D projection with magnetic field gradient demonstrated that the method can potentially be used for EPR imaging. The resonator Q limits the excitation bandwidth. For the same Q the bandwidth increases proportional to frequency so for the same Q wider spectra can be excited at higher frequencies. However, implementation of continuous polyphase excitation at higher frequencies puts additional speed requirements on the AWG and the digitizer.

Research Highlights.

  • The use of a cross-loop resonator (CLR) allows continuous detection of the EPR signal during polyphase excitation.

  • Polyphase continuous excitation based on the Frank sequence minimizes power used.

  • Increasing excitation power results in a stronger EPR signal at the expense of line distortion.

  • For the relaxation times of LiPc the maximum signal is reached at turning angle α = 5°, and the line is broadened by 5%.

Acknowledgments

Funding for this work from NIH NIBIB grant EB000557 (GRE and SSE) and NIH NIBIB P41 EB002034 (Howard J. Halpern, PI) is gratefully acknowledged. Samples of OX63 and OX63-d24 were provided by Howard Halpern, University of Chicago. The L-band cross loop resonator was constructed by George A. Rinard (University of Denver). The loan of the AWG from Tektronix and the digitizer card from Agilent Acqiris is gratefully acknowledged.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Blümich B, Ziessow D. Saturation in Hadamard NMR spectroscopy and its description by a correlation function. J. Magn. Reson. 1982;46:385–405. [Google Scholar]
  • 2.Prisner T, Dinse K-P. ESR with Stochastic Excitation. J. Magn. Reson. 1989;84:296–308. [Google Scholar]
  • 3.Pursley RH, Kakareka J, Salem G, Devasahayam N, Subramanian S, Tschudin RG, Krishna MC, Pohida TJ. Stochastic excitation and Hadamard correlation spectroscopy with bandwidth extension in RF FT-EPR. J. Magn. Reson. 2003;162:35–45. doi: 10.1016/s1090-7807(03)00050-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Ernst RR. Magnetic resonance with stochastic excitation. J. Magn. Reson. 1970;3:10–27. [Google Scholar]
  • 5.Blümich B, Gong Q, Byrne E, Greferath M. NMR with excitation by Frank sequences. J. Magn. Reson. 2009;199:18–24. doi: 10.1016/j.jmr.2009.03.009. [DOI] [PubMed] [Google Scholar]
  • 6.Rinard GA, Quine RW, Eaton SS, Eaton GR, Froncisz W. Relative benefits of overcoupled resonators vs. inherently low-Q resonators for pulsed magnetic resonance. J. Magn. Reson. A. 1994;108:71–81. [Google Scholar]
  • 7.Tseitlin M, Quine RW, Eaton SS, Eaton GR, Halpern HJ, Ardenkjaer-Larsen JH. Use of the Frank Sequence in Pulsed EPR. J. Magn. Reson. 2011;209:306–309. doi: 10.1016/j.jmr.2011.01.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Ardenkjaer-Larsen JH, Laursen I, Leunbach I, Ehnholm G, Wistrand L-G, Petersson JS, Golman K. EPR and DNP properties of certain novel single electron contrast agents intended for oximetric imaging. J. Magn. Reson. 1998;133:1–12. doi: 10.1006/jmre.1998.1438. [DOI] [PubMed] [Google Scholar]
  • 9.Turek P, Andre JJ, Giraudeau A, Simon J. Preparation and study of a lithium phthalocyanine radical: optical and magnetic properties. Chem. Phys. Lett. 1987;134:471–476. [Google Scholar]
  • 10.Grinberg VO, Smirnov AI, Grinberg OY, Grinberg SA, O'Hara JA, Swartz HM. Practical conditions and limitations for high spatial resolution of multi-site EPR oximetry. Appl. Magn. Reson. 2005;28:69–78. [Google Scholar]
  • 11.Rinard GA, Quine RW, Eaton SS, Eaton GR. Frequency dependence of EPR signal intensity, 250 MHz to 9.1 GHz. J. Magn. Reson. 2002;156:113–121. doi: 10.1006/jmre.2002.2530. [DOI] [PubMed] [Google Scholar]
  • 12.Rinard GA, Quine RW, Ghim BT, Eaton SS, Eaton GR. Easily tunable crossed-loop (bimodal) EPR resonator. J. Magn. Reson. A. 1996;122:50–57. [Google Scholar]
  • 13.Rinard GA, Quine RW, Eaton GR, Eaton SS. 250 MHz crossed loop resonator for pulsed electron paramagnetic resonance. Magn. Reson. Engineer. 2002;15:37–46. [Google Scholar]
  • 14.Rinard GA, Quine RW, Eaton GR. An L-Band crossed-loop (bimodal) EPR resonator. J. Magn. Reson. 2000;144:85–88. doi: 10.1006/jmre.2000.2014. [DOI] [PubMed] [Google Scholar]
  • 15.Tseitlin M, Eaton SS, Eaton GR. Reconstruction of the first derivative EPR spectrum from multiple harmonics of the field-modulated CW signal. J. Magn. Reson. 2011;209:277–281. doi: 10.1016/j.jmr.2011.01.027. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ahmad R, Som S, Kesselring E, Kuppusamy P, Zweier JL, Potter LC. Digital detection and processing of multiple quadrature harmonics for EPR spectroscopy. J. Magn. Reson. 2010;207:322–331. doi: 10.1016/j.jmr.2010.09.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Tseitlin MP, Tseitlin OA. Using of digital demodulation of multiharmonic overmodulated EPR signals to improve EPR oximetry reliability Appl. Magn. Reson. 2009;36:25–34. [Google Scholar]
  • 18.Hyde JS, Camenisch TG, Ratke JJ, Strangeway RA, Froncisz W. Digital detection by time-locked sampling in EPR. Biol. Magn. Reson. 2005;24:199–222. [Google Scholar]
  • 19.Tseitlin M, Iyudin VS, Tseitlin OA. Advantages of Digital Phase-Sensitive Detection for Upgrading an Obsolete CW EPR Spectrometer. Appl. Magn. Reson. 2009;35:569–580. [Google Scholar]
  • 20.Schweiger A, Jeschke G. Principles of Pulse Electron Paramagnetic Resonance. Oxford University Press; Oxford: 2001. [Google Scholar]
  • 21.Stoner JW, Szymanski D, Eaton SS, Quine RW, Rinard GA, Eaton GR. Direct-detected rapid-scan EPR at 250 MHz. J. Magn. Res. 2004;170:127–135. doi: 10.1016/j.jmr.2004.06.008. [DOI] [PubMed] [Google Scholar]
  • 22.Huisjen J, Hyde JS. A pulsed EPR spectrometer. Rev. Sci. Instrum. 1974;45:669–675. [Google Scholar]
  • 23.Joshi JP, Ballard JR, Rinard GA, Quine RW, Eaton SS, Eaton GR. Rapid-Scan EPR with Triangular Scans and Fourier Deconvolution to Recover the Slow-Scan Spectrum. J. Magn. Reson. 2005;175:44–51. doi: 10.1016/j.jmr.2005.03.013. [DOI] [PubMed] [Google Scholar]

RESOURCES