The Texas Heart Institute Journal logoLink to The Texas Heart Institute Journal
. 1987 Mar;14(1):72–76.

Assessment of Björk-Shiley and Porcine Mitral Valve Dysfunction by M-Mode Echocardiographic Left Ventricular Dimension

Manuel G Eguaras 1, Isabel M Luque 1, Manuel Pan 1, Anastasio Montero 1, Ignacio Moriones 1, Jorge Granados 1, Miguel A Garcia 1, Manuel Concha 1
PMCID: PMC324696  PMID: 15227333

Abstract

In this study of 46 patients with mitral valve prostheses, we report on the accuracy of the left ventricular maximum filling velocity (LVMFV) index as a diagnostic tool for valve dysfunction. Echocardiographic left ventricular internal dimensions were measured every 0.05 seconds, and every measure was transferred to an axis system. The tangent to this curve was traced at the point of maximum rate of increase of dimension, and the slope in mm/sec represents the LVMFV index.

Six patients, each with an obstructed mitral valve prosthesis, had a reduced LVMFV index (p < 0.001). On the other hand, eight patients with a paravalvular leak showed a LVMFV index significantly higher (p < 0.05) than patients with a normally functioning mitral valve prosthesis.

The results of this study suggest that analysis of the LVMFV index can be a useful and reliable substitute for other noninvasive diagnostic techniques in evaluating certain patients with suspected prosthetic mitral valve malfunction. (Texas Heart Institute Journal 1987; 14:72-76).

Keywords: Mitral valve

Keywords: heart valve prosthesis

Keywords: prosthesis failure

Keywords: echocardiography

Full text

PDF
72

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Alam M., Madrazo A. C., Magilligan D. J., Goldstein S. M mode and two dimensional echocardiographic features of porcine valve dysfunction. Am J Cardiol. 1979 Mar;43(3):502–509. doi: 10.1016/0002-9149(79)90006-7. [DOI] [PubMed] [Google Scholar]
  2. Bernal-Ramirez J. A., Phillips J. H. Echocardiographic study of malfunction of the Bjork-Shiley prosthetic heart valve in the mitral position. Am J Cardiol. 1977 Sep;40(3):449–453. doi: 10.1016/0002-9149(77)90170-9. [DOI] [PubMed] [Google Scholar]
  3. Borow K. M. The need for an integrated noninvasive approach to valvular heart disease. Am Heart J. 1983 Nov;106(5 Pt 1):1177–1180. doi: 10.1016/0002-8703(83)90679-8. [DOI] [PubMed] [Google Scholar]
  4. Copans H., Lakier J. B., Kinsley R. H., Colsen P. R., Fritz V. U., Barlow J. B. Thrombosed Björk-Shiley mitral prostheses. Circulation. 1980 Jan;61(1):169–174. doi: 10.1161/01.cir.61.1.169. [DOI] [PubMed] [Google Scholar]
  5. Cunha C. L., Giuliani E. R., Callahan J. A., Pluth J. R. Echophonocardiographic findings in patients with prosthetic heart valve malfunction. Mayo Clin Proc. 1980 Apr;55(4):231–242. [PubMed] [Google Scholar]
  6. Eguaras M. G., Pasalodos J., Tejero I., Montero A., Franco M., García Jiménez M. A., Moriones I., Granados J., Concha M., Valles F. Velocidad de cambio en las dimensiones ecocardiográficas del ventrículo izquierdo: su valor diagnóstico en la estenosis o insuficiencia mitral severas. Rev Esp Cardiol. 1985 Jan-Feb;38(1):34–37. [PubMed] [Google Scholar]
  7. Gibson D. G., Brown D. Measurement of instantaneous left ventricular dimension and filling rate in man, using echocardiography. Br Heart J. 1973 Nov;35(11):1141–1149. doi: 10.1136/hrt.35.11.1141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Gibson D. G., Traill T. A., Hall R. J., Brown D. J. Echocardiographic features of secondary left ventricular hypertrophy. Br Heart J. 1979 Jan;41(1):54–59. doi: 10.1136/hrt.41.1.54. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Horowitz M. S., Goodman D. J., Hancock E. W., Popp R. L. Noninvasive diagnosis of complications of the mitral bioprosthesis. J Thorac Cardiovasc Surg. 1976 Mar;71(3):450–457. [PubMed] [Google Scholar]
  10. Jacovella G., Giampaolo P., Gambelli G. Echocardiographic patterns of prosthetic valve malfunction. Am J Cardiol. 1978 Nov;42(5):876–876. doi: 10.1016/0002-9149(78)90118-2. [DOI] [PubMed] [Google Scholar]
  11. Martin R. P., Rakowski H., Kleiman J. H., Beaver W., London E., Popp R. L. Reliability and reproducibility of two dimensional echocardiograph measurement of the stenotic mitral valve orifice area. Am J Cardiol. 1979 Mar;43(3):560–568. doi: 10.1016/0002-9149(79)90014-6. [DOI] [PubMed] [Google Scholar]
  12. Mintz G. S., Carlson E. B., Kotler M. N. Comparison of noninvasive techniques in evaluation of the nontissue cardiac valve prosthesis. Am J Cardiol. 1982 Jan;49(1):39–44. doi: 10.1016/0002-9149(82)90275-2. [DOI] [PubMed] [Google Scholar]
  13. Moreno-Cabral R. J., McNamara J. J., Mamiya R. T., Brainard S. C., Chung G. K. Acute thrombotic obstruction with Björk-Shiley valves: diagnostic and surgical considerations. J Thorac Cardiovasc Surg. 1978 Mar;75(3):321–330. [PubMed] [Google Scholar]
  14. Pluth J. R., McGoon D. C. Current status of heart valve replacement. Mod Concepts Cardiovasc Dis. 1974 Jan;43(1):65–70. [PubMed] [Google Scholar]
  15. Salati A., Narducci C., Pino P. G., Chidichimo G., Jacovella G. Patologia delle protesi mitraliche porcine. Rilievi ecocardiografici in cinque casi con rigurgito intra-para-protesico. G Ital Cardiol. 1980;10(5):605–609. [PubMed] [Google Scholar]
  16. Slavich G., Pinamonti B., De Piccoli B. Echocardiographic diagnosis of porcine mitral prosthesis dehiscence. A new finding. Am Heart J. 1984 Jan;107(1):172–174. doi: 10.1016/0002-8703(84)90155-8. [DOI] [PubMed] [Google Scholar]
  17. Upton M. T., Gibson D. G. The study of left ventricular function from digitized echocardiograms. Prog Cardiovasc Dis. 1978 Mar-Apr;20(5):359–384. doi: 10.1016/0033-0620(78)90003-8. [DOI] [PubMed] [Google Scholar]

Articles from Texas Heart Institute Journal are provided here courtesy of Texas Heart Institute

RESOURCES