Effect of inter-reader variability on outcomes in studies using carotid intima media thickness quantified by carotid ultrasonography

Joseph A C Delaney 1,2,, Rebecca Scherzer 3,4, Joseph Polak 5, Mary Lou Biggs 6, Richard Kronmal 7, Haiying Chen 8, Stephen Sidney 9, Carl Grunfeld 10,11
PMCID: PMC3161119  NIHMSID: NIHMS204709  PMID: 20309612

Abstract

Systematic differences between readers or equipment in imaging studies are not uncommon; failure to account for such differences when using Carotid Ultrasonography may introduce bias into associations between carotid intima media thickness (cIMT) and outcomes. We demonstrate the impact of this source of systematic measurement error (SME) using data on 5,521 participants from the Multi-Ethnic Study of Atherosclerosis (MESA) and 661 participants from the Study of Fat Redistribution and Metabolic Change in HIV Infection (FRAM). Participants were between 37 and 78 years old. Two outcomes were considered: (1) the effect of HIV infection on cIMT (between study) and (2) the association of cIMT with cardiovascular events (within study). All estimates were adjusted for demographics (age, gender, and ethnicity) and for traditional cardiovascular disease risk factors (smoking, blood pressure, diabetes and cholesterol). When comparing the FRAM and MESA cohorts to estimate the association of HIV infection on common cIMT, accounting for machine and reader variability (between study variability) reduced the difference associated with HIV infection from +0.080 mm (95% Confidence Interval (CI):0.065–0.095) to +0.037 mm (95% CI:0.003 to 0.072) while internal cIMT declined from +0.254 mm (95% CI:0.205–0.303) to +0.192 mm (95% CI:0.076–0.308). Attenuation of the association between cIMT and cardiovascular endpoints occurred when within study reader variability was not accounted for. The effect of SME due to use of multiple readers or machines is most important when comparisons are made between two different study populations. Within-cohort measurement error dilutes the association with events.

Keywords: Carotid intima media thickness, Measurement error, Bias, Carotid ultrasonography

Introduction

Carotid intima media thickness (cIMT) measured by ultrasonography has been used in many studies to assess factors contributing to atherosclerosis [1]. Therefore, carotid IMT has been used to assess whether HIV infection is associated with increased atherosclerosis [28], but few studies included adequate control groups. While some of those studies found an independent association of HIV infection with carotid lesions [24], others did not [58]. To better determine whether HIV infection is associated with increased pre-clinical atherosclerosis after adjusting for traditional cardiovascular disease (CVD) risk factors, cIMT was measured in the second examination of the Study of Fat Redistribution and Metabolic Change in HIV Infection (FRAM) and the results in 433 HIV-infected subjects were compared to those from a large number (5,749) of similarly aged control subjects from two studies: Coronary Artery Risk Development in Young Adults (FRAM controls) and the Multi-Ethnic Study of Atherosclerosis (MESA). Results from the FRAM study demonstrated that, even after adjusting for traditional CVD risk factors and ultrasound reader, HIV infection remains associated with increased atherosclerosis [9].

There is always some degree of error in any measurement made by a specific reader or laboratory [1013]. Within a single study, it is likely that the patients assigned to any one reader or laboratory will not depend on the patient characteristics. This with-in cohort variability due to reader measurement error will be non-differential and leads to a small attenuation of the association. As a result, most studies of cIMT appropriately decide to ignore within cohort measurement error effects [14].

However, when cohorts are compared (between cohort variability) the effect of reader is unclear as studies will have different participants. As the participants in each study cohort are different, it is no longer reasonable to assume that measurement differences will be unrelated to patient characteristics. The effect of this between cohort variability may be to increase or decrease the magnitude of the association under study. Under these conditions, the possibility for bias due to differential or systematic measurement error exists and the resulting changes in associations may be large. Study validity may, therefore, be more seriously impacted and to ensure that this between-study variability did not impact the validity of the reported results, the FRAM study adjusted for reader effects to remove this important source of potential bias [9].

This paper had two objectives in terms of quantifying the impact of neglecting sources of variation due to differences in both reader and/or ultrasound machine. First, we present data from the FRAM study demonstrating that confounding by reader/machine type may bias the results of a comparison of carotid IMT between cohorts if one does not properly account for between study variability due to systematic measurement error (SME). Therefore, we compared cIMT in the MESA and FRAM cohorts before and after adjustment. Second, we explored the with-in study measurement error in the MESA study, by demonstrating the effect on the association between cIMT and CVD endpoints due to SME introduced when failing to account for reader effects within a single cohort.

Methods

MESA is a population-based, prospective cohort study designed to determine risk factors for the development and progression of subclinical and clinical CVD [9]. The MESA study consisted of 6,814 participants between the ages of 45 and 84 years old at baseline. Participants were recruited from six field centers across the United States: Baltimore, MD; Chicago, IL; Forsyth County, NC; Los Angeles County, CA; New York, NY and St. Paul, MN.

The FRAM study was initially designed to evaluate the prevalence and correlates of changes in fat distribution, insulin resistance, and dyslipidemia in HIV-infected individuals. At the follow-up exam, FRAM also included measurements of cIMT to study sub-clinical CVD. FRAM HIV-infected participants were initially recruited from 16 HIV or infectious disease clinics or cohorts; FRAM control participants were recruited from two centers of the Coronary Artery Risk Development in Young Adults (CARDIA) study. The FRAM HIV-infected and FRAM control participants have been previously discussed in detail elsewhere [9].

The MESA, CARDIA and FRAM studies were approved by the institutional review boards of the participating study sites and the data coordinating center.

For all analyses, age was restricted to 37–78 years old and ethnicity was restricted to Caucasians, Hispanics and African-Americans. This restriction ensured comparability among FRAM HIV-infected subjects, FRAM controls and MESA participants. Participants from the FRAM cohorts with any prior history of CVD (physician-diagnosed heart attack, angina, stroke, TIA, heart failure, cardiac arrest or having undergone procedures related to CVD) were excluded as this was an exclusion criterion for the MESA cohort [9].

Carotid IMT assessment

Technicians, trained by Tufts Medical Center, at each of the 22 MESA or FRAM study sites performed B-mode ultrasonography of the right and left near and far walls of the internal carotid and common carotid arteries. A single ultrasound reading center (Department of Radiology, Tufts Medical Center) measured maximal intima media thickness of the internal (including the bulb) and common carotid sites as the mean of the maximum intima media thickness of the near and far walls of the right and left sides [15].

Ultrasound readers

Measurements of cIMT in the FRAM and MESA cohorts were taken at different time periods, ~2 years apart. There were six different ultrasound image readers for the MESA study, two for the FRAM HIV-infected cohort and three for the FRAM controls. One of the readers was in common between the FRAM controls and the MESA participants. For quality control, one of the readers for the FRAM HIV-infected subjects reread 134 MESA ultrasound images to create overlap between these populations as well. It was therefore possible to directly compare MESA to the FRAM HIV-infected cohort and MESA to the FRAM controls. However, because the FRAM HIV-infected cohort and FRAM controls had no ultrasound readers in common, it was not possible to directly calibrate these two cohorts. Instead, we assumed that the cohorts should yield equal mean cIMT conditional on demographic characteristics, traditional CVD risk factors and reader variation—a generally reasonable assumption.

Ultrasound machines

All MESA readings were made using a single machine (Logiq 700 ultrasound device). However, the FRAM cohort used a total of six different machines (GE Logiq 9, GE Logiq 700, Phillips (ATL) HDI 5000, Acuson Sequoia, HP 7500, Siemens). The acquisition protocol and sonographer certification process were the same for FRAM and MESA with certification by a single author [JP]. As different machines may produce slightly different images, these different machines were controlled for using indicator variables with the GE Logiq 700 as reference. Each site used only one type of machine which was used by all sonographers at that site. Therefore, it was not possible to simultaneously account for site or sonographer effects at the same time as we accounted for machine effects. Therefore, some of the variability associated with machine type could be due to differences in the sonographers using the machines.

MESA endpoints

The MESA cohort was followed for incident CVD events for a median of 4.6 years (max 6.5 y). At intervals of 9–12 months, a telephone interviewer contacted each participant to inquire about all interim hospital admissions, CVD outpatient diagnoses and procedures, and deaths. In addition, MESA occasionally identified additional medical encounters through cohort clinic visits, participant call-ins, medical record abstractions or obituaries. In order to verify self-reported diagnoses, copies of all death certificates and medical records were requested for all hospitalizations and selected outpatient CVD diagnoses and procedures. Next of kin were interviewed for out of hospital CVD deaths. Hospital records were obtained for an estimated 98% of hospitalized CVD events, and at least some information was obtained for 95% of outpatient diagnostic encounters.

Trained personnel abstracted any hospital records that suggested possible CVD events. Abstractors recorded symptoms, history, biomarkers, scanned ECGs, echocardiograms, cathetherization reports, outpatient records, and other relevant diagnostic and procedure reports; and transmitted these to the data coordinating center. The coordinating center collated the abstracted or original endpoint records and sent them to two paired physicians for independent endpoint classification and assignment of incidence dates. Cardiologists or cardiovascular physician epidemiologists reviewed non-neurovascular endpoints; neurologists reviewed all neurovascular endpoints. If the reviewing pair disagreed on classification, they adjudicated differences. If disagreements persisted, the full review committee made the final classification.

A composite endpoint of all CVD was defined to include Myocardial Infarction, Angina, Resuscitated Cardiac Arrest, Stroke (but not Transient Ischemic Attack), Coronary Heart Disease implicated Death, Stroke Death, Other Atherosclerotic Death, or Other Death related to CVD.

Statistical analysis

For the first study objective (comparing MESA and FRAM), we fit a multivariable linear generalized estimating equations regression model (to handle repeated measures of cIMT) in order to estimate the mean difference in cIMT associated with HIV infection. We compared estimates from models excluding covariates for both readers and machines to ones that included covariates for either reader and machine or both reader and machine. Reader and machine effects were included using indicator variables for readers, with the common reader between FRAM HIV infected subjects and MESA participants used as the reference category. Using indicator variables for reader and machine clusters is the simplest form of hierarchical model [16]. The goodness of fit of these models was evaluated using a Likelihood-ratio test comparing differences in log-likelihoods between models with and without reader or machine effects.

For the second objective (MESA only), we performed a time to event analysis (Cox proportional hazards model) to estimate the risk of CVD events by quartile of common and internal cIMT. We then compared models in which the differences due to reader effects are accounted for with indicator variables and those models where they are not. Pseudo-bias was calculated as the difference in estimates between models with and without reader adjustment, divided by the estimate from the model with reader adjustment.

All analyses were conducted using the SAS system, version 9.2 (SAS Institute, Inc., Cary, North Carolina, USA).

Results

The characteristics of the participants are presented in Table 1. FRAM HIV+ participants differed in many important respects from FRAM and MESA controls, as previously reported [9]. Because of these imbalances, a unadjusted differences in mean levels of cIMT should be interpreted with caution and attention should focus on estimates adjusted for differences in demographic factors.

Table 1.

Descriptive statistics [either mean (standard deviation) or percentage] for the multi-ethnic study of atherosclerosis (MESA), the study of fat redistribution and metabolic change in HIV infection (FRAM) and the FRAM controls for all participants with at least one valid carotid intimal medial thickness ultrasound reading

MESA Controls (n = 5,521) FRAM HIV+ (n = 433) FRAM Controls (n = 228)
Number of images read by readera
    Reader # 810 1,253 (22.2%) 0 (0%) 104 (45.6%)
    Reader # 815 2,224 (39.3%) 0 (0%) 0 (0%)
    Reader # 816 216 (3.8%) 0 (0%) 0 (0%)
    Reader # 817 651 (11.5%) 0 (0%) 0 (0%)
    Reader # 818 919 (16.6%) 0 (0%) 0 (0%)
    Reader # 819 236 (4.2%) 0 (0%) 0 (0%)
    Reader # 824 134 (2.4%) 175 (34.7%) 0 (0%)
    Reader # 825 0 (0%) 0 (0%) 73 (32.0%)
    Reader # 827 0 (0%) 0 (0%) 51 (22.4%)
    Reader # 828 0 (0%) 330 (65.3%) 0 (0%)
Cohort characteristics
    Age (years) 61.3 (9.3) 49.0 (7.7) 45.2 (3.7)
    Male 47.2% 60.5% 50.1%
    African American 31.4% 42.3% 43.4%
    Hispanic American 25.1% 7.2% 8.8%
    Known HIV infection 0% 100% 0%
    Current smoker 14.7% 36.3% 11.4%
    Past smoker 38.8% 24.3% 21.5%
    Diabetic 13.7% 8.7% 7.5%
    Systolic blood pressure (mmHG) 125.8 (20.9) 125.6 (17.6) 120.7 (16.5)
    Diastolic blood pressure (mmHG) 72.1 (10.2) 77.7 (10.1) 75.3 (9.8)
    Total cholesterol (mg/dl) 194.6 (36.3) 188.2 (47.8) 198.2 (39.4)
    HDL cholesterol (mg/dl) 51.0 (15.0) 46.5 (16.8) 54.7 (16.1)
    Internal carotid intima media thickness (mm) 1.07 (0.59) 1.17 (0.50) 0.89 (0.20)
    Common carotid intima media thickness (mm) 0.87 (0.19) 0.88 (0.16) 0.80 (0.13)
a

Some images were read more than once so number of readings > number of subjects

In models controlling for both reader and machine, estimates of the mean difference in cIMT associated with HIV infection were 0.037 mm (95% Confidence interval (CI): 0.003–0.072) for common cIMT (Table 2) and 0.192 mm (95% CI: 0.076–0.308) for internal. In models that failed to account for reader or machine, the HIV effect was inflated (pseudo-bias: 116% for common, 32% for internal). Models that controlled for machine but not reader also produced inflated HIV effects (151% for common, 67% for internal). Finally, when reader effects were incorporated but machine was ignored, the HIV effects appeared to be slightly attenuated (-14% for common, -23% for internal) but this attenuation was much less than sampling error.

Table 2.

Adjusted estimates of the mean difference in cIMT associated with HIV infection accounting for the two sources of principal measurement error (machine type and ultrasound reader) among 433 HIV positive participants and 5,749 controls: data from FRAM, MESA and CARDIA

Adjustment for machine or reader? Adjusted mean difference in cIMT associated with HIV infectiona 95% Confidence interval P-value % Pseudo-bias
Common cIMT
    None 0.080 0.065–0.095 <0.0001 +116
    Machine 0.093 0.064–0.122 <0.0001 +151
    Reader 0.032 0.010–0.056 0.005 –14
    Machine and reader 0.037 0.003–0.072 0.033 Reference
Internal cIMT
    None 0.254 0.205–0.303 <0.0001 +32
    Machine 0.321 0.222–0.420 <0.0001 +67
    Reader 0.148 0.072–0.224 0.0001 –23
    Machine and Reader 0.192 0.076–0.308 0.001 Reference
a

Estimates are also adjusted for age, sex, ethnicity, smoking status, diabetes, blood pressure and cholesterol

The addition of machine to models containing reader did not significantly improve goodness of fit for either common (P = 0.8) or internal (P = 0.2) cIMT. In contrast, adding reader to models with only machine in them resulted in a statistically significant improvement for goodness of fit for both common (P < 0.0001) and internal P < 0.0001) cIMT. Clearly adding both machine and reader to the model at once also improved goodness of fit for both common (P < 0.0001) and internal (P < 0.0001) cIMT.

Within the MESA cohort, we considered the association of internal and common cIMT (quartiles) with CVD events (Tables 3, 4) in models excluding and including reader effects (the same machine was used for all MESA readings). In both cases, the association of cIMT with CVD events was weaker in models that excluded reader effects. In common cIMT, the estimate of the effect of being in the fourth quartile (compared with the first quartile) on the rate of all CVD events is increased from Hazard Ratio (HR) 1.62 (95% Confidence Interval (CI): 1.02–2.58) to HR 1.84 (95% CI: 1.12–3.03) which is a modest strengthening of the association due to properly accounting for with-in study measurement error. For internal cIMT, the estimate of the effect of being in the fourth quartile (compared with the first quartile) on the rate of all CVD events has a Hazard Ratio (HR) of 1.97 (95% CI: 1.32–2.95) when adjusted for reader, but the HR was 1.89 (95% Confidence Interval (CI): 1.28–2.80) with no adjustment, which is consistent with the effects seen with common cIMT.

Table 3.

Association between quartiles of common carotid intima-media thickness (cIMT) and the 243 reported “all cardiovascular” (CVD) events in the MESA study after median of 4.6 years (max 6.5 years) of follow-up with adjustment for reader effects

Adjusted hazard ratioa 95% Confidence interval P-value % Pseudo-bias
Not adjusted for reader
        Q1 common cIMT 1.00 Reference NA
        Q2 common cIMT 1.25 0.77–2.03 0.36 5
        Q3 common cIMT 1.28 0.79–2.06 0.32 8
        Q4 common cIMT 1.62 1.02–2.58 0.04 12
Adjusted for reader
        Q1 common cIMT 1.00 Reference NA
        Q2 common cIMT 1.31 0.80–2.13 0.28 Reference
        Q3 common cIMT 1.39 0.85–2.27 0.19 Reference
        Q4 common cIMT 1.84 1.12–3.03 0.02 Reference

Quartiles are cut at 1.278, 0.860 and 0.689 mm. Data from the Multi-Ethnic Study of Atherosclerosis (restricted to 5,521 subjects in the FRAM comparison population)

a

Estimates are also adjusted for age, sex, ethnicity, smoking status, diabetes, blood pressure and cholesterol. Estimates are not adjusted for machine, as only one machine type was used in MESA

Table 4.

Association between quartiles of internal carotid intima-media thickness (cIMT) and the 243 reported “all cardiovascular” (CVD) events in the MESA study after median of 4.6 years (max 6.5 years) of follow-up with adjustment for reader effects

Adjusted hazard ratioa 95% Confidence interval P-value % Pseudo-bias
Not adjusted for reader
        Q1 internal cIMT 1.00 Reference NA
        Q2 internal cIMT 1.03 0.65–1.63 0.92 1
        Q3 internal cIMT 1.10 0.71–1.69 0.68 4
        Q4 internal cIMT 1.89 1.28–2.80 0.001 4
Adjusted for reader
        Q1 internal cIMT 1.00 Reference NA
        Q2 internal cIMT 1.04 0.65–1.69 0.88 Reference
        Q3 internal cIMT 1.14 0.72–1.79 0.58 Reference
        Q4 internal cIMT 1.97 1.32–2.95 0.001 Reference

Quartiles are cut at 1.278, 0.860 and 0.689 mm. Data from the multi-ethnic study of atherosclerosis (restricted to 5,521 subjects in the FRAM comparison population)

a

Estimates are also adjusted for age, sex, ethnicity, smoking status, diabetes, blood pressure and cholesterol. Estimates are not adjusted for machine, as only one machine type was used in MESA

Discussion

Measurement error (either from general variability or due to systematic differences between groups) can introduce an important amount of bias into estimates, even under ideal conditions. In this paper we demonstrated the bias introduced by the use of multiple readers or machines for studies using carotid ultrasonography. The effect of systematic measurement error due to use of multiple readers or machines is most important when comparisons are made between two different study populations that were measured separately. However, we have shown that even within a single cohort, the failure to account for measurement error resulted in a change in the estimate of the risk of serious CVD events associated with increased levels of cIMT.

Pooling separate study cohorts may be unavoidable in several different types of studies. One possibility, as seen in the first example presented here from the FRAM Study, is the need to estimate the difference in cIMT between cohorts of differing characterizes. In such a case, comparisons with existing cohort studies are an attractive alternative to recruiting a new control series in parallel, due to issues of cost and feasibility. The failure to account for reader and machine differences between cohorts could create potential issues in the pooling of these studies.

We also show here that failure to account for reader measurement error with-in a single cohort on measurement of cIMT can distort the association between cIMT and important endpoints. Typically, within-cohort measurement error is not associated with participant characteristics, but dilutes associations. While the effect sizes seen in this example are modest, they do rise to the level of practical significance for common cIMT where the difference in estimates (pseudo-bias) is greater than 20%. This could make the measure less or more predictive when compared to other candidate measures of sub-clinical atherosclerosis [14].

Our results are compatible with Espeland et al. who found that reader effects made an 11% contribution to variations in cIMT measurements in the ACAPS study [12]. They suggested that the true correlation of cIMT with risk factors is about twice what is commonly observed [12] but that this association is diluted by uncontrolled measurement error when reader effects are not accounted for. Controlling for reader effects in the analysis of any cIMT endpoint should improve estimates of association between cIMT and predictors as we are reducing one source of non-differential misclassification.

Measurements of cIMT in the FRAM and MESA cohorts were taken at different time periods, ~2 years apart. Temporal bias has been observed in previous studies, and the approach used here (having ultrasounds reread) has previously been used to correct these issues [17]. Temporal drift may explain why reader effects seen in this study are larger than those seen in studies such as ELSA (European Lacidipine Study on Atherosclerosis) where ultrasound scans were all read contemporaneously by multiple readers [8, 18].

In this study, the large differences observed in residual reader and machine effects demonstrate that pooling measurements from different readers and machines will result in misclassification of participants. That this type of systematic measurement error can lead to a modest dilution of the estimates of the association between cIMT and CVD endpoints is a known phenomenon [1921], but often considered of little practical importance. However, the sheer magnitude of the effects seen when comparing two cohorts suggests that extra care should be taken in studies of cIMT when comparing cohort differences is the primary objective. Previous studies have often failed to account for reader effects [22] which can create vulnerability to high levels of systematic measurement error if, by bad luck, the readers in the two studies have very different mean estimates of cIMT. In the current study, a failure to account for reader effects would have dramatically overestimated the effect of HIV infection on cIMT due to the pattern of reader differences between the FRAM and MESA studies.

It is unclear if it is appropriate to adjust for machine type in this type of analysis and our previous FRAM comparative study only adjusted for reader [9]. Other studies have used careful protocols that avoid this issue entirely [23]. Unlike reader effects, which are strong and for which we have rereads to assess our model validity, machine effects are weaker and need to be estimated from data. However, there have been studies in which adjustment for a large number of machines was essential to seeing a clear result. For example, the EDIC study used 12 different ultrasound machines at 28 different sites and found it important to adjust for machine type when estimating progression of cIMT [24]. In the FRAM population, the effect of machine type is small and the benefits of including machine adjustments in the analysis are dubious. Not only is the effect small, but the MESA cohort used a single machine. This raises the concern that machine variability could be acting as a marker for HIV infection—a concern that could only be directly addressed by cross-calibration (which is not typically feasible). Therefore, the more conservative approach to presenting between cohort estimates of the HIV effect would adjust for reader but not machine type [9].

When modeling the association between cIMT and candidate risk factors, it is important to account for differences in readers as a form of measurement error. Current practice in studies that use IMT may be very suboptimal in terms of efficiently mitigating the reader error problem. Failing to account for measurement error due to reader or machine differences can result in important levels of bias and either a common reader or the use of rereads is essential to the calibration of differences between readers when comparing cohorts.

Acknowledgments

This research was supported by contracts N01-HC-95159 through N01-HC-95165 and N01-HC-95169 from the National Heart, Lung, and Blood Institute and by NIH grants RO1-DK57508, RO1-HL74814, RO1-HL 53359, M01-RR00036, M01-RR00051, M01-RR00052, M01-RR00054, M01-RR00083, M01-RR0636, M01-RR00865 and UL1-RR024131. The authors thank the other investigators, the staff, and the participants of the MESA study for their valuable contributions. A full list of participating MESA investigators and institutions can be found at http://www.mesa-nhlbi.org. The funding agency had no role in the collection or analysis of the data.

Contributor Information

Joseph A. C. Delaney, Collaborative Health Studies Coordinating Center, Department of Biostatistics, University of Washington, Bldg.29, Suite 310, 6200 NE 74th Street, Seattle, WA 98115, USA College of Pharmacy, University of Florida, Gainesville, FL 32610, USA.

Rebecca Scherzer, University of California, San Francisco, CA, USA; Medical Service, Department of Veterans Affairs Medical Center, San Francisco, CA, USA.

Joseph Polak, Tufts Medical Center, Boston, MA, USA.

Mary Lou Biggs, Collaborative Health Studies Coordinating Center, Department of Biostatistics, University of Washington, Bldg.29, Suite 310, 6200 NE 74th Street, Seattle, WA 98115, USA.

Richard Kronmal, Collaborative Health Studies Coordinating Center, Department of Biostatistics, University of Washington, Bldg.29, Suite 310, 6200 NE 74th Street, Seattle, WA 98115, USA.

Haiying Chen, Wake Forest University School of Medicine, Winston-Salem, NC, USA.

Stephen Sidney, Division of Research, Kaiser Permanente, Oakland, CA, USA.

Carl Grunfeld, University of California, San Francisco, CA, USA; Medical Service, Department of Veterans Affairs Medical Center, San Francisco, CA, USA.

References

  • 1.O'Leary DH, Polak JF. Intima-media thickness: a tool for atherosclerosis imaging and event prediction. Am J Cardiol. 2002;90(10C):18L–21L. doi: 10.1016/s0002-9149(02)02957-0. [DOI] [PubMed] [Google Scholar]
  • 2.Hsue PY, Squires K, Bolger AF, Capili B, Mensah GA, Temesgen Z, et al. Screening and assessment of coronary heart disease in HIV-infected patients. Circulation. 2008;118(2):e41–7. doi: 10.1161/CIRCULATIONAHA.107.189626. [DOI] [PubMed] [Google Scholar]
  • 3.Depairon M, Chessex S, Sudre P, Rodondi N, Doser N, Chave JP, et al. Premature atherosclerosis in HIV-infected individuals–focus on protease inhibitor therapy. AIDS. 2001;15(3):329–34. doi: 10.1097/00002030-200102160-00005. [DOI] [PubMed] [Google Scholar]
  • 4.Hsue PY, Lo JC, Franklin A, Bolger AF, Martin JN, Deeks SG, et al. Progression of atherosclerosis as assessed by carotid intima-media thickness in patients with HIV infection. Circulation. 2004;109(13):1603–8. doi: 10.1161/01.CIR.0000124480.32233.8A. [DOI] [PubMed] [Google Scholar]
  • 5.Currier JS, Kendall MA, Zackin R, Henry WK, Alston-Smith B, Torriani FJ, et al. Carotid artery intima-media thickness and HIV infection: traditional risk factors overshadow impact of protease inhibitor exposure. AIDS. 2005;19(9):927–33. doi: 10.1097/01.aids.0000171406.53737.f9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Currier JS, Kendall MA, Henry WK, Alston-Smith B, Torriani FJ, Tebas P, et al. Progression of carotid artery intima-media thickening in HIV-infected and uninfected adults. AIDS. 2007;21(9):1137–45. doi: 10.1097/QAD.0b013e32811ebf79. [DOI] [PubMed] [Google Scholar]
  • 7.Johnsen S, Dolan SE, Fitch KV, Kanter JR, Hemphill LC, Connelly JM, et al. Carotid intimal medial thickness in human immunodeficiency virus-infected women: effects of protease inhibitor use, cardiac risk factors, and the metabolic syndrome. J Clin Endocrinol Metab. 2006;91(12):4916–24. doi: 10.1210/jc.2006-1140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Tang R, Hennig M, Bond MG, Hollweck R, Mancia G, Zanchetti A, et al. Quality control of B-mode ultrasonic measurement of carotid artery intima-media thickness: the European Lacidipine Study on Atherosclerosis. J Hypertens. 2005;23(5):1047–54. doi: 10.1097/01.hjh.0000166846.87968.85. [DOI] [PubMed] [Google Scholar]
  • 9.Grunfeld C, Delaney JAC, Wanke C, Currier J, Scherzer R, Biggs ML, Shlipak M, Polak J, O'Leary D, Sidney S, Bacchetti P, Kronmal R. Pre-clinical atherosclerosis due to HIV infection: carotid intima-medial thickness measurements from the study of fat redistribution and metabolic change in HIV infection (FRAM). AIDS. 2009;23(14):1841–9. doi: 10.1097/QAD.0b013e32832d3b85. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.O'Leary DH, Bryan FA, Goodison MW, Rifkin MD, Gramiak R, Ball M, et al. Measurement variability of carotid atherosclerosis: real-time (B-mode) ultrasonography and angiography. Stroke. 1987;18(6):1011–7. doi: 10.1161/01.str.18.6.1011. [DOI] [PubMed] [Google Scholar]
  • 11.Furberg CD, Byington RP, Craven TE. Lessons learned from clinical trials with ultrasound end-points. J Intern Med. 1994;236:575–80. doi: 10.1111/j.1365-2796.1994.tb00848.x. [DOI] [PubMed] [Google Scholar]
  • 12.Espeland MA, Craven TE, Riley WA, Corson J, Romont A, Furberg CD. Reliability of longitudinal ultrasonographic measurements of carotid intimal-medial thicknesses. Asymptomatic Carotid Artery Progression Study Research Group. Stroke. 1996;27(3):480–5. doi: 10.1161/01.str.27.3.480. [DOI] [PubMed] [Google Scholar]
  • 13.Kaplan RC, Kingsley LA, Gange SJ, Benning L, Jacobson LP, Lazar J, et al. Low CD4 + T-cell count as a major atherosclerosis risk factor in HIV-infected women and men. AIDS. 2008;22(13):1615–24. doi: 10.1097/QAD.0b013e328300581d. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Folsom AR, Kronmal RA, Detrano RC, O'Leary DH, Bild DE, Bluemke DA, et al. Coronary artery calcification compared with carotid intima-media thickness in the prediction of cardiovascular disease incidence: the multi-ethnic study of atherosclerosis (MESA). Arch Intern Med. 2008;168(12):1333–9. doi: 10.1001/archinte.168.12.1333. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.O'Leary DH, Polak JF, Wolfson SK, Jr, Bond MG, Bommer W, Sheth S, et al. Use of sonography to evaluate carotid atherosclerosis in the elderly: the cardiovascular health study. Stroke. 1991;22:1155–63. doi: 10.1161/01.str.22.9.1155. [DOI] [PubMed] [Google Scholar]
  • 16.Gelman A, Hill J. Data analysis using regression and multilevel/hierarchical models. Cambridge University Press; New York: 2007. [Google Scholar]
  • 17.Riley WA, Craven T, Romont A, Furberg CD. Assessment of temporal bias in longitudinal measurements of carotid intimal-medial thickness in the asymptomatic carotid artery progression study (ACAPS). ACAPS Research Group. Ultrasound Med Biol. 1996;22(4):405–11. doi: 10.1016/0301-5629(96)00027-0. [DOI] [PubMed] [Google Scholar]
  • 18.Tang R, Hennig M, Thomasson B, Scherz R, Ravinetto R, Catalini R, et al. Baseline reproducibility of B-mode ultrasonic measurement of carotid artery intima-media thickness: the European Lacidipine Study on Atherosclerosis (ELSA). J Hypertens. 2000;18(2):197–201. doi: 10.1097/00004872-200018020-00010. [DOI] [PubMed] [Google Scholar]
  • 19.Strachan D, Rose G. Strategies of prevention revisited: effects of imprecise measurement of risk factors on the evaluation of “high-risk” and “population-based” approaches to prevention of cardiovascular disease. J Clin Epidemiol. 1991;44(11):1187–96. doi: 10.1016/0895-4356(91)90151-x. [DOI] [PubMed] [Google Scholar]
  • 20.Nakamura T. Proportional hazards model with covariates subject to measurement error. Biometrics. 1992;48(3):829–38. [PubMed] [Google Scholar]
  • 21.Prentice RL. Covariate measurement errors and parameter estimation in a failure time regression model. Biometrika. 1982;69:331–42. [Google Scholar]
  • 22.Howard G, Manolio TA, Burke GL, Wolfson SK, O'Leary DH. Does the association of risk factors and atherosclerosis change with age? An analysis of the combined ARIC and CHS cohorts. The Atherosclerosis Risk in Communities (ARIC) and Cardiovascular Health Study (CHS) Investigators. Stroke. 1997;28(9):1693–701. doi: 10.1161/01.str.28.9.1693. [DOI] [PubMed] [Google Scholar]
  • 23.Touboul PJ, Vicaut E, Labreuche J, Belliard JP, Cohen S, Kownator S, et al. Design, baseline characteristics and carotid intima-media thickness reproducibility in the PARC study. Cerebrovasc Dis. 2005;19(1):57–63. doi: 10.1159/000081913. [DOI] [PubMed] [Google Scholar]
  • 24.Nathan DM, Lachin J, Cleary P, Orchard T, Brillon DJ, Backlund JY, et al. Intensive diabetes therapy and carotid intima-media thickness in type 1 diabetes mellitus. N Engl J Med. 2003;348(23):2294–303. doi: 10.1056/NEJMoa022314. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES