Assessment of cardiac function during mechanical circulatory support: the quest for a suitable clinical index

Antonio L Ferreira 1, Yajuan Wang 2, John Gorcsan III 3, James F Antaki 4
PMCID: PMC3265331  NIHMSID: NIHMS347404  PMID: 22254290

Abstract

A new index to assess left ventricular (LV) function in patients implanted with continuous flow left-ventricular assist devices (LVADs) is proposed. Derived from the pump flow signal, this index is defined as the coefficient (k) of the semilogarithmic relationship between “pseudo-ejection” fraction (pEF) and the volume discharged by the pump in diastole, (V d). pEF is defined as the ratio of the “pseudo-stroke volume” (pSV) to V d. The pseudo-stroke volume is the difference between V d and the volume discharged by the pump in systole (V s), both obtained by integrating pump flow with respect to time in a cardiac cycle. k was compared in-vivo with others two indices: the LV pressure-based index, MTP, and the pump flow-based index, IQ. MTP is the slope of the linear regression between the “triple-product” and end-diastolic pressure, EDP. The triple-product, TP = LV SP.dP=dtmax.HR, is the product of LV systolic pressure, maximum time-derivative of LV pressure, and heart rate. IQ is the slope of the linear regression between maximum time-derivative of pump flow, dQ/dtmax, and pump flow peak-to-peak amplitude variation, QP2P. To test the response of k to contractile state changes, contractility was altered through pharmacological interventions. The absolute value of k decreased from 1.354±0.25 (baseline) to 0.685±0.21 after esmolol infusion. The proposed index is sensitive to changes in inotropic state, and has the potential to be used clinically to assess contractile function of patients implanted with VAD.

I. INTRODUCTION

Left ventricular assist devices (LVADs) have emerged as the standard of care for advanced heart failure (HF) patients, requiring mechanical circulatory support [1]. The third generation, non-pulsatile flow devices are more reliable and safer, presenting a low incidence of device related adverse events, with enhanced patient’s quality of life when compared to their pulsatile counterparts.

The remodeling process resultant from HF was generally believed to be irreversible. However, there has been compelling evidence of myocardial recovery in a small - yet significant - group of patients, especially those with idiopathic cardiomyopathy [2]. Both basic and clinical research findings [3] suggested that left ventricular unloading through VAD support lead to cellular and functional recovery, with consequent LVAD explantation.

Identification of potential candidates to recovery entails assessment of LV function of the assisted heart. Proper evaluation of LV contractility has to be load-independent, and sensitive to inotropic changes. The slope of the linear regression between the LV “triple product” (TP) and the LV end-diastolic pressure (EDP), MTP, was evaluated by McConnell and collaborators [4] in a series of in-vivo experiments. TP is the product of LV systolic pressure, heart rate and the maximal time-derivative of LV pressure, i.e., TP = LVSP.(dP=dt)max.HR. Because this index requires invasive measurements of LV pressure, its application on the clinical setting would be limited.

Recently, Schima and colleagues proposed the IQ [5], an index that can be derived from pump flow. The IQ was defined as the slope of the linear regression between the maximal time-derivative of pump flow (dQ/dtmax) and its peak-to-peak amplitude variation (QP2P). Tests in-silico and in-vivo demonstrated that the IQ is sensitive to contractility changes, with potential to clinical applications.

This paper presents a novel method to assess cardiac contractility from pump flow, based on a curvilinear relationship between pEF and pump diastolic volume, V d. For a pulsatile pump, one can define pump stroke volume per beat, either by integrating outlet and inlet flows and subtracting the latter from the former, or by a fixed amount of stroke volume at each pump beat. For continuous-flow devices, we propose the pSV. The pump flow signal is integrated over time and a reference signal (e.g., ECG) would make possible to calculate the amount of blood generated (or discharged) by the pump in systole and in diastole [6].

Pump flow pulsaility has been used in the clinical realm as a surrogate to cardiac contractility. Surely, pulsatility may increase as a result of ventricular recovery. However, pulsatility is preload dependent, and even intermittent episodes of inlet cannula tip moving against the ventricular wall can be confounded with increased pulsatility. This report addresses the need for an index that can be used in the clinical setting, especially at low speeds. This paper is organized as follows: Section 2 describes the k index derivation. Section 3 presents the experimental results of an in-vivo study performed in a calf. Concluding remarks are presented in section 4.

II. INDEX DEFINITION

Because of the residual left-ventricular function of the assisted heart, continuous-flow devices generate flow that may present pulsatility. Indeed, presence (or absence) of pulsatility on pump signals (flow, current) has been used as an anecdotal “marker” of myocardial contractile state in the clinical realm. As opposed to pulsatile devices, rotary pumps are in synchrony with the natural heart, as shown in Figure 1.

Fig. 1.

Fig. 1

Pump hemodynamics at 1800 rpm. Ts is systolic time; Td is diastolic time; Vs is the volume of pump flow in Ts, and Vd is the volume of pump flow in Td. AoP is aortic pressure, LVP is left ventricular pressure and Q(t) is pump flow.

The pEF is defined as

pEF=pSVVd=VdVsVd (1)

where V d and V s are defined as

Vd=TdQ(t)dt (2)
Vs=TsQ(t)dt (3)

and Q(t) is pump flow, as shown in Figure 1.

Besides contractility of the native heart, pump speed also influences pulsatility: as pump speed increases, pump flow pulsatility diminishes. Figure 2 illustrates this phenomena in-vivo. The upper panel of Fig. 2 shows that pump speed was changed in a stair step fashion. Three increasingly speed set points are of interest: 1800 rpm, 2100 rpm, and 2500 rpm, respectively.

Fig. 2.

Fig. 2

Pump flow analysis for a stair steep speed experimental study, at A) low support (1800 rpm); B) moderate support (2100 rpm), and C) higher support at 2500 rpm. In each case, panels from the top are pump flow, Vd and Vs, and pump flow derivative, dQ/dt

The second panel of Fig. 2 shows how the pump flow signal changes as pump speed is increased. For each one of the speed settings described, Vd and Vs were calculated and the derivative of pump flow with respect to time. When speed is low (1800 rpm, in this example), most of the pump flow volume in a beat would occur in systole (Vs ≫ Vd), for a recovered or healthier heart. At 2100 rpm, the pump flow volume in systole and diastole would be about the same, regardless of the cardiac phase (Vs ≈ Vd), and at higher speeds, 2500 rpm, Vd would surpass Vs (Vs < Vd). If the heart function is depleted (“weak” heart, with low contratile strength), even at lower speeds, Vs would not be higher than Vd, because of the “flattened” pump flow waveform.

Since pEF (see eq(1)) measures the amount of blood volume generated by the VAD in one cardiac cycle, pEF may carry information on the heart contractile function. Thus, the relationship between pEF and V d can be used to assess contractility. Figure 3 shows that, at a constant inotropic state, the (V d, pEF) coordinates move along a curvilinear relation that can be written as

Fig. 3.

Fig. 3

Relationship between pEF and V d from experimental data

pEF=klog(Vd)+a (4)

where a, k are constants. In summary, the constant k will change according to the contratile state of the heart, and a accounts for the fact that pEF may go negative.

III. EXPERIMENTAL RESULTS

The study protocol used here was approved by WorldHeart Inc. (Salt Lake city, UT). Those studies were performed in collaboration with LaunchPoint Technologies (Goleta, CA). A Levacor pump was implanted in a health calf. Left ventricular pressure were acquired with a pressure-tip catheter, and ultrasonic flow probes were used to measure pump flow and pulmonary artery flow. Pump signals (current, speed) were also recorded at a rate of 500Hz. Data Analysis wass performed using MATLAB 1. This study aimed to test a physiologic controller, evaluating its response to contractility changes and venous return depletion. Contractility changes were done by pharmacological interventions with esmolol, 0.1 mg/kg/min. An occluder was used to decrease venous return, inducing suction events.

Before and after each pharmacological intervention, a stair step speed profile was used (see Fig. 2). The three indices, TP, IQ and k were calculated for each inotropic condition. Figure 4A shows the results for TP. At baseline, MTP = 1.55 × 105 mmHg s−1 bpm, and after esmolol intervention MTP = 2.02 × 105 mmHg s−1 bpm. Figure 4B shows the results for IQ. This index was 9.78s−1 at baseline, and 8.63s−1 after esmolol infusion.

Fig. 4.

Fig. 4

Indices results on experimental data: A) triple product (MTP), B) IQ index

Figure 5A shows the results for the proposed index. At baseline, k = 1.354 ml−1, and after esmolol intervention it was k = 0.685 ml−1. As expected, k decreased at a lower contratile state. Figure 5B shows the results obtained from a pulsatile index, defined as

Fig. 5.

Fig. 5

A) Results of the proposed index, k; B) Pulsatility index calculated as per eq. 5.

P=QmaxQminQavg (5)

where, Qmax is the maximum of pump flow, Qmin is its minimum, and Qavg is the mean value of pump flow. In Fig. 5B, P vs. Qmin relationship also could be fitted by an exponential. However, all (Qmin, P) coordinates would lie in the same curve, irrespective of the inotropic state, considering the overlapping of the 95% confidence interval. This observation corroborates the fact that a pulsatile index defined as in eq. (5), would not convey reliable information on myocardial contractility. Table I summarizes the results for all three indices.

TABLE I.

Summary of results

Baseline Esmolol
MTP (mmHg s−1 bpm*105) 1.55 ± 1.17 2.02 ± 1.4
IQ (s −1) 9.78 ± 0.2 8.63 ± 1.12
k (ml −1) 1.354 ± 0.25 0.685 ± 0.21

Data expressed as mean ± standard deviation

IV. DISCUSSION

The assessment of residual LV function on assisted hearts is still a challenge [8], [9]. Fewer indices have been proposed in recent years [4], [5] to assess LV function in assisted hearts. Traditional indices, such as ESPVR may not be directly applied to the assisted heart. As demonstrated by Vandenberghe and colleagues [10], the original time-varying elastance theory insufficiently models the complex hemodynamic behavior of a mechanically assisted LV. Moreover, LV unloading renders Ees curvilinear, and therefore is less adequately defined by a linear relationship [7]. Thus, continuous LV unloading limits applicability of indices based on ventricular volume measurements, because continuous unloading influences the definition of systolic ventricular volume and ejection fraction [5].

The slope of TP vs EDP (MTP) should have decreased when esmolol was administrated to the animal, with respect to its baseline value. Several reasons may explain this unexpected result. First and foremost, during the axial unloading reported in [4], EDP ranged from 8–20mmHg (baseline) and 13–26mmHg (esmolol). In our case, the centrifugal unloading employed gave a much narrow range, 5–15mmHg for EDP. This may explain the low R2 coefficients in Fig. 4 (0.40 at baseline and 0.49 with esmolol). It remains unclear how axial vs centrifugal flow pumps affect the circulatory system. In addition, in our in-vivo experiment, the pump was interposed between the LV apex and ascending aorta. In the experiments reported in [4], LV unloading was to the descendent thoracic aorta, and differences may exist when LVAD outflow is to the ascending aorta.

The IQ index decreased when esmolol was administrated to the animal, as expected. Even though the relationship between dQ/dtmax and QP2P was highly linear in both cases (R2 = 0.98, baseline, and R2 = 0.96 esmolol), the 95% confidence intervals overlapped when comparing baseline and esmolol groups, 9.78 ± 0.2 and 8.63 ± 1.12 respectively. (see Table I).

The k index proposed, showed a non-overlapping 95% confidence interval between baseline (k = 1.354 ± 0.25 ml−1), and after esmolol infusion (k = 0.685 ± 0.21 ml−1), as shown in Fig. 5A. Thus, k would correctly stratify “recovering” hearts (baseline), from a “weak” heart (esmolol).

Current assessment of cardiac function in VAD implanted patients is done via pump speed reduction to minimal support. Usually, pump speed is maintained only high enough to avoid regurgitant flow, and echocardiographic analysis is employed to assess LV area [11]. The proposed index could be calculated through this procedure, by Doppler wave techniques.

Some limitations of our study are noteworthy. First, more in-vivo studies are needed to verify that the k index is a viable option in assessing cardiac contractility. Finally, comparisons with others traditional indices (e.g., preload recruitable stroke work or maximal time-derivative of LV pressure vs end diastolic volume, (dP/dt)max vs EDV) will be done soon, as part of our current ongoing research.

Acknowledgments

This work has been supported by NIH grant 1R01HL086918-01A1

Footnotes

1

The MathWorks Inc., Natik, MA

Contributor Information

Antonio L. Ferreira, Email: aferreir@andrew.cmu.edu, Department of Mathematics, Federal University of Maranhao, Sao Luis MA 65080-040, Brazil and a researcher at the Department of Biomedical Engineering, Carnegie Mellon University, Pennsylvania, PA 15219, USA

Yajuan Wang, Email: yajuanw@andrew.cmu.edu, PhD candidate at the Department of Electrical Engineering, Carnegie Mellon University, Pennsylvania, PA 15219, USA.

John Gorcsan, III, Email: gorcsanj@upmc.edu, Cardiovascular Institute, University of Pittsburgh Medical Center, Pittsburgh, PA 15213, USA.

James F. Antaki, Email: antaki@andrew.cmu.edu, Professor of Biomedical Engineering at Carnegie Mellon University, Pennsylvania, PA 15213, USA

References

  • 1.Slaughter MS, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361(23):2241–51. doi: 10.1056/NEJMoa0909938. [DOI] [PubMed] [Google Scholar]
  • 2.Maybaum S, Kamalakannan G, Murthy S. Cardiac recovery during mechanical assist device support. Thoracic and Cardiovascular Surgery. 2008;20(3):234–46. doi: 10.1053/j.semtcvs.2008.08.003. [DOI] [PubMed] [Google Scholar]
  • 3.Klotz S, Danser AH, Burkhoff D. Impact of left ventricular assist device (LVAD) support on the cardiac reverse remodeling process. Progress in Biophysics & Molecular Biology. 2008;97:479–496. doi: 10.1016/j.pbiomolbio.2008.02.002. [DOI] [PubMed] [Google Scholar]
  • 4.McConnell Patrick I, Del Rio Carlos L, Kwiatkowski Pawel, Farrar David. Assessment of Cardiac Function During Axial-flow Left Ventricular Assist Device Support Using a Left Ventricular Pres-sure-derived Relationship: Comparison With Pre-load Recruit able Stroke Work. J Heart Lung Transplant. 2007 Feb;26(2):159–166. doi: 10.1016/j.healun.2006.11.010. [DOI] [PubMed] [Google Scholar]
  • 5.Naiyanetr Phornphop, Moscato Francesco, Vollkron Michael, Zimpfer Daniel, Wieselthaler Georg, Schima Heinrich. Continuous assessment of cardiac function during rotary blood pump support: A contractility index derived from pump flow. J Heart Lung Transplant. 2010 Jan;29(1):37–44. doi: 10.1016/j.healun.2009.05.032. [DOI] [PubMed] [Google Scholar]
  • 6.Akimoto T, Yamazaki K, Litwak P, et al. Relationship of blood pressure and pump flow in an implantable centrifugal blood pump during hypertension. ASAIO J. 2000;46:596–599. doi: 10.1097/00002480-200009000-00018. [DOI] [PubMed] [Google Scholar]
  • 7.McConnell P. Pressure-volume analysis during axial flow ventricular assist device support. J Heart Lung Transplant. 2006;25(2):256–7. doi: 10.1016/j.healun.2005.05.010. [DOI] [PubMed] [Google Scholar]
  • 8.Sakamoto T. Evaluation of native left ventricular function during mechanical circulatory support: theoretical basis and clinical limitations. Ann Thorac Cardiovasc Surg. 2002;8(1):1–6. [PubMed] [Google Scholar]
  • 9.Ferrari M, Kadipasaoglu K, Croitoru M, et al. Evaluation of myocardial function in patients with end-stage heart failure during support with the Jarvik 2000 left ventricular assist device. J Heart Lung Transplant. 2005;24(2):226–8. doi: 10.1016/j.healun.2003.09.044. [DOI] [PubMed] [Google Scholar]
  • 10.Vandenberghe S, Segers P, Steendijk P, et al. Modeling Ventricular Function during Cardiac Assist: Does Time-Varying Elastance Work? ASAIO J. 2006;52:4–8. doi: 10.1097/01.mat.0000196525.56523.b8. [DOI] [PubMed] [Google Scholar]
  • 11.George R, Sabharwal N, Webb C, et al. Echocardiographic assessment of flow across continuous-flow ventricular assist devices at low speeds. J Heart Lung Transplant. 2010;29(11):1245–52. doi: 10.1016/j.healun.2010.05.032. [DOI] [PubMed] [Google Scholar]