Skip to main content
The Texas Heart Institute Journal logoLink to The Texas Heart Institute Journal
. 1993;20(3):213–217.

Use of polytetrafluoroethylene surgical membrane as a pericardial substitute in the correction of congenital heart defects.

M Loebe 1, V Alexi-Meskhishvili 1, Y Weng 1, G Hausdorf 1, R Hetzer 1
PMCID: PMC325093  PMID: 8219824

Abstract

Patients who undergo repair of congenital heart defects have a high probability of eventually needing a repeat operation. In such cases, use of a pericardial substitute can preclude adhesion formation, thereby helping to avert the complications of a repeat sternotomy. This report describes our use of the GORE-TEX polytetrafluoroethylene surgical membrane in 321 patients who underwent repair of congenital heart disease at our hospital between May of 1989 and December of 1992 (these cases account for 37.1% of all such repairs performed during this period). The patients' ages ranged from 2 days to 31 years (mean, 5.9 years). Eighty-eight patients (27.4%) had at least 1 previous median sternotomy, and 234 (73.0%) had a moderate to high probability of reoperation. Seventy-eight patients (24.3%) received a homograft, which was positioned extracardially in 63 cases. During a maximum follow-up period of 43 months (mean, 26 +/- 5.6 months), 11 patients died, for an overall mortality of 3.4%. In no case was death attributable to membrane-related complications. Postoperative complications were encountered in 12 patients (3.7%), 10 of whom required an early repeat thoracotomy. One patient had mediastinitis, for an incidence of 0.3% (in contrast to 0.2% in 864 patients who underwent similar repair without the use of a surgical membrane). No complications were encountered in the 27 patients in whom the membrane was implanted during delayed sternal closure. Seventeen patients (5.5% of the survivors) underwent a repeat sternotomy, for further correction of congenital heart defects, a mean of 20 months after the original repair. At reoperation, major complications were encountered. Although the number of reoperations was low, we conclude that use of the polytetrafluoroethylene surgical membrane for pericardial closure in children is a safe procedure that helps prevent complications at reoperation.

Full text

PDF
213

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Amato J. J., Cotroneo J. V., Galdieri R. J., Alboliras E., Antillon J., Vogel R. L. Experience with the polytetrafluoroethylene surgical membrane for pericardial closure in operations for congenital cardiac defects. J Thorac Cardiovasc Surg. 1989 Jun;97(6):929–934. [PubMed] [Google Scholar]
  2. Bailey L. L., Ze-jian L., Schulz E., Roost H., Yahiku P. A cause of right ventricular dysfunction after cardiac operations. J Thorac Cardiovasc Surg. 1984 Apr;87(4):539–542. [PubMed] [Google Scholar]
  3. Bryan A. J., Angelini G. D. Pericardial effusion after open heart surgery. Thorax. 1990 Sep;45(9):655–656. doi: 10.1136/thx.45.9.655. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Castaneda A. R., Mayer J. E., Jr, Jonas R. A., Lock J. E., Wessel D. L., Hickey P. R. The neonate with critical congenital heart disease: repair--a surgical challenge. J Thorac Cardiovasc Surg. 1989 Nov;98(5 Pt 2):869–875. [PubMed] [Google Scholar]
  5. Cimino J. J., Kogan A. D. Constrictive pericarditis after cardiac surgery: report of three cases and review of the literature. Am Heart J. 1989 Dec;118(6):1292–1301. doi: 10.1016/0002-8703(89)90021-5. [DOI] [PubMed] [Google Scholar]
  6. Cunningham J. N., Jr, Spencer F. C., Zeff R., Williams C. D., Cukingnan R., Mullin M. Influence of primary closure of the pericardium after open-heart surgery on the frequency of tamponade, postcardiotomy syndrome, and pulmonary complications. J Thorac Cardiovasc Surg. 1975 Jul;70(1):119–125. [PubMed] [Google Scholar]
  7. D'Cruz I. A., Overton D. H., Pai G. M. Pericardial complications of cardiac surgery: emphasis on the diagnostic role of echocardiography. J Card Surg. 1992 Sep;7(3):257–268. doi: 10.1111/j.1540-8191.1992.tb00811.x. [DOI] [PubMed] [Google Scholar]
  8. Dobell A. R., Jain A. K. Catastrophic hemorrhage during redo sternotomy. Ann Thorac Surg. 1984 Apr;37(4):273–278. doi: 10.1016/s0003-4975(10)60728-x. [DOI] [PubMed] [Google Scholar]
  9. Harada Y., Imai Y., Kurosawa H., Hoshino S., Nakano K. Long-term results of the clinical use of an expanded polytetrafluoroethylene surgical membrane as a pericardial substitute. J Thorac Cardiovasc Surg. 1988 Nov;96(5):811–815. [PubMed] [Google Scholar]
  10. Heydorn W. H., Daniel J. S., Wade C. E. A new look at pericardial substitutes. J Thorac Cardiovasc Surg. 1987 Aug;94(2):291–296. [PubMed] [Google Scholar]
  11. Heydorn W. H., Ferraris V. A., Berry W. R. Pericardial substitutes: a survey. Ann Thorac Surg. 1988 Nov;46(5):567–569. doi: 10.1016/s0003-4975(10)64699-1. [DOI] [PubMed] [Google Scholar]
  12. Laks H., Hammond G., Geha A. S. Use of silicone rubber as a pericardial substitute to facilitate reoperation in cardiac surgery. J Thorac Cardiovasc Surg. 1981 Jul;82(1):88–92. [PubMed] [Google Scholar]
  13. Londe S., Sugg W. L. The challenge of reoperation in cardiac surgery. Ann Thorac Surg. 1974 Feb;17(2):157–162. doi: 10.1016/s0003-4975(10)65627-5. [DOI] [PubMed] [Google Scholar]
  14. Loop F. D. Catastrophic hemorrhage during sternal reentry. Ann Thorac Surg. 1984 Apr;37(4):271–272. doi: 10.1016/s0003-4975(10)60727-8. [DOI] [PubMed] [Google Scholar]
  15. Malm T., Bowald S., Bylock A., Busch C. Prevention of postoperative pericardial adhesions by closure of the pericardium with absorbable polymer patches. An experimental study. J Thorac Cardiovasc Surg. 1992 Sep;104(3):600–607. [PubMed] [Google Scholar]
  16. Merin G., McGoon D. C. Reoperation after insertion of aortic homograft as a right ventricular outflow tract. Ann Thorac Surg. 1973 Aug;16(2):122–126. doi: 10.1016/s0003-4975(10)65826-2. [DOI] [PubMed] [Google Scholar]
  17. Minale C., Hollweg G., Nikol S., Mittermayer C., Messmer B. J. Closure of the pericardium using expanded polytetrafluoroethylene GORE-TEX-Surgical Membrane: clinical experience. Thorac Cardiovasc Surg. 1987 Oct;35(5):312–315. doi: 10.1055/s-2007-1020253. [DOI] [PubMed] [Google Scholar]
  18. Minale C., Nikol S., Hollweg G., Mittermayer C., Messmer B. J. Clinical experience with expanded polytetrafluoroethylene Gore-Tex surgical membrane for pericardial closure: a study of 110 cases. J Card Surg. 1988 Sep;3(3):193–201. doi: 10.1111/j.1540-8191.1988.tb00239.x. [DOI] [PubMed] [Google Scholar]
  19. Revuelta J. M., Garcia-Rinaldi R., Johnston R. H., Jr, Vaughan G. D., 3rd Implantation of pericardial substitutes. Ann Thorac Surg. 1985 Feb;39(2):190–191. doi: 10.1016/s0003-4975(10)62566-0. [DOI] [PubMed] [Google Scholar]
  20. Revuelta J. M., Garcia-Rinaldi R., Val F., Crego R., Duran C. M. Expanded polytetrafluoroethylene surgical membrane for pericardial closure. An experimental study. J Thorac Cardiovasc Surg. 1985 Mar;89(3):451–455. [PubMed] [Google Scholar]
  21. von Segesser L., Jornod N., Faidutti B. Repeat sternotomy after reconstruction of the pericardial sac with glutaraldehyde-preserved equine pericardium. J Thorac Cardiovasc Surg. 1987 Apr;93(4):616–619. [PubMed] [Google Scholar]

Articles from Texas Heart Institute Journal are provided here courtesy of Texas Heart Institute