Abstract
Cerebral aneurysms are abnormal focal dilatations of artery walls. The interest in virtual tools to help clinicians to value the effectiveness of different procedures for cerebral aneurysm treatment is constantly growing. This study is focused on the analysis of the influence of different stent deployment approaches on intra-aneurysmal haemodynamics using computational fluid dynamics (CFD). A self-expanding stent was deployed in an idealized aneurysmatic cerebral vessel in two initial positions. Different cases characterized by a progression of simplifications on stent modelling (geometry and material) and vessel material properties were set up, using finite element and fast virtual stenting methods. Then, CFD analysis was performed for untreated and stented vessels. Haemodynamic parameters were analysed qualitatively and quantitatively, comparing the cases and the two initial positions. All the cases predicted a reduction of average wall shear stress and average velocity of almost 50 per cent after stent deployment for both initial positions. Results highlighted that, although some differences in calculated parameters existed across the cases based on the modelling simplifications, all the approaches described the most important effects on intra-aneurysmal haemodynamics. Hence, simpler and faster modelling approaches could be included in clinical workflow and, despite the adopted simplifications, support clinicians in the treatment planning.
Keywords: computational fluid dynamics, virtual stenting, self-expanding stent, cerebral aneurysm
1. Introduction
Cerebral aneurysms are abnormal focal dilatations of the artery walls, typically present near or at the bifurcations of the circle of Willis. It is estimated that worldwide 2–5% of the population harbours one or more cerebral aneurysms [1]. Statistics indicate that around 2 per cent of cerebral aneurysms will rupture, resulting in subarachnoid haemorrhage, with 50 per cent of these leading to death [2]. Aneurysms are usually asymptomatic, but the greater improvement and availability of non-invasive neuroradiological imaging techniques make their incidental discovery more and more frequent. Once an aneurysm is detected, preventive and elective intervention is usually followed to avoid its rupture [3]. Among treatment options, such as clipping, coiling or stent-supported coiling, clinicians select the most appropriate therapy for the patient to exclude the aneurysm from the blood circulation.
Several computational tools are being developed to help clinicians in the decision-making and treatment planning process [4–8]. These tools should produce accurate and clinically meaningful results, with a high level of usability and automation at low computational costs. Some of these tools allow endovascular devices to be deployed virtually inside patient-specific vascular geometries [9–12]. They can allow computational fluid dynamics (CFD) to be performed simulations before and after treatment to analyse the haemodynamic alterations induced by the implantation of the device.
Normally, the outcome of endovascular treatment (including stenting) is assessed by the use of angiographic images [13]. This imaging method provides a visualization of the flow field with which the clinicians are familiar. Also, the main effects of the device on intra-aneurismal flow are visible. In this way, the clinician can assess at an early stage if the treatment has a positive effect or not, thus requiring an additional treatment, specific follow-up or if the patient can be discharged. Recently, the simulation of an angiography trough the virtual injection of a contrast agent (dye) has been included in CFD models to give clinicians and researchers a closer and additional instrument for treatment evaluation [11,14,15].
Regarding the treatment of cerebral aneurysms with stents, these are frequently used to provide support for the coil in wide-necked aneurysms. In fact, it was proposed that stents could prevent aneurysm re-canalization after coiling by reducing coil migration and residual ostium [16–19]. Moreover, it has been found that the position of struts and links above the ostium (configuration of the released stent) has a considerable impact on the resulting blood flow [19–25].
The mechanics of the stent release can be mathematically described by mean of partial differential equations that can be solved using finite-element (FE) methods. These methods are used extensively in simulating the deployment of stents in stenotic arteries [26–31], where the forces exchanged between the stent and the vessel and the stress field generated by this interaction play an important role. Nevertheless, these methods are not user-friendly for clinicians and are too much computationally demanding for an operating room. Furthermore, their application in clinical practice on patient-specific anatomies poses major challenges. On the other hand, Larrabide et al. [11] recently proposed a method, called fast virtual stenting (FVS), capable of simulating the expansion of stents into patient-specific aneurysmatic vessels. This method is based on deformable meshes [32] and simple geometrical constraints, which neglect material properties. It is semi-automatic, easy to implement and drastically reduces the computational cost of virtual stenting.
Bernardini et al. [33] investigated the effect of simplifications in computational approaches for simulating the stenting on computational time and accuracy of results, using both FE and FVS methods. An open-cell self-expanding stent was released in an idealized geometry of a cerebral vessel with an aneurysm neck by progressively simplified stent deployment approaches. The time for which a central processing unit (CPU) was used for processing the simulations (CPU time) decreased from tens of thousands to hundreds of seconds by simplifying the stent morphology using only FE methods, and down to few seconds using the FVS method. They also identified that there is a loss of accuracy in the released stent configuration by neglecting material properties. Nevertheless, the relevance of the differences found throughout the approaches was compared with different equally probable initial positions of the stent but not evaluated from the haemodynamic point of view.
Alternatively to this previous work, in the current study, FE and FVS methods for stent deployment were used to obtain the deployed stent representation and the main focus was on CFD simulations performed for each approach. The purpose of this work is to understand the influence of the stent modelling approach on intra-aneurysmal haemodynamics.
2. Material and methods
Models of a self-expanding stent and of an idealized aneurysmatic cerebral vessel were considered. First, the stent was virtually placed using five progressively simplified deploying approaches, four using the FE method and one using the FVS method. Then, CFD simulations were performed in stented vessels and in the untreated vessels (UVs).
2.1. Geometrical models
The geometrical models were generated using the commercial computer-aided design (CAD) software Rhinoceros v. 4.0 (McNeel & Associates, Indianapolis, IN, USA). A geometrical model of a self-expanding Neuroform stent (Boston Scientific, Natick, MA, USA) obtained from a micro-CT scan was created. This device is a high-porosity non-symmetric open-cell stent, and it has been selected because these features better highlight the differences among different released positions when compared with symmetric closed-cell stents. The stent is composed of four rings of 14 struts each, connected by 12 links, three between pairs of rings. Its total length is 10 mm and the nominal diameter is 4.5 mm. Struts and links have a length of 2.43 and 0.26 mm, respectively, with a rectangular cross section of 0.1 × 0.08 mm2, as presented in figure 1a. This model was represented in two ways: keeping the morphology (cross section and shape of strut endings) of the three-dimensional components (three-dimensional-strut stent model, figure 1a), and giving a uni-dimensional description of the stent struts (one-dimensional-strut stent model, figure 1c). One-dimensional-strut stent considers only the centrelines of struts and links, and approximates them as straight lines (figure 1b).
Figure 1.
(a) Three-dimensional-strut stent (rectangular cross section: a = 0.1 mm and b = 0.08 mm for the struts; a = 0.34 mm and b = 0.08 mm for the links). (b) Details of the approximation of one-dimensional-strut stent. (c) One-dimensional-strut stent (circular cross section: d = 0.1 mm for struts and links). (d) Geometrical model of the idealized aneurysmatic cerebral vessel (Dv = 3.6 mm; R = 30 mm; H = 8.3 mm; Dn = 4.5 mm). (e) Relative positions of the stent with respect to the aneurysm ostium: asymmetric (posA) and symmetric (posB) link.
The vascular geometrical model corresponds to an idealized aneurysmatic vessel previously used by Bernardini et al. [33]. The vessel geometry was completed with an extrusion of its endings, keeping the vessel diameter of 3.6 mm and its curvature radius of 30 mm. A side-wall aneurysm was also added. The aneurysm had an aspect ratio (AR), dome-depth to neck ratio, of 1.8 (figure 1d); treatment is highly recommended for this type of aneurysm [34].
2.2. Stent deployment
Five modelling approaches were adopted to deploy the stent in the vessel. A progression of simplifications on the morphology of the stent and on material properties was considered. Table 1 summarizes the main features of the five cases. Two relative positions of the stent struts with respect to the ostium were considered: asymmetrical (posA) and symmetrical (posB) (figure 1e), both of them obtained by translating and rotating the stent at the un-deployed position. These two positions were chosen since they showed the greatest difference in the released stent configuration [33].
Table 1.
Main features of the analysed cases.
| case abbreviation | FE3H | FE3R | FE1H | FE1R | FVS |
|---|---|---|---|---|---|
| method | FEM | FEM | FEM | FEM | FVS |
| strut and link description | 3D | 3D | 1D | 1D | 1D |
| stent material | Nitinol | Nitinol | Nitinol | Nitinol | — |
| vessel material | deformable, hyperelastic | rigid body | deformable, hyperelastic | rigid body | rigid body |
Four cases were deployed using the FE method. Three-dimensional- and one-dimensional-strut stents were deployed in deformable (FE3H, FE1H) and rigid (FE3R, FE1R) vascular walls. The self-expanding Neuroform stent is made of Nitinol, a nickel–titanium shape memory alloy. The pseudo-elastic material property of the Nitinol was described using average values taken from literature [29,31] by introducing a previously developed user subroutine [35] (figure 2a). The vessel was modelled as either a deformable or rigid body. The deformable wall was described as a single homogeneous isotropic hyperelastic layer of 0.3 mm of thickness. The material parameters were derived from experimental data on cerebral vessels [36] (figure 2b). The simplification of characterizing the vessel wall by its whole stiffness is reasonable as it was clinically observed that stent insertion does not modify the shape and dimension of the vessel [16], suggesting a modest influence of the stent on the vascular wall deformation.
Figure 2.
(a) Pseudo-elastic model for Nitinol. (b) Hyperelastic constitutive law for the vessel, based on a second-order reduced polynomial strain energy density function (parameters: ν = 0.49; D1 = D2 = 0, C10 = 3.3096, C20 = 0.6353, C01 = −3.0197, C11 = 0.5814, C02 = −0.2745).
The commercial code ABAQUS/Standard 6.8EF (Simulia Corp, Providence, RI, USA) was used to perform a large deformation analysis for the FE-based simulations. The classical displacement formulation solved by Newton's method was adopted. The stent deployment was simulated in two steps: crimping and release. A uniform inward radial displacement was applied to the nodes of the internal surface of the stent to fit it completely inside the vessel (crimping step). The stent was constrained at nodes in the endings of each strut to avoid rotation and at nodes in the middle of the three links of the central section to avoid translation. Then, by removing the constraints in the radial direction, the stent was allowed to self-expand until it contacted the vessel wall (release step). A stent–vessel wall contact was managed using a soft exponential frictionless interaction [37]. To prevent rigid motions, the vessel was fixed at its endings during the whole deployment. The three-dimensional- and one-dimensional-strut stent models were discretized in 8 × 104 8-node brick elements and 4 × 102 2-node linear beam elements, respectively. The beam elements of the one-dimensional-strut stents were modelled with circular cross section of 0.1 mm of diameter (figure 1c). The vessel wall was discretized in 104 3-node shell elements.
The fifth stent case used the FVS method of deployment (figure 3). The struts and their connectivity were defined over a subset of the points of a 2-simplex mesh with a size of 32 × 20. No material properties were defined for the stent, and the vessel was described as a rigid body. The simplex mesh was arranged as a cylinder of 2.6 mm of diameter and aligned along the curved vessel centreline. The expansion of the stent is governed by a system of second-order differential equations (2.1) discretized with finite differences:
![]() |
2.1 |
where t is an artificial time describing time evolution and pi refers to the ith point of the simplex mesh, fint represents the internal force accounting for mesh smoothness and the mesh expansion force, flength represents the stent-shape constraining force, and fext accounts for the interaction between the stent and the vessel wall. The constants α and χ are weighting parameters for the internal–external forces and constraints forces, respectively, set as α = 0.5 and χ = 0.9 according to the work of Larrabide et al. [11]. The interaction between the stent and the wall is defined by the interruption of the expanding force when the mesh reaches a distance of half the strut radial thickness from the vessel wall.
Figure 3.
FVS methodology. First the stent geometry is created in the initial undeformed configuration over a subset of points constituting the simplex mesh. The simplex mesh is deformed to obtain a curved tube with its axis along the vessel centreline.
2.3. Haemodynamic model
Before CFD simulations, three geometrical processes were applied. First, an ostium plane that distinguishes the aneurysm dome from the vessel lumen was defined (figure 4a). This plane was the same for all the CFD cases and did not intersect stent struts. Second, an automatic tubular three-dimensional surface of 0.1 mm of diameter was automatically created around the lines of the deformed one-dimensional-strut stents. Third, the stent was cut and the struts lying entirely on the vessel wall were removed (figure 4b). In literature it has been suggested that using the patch of the stent over the ostium instead of the whole stent reduces the computational cost of the CFD simulations without inducing important intra-aneurysmal haemodynamic variations [20].
Figure 4.
(a) Haemodynamic model of the idealized aneurysmatic cerebral vessel. Definition of dome, ostium plane and vessel. (b) Operation of cutting the struts of the stent lying entirely on the vessel wall. Arrows indicate the direction of the flow. (c) Volumetric mesh and details of the stent mesh. (d) Superimposition of three-dimensional reconstruction of the one-dimensional-strut stent on the three-dimensional-strut stent.
The CFD analysis was performed using ANSYS CFX v. 12.0 (ANSYS Inc., Canonsburg, PA, USA) to solve the unsteady three-dimensional Navier–Stokes equations, consisting of the continuity equation (2.2) and the momentum equation (2.3):
| 2.2 |
and
![]() |
2.3 |
where υ is the velocity vector, ρ the fluid density, t the time, μ the fluid dynamic viscosity and p the pressure. The blood was assumed to be a Newtonian incompressible fluid, with μ =0.004 Pa × s and ρ = 1060 kg m−3.
To simulate an angiography, the injection of the dye was performed by solving a transport (convection-diffusion) equation (2.4):
| 2.4 |
where C is the scalar field of dye concentration and k is the diffusion coefficient. The dye was modelled as a massless scalar passively transported by the fluid using a negligible diffusion coefficient (k = 10−15 m2 s−1). No flow rate change owing to the injection of the dye was considered. The temporal terms in the equations were discretized by using an Euler implicit scheme, and the spatial terms were discretized with second-order accuracy.
The assumption of a rigid wall with no-slip boundary condition was imposed on the vessel wall and stent. A constant pressure was applied to the outlet. At the inlet, a physiological curve of mean cross-sectional velocity was imposed [15] (figure 5). This velocity was characterized by a parabolic profile at the inlet and a time-averaged Reynolds number of 250. Four cardiac cycles were computed in order to simulate the angiography. For the quantitative analysis, only the results of the last cardiac cycle were taken into account.
Figure 5.
Four cardiac cycles of a physiological mean cross-sectional velocity at the inlet (solid line). Dye injection at 0.75 s for the whole second cardiac cycle (dashed line).
A time independence analysis has been performed. Four time steps (0.01 s, 0.005 s, 0.001 s and 0.0005 s) were tested. A difference of less than 1 per cent of (2.5) and (2.9) between the three smallest time steps has been observed, and the value of 0.005 s of time step has been used.
A root mean square (r.m.s.) of the residuals equal to 10−5 has been imposed as a criterion of convergence of the numerical solution.
The CFD cases were discretized with unstructured grids composed of tetrahedral elements, generated by ANSYS ICEM CFD v. 12.0 (ANSYS Inc). Three prism layers were applied at the aneurysm wall. A mesh independency analysis was performed, testing six different mesh sizes. To choose the mesh size, the criteria of reaching the accuracy of more than 99.5 per cent on (2.5) and (2.9) was used. From this, a global element size of 0.3 mm was selected, except at the vessel and aneurysm wall (0.15 mm) and around the stent struts (0.02 mm) [38] (figure 4c). Finally, UV and stented cases resulted in around 4 × 106 and 7.9 × 106 mesh elements, respectively.
To study the flow in the aneurysm, for each case, the average WSS [
, (2.5)], the average velocity ν [
(2.7)] and the mass inflow rate [
(2.9)] over the inflow region were considered. The inflow region was defined as the area on the ostium plane where the flow enters the aneurysm [22]. The WSS% (2.6), v% (2.8) and
% (2.10) are the percentages, averaged over time, of (2.5), (2.7) and (2.9) in the stented cases (S) with respect to the corresponding in UV. These parameters are widely used in literature to characterize the intra-aneurysmal haemodynamic conditions regarding aneurysm genesis, growth and rupture [19,39].
![]() |
2.5 |
| 2.6 |
![]() |
2.7 |
| 2.8 |
| 2.9 |
| 2.10 |
(W, aneurysm wall; D, aneurysm dome; IR, inflow region; cc, cardiac cycle).
3. Results
3.1. Geometrical findings
Before CFD analysis, a qualitative geometrical comparison across the five stented cases (inter-case) and between the two stent positions (inter-position) was performed. First, the hyperelastic properties of the vessel produced a change of the aneurysm ostium to a more circular shape, and a longitudinal straightening of the vessel in the stented part. Second, FVS produced a different configuration of the stent in both positions compared with the other stented cases, with a greater ring opening and enlarged links (the link above the ostium is four times longer than the original length). Third, the stent position produced a different stent configuration above the ostium. For example, the asymmetry of the link in posA (figure 6) allowed the struts to protrude towards the aneurysm, as it is represented for FE3H, FE3R and FE1H cases, but not for FE1R and FVS cases. On the other hand, the symmetric link in posB prevented the struts from protruding and none of the FE cases showed significant differences (see [33]).
Figure 6.
Configuration of the released stent in posA into the vessel for the five cases. Top and front view of the vessel.
A quantitative comparison of the stent screen effect was done after the three-dimensional reconstruction of the one-dimensional-strut stents (§2.3). To calculate the screen effect, the pixels of the image of the projection of the struts on the ostium plane were counted using the open source software Image J v. 1.43. The screen effect was computed as the percentage of the ratio of the area covered by the stent over the total area. The cases with three-dimensional-strut stent had the largest screen effect in both posA (20%) and posB (around 17%). The screen effect of one-dimensional-strut FE stents was approximately 6 per cent lower than the three-dimensional-strut one. FVS cases showed the lowest values (12% for posA and 10% for posB), because the three-dimensional reconstruction neglects morphological details (such as the rectangular cross-sectional area of the strut) and the released stent configuration is different. A difference of approximately 6 per cent between FE3H and FE1H and between FE3R and FE1R was observed. PosB produced a mean screen effect of 2.8 per cent lower than posA.
3.2. Analysis of computational fluid dynamics results
Figure 7 shows the reduction of intra-aneurysmal haemodynamic parameters owing to the presence of the stents. In posA, the mean values of WSS%, v% and
% of the five stented cases were 43 per cent, 40 per cent and 51 per cent, respectively. The inter-case variability of these parameters ranged between 10 and 15 per cent. In posB, while WSS% and v% were comparable with those of posA,
% was almost 66 per cent. The inter-case variability of these parameters ranged from 12 (WSS%) up to 25 per cent (
%).
Figure 7.

WSS%, v% and
% of each stented case for posA and posB; 100% refers to UV case. Filled squares, FE3H; open squares, FE3R; filled triangles, FE1H; open triangles, FE1R; circles, FVS.
All the cases highlighted an increase of
% for posB with respect to posA, while not all the cases predicted a unique relationship between the two positions concerning WSS% and v%. In fact, FE3H, FE3R and FE1H indicated a decrease of WSS% and v% in posB, while an opposite trend is shown by FE1R and FVS.
The inter-position variability varied both across the cases considering each parameter (e.g. v% shows a difference between posA and posB from 0.9% in FE3H case to 18.7% in FVS case) and across the parameters considering each case (e.g. FE3H case shows a difference between posA and posB of 2.5% for WSS%, 0.9% for v% and 19.8% for
%). The inter-case and inter-postion variabilities results are comparable (around 15%).
The effect of using a patch of the stent over the ostium on intra-aneurysmal haemodynamic parameters was investigated in the most complex case, FE3H, performing a simulation that retained the whole stent. Results showed that using the patch, the intra-aneurysmal haemodynamic parameters are underestimated by around 1 per cent. Furthermore, a percentile analysis at peak systole (2.4 s) showed that the presence of the stent (either the entire stent—additional case FE3Hc—or the patch—FE3H, FE3R, FE1H, FE1R and FVS) decreases the WSS on the aneurysm wall with respect to UV (figure 8).
Figure 8.
Percentile analysis of the WSS on the aneurysm wall at peak systole (2.4 s) of the stented cases with the stent patch, with the entire stent (FE3Hc), and UV in posA. Thick line, UV; crosses, FE3Hc; filled squares, FE3H; open squares, FE3R; filled triangles, FE1H; open triangles, FE1R; circles, FVS.
3.3. Impact zone and dye concentration
Figure 9 represents the area of vessel and aneurysm wall where WSS is lower than 2 Pa and the area where is higher than 2 Pa at peak systole (2.4 s). The value of the threshold used for distinguishing the two areas was selected to compare the cases for its physiological meaning. In fact, Hoi et al. [22] defined this threshold to identify the area of impact of blood (impact zone) on the aneurysm wall. With respect to UV, all stented cases showed a reduction of the impact zone. Depending on stent position, the shape of the impact zone could be different. PosB produced a wider and flatter high WSS area when compared with posA. For both positions, cases FE1R and FVS showed more evident differences with respect to the other stented cases.
Figure 9.
Impact zone at the aneurysm wall: in white WSS < 2 Pa, in black WSS > 2 Pa. Distal side view of the vessel.
The concentration of the dye in the vessel was measured as the integral of the transported scalar in the volume. Figure 10 shows the dye concentration distribution (a colour map of the concentration of the dye) of a longitudinal and a transversal middle plane of the aneurysmatic vessel at three time steps. As can be seen, aneurysm filling (1.125 s) started from the distal part of the ostium and produced a counter-clockwise vortex. Then, during the wash-out phase (1.7 s), the remaining dye inside the dome was diluted and more layers of the vortex appeared towards the aneurysm tip. After the end of the injection (3 s) in two cardiac cycles, the dye was completely washed out. However, an average of 0.1 per cent of dye concentration still remained in the aneurysm dome. The presence of the stent screened the flow entering the aneurysm, decreasing the concentration of dye in the aneurysm. This phenomenon was well represented by all the cases. At 1.7 s, FE1R and FVS of posA had some differences from the other cases in terms of amount and distribution of the dye. In general, posB showed a higher amount of dye inside the aneurysm compared with posA at the same time (1.7 s). In posB, the dye seemed to have the same distribution in the FE cases with the same vessel properties. Minor differences were visible for FVS.
Figure 10.
Dye concentration in the middle cut planes of the frontal and distal side of the vessel.
4. Discussion
This study was performed to evaluate the influence of different stent deployment approaches on intra-aneurysmal haemodynamics. Our results indicate that simulating the deployment of a stent in aneurysmatic vessels, besides the computational approach adopted, enables the representation of important effects on intra-aneurysmal haemodynamics in the successful CFD simulations. It was shown that, independently from the stent deployment approach, the insertion of the Neuroform stent model in an idealized aneurysmatic cerebral vessel produced a reduction of average WSS and average velocity inside the aneurysm of almost 50 per cent.
The five computational approaches investigated in this study were similar to those described in Bernardini et al. [33], although in the present study, a more comprehensive description of the aneurysmatic vessel was adopted. The obtained released stent and vessel configurations confirmed the previous findings [33]. In particular, there are observable differences in released stent and vessel configuration when neglecting the deformability of the vessel. Even larger differences appear increasing the level of simplifications (FE1R and FVS).
Regarding intra-aneurysmal haemodynamics, two factors influenced the results: the released geometries of the stented cases and the morphology of the stent. These two factors can be evaluated separately considering FE3H, FE3R and FE1H cases. The differences between FE3H and FE3R cases are influenced by the vessel model. It has been previously observed that the curvature and neck size of the vessel have an influence on intra-aneurysmal haemodynamics [19,22]. Comparing FE3H and FE1H cases, the morphology of the stent is the reason for their haemodynamic differences. To better understand the role played by the simplified reconstruction of the one-dimensional-strut stents on intra-aneurysmal haemodynamics, an additional simulation was performed with a simplified three-dimensional-strut stent automatically generated on the bases of the centreline of the deployed FE3H stent. Namely, this new deployed stent had the same configuration of FE3H but the same thickness of FE1H (figure 4d), and resulted in a lower screen effect (almost 6% less than FE3H). Compared with FE3H, the automatic stent reconstruction induced an overestimation of the average WSS, velocity and mass inflow rate of 1.35 per cent, 4.13 per cent and 9.3 per cent, respectively. FE1R and FVS are the two cases that combine the simplifications adopted for the stent deployment and the CFD simulations. Results showed that these two cases have more evident differences, both quantitatively and qualitatively, compared with the others.
The effectiveness of the treatment is usually evaluated by clinicians through qualitative analysis of the concentration of the dye in the aneurysm. Simulating an angiography can give clinicians a more familiar instrument to compare pre- and post-treatment. For this reason, it is important that all the computational approaches should highlight the effect of the stenting procedure. The present study showed that the concentration of the dye inside the aneurysm visibly decreased in all the stented cases.
The Neuroform is an open-cell stent with no symmetry in the disposition of the links, so the screen effect to blood flow and the possibility of protrusion of struts towards the aneurysm depends on the orientation and position of the stent in the vessel [40]. Hirabayashi et al. [41] found that the orientation and position of high-porosity stents influence the amount of flow reduction in the aneurysm. In our study, all the cases highlighted inter-position differences in intra-aneurysmal haemodynamics, even though those which combine simplifications (FE1R and FVS) did not show the same trend between posA and posB in all the parameters analysed, as the others did. Furthermore, they resulted in having greater inter-position variability.
In posB, where the protrusion of struts is prevented, there is a concordance between the haemodynamic parameters and the observed screen effect of the stent. Comparing between cases with the same vessel wall characteristics (FE3H with FE1H, FE3R with FE1R and FVS), the lower the screen effect was, the greater values for the different parameters were observed. On the other hand, for posA, the protrusion of the struts made the correlation of the haemodynamics with the screen effect more complex because it was only presented in some of the cases (FE3H, FE3R and FE1H).
Since the orientation and position of the stent are not completely controlled by the clinician, posA and posB are both probable during the intervention. In this regards, the inter-case and inter-position variabilities were comparable in our results, suggesting a sufficient accuracy for all the approaches adopted for stent deployment. Nevertheless, additional investigations with other stents for cerebral aneurysm treatment are mandatory to verify if, despite the observed inter-case variabilities, all the computational approaches allow the performances of various treatments to be distinguished (e.g. different stents).
Frictional forces of the stent on the wall have not been taken into account in the present study and simplified vessel geometry with a regular and smoothed wall was used. The authors think that the main findings of the present study, focused on the intra-aneurismal haemodynamics, are not significantly affected by these simplifications. However, future studies will consider a more realistic representation of the vessel and wall–stent interaction. A validation in real cases using medical images would allow a better judgement on the differences between models and reality. Virtual stent deployment together with CFD simulation can provide useful information for cerebral aneurysm interventional planning. It is possible to reduce the computational costs for the stent deployment step by simplifying the approach, keeping acceptable accuracy [33], but still the CFD step is the more time-consuming. Unsteady simulations to capture the pulsatility of the blood flow and obtain the virtual angiography were performed. Their computational time was 105 s, using 15 processors distributed over 8-nodes on a cluster with two Quad-Core Intel Xeon X5355 processors (2.66 GHz) per node sharing 16 GB RAM. This time is compatible with the time available for decision-making in elective interventions, because the treatment is usually scheduled some weeks in advance. To be able to apply these tools just before or during the intervention, the development of more efficient CFD tools or acceptable simplifications in simulations would be mandatory. Currently, some studies are going in this direction, for example, comparing steady state with transient simulations [42,43]. Steady-state simulations might facilitate the inception of haemodynamic simulations into clinical practice by providing the critical information for interventional planning at a lower computational cost.
In summary, five computational approaches for simulating stent deployment have been developed and compared. Below, an outline of the most outstanding features observed with regard to their application in clinical practice is provided:
— FE3H: is the most complex case, considering details of the stent shape and material. It also considers mechanical deformation of the vessel on its interaction with the stent, giving a more accurate description of the procedure, but its set-up and simulation time make it impractical for clinical practice today.
— FE3R: considers the detailed shape and material of the stent but neglects the mechanical behaviour of the wall, which could be an acceptable simplification in case of aneurysm treatment using Neuroform stents [16]. Its set-up and simulation time are expensive, making it impractical for clinical practice today.
— FE1H: simplifies the geometry of the stent but considers its material properties as well as the mechanical behaviour of the vessel wall. It is efficient regarding computational time, but its set-up is time-consuming.
— FE1R: stent geometry and vessel wall properties are simplified but the stent material is accurately represented. It is time efficient, but its set-up is time-consuming.
— FVS: stent geometry and material properties are simplified. Also, vessel material properties are neglected. On the other hand, it is time efficient and simple to set-up.
Furthermore, the approaches based on FE methods are not always applicable to patient-specific geometries, while the FVS algorithm enables easy stent deployment in anatomically accurate vasculature.
5. Conclusion
The present numerical simulation study shows how the choice of a computational approach for deploying the stent influences the CFD analysis, particularly in terms of intra-aneurysmal flow. Variability in intra-aneurysmal haemodynamics existed across the cases, varying according to the parameter analysed and the initial position of the stent considered. The reasons for this variability were discussed and analysed. Average WSS and average velocity were the parameters less affected by the computational approach for deployment and different position of the stent. Nevertheless, in all the analyses carried out, an important reduction of intra-aneurysmal flow was clearly induced in all modelled stent–vessel scenarios.
Acknowledgements
This work was partially supported within the CENIT programme, as part of CDTEAM and cvREMOD projects funded by the Spanish CDTI and partly within the framework of the @neurIST Project (IST-2005-027703), which is co-financed by the European Commission within the IST Programme of Sixth Framework Programme.
Footnotes
One contribution of 17 to a Theme Issue ‘The virtual physiological human’.
References
- 1.van Gijn J., Rinkel G. J. E. 2001. Subarachnoid hemorrhage: diagnosis, causes and management. Brain 124, 249–278 10.1093/brain/124.2.249 (doi:10.1093/brain/124.2.249) [DOI] [PubMed] [Google Scholar]
- 2.Juvela S. 2004. Treatment options of unruptured intracranial aneurysms. Stroke 35, 372–374 10.1161/01.STR.0000115299.02909.68 (doi:10.1161/01.STR.0000115299.02909.68) [DOI] [PubMed] [Google Scholar]
- 3.Wardlaw J. M., White P. M. 2000. The detection and management of unruptured intracranial aneurysms. Brain 123, 205–221 10.1093/brain/123.2.205 (doi:10.1093/brain/123.2.205) [DOI] [PubMed] [Google Scholar]
- 4.Dawson D. L., Meyer J., Lee E. S., Pevec W. C. 2007. Training with simulation improves residents' endovascular procedure skills. J. Vasc. Surg. 45, 149–154 10.1016/j.jvs.2006.09.003 (doi:10.1016/j.jvs.2006.09.003) [DOI] [PubMed] [Google Scholar]
- 5.Sarker S. K., Patel B. 2007. Simulation and surgical training. Int. J. Clin. Pract. 61, 2120–2125 10.1111/j.1742-1241.2007.01435.x (doi:10.1111/j.1742-1241.2007.01435.x) [DOI] [PubMed] [Google Scholar]
- 6.Tsang J. S., Naughton P. A., Leong S., Hill A. D. K., Kelly C. J., Leahy A. L. 2008. Virtual reality simulation in endovascular surgical training. Surgeon 6, 214–220 10.1016/51479-666X(08)80031-5 (doi:10.1016/51479-666X(08)80031-5) [DOI] [PubMed] [Google Scholar]
- 7.Villa-Uriol M. C., et al. 2010. AngioLab: integrated technology for patient-specific management of intracranial aneurysms. In 32nd Annual Conf. IEEE Engineering in Medicine and Biology Society (EMBS), 31 August–4 September, Buenos Aires, Argentina [DOI] [PubMed] [Google Scholar]
- 8.Villa-Uriol M. C., et al. 2010. Toward integrated management of cerebral aneurysms. Phil. Trans. R. Soc. A 368, 2961–2982 10.1098/rsta.2010.0095 (doi:10.1098/rsta.2010.0095) [DOI] [PubMed] [Google Scholar]
- 9.Appanaboyina S., Mut F., Löhner R., Putman C. M., Cebral J. R. 2008. Computational fluid dynamics of stented intracranial aneurysms using adaptive embedded unstructured grids. Int. J. Numer. Method Fluids 57, 475–493 10.1002/fld.1590 (doi:10.1002/fld.1590) [DOI] [Google Scholar]
- 10.Dequidt J., Marchal M., Duriez C., Kerien E., Cotin S. 2008. Interactive simulation of embolization coils: modeling and experimental validation. Med Image Comput. Assist Interv. 11, 695–702 [DOI] [PubMed] [Google Scholar]
- 11.Larrabide I., Kim M., Augsburger L., Villa-Uriol M. C., Rüfenacht D., Frangi A. F. 2010. Fast virtual deployment of self-expandable stents: method and in-vitro validation for intracranial aneurysmal stenting. Med Image Anal. 10.1016/j.media.2010.04.009 (doi:10.1016/j.media.2010.04.009) [DOI] [PubMed] [Google Scholar]
- 12.Morales H. G., Kim M., Villa-Uriol M. C., Diaz E. V., Frangi A. F. 2010. Influence of coil packing rate and configuration on intracranial aneurysm hemodynamics. IFMBE Proc. 25, 2291–2294 10.1007/978-3-642-03882-2_609 (doi:10.1007/978-3-642-03882-2_609) [DOI] [Google Scholar]
- 13.Yoon W. K., Kim Y. W., Kim S. R., Park I. S., Kim S. D., Jo K. W., Baik M. W. 2010. Angiographic and clinical outcomes of stent-alone treatment for spontaneous vertebrobasilar dissecting aneurysm. Acta Neurochirurgica 152, 1477–1486 10.1007/s00701-010-0693-7 (doi:10.1007/s00701-010-0693-7) [DOI] [PubMed] [Google Scholar]
- 14.Cebral J. R., Lohner R. 2005. Efficient simulation of blood flow past complex endovascular devices using an adaptative embedding technique. IEEE Trans. Med. Imaging 24, 468–476 10.1109/TMI.2005.844172 (doi:10.1109/TMI.2005.844172) [DOI] [PubMed] [Google Scholar]
- 15.Ford M. D., Sthune G. R., Nikolov H. N., Habets D. F., Lownie S. P., Holdsworth D. W., Steinman D. A. 2005. Virtual angiography for visualization and validation of computational models of aneurysm hemodynamics. IEEE Trans. Med. Imaging 24, 1586–1592 10.1109/TMI.2005.859204 (doi:10.1109/TMI.2005.859204) [DOI] [PubMed] [Google Scholar]
- 16.Alfke K., Straube T., Dörner L., Mehdorn H. M., Jansen O. 2004. Treatment of intracranial broad-neck aneurysms with a new self-expanding stent and coil embolization. AJNR Am. J. Neuroradiol. 25, 584–591 [PMC free article] [PubMed] [Google Scholar]
- 17.Bhatti M. T., Peters K. R., Firment C., Mericle R. A. 2004. Delayed exacerbation of third nerve palsy due to aneurysmal regrowth after endovascular coil embolization. J. NeuroOphthalmol. 24, 3–10 10.1097/00041327-200403000-00002 (doi:10.1097/00041327-200403000-00002) [DOI] [PubMed] [Google Scholar]
- 18.Lanzino G., Kanaan Y., Perrini P., Dayoub H., Fraser K. 2005. Emerging concepts in the treatment of intracranial aneurysms: stents, coated coils, and liquid embolic agents. Neurosurgery 57, 449–459 10.1227/01.NEU.0000170538.74899.7F (doi:10.1227/01.NEU.0000170538.74899.7F) [DOI] [PubMed] [Google Scholar]
- 19.Meng H., Wang Z., Kim M., Ecker R. D., Hopkins L. N. 2006. Saccular aneurysms on straight and curved vessels are subject to different hemodynamics: implications of intravascular stenting. AJNR Am. J. Neuroradiol. 27, 1861–1865 [PMC free article] [PubMed] [Google Scholar]
- 20.Appanaboyina S., Mut F., Löhner R., Putman C. M., Cebral J. R. 2009. Simulation of intracranial aneurysm stenting: techniques and challenges. Comput. Methods Appl. Mech. Eng. 198, 3567–3582 10.1016/j.cma.2009.01.017 (doi:10.1016/j.cma.2009.01.017) [DOI] [Google Scholar]
- 21.Fu W., Gu Z., Meng X., Chu B., Qiao A. 2010. Numerical simulation of hemodynamics in stented internal carotid aneurysm based on patient-specific model. J. Biomech. 43, 1337–1342 10.1016/j.jbiomech.2010.01.009 (doi:10.1016/j.jbiomech.2010.01.009) [DOI] [PubMed] [Google Scholar]
- 22.Hoi Y., Meng H., Woodward S. H., Bendok B. R., Hanel R. A., Guterman L. R., Hopkins L. N. 2004. Effects of arterial geometry on aneurysm growth: three-dimensional computational fluid dynamics study. J. Neurosurg. 101, 676–681 10.3171/jns.2004.101.4.0676 (doi:10.3171/jns.2004.101.4.0676) [DOI] [PubMed] [Google Scholar]
- 23.Kim M., Levy E. I., Meng H., Hopkins L. N. 2007. Quantification of hemodynamic changes induced by virtual placement of multiple stents across a wide-necked basilar trunk aneurysm. Neurosurgery 61, 1305–1313 10.1227/01.NEU.0000280168.25968.49 (doi:10.1227/01.NEU.0000280168.25968.49) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Kim M., Taulbee D. B., Tremmel M., Meng H. 2008. Comparison of two stents in modifying cerebral aneurysm hemodynamics. Ann. Biomed. Eng. 36, 726–741 10.1007/s10439-008-9449-4 (doi:10.1007/s10439-008-9449-4) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Liou T. M., Li Y. C., Wang T. C. 2008. Hemodynamics altered by placing helix stents in an aneurysm at a 45° angle to the curved vessel. Phys. Med. Biol. 53, 3763–3776 10.1088/0031-9155/53/14/004 (doi:10.1088/0031-9155/53/14/004) [DOI] [PubMed] [Google Scholar]
- 26.De Beule M., Mortier P., Carlier S. G., Verhegghe B., van Impe R., Verdonck P. 2008. Realistic finite element-based stent design: the impact of balloon folding. J. Biomech. 41, 383–389 10.1016/j.jbiomech.2007.08.014 (doi:10.1016/j.jbiomech.2007.08.014) [DOI] [PubMed] [Google Scholar]
- 27.Early M., Lally C., Prendergast P. J., Kelly D. J. 2009. Stresses in peripheral arteries following stent placement: a finite element analysis. Comput. Methods Biomech. Biomed. Eng. 12, 25–33 10.1080/10255840802136135 (doi:10.1080/10255840802136135) [DOI] [PubMed] [Google Scholar]
- 28.Gijsen F. J. H., et al. 2008. Simulation of stent deployment in a realistic human coronary artery. Biomed. Eng. OnLine 6, 7–23 10.1186/1475-925X-7-23 (doi:10.1186/1475-925X-7-23) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Migliavacca F., Petrini L., Massarotti P., Schievano S., Auricchio F., Dubini G. 2004. Stainless and shape memory alloy coronary stents: a computational study on the interaction with the vascular wall. Biomech. Model Mechanobiol. 2, 205–217 10.1007/s10237-004-0039-6 (doi:10.1007/s10237-004-0039-6) [DOI] [PubMed] [Google Scholar]
- 30.Mortier P., Holzapfel G. A., De Beule M., van Loo D., Taeymans Y., Segers P., Verdonck P., Verhegghe B. 2010. A novel simulation strategy for stent insertion and deployment in curved coronary bifurcations: comparison of three drug eluting stents. Ann. Biomed. Eng. 38, 88–99 10.1007/s10439-009-9836-5 (doi:10.1007/s10439-009-9836-5) [DOI] [PubMed] [Google Scholar]
- 31.Wu W., Qi M., Liu X. P., Yang D. Z., Wang W. Q. 2007. Delivery and release of nitinol stent in carotid artery and their interactions: a finite element analysis. J. Biomech. 40, 3034–3040 10.1016/j.jbiomech.2007.02.024 (doi:10.1016/j.jbiomech.2007.02.024) [DOI] [PubMed] [Google Scholar]
- 32.Montagnat J., Delingette H. 1998. Globally constrained deformable models for 3D object reconstruction. Signal Process 71, 173–186 10.1016/S0165-1684(98)00143-1 (10.1016/S0165-1684(98)00143-1) [DOI] [Google Scholar]
- 33.Bernardini A., Larrabide I., Petrini L., Pennati G., Flore E., Kim M., Frangi A. F. 2010. Deployment of self-expandable stents in aneurysmatic cerebral vessels: comparison of different computational approaches for interventional planning. Comput. Methods Biomech. Biomed. Eng. 10.1080/10255842.2010.527838 (doi:10.1080/10255842.2010.527838) [DOI] [PubMed] [Google Scholar]
- 34.Ujiie H., Tamano Y., Sasaki K., Hori T. 2001. Is the aspect ratio a reliable index for predicting the rupture of a saccular aneurysm? Neurosurgery 48, 495–503 10.1097/00006123-200103000-00007 (doi:10.1097/00006123-200103000-00007) [DOI] [PubMed] [Google Scholar]
- 35.Auricchio F., Petrini L. 2004. A three-dimensional model describing stress-temperature induced solid phase transformations: solution algorithm and boundary value problems. Int. J. Numer. Method Eng. 61, 807–836 10.1002/nml.1086 (doi:10.1002/nml.1086) [DOI] [Google Scholar]
- 36.Monson K. L., Barbaro N. M., Manley G. T. 2006. Multiaxial response of human cerebral arteries. In Annual Meeting of the American Society of Biomechanics Blacksburg, VA: Virginia Polytechnic Institute and State University [Google Scholar]
- 37.ABAQUS Theory Manual 2008. Version 6.8EF, Hibbit, Karlsson and Sorensen, Inc. [Google Scholar]
- 38.Stuhne G., Steinman D. A. 2004. Finite-Element modeling of the hemodynamics of stented aneurysms. J. Biomech. Eng. 126, 382–388 10.1115/1.1762900 (doi:10.1115/1.1762900) [DOI] [PubMed] [Google Scholar]
- 39.Jou L. D., Lee D. H., Morsi H., Mawad M. E. 2008. Wall shear stress on ruptured and unruptured intracranial aneurysms at the internal carotid artery. AJNR Am. J. Neuroradiol. 29, 1761–1767 10.3174/ajnr.A1180 (doi:10.3174/ajnr.A1180) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Hsu S. W., Chaloupka J. C., Feekes J. A., Cassell M. D., Cheng Y. F. 2006. In vitro studies of the Neuroform microstent using transparent human intracranial arteries. AJNR Am. J. Neuroradiol. 27, 1135–1139 [PMC free article] [PubMed] [Google Scholar]
- 41.Hirabayashi M., Ohta M., Rüfenacht D. A., Chopard B. 2004. A lattice Boltzmann study of blood flow in stented aneurysm. Future Gener. Comput. Syst. 20, 925–934 10.1016/j.future.2003.12.004 (doi:10.1016/j.future.2003.12.004) [DOI] [Google Scholar]
- 42.Geers A. J., Larrabide I., Morales H. G., Frangi A. F. 2010. Comparison of steady-state and transient blood flow simulations of intracranial aneurysms. In 32nd Annual Conf. IEEE Engineering in Medicine and Biology Society (EMBS), 31 August–4 September, Buenos Aires, Argentina [DOI] [PubMed] [Google Scholar]
- 43.Mantha A. R., Benndorf G., Hernandez A., Metcalfe R. W. 2009. Stability of pulsatile blood flow at the ostium of cerebral aneurysms. J. Biomech. 42, 1081–1087 10.1016/j.jbiomech.2009.02.029 (doi:10.1016/j.jbiomech.2009.02.029) [DOI] [PubMed] [Google Scholar]