Abstract
This article reports on the fabrication of a disposable bio-nano-chip (BNC), a microfluidic device composed of polydimethylsiloxane (PDMS) and thiolene-based optical epoxy which is both cost-effective and suitable for high performance immunoassays. A novel room temperature (RT) bonding technique was utilized so as to achieve irreversible covalent bonding between PDMS and thiolene-based epoxy layers, while at the same time being compatible with the insertion of agarose bead sensors, selectively arranged in an array of pyramidal microcavities replicated in the thiolene thin film layer. In the sealed device, the bead-supporting epoxy film is sandwiched between two PDMS layers comprising of fluidic injection and drain channels. The agarose bead sensors used in the device are sensitized with anti-C-reactive protein (CRP) antibody, and a fluorescent sandwich-type immunoassay was run to characterize the performance of this device. Computational fluid dynamics (CFD) was used based on the device specifications to model the bead penetration. Experimental data revealed analyte penetration of the immunocomplex to 100 μm into the 280 μm diameter agarose beads, which correlated well with the simulation. A dose–response curve was obtained and the linear dynamic range of the assay was established over 1 ng/mL to 50 ng/mL with a limit of detection less than 1 ng/mL.
Keywords: Agarose bead, PDMS, Microfluidic, Disposable, Immunoassay, CRP
1. Introduction
General immunoassays typically require time-consuming procedures, expensive instruments, and highly trained technicians. In implementing microfabrication techniques, microfluidic technology has shown significant potential to decrease analysis time and miniaturize the large, sophisticated detection instrumentation (Walt, 2005; Sia and Kricka, 2008). Microfluidic immunoassay systems can target small volumes of biological solutions with highly sensitive integrated sensors (Bange et al., 2005). In efforts to develop portable prototype sensors used in microfluidic systems, researchers have explored various types of functional materials and techniques, such as nitrocellulose paper (Fu et al., 2010), silver reduction (Sia et al., 2004), encoded particles (Pregibon et al., 2007) and gel electrophoresis (Meagher et al., 2008).
Recently, incorporating functionalized colloidal microbeads into microfluidic channels has been reported as a novel miniaturized means to perform immunoassay tests (Lim and Zhang, 2007; Shin et al., 2007). As the physical support for capture antibodies (Nolan and Sklar, 2002), beads are suitable as sensors for immunoassays. First of all, they possess high surface to volume ratios and thus are capable to immobilize a large number of capture antibodies. Furthermore, because of the fact that beads are addressable and can be pre-programmed by selective conjugation to various probe molecules, bead-based systems are capable of multiplexed assay tests. While many groups use solid phase beads, such as polystyrene beads, some groups utilize agarose beads (Goodey et al., 2001; Christodoulides et al., 2002; Ali et al., 2003; Jokerst et al., 2008; Thompson and Bau, 2011), because the interior porous structure of agarose beads, further contributes to increase the number of capture antibodies, and hence antigen antibody interactions (Jokerst et al., 2011).
However, the optimal strategy to integrate agarose beads in microfluidic systems is still not well established yet. As a particular type of gel, agarose beads cannot keep their original shapes if immobilized by clamping in two dimension structures (Thompson et al., 2010). Further, agarose beads need to be kept moist after preparation to prevent interior pore size degradation upon bead drying (Wong, 2007), and it is not quite clear yet whether this will affect final signal stabilities. Moreover, instead of using lateral flow over the top of beads or diffusion driven delivery methods, more complex fluidic patterns such as convective flow are necessary for analytes to quickly and effectively penetrate into the cores of the beads to take full advantage of the high avidity and large density of capture antibodies in three dimensional porous structures (Jokerst et al., 2011). To overcome these issues, several studies have employed anisotropically etched silicon to both physically protect beads within dedicated areas of the array and generate convective flow to efficiently deliver analytes to beads (Kirby et al., 2004; Li et al., 2005; Christodoulides et al., 2005b, 2007; Jokerst et al., 2010). Nevertheless, fabrication issues and cost constraints limit the direct use of silicon into practical systems, and new materials and designs are needed to optimize agarose bead-based immunoassays in microfluidic systems.
This paper focuses on the development of a microfluidic design, termed bio-nano-chip (BNC), composed of PDMS and thiolene-based epoxy, and evaluates its performance with a bead-based C-reactive protein (CRP) immunoassay. The low cost chip was integrated by permanently and irreversibly bonding PDMS and thiolene film at room temperature (RT).The seal-proof chip is disposable and potentially free of contaminations. These features are expected to accelerate the translation of the BNC from the laboratories to point of care applications for global public health (Yager et al., 2006). To characterize the performance of new structure, we carried out a sandwich-type immunoassay for CRP, an inflammation and cardiac risk biomarker (Christodoulides et al., 2005a).
2. Materials and methods
2.1. Materials and reagents
Silicon wafers (4 in.) were purchased from Nova Electronic Materials (Richardson, TX, USA). Photolithography masks were printed on mylar films at 10,160 dpi by Fine Line Imaging (Colorado Springs, Colorado, USA). SU8-3035 photoresist was obtained from MicroChem (Newton, MA, USA). Sylgard 184 PDMS kits were manufactured by Dow Corning (Midland, MI, USA) and 10:1 weight ratio (pre-polymer to curing agent) was used in this work. Norland optical adhesive 81 (NOA81) was ordered from Norland Products (Cranbury, NJ, USA). 3-Aminopropyltriethoxysilane (APTES, 97%) and 2-hydroxyethyl methacrylate (HEMA) were received from Sigma–Aldrich.
Agarose beads were prepared by emulsification of 2% type I-B agarose purchased from Sigma. Briefly, 1 g of agarose was stirred and dissolved in 50 mL of nano-pure water at 60°C. A suspending solution of 10 mL of Span85 (Sigma) and 90 mL hexanes was heated to 60° C and stirred at 900 rpm. The agarose solution was poured into the suspending solution followed by stirring at 900 rpm at 59° C for 1 min. The stirring was then adjusted to 600 rpm with heat off to allow the agarose to gel to 25°C. Beads were collected, washed with 50/50 mixture of ethanol/water and sorted using a sieve that screens out beads of 250–280 μm in diameter, which were then glyoxylated to transform the hydroxyl groups to aldehyde groups.
Rabbit anti-human CRP antibody used both as capture and detection antibody, was purchased from Accurate Chemical Corp (Westbury, NY, USA). AlexaFluor®488 protein labeling kit from Invitrogen was conjugated to the detection antibody following instructions from the manufacturer. CRP antigen was from Fitzgerald (Acton, MA, USA), and the Helicobacter pylori antibody for the control beads, was purchased from Meridian Life Science (Memphis, TN, USA). Prior to the assay, CRP antigen was diluted with phosphate buffered saline (PBS) blocking buffer containing 1% bovine serum albumin (BSA). Secondary detection antibody was diluted with PBS in 0.4% (v/v). A 500 μL volume of homemade glyoxylated 2% agarose beads were coupled to 9 mg/mL polyclonal rabbit anti human CRP antibody in a 1.5 mL solution overnight and blocked with tris solution for 1 h prior to final wash. Negative control beads were prepared similarly, by incubating 2% agarose beads with a polyclonal antibody irrelevant to the CRP target and specific to H. pylori.
2.2. Instrumentation
The BNC was placed under a BX2 Olympus microscope, equipped with a mercury lamp, a 0.13 NA 4× UPlanFlair objective, and a FITC filter cube (fluoroisothiocyanate, 480 nm excitation, 505 nm long-pass beam splitter dichroic mirror, and 535 ± 25 nm emissions). Fluorescent images were captured by a DVC 1312C CCD camera (Digital Video Camera, Austin, TX, USA).
Confocal images were acquired on an LSM 510 META laser scanning microscope system equipped with a 10×/0.45 Plen-APOCHROMAT objective lens from Zeiss (Germany). From the excitation and emission maxima of the AlexaFluor®488 dye at 495 nm and 519 nm, respectively, the proper configuration controls were applied. Argon laser, Lasos Carl Zeiss promenade 10, with output of 50% was used for excitation at 488 nm. Fluorescence emission was detected in the interval between 500 nm and 550 nm.
2.3. Assay procedure
In this study, the chip was then structurally and functionally evaluated with the completion of a series of assays targeting detection of CRP. Prior to the assay, a 3 × 4 microarray of wells on the chip was manually loaded with agarose beads. The first three columns on the array were loaded with beads coupled to anti-CRP antibody, while the fourth column hosted negative control beads, coupled to an irrelevant antibody to CRP antigen. A syringe pump (NE-1000, New Era Pump Systems) was employed to deliver the samples and reagents for the assay. After exciting the beads at 480 nm, the CRP detection beads exhibited green fluorescence signals in contrast to the negative counterparts, confirming the specificity of the antigen–antibody reactions and negligible non-specific binding within the BNC device (Supplement Fig. 2).
2.4. Data analysis
The beads were optically excited at 480 nm and resulting images were captured by CCD and further processed to quantitate the signals derived from the beads. More specifically, custom-built image analysis routines written within the ImageJ environment were used to process the captured images and analyze the fluorescence data. Each bead in the array as visualized in the green color was scanned with a series of line profiles spanning 80% of the bead diameter and the average of the maximum obtained for each profile line was recorded into a list. An outlier routine rejected measurements outside two standard deviations (SD) from the median. The fluorescence signal of a bead was calculated by averaging the fluorescence intensity from the remaining measurements. The fluorescence intensity of an array was defined as the average of all fluorescent beads of the same type (i.e., CRP) and the background was defined as the average of the negative control beads.
2.5. Computational modeling
Computational simulations were run in COMSOL 3.5a (Burlington, MA). An array of 280 μm beads resting in individual wells was constructed using AutoCAD and then imported into COMSOL. Navier–Stokes and convection and diffusion application modes were added from the chemical engineering module. The subdomain properties of the beads were loaded from the water library and Brinkman's equation was enabled to allow for flow through a 243 nm pore size bead with 96% porosity. The inlet velocity was set to 80 μL/min and outlet pressure was set to atmospheric pressure. The reaction rate governing the system is given by cAb + cAg < = > cAbAg, where cAb is the loaded antibody concentration, cAg is the delivered analyte concentration, and cAbAg is the coupled analyte–antibody pair. The reaction association and dissociation rates are 105 Lmol−1 s−1 and 10−5 s−1, respectively. The analyte concentration was set to 300 ng/mL and antibody concentration was 9.01 mg/mL. For simplicity, the loaded antibody and bound pair concentrations were normalized to the inlet analyte concentration. Particle image velocimetry (PIV) studies on the platform have been conducted to confirm the flow profiles from CFD and the flow patterns presented here compare well with those obtained from the experimental PIV (data not shown).
3. Results and discussions
3.1. Chip design and fabrication
The hybrid chip is composed of three layers as shown in Fig. 1 (see also Supplement Fig. 1). The top and bottom PDMS layers contain the injection and drain channels respectively. The channels (300 μm wide, 50 μm high) were molded with SU8-2015 and cast into PDMS. The middle layer was an optical epoxy film with square pyramidal cavities, where the beads were physically constrained. Optically transparent epoxy provides light transmission in a wide spectrum range. The low background noise becomes an important feature when dealing with complex biological matrices, such as blood and saliva.
Fig. 1.

(A) Thiolene-based epoxy bead array layer is sandwiched between two PDMS microfluidic layers. (B) Each 280 μm bead rests in a pyramidal pit well replicated from anisotropically etched silicon. (C) A higher pressure exists at the initial bead column closest to the inlet. (D) Computational fluidic dynamics simulations show analyte binding onto the bead array. (E) Schematic representation of a sandwiched-type immunoassay on agarose beads. Individual agarose beads conjugated on their surface and in their interior with a capture antibody are exposed to samples containing the target antigen. The unbound antigen is then washed away and a fluorescence-labeled detection antibody specific to the antigen is added that provides the tic signals when bound to the antigen throughout the beads imaged with a CCD camera.
The following steps were taken to achieve the pyramidal holes in the thiolene-based epoxy, as in Fig. 2(A). First, a 400 μm-thick silicon wafer with the 〈100〉 crystal surface orientation was prepared through anisotropic etching, as reported previously (Christodoulides et al., 2007), to create square pyramids with dimension at the top of 670 μm and bottom 80 μm as in Fig. 2(B). Second, PDMS is casted on the silicon master to get a replica with positive features of the square pyramids. Then, NOA81 optical liquid adhesive is dispensed on the PDMS surface that has positive features and spun at 3.3 g units for 30 s. The NOA81 chosen in this study has relatively low viscosity (300 cps), and therefore it easily spreads to a thin film without any air bubbles by spinning. However, because the PDMS surface is naturally hydrophobic, the liquid adhesive tends to aggregate together and form droplets on its surface, rather than spread out evenly, before spinning we need first utilize oxygen plasma (PE-50, Plasma Etch) and 2-hydroxyethyl methacrylate polymer grafting to render its surface permanently hydrophilic (Bodas and Khan-Malek, 2006). Finally, the composite was exposed in UV light for 1 min (Polylux 500, Norland Products). After the UV treatment, the liquid epoxy turns to a hard resilient solid film that could be peeled off easily from the PDMS. As displayed in Fig. 2(C), the replication from silicon to epoxy was accurate in terms of the square pyramidal shape. The bottom of the pyramid was found to be enlarged by about 50 μm due to the need to keep the level of epoxy liquid lower than the peak of positive PDMS replica after spinning, in order to produce through holes. A single piece of PDMS replica could be reused to produce a number of epoxy copies.
Fig. 2.

(A) Fabrication process of bead holders with the optical epoxy. (1) Spreading the epoxy over the PDMS replica of anisotropically etched silicon wafer. (2) Curing epoxy under the UV treatment. (3) Taking the PDMS and epoxy apart. (4) Attaching the bottom of epoxy film to the PDMS layer containing the drain. (5) Arranging the beads into the wells by tweezers. (6) Sealing the bead array by the PDMS layer the sample delivery channel. (B) SEM image on a silicon pyramidal cavity. The scale bar is 500 μm. (C) SEM image on the epoxy containing cavity replicated from the silicon mold. The scale bar is 500 μm.
The thiolene based epoxy could be bonded to PDMS by plasma cleaning (Hung et al., 2008). However, the general plasma cleaning protocol of bonding PDMS may not be suitable when agarose beads located on its surface. To ensure the polymeric bead structure stability, the protein-derivatized agarose beads need be kept wetted all the time. On one hand, if the beads were loaded before plasma cleaning, the vacuum environment would dry out the agarose beads and denature the antibodies. On the other hand, if beads were loaded after plasma cleaning, the time required to load beads onto the array, would make it hard to exploit the best timeframe for bonding to take effect, which is immediately after treatment of the epoxy surface. Recently, a new way of bonding PDMS to thermoplastics at RT has been reported (Sunkara et al., 2010; Tang and Lee, 2010). It takes advantage of the irreversible covalent bonding between plasma treated PDMS and the 3-aminopropyltriethoxysilane (APTES) modified thermoplastic surface (Supplement Fig. 3). Here, we applied a similar technique to the surface of epoxy. First, the surface of epoxy was cleaned by oxygen plasma and then soaked in 1% (v/v) APTES solution for 30 min and dried with nitrogen. Second, agarose beads conjugated with anti-CRP antibodies were loaded into the wells manually with tweezers. Finally, the epoxy surface was brought into contact with the plasma cleaned PDMS surface, resulting into irreversible covalent bonding at RT.
3.2. CRP immunoassay within BNC
The inverted pyramidal microwells within the BNC serve both as reaction and analysis chambers. In a typical assay procedure, sample and reagent solutions are delivered around and through the beads, and then exit the chip through a drain, as illustrated in Fig. 1. The microchip architecture largely enhances the mixing and reactions between analytes/detection antibodies and therefore sensitivities (Goodey et al., 2001; Li et al., 2005; Jokerst et al., 2009).
The assay procedure was optimized by varying the sample and reagent volumes as well as the assay incubation times. Optimal conditions identified for this setup included incubation of CRP antigen for 30 min at flow rate of 80 μL/min, followed by 30 min incubation with AlexaFluor®488-conjugated detection antibody also delivered at 80 μL/min for 30 min. Following the incubation of detection antibody, the chamber and beads contained within were rinsed with PBS at 120 μL/min for 3 min to remove non-specific binding of reagents.
The observed small intra-assay bead to bead variations arise from three main sources. First, according to the simulation data of computational fluid dynamics, as shown in Fig. 1(C) and (D), there is a minor pressure gradient across the bead array, with pressures dropping from the column closest to the inlet to the column closest to the outlet. Higher pressure results in more penetrations of both the antigen and detecting antibody into the agarose bead matrix, and hence stronger signals. Second, within the same column, signals could vary because of the slightly different drain sizes below beads, in other words, the bottom hole size of the reaction chamber. Drain size is important to the signal of bead since it also affects the pressure applied on the bead by creating complex local fluidic fields. In other words, both the location of bead in the array and the pressure on the bead could play important roles on the detected fluorescence intensity. In our experiments, and especially for multiplex experiments we always keep the arrangement of beads consistent with respect to analyte groups, and observe great reproducibility. Last, while tightly sieved, the agarose bead size distribution still displays a 5% CV (Wong, 2007), contributing to signal variations in our experiments. Because smaller beads sit lower in the pyramidal wells than the larger beads, they receive higher pressures and therefore show stronger signals.
3.3. Signals in agarose beads
The signal distribution was investigated by taking confocal images on a CRP bead at the end of the assay (Fig. 3(A)). The PDMS chip was mounted on a glass slide and inverted on the stage for imaging from the objective from below. Medial slices of the beads in the x–y plane, with a 1 μm thickness, at a distance of 140 μm from the top of the bead, were recorded as 8-bit in the green pseudo-color image of 510 × 510 pixels. The confocal data showed that signal mainly formed on the peripherals of beads and penetrated about 30% to the core of the beads (Supplement Fig. 4). As compared to the solid phase polystyrene beads, where antigen–antibody interactions are confined on their 2D microsphere surfaces, the porous agarose beads are capable to capture analytes in 3D shells with a thickness of about 100 μm, as we have previously demonstrated (Jokerst et al, 2011).
Fig. 3.

(A) Confocal fluorescence image of a bead after 25 ng/mL CRP assays. (C) Dose-response curve of CRP based on epi-fluorescence data.
3.4. Dose–response curve
To gain a thorough understanding of the behavior of the platform when used in the multiplex, a necessary step is to investigate the behavior of array composed of identical beads, as to eliminate any possible influence of the differences between beads conjugated to different biological entities, with potentially different concentrations of capture antibody depending on the analytes, and possible interferences due to cross talk between the immunoassays. In Fig. 3(B) is displayed a dose–response curve of CRP, where each data point obtained by averaging nine beads for that concentration with error bars shown at ±1 × SD. The limit of detection (LOD) was calculated at 1 ng/mL, as the mean of the 0 antigen control run ±3 × SD. While the physiological range for CRP measurements in blood extends to higher concentrations than the standards used in the dose–response curve (Liu et al., 2010), we are targeting the detection of biomarkers in both blood and saliva (Floriano et al., 2009) in our clinical studies. As such, further dilutions by several orders of magnitude are often required for the more viscous and mucinous oral fluid, making it more strategic in assay development steps to target a wide range of lower concentrations. If necessary, a larger dynamic range is achieved easily by lowering or increasing the concentrations of detection antibody, as well as the exposure time for the CCD camera.
4. Conclusions
In summary, we have demonstrated the suitability of a cost-effective microfabrication methodology based on PDMS and thiolene-based optical epoxy to produce a new disposable bio-nano-chip design for lab-on-a-chip immunoassays. In this example demonstrated with a CRP immunoassay, analytes penetrated 100 μm into the core of 280 μm porous agarose beads and resulted in high sensitivity. Moreover, the fabrication process of the bio-nano-chip is repeatable and amenable to scale-up, thereby affording applications targeting resource-scarce settings. In addition, the PDMS, a standard material in microfluidics, allows for a wide range of integration with new features and advances, which are constantly emerging from the microfluidic.
Supplementary Material
Acknowledgments
Funding for this work was provided by National Institutes of Health through the National Institute of Dental and Craniofacial Research (U01 DE15017 and U01 DE017793). The content is solely the responsibility of the authors and does not necessarily represent or reflect views of the NIH or the United States Government.
Footnotes
Appendix A. Supplementary Data: Supplementary data associated with this article can be found, in the online version, at doi:10.1016/j.bios.2011.07.027.
References
- Ali MF, Kirby R, Goodey AP, Rodriguez MD, Ellington AD, Neikirk DP, McDevitt JT. Analytical Chemistry. 2003;75:4732–4739. doi: 10.1021/ac034106z. [DOI] [PubMed] [Google Scholar]
- Bange A, Halsall HB, Heineman WR. Biosensors and Bioelectronics. 2005;20:2488–2503. doi: 10.1016/j.bios.2004.10.016. [DOI] [PubMed] [Google Scholar]
- Bodas D, Khan-Malek C. Microelectronic Engineering. 2006;83:1277–1279. [Google Scholar]
- Christodoulides N, Dharshan P, Wong J, Floriano PN, Neikirk D, McDevitt JT. Methods in Molecular Biology. 2007;385:131–144. doi: 10.1007/978-1-59745-426-1_10. [DOI] [PubMed] [Google Scholar]
- Christodoulides N, Floriano PN, Acosta SA, Ballard KLM, Weigum SE, Mohanty S, Dharshan P, Romanovicz D, McDevitt JT. Clinical Chemistry. 2005a;51a:2391–2395. doi: 10.1373/clinchem.2005.054882. [DOI] [PubMed] [Google Scholar]
- Christodoulides N, Mohanty S, Miller CS, Langub MC, Floriano PN, Dharshan P, Ali MF, Bernard B, Romanovicz D, Anslyn E, Fox PC, McDevitt JT. Lab on a Chip. 2005b;5:261–269. doi: 10.1039/b414194f. [DOI] [PubMed] [Google Scholar]
- Christodoulides N, Tran M, Floriano PN, Rodriguez M, Goodey A, Ali M, Neikirk D, McDevitt JT. Analytical Chemistry. 2002;74:3030–3036. doi: 10.1021/ac011150a. [DOI] [PubMed] [Google Scholar]
- Floriano PN, Christodoulides N, Miller CS, Ebersole JL, Spertus J, Rose BG, Kinane DF, Novak MJ, Steinhubl S, Acosta S, Mohanty S, Dharshan P, Yeh Ck, Redding S, Furmaga W, McDevitt JT. Clinical Chemistry. 2009;55:1530–1538. doi: 10.1373/clinchem.2008.117713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fu E, Lutz B, Kauffman P, Yager P. Lab on a Chip. 2010;10:918–920. doi: 10.1039/b919614e. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goodey A, Lavigne JJ, Savoy SM, Rodriguez MD, Curey T, Tsao A, Simmons G, Wright J, Yoo SJ, Sohn Y, Anslyn EV, Shear JB, Neikirk DP, McDevitt JT. Journal of the American Chemical Society. 2001;123:2559–2570. doi: 10.1021/ja003341l. [DOI] [PubMed] [Google Scholar]
- Hung LH, Lin R, Lee AP. Lab on a Chip. 2008;8:983–987. doi: 10.1039/b717710k. [DOI] [PubMed] [Google Scholar]
- Jokerst JV, Chou J, Camp JP, Wong J, Lennart A, Pollard AA, Floriano PN, Christodoulides N, Simmons GW, Zhou Y, Ali MF, McDevitt JT. Small. 2011;7:613–624. doi: 10.1002/smll.201002089. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jokerst JV, Floriano PN, Christodoulides N, Simmons GW, McDevitt JT. Lab on a Chip. 2008;8:2079–2090. doi: 10.1039/b817116e. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jokerst JV, Jacobson JW, Bhagwandin BD, Floriano PN, Christodoulides N, McDevitt JT. Analytical Chemistry. 2010;82:1571–1579. doi: 10.1021/ac901743u. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jokerst JV, Raamanathan A, Christodoulides N, Floriano PN, Pollard AA, Simmons GW, Wong J, Gage C, Furmaga WB, Redding SW, McDevitt JT. Biosensors and Bioelectronics. 2009;24:3622–3629. doi: 10.1016/j.bios.2009.05.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kirby R, Cho EJ, Gehrke B, Bayer T, Park YS, Neikirk DP, McDevitt JT, Ellington AD. Analytical Chemistry. 2004;76:4066–4075. doi: 10.1021/ac049858n. [DOI] [PubMed] [Google Scholar]
- Li SF, Floriano PN, Christodoulides N, Fozdar DY, Shao DB, Ali MF, Dharshan P, Mohanty S, Neikirk D, McDevitt JT, Chen SC. Biosensors and Bioelectronics. 2005;21:574–580. doi: 10.1016/j.bios.2004.12.010. [DOI] [PubMed] [Google Scholar]
- Lim C, Zhang Y. Biosensors and Bioelectronics. 2007;22:1197–1204. doi: 10.1016/j.bios.2006.06.005. [DOI] [PubMed] [Google Scholar]
- Liu A, Bui T, Van Nguyen H, Ong B, Shen Q, Kamalasena D. Age and Ageing. 2010;39:559–565. doi: 10.1093/ageing/afq067. [DOI] [PubMed] [Google Scholar]
- Meagher RJ, Hatch AV, Renzi RF, Singh AK. Lab on a Chip. 2008;8:2046–2053. doi: 10.1039/b815152k. [DOI] [PubMed] [Google Scholar]
- Nolan JP, Sklar LA. Trends in Biotechnology. 2002;20:9–12. doi: 10.1016/s0167-7799(01)01844-3. [DOI] [PubMed] [Google Scholar]
- Pregibon DC, Toner M, Doyle PS. Science. 2007;315:1393–1396. doi: 10.1126/science.1134929. [DOI] [PubMed] [Google Scholar]
- Shin KS, Lee SW, Han KC, Kim SK, Yang EK, Park JH, Ju BK, Kang JY, Kim TS. Biosensors and Bioelectronics. 2007;22:2261–2267. doi: 10.1016/j.bios.2006.11.011. (Selected Papers from the Ninth World Congress On Biosensors. Toronto, Canada 10–12 May 2006, Alice X. J. Tang) [DOI] [PubMed] [Google Scholar]
- Sia SK, Kricka LJ. Lab on a Chip. 2008;8:1982–1983. doi: 10.1039/b817915h. [DOI] [PubMed] [Google Scholar]
- Sia SK, Linder V, Parviz BA, Siegel A, Whitesides GM. Angewandte Chemie International Edition. 2004;43:498–502. doi: 10.1002/anie.200353016. [DOI] [PubMed] [Google Scholar]
- Sunkara V, Park DK, Hwang H, Chantiwas R, Soper SA, Cho YK. Lab on a Chip. 2010;11:962–965. doi: 10.1039/c0lc00272k. [DOI] [PubMed] [Google Scholar]
- Tang L, Lee NY. Lab on a Chip. 2010;10:1274–1280. doi: 10.1039/b924753j. [DOI] [PubMed] [Google Scholar]
- Thompson JA, Bau HH. Analytical Chemistry. 2011;83:2858–2861. doi: 10.1021/ac200410v. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Thompson JA, Du XG, Grogan JM, Schrlau MG, Bau HH. Journal of Micromechanics and Microengineering. 2010;20:115017–115025. [Google Scholar]
- Walt DR. Science. 2005;308:217–219. doi: 10.1126/science.1108161. [DOI] [PubMed] [Google Scholar]
- Wong J. Ph D Thesis. The University of Texas at Austin; 2007. Bead based microreactors for sensing applications. [Google Scholar]
- Yager P, Edwards T, Fu E, Helton K, Nelson K, Tam MR, Weigl BH. Nature. 2006;442:412–418. doi: 10.1038/nature05064. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.