Abstract
Heart rate variability (HRV) analysis is affected by ectopic beats. An efficient method was proposed to deal with the ectopic beats. The method was based on trend correlation of the heart timing signal. Predictor of R-R interval (RRI) value at ectopic beat time was constructed by the weight calculation and the slope estimation of preceding normal RRI. The type of ectopic beat was detected and replaced by the predictor of RRI. The performance of the simulated signal after ectopic correction was tested by the standard value using power spectrum density (PSD) estimation, whereas the results of clinical data with ectopic beats were compared with the adjacent ectopic-free data. The result showed the frequency indexes after ectopy corrected had less error than other methods with the test of simulated signal and clinical data. It indicated our method could improve the PSD estimation in HRV analysis. The method had advantages of high accuracy and real time properties to recover the sinus node modulation.
Keywords: Frequency domain, Ectopic beat, Heart rate variability (HRV), Power spectrum density
1. Introduction
Heart rate variability (HRV) analysis is widely used to evaluate the fluctuations of heart beat signals in many fields (Camm et al., 1996; Bettoni and Zimmermann, 2002; Maestri et al., 2010). Frequency domain analysis of HRV is efficient to reflect the automatic nerves system (ANS) modulation containing sympathetic and parasympathetic nerve activities by the indexes of low frequency (LF, 0.04–0.15 Hz) and high frequency (HF, 0.15–0.40 Hz) (Myers et al., 1986; Li et al., 2009). The indexes of frequency analysis are obtained from power spectrum density (PSD) to estimate ANS activity modulated by the sinus node (SN).
The HRV signal is expected to include normal R-R interval (RRI) modulated by SN, but will be disturbed by ectopic beats that are not originating from SN. The ectopy beats are mainly caused by premature beats, tachycardia, and other arrhythmias and even by QRS wave misdetections (Sapoznikov et al., 1992). They distort HRV signal with spurious frequency components and random spectra in the LF and HF bandwidths. The ectopic beats could be detected from the irregular RRI that results in a sharp transient of time domain indexes at the ectopic beat.
Some techniques are introduced to correct the ectopic confusion by deleting the beat or replacing the beat by a middle value. When the ectopic beat is deleted, the RRI is twice the mean normal heart period (HP); replacing the beat with a middle value is proper in the case of a ventricular beat. Another correction is performed by spline interpolation on the erroneous beats to aid in the avoidance of substantial disturbance of PSD estimation (Acar et al., 2000). Unfortunately, when the atrial premature contraction (APC) occurs, the SN rhythm is reset and the coupled RRI and compensational RRI are shorter than two normal RRIs. The interpolation on the period of APC may not reflect the SN rhythm properly. These methods are based on the beat differences as HP or heart rate (HR). The heart timing (HT) method is oriented to the beat location band limited by SN modulation. The HT method (Mateo and Laguna, 2003; Solem et al., 2006) is proposed to correct the ectopic beats. The ectopic RRI on the ectopic beat was removed and then replaced by a proper one in order to restore the SN modulation. The solution was presented to minimize the effect of these types of anomalies in the series of beats and the corresponding PSD-derived clinical indexes. These methods showed a better ability to reset the activity of the SN by PSD test. While providing excellent results, HT-based correction was also associated with heavy computation in prior work.
Our work adopted a trend-predict correction (TPC) method to detect the ectopic beats in RRI sequence. On the presence of ectopic beats, the TPC method based on HT signal was proposed to handle the ectopic beats. The TPC method was tested by a simulated HRV signal compared with other methods. The TPC method was presented to resolve the PSD of actual electrocardiograph (ECG) signal containing ectopic beats from European ST-T database.
2. Materials and methods
2.1. Ectopic correction algorithm
An algorithm was designed to detect the ectopic beats (Mateo and Laguna, 2003), and the deviation value in time tk was:
,
|
(1) |
where tk is the time of the kth beat following the initial event, tk −1 and tk +1 are the front and back time, respectively, and U is the suitable threshold which is calculated as function of the standard deviation (SD) to detect the ectopic beat. With this method, the time of ectopic beat occurring could be detected.
The correction method is based on the correlation of preceding sinus beats on the presence of ectopic beat, and is called the TPC method.
If the normal RRI exists before the time tk when ectopic beat occurs, the predictive RRI (RRIpred) in tk is constructed as:
| RRIpred=RRItrend+RRIturb, | (2) |
where RRItrend is reflected by trending component, and RRIturb is the turbulence value between adjacent RRIs. RRItrend is calculated based on weighted value of preceding RRI sequence.
,
|
(3) |
where y(t) is RRI value at time t, n is the number of prior RRI, and w(t) is the weight at time t, which is higher near tk.
Then RRIturb is calculated by I[tk] and E[tk]:
| RRIturb=I[tk]×E[tk]. | (4) |
I[tk] is judged by the signs of k 1 and k 2. A tiny value will be added to k 1 and k 2 to avoid zero.
| I[tk]=k1×k2×(k1+k2)/|k1×k2×(k1+k2)|. | (5) |
k 1 and k 2 are calculated to present the slopes at tk −1 and tk −2:
| k1=(y(tk−1)−y(tk−2))/(y(tk−1)+y(tk−2)), | (6) |
| k2=(y(tk−2)−y(tk−3))/(y(tk−2)+y(tk−3)). | (7) |
E[tk] is determined by k 1, k 2, RRItrend, and SDRRI:
,
|
(8) |
where SDRRI is the SD of preceding RRI, and a and b are predictive coefficients.
If the ectopic beats occur at time tk, the characteristics of ectopic RRI before and after tk are judged by RRIpred. The types of ectopic beats such as APC and premature ventricular contraction (PVC) are determined from their characteristics of RRI. For example, the threshold of coupled RRI of APC is lower than 0.85×RRIpred. The sums of coupled RRI and compensatory RRI of APC are lower than 2×RRIpred. When the APC is detected, the coupled RRI is replaced mainly by RRIpred, and the compensatory RRI at tk +1 is replaced by next predictive RRI.
2.2. Simulated HRV signal
The simulated signal is:
| y(t)=4cos(2πt×0.02)+cos(2πt×0.095)+cos(2πt×0.275)+C, | (9) |
where y(t) is the simulated RRI of HRV signal, t is the time of each RRI, and C is the constant value. y(t) consists of several items that represent different frequency variables. The three cosine components express three frequency components with a central frequency at 0.020, 0.095, and 0.275 Hz, corresponding to indexes of very low frequency (VLF), LF, and HF in frequency domain. It is a simple simulated HRV signal including frequency components with enhancing VLF power (direct current (DC) component).
y(t) reflects the regular heart beat period from SN modulation except that ectopic beats happen. It is hypothesized that one ectopic beat occurred at the time tk. The RRI at the ectopic time is disturbed by more than 20% alteration of normal RRI.
On the case of PVC, the coupled RRI y(tk)PVC and the compensatory RRI y(tk+ 1)PVC are simulated:
![]() |
(10) |
The two RRIs of APC below are obtained, when y(tk+ 1)APC is not fully compensated:
![]() |
(11) |
The simulated HRV signals containing ectopic beats are achieved as Eqs. (10) and (11). The signals are corrected by the TPC method in Section 2.1 using the preceding normal RRI before the ectopic beat(s), where the positions of ectopic beats are detected with Eq. (1). The spline interpolation correction (SIC) method and middle value replacing (MVR) method are as contrast. In SIC, the ectopic RRI is removed, and the new RRI is got on the period of ectopic beats by cubic spline interpolation with resample frequency at 4 Hz. In MVR method the ectopic RRI is replaced by the middle value of the two abnormal RRIs. The signals after TPC and MVR are also resampled at 4 Hz.
2.3. Clinical data correction
The evaluation protocol of correction method for clinical data is based on the hypothesis that the frequency indexes at the nearby segments A, B, and C in limited time are similar or are stably changing after SN modulation (Fig. 1). Thus, the differences of frequency indexes between segments A and B, B and C were smaller. It will be possible to evaluate the behavior of ectopic correction method when comparing ectopic-beats-free behavior in segments A and C with the ectopic beats-treated behavior in B (Mateo and Laguna, 2003). The differences of PSD estimation between segments are shown as mean±SD, where mean reflects the accuracy of correction result, and SD reflects the precision. The smaller the mean and SD, the higher the accuracy and precision that will be achieved.
Fig. 1.
HRV signals of test window, segments A, B, and C
A and C have no ectopic beats, segment B has at least one. A, B and C are set to four minutes. Segments A and B, C and B have overlapping episodes
A sum of 132 pieces of ECG episodes were gained from the first 48 of 90 data in the European ST-T database (ESDB) previously studied by Taddei et al. (1992). These data contained 132 ectopic episodes from annotation files as follows: 91 episodes containing one ectopic beat, 23 containing two, 18 containing three or more. Each ECG episode was divided into three overlapping, 4-min segments: A, B, and C (Fig. 1). Segments A and C had no ectopic beats, whereas B contained the ectopic beat(s). Segment B was between segments A and C, having the overlapping episodes with A and C. The ectopic beats during segment B were handled with TPC method or other methods.
Significant difference between two methods or between two segments was evaluated by paired-sample t-test (P<0.05).
2.4. Advanced detrending method
The real HRV signal composed of RRI sequence is usually nonstationary due to hormone modulation and noises (Tarvainen et al., 2002), which are reflected by the VLF and ultra low frequency (ULF). VLF and ULF, called as DC components or trending components by us, have uncertain influence on the PSD estimation (Myers et al., 1992; van Schelven et al., 2000). As the sample length of HRV signal is short (such as four minutes duration), the differential ratio of frequency analysis is unable to distinguish the DC component and other frequency bandwidths, and then the frequency confusion occurs. If the DC component is not removed, it will enhance the LF and HF components. Some works have presented detrending techniques to eliminate the component (Mitov, 1998; Tarvainen et al., 2002; Colak, 2009). The detrending preprocessing is carried out together with ectopic correction to avoid the influence of DC after spline interpolation.
The RRI sequences are presented by variable Z:
| Z=RRI1, RRI2, RRI3, ..., RRIn, | (12) |
where n is the number of RRI.
Z is divided to Z stat and Z trend (Z=Z stat +Z trend), where Z stat contains the components of LF and HF. Z trend reflecting the trending component (DC) should be removed.
With the advanced detrending method, Z stat is calculated as:
.
|
(13) |
Here, parameter I is the observation matrix, D 2 is the discrete approximation of the second derivative operator. λ is regularization parameter which could be adjusted. When λ is set as 20 and 300, the cut-off frequency is 0.041 and 0.011 times the sampling frequency, respectively. Z stat is a new HRV signal sequence, which contains only the useful frequency components. The method will show its performance with the simulated signal Eq. (9) and be applied to the actual HRV signal.
3. Results
3.1. Ectopic beat simulation
The simulated HRV signals from Eqs. (9)–(11) are introduced to test the ectopic correction methods. In Table 1, the initial value is the initial result of PSD estimation of the simulated HRV signal containing only normal RRI. The detrending value is the PSD estimation of the simulated normal signal with the processing of detrending and resample at 4 Hz. The other estimations of the signals containing ectopic beats are made by the process of detrending. The detrending value is regarded as the standard value for correction methods.
Table 1.
PSD estimations of simulated HRV signals with ectopic correction methods
| Simulated HRV signal | UF | LF | HF | LVH |
| Initial value | 1.00 | 0.50 | 0.50 | 1.00 |
| Detrending value | 0.963±0.000 | 0.458±0.000 | 0.505±0.000 | 0.908±0.000 |
| One APC existing | 1.820±0.200 | 0.549±0.048 | 1.271±0.186 | 0.440±0.069 |
| One PVC existing | 2.378±0.321 | 0.570±0.054 | 1.808±0.300 | 0.323±0.053 |
| APC+PVC by TPC | 0.964±0.015 | 0.459±0.011 | 0.505±0.010 | 0.910±0.026 |
| APC+PVC by MVR | 1.078±0.053 | 0.507±0.038 | 0.571±0.033 | 0.890±0.080 |
| APC+PVC by SIC | 0.959±0.013 | 0.463±0.012 | 0.496±0.002 | 0.933±0.023 |
Data are expressed as mean±SD. PSD is estimated using Welch’s periodogram. UF=LF+HF, LVH=LF/HF. The simulated HRV signal is from Eq. (9), with t=1 to 300 s, C=100. Simulated APC or PVC is at some time (tk, k=51–250) in the signal. On the case of APC+PVC, the distance between APC and PVC is changing from −200 to 200 with steps of 2 beats
Table 1 shows the results of PSD indexes with the influence of PVC or APC simulated 200 times, where the four indexes (UF, LF, HF, and LVH) are disturbed largely in one APC existing or one PVC existing. It indicates that HF enhances more largely than LF when ectopic beats occur. It shows the results when the signals containing both APC and PVC are corrected by the TPC, MVR and SIC methods. The result of TPC method is better than those of the methods of MVR and SIC. It had an advantage to recover the original indexes of HRV PSD.
3.2. Clinical data results
Table 2 shows the differences between the clinical indexes at LF and HF bands by TPC and SIC methods during segments A, B, and C as defined in Section 2.3. ΔLFA is the difference of LF during segment A between TPC and SIC methods. It is shown during all segments that the SIC method obtains more power by means of LF and HF than TPC when all the means of the differences are negative. The two methods had more differences on correcting segment B presented by ΔLFB and ΔHFB, whereas their differences are negligible during segments A and C.
Table 2.
Differences between the clinical indexes at LF and HF bands by means of TPC and SIC methods during segments A, B, and C
| Diff | Mean±SD | P |
| ΔLFA | −11±65 | 0.057 |
| ΔLFB | −93±502* | 0.034* |
| ΔLFC | −15±82* | 0.033* |
| ΔHFA | −8±101 | 0.359 |
| ΔHFB | −71±574 | 0.157 |
| ΔHFC | −7±53 | 0.131 |
Significant difference (P<0.05)
Table 3 shows the differences between each two segments with TPC and SIC methods. For example ΔLFAB is the difference between segments A and B with TPC or SIC method. It can be observed that when the segment B is involved, the mean and SD by SIC are larger than those by TPC, whereas the SD is higher involving segment B than that between the normal segments A and C with SIC method.
Table 3.
Differences between different segments with SIC and TPC methods
| Method | ΔLFAB | ΔLFBC | ΔLFAC | ΔHFAB | ΔHFBC | ΔHFAC |
| SIC | −111±560* | 85±557 | −26±394 | −101±584* | 52±601 | −50±279 |
| TPC | −28±245 | 7±296 | −21±378 | −38±147* | −13±197 | −51±266* |
Significant difference (P<0.05)
4. Discussion
The detrending method is introduced by Tarvainen et al. (2002). It was adopted in this research as a necessary preprocessing technique. Because the DC component exists in the normal RRI sequence, its influence is only seen in simulated HRV signals. Kim et al. (2009) found the Welch method and autoregressive (AR) model were not appropriate for calculating the LF and HF tested by the real HRV signals resampled at 4 Hz. However, after the processing of detrending, it showed that Welch and AR were able to estimate PSD with less error (Table 1).
Some researchers (Dimmer et al., 1998; Lombardi et al., 2004) had different findings in the LF index, while some (Zimmermann and Kalusche, 2001; Bettoni and Zimmermann, 2002; Amar et al., 2003; Ovreiu et al., 2008) thought that the HF was dominant in paroxysmal atrial fibrillation (PAF) onset or HF was the independent predictor to PAF recurrence (Yamada et al., 2009). Others had their own opinions on LF or HF (Huang et al., 1998; Miyakoshi et al., 2009). These may be due to the different preprocessing of DC and ectopic beats that enhance LF and HF separately.
Some works (Mateo and Laguna, 2003; Solem et al., 2006) used new methods for ectopy correction based on HT signals referred to as the DFT of the interpolated HT signal (FHTIS) method (Table 4), comparing spline interpolation of the 14th order with the incorrect RRI removed (DFT of the interpolated HR signals (FHRIS)). The 132 pieces of ECG episodes from ESDB were obtained to test their method. The work of FHTIS by Solem et al. (2006) continued the work by Mateo and Laguna (2003) to decrease the computation amount for real-time analysis. It was mentioned that the LF index estimation had different precision in their methods when they obtained satisfactory results. It shows FHRIS had poor performance for ectopy correction, whereas SIC was using the cubic spline interpolation.
Table 4.
Differences between segments by FHTIS, FHRIS, SIC, and TPC methods using European ST-T database
| Method | ΔLFAB | ΔLFBC | ΔLFAC | ΔHFAB | ΔHFBC | ΔHFAC |
| FHTIS1 | −71±575* | 103±415* | 32±670 | −58±213* | 38±177* | −20±186 |
| FHTIS2 | −71±575 | 103±415 | 32±670 | −58±213 | 38±177 | −20±186 |
| FHRIS12 | −1256±11767 | 1288±11657 | 31±658 | −331±2323 | 314±2292 | −17±140 |
| SIC | −111±560* | 85±557 | −26±394 | −101±584* | 52±601 | −50±279 |
| TPC | −28±245 | 7±296 | −21±378 | −38±147* | −13±197 | −51±266* |
The HRV signal from ESDB had some acceleration and deceleration of HR. Therefore, the mean and SD are larger in ESDB data than those of simulated signals with TPC method. It showed that the differences of LF and HF between segments A and B and between B and C were lower than the differences between A and C with TPC. The TPC could minimize the error in correcting segment B. In addition, the ectopic beats were detected before each correction method. The ectopic detecting method such as Eq. (1) is not a point in our work.
Detrending decreased the LF by approximately 9% and increased HF by approximately 1% as shown in Table 1. The correction result of the TPC contained a lesser error ratio when considering the effect of detrending. The SIC method had superior performance when dealing with the simulated HRV signal with ectopic beats. SIC had small differences from TPC during the segments A and C of clinical data and may be due to the fact that some RRIs were corrected by TPC. Otherwise, the RRI after SIC may be not suitable to calculate the time domain indexes of HRV.
The preceding normal RRI with the numbers of 8 to 16 is applied to calculate the predictive RRI in TPC, and sometimes the RRI sequence behind the ectopic beat could be used. The total time of the procedure to deal with 300-point episode of simulated HRV signal is no more than 1 s where it requires 114 s to deal with 200 simulated signal numbers. According to the simulation and clinical test, the TPC had a real-time advantage to recover the SN modulation. The indexes of HRV analysis are widely used; however, the HRV estimation could be more accurate with efficient preprocessing methods.
5. Conclusions
The new ectopic correction method was proposed to deal with ectopic beats in the HRV frequency analysis. This method has an advantage to recover the SN rhythm through the PSD estimation by simulated data and clinical data when compared with other methods. The correction method is useful to the real-time HRV analysis. Lastly, this preprocessing method requires further investigation.
Acknowledgments
The authors would like to thank Drs. Qiu-ping XU and Xiao-mei YIN (Sir Run Run Shaw Hospital, Zhejiang University, China) for clinical support.
Footnotes
Project supported by the Key Industrial Projects of Science of Zhejiang Province (No. 2007C21079), and the National Key Technologies R&D Program (No. 2007BAI07A28) of China
References
- 1.Acar B, Savelieva I, Hemingway H, Malik M. Automatic ectopic beat elimination in short-term heart rate variability measurement. Comput Methods Programs Biomed. 2000;63(2):123–131. doi: 10.1016/S0169-2607(00)00081-X. [DOI] [PubMed] [Google Scholar]
- 2.Amar D, Zhang H, Miodownik S, Kadish AH. Competing autonomic mechanisms precede the onset of postoperative atrial fibrillation. J Am Coll Cardiol. 2003;42(7):1262–1268. doi: 10.1016/S0735-1097(03)00955-0. [DOI] [PubMed] [Google Scholar]
- 3.Bettoni M, Zimmermann M. Autonomic tone variations before the onset of paroxysmal atrial fibrillation. Circulation. 2002;105(23):2753–2759. doi: 10.1161/01.CIR.0000018443.44005.D8. [DOI] [PubMed] [Google Scholar]
- 4.Camm AJ, Malik M, Bigger JT, Breithardt G, Cerutti S, Cohen RJ, Coumel P, Fallen EL, Kennedy HL, Kleiger RE, et al. Heart rate variability: standards of measurement, physiological interpretation, and clinical use. Circulation. 1996;93(5):1043–1065. doi: 10.1161/01.CIR.93.5.1043. [DOI] [PubMed] [Google Scholar]
- 5.Colak OH. Preprocessing effects in time-frequency distributions and spectral analysis of heart rate variability. Digital Signal Process. 2009;19(4):731–739. doi: 10.1016/j.dsp.2008.09.004. [DOI] [Google Scholar]
- 6.Dimmer C, Tavernier R, Gjorgov N, van Nooten G, Clement DL, Jordaens L. Variations of autonomic tone preceding onset of atrial fibrillation after coronary artery bypass grafting. Am J Cardiol. 1998;82(1):22–25. doi: 10.1016/S0002-9149(98)00231-8. [DOI] [PubMed] [Google Scholar]
- 7.Huang JL, Wen ZC, Lee WL, Chang MS, Chen SA. Changes of autonomic tone before the onset of paroxysmal atrial fibrillation. Int J Cardiol. 1998;66(3):275–283. doi: 10.1016/S0167-5273(98)00241-1. [DOI] [PubMed] [Google Scholar]
- 8.Kim KK, Kim JS, Lim YG, Park KS. The effect of missing RR-interval data on heart rate variability analysis in the frequency domain. Physiol Meas. 2009;30(10):1039–1050. doi: 10.1088/0967-3334/30/10/005. [DOI] [PubMed] [Google Scholar]
- 9.Li H, Ma XQ, Ye F, Zhang J, Zhou X, Wang ZH, Li YM, Zhang GY. Expressions of cardiac sympathetic norepinephrine transporter and β1-adrenergic receptor decreased in aged rats. J Zhejiang Univ-Sci B. 2009;10(3):203–210. doi: 10.1631/jzus.B0820213. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Lombardi F, Tarricone D, Tundo F, Colombo F, Belletti S, Fiorentini C. Autonomic nervous system and paroxysmal atrial fibrillation: study based on the analysis of RR interval changes before, during and after paroysmal atrial fibrillation. Eur Heart J. 2004;25(14):1242–1248. doi: 10.1016/j.ehj.2004.05.016. [DOI] [PubMed] [Google Scholar]
- 11.Maestri R, Raczak G, Danilowicz-Szymanowicz L, Torunski A, Sukiennik A, Kubica J, La Rovere MT, Pinna GD. Reliability of heart rate variability measurements in patients with a history of myocardial infarction. Clin Sci. 2010;118(3):195–201. doi: 10.1042/CS20090183. [DOI] [PubMed] [Google Scholar]
- 12.Mateo J, Laguna P. Analysis of heart rate variability in the presence of ectopic beats using the heart timing signal. IEEE Trans Biomed Eng. 2003;50(3):334–343. doi: 10.1109/TBME.2003.808831. [DOI] [PubMed] [Google Scholar]
- 13.Mitov IP. A method for assessment and processing of biomedical signals containing trend and periodic components. Med Eng Phys. 1998;20(9):660–668. doi: 10.1016/S1350-4533(98)00077-0. [DOI] [PubMed] [Google Scholar]
- 14.Miyakoshi M, Ikeda T, Miwa Y, Sakaki K, Ishiguro H, Abe A, Tsukada T, Mera H, Yusu S, Yoshino H. Quantitative assessment of cibenzoline administration for vagally mediated paroxysmal atrial fibrillation using frequency-domain heart rate variability analysis. J Cardiol. 2009;54(1):86–92. doi: 10.1016/j.jjcc.2009.04.009. [DOI] [PubMed] [Google Scholar]
- 15.Myers GA, Martin GJ, Magin NM, Barnett PS, Schaad JW, Weiss JS, Lesch M, Sigher DH. Power spectral analysis of heart rate variability in sudden cardiac death: comparison to other methods. IEEE Trans Biomed Eng. 1986;33(12):1149–1156. doi: 10.1109/TBME.1986.325694. [DOI] [PubMed] [Google Scholar]
- 16.Myers G, Workman M, Birkett C, Ferguson D, Kienzle M. Problems in measuring heart rate variability of patients with congestive heart failure. J Electrocardiol. 1992;25:214–219. doi: 10.1016/0022-0736(92)90105-9. [DOI] [PubMed] [Google Scholar]
- 17.Ovreiu M, Nair BG, Xu M, Bakri MH, Li L, Wazni O, Fahmy T, Petre J, Starr NJ, Sessler DI, et al. Electrocardiographic activity before onset of postoperative atrial fibrillation in cardiac surgery patients. Pacing Clin Electrophysiol. 2008;31(11):1371–1382. doi: 10.1111/j.1540-8159.2008.01198.x. [DOI] [PubMed] [Google Scholar]
- 18.Sapoznikov D, Luria MH, Mahler Y, Gostman MS. Computer-processing of artifact and arrhythmias in heart-rate-variability analysis. Comput Methods Programs Biomed. 1992;39(1-2):75–84. doi: 10.1016/0169-2607(92)90060-K. [DOI] [PubMed] [Google Scholar]
- 19.Solem K, Laguna P, Sornmo L. An efficient method for handling ectopic beats using the heart timing signal. IEEE Trans Biomed Eng. 2006;53(1):13–20. doi: 10.1109/TBME.2005.859780. [DOI] [PubMed] [Google Scholar]
- 20.Taddei A, Distante G, Emdin M, Pisani P, Moody GB, Zeelenberg C, Marchesi C. The European ST-T database: standard for evaluating systems for the analysis of ST-T changes in ambulatory electrocardiography. Eur Heart J. 1992;13(9):1164–1172. doi: 10.1093/oxfordjournals.eurheartj.a060332. [DOI] [PubMed] [Google Scholar]
- 21.Tarvainen MP, Ranta-aho PO, Karjalainen PA. An advanced detrending method with application to HRV analysis. IEEE Trans Biomed Eng. 2002;49(2):172–175. doi: 10.1109/10.979357. [DOI] [PubMed] [Google Scholar]
- 22.van Schelven LJ, Oey PL, Klein IHI, Barnas MGW, Blankestijn PJ, Wieneke GH. Observer variations in short period spectral analysis of heart rate variability. J Autonom Nerv Syst. 2000;79(2-3):144–148. doi: 10.1016/S0165-1838(99)00106-X. [DOI] [PubMed] [Google Scholar]
- 23.Yamada T, Yoshida N, Murakami Y, Okada T, Yoshida Y, Muto M, Inden Y, Murohara T. Vagal modification can be a valid predictor of late recurrence of paroxysmal atrial fibrillation independent of the pulmonary vein isolation technique. Circ J. 2009;73(9):1606–1611. doi: 10.1253/circj.CJ-09-0158. [DOI] [PubMed] [Google Scholar]
- 24.Zimmermann M, Kalusche D. Fluctuation in autonomic tone is a major determinant of sustained atrial arrhythmias in patients with focal ectopy originating from the pulmonary veins. J Cardiovasc Electrophysiol. 2001;12(3):285–291. doi: 10.1046/j.1540-8167.2001.00285.x. [DOI] [PubMed] [Google Scholar]