Outlet-Strut Fracture and Disc Embolization in 27-mm and Smaller Björk-Shiley Convexo-Concave Valves

Historical Overview and 3 Case Reports

Carlos G Montero 1, Raul Burgos 1, Jose E Rodriguez 1, Diego Figuera 1
PMCID: PMC324840  PMID: 15227234

Abstract

Outlet-strut fracture leading to primary failure of Björk-Shiley convexo-concave disc valves has only recently been reported in valves of 27-mm size and smaller. We now report an additional 3 cases of outlet strut failure in smaller convexo-concave valves, in a series of 899 implantations (overall incidence, 0.33%) of convexo-concave valves in various sizes and in 2 variant opening angles (60° and 70°), during a 6-year period. All failures were confined to lots implanted during 1981 and 1982. We describe these 3 cases within the context of an historical overview of the Björk-Shiley tilting-disc valve, and propose methods of surgical intervention when this life-threatening complication arises. (Texas Heart Institute Journal 1989; 16:32-35)

Keywords: Björk-Shiley prosthetic valve

Keywords: heart valve replacement

Keywords: embolization

Full text

PDF
32

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Björk V. O., Henze A. Ten years' experience with the Björk-Shiley tilting disc valve. J Thorac Cardiovasc Surg. 1979 Sep;78(3):331–342. [PubMed] [Google Scholar]
  2. Brubakk O., Simonsen S., Källman L., Fredriksen A. Strut fracture in the new Bjørk-Shiley mitral valve prosthesis. Thorac Cardiovasc Surg. 1981 Apr;29(2):108–109. doi: 10.1055/s-2007-1023454. [DOI] [PubMed] [Google Scholar]
  3. Davis P. K., Myers J. L., Pennock J. L., Thiele B. L. Strut fracture and disc embolization in Björk-Shiley mitral valve prostheses: diagnosis and management. Ann Thorac Surg. 1985 Jul;40(1):65–68. doi: 10.1016/s0003-4975(10)61172-1. [DOI] [PubMed] [Google Scholar]
  4. Elliot R. A. Letter: Bjork-Shiley prosthesis recall. Circulation. 1976 Jan;53(1):206–206. doi: 10.1161/01.cir.53.1.206. [DOI] [PubMed] [Google Scholar]
  5. Fennerty M. B., Pick R. A., Clark M. A. Outlet strut fracture of a Björk-Shiley prosthesis in the mitral position. Am J Cardiol. 1984 Oct 1;54(7):923–924. doi: 10.1016/s0002-9149(84)80238-6. [DOI] [PubMed] [Google Scholar]
  6. Hiratzka L. F., Kouchoukos N. T., Grunkemeier G. L., Miller D. C., Scully H. E., Wechsler A. S. Outlet strut fracture of the Björk-Shiley 60 degrees Convexo-Concave valve: current information and recommendations for patient care. J Am Coll Cardiol. 1988 May;11(5):1130–1137. doi: 10.1016/s0735-1097(98)90075-4. [DOI] [PubMed] [Google Scholar]
  7. Ibarra F., Gutierrez A., Martinez F., Carreras L., Lopez C., Alonso-Lej F. Fracture of the outlet strut of a Björk-Shiley mitral prosthesis: emergency operation with survival. J Thorac Cardiovasc Surg. 1984 Feb;87(2):315–318. [PubMed] [Google Scholar]
  8. Lindblom D., Björk V. O., Semb B. K. Mechanical failure of the Björk-Shiley valve. Incidence, clinical presentation, and management. J Thorac Cardiovasc Surg. 1986 Nov;92(5):894–907. [PubMed] [Google Scholar]
  9. Mok C. K., Lee J. W., Kong S. M., Hui K. K. Experience with outlet strut fracture of the Björk-Shiley convexoconcave mitral valve prosthesis. Am Heart J. 1985 Oct;110(4):814–818. doi: 10.1016/0002-8703(85)90462-4. [DOI] [PubMed] [Google Scholar]
  10. Montero C. G., Tellez G., Burgos R., Rufilanchas J. J., Agosti J., Juffe A., Figuera D. Dysfunction of the Björk-Shiley prosthesis. Report of 32 cases. Scand J Thorac Cardiovasc Surg. 1982;16(1):9–15. doi: 10.3109/14017438209100602. [DOI] [PubMed] [Google Scholar]
  11. Nakano S., Kawashima Y., Hirose H., Matsuda H., Shirakura R., Kawamoto T. Strut fracture of a Björk-Shiley mitral valve prosthesis with successful reoperation. J Cardiovasc Surg (Torino) 1986 Nov-Dec;27(6):685–687. [PubMed] [Google Scholar]
  12. Ostermeyer J., Horstkotte D., Bennett J., Huysmans H., Lindblom D., Olin C., Orinius E., Semb G. The Björk-Shiley 70 degree convexo-concave prosthesis strut fracture problem (present state of information). Thorac Cardiovasc Surg. 1987 Apr;35(2):71–77. doi: 10.1055/s-2007-1020200. [DOI] [PubMed] [Google Scholar]
  13. Sethia B., Quin R. O., Bain W. H. Disc embolisation after minor strut fracture in a Björk-Shiley mitral valve prosthesis. Thorax. 1983 May;38(5):390–391. doi: 10.1136/thx.38.5.390. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Texas Heart Institute Journal are provided here courtesy of Texas Heart Institute

RESOURCES