. 2011 Oct 10;4:391. doi: 10.1186/1756-0500-4-391

An inter-laboratory comparison of urinary 3-hydroxypropylmercapturic acid measurement demonstrates good reproducibility between laboratories

Emmanuel Minet 1,, Graham Errington 1, Gerhard Scherer 2, Kirk Newland 3, Mehran Sharifi 4, Brian Bailey 5, Mike McEwan 1, Francis Cheung 1
PMCID: PMC3208382  PMID: 21985092

Abstract

Background

Biomarkers have been used extensively in clinical studies to assess toxicant exposure in smokers and non-smokers and have recently been used in the evaluation of novel tobacco products. The urinary metabolite 3-HPMA, a metabolite of the major tobacco smoke toxicity contributor acrolein, is one example of a biomarker used to measure exposure to tobacco smoke. A number of laboratories have developed liquid chromatography with tandem mass spectrometry (LC-MS/MS) based methods to measure urinary 3-HPMA; however, it is unclear to what extent the data obtained by these different laboratories are comparable.

Findings

This report describes an inter-laboratory comparison carried out to evaluate the comparability of 3-HPMA measurement between four laboratories. A common set of spiked and authentic smoker and non-smoker urine samples were used. Each laboratory used their in-house LC-MS/MS method and a common internal standard. A comparison of the repeatability ('r'), reproducibility ('R'), and coefficient of variation for 3-HPMA demonstrated that within-laboratory variation was consistently lower than between-laboratory variation. The average inter-laboratory coefficient of variation was 7% for fortified urine samples and 16.2% for authentic urine samples. Together, this represents an inter-laboratory variation of 12.2%.

Conclusion

The results from this first inter-laboratory comparison for the measurement of 3-HPMA in urine demonstrate a reasonably good consensus between laboratories. However, some consistent measurement biases were still observed between laboratories, suggesting that additional work may be required to further reduce the inter-laboratory coefficient of variation.

Background

Cigarette smoke contains thousands of chemicals, including toxicants, which can be categorized as either gases, semi-volatiles (gas/vapor phase), or particles ("tar" phase) [1]. Machine-measured cigarette yields under the ISO testing regimen do not provide an accurate estimate of human exposure to cigarette smoke toxicants [2]. These limitations have led to the development of methods to quantify biomarkers for specific toxicants in biological fluids such as urine, saliva, and plasma [3].

The gas phase, tobacco smoke toxicant acrolein [CAS:107-02-8] (Figure 1A) has been identified by the World Health Organization (WHO) study group on Tobacco Product Regulation (TobReg) as a major contributor to smoke toxicity [4]. This evaluation was based on the concentration of acrolein in smoke and its toxicity potency factor (cancer and non-cancer), established using various models. 3-hydroxypropylmercapturic acid (3-HPMA) is the major urinary metabolite of acrolein (Figure 1B) [5], and it can be quantified using LC-MS based methods [3,5].

Figure 1.

Figure 1

Chemical structure of acrolein (A), 3-HPMA (B), and 3-HPMA-13C3-15N (C). 3-HPMA and 3-HPMA-13C3-15N molecular weights are also indicated.

One critical element in the measurement and interpretation of biomarker data (including 3-HPMA) is the comparability in method analysis between different laboratories, which can use different methodologies. For instance, a study conducted by Biber and colleagues on a common set of urinary and plasma samples comparing nicotine and cotinine data from eleven laboratories concluded that individual values could vary significantly between laboratories [6]. In a more recent study, Bernert and colleagues showed that good measurement reproducibility for cotinine in a common set of samples could be achieved between six laboratories, when a standardized HPLC-UV method was used [7].

In this study we tested the reproducibility of 3-HPMA measurement between four laboratories using their in-house method and a common set of fortified and authentic urine samples. Each laboratory used a common reference compound and the internal standard 3-HPMA-13C3-15N.

Material and methods

Reagents and samples

Synthetic 3-HPMA (reference compound, Figure 1B) and 3-HPMA-13C3-15N (internal standard, Figure 1C) were obtained from AptoChem (Montreal, Canada). 3-HPMA-13C3-15N was ordered as a custom synthesis and the same lot was used by each laboratory. 3-HPMA-d3 was supplied by Toronto Research Chemicals (North York, Canada). Pooled non-smoker urine samples were supplied fortified with 3-HPMA by RECIPE Chemicals (Munich, Germany). Four concentrations of synthetic 3-HPMA were used: unspiked (background ≈30-50 ng/ml), 400 ng/ml, 1200 ng/ml, and 3600 ng/ml 3-HPMA. Prior to distribution, the samples were quantified using one of the participating laboratories (Lab 2) in order to ensure the quality of the preparation. The samples were then portioned into 5-ml-aliquots, lyophilized and shipped to the laboratories in triplicates of each sample (3 × 4 vials with lyophilized urine). The laboratories were advised to reconstitute the samples with 5 ml water.

Five authentic urine samples, covering a 3-HPMA concentration range which reflects typical levels in non-smokers to heavy smokers, were aliquoted in triplicates and sent to the participating laboratories (3 × 5 vials). The lyophilized non-smoker urine samples were supplied by RECIPE® (Munich, Germany), a supplier of samples used for quality assurance testing. The smoker samples were obtained as part of a biomarker study conducted previously by BAT. The corresponding study protocol and informed consent forms were approved by the Ethics Committee of the Bayerische landesarztekammer Munich, Germany (v. 18.02.2008), which contained a provision for revisiting the samples for the purpose of biomarker method development. The clinical study was conducted in accordance with the World Medical Association Declaration of Helsinki (World Medical Association, 2004) and International Conference on Harmonisation (ICH) Guidelines for Good Clinical Practice (GCP) (International Conference on Harmonization, 1996).

Analytical procedure

The samples were labelled with a code number and randomized prior to distribution. 10 mg of each reference compound and the internal standard were dispatched as a dried powder. Samples were received in one batch by each laboratory in September 2009, and all analyses were completed by November 2009. A summary of the analytical methods is given in Table 1, which is based on the performance of QC samples measured prior to the initiation of the study. All four laboratories used their in-house protocol, which included a solid phase extraction step and analysis by LC-ESI-MS/MS. The methods were validated according to FDA guidelines.

Table 1.

Analytical method characteristics

Laboratories 1 2 3 4
Method LC-ESI-MS/MS LC-ESI-MS/MS LC-ESI-MS/MS LC-ESI-MS/MS
mode positive negative positive negative
SPE (column) Phenomenex Strata-X Isolute ENV+ Waters OASIS Waters OASIS
SPE recovery (%) 74 68 75 78
HPLC column (make) Waters Xterra MS C18 Waters HILIC-Silica Waters Acquity Phenyl Thermo BioBasic AX
HPLC column (size) 50 × 2.1 mm, 2.5 μm 150 × 2.1 mm, 3 μm 100 × 2.1 mm, 1.7 μm 50 × 3 mm, 5 μm
Quantifier ion (mass) m/z 222 - 163 m/z 220 - 89 m/z 222 - 117 m/z 220 -91
Qualifier ion (mass) m/z 222 - 117 m/z 220 - 91
precision intra-day (%) 1.1 to 5.9 1.4 to 8.6 1.1 to 1.5 1.2 to 6.5
precision inter-day (%)a 5.1 to 5.3 3.3 to 7 1.7 to 3.9 3.3 to 7.5
accuracy (%) 93.2 to 102 83.9 to 102 97.6 to 102 96.8 to 101
LOD (ng/ml) 2.21 NDd ND NDd
LOQ (ng/ml) 7 25 50f 35f
Linearity (ng/ml) 7 to 5400 25 to 10000 50 to 5000 35 to 5000
Matrix effect (%) ND 7.4 to 17 -1.7 to 19.6 -5.5 to 6.3

aValues represent the precision range obtained for low, medium, and high concentrations at the time the methods were developed, except for Lab 1 where only low and high concentrations were tested.

bBased on calibration standards.

cLOD was an estimate based on spiked water, water being used as the SPE solvent.

dNot determined

eLOQ was established in spiked water, water being used as SPE solvent.

fMatrices used were either diluted non-smoker urine or non-smoker urine with very low 3-HPMA background.

Statistical analysis

Basic statistical analysis was carried out with MINITAB v15.1. Individual value plots were produced to test inter-laboratory variation of 3-HPMA concentrations. A non-parametric Wilcoxon paired t-test was performed to compare preliminary test analyses conducted with different internal standards. Analysis of covariance was used to compare the analytical methods at the four laboratories using the same internal standard (3-HPMA-13C3-15N). Precision statistics, as defined in ISO 5725-2 [8], were used as a measure of random errors, and expressed as repeatability ('r') and reproducibility ('R'). For the purposes of this study, in which each laboratory used its own method, 'R' refers to inter-laboratory variation.

Results and discussion

Data for both fortified and authentic urine, using 3-HPMA-13C3-15N as internal standard, were reported by each lab and the corresponding 3-HPMA concentrations (ng/ml) are shown in Table 2. In addition, laboratory 1 repeated the measurements using two different internal standards - 3-HPMA-d3 and 3-HPMA-13C3-15N - which were prepared and analyzed on the same day, in order to investigate the potential confounding effects of using different standards under the same analytical conditions. A non-parametric paired t-test showed that the use of 3-HPMA-d3 gave consistently higher concentrations than 3-HPMA-13C3-15N (Figure 2), highlighting the importance of standardizing the use of internal standards across each laboratory throughout the study.

Table 2.

All urinary 3-HPMA data in ng/ml

3-HPMA (ng/ml)

Samples Lab1 Lab2 Lab3 Lab4
1 a pooled NSa urine 40.6 31.9 < LOQb < LOQ
b pooled NS urine 29.8 30.8 < LOQ < LOQ
c pooled NS urine 40.0 30.7 < LOQ < LOQ

2 a fortified NS urine 466 492 396 405
b fortified NS urine 402 470 400 403
c fortified NS urine 431 471 401 402

3 a fortified NS urine 1302 1340 1180 1070
b fortified NS urine 1230 1270 1160 1140
c fortified NS urine 1140 1340 1150 1090

4 a fortified NS urine 3624 3780 3470 3220
b fortified NS urine 3714 3820 3420 3210
c fortified NS urine 3504 3970 3370 3240

5 a NS urine 39.8 48.3 < LOQ < LOQ
b NS urine 48.2 46.1 < LOQ 36.6
c NS urine 62.4 46.9 < LOQ 35.1

6 a smoker urine 371 376 269 321
b smoker urine 376 382 294 269
c smoker urine 371 370 300 258

7 a smoker urine 870 842 613 556
b smoker urine 960 874 673 659
c smoker urine 840 830 721 619

8 a smoker urine 1080 1200 862 969
b smoker urine 1122 1160 925 914
c smoker urine 1044 1180 878 929

9 a smoker urine 1482 1390 1200 1100
b smoker urine 1296 1370 1210 1130
c smoker urine 1260 1390 1100 894

Samples are numbered from 1 to 9. Samples 1 non-smoker urine, samples 2-4 = fortified samples, and samples 5-9 = authentic urine samples. Each sample was aliquoted in triplicates labeled a, b, and c.

aNS: non smokers

b< LOQ: below limit of quantification

Figure 2.

Figure 2

Boxplot of non-parametric differences between 3-HPMA measured in all samples using 3-HPMA-d3 and 3-HPMA-13C3-15N internal standards. The hypothesis (Ho) is based on no difference (0) between the 3-HPMA-d3 measures minus the 3-HPMA-13C3-15N measures. The box plot shows a clear positive difference with p = 0 based on a Wilcoxon paired t-test with a 95% confidence interval for the mean difference (x).

A background level of 40 to 60 ng/ml 3-HPMA was observed in the non-smoker samples selected for this study. This is expected given that acrolein is also the product of lipid peroxidation, fossil fuel combustion, and is found in cooked food [9].

As a quality control check, data from the fortified urine samples (Table 2) were plotted to generate a regression line and the corresponding equation. Using this, the values for the urine samples (Table 2) were recalculated based on the 3-HPMA peak area and the 3-HPMA-13C3-15N internal standard. The calculated concentrations were consistent with the reported concentrations from each lab (Additional file 1).

Individual value plots constructed using the sample data indicate a close similarity in the measurements, across the broad range of 3-HPMA concentrations, for all four laboratories using 3-HPMA-13C3-15N (Figure 3). However, an analysis of covariance (ANOVA) indicated a significant variation between laboratories still existed (Table 3). A closer fit could be observed between lab 1 and 2, and between lab 3 and 4. The coefficients of variation, giving an estimate of the imprecision for repeated measures at different concentration ranges, are also reported in Table 3. The imprecision for each concentration range should be interpreted carefully since the replicate measures were obtained from three aliquots from a single solution rather than a triplicate measure of a unique sample. The fortified samples (samples 2 to 4), were used as an internal calibration reference to calculate accuracies (Table 4).

Figure 3.

Figure 3

Individual value plots for 3-HPMA (blue circles) (ng/ml). A. Value plot for the fortified samples in four laboratories. B. Value plot for authentic urine samples in four laboratories. Missing values were below the LOQ.

Table 3.

One way ANOVA for 3-HPMA vs laboratories (lab1, 2, 3, and 4) for each set of samples

Mean 3-HPMA (ng/ml) StDev CoV P value (one-way Anova)
Sample 1 lab1 36.8 6.1 16.5 0.186
lab2 31.13 0.7 2.1
lab3 < LOQ NA NA
lab4 < LOQ NA NA

Sample 2 lab1 433 32.2 7.4 0.002
lab2 477.7 12.4 2.6
lab3 399 2.6 0.7
lab4 403.3 1.5 0.4

Sample 3 lab1 1224 81.2 6.6 0.004
lab2 1316.7 40.4 3.1
lab3 1163.3 15.3 1.3
lab4 1100 36.1 3.3

Sample 4 lab1 3614 105.4 2.9 0.000
lab2 3856.7 100.2 2.6
lab3 3420 50 1.5
lab4 3223.3 15.3 0.5

Sample 5 lab1 50.1 11.4 22.8 0.182
lab2 47.1 1.1 2.4
lab3 < LOQ NA NA
lab4 35.8 1.1 3

Sample 6 lab1 372.4 2.77 0.7 0.000
lab2 376 6 1.6
lab3 287.7 16.4 5.7
lab4 282.7 33.6 11.9

Sample 7 lab1 890 62.4 7 0.000
lab2 848.7 22.7 2.7
lab3 669 54.1 8.1
lab4 611.3 51.9 8.5

Sample 8 lab1 1082 39 3.6 0.000
lab2 1180 20 1.7
lab3 888.3 32.7 3.7
lab4 937.3 28.4 3

Sample 9 lab1 1346 119.1 8.85 0.006
lab2 1383.3 11.5 0.8
lab3 1170 60.8 5.2
lab4 1041.3 128.5 12.3

gNA: not applicable, fewer than three data points due to at least one measure < LOQ.

Table 4.

Accuracies calculated for each laboratory based on 3-HPMA fortified samples at 400 ng/ml (samples 2), 1200 ng/ml (samples 3), and 3600 ng/ml (samples 4)

Accuracies (%)
Fortified sample (ng/ml) Lab1 Lab2 Lab3 Lab4

400 ng/ml 108 119 99 101
1200 ng/ml 102 110 97 92
3600 ng/ml 100 107 95 90

aAverage calculated over three independent measures

A comparison of the repeatability ('r'), reproducibility ('R'), and coefficient of variation for 3-HPMA demonstrated that within-laboratory variation was consistently lower than between-laboratory variation. The average intra-laboratory CoV was 5%, while the average inter-laboratory CoV was 12.2% (Table 5). The average inter-laboratory coefficient of variation was 7% for the fortified urine samples and 16.2% for the authentic urine samples. These results show close comparability with those observed by Biber and colleagues, where samples spiked with cotinine had an inter-laboratory CoV ranging from 3 to 19%, while a CoV range of 4 to 59% was reported for authentic urine samples of smokers [10].

Table 5.

Repeatability, reproducibility, and intra, inter-laboratory coefficient of variation for 3-HPMA between the four participating laboratories

Samples Mean 3-HPMA (ng/ml)a rb Rc CoV within (%)d CoV between (%)e
1 34 5.8 13.9 6.1 NAf
2 428 50 100.8 4.2 8.4
3 1201 142.7 261.8 4.2 7.8
4 3528 268.4 460.3 2.7 4.7
5 45 7.6 24.2 5.9 NAf
6 330 54.9 138 6 15
7 755 143.3 363.4 6.8 17.2
8 1021 89.1 346.4 3.1 12.1
9 1235 223.1 710.1 6.5 20.5

aMean of individual 3-HPMA values for all participating laboratories and for the corresponding sample set

bReproducibility

cRepeatability

dcomposite intralaboratory coefficient of variation

einter-laboratory coefficient of variation for the four participating laboratories

fNA: not applicable due to data < LOQ

The overall average inter-laboratory coefficient of variation for all samples in this study was 12.2%. A CoV value higher than 10% might indicate that there is still some room for improvement; however, this seems to be in line with WHO standardized clinical methods, which in previous studies have reported average inter-laboratory coefficients of variation (CoV) above 10% [11-13].

The results from this first inter-laboratory comparison for the measurement of 3-HPMA in urine demonstrate a reasonably good consensus between laboratories, with an average CoV of 12.2%. However, some consistent measurement biases were still observed between laboratories, suggesting that additional work may be required to reduce the inter-laboratory CoV even further.

Abbreviations

3-HPMA: 3-hydroxypropylmercapturic acid; CoV: coefficient of variation; ESI: electrospray; LC-MS: liquid chromatography-mass spectrometry; r: repeatability; R: reproducibility; TobReg: tobacco product regulation; UV: ultraviolet; WHO: world health organization

Competing interests

British American Tobacco (BAT) funded and designed the study.

The participating laboratories are service providers for Biomarkers analysis and have previously conducted biomarkers analyses for BAT.

Authors' contributions

EM, MM, FC, and GS designed the study. GS, KN, BB, and MS conducted the sample analyses. GE performed the statistical analysis. EM and FC drafted the manuscript. All authors read and approved the final manuscript.

Written informed consent was obtained from the urine donors for publication of this manuscript and accompanying images. A copy of the form was made available for review by the by the Editor-in-Chief of this journal.

Supplementary Material

Additional file 1

Authentic urine 3-HPMA concentrations recalculated using a calibration curve derived from the urine samples spiked with synthetic 3-HPMA. The 3-HPMA fortified non-smokers urine samples were used to establish 3-HPMA calibration curves for each laboratory. The concentration of 3-HPMA in authentic smokers urine was back-calculated based on the calibration curve and peak areas. The values were compared with the values obtained by each lab with their own calibration curve.

Click here for file (27.5KB, DOC)

Contributor Information

Emmanuel Minet, Email: emmanuel_minet@bat.com.

Graham Errington, Email: graham_errington@bat.com.

Gerhard Scherer, Email: Gerhard.Scherer@abf-lab.com.

Kirk Newland, Email: kirk.newland@celerion.com.

Mehran Sharifi, Email: msharifi@labstat.com.

Brian Bailey, Email: Brian.Bailey@covance.com.

Mike McEwan, Email: mike_mcewan@bat.com.

Francis Cheung, Email: francis_cheung@bat.com.

References

  1. Rogdman A, Perfetti TA. The chemical components of tobacco and tobacco smoke. Baton Rouge, CRC Press; 2008. [Google Scholar]
  2. WHO Scientific Advisory Committee on Tobacco Product Regulation (SacTob) World Health Organization. Geneva; 2002. Conclusions and recommendations on health claims derived from ISO/FTC method to measure cigarette yield. [Google Scholar]
  3. Scherer G, Engl J, Urban M, Gilch G, Janket D, Riedel K. Relationship between machine-derived smoke yields and biomarkers in cigarette smokers in Germany. Regul Toxicol Pharmacol. 2007;47:171–183. doi: 10.1016/j.yrtph.2006.09.001. [DOI] [PubMed] [Google Scholar]
  4. Burns DM, Dybing E, Gray N, Hecht S, Anderson C, Sanner T, O'Connor R, Djordjevic M, Dresler C, Hainaut P, Jarvis M, Opperhuizen A, Straif K. Mandated lowering of toxicants in cigarette smoke: a description of the World Health Organization TobReg proposal. Tob Control. 2008;17:132–141. doi: 10.1136/tc.2007.024158. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Carmella SG, Chen M, Zhang Y, Zhang S, Hatsukami DK, Hecht SS. Quantitation of acrolein-derived (3-hydroxypropyl)mercapturic acid in human urine by liquid chromatography-atmospheric pressure chemical ionization tandem mass spectrometry: effects of cigarette smoking. Chem Res Toxicol. 2007;20:986–990. doi: 10.1021/tx700075y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Biber A, Scherer G, Hoepfner I, Adlkofer F, Heller WD, Haddow JE, Knight GJ. Determination of nicotine and cotinine in human serum and urine: an interlaboratory study. Toxicol Lett. 1987;35:45–52. doi: 10.1016/0378-4274(87)90084-1. [DOI] [PubMed] [Google Scholar]
  7. Bernert JT, Jacob P, Holiday DB, Benowitz NL, Sosnoff CS, Doig MV, Feyerabend C, Aldous KM, Sharifi M, Kellogg MD, Langman LJ. Interlaboratory comparability of serum cotinine measurements at smoker and nonsmoker concentration levels: a round-robin study. Nicotine Tob Res. 2009;11:1458–1466. doi: 10.1093/ntr/ntp161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. ISO 5725-2. Accuracy (trueness and precision) of measurement methods and results: Part 2. Basic method for the determination of repeatability and reproducibility of a standard measurement method. International Organization for Standardization. Geneva, Switzerland; 1994. [Google Scholar]
  9. Stevens JF, Maier CS. Acrolein: sources, metabolism, and biomolecular interactions relevant to human health and disease. Mol Nutr Food Res. 2008;52(1):7–25. doi: 10.1002/mnfr.200700412. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Biber A, Scherer G, Hoepfner I, Adlkofer F, Heller WD, Haddow JE, Knight GJ. Determination of nicotine and cotinine in human serum and urine: an interlaboratory study. Toxicol Lett. 1987;35:45–52. doi: 10.1016/0378-4274(87)90084-1. [DOI] [PubMed] [Google Scholar]
  11. Coucke W, Devleeschouwer N, Libeer JC, Schiettecatte J, Martin M, Smitz J. Accuracy and reproducibility of automated estradiol-17beta and progesterone assays using native serum samples: results obtained in the Belgian external assessment scheme. Hum Reprod. 2007;22:3204–3209. doi: 10.1093/humrep/dem322. [DOI] [PubMed] [Google Scholar]
  12. Siekmann L. Requirements for reference (calibration) laboratories in laboratory medicine. Clin Biochem Rev. 2007;28:149–154. [PMC free article] [PubMed] [Google Scholar]
  13. Thorpe SJ, Heath A, Blackmore S, Lee A, Hamilton M, O'broin S, Nelson BC, Pfeiffer C. International Standard for serum vitamin B(12) and serum folate: international collaborative study to evaluate a batch of lyophilised serum for B(12) and folate content. Clin Chem Lab Med. 2007;45:380–386. doi: 10.1515/CCLM.2007.072. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Additional file 1

Authentic urine 3-HPMA concentrations recalculated using a calibration curve derived from the urine samples spiked with synthetic 3-HPMA. The 3-HPMA fortified non-smokers urine samples were used to establish 3-HPMA calibration curves for each laboratory. The concentration of 3-HPMA in authentic smokers urine was back-calculated based on the calibration curve and peak areas. The values were compared with the values obtained by each lab with their own calibration curve.

Click here for file (27.5KB, DOC)

Articles from BMC Research Notes are provided here courtesy of BMC