Abstract
BACKGROUND
Oral fluid (OF) is an exciting alternative matrix for monitoring drugs of abuse in workplace, clinical toxicology, criminal justice, and driving under the influence of drugs (DUID) programs. During the last 5 years, scientific and technological advances in OF collection, point-of-collection testing devices, and screening and confirmation methods were achieved. Guidelines were proposed for workplace OF testing by the Substance Abuse and Mental Health Services Administration, DUID testing by the European Union’s Driving under the Influence of Drugs, Alcohol and Medicines (DRUID) program, and standardization of DUID research. Although OF testing is now commonplace in many monitoring programs, the greatest current limitation is the scarcity of controlled drug administration studies available to guide interpretation.
CONTENT
This review outlines OF testing advantages and limitations, and the progress in OF that has occurred during the last 5 years in collection, screening, confirmation, and interpretation of cannabinoids, opioids, amphetamines, cocaine, and benzodiazepines. We examine controlled drug administration studies, immunoassay and chromatographic methods, collection devices, point-of-collection testing device performance, and recent applications of OF testing.
SUMMARY
Substance Abuse and Mental Health Services Administration approval of OF testing was delayed because questions about drug OF disposition were not yet resolved, and collection device performance and testing assays required improvement. Here, we document the many advances achieved in the use of OF. Additional research is needed to identify new bio-markers, determine drug detection windows, characterize OF adulteration techniques, and evaluate analyte stability. Nevertheless, there is no doubt that OF offers multiple advantages as an alternative matrix for drug monitoring and has an important role in DUID, treatment, workplace, and criminal justice programs.
Oral fluid (OF)3 is a suitable alternative matrix to test drugs of abuse in clinical, workplace, driving under the influence of drugs (DUID), drug treatment, and criminal justice settings. The main advantages of OF are the simplicity and noninvasiveness of sample collection, which can be easily observed (1), obviating the need for special restroom facilities and same-sex collectors and making adulteration more difficult (2). Infection risk is lower than for blood (3), and OF may better reflect recent drug use. The parent drug is frequently prominent in OF and may reflect free plasma concentrations, providing a better correlation with pharmacodynamic effects, such as impaired performance (4). It is difficult to differentiate heroin from morphine or codeine ingestion with urine drug testing, whereas 6-acetylmorphine (6AM) and heroin are frequently present in OF, clearly delineating heroin abuse.
Weak bases are detected in higher concentrations and for longer times in OF than in plasma because of ion trapping. Depending on a drug’s pKa and lipophilicity, ion trapping occurs because of pH differences in blood (7.4) and OF (4–6). Free uncharged drug is in equilibrium across membranes separating blood and OF. At the lower OF pH, weak bases ionize, increasing total OF drug concentrations.
For clinical toxicology applications, including drug treatment, physician office, and emergency room testing, onsite OF testing offers rapid availability of results for diagnostic purposes and ability to confront treatment patients with immediate indicators of drug relapse. In 2004, Substance Abuse and Mental Health Services Administration (SAMHSA) proposed recommended guidelines (Table 1) for mandated federal workplace OF testing (5). Lack of resolution of important scientific questions delayed final approval, but SAMHSA guidelines, including low cutoff concentrations, are frequently applied in nonregulated settings. The guidelines encompass mandated cutoffs and procedures for sample collection, custody, and control to ensure sample identity and integrity; sample validity tests; testing facility requirements; analytical method result review and reporting; alternative medical explanations; and laboratory certification (6). In the US, OF testing is expanding at a rapid pace in nonregulated workplace testing, treatment, and driving under the influence of drugs (DUID) programs.
Table 1.
Drug/analyte | SAMHSA screen, μg/L | SAMHSA confirmation, μg/L | DRUID confirmation, μg/L | Talloires confirmation, μg/L |
---|---|---|---|---|
Cannabinoids | 4 | 2b | 1b | 2 |
Opiates | 40 | |||
Morphine | 40 | 20 | 20 | |
Codeine | 40 | 20 | 20 | |
6AM | 4 | 4 | 5 | 5 |
Methadone | — | 20 | 20 | |
Phencyclidine | 10 | 10 | — | — |
Amphetamines | 50 | |||
Amphetamine | 50 | 25 | 20 | |
Methamphetamine | 50c | 25 | 20 | |
MDMA | 50 | 50 | 25 | 20 |
MDA | 50 | 25 | 20 | |
MDEA | 50 | 25 | 20 | |
Cocaine or benzoylecgonine | 20 | 8 | 10 | 10 |
Benzodiazepines | — | |||
Flunitrazepam | — | 1 | — | |
Diazepam | — | 5 | — | |
Alprazolam | — | 1 | — | |
Oxazepam | — | 5 | — | |
Nordiazepam | — | 1 | — | |
Lorazepam | — | 1 | — | |
Clonazepam | — | 1 | — |
One major incentive for improving OF testing is the increasing problem of DUID. Drugs and alcohol were contributing factors in up to 22% of motor vehicle crashes in the US (7). Unfortunately, an even higher prevalence is likely, because additional analyses are usually not conducted for drugs if alcohol is above the legal limit. OF testing offers a new tool to improve traffic safety with rapid, easy roadside drug testing of drivers. ROadSIde Testing Assessment (ROSITA) was the first European Union effort to determine the best means of identifying drugged drivers at the roadside. OF was selected over urine or sweat as the best matrix, although OF collection and testing was then considered inadequate (8). ROSITA-2 in Europe and the US specifically evaluated available OF point-of-collection testing (POCT) devices, concluding that none were yet adequately reliable (9). The latest European initiative, Driving Under the Influence of Drugs, Alcohol, and Medicines (DRUID), promotes research and scientific support to reduce European Union road deaths by 50%. Guidelines were proposed for DUID testing in the DRUID program, as shown in Table 1. In Victoria, Australia, OF DUID testing began in 2003, with public acceptance and reductions in impaired driving, despite less than optimal performance of collection devices and assays (10). The Talloires recommendations for OF cutoff concentrations (11) also are included in Table 1. In 2006, international experts met to draft guidelines for conducting research on drugged driving. Currently, it is difficult to compare results across studies owing to the lack of standardization in experimental methods. Recommendations focused on 3 broad areas, i.e., behavior, epidemiology, and toxicology. Included among the 80 recommendations for toxicological contributions were suggested OF drug cutoff concentrations.
OF testing does have disadvantages. Drugs may reduce salivation, limiting sample volume and necessitating sensitive analytical methods to quantify multiple analytes in 1 assay. Drug concentrations are frequently lower in OF than in urine (4). Other issues include contamination from ingested food and beverages, and unknown adulteration techniques (12). Drugs that are smoked, inhaled, insufflated, or taken orally also may contaminate the oral mucosa and OF, increasing detection, but reducing correlation with blood concentrations for 30 – 60 min. We observed that OF concentrations of Δ9-tetrahydrocannabinol (THC) greatly exceeded plasma concentrations for approximately 30 min after cannabis smoking (13). The magnitude and duration of contamination have not been clearly defined for many drugs, and require additional controlled drug administration studies. Stimulation of salivary flow leads to increased OF pH due to increased bicarbonate excretion and reduced drug concentrations due to dilution (12).
In this review, we focus on advances, primarily those occurring in the last 5 years, in OF measurement and interpretation of OF concentrations of cannabinoids, opioids, amphetamines, cocaine, and benzodiazepines. We also summarize progress in OF testing and unresolved issues. Our aim is to inform laboratories adding OF collection and testing, improve interpretation of OF test results, and promote evidence-based drug policies and relevant traffic safety legislation.
OF Collection
There are 2 main approaches to OF testing. One approach involves OF collection by expectoration or with a specialized device, and transfer of the sample for conventional laboratory-based immunoassay screening and chromatographic confirmation. The other approach is a system for collection and initial screening of OF at the site of collection (POCT), followed by a laboratory-based chromatographic confirmation at a later time. Collecting OF without a specialized device can be achieved by the passive drool technique or by expectoration. The passive drool technique best reflects drug concentrations in excreted saliva, because expectoration increases the rate of salivary excretion to a minor extent. Although samples collected without a specialized device are useful from a scientific perspective, the collection process is distasteful for donors and collectors. Thus OF collection devices are now used for large-scale testing applications. As interest grew in testing this alternative matrix, a wide variety of OF collection devices were developed, with variable success. Major problems are variability of OF collected and deficiency of sample amounts.
Many devices collect <1 mL OF, limiting available sample for multiple drug confirmations. However, as chromatographic methods and instruments improve in analytical sensitivity, smaller sample volumes are being required. In addition, the trend is to develop assays that quantify multiple drug classes simultaneously, reducing the needed sample size. Another factor affecting adequate sample volume is the expected OF drug concentrations. Depending on a drug’s potency and physiochemical characteristics, concentrations range from nanograms per liter to micrograms per liter. Therefore, the determination of adequate sample volume must take into consideration the drug being tested.
Some OF collection devices have built-in volume-adequacy indicators, facilitating appropriate collection, whereas other manufacturers report only approximate amounts collected. The Arco Biotech device has a mark on the vial indicating the amount collected (3), and the Immunalysis Quantisal™ and StatSure Saliva Sampler™ contain indicators for collection of 1 mL. With the use of these indicators, the collector can keep the device in the donor’s mouth for the time required to collect an adequate volume, despite individual variability in OF excretion. A similar indicator on the Cozart® Drug Detection System (DDS) documents 340 (60) μL OF (14). Another approach is gravimetric determination by weighing the device before and after OF collection (15, 16). With the Greiner device a novel tactic is employed that determines OF amount based on dilution of a dye in the extraction solution (3). The Greiner device differs from other collection devices in that the sample donor rinses the mouth with the extraction solution, which is then expectorated with the OF into a collection beaker. The absorbance of the resulting solution is read in a spectrophotometer to calculate collected OF amount.
Even more problematic than the amount collected is variability in OF volume obtained within and between devices. The Orasure Intercept® Drugs Of Abuse oral sample collection device was reported to collect from 0.38–1.53 g (16). A recent evaluation of 3 collection devices provided within and between collection device variability ranging from 1.045–1.667 g (15). This variability reflects imprecision in elution buffer volume included in the device, and more importantly, inconsistency in donor OF amount. Obviously, drug concentration is dependent upon the degree of dilution. Some manufacturers now provide assurances of the degree of variability in OF collection; within-device variability of <10% for the Quantisal collection device and <5% for the Cozart and StatSure devices were reported (3).
OF drug concentration also is dependent on OF excretion stimulation, which may occur even to a small extent by placement of a collector in the mouth. Thus, it is impossible to prevent an increase in OF excretion during collection, except by employing passive drool. Older methods of increasing OF excretion included chewing on paraffin, and newer approaches embed citric acid or other chemicals onto collection devices. Early studies, including those in our laboratory, used devices that stimulated OF excretion; however, we learned that stimulation ultimately lowered rather than increased drug concentrations and complicated interpretation of results.
One of the most important limitations of OF collection devices is adsorption of drugs to the device, which frequently leads to false-negative test results (2). This effect was unknown early in OF test development, leading to low sensitivity. Manufacturers addressed the problem, with mixed results, by eluting drugs off the device with buffers. Buffers also reduce OF viscosity, improving measurement accuracy, but also dilute analyte concentrations. Many buffers and surfactants also interfere with direct injection techniques for LC-MS methods by increasing matrix effect. Generally, the more lipophilic a drug, the greater the adsorptivity to the collection device, but recoveries must be empirically determined. Device materials and buffers are proprietary, and it is not possible to predict when recovery might be an issue.
As is true for all toxicological analyses, the analysis is only as good as the sample. Specific challenges for OF collection devices include collection of adequate volume with good imprecision; although readily achievable, as demonstrated with some currently available devices, the speed of collection also is important. For DUID testing, one of the major challenges for manufacturers, and the source of complaints by police, is the time required for collection and testing at the roadside. Although elution buffers appear necessary for effective removal of drugs from the collection device, the resulting drug dilution and presence of salts that must be removed before LC-MS analysis are disadvantages. Although variability is inherent in OF collection and contributes to the difficulty in interpreting OF drug concentrations, is it appropriate to hold OF testing to a higher standard than is achieved with the currently accepted urine testing technology?
Laboratory-Based Drug Screening
The importance of screening OF for parent drugs and metabolites has been clearly documented by controlled drug administration data. With most RIAs for commonly abused drugs removed from the market, new ELISA assays became available for screening drugs in blood; many were adapted for OF testing. Adaptation of blood rather than urine assays is preferable in many cases because of the presence of greater proportions of parent drug in blood and OF. Manufacturers that simply tried to adapt urine assays that primarily target metabolites had difficulty in meeting the required detection requirements, especially for cannabinoids. However, pH differences in blood and OF yield distinct biomarker disposition. For example, benzoylecgonine (BE) concentrations are higher than cocaine in blood, whereas the reverse is true in OF.
Another important consideration in the use of OF is expected drug concentrations. In general, drug concentrations in OF are much lower than in urine, although oral mucosal contamination for 30–45 min after smoking a drug can produce microgram per liter concentrations. Oral or sublingual administration also may contaminate the oral mucosa, although we observed little increase in OF concentrations after sample donors had ingested encapsulated or coated pills. Thus, challenges include achieving low limits of detection to extend the window of drug detection and expanding the linear range or validating dilution procedures to account for high concentrations found after smoked or oral doses.
In addition, it is essential that calibrators and controls be prepared in the same matrix as authentic samples. Matrix effects are important considerations, especially with LC-MS, and require documentation that analyte recovery and assay imprecision are equivalent for calibrator, QC, and authentic samples. Many manufacturers prepare calibrators and QC in synthetic OF; laboratories should validate performance against calibrators prepared in human OF.
Most OF assays are nonhomogeneous, presenting challenges for large-scale automation. To meet the high workplace testing demands in the US, rapid throughput and automation have enabled the achievement of low costs per test for urine analysis. Fully automated procedures, including pipetting, incubation, washing, and absorbance measurement, are available for 96-well ELISA plates. Evaluations of drug immunoassays are included in Table 2.
Table 2.
Method | Type | Confirm | Cutoffs, μg/L
|
SN, % | SP, % | EFF, % | Reference | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cannabinoids
|
Opiates
|
AMPa |
COC
|
Benzodiazepines
|
||||||||||||
Screen | Confirm | Screen | Confirm | Screen | Confirm | Screen | Confirm | Screen | Confirm | |||||||
OraSure Intercept ELISA | — | GC-MS/MS LC-MS/MS |
THC 1 | — | Methadone 5 Opiates 10 |
— | MAMP 40; AMP 100 | — | BE 5 | — | nDZP 1 | — | 50–100 | 99–100 | — | Clarke and Wilson (76) |
| ||||||||||||||||
Roche KIMS | — | — | Opiates 40 | — | MAMP & AMP 50 | — | — | — | — | Ananias et al. (77) | ||||||
| ||||||||||||||||
Cozart Microplate ELISA | E | GC-MS | Opiates 2.5, 10, 20, 30, 40 | COD, nCOD, MOR, nMOR 2.5, 20, 30, 40 | 73.4–91.5 | 88.6–99.2 | 89.3–95.6 | Barnes et al. (78) | ||||||||
| ||||||||||||||||
Cozart RapiScan | E | GC-MS | Opiates 30 | COD, nCOD, MOR, nMOR 2.5, 15, 30, 40 | 56.1–99.0 | 96.2–99.9 | 85.7–98.2 | Kacinko et al. (49) | ||||||||
| ||||||||||||||||
Cozart Microplate ELISA | T | GC-MS | Opiates 30 | Opiates 30 | 99.1 | 94.4 | 96.8 | Cooper et al. (22) | ||||||||
| ||||||||||||||||
Cozart Microplate ELISA | T | GC-MS | Methadone 20 & 30 | Methadone 30 | 91.3–92.2 | 97.9–100 | 94.9–95.5 | Cooper et al. (79) | ||||||||
| ||||||||||||||||
IDS BUP One-step ELISA | T | GC-MS | BUP 0.5 | — | 100 | 100 | 100 | De Giovanni et al. (47) | ||||||||
| ||||||||||||||||
Cozart Microplate ELISA | E | LC-MS/MS | MDMA & MDA 51 | MDMA & MDA 10 | 98.6 | 98.6 | 98.6 | Laloup et al. (53) | ||||||||
| ||||||||||||||||
Cozart Microplate ELISA | F | GC-MS | AMP 45 | AMP 30 | 87.2 | 97.8 | 93.5 | Cooper et al. (24) | ||||||||
| ||||||||||||||||
Bio-Quant Direct ELISA | F | MAMP & AMP ≤ 6 (1 in PBS) | — | — | — | — | Apollonio et al. (80) | |||||||||
| ||||||||||||||||
Cozart Microplate ELISA | E | GC-MS | BE 10, 20, 30 | COC, BE, EME 2.5, 8, 10, 15 | 89.8–99.0 | 58.8–89.6 | 80.0–89.7 | Kim et al. (81) | ||||||||
| ||||||||||||||||
Cozart RapiScan | E | GC-MS | BE 20, 30 | COC, BE, EME 2.5, 8, 15, 30 | 66.5–95.5 | 82.7–98.1 | 74.1–92.0 | Kolbrich et al. (82) | ||||||||
| ||||||||||||||||
Cozart Microplate ELISA | T | GC-MS | BE 20 | COC, BE, CE 30 | 99.0 | 100 | 99.5 | Cooper et al. (23) | ||||||||
| ||||||||||||||||
EMIT II Plus | D | LC-MS/MS | LMP 10 | NZP 0.1; APZ, LZP 1.2; LMP 1.3; MZL 2.3; DZP 6.9; nDZP 7.6; TZP 8.1; OZP 13.1 | 75 | 96.9 | 96.0 | Smink et al. (83) |
AMP, amphetamine; COC, cocaine; MAMP, methamphetamine; nDZP, nordiazepam; E, experimental; COD, codeine; nCOD, norcodeine; MOR, morphine; nMOR, normorphine; T, treatment; BUP, buprenorphine; F, fortified; CE, cocaethylene; LMP, lormetazepam; NZP, nitrazepam; APZ, alprazolam; LZP, lorazepam; MZL, midazolam; DZP, diazepam; TZP, temazepam; OZP, oxazepam.
POCT
The promise of worldwide OF testing spurred commercial research and development of POCT devices, and commercial devices were rushed to market before much of the basic science of drug excretion into OF was known. A POCT device includes OF collection and a built-in system for screening multiple drug classes. A large number of POCT devices have been evaluated (Table 3), although many are no longer available or were substantially modified (9, 12). The major problems with early generation OF POCT included inadequate limits of detection, specificity, and efficiency for identifying cannabinoids, amphetamines, and benzodiazepines; poor performance in bad weather; difficult-to-read results; complicated testing procedures; insufficient sample volume and prevalent device failures (9). Many deficiencies were revealed, leading to additional research and modifications, which led to improvement of available products. The ROSITA-2 project evaluated the performance of multiple POCT devices and set acceptance criteria for diagnostic sensitivity, diagnostic specificity, positive and negative predictive values of ≥90%, and efficiency of ≥95% (9).
Table 3.
Type | THC | OPa | AMP | COC | BENZO | Confirmation | SN (%) | SP (%) | EFF (%) | Recovery (%) | Failures (%) | Reference | |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
POCT devices | |||||||||||||
American Bio Medica OralStat | F | x | x | x | x | x | HPLC GC-MS | 0–100 | 100 | 71.4–100 | — | Walsh et al. (20) | |
Biomar Toxiquick | D | x | x | x | x | GC-MSb HPLC-DAD | 42.1–100 | 63.9–88.7 | 64.3–86.1 | — | Biermann et al. (19) | ||
Branan Oratect | F | x | x | x | x | GC-MS or LC-MS/MS | 0–100 | 33.3–100 | 71.4–100 | — | Crouch et al. (84) | ||
Cozart DDSV | T | x | GC-MS | 44.4 | 100 | 60 | — | 0 | Kintz et al. (14) | ||||
Cozart RapiScan | T | x | x | GC-MS | Opiates 96.7 COC 98.2 | 100 | Opiates 98.0 COC 99.0 | — | Cooper et al. (21) | ||||
Cozart RapiScan | E | x | ELISA & GC-MS | ELISA 98.6 GC- MS 91.3 | ELISA/GC-MS: 98.1/98.9 | ELISA/GC-MS: 98.2/97.5 | — | Kacinko et al. (49) | |||||
Cozart RapiScan | F | x | x | x | x | x | GC-MS | 5.0–100 | 40.0–100 | 45.7–100 | — | Walsh et al. (20) | |
Cozart RapiScan | — | x | GC-MS | 98.3 | 96.9 | 97.3 | — | Wilson et al. (85) | |||||
Dräger Drug Test | D | x | x | x | x | LC-MS | 22.2–90.1 | 89.8–100 | 77.7–97.3 | — | 10.2 | Concheiro et al. (17) | |
Dräger Drug Test | D | x | LC-MS/MS | 49.5 | 100 | 55.0 | — | Laloup et al. (86) | |||||
Envitec SmartClip | F | x | x | x | HPLC or GC-MS | 65–100 | 33.3–100 | 65.7–97.1 | — | Walsh et al. (20) | |||
LifePoint Impact | F | x | x | x | x | HPLC or GC-MS | 70.0–100 | 33.3–100 | 71.4–100 | — | Walsh et al. (20) | ||
OraSure Uplink | F | x | x | x | x | MS | 10.0–100 | 33.3–100 | 74.3–100 | — | Crouch et al. (84) | ||
Securetec DrugWipe | D | x | x | x | x | ELISA/MSb | 7.7–42.9 | 97.4–99.6 | 86.1–98.1 | — | 0.7 | Crouch et al. (18) | |
Securetec DrugWipe | D | x | x | x | x | x | GC-MS | 50.0–100 | 84.2–99.3 | 79.2–98.6 | — | 5.3 | Pehrsson et al. (67) |
Securetec DrugWipe | F | x | x | x | x | GC-MS | 50.0–100 | 100 | 71.4–100 | — | Walsh et al. (20) | ||
Sun OraLine IV s.a.t. | T | x | GC-MS | 69.2 | 91.7 | 80.0 | — | 10.3 | Cirimele et al. (87) | ||||
Sun OraLine IV s.a.t. | F | x | x | x | x | HPLC or GC-MS | 100 | 0–100 | 54.3–100 | — | Walsh et al. (20) | ||
Ulti-Med SalivaScreen | F | x | x | x | x | GC-MS | 65.0–100 | 33.3–100 | 71.4–94.3 | — | Walsh et al. (20) | ||
Varian OraLab | D | x | x | x | x | LC-MS | 50.0–100 | 97.9–100 | 71.4–100 | — | 24.5 | Concheiro et al. (17) | |
Ansys OralLab | F | x | x | x | x | GC-MS | 40.0–100 | 33.3–100 | 71.4–100 | — | Walsh et al. (20) | ||
Collection devices | |||||||||||||
Acro Biotech Salicule | F | x | x | x | x | x | GC-MS | >90 except THC 45.9 | Langel et al. (3) | ||||
Cozart | F | x | x | x | x | x | GC-MS | — | — | — | 66.0–91.6 | — | Langel et al. (3) |
Cozart DDS | T | x | x | x | x | x | GC-MS or LC-MS | — | — | — | 91.1–100 | — | Speedy et al. (32) |
Dräger DCD 5000 | F | x | LC-MS/MS | — | — | — | >90 | — | Kempf et al. (88) | ||||
Greiner Bio-One Saliva Collection System | F | x | x | x | x | x | GC-MS | — | — | — | 73.6–98.5; >80, except THC | — | Langel et al. (3) |
Immunalysis Quantisal | F | x | x | x | x | x | GC-MS | — | — | — | >80, except THC 55.8 | — | Langel et al. (3) |
Immunalysis Quantisal | F | x | GC-MS | — | — | — | 86.7–96.6 | — | Moore et al. (46) | ||||
Immunalysis Quantisal | Fs | x | x | x | x | x | LC-MS/MS | — | — | — | 81.3–109.4 | — | Quintela et al. (33) |
Malvern Medical OraCol | F | x | x | x | x | x | GC-MS | — | — | — | <12.5–81.5 | — | Langel et al. (3) |
Orasure Intercept | F | x | GC-MS | — | — | — | 53.0–83.4 | — | Kauert et al. (16) | ||||
Orasure Intercept | F | x | x | x | x | x | GC-MS | — | — | — | >80, except THC 37.6 | — | Langel et al. (3) |
Sarstedt Salivette | T | x | x | LC-MS/MS | — | — | — | OP 73.9–95.3 except PV 30.8–35.2, NC 56.4–63.4 COC 74.3–114.9 except pOHBE 47.4–49.6 | — | Dams et al. (89) | |||
Sarstedt Salivette | F | x | x | x | x | x | GC-MS | — | — | — | <12.5–51.8 | — | Langel et al. (3) |
StatSure Saliva Sampler | F | x | x | x | x | x | GC-MS | — | — | — | 81.3–91.1 | — | Langel et al. (3) |
Varian OraTube | F | x | x | x | x | x | GC-MS | — | — | — | 39.8–86.7 | — | Langel et al. (3) |
OP, opioids; AMP, amphetamines; COC, cocaine; BENZO, benzodiazepines; F, fortified; T, treatment; D, DUID; DAD, diode array detection; E, experimental; s, synthetic OF; PV, papaverine; NC, noscapine; pOHBE, p-hydroxybenzoylecgonine.
Confirmation in blood.
For all evaluations reported since 2004, Table 3 contains POCT devices, drug classes evaluated, confirmation methods employed, diagnostic sensitivities, diagnostic specificities, efficiencies, recoveries, and device failures, when reported. In addition, the types of samples tested are described, including fortified blank OF or authentic samples from drug treatment, DUID, or after experimental controlled drug administration. Critical factors must be taken into account when referencing these data: (a) Data are relevant for devices available at the time of testing. (b) Many early devices simply failed to operate correctly. (c) Only a few devices included adequate volume indicators. Collected OF volume varied substantially within and between devices, limiting ability to quantify OF concentrations. (d) Devices may perform differently with fortified authentic or synthetic OF than with authentic OF samples collected after drug ingestion, because the spectrum of analytes present may differ. Knowledge of antibody cross-reactivity data is key for understanding device performance. (e) Drug recovery from the device is a major issue, especially for lipophilic drugs like cannabinoids. (f) Many evaluations judged device performance only against stated manufacturer cutoff concentrations that may be unreasonably high compared to proposed SAMHSA, DRUID, or DUID research recommendations. Other evaluations assessed performance against low laboratory limits of quantification (LOQs), providing a better idea of false-negative rates. (g) Some devices performed poorly at night or in poor weather conditions, or had endpoints that were difficult to discriminate. (h) Some investigators compared OF POCT device performance to serum drug concentrations collected up to several hours after OF. (i) In some investigations, for later confirmation additional OF samples were collected by expectoration or with the Intercept® device.
Most POCT devices obtained good results for opioids with cutoffs at or below SAMHSA cutoffs, although false-negative results may have been missed if all negative screening tests were not subjected to confirmation (Table 3). Reported efficiencies do not always agree, owing to differences in populations, analytes, and concentrations tested and cutoffs used. Opioids have a wide range of potency, leading to differences in sensitivities for various analytes.
Many POCT devices (Table 3) performed well for amphetamines, with diagnostic sensitivities, diagnostic specificities, and efficiencies in the range of 70%–100%. Notable exceptions occurred, however: diagnostic sensitivity was too low for the Varian OraLab® and Dräger Drug Test® (17) and the Securetec Drug-Wipe® (18). The Biomar Toxiquick® (19) and 4 devices evaluated by Walsh et al. (20) performed poorly, with low diagnostic sensitivities, diagnostic specificities, and efficiencies. Development of a POCT for amphetamines is challenging owing to the large number of over-the-counter and prescription drugs sharing similar sympathomimetic amine structures. The challenge is to detect drugs of abuse without detecting multiple therapeutic drugs.
POCT devices (Table 3) performed variably for cocaine. The Cozart RapiScan had diagnostic sensitivities ranging from 5% (20) to 98.2% (21) and efficiencies of 45.7% (20) to 99% (21). Diagnostic specificity was generally good, with a range from 88.7% for cocaine/BE with the Biomar Toxiquick® (cutoff 50 μg/L) (19) to 100% for 6 different devices (20). Diagnostic specificities were poor for the Sun OraLine® IV substance of abuse test and Ansys OralLab® (20).
Other important OF testing issues have been poorly investigated. For example, adulteration of OF samples and potential interferences with POCT devices have not been systematically investigated. It is clear that considerable effort is expended to produce false-negative urine test results, with numerous products available. Certainly similar efforts will be directed at falsifying OF tests. The risk of test falsification requires initial characterization and continuous surveillance of potential interfering substances and adulterants. Specified foods, drinks, mouthwash, and cigarettes did not interfere with the Cozart RapiScan and the Cozart Microplate ELISA assays for opiates (22), cocaine (23), and amphetamines (24), nor with the Branan Oratect® collection device (25). Alcohol and hemoglobin did not affect the Cozart assays and the RapiScan collection device (22–24). Toothpaste, lipstick, gum, and 2 commercially available adulterants also had no effect on Branan Oratect results (25). These few reports provide to date the only available data on adulterants and interferences. Additional research is critically needed to characterize potential problems with OF collection devices and immunological and chromatographic assays.
Chromatographic Confirmation and Quantification of Drugs in OF
From 2004 to 2008, reports of 71 OF drug assays were published, documenting increasing interest in OF as an alternative matrix for identifying and quantifying drug exposure. Chromatographic methods, analytes, extraction procedures, and LOQs for quantifying OF drug biomarkers are described in Table 4. Most new methods utilize liquid chromatography-tandem mass spectrometry (LC-MS/MS), which permits simultaneous analysis of multiple, nonvolatile, labile, polar, and/or high molecular weight compounds in a limited OF volume (26–29). Comprehensive methods for multiple drugs also limit optimization for individual drugs, sometimes yielding insufficient detection capabilities (88, 90). The major disadvantage of LC-MS is matrix enhancement or suppression, which is best managed by inclusion of deuterated internal standards for all analytes; however, this technique may not fully compensate for matrix effects (30). OF is less complex than blood, with fewer proteins, characteristics suggesting the possibility of successful measurement with limited sample preparation, such as dilution, protein precipitation, or centrifugation and direct injection. Unfortunately, more extensive sample preparation is frequently required to limit matrix suppression. Our laboratory found atmospheric pressure chemical ionization (APCI) to be less susceptible to matrix effects than electrospray ionization (ESI) (30). Owing to more efficient ionization, however, ESI is preferred for some analytes, i.e., glucuronides. Many laboratories use ESI routinely and do not consider the advantages of APCI in reducing matrix effects. In addition to LC-MS, many GC-MS assays for drugs in OF also are available, but generally require biomarker derivatization, which increases cost and analysis time.
Table 4.
Method | Extraction | LOQ, μg/L | Reference | ||||
---|---|---|---|---|---|---|---|
Cannabinoids | Opiates | Amphetamines | Cocaine | Benzodiazepines | |||
GC-MS | SPEa | THC 2 | 6AM, FEN 2; BUP, nBUP 4; Pholcodine 5; Methadone, MOR, COD, EtMOR 20 | AMP, MAMP, MDMA, MDA, MDEA 25 | COC, BE 8 | DZP, nDZP, LZP, MZL, OZP, PZP, TZP 2, APZ, CZP, NZP 5 | Gunnar et al. (90) |
LC-MS/MS (ESI) | SPE | MOR, 6AM, COD 2 | AMP, MAMP, MDMA, MDA 2 | COC, BE 2 | Wood et al. (54) | ||
LC-MS/MS (ESI) | SPE | Methadone 2.2; EDDP 0.8; 6AM 1.0; nBUP 1.6; MOR 1.7; COD, BUP 1.8 | AMP1.2; MAMP 1.3; MDMA 2.5; MDA 3.3; MDEA 4 | EME 0.9; COC 3.1; CE 3.4; BE 11.4 | DZP 0.9; nDZP 1.2; CLP, TZP 1.6; OZP 2.5 | Wylie et al. (91) | |
LC-MS/MS (APCI) | PP | nMOR, MOR, nCOD, COD, 6AM, Heroin, AC 1 | COC, EME, BE, CE 1 | Dams et al. (89) | |||
LC-MS/MS (ESI) | LLE | THC <0.16 | 6AM 0.82; BUP 1.2 Methadone 4.0; COD 6.5; MOR 7.1 | MAMP 3.0; MDA <3.6 MDMA <3.9; MDEA <4.1 AMP < 6.8 | COC <0.78 BE 7.2 | APZ 0.093; OZP <1.4; FTP 0.15 NZP 0.23; CZP 0.26; LZP 0.32 DZP 0.41; FZP 0.75; BZP 7.9 | Øiestad et al. (92) |
GC-MS | SPE | CBD 0.9; THC 1.9; THCCOOH 4.8; CBN 5.6; 11- OH-THC 12.7 | 6AM 2.9; MOR 6.5; COD 6.6 | MDMA 8.9; MDA 14.4; MDEA 15.0 A 20.6; MAMP 20.9 | COC 4.1; BE 8; CE 7.2; EME 12.7 | FLP 10.9; DZP 16.3 | Pujadas et al. (93) |
LC-MS/MS (ESI) | SPE | THC 1 | 6AM 1; MOR, COD 5 | AMP, MAMP, MDMA, MDA, MDEA 5 | COC, BE 5 | APZ, CZP, DZP, nDZP, FZP, LZP, OZP 1 | Concheiro et al. (94) |
LC-DAD | LLE/MAE | MOR, 6AM, Methadone, EDDP 50 | COC, BE, CE 50 | Fernandez et al. (95) | |||
LC-MS/MS (ESI) | LLE/SPE | COD, MOR, HCOD, HMOR, OCOD, OMOR 10 | AMP, MAMP, MDMA, MDA 10 | COC, BE 2 | Kala et al. (96) | ||
LC-MS/MS (ESI) | PP | THCCOOH 3.5; THC 3.7 | MOR 1.2; 6AM 1.9 | MAMP 0.4; MDE 0.5; AMP, MDMA, MDA 1.5 | COC, BE 0.5 | Sergi et al. (97) | |
LC-MS (ESI) | LLE | THC 2 | Concheiro et al. (98) | ||||
GC-MS/MS | SPE | THC 0.75 | Niedbala et al. (38) | ||||
LC-MS | SPE | THC 2 | Teixeira et al. (99) | ||||
LC-MS/MS (ESI) | LLE | THC 0.1 (500 μL), 0.5 (100 μL) | Laloup et al. (100) | ||||
LC-MS (ESI) | SPE | THC 2 | Teixeira et al. (101) | ||||
GC-MS/MS (NCI) | SPE | THCCOOH 0.01 | Day et al. (102) | ||||
GC-MS | SPE | THC 2.4 | Kauert et al. (16) | ||||
GC-GC-MS (ECCI) | SPE | THCCOOH 0.02 | Moore et al. (40) | ||||
GC-MS | SPE | 11-OH-THC 0.3; THC 0.5; THCCOOH 6.2 | Kauert et al. (37) | ||||
GC-MS | SPE | Conjugated THCCOOH 0.5 | Moore et al. (42) | ||||
GC-MS | SPE | THC, CBN 0.5; CBD, 2-carboxyTHC 1 | Moore et al. (103) | ||||
LC-MS/MS (ESI) (synthetic OF) | LLE | THC 0.1; THCCOOH 0.5 | Quintela et al. (104) | ||||
LC-MS (ESI) | SPE | THC 5 | Teixeira et al. (105) | ||||
GC-MS (PCI) | LLE | 6AM 2; COD 2.3; MOR 6.7 | Campora et al. (106) | ||||
LC-MS/MS (ESI) | None | 6AM, COD, AC 2; MOR 6; Heroin 10 | Phillips and Allen (107) | ||||
GC-MS | SPE | Propoxyphene 5 | Rana et al. (108) | ||||
GC-MS | SPE | MEP, TRA, OCOD 10 | Moore et al. (46) | ||||
CEP | LLE | EDDP 7.6; Methadone 8.1 | Martins et al. (109) | ||||
GC-MS | LLE | AMP, MAMP, MDMA, MDA, MDEA 20 | Kankaanpää et al. (110) | ||||
LC-FD | LLE | MDMA, MDA, MDEA 10 | Concheiro et al. (111) | ||||
GC-MS | SPE | MDMA, MDA 5; HMMA, HMA 25 | Scheidweiler and Huestis (112) | ||||
GC-MS | SPE | AMP, MAMP, MDMA, MDA, MDEA 5; HMA, HMMA 25 | Scheidweiler and Huestis (113) | ||||
GC-MS | SPE | MAMP 2.5 | Huestis and Cone (52) | ||||
GC-MS | SPE | AMP, MAMP, MDMA, MDA, MDEA 25 | Moore et al. (114) | ||||
GC-MS (NCI) | LLE | MDA 5; AMP, MAMP, MDMA, MDEA 25 | Peters et al. (55) | ||||
LC-MS/MS (ESI) | SPE | COC, BE, CE 10 | Clauwaert et al. (115) | ||||
GC-MS | SPE | BE 0.75 | Robarge et al. (116) | ||||
GC-MS/MS (PCI) | SPE | COC, CE 2; EME 5 | Cognard et al. (117) | ||||
LC-MS (ESI) | LLE | MZL 0.2 | Quintela et al. (118) | ||||
LC-MS/MS (ESI) | LLE | APZ, BZP, DZP, nDZP, LZP, LMP, MZL, OZP, TZP, TRZ, TZL 0.1; CBZ 0.2 | Kintz et al. (119) | ||||
LC-MS (ESI) | LLE | DZP, FTP, MZL, TRZ, TZL 0.2; APZ, BZP, LZP 0.5 | Quintela et al. (120) | ||||
LC-MS/MS (APCI) | LLE | NZP 0.1; APZ, LZP 1.2; LMP 1.3; MZL 2.3; DZP 6.9; nDZP 7.6; TZP 8.1; OZP 13.1 | Smink et al. (83) | ||||
LC-MS/MS (ESI) | LLE | DZP, nDZP, OZP, 0.05; TZP, TRZ 0.2 | Laloup et al. (62) | ||||
LC-MS/MS (ESI) | LLE | MZL 0.05 | Link et al. (121) | ||||
LC-MS/MS (ESI) | SPE | APZ, nDZP, FTP, FLP, MZL, TZP, TZL 0.5; BZP, CZP, DZP, NZP 1; LZP, OZP 5 | Moore et al. (122) | ||||
LC-MS/MS (APCI) | SPE | APZ, CZP, DZP, nDZP, FTP, LZP, OZP, TZP 0.1 | Ngwa et al. (123) | ||||
LC-MS/MS | PP | OZP glucuronide 0.25; OZP 0.5 | Smink et al. (63) | ||||
LC-DAD | SPE | LZP 130; DZP 290; BZP 370; CZP, FTP 390; APZ 450 | Uddin et al. (124) |
SPE, solid phase extraction; FEN, fentanyl; BUP, buprenorphine; nBUP, norbuprenorphine; MOR, morphine; COD, codeine; EtMOR, ethylmorphine; AMP, amphetamine; MAMP, methamphetamine; MDEA, 3,4-methylenedioxy-ethylamphetamine; COC, cocaine; DZP, diazepam; nDZP, nordiazepam; LZP, lorazepam; MZL, midazolam; OZP, oxazepam; PZP, phenazepam; TZP, temazepam; APZ, alprazolam; CZP, clonazepam; EDDP, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine; CE, cocaethylene; CLP, citalopram; PP, protein precipitation; nMOR, normorphine; nCOD, norcodeine; AC, acetylcodeine; LLE, liquid liquid extraction; FLP, flurazepam; FZP, fenazepam; BZP, bromazepam; DAD, diode array detection; MAE, microwave-assisted extraction; HCOD, hydrocodone; HMOR, hydromorphone; OCOD, oxycodone; OMOR, oxymorphone; MDE, methylenedioxy-N-ethylamphetamine; NCI, negative chemical ionization; ECCI, electron capture chemical ionization; PCI, positive chemical ionization; MEP, meperidine; TRA, tramadol; CEP, capillary electrophoresis; HMMA, 3-hydroxy-4-methoxymethamphetamine; HMA, 3-hydroxy-4-methoxyamphetamine; FD, fluorescence detection; LMP, lormetazepam; TRZ, tetrazepam; TZL, triazolam; CBZ, clobazam; FTP, flunitrazepam; NZP, nitrazepam.
OF Testing of Specific Drug Classes
With various classes of drugs that are measured in OF, including cannabinoids, amphetamines, opiates, cocaine, and benzodiazepines, specific issues arise with regard to collection, screening, POCT, and result confirmation and interpretation. Controlled drug administration provides a scientific database to guide interpretation of OF tests. The disposition of illicit drugs in OF is affected by metabolic, physiological, and chemical processes (31), including biomarkers of interest, minimal detectable doses, expected concentrations, detection windows, and collection device requirements. The major research findings are detailed by drug class. Specific studies are referenced in Table 5.
Table 5.
THC (cannabinoids) | Opioids | Amphetamines | Cocaine | BENZOa | Doses | Study participants, N | Reference |
---|---|---|---|---|---|---|---|
x | 1 Cannabis joint, potency unknown | 3 | Jehanli et al. (125) | ||||
x | 20–25 mg Smoked THC 20–25 mg THC (brownies) |
15 3 |
Niedbala et al. (44) | ||||
x | 0, 1.75, and 3.55% Smoked THC | 6 | Huestis and Cone (13) | ||||
x | Passive cannabis smoke exposure | 4 | Niedbala et al. (38) | ||||
x | Passive cannabis smoke exposure | 4 | Niedbala et al. (39) | ||||
x | 0, 250, and 500 μg/kg Smoked THC | 10 | Ramaekers et al. (36); Kauert et al. (37) | ||||
x | 1 Cannabis joint, potency unknown | 1 | Moore et al. (40) | ||||
x | 30 mg Oral codeine | 3 | Sharp et al. (126) | ||||
x | 60 and 120 mg Intramuscular codeine | 1 | Cone (127) | ||||
x | Oral codeine | 8 | Chen et al. (128) | ||||
x | 30 mg Oral liquid codeine phosphate | 17 | O’Neal et al. (129) | ||||
x | 30 mg Oral liquid codeine phosphate | 37 | O’Neal et al. (130) | ||||
x | 16 mg Oral codeine | 5 | Jehanli et al. (125) | ||||
x | Single and multiple 60 and 120 mg/70 kg codeine | 19 | Kim et al. (48); Kacinko et al. (49) | ||||
x | Poppy seed bagels and poppy seeds | 7 | Rohrig and Moore (50) | ||||
x | 10 or 20 mg MAMP | 8 | Schepers et al. (51) | ||||
x | 0, 75, and 100 mg Oral MDMA | 18 | Laloup et al. (53); Wood et al. (54) | ||||
x | Single and multiple 10 and 20 mg MAMP | 8 | Huestis and Cone (52) | ||||
x | 75 mg MDMA | 12 | Peters et al. (55) | ||||
x | 25 mg Smoked, 32 mg intravenous, 42 mg insufflated cocaine, multiple oral cocaine up to 2000 mg/day | 66 | Jufer et al. (59) | ||||
x | 1 mg Oral flunitrazepam | 4 | Samyn et al. (61) | ||||
x | 50 mg Oral TZP or 10 mg oral DZP or repeated 25 mg oral TZP | 7 | Laloup et al. (62) | ||||
x | 15 and 30 mg Oxazepam | 8 | Smink et al. (63) |
BENZO, benzodiazepines; MAMP, methamphetamine; MDMA, 3,4-methylenedioxymethamphetamine; DZP, diazepam; TZP, tetrazepam.
CANNABINOIDS
THC is the primary analyte for cannabinoids in OF, whereas 11-nor-9-carboxy-THC (THCCOOH) predominates in urine. Most available antibodies from urine immunoassays are directed toward THCCOOH and have poor cross-reactivity with THC, limiting diagnostic sensitivity.
Recovery of the sample from the collection device is a major issue, especially for lipophilic drugs like cannabinoids, although 94.5% THC recovery from the Cozart DDS device (32) and up to 91.4% recovery from the Quantisal device at a concentration of 4 μg/L (33) were recently reported. It is suggested that manufacturers evaluate each lot of OF collection devices and publish the data in the package insert, which would alleviate each laboratory from performing the same function. To our knowledge, only THC recovery from devices has been examined, although THCCOOH recovery may be equally important.
Initially, THC was thought to be present in OF only from oral mucosal contamination from cannabis smoke. No radiolabel was detected in human OF following radiolabeled intravenous THC administration (34). Using GC-MS with 0.5 μg/L LOQs, we detected no THC metabolites, 11-hydroxy-THC (11-OH-THC) or THCCOOH, in OF after 6 participants smoked a 1.75% or 3.55% THC cigarette (35). However, later we observed that after approximately 30–60 min, gross oral mucosal contamination following cannabis smoking was cleared and OF and plasma concentrations were temporally correlated (13). THC concentrations as high as 5800 μg/L were observed 12 min after cannabis smoking, falling to 81 μg/L by 20 min, documenting rapid drug clearance. Mean (SD) THC OF/plasma ratio in simultaneously collected samples was 1.18 (0.62) (range 0.5–2.2) 0.3 to 4 h after cannabis smoking. Similarity in OF and plasma THC concentrations following contamination dissipation likely indicates a physiological link (transmucosal THC absorption into blood) between these matrices.
Other investigators recently reported a strong linear correlation (r = 0.84, P < 0.001) in OF and serum THC concentrations up to 6 h after sample donors had smoked 250 or 500 μg/kg THC (36). Mean (SD) THC OF concentrations were 900 (589) and 1041 (652) μg/L, respectively, in samples collected 15 min after donors had smoked, decreasing to 18 (12) μg/L during the next 6 h (37). OF/serum ratios were markedly higher than previously reported at 46 (27) (250 μg/kg) and 36 (20) (500 μg/kg), with a longer period of THC OF detection in OF compared to serum samples. It is not yet clear why the OF/plasma (13) and OF/serum (37) ratios are so discrepant between these studies; additional research is underway. We noted that although plasma and OF THC concentrations were temporally correlated, there was too much variability to predict plasma concentrations from single OF results.
Another major issue is the possibility of false-positive OF test results in sample donors exposed to environmental cannabis smoke. With a 0.5 μg/L gas chromatography–tandem mass spectrometry (GC-MS/MS) cutoff, passive cannabis smoke exposure produced positive THC OF test results in 4 nonsmokers after they spent 15–30 min in a sealed room with 5 cannabis smokers (38). Contamination cleared rapidly, with negative results for all later samples. In another passive exposure study in which sample donors were exposed to cannabis smoke while in an unventilated van, OF samples were collected with the Intercept device inside the van 1 h after smoking cessation (study 1) and outside the van after up to 72 h (study 2), with 4 study participants smoking 39.5 mg THC mixed with tobacco (study 1) or 83.2 mg THC (study 2) cigarettes (39). Peak OF THC concentrations in passively exposed sample donors were up to 7.5 μg/L (study 1) and 1.2 μg/L (study 2); THC was not quantifiable by 30–45 min after exposure. Collection devices environmentally exposed in the van contained 3–14 μg/L in study 1. When OF samples were collected outside the van after the 1 h exposure, all OF samples were negative, suggesting that earlier studies may have used environmentally contaminated collection devices. OF results were positive for active smokers for 0–8 h.
More recently, THCCOOH (40, 41) and its glucuronide metabolite (42) were quantified in OF in nanogram per liter concentrations. This finding was important, because THCCOOH is not in cannabis smoke. Demonstrating THCCOOH in OF could clearly differentiate passive exposure from active smoking. Until recently, no analytical procedures existed for simultaneously extracting and quantifying THC and THCCOOH from the same OF sample. We recently developed and validated an assay for simultaneous identification and quantification of THC, cannabidiol (CBD), cannabinol (CBN), 11-OH-THC, and THCCOOH in OF collected with the Quantisal device (43). Simultaneous analysis was problematic owing to varying physiochemical characteristics and concentration ranges. THC, 11-OH-THC, CBD, and CBN were eluted with hexane/acetone/ethyl acetate (60:30:20 vol/vol/vol), derivatized with N, O-bis-(trimethylsilyl) tri-fluoroacetamide and quantified by 2-dimensional GC electron-impact MS with cryotrapping. Acidic THC-COOH was separately eluted with hexane/ethyl acetate/acetic acid (75:25:2.5 vol/vol/vol), derivatized with tri-fluoroacetic anhydride and hexafluoroisopropanol, and quantified by the more sensitive 2-dimensional GC-MS–electron capture negative chemical ionization method. Linearity was 0.5–50 μg/L for THC, 11-OH-THC, and CBD and 1–50 μg/L for CBN. The linear dynamic range for THCCOOH was 7.5–500 ng/L. This new analytical method is being applied to OF samples collected in our cannabinoid controlled drug administration studies, and we hope it will provide new insights into cannabinoid disposition in OF.
Acceptance of OF testing for cannabinoids has been difficult for multiple reasons, including short detection times. The mean time for the last positive result for THC in OF, collected with the Intercept OF collection device and analyzed by GC-MS/MS (LOQ 0.5 μg/L), was 34 (11) h (range 1–72 h) after the sample donor had smoked a 20–25 mg THC cigarette (44). However, study participants left the laboratory after 8 h and were instructed not to smoke cannabis, but they were not monitored. Mean OF THC concentrations 1 h after smoking were 27.8 (6.2) and 22.6 (6.2) μg/L from the right and left sides of the mouth, with good concordance throughout 72 h after smoking (44). OF samples were positive up to 72 h after eating 20–25 mg THC-laced brownies; concentrations were <6.9 μg/L THC (44). Currently, few data are available for detection windows after acute cannabis smoking, and no data are available for detection after ingestion of THC-containing pharmacotherapies or after chronic, heavy cannabis smoking. The lack of such data limits our ability to interpret OF test results.
Cannabinoid cutoffs were proposed before THC-COOH was detected in OF. Certainly quantification of THCCOOH requires a highly sensitive assay, but such an assay may prove to be the answer for cannabinoid OF testing. The window of detection and prevalence of THCCOOH in OF is still unknown. Perhaps an immunoassay screen for THC and confirmation for THC-COOH would provide definitive identification of cannabis use, although this must be proven. Because THC and cannabinoids are the most prevalent drugs in clinical, workplace, and DUID testing (45), it is difficult to move to an alternative matrix if cannabinoids are not adequately and specifically identified. These limitations have slowed acceptance of OF for multiple drug testing applications.
OPIOIDS
Opioids have many therapeutic applications, but these drugs also are abused. Heroin has a short half-life and is rapidly metabolized to 6AM and then morphine. Although 6AM can be detected in urine for a short time after exposure, one of the main advantages of OF over urine testing for opiates is that 6AM and sometimes heroin are frequently detected in OF. SAMHSA increased the mandated urine testing cutoff from 300 to 2000 μg/L because Medical Review Officers acted on so few opiate-positive tests; most were attributed to codeine use or poppy seed ingestion rather than heroin abuse.
Recoveries of opioids from collection devices were approximately 80% for most devices, except the Sarstedt Salivette® (35.2% for morphine and 39.0% for codeine) (3). Existing immunoassay antibodies adequately target morphine, codeine, and 6AM in OF. A problem for opioid OF and urine testing is the wide range of analytes that are abused, including heroin, codeine, oxycodone, buprenorphine, hydromorphone, and hydrocodone. Additional specific assays are available that target oxycodone (46) and buprenorphine (47). Multiple screening assays must be conducted to adequately cover this large class of abused drugs.
There are few controlled opiate administration studies to guide interpretation of OF tests. Our laboratory administered placebo, low (60 mg/70 kg) and high (120 mg/70 kg) oral codeine sulfate to 19 participants and then collected OF for up to 72 h by citric acid–stimulated expectoration and by stimulated and un-stimulated Salivette. Samples were analyzed by GC-MS with LOQs of 2.5 μg/L for codeine, norcodeine, morphine, and normorphine (48, 49). At 0.08–1.0 h after administration, codeine was detected with all collection methods after both doses. Mean peak OF codeine concentrations did not differ significantly between collection methods and were 638.4 (64.4) μg/L at 1.7 (0.23) h (low) and 1599.3 (241.0) μg/L at 1.6 (0.14) h (high). During the 1–12-h period after administration, the OF/plasma ratio was approximately 4. Codeine is a basic drug and is ion trapped in OF, increasing its detection. OF and plasma codeine concentrations were significantly correlated (r = 0.22, P < 0.0001), but variability precluded predicting plasma concentrations from OF concentrations. Mean codeine half-lives were 2.5 (0.21) h and 1.8 (0.19) h, similar to plasma half-lives. Mean codeine OF detection time was 21 h (using the LOQ), but only 7 h at proposed SAMHSA cutoffs. Norcodeine/codeine ratios increased over time from 0.5–8 h, ranging from 0.3%–31%. Norcodeine OF/plasma ratios were lower than those of codeine owing to increased polarity and a lower pKa, conditions that decrease ion trapping. Morphine and normorphine have not been detected in OF or plasma after codeine administration, whereas late in the time course of urinary excretion of codeine, low morphine concentrations may be present without measurable codeine, making interpretation difficult. This does not occur in OF after codeine ingestion, another advantage of OF compared to urine testing for opiates.
These data document similar opiate detection windows for OF and plasma, making OF a good matrix for DUID, because the OF detection window better reflects drug intoxication than urinary detection windows. OF also is a good alternative matrix for drug treatment, owing to the shorter opiate half-lives in OF than urine. For the same reason, consecutive OF samples are less likely to be positive after a single opiate exposure than consecutive urine samples. A difficult problem in drug treatment programs is differentiating new opiate use from residual opiate excretion in urine. Our controlled drug administration data support that such differentiation would be much less problematic with OF than with urine testing.
Another problem with opiate urine testing is that positive results may occur after sample donors have ingested poppy seed foodstuffs. A recent study showed that ingestion of poppy seed bagels and poppy seeds (9.8–20.8 g) also can produce positive OF tests, albeit only for up to 1 h after ingestion, when a morphine OF cutoff of 40 μg/L is applied (50).
Thus, OF opiate testing offers multiple advantages to urine testing, including observed, easy collection, even at the roadside; enhanced identification of heroin abuse; better differentiation of new opiate exposure from residual drug excretion; and reduced possibility of positive tests following ingestion of opiate-containing food. Urine testing for opiates offers the advantage of higher drug concentrations and greater availability of data on controlled administration.
AMPHETAMINES
Amphetamine is the primary abused sympathomimetic amine in Europe, whereas methamphetamine is more prominent in the Americas, Asia, and Australia. Cross-reactivity with methamphetamine, amphetamine, and 3,4-methylenedioxymethamphetamine (MDMA) (ecstasy) is considered an advantage because only one screen adequately covers the three most commonly abused drugs, while cross-reactivity with phenylpropanolamine, ephedrines and other common cold medication ingredients is a disadvantage, due to the need for expensive and time-consuming confirmatory procedures to rule out these drugs. Analytes of interest are the same in urine and OF, thus, cross-reactivity is a major testing issue for both. Also, drug potencies and doses are similar for the analytes of interest, making sensitivity less of a problem.
One of the problems encountered in OF testing after stimulant abuse is dry mouth. During routine OF testing and controlled dosing research, mouth dryness in sample donors may lead to low sample volumes and/or long collection times. Recovery from collection devices was rarely problematic for amphetamines, with percentages generally above 75%. However, the Sarstedt Salivette had low recoveries of 51.8%/26.5% for amphetamine/MDMA, as did the Malvern Medical OraCol (69.1%/52.0%) for the same analytes (3).
Amphetamines are weak bases and are subject to ion trapping in OF, providing the advantage of higher concentrations in OF than plasma. We first detected methamphetamine in expectorated OF with citric acid stimulation within 0.08–2 h after sample donors had ingested 10 or 20 mg sustained-release oral S-(+)-methamphetamine hydrochloride (GC-MS LOQs 2.5 μg/L) (51). We found that samples collected with the use of citric acid stimulation had amphetamine concentrations 1.5-fold lower than samples collected without stimulation. The decrease in OF pH that occurs with citric acid should increase basic amphetamine concentrations, but stimulation had a greater effect on salivary volume. For samples collected by expectoration, mean peak OF concentrations after single 10- or 20-mg doses were 106.1 (24.7) μg/L (range 25–312 μg/L) during the period of 4–8 h after dosing and 192.2 (120.8) μg/L (range 75–322) 2–12 h after dosing. The linear correlation between OF and plasma concentrations, although significant, was only 0.222. The median OF/plasma methamphetamine ratio was 2.0 and was highly variable within and between participants. Methamphetamine was detected in the OF of all study participants for 24 h; amphetamine was detected in only 62.5% (low) and 100% (high) of samples after administration at 8.6 (6.5) μg/L (range 4–21 μg/L) and 14.3 (6.1) μg/L (range 3–20 μg/L). Interestingly, OF was not contaminated by oral administration of this preparation, as noted by a lag time before first drug detection. It was thought that swallowing the methamphetamine pill would lead to contamination of the oral mucosa, but such contamination did not occur. If methamphetamine was smoked, extensive contamination of the oral cavity would be expected. The OF detection rate at 50 μg/L (proposed SAMHSA cutoffs) was only 4% after 10 mg and 28.6% after 20 mg methamphetamine, which are therapeutic doses and quite low compared to abuse doses (52). Mean detection times until last positive OF after the last of 4 daily 10- or 20-mg methamphetamine doses (using the LOQ) were 24.6 (9.9) h and 74.1 (17.0) h, respectively, and with a 50-μg/L methamphetamine and 2.5-μg/L amphetamine cutoff were 4.0 (1.8) h and 20.6 (3.2) h. Detection times for OF were much shorter than those for urine.
In contrast to the controlled dosing data available for methamphetamine, MDMA doses spanning the recreational range have been administered. Median MDMA and 3,4-methylenedioxyamphetamine (MDA) OF concentrations collected with the OraSure Intercept® device 1.5 h after sample donors had received 75 or 100 mg oral MDMA were 447.6 μg/L (range 44.1–3993.1 μg/L) and 7.7 μg/L (range 1–42.4 μg/L), respectively (53, 54). By 5.5 h, median MDMA concentrations were 315.7 μg/L (range 42.4–3078.9 μg/L) and MDA concentrations were 16.5 μg/L (range 1.5–102.6 μg/L). Unfortunately, no later samples were collected, negating the determination of MDMA OF detection windows.
Determination of MDMA detection windows also was not possible in a study in which racemic MDMA (75 mg) was administered to 9 participants and OF was collected by expectoration from 1–5 h after administration (55). Interestingly, however, MDMA and MDA enantiomers were quantified by GC-MS. Mean peak OF concentrations 1–4 h after ingestion were approximately 600 μg/L, with R-(−)-MDMA significantly (P < 0.05) exceeding S-(+)-MDMA (approximately 400 μg/L) concentrations in all samples. Enantiomer ratios (R/S) increased over time and ranged from 1.04–1.92. R-(−)-MDA concentrations were <5 μg/L in 78% of samples and never >21.4 μg/L, whereas S-(+)-MDA concentrations were quantified up to 74.7 μg/L in 72% of samples.
As weak bases, amphetamines are ion trapped in OF, leading to increased concentrations and windows of drug detection. OF testing appears to be a good alternative matrix to urine, although data informing interpretation of methamphetamine, amphetamine, and MDMA OF concentrations, including windows of drug detection, are limited.
COCAINE
Cocaine is a highly abused drug in the US and South America, and prevalence of abuse has increased in Europe in recent years (56). Because cocaine is a weak base and stimulant, ion trapping in OF is expected, as is the incidence of dry mouth and restricted sample volume. In simultaneously collected plasma and OF samples, we found higher concentrations of cocaine in plasma than OF, and the reverse for the primary, more polar, BE metabolite (57, 58). Recovery of cocaine from collection devices was reported as good (85.6%–100%), except for the Malvern Medical OraCol device (3).
Although cocaine is an important analyte in emergency toxicology, drug treatment, criminal justice, and workplace drug testing programs, there are only 2 reports of cocaine OF concentrations after controlled drug administration. In the first, cocaine and metabolite disposition in OF after single 25-mg intravenous, 32-mg intranasal, and 42-mg smoked doses of cocaine was determined in OF collected by stimulated expectoration before and up to 12 h after dosing (59). Mean last detection times for cocaine were short, 4.7 (0.6), 6.3 (1.2), and 4.1 (1.0) h with an 8 μg/L proposed SAMHSA cutoff. However, these administered doses are low compared to recreational doses and doses self-administered by tolerant users. BE mean last detection times were 6.7 (1.9), 8.7 (1.5), and 5.0 (2.1) h at the same cutoff. Cocaine and BE OF disposition also were evaluated following up to 5 daily escalating doses of oral cocaine, beginning with 100 mg and increasing by 25 mg to a maximum single dose of 400 mg and a maximum total dose of 2000 mg/day for up to 16 days (59). OF detection times with stimulated expectoration after repeated dosing were 4-fold and 7-fold longer for cocaine and BE compared to single doses. The authors suggested that detection times based on single doses underestimate the utility of OF for identifying cocaine use because high, repeated doses are frequently self-administered.
In the other investigation of controlled cocaine administration, participants each received placebo or 75 and 150 mg/70 kg cocaine sulfate on different days, administered subcutaneously for safety purposes (58). OF collection occurred for 48 h after each dose and samples were obtained by expectoration with citric acid candy stimulation. OF was analyzed by GC-MS with 2.5 μg/L LOQs for cocaine, BE, and ecgonine methylester (EME). OF was positive for cocaine within 5 min in 65% of low and 100% of high-dose samples. Mean onset times of EME and BE were similar, approximately 0.1 h in plasma and 0.3 h in OF. OF cocaine maximum concentrations were 1322.3 (848) and 3130.7 (2228.2) μg/L after the 75 and 150 mg/70 kg doses. Cocaine and EME maximum concentrations were significantly higher in OF than plasma, whereas BE maximum concentrations were higher in plasma. Mean BE OF maximum concentrations were 154.7 (100.3) and 308.0 (163.8) μg/L. Cocaine maximum concentrations was slightly lower in plasma than in OF. Although cocaine concentrations exceeded those of BE in OF, detection times were generally longer for BE in this matrix. Cocaine, EME, and BE were ≥LOQ in 93% of participant samples at 8 h. After study participants received the high dose, cocaine OF/plasma ratios tended to be 1–2 up to 0.25 h and >3 at 0.5–8 h; about half remained >3 up to 48 h.
Cocaine is a weak base and thereby is subject to OF ion trapping, which leads to increased concentrations and windows of drug detection. Reduced salivary volume follows the use of this stimulant. Abuse of smoked crack cocaine, insufflation of cocaine hydrochloride, and oral cocaine abuse will contaminate the oral cavity, with resulting high initial OF concentrations compared to concentration that occur after intravenous cocaine abuse. There are no data describing how long cocaine concentrations remain increased after mucosal contamination. OF appears to be a good alternative matrix to urine, with reasonable detection windows of 1–2 days, even after recreational single 100-mg doses.
BENZODIAZEPINES
Benzodiazepines encompass a large variety of drugs with differing pharmacodynamic effects and pharmacokinetic profiles. SAMHSA does not include benzodiazepines in workplace drug testing, but DRUID recommends cutoffs (Table 1). Benzodiazepines have a high prevalence in DUID cases (4) and appear to be especially impairing during the first few weeks of therapy (60). Monitoring benzodiazepine exposure with OF is especially challenging owing to the wide range of available benzodiazepines with highly variable potencies, low concentrations, legal prescription status, high lipophilicity, analyte instability in nonpreserved OF, and variable cross-reactivities with antibodies. Many jurisdictions, especially in Europe, strongly advocate inclusion of benzodiazepines in testing strategies because of their contribution to DUID. However, few data are available for OF benzodiazepines.
Research on collection devices for benzodiazepines is limited and results are more variable (Table 3). Langel et al. reported 97.1% and 96.9% recoveries for diazepam and alprazolam from OF collected by expectoration, and 95.7% and 95.8% with the Arco Biotech Salicule™ device (3). The Varian OraTube, Sarstedt Salivette, and Malvern Medical OraCol were not suitable for benzodiazepine collection, with recoveries <49% (3). Recoveries for the Cozart device were 66.0% for alprazolam but 91.6% for diazepam, results that highlight the variability within this class (3). Speedy et al. found a recovery of 92.1% for temazepam with Cozart DDS (32) and Quintela et al. 101.3% for oxazepam with Immunalysis Quantisal (33).
Flunitrazepam was detected only up to 6 h in expectorated OF after oral dosing of 1 mg. Concentrations were <0.6 μg/L if OF was preserved with 2% sodium fluoride. Concentrations of flunitrazepam’s metabolite 7-aminoflunitrazepam were <3.1 μg/L(61). OF was positive for tetrazepam from 0.25–10 h after a 50-mg dose, and no diazepam or nordiazepam metabolites were measurable in OF, but these metabolites were positive in urine (62). Diazepam was found as a pill contaminant, but also may have derived from metabolism of tetrazepam. Three study participants receiving 10 mg diazepam in the same study had positive OF samples for >10 h. Other investigators reported that after study participants received 15 or 30 mg oxazepam, oxazepam and oxazepam glucuronide were detected in expectorated OF for >8.5 h, in much lower concentrations than in simultaneously collected blood (63). OF/blood ratios ranged from 0.04–0.07 for oxazepam and from 0.002–0.006 for oxazepam glucuronide.
These data indicate that OF testing for benzodiazepines will be challenging indeed. Low OF concentrations, a multitude of drugs within the class, instability of some parent drugs, and potential recovery issues from collection devices must be addressed. Although a single study has addressed glucuronide-benzodiazepine conjugates, the available data suggest that glucuronide concentrations may be an order of magnitude lower than free drug in OF, and hydrolysis of OF samples may not be necessary. Alternatively, LC-MS/MS may permit direct measurement of glucuronide conjugates; however, matrix suppression and coelution with endogenous matrix components are adverse events that frequently occur with early elution of conjugates on the most common reversed-phase columns.
Applications
Applications of OF testing in DUID, treatment, and workplace settings reported in the last 5 years are listed in Table 6. The number of such reports is increasing, providing excellent data on the usefulness of this new technology, the prevalence of drugs in various populations, and the performance of various onsite tests. It is hoped that recommendations proposed at the Talloires meeting of experts for standardizing research DUID methods will be followed for future studies, allowing better comparison of data across studies than can be currently accomplished.
Table 6.
Type | No. of samples | Screening | Confirmation | Prevalence, % | Reference | ||||
---|---|---|---|---|---|---|---|---|---|
THC | Opiates | Amphetamines | Cocaine | Benzodiazepines | |||||
D: I/Sa | 302 | Toxiquick | GC-MSb | 56.8 | 8.0 | 48.3 | 5.7 | Biermann et al. (19) | |
D: Hos | 300 | Cozart Microplate ELISA | GC-MS & LC-MS/MS | ≥1.7 | ≥1.4 | ≥1.4 | ≥0.4 | ≥6.7 | Bernhoft et al. (70) |
D | 61 | DrugWipe & Intercept | GC-MS | 29.5 | Kintz et al. (71) | ||||
D: I/S | 131 | Dräger DrugTest | GC-MS | 54.2 | 17.6 | 32.8 | 14.5 | Toennes et al. (64) | |
D: I/S | 177 | Dräger DrugTest | GC-MS | 54.2 | 19.8 | 44.1 | 16.4 | Toennes et al. (65) | |
D | 55 | — | LC-MS/MS | 3.6 | 27.3 | 20.0 | Wood et al. (54) | ||
D: Ra | 1396 | Cozart Microplate ELISA | GC-MS | 3.7 | ≥3.2 | ≥4.6 | 1.6 | 2.0 | Wylie et al. (69) |
D: Rd | 468 | Dräger DrugTest | LC-MS | ≥38.7 | ≥3.5 | ≥7.0 | ≥35.5 | Concheiro et al. (17) | |
D: Rd | 468 | OraLab | LC-MS | ≥9.5 | ≥3.5 | ≥1.2 | ≥18.5 | Concheiro et al. (17) | |
D: R | 13 176 | DrugWipe & RapiScan | GC-MS | ≥0.7 | ≥2.0 | Drummer et al. (10) | |||
D: I/S | 114 | DrugWipe | GC-MS | 87.7 | Engblom et al. (66) | ||||
D: I/S | 40 | DrugWipe | ELISA & MSb | 37.5 | 12.5 | 5.0 | 20.0 | Crouch et al. (18) | |
D: I/S | 267 | DrugWipe | ELISA & MSb | 14.6 | 2.6 | 4.1 | 14.6 | Crouch et al. (18) | |
D: R | 10 816 | — | LC-MS/MS | 0.6 | 1.1 | 0.3 | 0.1 | 1.4 | Gjerde et al. (68) |
D: I/S | 148 | DrugWipe | GC-MS | 15.5 | 2.7 | 90.5 | 1.4 | 50.6 | Pehrsson et al. (67) |
T | 217 | Cozart Microplate ELISA | GC-MS | 53.5 | Cooper et al. (23) | ||||
T | 358 | Cozart RapiScan | GC-MS | 25.4 | 31.6 | Cooper et al. (21) | |||
T | 198 | Cozart Microplate ELISA | GC-MS | 57.1 | Cooper et al. (79) | ||||
T | 216 | Cozart Microplate ELISA | GC-MS | 50.5 | Cooper et al. (22) | ||||
T | 27 | Sun OraLine IV s.a.t. | GC-MS | 52.0 | Cirimele et al. (87) | ||||
T | 135 | Cozart Microplate ELISA | GC-MS | 57.8 | Cooper et al. (24) | ||||
T | 513 | Orasure Intercept | LC-MS/MS | 57.9 | Phillips and Allen (107) | ||||
T&W | 635 000 | Orasure Intercept | GC-MS/MS | ≥0.1 | ≥1.1 | ≥0.1 | ≥0.2 | ≥0.1 | Cone et al. (74) |
T | 403 | Sarstedt Salivette | LC-MS/MS | 15.4 | 32.3 | Dams et al. (89) | |||
T | 134 | Cozart Microplate ELISA | GC-MS & LC-MS | 73.1 | 23.9 | Speedy et al. (32) | |||
T | 20 | Cozart DDSV | GC-MS | 90.0 | Kintz et al. (14) |
D, DUID; Hos, hospitalized drivers; I/S, impaired or suspected drivers; R, random; Ra, random at high-risk accident area; Rd, random near discotheques; T, treatment; W, workplace testing;
Screening with OF, but confirmation with blood.
DUID studies have tested impaired or suspected drivers (18, 19, 64–67), random drivers at road-blocks (10, 68), and drivers near discotheques (17) and high-risk accident areas (69), and have evaluated injured hospitalized drivers (70). Percentages of positive tests were highly variable (range 0.1%–90.5%) owing to differences in study populations, collection and analytical methods, LOQs, times between stop and OF collection, and DUID laws (69).
In 46.7% of reports, cannabinoids were the most prevalent drug. THC concentrations ranged from 0.7 (71) to 6484 μg/L (10). The higher concentrations appear to be attributable to contamination of the oral mucosa from cannabis smoke immediately after use; median concentration was 81 μg/L (10). Concheiro et al. evaluated the Varian OraLab® and Dräger Drug Test®, but also collected OF for confirmation by expectoration, eliminating the variable of cannabinoid recovery from the collection device (17).
Opioids are one of the least commonly identified drug classes in DUID. Detected concentrations have ranged from 4 μg/L for codeine to 7600 μg/L for 6AM, the latter possibly due to oral contamination, because high-purity heroin may be smoked, inhaled, or snorted (69). Opioids are one of the most investigated drug classes in treatment settings (54.5% of studies).
Amphetamine prevalence was as high as 90.5% (67) and 87.7% (66) in 2 Finnish studies of DUID suspected drivers. In a study by Engblom et al. OF concentrations up to 131 000 μg/L for amphetamine were observed, most likely due to oral contamination, because the median concentration was 7440 μg/L (66).
Cocaine was one of the least commonly reported drugs in DUID, having the lowest prevalence rate in 50% of worldwide studies. In 2 populations, however, Spain 18.5% (17) and Utah 14.6% (18), it was the most commonly found drug.
Benzodiazepines are an important drug class to monitor because these drugs are frequently reported in DUID cases (72), yet they are some of the most commonly prescribed medications (73). Another difficulty is that an individual may be highly impaired when first taking a benzodiazepine, but rapidly develop tolerance to the drug’s impairing effects (60). Thus, determining an appropriate cutoff concentration that reflects impairment is difficult for this and most other drug classes. Reported concentrations ranged from 4 μg/L for temazepam, nordiazepam, or oxazepam to 221 μg/L for nordiazepam in DUID (69). In treatment and workplace settings, concentrations ranged from 0.5 μg/L for diazepam, nordiazepam, oxazepam, and temazepam to 14 301 μg/L for diazepam, with a median of 2.8 μg/L (74). In this interesting study, which compared the sensitivity of drug detection in 635 000 treatment and workplace urine and OF samples, investigators concluded that OF was as efficient as urine for identifying drug use.
In conclusion, the technology of OF collection devices, laboratory-based assays, on-site testing devices, and chromatographic confirmation methods has greatly advanced in the last 5 years. POCT devices are improving, with better recoveries, especially for THC, and more accurate performance. However, successful development of a POCT device that performs acceptably for all drug classes is a challenge. Currently available chromatographic methods quantify multiple analytes at increasingly lower concentrations, and important new analytes (e.g., THCCOOH) have been identified. Perhaps the greatest current limitation for OF testing is the small number of controlled drug administration studies available to inform interpretation of OF tests. Additional research is needed to identify new biomarkers, determine drug detection windows, characterize OF adulteration techniques, and evaluate analyte stability in OF. There is no doubt that OF offers multiple advantages as an alternative matrix for monitoring licit and illicit drug use, and that OF testing has an important role in DUID, treatment, workplace, and criminal justice programs.
Acknowledgments
Research Funding: NIH, Intramural Research Program of the National Institute on Drug Abuse.
Role of Sponsor: The funding organizations played no role in the design of study, choice of enrolled patients, review and interpretation of data, or preparation or approval of manuscript.
Footnotes
Nonstandard abbreviations: OF, oral fluid; DUID, driving under the influence of drugs; 6AM, 6-acetylmorphine; SAMHSA, Substance Abuse and Mental Health Services Administration; ROSITA, ROadSIde Testing Assessment; POCT, point-of-collection testing; DRUID, DRiving Under the Influence of Drugs, alcohol, and medicines; THC, Δ9-tetrahydrocannabinol; DDS, Drug Detection System; BE, benzoylecgonine; LOQ, limit of quantification; LC-MS/MS, liquid chromatography-tandem mass spectrometry; APCI, atmospheric pressure chemical ionization; ESI, electrospray ionization; THCCOOH, 11-nor-9-carboxy-THC; 11-OH-THC, 11-hydroxy-THC; GC-MS/MS, gas chromatography–tandem mass spectrometry; CBD, cannabidiol; CBN, cannabinol; MDMA, 3,4-methylenedioxymethamphetamine; MDA, 3,4-methylenedioxyamphetamine; EME, ecgonine methylester.
Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article.
Authors’ Disclosures of Potential Conflicts of Interest: Upon manuscript submission, all authors completed the Disclosures of Potential Conflict of Interest form. Potential conflicts of interest:
Employment or Leadership: None declared.
Consultant or Advisory Role: None declared.
Stock Ownership: None declared.
Honoraria: None declared.
Expert Testimony: None declared.
References
- 1.Choo RE, Huestis MA. Oral fluid as a diagnostic tool. Clin Chem Lab Med. 2004;42:1273–87. doi: 10.1515/CCLM.2004.248. [DOI] [PubMed] [Google Scholar]
- 2.Gallardo E, Queiroz JA. The role of alternative specimens in toxicological analysis. Biomed Chromatogr. 2008;22:795–821. doi: 10.1002/bmc.1009. [DOI] [PubMed] [Google Scholar]
- 3.Langel K, Engblom C, Pehrsson A, Gunnar T, Ariniemi K, Lillsunde P. Drug testing in oral fluid-evaluation of sample collection devices. J Anal Toxicol. 2008;32:393–401. doi: 10.1093/jat/32.6.393. [DOI] [PubMed] [Google Scholar]
- 4.Verstraete AG. Oral fluid testing for driving under the influence of drugs: history, recent progress and remaining challenges. Forensic Sci Int. 2005;150:143–50. doi: 10.1016/j.forsciint.2004.11.023. [DOI] [PubMed] [Google Scholar]
- 5.Department of Health and Human Services, SAMHSA. Proposed revisions to mandatory guidelines for federal workplace drug testing programs. Federal Register. 2004;69:19673–732. [Google Scholar]
- 6.Bush DM. The US mandatory guidelines for federal workplace drug testing programs: current status and future considerations. Forensic Sci Int. 2008;174:111–9. doi: 10.1016/j.forsciint.2007.03.008. [DOI] [PubMed] [Google Scholar]
- 7.National Institute on Drug Abuse. [Accessed October 2008];Drugged driving info facts. www.nida.nih.gov.
- 8.Verstraete A, Puddu M. [Accessed February 2009];Evaluation of different roadside drug tests. www.rosita.org.
- 9.Verstraete A, Raes E. [Accessed February 2009];Rosita-2 project: final report. www.rosita.org.
- 10.Drummer OH, Gerostamoulos D, Chu M, Swann P, Boorman M, Cairns I. Drugs in oral fluid in randomly selected drivers. Forensic Sci Int. 2007;170:105–10. doi: 10.1016/j.forsciint.2007.03.028. [DOI] [PubMed] [Google Scholar]
- 11.Walsh JM, Verstraete AG, Huestis MA, Morland J. Guidelines for research on drugged driving. Addiction. 2008;103:1258–68. doi: 10.1111/j.1360-0443.2008.02277.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Crouch DJ. Oral fluid collection: the neglected variable in oral fluid testing. Forensic Sci Int. 2005;150:165–73. doi: 10.1016/j.forsciint.2005.02.028. [DOI] [PubMed] [Google Scholar]
- 13.Huestis MA, Cone EJ. Relationship of Delta 9-tetrahydrocannabinol concentrations in oral fluid and plasma after controlled administration of smoked cannabis. J Anal Toxicol. 2004;28:394–9. doi: 10.1093/jat/28.6.394. [DOI] [PubMed] [Google Scholar]
- 14.Kintz P, Brunet B, Muller JF, Serra W, Villain M, Cirimele V, Mura P. Evaluation of the Cozart DDSV test for cannabis in oral fluid. Ther Drug Monit. 2009;31:131–4. doi: 10.1097/FTD.0b013e3181947916. [DOI] [PubMed] [Google Scholar]
- 15.Dickson S, Park A, Nolan S, Kenworthy S, Nicholson C, Midgley J, et al. The recovery of illicit drugs from oral fluid sampling devices. Forensic Sci Int. 2007;165:78–84. doi: 10.1016/j.forsciint.2006.03.004. [DOI] [PubMed] [Google Scholar]
- 16.Kauert GF, Iwersen-Bergmann S, Toennes SW. Assay of Delta9-tetrahydrocannabinol (THC) in oral fluid-evaluation of the OraSure oral specimen collection device. J Anal Toxicol. 2006;30:274–7. doi: 10.1093/jat/30.4.274. [DOI] [PubMed] [Google Scholar]
- 17.Concheiro M, de Castro A, Quintela O, Cruz A, Lopez-Rivadulla M. Confirmation by LC-MS of drugs in oral fluid obtained from roadside testing. Forensic Sci Int. 2007;170:156–62. doi: 10.1016/j.forsciint.2007.03.034. [DOI] [PubMed] [Google Scholar]
- 18.Crouch DJ, Walsh JM, Cangianelli L, Quintela O. Laboratory evaluation and field application of roadside oral fluid collectors and drug testing devices. Ther Drug Monit. 2008;30:188–95. doi: 10.1097/FTD.0b013e3181679249. [DOI] [PubMed] [Google Scholar]
- 19.Biermann T, Schwarze B, Zedler B, Betz P. On-site testing of illicit drugs: the use of the drug-testing device “Toxiquick. Forensic Sci Int. 2004;143:21–5. doi: 10.1016/j.forsciint.2004.01.013. [DOI] [PubMed] [Google Scholar]
- 20.Walsh JM, Crouch DJ, Danaceau JP, Cangianelli L, Liddicoat L, Adkins R. Evaluation of ten oral fluid point-of-collection drug-testing devices. J Anal Toxicol. 2007;31:44–54. doi: 10.1093/jat/31.1.44. [DOI] [PubMed] [Google Scholar]
- 21.Cooper G, Wilson L, Reid C, Main L, Hand C. Evaluation of the Cozart RapiScan drug test system for opiates and cocaine in oral fluid. Forensic Sci Int. 2005;150:239–43. doi: 10.1016/j.forsciint.2004.12.042. [DOI] [PubMed] [Google Scholar]
- 22.Cooper G, Wilson L, Reid C, Baldwin D, Hand C, Spiehler V. Validation of the Cozart® microplate EIA for analysis of opiates in oral fluid. Forensic Sci Int. 2005;154:240–6. doi: 10.1016/j.forsciint.2004.10.013. [DOI] [PubMed] [Google Scholar]
- 23.Cooper G, Wilson L, Reid C, Baldwin D, Hand C, Spiehler V. Validation of the Cozart microplate EIA for cocaine and metabolites in oral fluid. J Anal Toxicol. 2004;28:498–503. doi: 10.1093/jat/28.6.498. [DOI] [PubMed] [Google Scholar]
- 24.Cooper G, Wilson L, Reid C, Hand C, Spiehler V. Validation of the Cozart® Amphetamine Microplate EIA for the analysis of amphetamines in oral fluid. Forensic Sci Int. 2006;159:104–12. doi: 10.1016/j.forsciint.2005.06.011. [DOI] [PubMed] [Google Scholar]
- 25.Wong RC, Tran M, Tung JK. Oral fluid drug tests: effects of adulterants and foodstuffs. Forensic Sci Int. 2005;150:175–80. doi: 10.1016/j.forsciint.2005.02.023. [DOI] [PubMed] [Google Scholar]
- 26.Dams R, Murphy CM, Choo RE, Lambert WE, De Leenheer AP, Huestis MA. LC-atmospheric pressure chemical ionization-MS/MS analysis of multiple illicit drugs, methadone, and their metabolites in oral fluid following protein precipitation. Anal Chem. 2003;75:798–804. doi: 10.1021/ac026111t. [DOI] [PubMed] [Google Scholar]
- 27.Maurer HH. Advances in analytical toxicology: the current role of liquid chromatography-mass spectrometry in drug quantification in blood and oral fluid. Anal Bioanal Chem. 2005;381:110–8. doi: 10.1007/s00216-004-2774-z. [DOI] [PubMed] [Google Scholar]
- 28.Samyn N, Laloup M, De Boeck G. Bioanalytical procedures for determination of drugs of abuse in oral fluid. Anal Bioanal Chem. 2007;388:1437–53. doi: 10.1007/s00216-007-1245-8. [DOI] [PubMed] [Google Scholar]
- 29.Wood M, Laloup M, Samyn N, Fernandez MDR, de Bruijn EA, Maes RAA, De Boeck G. Recent applications of liquid chromatography-mass spectrometry in forensic science. J Chromatogr. 2006;1130:3–15. doi: 10.1016/j.chroma.2006.04.084. [DOI] [PubMed] [Google Scholar]
- 30.Dams R, Huestis MA, Lambert WE, Murphy CM. Matrix effect in bio-analysis of illicit drugs with LC-MS/MS: influence of ionization type, sample preparation, and biofluid. J Am Soc Mass Spectrom. 2003;14:1290–4. doi: 10.1016/S1044-0305(03)00574-9. [DOI] [PubMed] [Google Scholar]
- 31.Cone EJ, Huestis MA. Interpretation of oral fluid tests for drugs of abuse. Ann NY Acad Sci. 2007;1098:51–103. doi: 10.1196/annals.1384.037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Speedy T, Baldwin D, Jowett G, Gallina M, Jehanli A. Development and validation of the Cozart® DDS oral fluid collection device. Forensic Sci Int. 2007;170:117–20. doi: 10.1016/j.forsciint.2007.03.038. [DOI] [PubMed] [Google Scholar]
- 33.Quintela O, Crouch DJ, Andrenyak DM. Recovery of drugs of abuse from the Immunalysis Quantisal oral fluid collection device. J Anal Toxicol. 2006;30:614–6. doi: 10.1093/jat/30.8.614. [DOI] [PubMed] [Google Scholar]
- 34.Hawks RL, Chiang NC. Urine testing for drugs of abuse. National Institute on Drug Abuse; 1986. [Accessed November 2008]. http://www.drugabuse.gov/pdf/monographs/73.pdf. [Google Scholar]
- 35.Huestis MA, Dickerson S, Cone EJ. Can saliva THC levels be correlated to behavior?. From: American Academy of Forensic Sciences annual meeting; 1992. p. 190. [Google Scholar]
- 36.Ramaekers JG, Moeller MR, van Ruitenbeek P, Theunissen EL, Schneider E, Kauert G. Cognition and motor control as a function of Delta9-THC concentration in serum and oral fluid: limits of impairment. Drug Alcohol Depend. 2006;85:114–22. doi: 10.1016/j.drugalcdep.2006.03.015. [DOI] [PubMed] [Google Scholar]
- 37.Kauert GF, Ramaekers JG, Schneider E, Moeller MR, Toennes SW. Pharmacokinetic properties of Delta9-tetrahydrocannabinol in serum and oral fluid. J Anal Toxicol. 2007;31:288–93. doi: 10.1093/jat/31.5.288. [DOI] [PubMed] [Google Scholar]
- 38.Niedbala S, Kardos K, Salamone S, Fritch D, Bronsgeest M, Cone EJ. Passive cannabis smoke exposure and oral fluid testing. J Anal Toxicol. 2004;28:546–52. doi: 10.1093/jat/28.7.546. [DOI] [PubMed] [Google Scholar]
- 39.Niedbala RS, Kardos KW, Fritch DF, Kunsman KP, Blum KA, Newland GA, et al. Passive cannabis smoke exposure and oral fluid testing, II: two studies of extreme cannabis smoke exposure in a motor vehicle. J Anal Toxicol. 2005;29:607–15. doi: 10.1093/jat/29.7.607. [DOI] [PubMed] [Google Scholar]
- 40.Moore C, Coulter C, Rana S, Vincent M, Soares J. Analytical procedure for the determination of the marijuana metabolite 11-nor-Delta9-tetrahydrocannabinol-9-carboxylic acid in oral fluid specimens. J Anal Toxicol. 2006;30:409–12. doi: 10.1093/jat/30.7.409. [DOI] [PubMed] [Google Scholar]
- 41.Moore C, Ross W, Coulter C, Adams L, Rana S, Vincent M, Soares J. Detection of the marijuana metabolite 11-nor-Delta9-tetrahydrocannabinol-9-carboxylic acid in oral fluid specimens and its contribution to positive results in screening assays. J Anal Toxicol. 2006;30:413–8. doi: 10.1093/jat/30.7.413. [DOI] [PubMed] [Google Scholar]
- 42.Moore C, Rana S, Coulter C, Day D, Vincent M, Soares J. Detection of conjugated 11-nor-Delta9-tetrahydrocannabinol-9-carboxylic acid in oral fluid. J Anal Toxicol. 2007;31:187–94. doi: 10.1093/jat/31.4.187. [DOI] [PubMed] [Google Scholar]
- 43.Milman G, Lowe RH, Huestis MA. Simultaneous identification and quantification of Cannabis sativa cannabinoids and metabolites in oral fluid. From: Society of Forensic Toxicologists annual meeting; 2008 Oct 27–31; Phoenix, AZ. [Google Scholar]
- 44.Niedbala RS, Kardos KW, Fritch DF, Kardos S, Fries T, Waga J, et al. Detection of marijuana use by oral fluid and urine analysis following single-dose administration of smoked and oral marijuana. J Anal Toxicol. 2001;25:289–303. doi: 10.1093/jat/25.5.289. [DOI] [PubMed] [Google Scholar]
- 45.Walsh JM. New technology and new initiatives in US workplace testing. Forensic Sci Int. 2008;174:120–4. doi: 10.1016/j.forsciint.2007.03.011. [DOI] [PubMed] [Google Scholar]
- 46.Moore C, Rana S, Coulter C. Determination of meperidine, tramadol and oxycodone in human oral fluid using solid phase extraction and gas chromatography-mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2007;850:370–5. doi: 10.1016/j.jchromb.2006.12.008. [DOI] [PubMed] [Google Scholar]
- 47.De Giovanni N, Fucci N, Scarlata S, Donzelli G. Buprenorphine detection in biological samples. Clin Chem Lab Med. 2005;43:1377–9. doi: 10.1515/CCLM.2005.235. [DOI] [PubMed] [Google Scholar]
- 48.Kim I, Barnes AJ, Oyler JM, Schepers R, Joseph RE, Jr, Cone EJ, et al. Plasma and oral fluid pharmacokinetics and pharmacodynamics after oral codeine administration. Clin Chem. 2002;48:1486–96. [PubMed] [Google Scholar]
- 49.Kacinko SL, Barnes AJ, Kim I, Moolchan ET, Wilson L, Cooper GA, et al. Performance characteristics of the Cozart® RapiScan Oral Fluid Drug Testing System for opiates in comparison to ELISA and GC/MS following controlled codeine administration. Forensic Sci Int. 2004;141:41–8. doi: 10.1016/j.forsciint.2003.12.003. [DOI] [PubMed] [Google Scholar]
- 50.Rohrig TP, Moore C. The determination of morphine in urine and oral fluid following ingestion of poppy seeds. J Anal Toxicol. 2003;27:449–52. doi: 10.1093/jat/27.7.449. [DOI] [PubMed] [Google Scholar]
- 51.Schepers RJF, Oyler JM, Joseph RE, Jr, Cone EJ, Moolchan ET, Huestis MA. Methamphetamine and amphetamine pharmacokinetics in oral fluid and plasma after controlled oral methamphet-amine administration to human volunteers. Clin Chem. 2003;49:121–32. doi: 10.1373/49.1.121. [DOI] [PubMed] [Google Scholar]
- 52.Huestis MA, Cone EJ. Methamphetamine disposition in oral fluid, plasma, and urine. Ann NY Acad Sci. 2007;1098:104–21. doi: 10.1196/annals.1384.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Laloup M, Tilman G, Maes V, De Boeck G, Wallemacq P, Ramaekers J, Samyn N. Validation of an ELISA-based screening assay for the detection of amphetamine, MDMA and MDA in blood and oral fluid. Forensic Sci Int. 2005;153:29–37. doi: 10.1016/j.forsciint.2005.04.019. [DOI] [PubMed] [Google Scholar]
- 54.Wood M, Laloup M, Ramirez Fernandez Mdel M, Jenkins KM, Young MS, Ramaekers JG, et al. Quantitative analysis of multiple illicit drugs in preserved oral fluid by solid-phase extraction and liquid chromatography-tandem mass spectrometry. Forensic Sci Int. 2005;150:227–38. doi: 10.1016/j.forsciint.2004.11.027. [DOI] [PubMed] [Google Scholar]
- 55.Peters FT, Samyn N, Kraemer T, Riedel WJ, Maurer HH. Negative-ion chemical ionization gas chromatography-mass spectrometry assay for enantioselective measurement of amphetamines in oral fluid: application to a controlled study with MDMA and driving under the influence cases. Clin Chem. 2007;53:702–10. doi: 10.1373/clinchem.2006.081547. [DOI] [PubMed] [Google Scholar]
- 56.European Monitoring Centre for Drugs and Drug Addiction. [Accessed December 2008];Annual report: the state of the drugs problem in Europe. www.emcdda.europa.eu.
- 57.Kolbrich E, Barnes A, Gorelick DA, Boyd SJ, Cone EJ, Huestis MA. Major and minor metabolites of cocaine in human plasma following controlled subcutaneous cocaine administration. J Anal Toxicol. 2006;30:501–10. doi: 10.1093/jat/30.8.501. [DOI] [PubMed] [Google Scholar]
- 58.Kolbrich EA, Kelly TL, Barnes AJ, Kim I, Moolchan ET, Cone EJ, Huestis MA. Saliva/plasma ratios of cocaine, ecgonine methyl ester and benzoylecgonine following controlled cocaine administration. From: Society of Forensic Toxicologists annual meeting; 2005 Oct 17–21; Nashville, TN. [Google Scholar]
- 59.Jufer R, Walsh SL, Cone EJ, Sampson-Cone A. Effect of repeated cocaine administration on detection times in oral fluid and urine. J Anal Toxicol. 2006;30:458–62. doi: 10.1093/jat/30.7.458. [DOI] [PubMed] [Google Scholar]
- 60.Busto UE, Kaplan HL, Wright CE, Gomez-Mancilla B, Zawertailo L, Greenblatt DJ, Sellers EM. A comparative pharmacokinetic and dynamic evaluation of alprazolam sustained-release, bromazepam, and lorazepam. J Clin Psychopharmacol. 2000;20:628–35. doi: 10.1097/00004714-200012000-00008. [DOI] [PubMed] [Google Scholar]
- 61.Samyn N, DeBoeck G, Cirimele V, Verstraete A, Kintz P. Detection of flunitrazepam and 7-aminoflunitrazepam in oral fluid after controlled administration of Rohypnol. J Anal Toxicol. 2002;26:211–5. doi: 10.1093/jat/26.4.211. [DOI] [PubMed] [Google Scholar]
- 62.Laloup M, Fernandez MDR, Wood M, Maes V, De Boeck G, Vanbeckevoort Y, Samyn N. Detection of diazepam in urine, hair and preserved oral fluid samples with LC-MS-MS after single and repeated administration of Myolastan (R) and Valium (R) Anal Bioanal Chem. 2007;388:1545–56. doi: 10.1007/s00216-007-1297-9. [DOI] [PubMed] [Google Scholar]
- 63.Smink BE, Hofman BJ, Dijkhuizen A, Lusthof KJ, de Gier JJ, Egberts AC, Uges DR. The concentration of oxazepam and oxazepam glucuronide in oral fluid, blood and serum after controlled administration of 15 and 30 mg oxazepam. Br J Clin Pharmacol. 2008;66:556–60. doi: 10.1111/j.1365-2125.2008.03252.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Toennes SW, Kauert GF, Steinmeyer S, Moeller MR. Driving under the influence of drugs—evaluation of analytical data of drugs in oral fluid, serum and urine, and correlation with impairment symptoms. Forensic Sci Int. 2005;152:149–55. doi: 10.1016/j.forsciint.2004.08.002. [DOI] [PubMed] [Google Scholar]
- 65.Toennes SW, Steinmeyer S, Maurer HJ, Moeller MR, Kauert GF. Screening for drugs of abuse in oral fluid— correlation of analysis results with serum in forensic cases. J Anal Toxicol. 2005;29:22–7. doi: 10.1093/jat/29.1.22. [DOI] [PubMed] [Google Scholar]
- 66.Engblom C, Gunnar T, Rantanen A, Lillsunde P. Driving under the influence of drugs—amphetamine concentrations in oral fluid and whole blood samples. J Anal Toxicol. 2007;31:276–80. doi: 10.1093/jat/31.5.276. [DOI] [PubMed] [Google Scholar]
- 67.Pehrsson A, Gunnar T, Engblom C, Seppä H, Jama A, Lillsunde P. Roadside oral fluid testing: comparison of the results of Drugwipe 5 and Drugwipe Benzodiazepines on-site tests with laboratory confirmation results of oral fluid and whole blood. Forensic Sci Int. 2008;175:140–8. doi: 10.1016/j.forsciint.2007.05.022. [DOI] [PubMed] [Google Scholar]
- 68.Gjerde H, Normann PT, Pettersen BS, Assum T, Aldrin M, Johansen U, et al. Prevalence of alcohol and drugs among Norwegian motor vehicle drivers: a roadside survey. Accid Anal Prev. 2008;40:1765–72. doi: 10.1016/j.aap.2008.06.015. [DOI] [PubMed] [Google Scholar]
- 69.Wylie FM, Torrance H, Seymour A, Buttress S, Oliver JS. Drugs in oral fluid—part II: investigation of drugs in drivers. Forensic Sci Int. 2005;150:199–204. doi: 10.1016/j.forsciint.2005.02.025. [DOI] [PubMed] [Google Scholar]
- 70.Bernhoft IM, Steentoft A, Johansen SS, Klitgaard NA, Larsen LB, Hansen LB. Drugs in injured drivers in Denmark. Forensic Sci Int. 2005;150:181–9. doi: 10.1016/j.forsciint.2004.12.039. [DOI] [PubMed] [Google Scholar]
- 71.Kintz P, Bernhard W, Villain M, Gasser M, Aebi B, Cirimele V. Detection of cannabis use in drivers with the Drugwipe device and by GC-MS after Intercept device collection. J Anal Toxicol. 2005;29:724–7. doi: 10.1093/jat/29.7.724. [DOI] [PubMed] [Google Scholar]
- 72.Ojaniemi KK, Lintonen TP, Impinen AO, Lillsunde PM, Ostamo AI. Trends in driving under the influence of drugs: a register-based study of DUID suspects during 1977–2007. Accid Anal Prev. 2009;41:191–6. doi: 10.1016/j.aap.2008.10.011. [DOI] [PubMed] [Google Scholar]
- 73.Cook JM, Marshall R, Masci C, Coyne JC. Physicians’ perspectives on prescribing benzodiazepines for older adults: a qualitative study. J Gen Intern Med. 2007;22:303–7. doi: 10.1007/s11606-006-0021-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Cone EJ, Clarke J, Tsanaclis L. Prevalence and disposition of drugs of abuse and opioid treatment drugs in oral fluid. J Anal Toxicol. 2007;31:424–33. doi: 10.1093/jat/31.8.424. [DOI] [PubMed] [Google Scholar]
- 75.Pil K, Raes E, Van den Neste T, Verstraete A. Toxicological analyses in the DRUID epidemiological studies: analytical methods, target analytes and analytical cut-offs. From: The International Association of Forensic Toxicology annual meeting; 2007 Aug 27–30; Seattle. [Google Scholar]
- 76.Clarke J, Wilson JF. Proficiency testing (external quality assessment) of drug detection in oral fluid. Forensic Sci Int. 2005;150:161–4. doi: 10.1016/j.forsciint.2004.11.025. [DOI] [PubMed] [Google Scholar]
- 77.Ananias D, Mountain L, Zhao Y, Blum K, Fritch D, Kardos K, Lee S. Automated immunoassays for the detection of opiates, amphetamines and methamphetamines in oral fluid on Roche instrument panels. Ther Drug Monit. 2007;29:493. [Google Scholar]
- 78.Barnes AJ, Kim I, Schepers R, Moolchan ET, Wilson L, Cooper G, et al. Sensitivity, specificity, and efficiency in detecting opiates in oral fluid with the Cozart opiate microplate EIA and GC-MS following controlled codeine administration. J Anal Toxicol. 2003;27:402–6. doi: 10.1093/jat/27.7.402. [DOI] [PubMed] [Google Scholar]
- 79.Cooper G, Wilson L, Reid C, Baldwin D, Hand C, Spiehler V. Comparison of GC-MS and EIA results for the analysis of methadone in oral fluid. J Forensic Sci. 2005;50:928–32. [PubMed] [Google Scholar]
- 80.Apollonio LG, Whittall IR, Pianca DJ, Kyd JM, Maher WA. Matrix effect and cross-reactivity of select amphetamine-type substances, designer analogues, and putrefactive amines using the Bio-Quant Direct ELISA presumptive assays for amphetamine and methamphetamine. J Anal Toxicol. 2007;31:208–13. doi: 10.1093/jat/31.4.208. [DOI] [PubMed] [Google Scholar]
- 81.Kim I, Barnes AJ, Schepers R, Moolchan ET, Wilson L, Cooper G, et al. Sensitivity and specificity of the Cozart microplate EIA cocaine oral fluid at proposed screening and confirmation cutoffs. Clin Chem. 2003;49:1498–503. doi: 10.1373/49.9.1498. [DOI] [PubMed] [Google Scholar]
- 82.Kolbrich EA, Kim I, Barnes AJ, Moolchan ET, Wilson L, Cooper GA, et al. Cozart RapiScan oral fluid drug testing system: an evaluation of sensitivity, specificity, and efficiency for cocaine detection compared with ELISA and GC-MS following controlled cocaine administration. J Anal Toxicol. 2003;27:407–11. doi: 10.1093/jat/27.7.407. [DOI] [PubMed] [Google Scholar]
- 83.Smink BE, Mathijssen MPM, Lusthof KJ, de Gier JJ, Egberts ACG, Uges DRA. Comparison of urine and oral fluid as matrices for screening of thirty-three benzodiazepines and benzodiazepine-like substances using immunoassay and LC-MS(-MS) J Anal Toxicol. 2006;30:478–85. doi: 10.1093/jat/30.7.478. [DOI] [PubMed] [Google Scholar]
- 84.Crouch DJ, Walsh JM, Flegel R, Cangianelli L, Baudys J, Atkins R. An evaluation of selected oral fluid point-of-collection drug-testing devices. J Anal Toxicol. 2005;29:244–8. doi: 10.1093/jat/29.4.244. [DOI] [PubMed] [Google Scholar]
- 85.Wilson L, Jehanli A, Hand C, Cooper G, Smith R. Evaluation of a rapid oral fluid point-of-care test for MDMA. J Anal Toxicol. 2007;31:98–104. doi: 10.1093/jat/31.2.98. [DOI] [PubMed] [Google Scholar]
- 86.Laloup M, Del Mar Ramirez Fernandez M, Wood M, De Boeck G, Maes V, Samyn N. Correlation of Delta9-tetrahydrocannabinol concentrations determined by LC-MS-MS in oral fluid and plasma from impaired drivers and evaluation of the on-site Drager DrugTest. Forensic Sci Int. 2006;161:175–9. doi: 10.1016/j.forsciint.2006.03.033. [DOI] [PubMed] [Google Scholar]
- 87.Cirimele V, Villain M, Mura P, Bernard M, Kintz P. Oral fluid testing for cannabis: on-site Ora-Line IV s.a.t. device versus GC/MS. Forensic Sci Int. 2006;161:180–4. doi: 10.1016/j.forsciint.2006.01.020. [DOI] [PubMed] [Google Scholar]
- 88.Kempf J, Wuske T, Schubert R, Weinmann W. Pre-analytical stability of selected benzodiazepines on a polymeric oral fluid sampling device. Forensic Sci Int. 2009;186:81–5. doi: 10.1016/j.forsciint.2009.02.004. [DOI] [PubMed] [Google Scholar]
- 89.Dams R, Choo RE, Lambert WE, Jones H, Huestis MA. Oral fluid as an alternative matrix to monitor opiate and cocaine use in substance-abuse treatment patients. Drug Alcohol Depend. 2007;87:258–67. doi: 10.1016/j.drugalcdep.2006.08.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Gunnar T, Ariniemi K, Lillsunde P. Validated toxicological determination of 30 drugs of abuse as optimized derivatives in oral fluid by long column fast gas chromatography/electron impact mass spectrometry. J Mass Spectrom. 2005;40:739–53. doi: 10.1002/jms.846. [DOI] [PubMed] [Google Scholar]
- 91.Wylie FM, Torrance H, Anderson RA, Oliver JS. Drugs in oral fluid, part I: validation of an analytical procedure for licit and illicit drugs in oral fluid. Forensic Sci Int. 2005;150:191–8. doi: 10.1016/j.forsciint.2005.02.024. [DOI] [PubMed] [Google Scholar]
- 92.Øiestad EL, Johansen U, Christophersen AS. Drug screening of preserved oral fluid by liquid chromatography-tandem mass spectrometry. Clin Chem. 2007;53:300–9. doi: 10.1373/clinchem.2006.074237. [DOI] [PubMed] [Google Scholar]
- 93.Pujadas M, Pichini S, Civit E, Santamariña E, Perez K, de la Torre R. A simple and reliable procedure for the determination of psychoactive drugs in oral fluid by gas chromatography-mass spectrometry. J Pharm Biomed Anal. 2007;44:594–601. doi: 10.1016/j.jpba.2007.02.022. [DOI] [PubMed] [Google Scholar]
- 94.Concheiro M, de Castro A, Quintela O, Cruz A, López-Rivadulla M. Determination of illicit and medicinal drugs and their metabolites in oral fluid and preserved oral fluid by liquid chromatography-tandem mass spectrometry. Anal Bioanal Chem. 2008;391:2329–38. doi: 10.1007/s00216-008-2135-4. [DOI] [PubMed] [Google Scholar]
- 95.Fernandez P, Morales L, Vazquez C, Lago M, Bermejo AM. Comparison of two extraction procedures for determination of drugs of abuse in human saliva by high-performance liquid chromatography. J Appl Toxicol. 2008;28:998–1003. doi: 10.1002/jat.1365. [DOI] [PubMed] [Google Scholar]
- 96.Kala SV, Harris SE, Freijo TD, Gerlich S. Validation of analysis of amphetamines, opiates, phencyclidine, cocaine, and benzoylecgonine in oral fluids by liquid chromatography-tandem mass spectrometry. J Anal Toxicol. 2008;32:605–11. doi: 10.1093/jat/32.8.605. [DOI] [PubMed] [Google Scholar]
- 97.Sergi M, Bafile E, Compagnone D, Curini R, D’Ascenzo G, Romolo FS. Multiclass analysis of illicit drugs in plasma and oral fluids by LC-MS/MS. Anal Bioanal Chem. 2009;393:709–18. doi: 10.1007/s00216-008-2456-3. [DOI] [PubMed] [Google Scholar]
- 98.Concheiro M, de Castro A, Quintela O, Cruz A, Lopez-Rivadulla M. Development and validation of a method for the quantitation of Delta9-tetrahydrocannabinol in oral fluid by liquid chromatography electrospray-mass-spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2004;810:319–24. doi: 10.1016/j.jchromb.2004.08.017. [DOI] [PubMed] [Google Scholar]
- 99.Teixeira H, Proenca P, Castanheira A, Santos S, Lopez-Rivadulla M, Corte-Real F, et al. Cannabis and driving: the use of LC-MS to detect delta9-tetrahydrocannabinol (delta9-THC) in oral fluid samples. Forensic Sci Int. 2004;146(Suppl):S61–3. doi: 10.1016/j.forsciint.2004.09.002. [DOI] [PubMed] [Google Scholar]
- 100.Laloup M, Ramirez Fernandez Mdel M, Wood M, De Boeck G, Henquet C, Maes V, Samyn N. Quantitative analysis of delta9-tetrahydrocannabinol in preserved oral fluid by liquid chromatography-tandem mass spectrometry. J Chromatogr. 2005;1082:15–24. doi: 10.1016/j.chroma.2005.03.031. [DOI] [PubMed] [Google Scholar]
- 101.Teixeira H, Proenca P, Verstraete A, Corte-Real F, Vieira DN. Analysis of Delta-9-tetrahydrocannabinol in oral fluid samples using solid-phase extraction and high-performance liquid chromatography-electrospray ionization mass spectrometry. Forensic Sci Int. 2005;150:205–11. doi: 10.1016/j.forsciint.2004.11.026. [DOI] [PubMed] [Google Scholar]
- 102.Day D, Kuntz DJ, Feldman M, Presley L. Detection of THCA in oral fluid by GC-MS-MS. J Anal Toxicol. 2006;30:645–50. doi: 10.1093/jat/30.9.645. [DOI] [PubMed] [Google Scholar]
- 103.Moore C, Rana S, Coulter C. Simultaneous identification of 2-carboxy-tetrahydrocannabinol, tetrahydrocannabinol, cannabinol and cannabidiol in oral fluid. J Chromatogr B Analyt Technol Biomed Life Sci. 2007;852:459–64. doi: 10.1016/j.jchromb.2007.02.016. [DOI] [PubMed] [Google Scholar]
- 104.Quintela O, Andrenyak DM, Hoggan AM, Crouch DJ. A validated method for the detection of Delta(9)-tetrahydrocannabinol and 11-nor-9-carboxy-Delta(9)-tetrahydrocannabinol in oral fluid samples by liquid chromatography coupled with quadrupole-time-of-flight mass spectrometry. J Anal Toxicol. 2007;31:157–64. doi: 10.1093/jat/31.3.157. [DOI] [PubMed] [Google Scholar]
- 105.Teixeira H, Verstraete A, Proenca P, Corte-Real F, Monsanto P, Vieira DN. Validated method for the simultaneous determination of Delta9-THC and Delta9-THC-COOH in oral fluid, urine and whole blood using solid-phase extraction and liquid chromatography-mass spectrometry with electrospray ionization. Forensic Sci Int. 2007;170:148–55. doi: 10.1016/j.forsciint.2007.03.026. [DOI] [PubMed] [Google Scholar]
- 106.Campora P, Bermejo AM, Tabernero MJ, Fernandez P. Use of gas chromatography/mass spectrometry with positive chemical ionization for the determination of opiates in human oral fluid. Rapid Commun Mass Spectrom. 2006;20:1288–92. doi: 10.1002/rcm.2449. [DOI] [PubMed] [Google Scholar]
- 107.Phillips SG, Allen KR. Acetylcodeine as a marker of illicit heroin abuse in oral fluid samples. J Anal Toxicol. 2006;30:370–4. doi: 10.1093/jat/30.6.370. [DOI] [PubMed] [Google Scholar]
- 108.Rana S, Moore C, Vincent M, Coulter C, Agrawal A, Soares J. Determination of propoxyphene in oral fluid. J Anal Toxicol. 2006;30:516–8. doi: 10.1093/jat/30.8.516. [DOI] [PubMed] [Google Scholar]
- 109.Martins LF, Yegles M, Wennig R. Simultaneous enantio selective quantification of methadone and of 2-ethylidene-1,5-dimethyl-3,3-diphenyl-pyrrolidine in oral fluid using capillary electrophoresis. J Chromatogr B Analyt Technol Biomed Life Sci. 2008;862:79–85. doi: 10.1016/j.jchromb.2007.11.001. [DOI] [PubMed] [Google Scholar]
- 110.Kankaanpää A, Gunnar T, Ariniemi K, Lillsunde P, Mykkänen S, Seppälä T. Single-step procedure for gas chromatography-mass spectrometry screening and quantitative determination of amphetamine-type stimulants and related drugs in blood, serum, oral fluid and urine samples. J Chromatogr B Analyt Technol Biomed Life Sci. 2004;810:57–68. doi: 10.1016/j.jchromb.2004.07.035. [DOI] [PubMed] [Google Scholar]
- 111.Concheiro M, de Castro A, Quintela O, Lopez-Rivadulla M, Cruz A. Determination of MDMA, MDA, MDEA and MBDB in oral fluid using high performance liquid chromatography with native fluorescence detection. Forensic Sci Int. 2005;150:221–6. doi: 10.1016/j.forsciint.2004.12.041. [DOI] [PubMed] [Google Scholar]
- 112.Scheidweiler KB, Huestis MA. Monitoring of ecstasy (MDMA) and three metabolites in oral fluid. From: The National Institute on Drug Abuse Frontiers in Addiction Research satellite event at the Society for Neuroscience annual meeting; 2005 Nov 11; Washington, DC. [Google Scholar]
- 113.Scheidweiler KB, Huestis MA. A validated gas chromatographic-electron impact ionization mass spectrometric method for methylenedioxymethamphetamine (MDMA), methamphetamine and metabolites in oral fluid. J Chromatogr B Analyt Technol Biomed Life Sci. 2006;835:90–9. doi: 10.1016/j.jchromb.2006.03.020. [DOI] [PubMed] [Google Scholar]
- 114.Moore C, Coulter C, Crompton K. Achieving proposal federal concentration using reduced specimen volume for the extraction of amphetamines from oral fluid. J Anal Toxicol. 2007;31:442–6. doi: 10.1093/jat/31.8.442. [DOI] [PubMed] [Google Scholar]
- 115.Clauwaert K, Decaestecker T, Mortier K, Lambert W, Deforce D, Van Peteghem C, Van Bocxlaer J. The determination of cocaine, benzoylecgonine, and cocaethylene in small-volume oral fluid samples by liquid chromatography-quadrupole-time-of-flight mass spectrometry. J Anal Toxicol. 2004;28:655–9. doi: 10.1093/jat/28.8.655. [DOI] [PubMed] [Google Scholar]
- 116.Robarge T, Phillips E, Edwards J, Seiter W, Presley L, Hollaway D. Positive ion chemical ionization and gas chromatography/quadrupole mass spectrometry for confirmation and quantitation of benzoylecgonine in oral fluid. LCGC Europe; 2005. Dec 2, [Google Scholar]
- 117.Cognard E, Bouchonnet S, Staub C. Validation of a gas chromatography–ion trap tandem mass spectrometry for simultaneous analyse of cocaine and its metabolites in saliva. J Pharm Biomed Anal. 2006;41:925–34. doi: 10.1016/j.jpba.2006.01.041. [DOI] [PubMed] [Google Scholar]
- 118.Quintela O, Cruz A, Concheiro M, de Castro A, Lopez-Rivadulla M. A sensitive, rapid and specific determination of midazolam in human plasma and saliva by liquid chromatography/electrospray mass spectrometry. Rapid Commun Mass Spectrom. 2004;18:2976–82. doi: 10.1002/rcm.1719. [DOI] [PubMed] [Google Scholar]
- 119.Kintz P, Villain M, Concheiro M, Cirimele V. Screening and confirmatory method for benzodiazepines and hypnotics in oral fluid by LC-MS/MS. Forensic Sci Int. 2005;150:213–20. doi: 10.1016/j.forsciint.2004.12.040. [DOI] [PubMed] [Google Scholar]
- 120.Quintela O, Cruz A, Castro A, Concheiro M, Lopez-Rivadulla M. Liquid chromatography-electrospray ionization mass spectrometry for the determination of nine selected benzodiazepines in human plasma and oral fluid. J Chromatogr B Analyt Technol Biomed Life Sci. 2005;825:63–71. doi: 10.1016/j.jchromb.2004.12.038. [DOI] [PubMed] [Google Scholar]
- 121.Link B, Haschke M, Wenk M, Krahenbuhl S. Determination of midazolam and its hydroxy metabolites in human plasma and oral fluid by liquid chromatography/electrospray ionization ion trap tandem mass spectrometry. Rapid Commun Mass Spectrom. 2007;21:1531–40. doi: 10.1002/rcm.2987. [DOI] [PubMed] [Google Scholar]
- 122.Moore C, Coulter C, Crompton K, Zumwalt M. Determination of benzodiazepines in oral fluid using LC-MS-MS. J Anal Toxicol. 2007;31:596–600. doi: 10.1093/jat/31.9.596. [DOI] [PubMed] [Google Scholar]
- 123.Ngwa G, Fritch D, Blum K, Newland G. Simultaneous analysis of 14 benzodiazepines in oral fluid by solid-phase extraction and LC-MS-MS. J Anal Toxicol. 2007;31:369–76. doi: 10.1093/jat/31.7.369. [DOI] [PubMed] [Google Scholar]
- 124.Uddin MN, Samanidou VF, Papadoyannis IN. Validation of SPE-HPLC determination of 1,4-benzodiazepines and metabolites in blood plasma, urine, and saliva. J Sep Sci. 2008;31:3704–17. doi: 10.1002/jssc.200800342. [DOI] [PubMed] [Google Scholar]
- 125.Jehanli A, Brannan S, Moore L, Spiehler VR. Blind trials of an onsite saliva drug test for marijuana and opiates. J Forensic Sci. 2001;46:1214–20. [PubMed] [Google Scholar]
- 126.Sharp ME, Wallace SM, Hindmarsh KW, Peel HW. Monitoring saliva concentrations of meth-aqualone, codeine, secobarbital, diphenhydramine and diazepam after single oral doses. J Anal Toxicol. 1983;7:11–4. doi: 10.1093/jat/7.1.11. [DOI] [PubMed] [Google Scholar]
- 127.Cone EJ. Testing human hair for drugs of abuse, I: individual dose and time profiles of morphine and codeine in plasma, saliva, urine, and beard compared to drug-induced effects on pupils and behavior. J Anal Toxicol. 1990;14:1–7. doi: 10.1093/jat/14.1.1. [DOI] [PubMed] [Google Scholar]
- 128.Chen ZR, Somogyi AA, Reynolds G, Bochner F. Disposition and metabolism of codeine after single and chronic doses in one poor and seven extensive metabolisers. Br J Clin Pharmacol. 1991;31:381–90. doi: 10.1111/j.1365-2125.1991.tb05550.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.O’Neal CL, Crouch DJ, Rollins DE, Fatah A, Cheever ML. Correlation of saliva codeine concentrations with plasma concentrations after oral codeine administration. J Anal Toxicol. 1999;23:452–9. doi: 10.1093/jat/23.6.452. [DOI] [PubMed] [Google Scholar]
- 130.O’Neal CL, Crouch DL, Rollins DE, Fatah AA. The effects of collection methods on oral fluid codeine concentrations. J Anal Toxicol. 2000;24:536–42. doi: 10.1093/jat/24.7.536. [DOI] [PubMed] [Google Scholar]