Abstract
We review our experience with the transvenous Medtronic Pacer Cardioverter Defibrillator System (Model 7217B), a multifunction implantable pacer defibrillator combined with a transvenous lead system (Transvene). From April 1991 to October 1992, we implanted this device in 19 consecutive patients (11 men and 8 women; average age, 56.5 years). Nine patients (47.4%), 5 with coronary artery disease and 4 with dilated cardiomyopathy, had an ejection fraction of < 30%. The average operative time was 129 minutes. In 18 patients (94.7%), the transvenous lead system provided effective sensing, pacing, and defibrillation during intraoperative testing. In each of these cases, the defibrillation threshold was less than 18 J. In 1 patient (5.3%), it was necessary to switch to epicardial leads, which were implanted through a left thoracotomy. All patients were extubated in the recovery room. The average hospital stay was 8 days. There was no early mortality or morbidity. During a maximum follow-up period of 17 months (mean, 9.2 months), no sudden death occurred. The implantable system terminated 245 ventricular tachycardia episodes in 14 patients (73.7%) and 82 ventricular fibrillation episodes in 13 patients (68.4%). Two hundred eleven (86.1%) of the ventricular tachycardia episodes were resolved by antitachycardia pacing alone. In 2 patients (10.5%), the caval electrode became dislocated; repositioning of the electrode was followed by repeat defibrillation threshold evaluation. Our experience shows that the transvenous Medtronic Pacer Cardioverter Defibrillator System provides safe, effective treatment of ventricular tachyarrhythmias. Because the perioperative mortality and morbidity are extremely low, use of this device may be particularly beneficial in patients with a high operative risk. Moreover, the lower number of unpleasant therapeutic shocks should increase patient a acceptance of the device.
Full text
PDF






Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Block M., Hammel D., Borggrefe M., Isbruch F., Hachenberg T., Scheld H. H., Breithardt G. Erste klinische Erfahrungen mit einem transvenös-subkutanen Defibrillationssystem. Z Kardiol. 1991 Nov;80(11):657–664. [PubMed] [Google Scholar]
- Block M., Hammel D., Isbruch F., Borggrefe M., Wietholt D., Hachenberg T., Scheld H. H., Breithardt G. Results and realistic expectations with transvenous lead systems. Pacing Clin Electrophysiol. 1992 Apr;15(4 Pt 3):665–670. doi: 10.1111/j.1540-8159.1992.tb05160.x. [DOI] [PubMed] [Google Scholar]
- Damiano R. J., Jr Implantable cardioverter defibrillators: current status and future directions. J Card Surg. 1992 Mar;7(1):36–57. doi: 10.1111/j.1540-8191.1992.tb00774.x. [DOI] [PubMed] [Google Scholar]
- Epstein A. E. Invited letter concerning: Automatic implantable cardioverter-defibrillator. J Thorac Cardiovasc Surg. 1991 Jul;102(1):160–163. [PubMed] [Google Scholar]
- Epstein A. E., Kay G. N., Plumb V. J., Shepard R. B., Kirklin J. K. Combined automatic implantable cardioverter-defibrillator and pacemaker systems: implantation techniques and follow-up. J Am Coll Cardiol. 1989 Jan;13(1):121–131. doi: 10.1016/0735-1097(89)90559-7. [DOI] [PubMed] [Google Scholar]
- Frame R., Brodman R., Gross J., Hollinger I., Fisher J. D., Kim S. G., Ferrick K., Roth J., Ben-Zur U., Furman S. Initial experience with transvenous implantable cardioverter defibrillator lead systems: operative morbidity and mortality. Pacing Clin Electrophysiol. 1993 Jan;16(1 Pt 2):149–152. doi: 10.1111/j.1540-8159.1993.tb01552.x. [DOI] [PubMed] [Google Scholar]
- Frank G., Lowes D. Implantable cardioverter defibrillators: surgical considerations. Pacing Clin Electrophysiol. 1992 Apr;15(4 Pt 3):631–635. doi: 10.1111/j.1540-8159.1992.tb05153.x. [DOI] [PubMed] [Google Scholar]
- Fromer M., Brachmann J., Block M., Siebels J., Hoffmann E., Almendral J., Ohm O. J., den Dulk K., Coumel P., Camm A. J. Efficacy of automatic multimodal device therapy for ventricular tachyarrhythmias as delivered by a new implantable pacing cardioverter-defibrillator. Results of a European multicenter study of 102 implants. Circulation. 1992 Aug;86(2):363–374. doi: 10.1161/01.cir.86.2.363. [DOI] [PubMed] [Google Scholar]
- Hammel D., Block M., Konertz W., Borggrefe M., Isbruch F., Asfour B., Geiger A., Breithardt G., Scheld H. H. Surgical experience with defibrillator implantation using nonthoracotomy leads. Ann Thorac Surg. 1993 Mar;55(3):685–693. doi: 10.1016/0003-4975(93)90276-n. [DOI] [PubMed] [Google Scholar]
- Hargrove W. C., 3rd, Josephson M. E., Marchlinski F. E., Miller J. M. Surgical decisions in the management of sudden cardiac death and malignant ventricular arrhythmias. Subendocardial resection, the automatic internal defibrillator, or both. J Thorac Cardiovasc Surg. 1989 Jun;97(6):923–928. [PubMed] [Google Scholar]
- Hauser R. G., Mower M. M., Mitchell M., Nisam S. Current status of the Ventak PRx pulse generator and Endotak nonthoracotomy lead system. Pacing Clin Electrophysiol. 1992 Apr;15(4 Pt 3):671–677. doi: 10.1111/j.1540-8159.1992.tb05161.x. [DOI] [PubMed] [Google Scholar]
- Horowitz L. N. The automatic implantable cardioverter defibrillator: review of clinical results, 1980-1990. Pacing Clin Electrophysiol. 1992 Apr;15(4 Pt 3):604–609. doi: 10.1111/j.1540-8159.1992.tb05149.x. [DOI] [PubMed] [Google Scholar]
- Kelly P. A., Cannom D. S., Garan H., Mirabal G. S., Harthorne J. W., Hurvitz R. J., Vlahakes G. J., Jacobs M. L., Ilvento J. P., Buckley M. J. The automatic implantable cardioverter-defibrillator: efficacy, complications and survival in patients with malignant ventricular arrhythmias. J Am Coll Cardiol. 1988 Jun;11(6):1278–1286. doi: 10.1016/0735-1097(88)90292-6. [DOI] [PubMed] [Google Scholar]
- Klein H., Trappe H. J. Implantable cardioverter defibrillator therapy: indications and decision making in patients with coronary artery disease. Pacing Clin Electrophysiol. 1992 Apr;15(4 Pt 3):610–615. doi: 10.1111/j.1540-8159.1992.tb05150.x. [DOI] [PubMed] [Google Scholar]
- Leitch J. W., Gillis A. M., Wyse D. G., Yee R., Klein G. J., Guiraudon G., Sheldon R. S., Duff H. J., Kieser T. M., Mitchell L. B. Reduction in defibrillator shocks with an implantable device combining antitachycardia pacing and shock therapy. J Am Coll Cardiol. 1991 Jul;18(1):145–151. doi: 10.1016/s0735-1097(10)80232-3. [DOI] [PubMed] [Google Scholar]
- Levine J. H., Mellits E. D., Baumgardner R. A., Veltri E. P., Mower M., Grunwald L., Guarnieri T., Aarons D., Griffith L. S. Predictors of first discharge and subsequent survival in patients with automatic implantable cardioverter-defibrillators. Circulation. 1991 Aug;84(2):558–566. doi: 10.1161/01.cir.84.2.558. [DOI] [PubMed] [Google Scholar]
- McCowan R., Maloney J., Wilkoff B., Simmons T., Khoury D., McAlister H., Morant V., Castle L. Automatic implantable cardioverter-defibrillator implantation without thoracotomy using an endocardial and submuscular patch system. J Am Coll Cardiol. 1991 Feb;17(2):415–421. doi: 10.1016/s0735-1097(10)80108-1. [DOI] [PubMed] [Google Scholar]
- Meesmann M. Factors associated with implantation-related complications. Pacing Clin Electrophysiol. 1992 Apr;15(4 Pt 3):649–653. doi: 10.1111/j.1540-8159.1992.tb05157.x. [DOI] [PubMed] [Google Scholar]
- Pinski S. L., Mick M. J., Arnold A. Z., Golding L., McCarthy P. M., Castle L. W., Maloney J. D., Trohman R. G. Retrospective analysis of patients undergoing one- or two-stage strategies for myocardial revascularization and implantable cardioverter defibrillator implantation. Pacing Clin Electrophysiol. 1991 Jul;14(7):1138–1147. doi: 10.1111/j.1540-8159.1991.tb02845.x. [DOI] [PubMed] [Google Scholar]
- Saksena S. Defibrillation thresholds and perioperative mortality associated with endocardial and epicardial defibrillation lead systems. The PCD investigators and participating institutions. Pacing Clin Electrophysiol. 1993 Jan;16(1 Pt 2):202–207. doi: 10.1111/j.1540-8159.1993.tb01562.x. [DOI] [PubMed] [Google Scholar]
- Schmitt C., Brachmann J., Saggau W., Beyer T., Waldecker B., Scharowski K., Hilbel T., Montero M., Offner B., Schöls W. Kombinierte antibradykarde/antitachykarde Schrittmacher-Kardioverter-Defibrillator-Systeme bei Patienten mit rezidivierenden ventrikulären Tachyarrhythmien. Z Kardiol. 1991 Nov;80(11):665–672. [PubMed] [Google Scholar]
- Shepard R. B., Goldin M. D., Lawrie G. M., Watkins L., Jr, Winkle R. A., Mower M. M., Thomas A. C., Nisam S. Automatic implantable cardioverter defibrillator: surgical approaches for implantation. J Card Surg. 1992 Sep;7(3):208–224. doi: 10.1111/j.1540-8191.1992.tb00804.x. [DOI] [PubMed] [Google Scholar]
- Siclari F., Klein H. Automatic implantable cardioverter defibrillator implantation after previous open heart surgery: subcostal incision and small left anterior thoracotomy. Pacing Clin Electrophysiol. 1990 Jun;13(6):715–718. doi: 10.1111/j.1540-8159.1990.tb02096.x. [DOI] [PubMed] [Google Scholar]
- Singer I., Hutchins G. M., Mirowski M., Mower M. M., Veltri E. P., Guarnieri T., Griffith L. S., Watkins L., Juanteguy J., Fisher S. Pathologic findings related to the lead system and repeated defibrillations in patients with the automatic implantable cardioverter-defibrillator. J Am Coll Cardiol. 1987 Aug;10(2):382–388. doi: 10.1016/s0735-1097(87)80022-0. [DOI] [PubMed] [Google Scholar]
- Spotnitz H. M., Ott G. Y., Bigger J. T., Jr, Steinberg J. S., Livelli F., Jr Methods of implantable cardioverter-defibrillator-pacemaker insertion to avoid interactions. Ann Thorac Surg. 1992 Feb;53(2):253–257. doi: 10.1016/0003-4975(92)91327-6. [DOI] [PubMed] [Google Scholar]
- Swerdlow C. D., Winkle R. A., Mason J. W. Determinants of survival in patients with ventricular tachyarrhythmias. N Engl J Med. 1983 Jun 16;308(24):1436–1442. doi: 10.1056/NEJM198306163082402. [DOI] [PubMed] [Google Scholar]
- Thomas A. C., Moser S. A., Smutka M. L., Wilson P. A. Implantable defibrillation: eight years clinical experience. Pacing Clin Electrophysiol. 1988 Nov;11(11 Pt 2):2053–2058. doi: 10.1111/j.1540-8159.1988.tb06349.x. [DOI] [PubMed] [Google Scholar]
- Troup P. J. Programmed stimulation-guided therapy compared with implantable cardioverter defibrillator device therapy in the treatment of ventricular tachyarrhythmias. Pacing Clin Electrophysiol. 1991 Feb;14(2 Pt 2):267–272. doi: 10.1111/j.1540-8159.1991.tb05105.x. [DOI] [PubMed] [Google Scholar]
- Waller T. J., Kay H. R., Spielman S. R., Kutalek S. P., Greenspan A. M., Horowitz L. N. Reduction in sudden death and total mortality by antiarrhythmic therapy evaluated by electrophysiologic drug testing: criteria of efficacy in patients with sustained ventricular tachyarrhythmia. J Am Coll Cardiol. 1987 Jul;10(1):83–89. doi: 10.1016/s0735-1097(87)80164-x. [DOI] [PubMed] [Google Scholar]
- Winkle R. A., Mead R. H., Ruder M. A., Gaudiani V. A., Smith N. A., Buch W. S., Schmidt P., Shipman T. Long-term outcome with the automatic implantable cardioverter-defibrillator. J Am Coll Cardiol. 1989 May;13(6):1353–1361. doi: 10.1016/0735-1097(89)90310-0. [DOI] [PubMed] [Google Scholar]