Skip to main content
Applied and Environmental Microbiology logoLink to Applied and Environmental Microbiology
. 2011 Sep;77(17):6225–6232. doi: 10.1128/AEM.00393-11

Total and Viable Legionella pneumophila Cells in Hot and Natural Waters as Measured by Immunofluorescence-Based Assays and Solid-Phase Cytometry ,

N Parthuisot 1,2, M Binet 3, A Touron-Bodilis 3, C Pougnard 3, P Lebaron 1,2, J Baudart 1,2,*
PMCID: PMC3165396  PMID: 21742913

Abstract

A new method was developed for the rapid and sensitive detection of viable Legionella pneumophila. The method combines specific immunofluorescence (IF) staining using monoclonal antibodies with a bacterial viability marker (ChemChrome V6 cellular esterase activity marker) by means of solid-phase cytometry (SPC). IF methods were applied to the detection and enumeration of both the total and viable L. pneumophila cells in water samples. The sensitivity of the IF methods coupled to SPC was 34 cells liter−1, and the reproducibility was good, with the coefficient of variation generally falling below 30%. IF methods were applied to the enumeration of total and viable L. pneumophila cells in 46 domestic hot water samples as well as in cooling tower water and natural water samples, such as thermal spring water and freshwater samples. Comparison with standard plate counts showed that (i) the total direct counts were always higher than the plate counts and (ii) the viable counts were higher than or close to the plate counts. With domestic hot waters, when the IF assay was combined with the viability test, SPC detected up to 3.4 × 103 viable but nonculturable L. pneumophila cells per liter. These direct IF methods could be a powerful tool for high-frequency monitoring of domestic hot waters or for investigating the occurrence of viable L. pneumophila in both man-made water systems and environmental water samples.

INTRODUCTION

Legionnaires' disease was first recognized in July 1976. The pathogenic agent isolated was identified as a bacterium and was called Legionella pneumophila. Since then, more than 50 species of Legionella and 64 serogroups have been identified, and at least 20 species have been associated with the disease in humans (15). Nevertheless, L. pneumophila is still the most pathogenic species, accounting for more than 90% of legionellosis cases, and serogroup 1 is the etiological agent responsible for more than 80% of the legionellosis cases diagnosed worldwide (30, 31). Legionella bacteria, described as opportunistic human pathogens, are widespread in natural aquatic environments and in artificial water systems and can also survive for a long time in a low-nutrient environment (7, 9, 29). The majority of outbreaks have been traced to aerosols contaminated with these organisms from either cooling towers or water distribution systems (i.e., drinking water distribution systems and hot sanitary waters) (16). The analysis of water samples collected from a source suspected of amplifying Legionella is a valuable means of preventing the health risk posed by potential sources of legionellosis. The rapid monitoring of Legionella in water systems has therefore become a priority in preventing and controlling the disease. The health risk can be measured by a microbiological laboratory experienced in Legionella detection by determining the number of organisms present in water samples. Culture methods are the most commonly used diagnostic methods for Legionella infections. Nevertheless, these methods have some recognized limitations: they are time-consuming (taking up to 10 days) (12), they are not very sensitive, microbial contamination may inhibit Legionella growth, and viable but nonculturable (VBNC) bacteria can be present but not detected (26). Consequently, although culture methods are still considered the “gold standard” for the detection of Legionella in water, they are not well suited for real-time monitoring or for risk assessment and management of Legionella in water systems. The development of quicker and more sensitive methods without the cultivation step is the main priority for water quality assessment. These methods should allow the detection of all viable cells, including VBNC Legionella, in water systems in a few hours, so that Legionella proliferation can be prevented and rapidly controlled.

Immunofluorescence (IF) assays have already been proposed as a faster method of monitoring Legionella at hot water facilities and in cooling towers (3, 11, 28). These methods have some advantages with respect to detecting specific microbes in environmental water samples: they are easy to implement in a laboratory, and results can be obtained within a few hours. The targeted cells can be concentrated efficiently on a membrane filter and directly counted at the cellular level by use of an epifluorescence microscope or solid-phase cytometry (SPC) (3, 19). However, some caution is required when IF analysis is applied to environmental water samples: the antibodies must be highly specific for Legionella, and the targeted cells must be quantified accurately within a wide range of concentrations (3, 14). A further advantage of IF assays is that they can be combined with both specific detection and viability criteria (10, 22).

This study aimed at assessing the performances of two IF-based assays for rapid quantification of total and viable L. pneumophila cells in waters by using SPC. Viable L. pneumophila cells were detected using the IF method proposed by Aurell et al. (3) combined with the use of ChemChrome V6 (CV6), a cellular esterase activity marker, for viability assessment (21). In this paper, the term “viable SPC counts” refers to cells detected by IF and with detectable SPC esterase activity, although there is no direct proof that the detected cells are able to grow and divide under favorable conditions. The use of esterase activity as a viability marker in combination with cell detection by SPC was proposed by several authors for the detection of viable cells in aqueous products or water samples (11, 21, 25). These assays were used to investigate the occurrence of L. pneumophila in samples from man-made water systems, such as hot water samples and cooling tower waters, and in natural water samples, such as hot spring waters and freshwaters, and the results were compared to standard plate counts.

MATERIALS AND METHODS

Specificity control of the IF protocol and bacterial strains.

Two monoclonal antibodies, specific for L. pneumophila serogroup 1 and L. pneumophila serogroups 2 to 15, were purchased from Microbiodetection (Commercy, France). Their specificity for L. pneumophila has been controlled by enzyme-linked immunosorbent assay (ELISA) and immunofluorescence methods in previous works (3, 17). In this study, 52 additional strains (Legionella and non-Legionella species), provided from reference and environmental collections, were tested to complete the specificity control (see Table S1 in the supplemental material). The Legionella strains were grown on buffered yeast extract containing α-ketoglutarate (BCYEα) supplemented with l-cysteine and ferric pyrophosphate (Oxoid, Dardilly, France) at 37°C for 48 to 72 h, and non-Legionella strains were grown at 30 or 37°C, depending on their optimum growth temperature, for 24 h on nutrient agar (Bio-Rad, Marnes la Coquette, France).

The L. pneumophila serogroup 1 strain Philadelphia (ATCC 33152) was also used for performance evaluations (sensitivity and recovery rate) of the IF methods.

Enumeration of total and viable L. pneumophila cells by solid-phase cytometry.

Total L. pneumophila cells were immunostained by fluorescein isothiocyanate (FITC)-conjugated antibodies according to the protocol proposed by Aurell et al. (3). Ten milliliters of each water sample was filtered through a black polyester membrane (0.4-μm pore size and 25-mm diameter) (CB04; AES-Chemunex, Ivry sur Seine, France). The membrane was then placed on 50 μl of PBS-T-BSA staining solution (0.1% [vol/vol] Tween 20, 2% [wt/vol] bovine serum albumin [BSA] [Sigma-Aldrich, Saint Quentin Fallavier, France] in phosphate-buffered saline [PBS] adjusted to pH 6.9 [Sigma-Aldrich]). The staining solution contained a primary monoclonal antibody cocktail (the final concentration of anti-L. pneumophila serogroup 1 was 4.5 μg ml−1, and that of anti-L. pneumophila serogroups 2 to 15 was 7.3 μg ml−1), and incubations were performed at 37°C for 60 min. After staining, the membrane was transferred onto a hydrating pad soaked in 500 μl of PBS-T-BSA and incubated for 2 min to eliminate excess antibodies. Primary antibodies were revealed by a secondary antibody solution containing FITC-conjugated sheep anti-mouse IgG (final concentration of 40 μg ml−1) (Sigma-Aldrich) diluted in PBS-T-BSA with 0.01% Evans blue (Sigma-Aldrich). The membrane was placed on 50 μl of secondary antibody solution in petri dishes and incubated at 37°C for 60 min. The membrane was then transferred onto a hydrating pad soaked in 500 μl of PBS-T-BSA and incubated for a minimum of 2 min at room temperature to eliminate excess antibodies before SPC analysis.

The analysis was performed in triplicate, and the results are expressed as total L. pneumophila SPC counts liter−1. The detection limit of the method was 34 total L. pneumophila cells liter−1.

Viable L. pneumophila cells were immunostained using the same primary antibodies as those used for the detection of total L. pneumophila cells, but using red fluorescence-conjugated secondary antibodies. Immunostaining was performed after sample filtration (10 ml) through a black polyester membrane (0.45-μm pore size and 25-mm diameter; Oxyphen AG, Lachen, Switzerland). Primary antibodies were revealed by a secondary antibody solution containing TriColor-conjugated goat anti-mouse IgG (Invitrogen, Molecular Probes Europe BV, Leiden, The Netherlands) (final concentration, 20 μg ml−1). The cellular viability was assessed after immunostaining by using the CV6 substrate (AES-Chemunex, Ivry-sur-Seine, France), a cellular esterase activity marker which is cleaved into a fluorescent green product inside viable cells (8, 11, 13, 21, 23). The viability of the immunostained cells was tested using a total viable count assay performed with the CV6 substrate and following the instructions provided by the manufacturer (AES-Chemunex). L. pneumophila hydrolyzing the substrate emitted both red and green fluorescence simultaneously, as shown in Fig. S1 in the supplemental material.

The analysis was performed in triplicate, and the results are expressed as viable L. pneumophila SPC counts liter−1. The detection limit of the method was 34 viable L. pneumophila cells liter−1.

SPC enumeration.

The ChemScanRDI system (AES-Chemunex) scans the sample on the membrane support pad with an air-cooled argon laser emitting at 488 nm, and the fluorescence emission is collected in the green FL1 channel (500 to 540 nm), the orange FL2 channel (540 to 570 nm), and the red FL3 channel (655 to 705 nm). Total FITC-labeled L. pneumophila cells were detected with the FL1 channel, using the TVC Bioburden application provided by the manufacturer (AES-Chemunex). This application discriminated fluorescent events detected by SPC as targeted fluorescent bacterial signals or as particles (autofluorescent particles) by using a set of discriminant parameters (20).

The fluorescent secondary antibodies used for the viable L. pneumophila immunoassay are characterized by a 488-nm excitation wavelength and a 670-nm emission wavelength. Viable L. pneumophila cells corresponding to red immunostained cells and green cells hydrolyzing the enzyme substrate (CV6) were detected with both the FL1 and FL3 channels at the same time, using a modified TVC Bioburden application. After membrane scanning, a ratio of tertiary (FL3) to primary (FL1) signals (T/P) of 0.2 was activated to detect double-stained cells.

A final validation step for the discriminated fluorescence events was performed using a BH2 epifluorescence microscope (Olympus, Rungis, France) with an FITC block filter and a tetramethyl rhodamine isocyanate (TRITC) block filter installed on a motorized stage driven by the ChemScanRDI system.

Enumeration of L. pneumophila by the culture method.

Plate counts were enumerated according to the French standard AFNOR T90-431 (2), which is in compliance with the international standard ISO 11731. One-hundred-milliliter to 1-liter water samples were concentrated by filtration through 0.45-μm-pore-size polycarbonate filters (Sartorius SAS, Palaiseau, France). Each membrane was then transferred into 5 ml of sterile Milli-Q water and treated with ultrasonic energy twice for 1 min each with a Branson sonicator (Fisher Bioblock Scientific, Illkirch, France) operating at 42 kHz. One hundred microliters of the concentrate was spread onto a selective GVPC plate (BCYEα supplemented with l-cysteine and ferric pyrophosphate, 3 g of glycine, 100,000 U of polymyxin B, 80 mg of cycloheximide, and 1 mg of vancomycin per liter) (Oxoid). Samples (100 μl) of the concentrate were also plated onto GVPC plates after heat treatment (50°C for 30 min) and acid treatment (0.2 M HCl, 0.2 M KCl, pH 2, for 5 min) to eliminate non-Legionella organisms. The inoculated plates were then incubated at 37 ± 2°C, and the colonies were counted after 3, 5, and 10 days. The colonies were then examined for fluorescence under a Wood lamp, and those exhibiting Legionella morphology were transferred to BCYEα medium, BCYEα medium without cysteine, and blood agar medium (bioMérieux, Marcy l'Etoile, France) as a control. At least five colonies per sample were identified by Legionella-specific immunolatex reagents (Oxoid). The selected colonies were also confirmed by IF staining. For filtration of 1 liter of water, the detection limit of the method was 1 colony per plate, equivalent to 50 CFU liter−1, whereas the statistically significant quantification limit was based on the AFNOR recommendation of counting 5 colonies per plate, equivalent to 250 CFU liter−1 (2).

Water samples.

The performances of IF methods for L. pneumophila quantification were first evaluated with hot sanitary water artificially contaminated with various concentrations (103 to 107 cells liter−1) of L. pneumophila serogroup 1 cells (ATCC 33152), and the results were compared to those of the culture method.

The cell suspension was created from a fresh colony resuspended in PBS, and viable cell concentrations were measured after CV6 staining.

Next, the occurrence of L. pneumophila was investigated in 46 hot shower water samples collected from May to December 2005 from shower water of the bathroom outlets of private homes representative of communities in the Paris area and more widespread areas in France.

In addition, 8 non-hot-water samples, comprising thermal spring waters (3 samples), cooling tower waters (3 samples), and freshwaters (2 samples), were also tested to investigate the robustness of IF assays for the detection of total and viable L. pneumophila cells in naturally contaminated water samples. Cooling tower samples were collected from an industrial site in Spain during the spring of 2007. Thermal spring waters and freshwaters were collected from sites located in the Pyrénées (France) in the summer of 2007.

For hot shower (including spiking experiment samples) and cooling water samples, residual free chlorine was neutralized by the addition of 20 mg liter−1 of sodium thiosulfate before the analysis was performed.

Additional treatments were applied on the non-hot-water samples before application of the viable L. pneumophila SPC protocol. A preliminary study showed that the associated microflora present in these samples interfered with the detection of viable L. pneumophila cells by SPC (data not shown). These water samples were characterized by the presence of abundant photosynthetic microeukaryotes showing an autofluorescent signal detected in channel FL3 by SPC. To eliminate these microeukaryotes, the samples were prefiltered through a polycarbonate membrane (0.8-μm pore size and 47-mm diameter) (Whatman, Dominique Dutscher, Brumath, France) before application of the viable L. pneumophila test. Two approaches were performed for the esterase activity measurement of L. pneumophila. The first consisted of the direct detection of cellular activity immediately after sample filtration (direct activity), and the second consisted of detecting cellular activity after the incubation of cells on a selective nutritive plate (activity after GVPC incubation). For the latter, activity was measured after incubation of the membrane on a GVPC plate for 4 h and 15 h at 37°C. In addition, acid treatment before GVPC incubation was applied to reduce the overgrowth of nontarget microorganisms, as recommended by the French guidelines for the standard culture method (2). The acid treatment of the sample was performed immediately after water sample filtration and before incubation on the GVPC agar plate. The membrane was placed on 200 μl of acid solution (0.2 M HCl-KCl buffer, pH 2.0) for 5 min at room temperature. The viable L. pneumophila cells were then immunostained as described previously.

Statistical tests.

The normal distribution of data was tested using the Shapiro-Wilk test, and statistical comparisons between the various cell concentrations were performed using the Wilcoxon test. Correlation tests were performed (Spearman test and Pearson test with an α level of 0.05), and the relations between SPC counts and standard plate counts were determined by linear regression models. All statistics were calculated using XL Stat software (Addinsoft, France).

RESULTS

Specificity tests.

The specificity of the monoclonal anti-L. pneumophila antibody cocktail used in this study has already been published (3, 17). The specificity list was completed in our study with 52 Legionella and non-Legionella strains, including 15 L. pneumophila (serogroups 1 to 15) and 24 Legionella non-pneumophila strains, as reported in Table S1 in the supplemental material. Specificity tests were performed by applying the IF protocol using the FITC-conjugated anti-L. pneumophila cocktail, as described above. All serogroups of L. pneumophila strains, except for serogroup 7 and serogroup 11, which are not targeted by the antibody cocktail, were recognized by the monoclonal antibodies, and no cross-reaction was observed for the non-L. pneumophila strains.

Accuracy of immunofluorescence assays for enumeration of L. pneumophila cells in artificially contaminated hot waters.

The accuracies of the IF-based assays for enumerating total and viable L. pneumophila cells were evaluated with hot waters artificially contaminated with known log decimal concentrations of L. pneumophila serogroup 1. The reproducibility of both methods for quantifying L. pneumophila was evaluated based on two independent assays and using a wide range of L. pneumophila concentrations. The recovery rate of viable L. pneumophila cells was measured and compared to the standard plate counts from a third experiment.

Total and viable L. pneumophila SPC counts were linearly correlated with the number of L. pneumophila cells added as the inoculum in a log-log plot, as reported in Fig. 1 A for total counts (n = 15, r2 = 0.994, P < 0.0001; Pearson test) and in Fig. 1B for viable counts (direct activity measurements) (n = 14, r2 = 0.990, P < 0.0001; Pearson test). The cell recoveries were 102.4% for total SPC counts and 73.4% for viable SPC counts (Fig. 1).

Fig. 1.

Fig. 1.

Total (A) and viable (B) SPC counts of L. pneumophila in artificially contaminated hot water samples at different log concentrations.

The reproducibility of the IF methods was investigated by determining the coefficient of variation (CV) for three replicates and was found to be globally lower than 30%. CVs ranged from 0.7 to 25.0% for total counts and from 3.5 to 29.8% for viable SPC counts, except for the lowest concentration (102 cells liter−1), where they increased considerably, up to 100%.

The viable L. pneumophila SPC counts and plate counts were compared, and the results are reported in Table 1. Counts measured by both approaches were positively correlated (r2 = 0.974 and P < 0.0001; Pearson test), and the recovery rates of SPC counts were higher than the plate counts for all cell concentrations.

Table 1.

Viable L. pneumophila SPC counts and standard plate counts quantified in artificially contaminated hot sanitary water samples

Theoretical count (log10 cells liter−1) Direct plate count (log10 CFU liter−1) Direct viable SPC count (log10 cells liter−1)
7.30 5.02 6.64
5.14 6.63
5.08 6.61
6.30 3.50 5.64
3.64 5.65
3.80 5.60
5.30 2.06 4.52
2.39 4.65
<1.67 4.41
4.30 <1.67 3.53
1.76 3.46
2.24 3.41
3.30 <1.67 2.60
<1.67 2.40
<1.67 2.30

Occurrence of L. pneumophila in hot waters.

The occurrence of total and viable L. pneumophila cells was investigated in 46 hot water samples, and SPC counts were compared to the standard plate counts (Table 2).

Table 2.

Comparison of total and viable L. pneumophila counts in hot water samples obtained by the reference culture method and IF assays

Sample Standard plate count (CFU liter−1) Mean (SD) SPC count (cells liter−1)
Ratio of viable to total SPC counts (%)
Total L. pneumophila Viable L. pneumophila
Direct activity measurement With GVPC incubationa
1 <2.5 × 102 <3.4 × 101 5.0 × 101 (10.0 × 101) ND
2 <2.5 × 102 <3.4 × 101 3.3 × 101 (5.8 × 101) ND
3 <2.5 × 102 5.8 × 103 (1.3 × 103) 3.4 × 103 (1.3 × 103) ND 58.8
4 <2.5 × 102 4.5 × 103 (0.9 × 103) 1.0 × 102 (1.0 × 102) ND 2.2
5 <2.5 × 102 7.3 × 103 (1.1 × 103) 1.2 × 103 (0.06 × 103) ND 16.9
6 <2.5 × 102 9.8 × 103 (0.8 × 103) 4.7 × 102 (0.6 × 102) ND 4.7
7 <2.5 × 102 3.0 × 102 (0.0) <3.4 × 101 ND
8 <2.5 × 102 4.6 × 103 (0.8 × 103) 1.1 × 103 (0.3 × 103) ND 23.2
9 <2.5 × 102 1.2 × 103 (0.1 × 103) <3.4 × 101 ND
10 <2.5 × 102 1.7 × 102 (1.1 × 102) <3.4 × 101 ND
11 <2.5 × 102 1.0 × 103 (0.4 × 103) 3.3 × 101 (5.8 × 101) ND 3.2
12 <2.5 × 102 9.0 × 103 (1.2 × 103) 6.7 × 101 (11.5 × 101) ND 0.7
13 <2.5 × 102 1.9 × 103 (0.6 × 103) 1.7 × 103 (0.3 × 103) ND 90.9
14 <2.5 × 102 2.0 × 102 (1.0 × 102) <3.4 × 101 ND
15 <2.5 × 102 4.3 × 102 (1.5 × 102) 3.3 × 101 (5.8 × 101) ND 7.7
16 <2.5 × 102 7.3 × 104 (0.1 × 104) 3.3 × 101 (5.8 × 101) ND 0.04
17 <2.5 × 102 3.3 × 101 (5.8 × 101) <3.4 × 101 ND
18 <2.5 × 102 4.1 × 104 (0.4 × 104) 1.0 × 102 (1.0 × 102) ND 0.24
19 <2.5 × 102 <3.4 × 101 <3.4 × 101 ND
20 <2.5 × 102 6.5 × 102 (0.7 × 102) 3.3 × 101 (5.8 × 101) ND 5.1
21 <2.5 × 102 1.3 × 104 (0.3 × 104) 3.3 × 101 (5.8 × 101) ND 0.26
22 <2.5 × 102 <3.4 × 101 <3.4 × 101 ND
23 <2.5 × 102 2.5 × 102 (0.7 × 102) <3.4 × 101 ND
24 <2.5 × 102 5.1 × 103 (1.9 × 102) <3.4 × 101 ND
25 <2.5 × 102 8.7 × 102 (1.5 × 102) <3.4 × 101 ND
26 <2.5 × 102 6.7 × 101 (11.5 × 101) <3.4 × 101 ND
27 <2.5 × 102 1.0 × 102 (1.0 × 102) <3.4 × 101 ND
28 <2.5 × 102 6.7 × 101 (0.6 × 101) 6.7 × 101 (11.5 × 101) ND 100.0
29 <2.5 × 102 8.5 × 103 (0.7 × 103) <3.4 × 101 ND
30 <2.5 × 102 8.0 × 103 (2.0 × 103) <3.4 × 101 ND
31 <2.5 × 102 1.9 × 104 (0.4 × 104) <3.4 × 101 ND
32 <2.5 × 102 1.3 × 102 (2.3 × 102) <3.4 × 101 ND
33 <2.5 × 102 3.3 × 101 (5.8 × 101) <3.4 × 101 ND
34 <1.0 × 104 8.9 × 105 (1.3 × 105) 1.3 × 104 (1.2 × 104) ND 1.5
35 <5.0 × 102 4.3 × 102 (0.6 × 102) <3.4 × 101 ND
36 <2.5 × 102 2.6 × 103 (0.5 × 103) <3.4 × 101 ND
37 <2.5 × 102 2.6 × 104 (0.6 × 104) <3.4 × 101 ND
38 <2.5 × 102 2.9 × 105 (0.6 × 105) <3.4 × 101 <3.4 × 101
39 <5.0 × 102 6.8 × 103 (1.6 × 103) <3.4 × 101 6.7 × 101 (11.5 × 101) 0.98
40 <2.5 × 102 1.5 × 105 (0.3 × 105) <3.4 × 101 <3.4 × 101
41 <2.5 × 102 3.7 × 105 (0.2 × 105) <3.4 × 101 <3.4 × 101
42 <2.5 × 102 1.4 × 104 (1.1 × 104) <3.4 × 101 <3.4 × 101
43 <2.5 × 102 4.7 × 103 (0.3 × 103) <3.4 × 101 <3.4 × 101
44 <2.5 × 102 3.8 × 103 (0.4 × 103) <3.4 × 101 <3.4 × 101
45 <2.5 × 102 2.6 × 103 (0.2 × 103) <3.4 × 101 <3.4 × 101
46 <2.5 × 102 4.5 × 102 (0.7 × 102) <3.4 × 101 <3.4 × 101
a

ND, not determined.

Although the counts for all samples were below the detection limit of the culture method, L. pneumophila cells were detected in 91.3% of the water samples (42/46 samples) for measuring total cells and in 36.9% of the water samples (17/46 samples) for measuring viable cells. Total cell concentrations were globally higher than those of viable cells and ranged from 3.3 × 101 to 8.9 × 105 cells per liter. For comparison purposes, the number of viable L. pneumophila cells per liter ranged from 3.3 × 101 to 1.3 × 104 cells per liter. Eleven of the 17 samples testing positive for viable cells showed very low viable SPC concentrations, and 6/17 samples (samples 3, 5, 6, 8, 13, and 34) showed higher viable SPC counts (up to 1.3 × 104 cells per liter). All viable L. pneumophila SPC-positive samples were also positive for total SPC counts, except for samples 1 and 2. For these samples, the counts measured by the viable method were close to the detection limit of the immunofluorescence assays.

The recovery rates of viable L. pneumophila cells detected with and without GVPC incubation were compared for 9 hot water samples (samples 38 to 46). For both conditions, viable SPC counts were below the detection limit, except for sample 39, in which viable L. pneumophila cells were counted only after a revivification step on a GVPC plate.

When total and viable L. pneumophila cells were detected in the same sample, the ratio of viable to total L. pneumophila counts varied between 0.04% (sample 16) and 100% (sample 28).

The reproducibility of the IF methods was investigated by determining the CVs measured for three replicates for all hot water samples. CVs were clearly dependent on cell concentrations, as reported in Fig. 2, and the relationships between cell concentration and CV were similar for the total and viable SPC counts. The lowest CV values were measured for cell concentrations above 5 × 102 cells per liter, but the values of CV increased at the lowest concentrations.

Fig. 2.

Fig. 2.

Coefficients of variation determined from total L. pneumophila SPC counts (filled circles) and viable L. pneumophila SPC counts (open circles) in hot sanitary water samples.

Enumeration of total and viable L. pneumophila cells in cooling tower water and natural waters by SPC and comparison with standard plate counts.

The performances of the IF methods in detecting and enumerating L. pneumophila cells by SPC were evaluated with cooling tower water samples and natural water samples, such as thermal spring waters and freshwaters (Table 3). These kinds of water samples are known as complex waters because of their more abundant and different bacterial communities compared to those found in hot sanitary waters. All of the samples tested positive with both the IF methods and the standard method. The mean total L. pneumophila SPC counts were systematically higher than the standard plate counts, and cultivable L. pneumophila cells accounted for 0.1% to 16% of the total cells. Viable SPC counts measured without GVPC incubation were either similar to the standard plate counts or lower. Nevertheless, for some water samples, the viable SPC counts increased after a short incubation time on GVPC. Taking into account the counts with and without GVPC incubation, viable SPC counts and plate counts were positively correlated (r2 = 0.862 and P = 0.002; Spearman test) (Fig. 3), and viable cell recovery was 103.9% for the IF assay.

Table 3.

Comparison of L. pneumophila counts obtained by the standard culture method and immunodetection methods (SPC) for thermal spring water (TSW), cooling tower water (CTW), and freshwater (FW) samples

Sample Water source Mean (SD) standard plate count (CFU liter−1) CV (%) Total L. pneumophila cells
Viable L. pneumophila cellsa
Mean (SD) SPC count (cells liter−1) CV (%) Direct activity measurement
GVPC incubation (4 h)
GVPC incubation (15 h)
Mean (SD) SPC count (cells liter−1) CV (%) Ratio of viable to total cells (%) Mean (SD) SPC count (cells liter−1) CV (%) Ratio of viable to total cells (%) Mean (SD) SPC count (cells liter−1) CV (%) Ratio of viable to total cells (%)
47 TSW 2.8 × 103 (0.1 × 103) 2.0 9.4 × 104 (3.0 × 104) 32.3 3.5 × 102 (3.5 × 102) 101.0 0.37 4.0 × 104 (0.2 × 104) 5.7 42.5 ND ND
48 TSW 2.3 × 103 (0.4 × 103) 15.7 5.4 × 104 (0.8 × 104) 15.4 2.5 × 102 (0.7 × 102) 28.3 0.46 3.6 × 103 (0.8 × 103) 21.9 66.6 ND ND
49 TSW 3.6 × 103 1.7 × 105 (0.1 × 105) 3.5 <5.0 × 101 1.7 × 103 (0.5 × 103) 30.0 1.0 1.5 × 103 (0.1 × 103) 4.9
50 CTW 4.9 × 104 (0.2 × 104) 4.4 2.9 × 105 (0.2 × 105) 5.3 5.5 × 103 (3.5 × 103) 64.3 1.89 ND 4.8 × 104 (0.6 × 104) 13.4 16.5
51 CTW 9.0 × 103 (0.5 × 103) 5.9 4.1 × 105 (1.3 × 105) 30.6 1.2 × 104 (0.6 × 104) 47.1 2.92 1.3 × 104 (0.1 × 104) 10.9 3.17 2.0 × 103 (1.4 × 103) 70.7 0.5
52 CTW 9.5 × 102 (3.5 × 102) 37.2 7.7 × 104 (2.8 × 104) 36.7 <5.0 × 101 <5.0 × 101 8.5 × 102 (2.1 × 102) 25.0 1.1
53 FW 1.9 × 102 (1.8 × 102) 94.3 1.1 × 105 (0.3 × 105) 23.9 1.0 × 102 (1.4 × 102) 141.4 0.09 ND 3.0 × 102 (2.8 × 102) 94.3 0.2
54 FW 2.0 × 102 (2.8 × 102) 141.4 2.0 × 105 (0.0 × 105) 1.1 5.0 × 102 (2.8 × 102) 56.6 0.25 ND 5.0 × 101 (7.1 × 101) 141.4 0.025
a

ND, not determined.

Fig. 3.

Fig. 3.

Logarithms of concentrations of L. pneumophila per liter in water samples (thermal spring water, cooling tower water, and freshwater), determined by the viable IF test and the standard culture method.

DISCUSSION

The monitoring of L. pneumophila concentrations in water is a major goal for stakeholders and industries if they are to improve the quality assessment of their water systems. The culture method is currently used for monitoring Legionella in water systems. However, culturing is too time-consuming (taking up to 10 days) to be used for diagnostic purposes and has some limitations in terms of detecting all viable L. pneumophila cells present in a water sample (27). The risk management strategy for Legionella detection requires more rapid methods for the real-time monitoring of all L. pneumophila cells. Moreover, it should include VBNC cells in water systems, as they can be resuscitated under some environmental conditions (1, 11, 25). This strategy may be considered a validation of quality assurance and quality control procedures (16). Various tests have already been proposed to detect directly bacterial viability or activity at the cellular level without a culture stage and to quickly provide information on the physiological state of cells. Most of the tests used in microbial ecology are enzyme activity tests (18). The esterase activity test offers some advantages for rapid cell viability measurement, because enzymatic activity and cell membrane integrity can be measured simultaneously in only a few minutes. Another advantage is the possible combination of the esterase test, which stains viable cells green, with immunofluorescence assays using red fluorescence-conjugated antibodies. The proposed test allows the simultaneous detection of L. pneumophila cells by SPC by combining taxonomical and physiological information in the same assay by use of appropriate discriminants (optimized FL3-to-FL1 ratio). The mix of monoclonal antibodies used was highly specific to the majority of L. pneumophila serogroups but did not recognize serogroups 7 and 11. However, the role of these serogroups in legionellosis outbreaks is minor (15).

Results obtained with spiking experiments in hot sanitary water clearly showed that immunofluorescence methods combined with SPC quantified L. pneumophila cells with accuracy and reproducibility for a wide range of concentrations (103 to 107 cells liter−1). The viable counts were positively correlated with plate counts and were systematically higher, with ratios ranging from 25 to 192. The results confirm that the viability of L. pneumophila can be tested at the cellular level by using a culture-independent method, such as the esterase activity test, which can be combined with an immunofluorescence method for more sensitive detection of viable L. pneumophila cells in waters (21).

Applied to naturally contaminated hot water samples, the methods showed the weakest accuracies for the lowest concentrations (below 5 × 102 cells liter−1). This can be explained by the heterogeneous distribution of L. pneumophila cells in drinking water distribution systems, which can be influenced by the presence of biofilms or free amoebas providing favorable ecological niches supporting the survival of Legionella (1, 5). Moreover, on spiked hot sanitary waters, the CV was the highest for concentrations above 102 cells liter−1. Thus, for these waters, the precision of the methods can be improved by increasing the volume of water analyzed or by increasing the number of filtered membranes analyzed.

The investigation of 46 hot sanitary water samples revealed a more frequent occurrence of L. pneumophila by IF methods than that obtained by culture, since 91% of samples were positive by the total L. pneumophila assay and 37% of them showed the presence of viable L. pneumophila cells, whereas all samples were negative by culture. Ours results confirmed those of Aurell et al. (3) showing a higher detection rate of L. pneumophila by using IF methods in combination with solid-phase cytometry. The discrepancy between IF counts and culture was not surprising for the total L. pneumophila assay, because the method is able to detect all Legionella cells, including dead, viable, and VBNC Legionella cells (3). However, the higher recovery rate of viable cells by the IF method supported the possible dissemination of viable or active forms of Legionella not detectable by standard culture methods in water systems, as previously reported (10, 13, 24).

Various environmental factors affect the physiological state of bacterial cells in aquatic ecosystems. Some biotic and abiotic factors are well known to promote or reduce the physiological activity of L. pneumophila cells. Consequently, depending on the relative contributions of favorable and unfavorable factors, a variable fraction of viable L. pneumophila cells can be present in waters. The ratio of viable cells to total cells is a good indicator of the physiological status of L. pneumophila populations in water samples. In hot water samples, we measured highly variable ratios ranging from 0.04% to 100%. This variability may be due to the complexity of the environmental factors characterizing drinking water distribution systems (i.e., a low-nutrient environment, the presence of biofilms promoting the regrowth and diversity of some microorganisms and communities, and a lethal or sublethal concentration of biocides) (6, 13, 24), which can support or reduce the physiological activity of L. pneumophila. Additional physicochemical and biological characterizations of water samples or water systems could be useful for achieving a better understanding of which environmental factors control the fraction of viable L. pneumophila cells in drinking water distribution systems.

Globally, the ratios measured for cooling tower water and natural water samples were low, but combining the viability test with an incubation step on GVPC improved the recovery rates of viable L. pneumophila cells. The intracellular accumulation of fluorescein generated by esterase hydrolysis of the CV6 substrate depends on its efflux and membrane integrity (18) and could be an energy-dependent process linked to a membrane potential (4). Consequently, this accumulation process could depend on the physiological state of the cells present in the water. Although few samples were analyzed, the higher recovery rates after incubation on GVPC supported the hypothesis that unfavorable factors affected fluorescein accumulation. While these were probably not lethal, they may have temporarily reduced the enzymatic activity of the cells. This activity recovered after a short incubation step on a nutritive medium, which provided the nutrients and energy required for enzyme synthesis and/or maintenance of the membrane potential (18). Consequently, in the absence of any information on the physiological status of cells before the test is performed, the protocol for detecting viable L. pneumophila cells should include an analysis performed directly and with an incubation step on GVPC varying from 4 h to 15 h.

To conclude, the IF methods combined with SPC made it possible to quantify accurately and rapidly the numbers of total and viable L. pneumophila serogroup 1 and serogroup 2 to 6, 8 to 10, and 12 to 15 cells present in diverse water samples.

Combined with a viability test, the technique showed a higher sensitivity than the standard culture method for counting viable cells and offered the possibility of detecting viable but nonculturable L. pneumophila. Since VBNC L. pneumophila may constitute a fairly significant proportion of Legionella cells in water systems, and because these cells may be infectious (26), the proposed methods could be a useful technique for assessing the health risks associated with these microorganisms. In fact, such techniques could be powerful tools, similar to what is now proposed for real-time PCR, for high-frequency monitoring of hot waters for routine investigation. Moreover, they could help to provide more information on natural reservoirs and ecological niches of Legionella in aquatic environments and consequently lead to a better understanding of their ecology and how to prevent their transfer to water systems.

Supplementary Material

[Supplemental material]

ACKNOWLEDGMENTS

We gratefully acknowledge help with sample collection and Legionella culture by Marion Mottier from Observatoire Océanologique of Banyuls-sur-Mer.

This work was supported by grants from Electricité de France (Research and Development, Laboratoire National d'Hydraulique et Environnement).

Footnotes

Supplemental material for this article may be found at http://aem.asm.org/.

Published ahead of print on 8 July 2011.

REFERENCES

  • 1. AbuKwaik Y., Gao L. Y., Stone B. J., Venkataraman C., Harb O. S. 1998. Invasion of protozoa by Legionella pneumophila and its role in bacterial ecology and pathogenesis. Appl. Environ. Microbiol. 64:3127–3133 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. AFNOR 2003. Water quality—detection and enumeration of Legionella spp. and L. pneumophila. Method by direct inoculation and after concentration by membrane filtration or centrifugation. AFNOR NF T90-431. Association Française de Normalisation, Paris, France [Google Scholar]
  • 3. Aurell H., et al. 2004. Rapid detection and enumeration of Legionella pneumophila in hot water systems by solid-phase cytometry. Appl. Environ. Microbiol. 70:1651–1657 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Battin T. J. 1997. Assessment of fluorescein diacetate hydrolysis as a measure of total esterase activity in natural stream sediment biofilms. Sci. Total Environ. 198:51–60 [Google Scholar]
  • 5. Bej A. K., Mahbubani M. H., Atlas R. M. 1991. Detection of viable Legionella pneumophila in water by PCR and gene probe methods. Appl. Environ. Microbiol. 57:597–600 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Bonadonna L., et al. 2009. Microbial characterization of water and biofilms in drinking water distribution systems at sport facilities. Cent. Eur. J. Public Health 17:99–102 [DOI] [PubMed] [Google Scholar]
  • 7. Borella P., et al. 2005. Legionella contamination in hot water of Italian hotels. Appl. Environ. Microbiol. 71:5805–5813 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Catala P., et al. 1999. Effectiveness of CSE to counterstain particles and dead bacterial cells with permeabilised membranes: application to viability assessment in waters. FEMS Microbiol. Lett. 178:219–226 [DOI] [PubMed] [Google Scholar]
  • 9. Chang C. W., Hwang Y. H., Cheng W. Y., Chang C. P. 2007. Effects of chlorination and heat disinfection on long-term starved Legionella pneumophila in warm water. J. Appl. Microbiol. 102:1636–1644 [DOI] [PubMed] [Google Scholar]
  • 10. Cooper L. R., White J., Mahenthiralingam E., Hanlon G. W. 2008. Long-term persistence of a single Legionella pneumophila strain possessing the mip gene in a municipal shower despite repeated cycles of chlorination. J. Hosp. Infect. 70:154–159 [DOI] [PubMed] [Google Scholar]
  • 11. Delgado-Viscogliosi P., et al. 2005. Rapid method for enumeration of viable Legionella pneumophila and other Legionella spp. in water. Appl. Environ. Microbiol. 71:4086–4096 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Ditommaso S., Giacomuzzi M., Gentile M., Zotti C. M. 2010. Evaluation of the usefulness of a new direct immunofluorescence assay (ScanVIT-LegionellaTM) for monitoring hospital water systems contaminated with Legionella spp. Lett. Appl. Microbiol. 50:341–346 [DOI] [PubMed] [Google Scholar]
  • 13. Dusserre E., et al. 2008. A PCR-based method for monitoring Legionella pneumophila in water samples detects viable but noncultivable legionellae that can recover their cultivability. Appl. Environ. Microbiol. 74:4817–4824inelevel0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Edelstein P. H. 2006. Deja vu all over again: rapid enumeration of Legionella pneumophila in water. Appl. Environ. Microbiol. 72:980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Fields B. S., Benson R. F., Besser R. E. 2002. Legionella and Legionnaires' disease: 25 years of investigation. Clin. Microbiol. Rev. 15:506–526 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Fliermans C. 1996. Ecology of Legionella: from data to knowledge with a little wisdom. Microb. Ecol. 32:203–228 [DOI] [PubMed] [Google Scholar]
  • 17. Helbig J. H., Kurtz J. B., Pastoris M. C., Pelaz C., Luck P. C. 1997. Antigenic lipopolysaccharide components of Legionella pneumophila recognized by monoclonal antibodies: possibilities and limitations for division of the species into serogroups. J. Clin. Microbiol. 35:2841–2845 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Joux F., Lebaron P. 2000. Use of fluorescent probes to assess physiological functions of bacteria at single-cell level. Microb. Infect. 2:1523–1535 [DOI] [PubMed] [Google Scholar]
  • 19. Lemarchand K., Parthuisot N., Catala P., Lebaron P. 2001. Comparative assessment of epifluorescence microscopy, flow cytometry and solid-phase cytometry used in the enumeration of specific bacteria in water. Aquat. Microb. Ecol. 25:301–309 [Google Scholar]
  • 20. Mignon-Godefroy K., Guillet J., Butor C. 1997. Solid phase cytometry for detection of rare events. Cytometry 27:336–344 [PubMed] [Google Scholar]
  • 21. Parthuisot N., Catala P., Lemarchand K., Baudart J., Lebaron P. 2000. Evaluation of ChemChrome V6 for bacterial viability assessment in waters. J. Appl. Microbiol. 89:370–380 [DOI] [PubMed] [Google Scholar]
  • 22. Pougnard C., et al. 2002. Rapid detection and enumeration of Naegleria fowleri in surface waters by solid-phase cytometry. Appl. Environ. Microbiol. 68:3102–3107 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Reynolds D. T., Fricker C. R. 1999. Application of laser scanning for the rapid and automated detection of bacteria in water samples. J. Appl. Microbiol. 86:785–795 [DOI] [PubMed] [Google Scholar]
  • 24. Roeder R. S., et al. 2010. Long-term effects of disinfectants on the community composition of drinking water biofilms. Int. J. Hyg. Environ. Health 213:183–189 [DOI] [PubMed] [Google Scholar]
  • 25. Smith R., Von Tress M., Tubb C., Vanhaecke E. 2010. Evaluation of the ScanRDI as a rapid alternative to the pharmacopoeial sterility test method: comparison of the limits of detection. PDA J. Pharm. Sci. Technol. 64:356–363 [PubMed] [Google Scholar]
  • 26. Steinert M., Emody L., Amann R., Hacker J. 1997. Resuscitation of viable but nonculturable Legionella pneumophila Philadelphia JR32 by Acanthamoeba castellanii. Appl. Environ. Microbiol. 63:2047–2053 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Wellinghausen N., Frost C., Marre R. 2001. Detection of legionellae in hospital water samples by quantitative real-time LightCycler PCR. Appl. Environ. Microbiol. 67:3985–3993 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Yamamoto H., Hashimoto Y., Ezaki T. 1993. Comparison of detection methods for Legionella species in environmental water by colony isolation, fluorescent antibody staining, and PCR. Microbiol. Immunol. 37:617–622 [DOI] [PubMed] [Google Scholar]
  • 29. Yamamoto H., Hashimoto Y., Ezaki T. 1996. Study of nonculturable Legionella pneumophila cells during multiple-nutrient starvation. FEMS Microbiol. Ecol. 20:149–154 [Google Scholar]
  • 30. Yanez M. A., Carrasco-Serrano C., Barbera V. M., Catalan V. 2005. Quantitative detection of Legionella pneumophila in water samples by immunomagnetic purification and real-time PCR amplification of the dotA gene. Appl. Environ. Microbiol. 71:3433–3441 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Yu V. L., et al. 2002. Distribution of Legionella species and serogroups isolated by culture in patients with sporadic community-acquired legionellosis: an international collaborative survey. J. Infect. Dis. 186:127–128 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

[Supplemental material]
supp_77_17_6225__1.pdf (337.2KB, pdf)

Articles from Applied and Environmental Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES