Abstract
Since its introduction as an ionization technique in mass spectrometry, matrix assisted laser desorption ionization (MALDI) has been applied to a wide range of applications. Quantitative small molecule analysis by MALDI, however, is limited due to the presence of intense signals from the matrix coupled with non-homogeneous surfaces. The surface used in nanostructured laser desorption ionization (NALDI) eliminates the need for a matrix and the resulting interferences, and allows for quantitative analysis of small molecules. This study was designed to analyze and quantitate phospholipid components of liposomes.
Here we have developed an assay to quantitate the DPPC and DC8,9PC in liposomes by NALDI following various treatments. To test our method we chose to analyze a liposome system composed of DPPC (1,2-dipalmitoyl-sn-glycero-3-phosphocholine) and DC8,9PC (1,2-bis (tricosa-10,12-diynoyl)-sn-glycero-3-phosphocholine), as DC8,9PC is known to undergo cross-linking upon treatment with UV (254 nm) and this reaction converts the monomer into a polymer.
First, calibration curves for pure lipids (DPPC and DC8,9PC) were created using DMPC (1,2-dimyristoyl-sn-glycero-3-phosphocholine) as an internal standard. The calibration curve for both DPPC and DC8,9PC showed an R2 of 0.992, obtained using the intensity ratio of analyte and internal standard. Next, DPPC:DC8,9PC liposomes were treated with UV radiation (254 nm). Following this treatment, lipids were extracted from the liposomes and analyzed. The analysis of the lipids before and after UV exposure confirmed a decrease in the signal of DC8,9PC of about 90%. In contrast, there was no reduction in DPPC signal.
Keywords: NALDI, MALDI, mass spectrometry, lipids, quantitation, drug delivery
Introduction
Chromatography has been the traditional technique of choice for the analysis of lipids. Thin-layer chromatography [1] and liquid chromatography [2] have been the most widely used methods in the field, especially in combination with UV detection [3]. Those techniques are well-established and robust but they are also time-consuming, because they often involve steps of derivatization and long separation times [4]. Spectroscopic methods, such as IR and NMR, have been applied in the lipid analysis field as well [3]. The first use of mass spectrometry for the analysis of lipids was reported by Gohlke in 1959 [5] using gas chromatography mass spectrometry (GC-MS). Mass spectrometry provides information on the molecular weight of the species of interest that can be useful for identification purposes. Since then GC-MS has been applied to many analytical questions regarding the biology and chemistry of lipids [6], however GC-MS may require derivatization when the analyte is not volatile and thermally stable [7]. While derivatization may help, in some cases it results in additional steps that can complicate the entire analysis [8].
In recent years with the introduction of the so-called “soft ionization” techniques, electrospray ionization and matrix-assisted laser desorption ionization, [9] that allow for the analysis of intact lipids [10], mass spectrometry has gained importance in the analysis of lipid mixtures [11] and in the rapidly growing field of mass spectrometry tissue imaging [12]. Electrospray ionization (ESI) is the most frequently used mass spectrometry technique for lipids analysis, but matrix-assisted laser desorption ionization (MALDI) represents a powerful tool in this field, because the time necessary for the analysis can be reduced to less than one minute per sample [10]. ESI mass spectrometry in association with liquid chromatography (LC) allows for separation and identification of molecular species, however analysis times can be long and the buffers used for LC/MS are often incompatible with lipid solubility.
In MALDI, the sample is mixed with a solution of matrix that absorbs the laser energy. The mixture is allowed to dry to form co-crystals on the surface of target. The homogeneity of the solid crystals on the plate is essential for the reproducibility of the analysis, therefore accurate sample preparation is necessary [10]. In most cases, the matrix is soluble in the same organic solvents as the lipid analytes, so very homogenous co-crystals are produced facilitating the analysis.
The choice of the most suitable matrix is often quite challenging. While water soluble analytes have a wide range of useful matrices [13], only a few matrices have been utilized for lipid analysis due to their lipophilic nature. Recently Teuber et al. [14] compared a series of matrices for lipids. Along with the most commonly used DHB (2,5 - dihydroxybenzoic acid), other matrices showed interesting characteristics, especially a newly synthesized compound, α-cyano-2,4-difluorocinnamic acid, for analysis of charged phospholipids, and 9-aminoacridine for negative ion analysis. For small molecule analysis it is necessary to choose a matrix that does not generate polymers by photochemical reactions (e.g. sinapinic acid), because the polymerization products will saturate the detector [4] and suppress the signal from the analytes. In addition the presence of interfering peaks due to matrix abducts (e.g. DHB with Na+), [15] in the lower m/z range is a crucial issue in the analysis of small molecules, therefore other approaches have been taken into consideration. Some studies suggested the use of high molecular weight matrices such as meso-tetrakis(pentafluorophenyl)porphyrin [16–18] for the analysis and quantification of fatty acids. In this case the signals due to molecular ions of the matrix or its impurities, adducts or clusters are in a higher mass range and do not overlap with the analyte signal.
Recently matrix-free laser desorption ionization (LDI) has become a promising approach in the field of small molecules. In this case the sample is applied onto a surface that absorbs the laser energy without the use of additional compounds. The direct deposition of the sample involves minimal sample preparation and practically no signal from the support surface [15]. Silicon has been one of the first materials to be used as a suitable support for matrix-free LDI [19]. Its surface can be easily oxidized and modified to trap the analyte and it has high absorptivity in the UV, therefore it can act like the matrix itself. This technique, termed desorption ionization on silica (DIOS), has been applied to a number of small molecule analytes [20]. The use of silicon as a LDI mass spectrometry support has evolved with the recent introduction of silicon nanowires [21]. Muck et al. provided evidence of the superiority of this approach compared with other traditional techniques that use energy absorbing matrices of questionable purity and stability [22]. Other nanostructures have interesting characteristics. Carbon nanotubes have shown low signal background and good capability to transfer energy to the analytes [23–25]. Carbon nanotubes often have solubility issues, however. Ren et al. used oxidization to improve solubility and produced a more homogeneous phase with reproducible analysis [25].
An issue with the use of nano-structured materials is that they are produced by a random process [20] that cannot always guarantee a homogeneous surface where the sample should be applied. In contrast, ZnO nanowires can be selectively grown on a substrate, are easy to synthesize, and their orientation can be controlled, so that on a flat substrate they can be organized as vertical wire arrays with defined dimension and structure [26]. ZnO presents characteristics that are particularly suitable for many industrial applications [27, 28]. It is a semi-conductive material that can absorb energy from the laser in wavelengths typically used in MALDI and transfer it to the analyte. ZnO nanoparticles have been successfully applied to the desorption/ionization of several small molecules, from drugs to lipids [29]. Kang et al. compared a suspension of analyte/ZnO nanowires and the direct deposition of the analyte onto the surface of a nanowire chip [28]. The nanowire chip showed better performance and a potential applicability to quantitive analysis.
In this study we demonstrate how a NALDI™ chip (Bruker Daltonics) can be used as a support for desorption/ionization applied for small molecule quantitative analysis. We have examined two phospholipids, DPPC and DC8,9PC, that have been recently used for development of photo-triggerable liposomes [30–32]. Our system represents a novel and elegant method to determinate the concentrations of the two phospholipids of interest in the liposomes. This methodology bears the potential for future applications for the analysis of molecules with similar chemical and physical properties.
Methods and Materials
Liposome preparation and treatment
Liposome Reagents
DPPC (1,2-dipalmitoyl-sn-glycero-3-phosphocholine) and DMPC (1,2-dimyristoyl-sn-glycero-3-phosphocholine) were purchased from Avanti Polar Lipids, Inc. (Alabaster, AL) as solutions in chloroform. DC8,9PC (1,2-bis (tricosa-10,12-diynoyl)-sn-glycero-3-phosphocholine) was synthesized by Dr. Alok Singh at the Naval Research Laboratory (Washington, DC) using a previously published procedure [30]. Methanol HPLC grade was purchased from OmniSolv (EMD Chemicals, Inc., Gibbstown, NJ), while Chloroform ACS was from Fischer Scientific (Pittsburg, PA).
Preparation of Liposomes
Liposomes were prepared using the probe sonication essentially as described [30]. Briefly, DPPC:DC8,9PC (80:20 mol%) lipid mixtures were mixed in a glass tube. The lipid film was formed by removing the solvent under nitrogen and any residual chloroform was removed by placing the film overnight in a vacuum desiccator. To encapsulate calcein, the lipid film was reconstituted in HEPES buffer (HBS, 10 mM HEPES, 140 mM NaCl, pH 7.5) containing self-quenched concentration of calcein (0.1 M at pH=7.2–7.6). Unilamellar Vesicles were formed by sonication at 4°C for 5–10 min (1 min pulses and 1 min rest) using a Probe Sonicator (W-375 Heat Systems-Ultrasonics, New York, USA). The samples were centrifuged to remove any titanium particles and larger aggregates. Solute-loaded liposomes were separated from unentrapped calcein using a size exclusion gel chromatography column (Bio Gel A0.5m, 1 × 40 cm, 40 ml bed volume).
UV Treatment
Liposomes were placed in a 96-well plate (total volume of 0.2 ml ) as described [23], and irradiated with a UV lamp (UVP, short wave assembly, 115V, 60Hz) at 254 nm in a distance of 1 inch for 45 minutes. We routinely examined calcein leakage and DC8,9PC photo-crosslinking to confirm the modification in our liposome preparations as previously described [30, 31]. For mass spectrometry analysis (see below) lipids were extracted according to the Bligh/Dyer method [33].
Mass Spectrometry Analysis
Reagents
Methanol was HPLC grade (OmniSolv, EMD Chemicals, Inc., Gibbstown, NJ). Chloroform was HPLC grade from Chromasolv (Sigma-Aldrich, St. Louis, MO). Water was LC-MS grade from J.T. Baker (Mallinckrodt Baker Inc., Phillipsburg, NJ). Cesium chloride (CsCl, 99.9%) was purchased from Sigma-Aldrich (St. Louis, MO).
Sample Preparation
Sample preparation used only glassware and Wiretrol disposable glass micropipettes (Drummond Scientific Company, Broomall, PA). The lipid films obtained from Bligh-Dyer extraction were reconstituted in 200 μL of chloroform, then diluted 1:10 in a mixture methanol/chloroform, 1/1, v/v. Standard solutions for each lipid were prepared in methanol/chloroform, 1/1, v/v. DMPC was used as an internal standard at a concentration of 25 ng/μL in methanol/chloroform, 1/1, v/v. A CsCl solution, 50 mM in methanol, was used as a matrix additive to promote formation of [M+Cs] pseudo molecular ions for the lipids studied. Calibration curves were created for the phospholipid combination by mixing DC8,9PC in equal volume (5 μL) with the corresponding dilution of DPPC, followed by addition of 5 μL of DMPC 25 ng/μL and 15 μL of CsCl 50mM. Concentration ranges for DPPC and DC8,9PC were 10–200 ng/μL and 2.5–100 ng/μL, respectively. We washed the NALDI target twice with HPLC grade chloroform and allowed to dry prior to use. The standard solutions were mixed and 3 μL spotted onto the surface of a NALDI chip (Bruker Daltonics, Billerica, MA). Standards were spotted in triplicate. In a similar fashion, 10 μL of the diluted sample solutions were mixed with DMPC 25 ng/μL (5 μL) and CsCl 50 mM (15 μL) and 3 μL of each solution spotted in triplicate onto the surface of the same target where the corresponding standard solutions were spotted.
Mass Spectrometry Analysis
All spectra were acquired on a Bruker Ultraflex III TOF/TOF mass spectrometer (Bruker Daltonics, Billerica, MA). Spectra of intact lipids were acquired in positive ion reflector mode, with no ion extraction delay. Ion source voltage 1 was 25kV, ion source voltage 2 was 21.30 kV and lens voltage was 9.52 kV. Reflectron and reflectron 2 were 29.50 kV and 13.84 kV, respectively. Monoisotopic masses were determined using FlexAnalysis 3.0 (Bruker Daltonics) with the SNAP peak picking algorithm. The instrument was calibrated in reflector mode using peptide standards of the Bruker peptide calibration kit (Bruker Daltonics, Billerica, MA). Data acquisition and peak analysis were completely automated with an accumulation of 1200 shots using a random walk pattern. To create the calibration curves, intensity ratios of the analyte signal relative to the intensity of the internal standard were calculated and averaged from the three measurements at each concentration.
Results and Discussion
Phospholipids are the basis of cellular structures and because of their amphiphilic properties, they have the natural tendency to form a bilayers in an aqueous system [3]. These unique properties allowed A.D. Bangham to formulate the first preparation of liposomes with encapsulated materials in 1965 [34]. Since then, scientists have concentrated their efforts to take advantage of the characteristics of liposomes to understand the biophysical properties of membranes, study lipid interactions and to create efficient drug delivery systems [35–38]. The evolution of this concept led Yavlovich et al. [30] to design an innovative liposome formulation for potential use as an efficient drug release system in appropriate organs and at the appropriate times. They were able to co-assemble several phospholipids (e.g. DPPC and EggPC) and DC8,9PC in the liposome bilayer. DC8,9PC, a phospholipid with reactive diacetylenic groups, is known to undergo a photo cross-linking reaction when treated with UV radiation (254nm) [39, 40]. Similar photo-crosslinking was observed when DC8,9PC was embedded in DPPC bilayers at mole ratios 1:9 and above leading to release of entrapped liposome contents (30). It was hypothesized that photo cross-linking generates local defects in the liposome membrane that can promote drug release.
The detailed analysis of the individual lipids in these liposomes is critical to understand the nature of molecular interactions that result in photo-crosslinking and such information may improve design of these formulations. Initially, we set out to develop a quantitative method for the analysis of these lipids by MALDI mass spectrometry. MALDI-TOF MS has not been considered the first choice for quantitative analysis of lipids or other small molecular weight molecules for two main reasons: first the presence of matrix peaks and adducts in the low molecular range can interfere with analyte peak detection and second, inhomogeneity of the matrix/sample crystallization process reduces shot-to-shot reproducibility. Therefore, a matrix-free approach was chosen to avoid the disadvantages of the matrix and selected nanowire targets which had been shown to be applicable to quantitative analysis of small molecules [28, 41].
In our initial method development we observed in the lipid spectra the presence of multiple peaks related to the analyte of interest – protonated and sodiated adducts – and realized this could present difficulties in data analysis and reproducibility of the assay. Considering the work of Shiller et al. [42], we decided to use CsCl as a cationization agent. Using CsCl, the MS signal of the phospholipids was shifted by 133 Da corresponding to the atomic mass of cesium. Other groups chose cesium acetate [18] and they were also able to obtain dominant cesiated adducts. Our analytes, dissolved in a mixture of chloroform/methanol, were mixed with a methanolic solution of cesium chloride prior to spotting on the NALDI target. Cesium chloride is soluble in water but not in chloroform/methanol; therefore we prepared a stock solution of 1M cesium chloride in water and diluted to our working concentration – 50mM – in methanol, so it would be miscible with our analyte dilution buffer (chloroform/methanol, 50/50, v/v). We tried to use other, less volatile solvents to facilitate the sample dilution procedure, but a decrease in the quality of the spots was observed, and consequently of the analysis, probably due to solubility issues related to the lipophilic nature of the analytes.
The choice of organic solvents such as chloroform and methanol necessitated the use of glassware and glass micropipettors for the sample preparation. We observed an advantage in the use of volatile solvent systems: the rapid evaporation of the solvent mixture promoted an efficient co-crystallization of lipids and cesium chloride, resulting in homogeneous spots, as previously observed by other investigators [43]. Homogenous spots, along with triplicate analyses, allowed us to control variability issues, such as shot-to-shot, region-to-region and sample-to-sample [43, 44] variability. The possibility of bias acquisition was considered and therefore we opted for a completely automated acquisition of data collecting 1200 shots, in a random walk pattern and with an intensity of laser between 15% and 60% in order to avoid saturation of the instrument. A better reproducibility was observed with the average of 1200 spectra, rather than with a smaller number of acquisitions for each spot.
Release of contents from DPPC: DC8,9PC 4:1 liposomes occurs as a result of UV-triggered DC8,9PC polymerization in the bilayer [30]. Therefore the liposomes were loaded with 50 mM calcein and its release was monitored after treatment with UV (254 nm) for 45 minutes using a fluorescent micro plate reader (Methods section). Before and after UV treatment liposome lipids were extracted and analyzed by the NALDI mass spectrometry method.
MALDI signals for DPPC and DC8,9PC were linear over the selected ranges and the calibration curves resulted in coefficient of determinations (R2) = 0.99, as shown in Figure 1. Figure 2 shows two examples of spectra obtained from the lipids extracted and treated as described above. The decrease in intensity of the signal of the pseudo molecular ion for DC8,9PC after UV treatment is evident. We were able to determine the concentration for both DPPC and DC8,9PC using the equation from the linear regression for each compound. Figure 3 compares the concentration, expressed in μM, for both species before and after UV treatment. The concentration of DPPC remained constant (260 μM) while DC8,9PC decreased from 120 μM in the starting liposome to 12 μM in the liposome treated with 254 nm irradiation.
Figure 1.
Calibration curves for DPPC and DC8,9PC. Concentration ranges for DPPC and DC8,9PC were respectively 10–200 ng/μL and 2.5–100 ng/μL.
Figure 2.
Mass spectra of phospholipids from liposome before (A) and after (B) UV treatment (254 nm).
Figure 3.
Quantitative comparison of DPPC and DC8,9PC in liposome before and after UV treatment. A decrease of DC8,9PC from 120 μM to 12 μM, before and after the UV treatment, was observed, while the concentration of DPPC remained unchanged following exposure of the intact liposome at 254 nm irradiation.
It is important to mention that the possibility of ion suppression and ionization efficiencies of the individual lipids we analyzed was considered. This is a common drawback in any mass spectrometry assay and we attempted to overcome this issue with the use of DMPC as the internal standard as it is structurally similar to the analytes of interest. The use of internal standards for quantitative analysis is standard practice [18, 45–47] and it is of crucial importance in any MALDI approach to these types of analyses. However in some cases a suitable internal standard may not be available, but reasonable quantitation may still be possible [48]. For optimal results, the use of a stable isotope analog of the analyte of interest would be the primary choice [47] for an internal standard, but in our case the available stable label analog of DPPC was not of sufficient purity for use as an internal standard so we choose a slightly smaller saturated PC as our internal standard.
By application of this novel MS method to the analysis of the phospholipids extracted from liposomes prepared and treated as described by Yavlovich et al. [30], we were able to confirm the decrease of DC8,9PC from 120 μM to 12 μM, before and after the UV treatment (Figure 3). As expected DPPC concentration did not change. It has been demonstrated that DC8,9PC liposomal photo-triggering by UV light occurs via direct photopolymerization mechanism [30].
Conclusion
The NALDI-based analysis of the phospholipids presented here provides a fast and reproducible approach for the quantitative analysis of lipids in relatively complex mixtures. We have demonstrated that our method is sufficiently specific for the lipids of interest because it was not necessary to use tandem mass spectrometry as suggested by other investigators [15, 46, 49]. The use of the NALDI targets allows for greatly simplified sample preparation and eliminates the matrix-associated issues of inhomogeneous crystals and matrix adducts. This approach should also be possible to for a wide range of lipids and structurally similar molecules.
Acknowledgments
This research was supported [in part] by federal funds from the Intramural Research Program of the NIH, National Cancer Institute, Center for Cancer Research. This project has been funded in part with federal funds from the National Cancer Institute, National Institutes of Health, under Contract No. HHSN261200800001E. The content of this publication does not necessarily reflect the views or policies of the Department of Health and Human Services, nor does mention of trade names, commercial products, or organizations imply endorsement by the U.S. Government.
Footnotes
List of abbreviation terms: DC8,9PC: (1,2-bis (tricosa-10,12-diynoyl)-sn-glycero-3-phosphocholine), DHB: 2,5 - dihydroxybenzoic acid, DIOS: desorption ionization on silica, DMPC: 1,2-dimyristoyl-sn-glycero-3-phosphocholine, DPPC: 1,2-dipalmitoyl-sn-glycero-3-phosphocholine, ESI: Electrospray ionization, GC: Gas chromatography, HPLC: High-performance liquid chromatography, LC: Liquid chromatography, LDI: laser desorption ionization, MALDI: Matrix-assisted laser desorption/ionization, MS: Mass Spectrometry, NALDI: Nano-Assisted Laser Desorption/Ionization, TOF: Time-of-flight
References
- 1.Touchstone JC. Thin-layer chromatographic procedures for lipid separation. J Chromatogr B Biomed Appl. 1995;671:169–195. doi: 10.1016/0378-4347(95)00232-8. [DOI] [PubMed] [Google Scholar]
- 2.McCluer RH, Ullman MD, Jungalwala FB. HPLC of glycosphingolipids and phospholipids. Adv Chromatogr. 1986;25:309–353. [PubMed] [Google Scholar]
- 3.Christie WW. Lipid Analysis. Bridgewater: Oily Press; 2003. [Google Scholar]
- 4.Schiller J, Süß R, Arnhold J, Fuchs B, et al. Matrix-assisted laser desorption and ionization time-of-flight (MALDI-TOF) mass spectrometry in lipid and phospholipid research. Progress in Lipid Research. 2004;43:449–488. doi: 10.1016/j.plipres.2004.08.001. [DOI] [PubMed] [Google Scholar]
- 5.Gohlke RS. Time-of-Flight Mass Spectrometry and Gas-Liquid Partition Chromatography. Analytical Chemistry. 1959;31:535–541. [Google Scholar]
- 6.Ryhage R. The Mass Spectrometry Laboratory at the Karolinska Institute 1944–1987. Mass Spectrometry Reviews. 1993;12:1–49. [Google Scholar]
- 7.Griffiths WJ. Tandem mass spectrometry in the study of fatty acids, bile acids, and steroids. Mass Spectrometry Reviews. 2003;22:81–152. doi: 10.1002/mas.10046. [DOI] [PubMed] [Google Scholar]
- 8.Harvey DJ, Bateman BH, Bordoli RS, Tyldesley R. Ionisation and fragmentation of complex glycans with a quadrupole time-of-flight mass spectrometer fitted with a matrix-assisted laser desorption/ionisation ion source. Rapid Communications in Mass Spectrometry. 2000;14:2135–2142. doi: 10.1002/1097-0231(20001130)14:22<2135::AID-RCM143>3.0.CO;2-#. [DOI] [PubMed] [Google Scholar]
- 9.Siuzdak G. The emergence of mass spectrometry in biochemical research. Proc Natl Acad Sci U S A. 1994;91:11290–11297. doi: 10.1073/pnas.91.24.11290. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Schiller J, Suss R, Fuchs B, Muller M, et al. MALDI-TOF MS in lipidomics. Front Biosci. 2007;12:2568–2579. doi: 10.2741/2255. [DOI] [PubMed] [Google Scholar]
- 11.Wolf C, Quinn PJ. Lipidomics: practical aspects and applications. Prog Lipid Res. 2008;47:15–36. doi: 10.1016/j.plipres.2007.09.001. [DOI] [PubMed] [Google Scholar]
- 12.Isaac G, Jeannotte R, Esch SW, Welti R. New mass-spectrometry-based strategies for lipids. Genet Eng (N Y) 2007;28:129–157. doi: 10.1007/978-0-387-34504-8_8. [DOI] [PubMed] [Google Scholar]
- 13.Fitzgerald MC, Parr GR, Smith LM. Basic matrices for the matrix-assisted laser desorption/ionization mass spectrometry of proteins and oligonucleotides. Anal Chem. 1993;65:3204–3211. doi: 10.1021/ac00070a007. [DOI] [PubMed] [Google Scholar]
- 14.Teuber K, Schiller J, Fuchs B, Karas M, Jaskolla TW. Significant sensitivity improvements by matrix optimization: a MALDI-TOF mass spectrometric study of lipids from hen egg yolk. Chem Phys Lipids. 2010 doi: 10.1016/j.chemphyslip.2010.04.005. [DOI] [PubMed] [Google Scholar]
- 15.van Kampen JJ, Burgers PC, de Groot R, Gruters RA, Luider TM. Biomedical application of MALDI mass spectrometry for small-molecule analysis. Mass Spectrom Rev. 2010 doi: 10.1002/mas.20268. [DOI] [PubMed] [Google Scholar]
- 16.Ayorinde FO, Garvin K, Saeed K. Determination of the fatty acid composition of saponified vegetable oils using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Rapid Commun Mass Spectrom. 2000;14:608–615. doi: 10.1002/(SICI)1097-0231(20000415)14:7<608::AID-RCM918>3.0.CO;2-4. [DOI] [PubMed] [Google Scholar]
- 17.Hlongwane C, Delves IG, Wan LW, Ayorinde FO. Comparative quantitative fatty acid analysis of triacylglycerols using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry and gas chromatography. Rapid Commun Mass Spectrom. 2001;15:2027–2034. doi: 10.1002/rcm.462. [DOI] [PubMed] [Google Scholar]
- 18.Yu H, Lopez E, Young SW, Luo J, et al. Quantitative analysis of free fatty acids in rat plasma using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry with meso-tetrakis porphyrin as matrix. Analytical Biochemistry. 2006;354:182–191. doi: 10.1016/j.ab.2006.04.050. [DOI] [PubMed] [Google Scholar]
- 19.Wei J, Buriak JM, Siuzdak G. Desorption-ionization mass spectrometry on porous silicon. Nature. 1999;399:243–246. doi: 10.1038/20400. [DOI] [PubMed] [Google Scholar]
- 20.Peterson DS. Matrix-free methods for laser desorption/ionization mass spectrometry. Mass Spectrometry Reviews. 2007;26:19–34. doi: 10.1002/mas.20104. [DOI] [PubMed] [Google Scholar]
- 21.Go EP, Apon JV, Luo G, Saghatelian A, et al. Anal Chem. 2005;77:1641–1646. doi: 10.1021/ac048460o. [DOI] [PubMed] [Google Scholar]
- 22.Muck A, Stelzner T, Hubner U, Christiansen S, Svatos A. Lithographically patterned silicon nanowire arrays for matrix free LDI-TOF/MS analysis of lipids. Lab Chip. 2010;10:320–325. doi: 10.1039/b913212k. [DOI] [PubMed] [Google Scholar]
- 23.Xu S, Li Y, Zou H, Qiu J, et al. Carbon Nanotubes as Assisted Matrix for Laser Desorption/Ionization Time-of-Flight Mass Spectrometry. Analytical Chemistry. 2003;75:6191–6195. doi: 10.1021/ac0345695. [DOI] [PubMed] [Google Scholar]
- 24.Ren SF, Zhang L, Cheng ZH, Guo YL. Immobilized carbon nanotubes as matrix for MALDI-TOF-MS analysis: applications to neutral small carbohydrates. J Am Soc Mass Spectrom. 2005;16:333–339. doi: 10.1016/j.jasms.2004.11.017. [DOI] [PubMed] [Google Scholar]
- 25.Ren SF, Guo YL. Oxidized carbon nanotubes as matrix for matrix-assisted laser desorption/ionization time-of-flight mass spectrometric analysis of biomolecules. Rapid Commun Mass Spectrom. 2005;19:255–260. doi: 10.1002/rcm.1779. [DOI] [PubMed] [Google Scholar]
- 26.Greene LE, Yuhas BD, Law M, Zitoun D, Yang P. Solution-Grown Zinc Oxide Nanowires. Inorganic Chemistry. 2006;45:7535–7543. doi: 10.1021/ic0601900. [DOI] [PubMed] [Google Scholar]
- 27.Fan Z, Lu JG. Zinc oxide nanostructures: synthesis and properties. J Nanosci Nanotechnol. 2005;5:1561–1573. doi: 10.1166/jnn.2005.182. [DOI] [PubMed] [Google Scholar]
- 28.Kang MJ, Pyun JC, Lee JC, Choi YJ, et al. Nanowire-assisted laser desorption and ionization mass spectrometry for quantitative analysis of small molecules. Rapid Communications in Mass Spectrometry. 2005;19:3166–3170. [Google Scholar]
- 29.Watanabe T, Kawasaki H, Yonezawa T, Arakawa R. Surface-assisted laser desorption/ionization mass spectrometry (SALDI-MS) of low molecular weight organic compounds and synthetic polymers using zinc oxide (ZnO) nanoparticles. Journal of Mass Spectrometry. 2008;43:1063–1071. doi: 10.1002/jms.1385. [DOI] [PubMed] [Google Scholar]
- 30.Yavlovich A, Singh A, Tarasov S, Capala J, et al. Design of Liposomes Containing Photopolymerizable Phospholipids for Triggered Release of Contents. J Therm Anal Calorim. 2009;98:97–104. doi: 10.1007/s10973-009-0228-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Yavlovich A, Singh A, Blumenthal R, Puri A. A novel class of photo-triggerable liposomes containing DPPC:DC(8,9)PC as vehicles for delivery of doxorubcin to cells. Biochim Biophys Acta. 2010 doi: 10.1016/j.bbamem.2010.07.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Puri A, Loomis K, Smith B, Lee JH, et al. Lipid-based nanoparticles as pharmaceutical drug carriers: from concepts to clinic. Crit Rev Ther Drug Carrier Syst. 2009;26:523–580. doi: 10.1615/critrevtherdrugcarriersyst.v26.i6.10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bligh EG, Dyer WJ. A rapid method of total lipid extraction and purification. Can J Biochem Physiol. 1959;37:911–917. doi: 10.1139/o59-099. [DOI] [PubMed] [Google Scholar]
- 34.Bangham AD, Standish MM, Watkins JC. Diffusion of univalent ions across the lamellae of swollen phospholipids. J Mol Biol. 1965;13:238–252. doi: 10.1016/s0022-2836(65)80093-6. [DOI] [PubMed] [Google Scholar]
- 35.Gregoriadis G, Neerunjun ED. Homing of liposomes to target cells. Biochem Biophys Res Commun. 1975;65:537–544. doi: 10.1016/s0006-291x(75)80180-x. [DOI] [PubMed] [Google Scholar]
- 36.Chen H, Torchilin V, Langer R. Lectin-bearing polymerized liposomes as potential oral vaccine carriers. Pharm Res. 1996;13:1378–1383. doi: 10.1023/a:1016030202104. [DOI] [PubMed] [Google Scholar]
- 37.Allen TM. Liposomes. Opportunities in drug delivery. Drugs. 1997;54(Suppl 4):8–14. doi: 10.2165/00003495-199700544-00004. [DOI] [PubMed] [Google Scholar]
- 38.Allen TM, Cullis PR. Drug delivery systems: entering the mainstream. Science. 2004;303:1818–1822. doi: 10.1126/science.1095833. [DOI] [PubMed] [Google Scholar]
- 39.Johnston DS, McLean LR, Whittam MA, Clark AD, Chapman D. Spectra and physical properties of liposomes and monolayers of polymerizable phospholipids containing diacetylene groups in one or both acyl chains. Biochemistry. 1983;22:3194–3202. doi: 10.1021/bi00282a024. [DOI] [PubMed] [Google Scholar]
- 40.Leaver J, Alonso A, Durrani AA, Chapman D. The physical properties and photopolymerization of diacetylene-containing phospholipid liposomes. Biochimica et Biophysica Acta (BBA) - Biomembranes. 1983;732:210–218. [Google Scholar]
- 41.Shin WJ, Shin JH, Song JY, Han SY. Effects of ZnO nanowire length on surface-assisted laser desorption/ionization of small molecules. J Am Soc Mass Spectrom. 2010;21:989–992. doi: 10.1016/j.jasms.2010.01.030. [DOI] [PubMed] [Google Scholar]
- 42.Schiller J, Süß R, Petkovic M, Hilbert N, et al. CsCl as an auxiliary reagent for the analysis of phosphatidylcholine mixtures by matrix-assisted laser desorption and ionization time-of-flight mass spectrometry (MALDI-TOF MS) Chemistry and Physics of Lipids. 2001;113:123–131. doi: 10.1016/s0009-3084(01)00188-8. [DOI] [PubMed] [Google Scholar]
- 43.Duncan MW, Roder H, Hunsucker SW. Quantitative matrix-assisted laser desorption/ionization mass spectrometry. Brief Funct Genomic Proteomic. 2008;7:355–370. doi: 10.1093/bfgp/eln041. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Nicola AJ, Gusev AI, Proctor A, Jackson EK, Hercules DM. Application of the fast-evaporation sample preparation method for improving quantification of angiotensin II by matrix-assisted laser desorption/ionization. Rapid Commun Mass Spectrom. 1995;9:1164–1171. doi: 10.1002/rcm.1290091216. [DOI] [PubMed] [Google Scholar]
- 45.Moskovets E, Chen HS, Pashkova A, Rejtar T, et al. Closely spaced external standard: a universal method of achieving 5 ppm mass accuracy over the entire MALDI plate in axial matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Rapid Communications in Mass Spectrometry. 2003;17:2177–2187. doi: 10.1002/rcm.1158. [DOI] [PubMed] [Google Scholar]
- 46.Hatsis P, Brombacher S, Corr J, Kovarik P, Volmer DA. Quantitative analysis of small pharmaceutical drugs using a high repetition rate laser matrix-assisted laser/desorption ionization source. Rapid Commun Mass Spectrom. 2003;17:2303–2309. doi: 10.1002/rcm.1192. [DOI] [PubMed] [Google Scholar]
- 47.Sleno L, Volmer DA. Assessing the properties of internal standards for quantitative matrix-assisted laser desorption/ionization mass spectrometry of small molecules. Rapid Communications in Mass Spectrometry. 2006;20:1517–1524. doi: 10.1002/rcm.2498. [DOI] [PubMed] [Google Scholar]
- 48.Nimptsch A, Schibur S, Schnabelrauch M, Fuchs B, et al. Characterization of the quantitative relationship between signal-to-noise (S/N) ratio and sample amount on-target by MALDI-TOF MS: Determination of chondroitin sulfate subsequent to enzymatic digestion. Anal Chim Acta. 2009;635:175–182. doi: 10.1016/j.aca.2009.01.025. [DOI] [PubMed] [Google Scholar]
- 49.Corr JJ, Kovarik P, Schneider BB, Hendrikse J, et al. Design considerations for high speed quantitative mass spectrometry with MALDI ionization. J Am Soc Mass Spectrom. 2006;17:1129–1141. doi: 10.1016/j.jasms.2006.04.026. [DOI] [PubMed] [Google Scholar]




